510(k) Data Aggregation

    K Number
    K173568

    Validate with FDA (Live)

    Date Cleared
    2018-09-04

    (288 days)

    Product Code
    Regulation Number
    862.1055
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeoBase™ 2 Non-derivatized MSMS kit is intended for the measurement and evaluation of amino acid, succinylacetone, free carnitine, acylcarnitine, nucleoside and lysophospholipid concentrations (Table 1) with a tandem mass spectrometer from newborn heel prick blood specimens dried on filter paper. Quantitative analytis of these analytes and their relationship with each other is intended to provide analyte concentration profiles that may aid in screening newborns for metabolic disorders.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes the acceptance criteria and study results for the NeoBase 2 Non-derivatized MSMS kit.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for the screening performance studies (e.g., minimum sensitivity or specificity targets). Instead, it states that "All verification studies were successfully concluded and met the respective study's predetermined acceptance criteria." The clinical studies for screening performance are presented as agreement between the new device (NeoBase 2) and the predicate device (NeoBase). The agreement is presented as contingency tables (e.g., "Screening positive" vs "Screening negative" for both devices).

    The performance is demonstrated by the agreement between the NeoBase 2 Non-derivatized MSMS kit and the predicate device, NeoBase Non-derivatized MSMS kit, in detecting various metabolic disorders in newborn screening. The results are presented in terms of the number of positive and negative screens detected by each device, along with the number of confirmed positive specimens.

    Summary of Device Performance (from Tables A, B, C, D):

    Disorder GroupCut-off Type (Percentile)NeoBase 2 Screening Positive (with Predicate Positive)NeoBase 2 Screening Negative (with Predicate Negative)Total SpecimensConfirmed Positive Specimens (detected by both methods)
    Study 1
    Amino acid disorders99th6211591175115
    Amino acid disorders99.5th4521645175115
    Amino acid disorders1st161168717371 (OTCD)
    Fatty acid oxidation99th8011581174610
    Fatty acid oxidation99.5th4511661174610
    Fatty acid oxidationLow Percentile1732138617382 (CUD)
    Organic acid condition99th5711660175115
    Organic acid condition99.5th3611697175115
    ADA-SCID99th2166117382
    ADA-SCID99.5th2170017382
    X-ALD99th2172417382
    X-ALD99.5th2173117382
    Study 2
    Amino acid disorders99th11612353264819
    Amino acid disorders99.5th7822474264818
    Amino acid disorders1st141257126312 (OTCD)
    Fatty acid oxidation99th16012326264112
    Fatty acid oxidation99.5th10812442264112
    Fatty acid oxidationLow Percentile1581236326323
    Organic acid condition99th8612479264213
    Organic acid condition99.5th4222561264212
    ADA-SCID99th2256326312
    ADA-SCID99.5th2257826312
    X-ALD99th2262626312
    X-ALD99.5th2262826312

    2. Sample Size Used for the Test Set and Data Provenance

    • Study 1 Sample Size:

      • Amino acid disorders, Fatty acid oxidation, Organic acid conditions: 1751 samples (for 99th and 99.5th percentile cut-offs) and 1737-1746 samples (for 1st and low percentile cut-offs).
      • ADA-SCID and X-ALD: 1738 samples.
    • Study 2 Sample Size:

      • Amino acid disorders, Fatty acid oxidation, Organic acid conditions: 2631-2648 samples.
      • ADA-SCID and X-ALD: 2631 samples.
    • Data Provenance: The data was obtained from "routine newborn screening" in "two CLIA-certified state laboratories." The confirmed positive specimens were described as "retrospective" for Study 2. This suggests a retrospective study design using existing samples and accompanying diagnostic information. The country of origin is not explicitly stated but is implied to be the US due to "CLIA-certified state laboratories."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. It mentions "confirmed positive specimens," implying a definitive diagnostic process was followed to establish the true disease status of these samples, but details on the experts involved are not provided.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set, such as 2+1 or 3+1. The acceptance is based on the agreement between the new device and the predicate device, using established cut-offs derived from routine newborn screening data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This device is a diagnostic kit measuring analyte concentrations, not an AI system assisting human readers. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The study described is a comparison of the new device (NeoBase 2) to a predicate device (NeoBase) in obtaining analyte concentrations. While not explicitly stated as an "algorithm only" study, it's a standalone performance comparison of two test kits. The results (analyte concentrations and screening positive/negative classifications) are derived directly from the kit's operation with a tandem mass spectrometer, without human interpretation being part of the primary measurement process itself. The interpretation of the analyte profiles to aid in screening for metabolic disorders would typically involve medical professionals, but the performance data presented is on the analytical and classification output of the device.

    7. Type of Ground Truth Used

    The ground truth for the test set was based on "confirmed positive specimens." This implies that the true disease status of these specimens was established through clinical diagnosis and follow-up, which would typically involve a combination of clinical outcomes, biochemical testing, and/or genetic testing, ultimately confirmed by clinical experts. For ADA-SCID and X-ALD, it explicitly states "comparing the result... to the clinical condition."

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. The term "cut-offs for both methods were determined by calculating the 99.5th and 99th percentile for all analytes" using "data from routine newborn screening." This large volume of routine newborn screening data could be considered analogous to a training or reference population used to establish the operating characteristics of the screening test. The specific sample size for this cut-off determination is not given, but it is implied to be a large dataset from the "two CLIA-certified state laboratories."

    9. How the Ground Truth for the Training Set Was Established

    As discussed in point 8, there isn't a traditional "training set" for an AI model. However, the cut-off values (e.g., 99th, 99.5th, 1st, 10th percentiles) used to define "screening positive" or "screening negative" were established using "data from routine newborn screening." This means the ground truth for establishing these cut-offs would inherently come from the statistical distribution of analyte levels in a large, presumably healthy and general newborn population, along with the understanding of what analyte levels are indicative of various metabolic disorders. The document states that the cut-off values "only apply to these studies."

    Ask a Question

    Ask a specific question about this device

    K Number
    K102419

    Validate with FDA (Live)

    Date Cleared
    2010-12-16

    (113 days)

    Product Code
    Regulation Number
    862.1725
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GSP Neonatal IRT kit is intended for the quantitative determination of IRT in blood specimens dried on filter paper as an aid in screening newborns for cystic fibrosis using the GSP® instrument.

    Device Description

    The GSP Neonatal IRT assay is a solid phase, two-site fluoroimmunometric assay based on the direct sandwich technique in which two monoclonal antibodies (derived from mice) are directed against two separate antigenic determinants on the IRT molecule. Calibrators, controls or test specimens containing IRT are reacted simultaneously with immobilized monoclonal antibodies directed against a specific antigenic site on the IRT molecule and europium-labeled monoclonal antibodies (directed against a different antigenic site) in assay buffer. The assay buffer elutes IRT from dried blood on filter paper disks. The complete assay requires only one incubation step. DELFIA Inducer dissociates europium ions from the labeled antibody into solution where they form highly fluorescent chelates with components of the DELFIA Inducer. The fluorescence in each well is then measured. The fluorescence of each sample is proportional to the concentration of IRT in the sample.

    AI/ML Overview

    The provided text describes the GSP Neonatal IRT kit, a device for screening newborns for cystic fibrosis. It compares the new device to a predicate device (AutoDELFIA Neonatal IRT kit) and presents information on its performance characteristics.

    Here’s a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Predicate Device K0003668)Reported Device Performance (GSP Neonatal IRT kit)
    Intended Use / Indications for UseQuantitative determination of human IRT in blood specimens dried on filter paper as an aid in screening newborns for cystic fibrosis using the 1235 AutoDELFIA automatic immunoassay system.Quantitative determination of IRT in blood specimens dried on filter paper as an aid in screening newborns for cystic fibrosis using the GSP® instrument.
    Instrument1235 AutoDELFIA InstrumentGSP Instrument
    Dissociation solutionEnhancement SolutionDELFIA Inducer
    Antibody Cross-Reactions
    α2-macroglobulin< 4 ng/ml blood0.000%
    α1-antitrypsin< 4 ng/ml blood0.000%
    Phospholipase A2< 4 ng/ml blood0.014%
    Chymotrypsin< 4 ng/ml blood0.959%
    Human IgG< 4 ng/ml blood0.000%
    Pepsinogen< 4 ng/ml blood (Uro)Pepsinogen-0.056%
    Complement Factor INA0.000%
    Measuring Range4 (as defined by LoB) to 500 (as defined by upper calibrator) ng/mL blood9 to 500 ng/mL blood
    TracerAnti-IRT-Eu tracer stock solution, approximate concentration of ~50 µg/ml mouse monoclonalAnti-IRT-Eu tracer stock solution, approximate concentration of ~40 µg/ml mouse monoclonal
    Analytical Sensitivity / Limit of Blank< 4 ng/mL bloodLimit of Blank: 0.76 ng/mL blood
    Limit of DetectionNot explicitly stated for predicate; implied by LoB2.2 ng/mL blood
    Limit of QuantitationNot explicitly stated for predicate; implied by LoB2.2 ng/mL blood
    Precision (Total Variation)42.6 ng/mL blood CV% 9.3; 98.8 ng/mL blood CV% 10.0; 266 ng/mL blood CV% 9.610.9 ng/mL blood CV% 7.3; 22.2 ng/mL blood CV% 7.2; 28.5 ng/mL blood CV% 7.0; 40.0 ng/mL blood CV% 8.2; 50.2 ng/mL blood CV% 8.0; 61.6 ng/mL blood CV% 7.8; 93.5 ng/mL blood CV% 7.2; 302.3 ng/mL blood CV% 7.4; 449 ng/mL blood CV% 7.5

    Study Proving Device Meets Acceptance Criteria:

    The document describes a substantial equivalence claim, not a separate clinical study with independent acceptance criteria beyond demonstrating equivalence to the predicate device. The "study" here is the comparison and characterization of the new device's performance against the established predicate device (AutoDELFIA Neonatal IRT kit, K0003668). The acceptance criteria are largely implicitly defined by the performance characteristics of the predicate device, where the new device is shown to have similar or improved performance in analytical parameters.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state a sample size for a test set in the context of a clinical study. The performance characteristics reported (e.g., precision, analytical sensitivity, measuring range) are typically derived from analytical verification and validation studies in a laboratory setting.

    • Sample Size: Not explicitly stated for a "test set" in a clinical context. The precision data lists various IRT concentrations tested, and each CV% (Coefficient of Variation) would have been derived from replicate measurements.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies generating these performance characteristics were retrospective or prospective. It describes the device's analytical performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    N/A. This is an in-vitro diagnostic device for measuring a biomarker (IRT). The "ground truth" for its performance is based on analytical measurements and comparison to a known standard, not on expert clinical evaluation or interpretation of images.

    4. Adjudication Method for the Test Set

    N/A. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of data (e.g., images) by multiple human readers, not in the analytical validation of an in-vitro diagnostic assay measuring a biomarker.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    N/A. An MRMC study is not applicable to the analytical performance evaluation of this type of in-vitro diagnostic device. This device measures a quantitative biomarker and does not involve human interpretation of cases or "readers" in the traditional sense, nor does it aim to improve human reader performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The device itself is a "standalone" assay in the sense that it quantifies IRT levels directly from a blood spot sample using an instrument. There is no human interpretation of an algorithm output, but rather a direct measurement that aids in screening. The GSP® instrument performs the assay automatically.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance (especially for parameters like measuring range, sensitivity, and precision) would be established by:

    • Reference materials/calibrators: Samples with known concentrations of IRT (calibrated using gravimetric methods, as mentioned).
    • Established analytical methods: Comparison to recognized analytical techniques for IRT measurement.
      This is not "expert consensus, pathology, or outcomes data" in the clinical sense, but rather a robust analytical validation against known standards.

    8. The Sample Size for the Training Set

    N/A. This document pertains to an in-vitro diagnostic kit for measuring a biomarker, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The device's performance is based on its chemical and immunological assay design.

    9. How the Ground Truth for the Training Set was Established

    N/A. As stated above, this is not an AI/ML device that uses a "training set."