Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K160805

    Validate with FDA (Live)

    Device Name
    NIO-P
    Manufacturer
    Date Cleared
    2016-06-23

    (92 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NIO-P for Pediatrics is intended to provide intraosseous access in the proximal tibia, as an alternative to IV access during emergencies. The device is for use in pediatric patients between 3 and 12 years of age.

    Device Description

    The NIO product family (Adult and Pediatric) is an upgrade in terms of design and structure to the company's B.I.G. Bone Injection Gun product family. The NIO product family has already one cleared model, for the adult population NIO Adult (K142086). Both product families NIO and BIG are composed of a trocar needle, spring, piston, and housing that contains safety mechanisms to prevent accidental device activation. Both devices are positioned onto the appropriate insertion site. Both product families have the same principle of operation. The device resembles a syringe and, when activated, a loaded spring is released and the device injects a needle to a predetermined depth into the bone marrow cavity. The user then pulls out the trocar needle, leaving the cannula inside the bone marrow cavity. Through the cannula, fluids and drugs can be delivered to the vascular system via the bone marrow cavity. The NIO-P device is designed to detect the proper injection point. In addition, the NIO-P is also equipped with a unique, adjustable needle penetration depth according to the age group, as part of the stabilizer. Both features are designed to improve accuracy in detecting the insertion point and needle penetration depth.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the NIO-P device, a hypodermic single lumen needle intended for pediatric intraosseous access. The submission aims to demonstrate substantial equivalence to existing predicate devices.

    Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria with numerical targets. Instead, it lists several validations and studies performed to demonstrate the device's performance characteristics. The "Reported Device Performance" is implied through the successful completion of these validations and bench testing, confirming the device functions as intended and meets safety and effectiveness requirements.

    Acceptance Criteria (Implied)Reported Device Performance
    Proper Insertion Point DetectionValidated by Insertion Point and Penetration Depth Validation
    Accurate Needle Penetration Depth (adjustable for age group)Validated by Insertion Point and Penetration Depth Validation
    Adequate Spring Force for Bone PenetrationValidated by Adequate Spring Force and Bone Integrity Following Activation Validation
    Preservation of Bone Integrity after ActivationValidated by Adequate Spring Force and Bone Integrity Following Activation Validation
    Needle Stabilizer EffectivenessValidated by Needle Stabilizer Validation
    Hub-Cannula/Hub-Needle Bond StrengthTested in Hub-Cannula/Hub-Needle Bond Strength Testing (According to ISO 7864)
    Structural Integrity during useTested in Structural Deformation and Needle Integrity Test
    Substantial Equivalence to predicate/reference devicesDemonstrated through performance testing and comparison to NIO-A and BIG-P, confirming that modifications do not raise new safety/effectiveness questions.

    2. Sample Size Used for the Test Set and Data Provenance

    The document primarily describes bench testing and validation studies. It does not specify sample sizes for these tests. The data provenance is implied to be from the manufacturer's internal testing (Waismed Ltd.). There is no mention of country of origin for test data, nor whether it is retrospective or prospective, as it pertains to engineering and product validation rather than clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The studies listed are bench tests evaluating physical properties and functionalities of the device (e.g., spring force, bond strength, structural deformation). These types of tests typically rely on engineering specifications and laboratory measurements rather than expert human assessment for ground truth.

    4. Adjudication Method for the Test Set

    Given that the studies are bench tests of physical characteristics, an adjudication method in the context of expert consensus or disagreement is not applicable and therefore not mentioned. The "ground truth" for these tests would be defined by the pre-established engineering specifications and validated measurement techniques.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence through bench testing of the device's physical and functional characteristics.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a mechanical hypodermic needle for intraosseous access. It is not an algorithm-based device or software, so the concept of "standalone (algorithm only)" performance is not applicable. The device's performance is inherently tied to its mechanical function, albeit with features designed to aid the human user (e.g., detecting proper injection point, adjustable penetration depth).

    7. The Type of Ground Truth Used

    The ground truth used for the performance studies is based on:

    • Engineering specifications and design requirements: For aspects like proper insertion point, penetration depth, spring force, bone integrity, bond strength, and structural deformation.
    • Established standards: E.g., ISO 7864 for Hub-Cannula/Hub-Needle Bond Strength Testing.
    • Comparison to predicate device characteristics: To demonstrate that the modified device's characteristics are consistent with or improved upon the predicate and reference devices, without raising new questions of safety or effectiveness.

    8. The Sample Size for the Training Set

    This information is not applicable. The NIO-P is a mechanical medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The design and validation are based on engineering principles and physical testing.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as #8. There is no training set for a mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142086

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-28

    (28 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NIO Adult is intended to provide intraosseous access in the proximal tibia, as an alternative to IV access during emergencies. The device is for use in adult patients only.

    The NIO Adult is indicated for use in providing intraosseous access as an alternative to IV access during emergencies. Humeral head IO access is indicated when rapid fluid or pharmacological resuscitation is required and intravenous access is not possible. The device is for use in adult patients only.

    Device Description

    The NIO Adult device is a modification to the company's cleared Adult B.I.G - Bone Injection Gun (K981853 and K062940) to permit intraosseous access through the Proximal Tibia and the Humeral Head.

    Like the cleared Adult B.I.G., the NIO Adult device is comprised of a trocar needle, spring, piston and housing. Likewise, both devices resemble a syringe and when activated, a loaded spring is released and the device injects the needle to a predetermined depth into the bone marrow cavity.

    The difference between the NIO Adult device and the cleared B.I.G. device is primarily the safety latch and trigger mechanism which is designed to provide a safer device and the minor design modifications to the plastic housing and needle dimensions in order to accommodate the modified safety mechanism.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called "NIO Adult," a hypodermic single lumen needle intended for intraosseous access. The document does not describe a study that uses acceptance criteria and reports device performance in the way a clinical trial or a machine learning algorithm validation would. Instead, it focuses on demonstrating substantial equivalence to a previously cleared predicate device (B.I.G. Adult Bone Injection Gun) through a series of engineering and performance tests.

    Therefore, many of the requested categories for acceptance criteria and study design are not applicable in the context of this 510(k) submission.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists several performance tests, but it does not explicitly state quantitative acceptance criteria or detailed numerical performance results for each test in a table format. It generally states that the device "functions as intended" and "inserts the needle to the same penetration depth as the predicate device."

    Test NameImplied Acceptance CriterionReported Device Performance
    Trigger Safety Mechanism TestDevice safety mechanism functions correctly.Functions correctly, improves device safety.
    Needle Stabilizer (Mechanical Stopper) Performance TestStabilizer performs as intended.Performs as intended.
    Needle Penetration Depth TestPenetration depth is consistent with the predicate device.Inserts the needle to the same penetration depth as the predicate device.
    Compressed Spring Force TestSpring force is within specified limits for proper needle insertion.Functions as intended.
    Bone Fracture TestDevice does not cause undue bone fracture during use.Functions as intended.
    Biocompatibility Testing (ISO 10993-5, -10, -11)Materials are biocompatible (non-cytotoxic, non-irritating, non-sensitizing).Meets ISO 10993 standards.
    Hub-Cannula/Hub-Needle Bond Strength Testing (ISO 7864)Bond strength meets ISO 7864 requirements to prevent separation.Meets ISO 7864 standards.
    Structural Deformation and Needle Integrity TestDevice structure and needle maintain integrity during use.Maintains integrity.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify general "sample sizes" in the way a clinical study would (e.g., number of patients or cases). The tests described are engineering and bench tests, likely performed on a small number of device units or components.

    • Sample Size: Not specified for each test. Typically, these types of tests involve a limited number of devices or components to demonstrate consistency and compliance.
    • Data Provenance: The tests were conducted to demonstrate substantial equivalence to the predicate device, also manufactured by Waismed Ltd. The document does not specify a country of origin for the testing data itself, but the applicant company (Waismed Ltd.) is located in Herzliya, Israel. The data is retrospective in the sense that it's comparing against an existing predicate, but the testing itself would have been conducted more recently.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study involving expert human interpretation or diagnosis. The "ground truth" for these engineering tests is established by objective measurements against predefined specifications or comparison to the predicate device's characteristics.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication process described, as the tests are objective engineering measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the subject of this medical device submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The NIO Adult is a physical medical device (a needle system), not an algorithm or software-based device.

    7. The type of ground truth used

    The "ground truth" for the performance tests is based on:

    • Engineering Specifications: Designed performance characteristics for components (e.g., spring force, depth of penetration).
    • International Standards: Compliance with standards like ISO 10993 (biocompatibility) and ISO 7864 (bond strength).
    • Predicate Device Performance: Direct comparison to the established performance characteristics of the B.I.G. Adult Bone Injection Gun in areas like needle penetration depth.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062940

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-12-22

    (85 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B.I.G. - Bone Injection Gun is indicated for use in providing intraosseous access as an alternative to IV access during emergencies. Humeral head IO access is indicated when rapid fluid or pharmacological resuscitation is required and intravenous access is not possible. The device is for use in adult patients only.

    Device Description

    The B.I.G. - Bone Injection Gun consists of a trocar needle held by a piston. The piston is surrounded by a compressed spring. All components are contained in the device's housing. When the device is operated, the compressed spring is released and propels the trocar needle into the bone marrow.

    AI/ML Overview

    The provided document, K062940, is a 510(k) summary for WaisMed's B.I.G.™ - Bone Injection Gun. This document primarily focuses on establishing substantial equivalence to predicate devices and does not contain detailed performance criteria or a study design description that would typically be found in a clinical trial report or a more comprehensive technical document.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

      The document states: "The device was tested for both performance and safety. In all instances, the B.I.G. - Bone Injection Gun functioned as intended."
      However, specific quantitative acceptance criteria (e.g., success rate, time to access, specific failure modes and their acceptable limits) and detailed reported device performance metrics are not provided in this summary. It only offers a general statement that the device functioned as intended.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      This information is not available in the provided 510(k) summary. The document does not describe any specific test set size for performance and safety testing, nor does it mention the provenance of any data used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      This information is not available. The summary does not describe a clinical study involving human experts establishing ground truth. The evaluation appears to be based on functional and safety testing of the device itself rather than an assessment by experts against a reference standard in a clinical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      This information is not available. No description of adjudication methods is present, as there's no mention of a human-reviewed test set requiring such a process.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      This information is not applicable to this device. The B.I.G. - Bone Injection Gun is an intraosseous infusion device, not an AI-powered diagnostic tool, and therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      This information is not applicable to this device. As a mechanical medical device, the concept of "algorithm only" performance without human-in-the-loop is not relevant. The device requires human operation for its intended use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      This information is not explicitly stated as part of a formal study with ground truth. For a mechanical device like the B.I.G., "ground truth" typically relates to its functional performance (e.g., successful insertion, fluid flow rate without leakage, structural integrity). The summary only states that the device "functioned as intended," implying that it met its functional design specifications, which would serve as the de-facto "ground truth" for its performance and safety tests.

    8. The sample size for the training set

      This information is not applicable to this device. The B.I.G. - Bone Injection Gun is a mechanical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

      This information is not applicable to this device, as it does not involve a training set.

    In summary, the provided 510(k) summary for the B.I.G. - Bone Injection Gun states that the device was tested for performance and safety and "functioned as intended." However, it lacks specific details about the acceptance criteria, study design, sample sizes, expert involvement, or any quantitative results that would typically be provided in a comprehensive study report. The premarket notification focuses on demonstrating substantial equivalence to predicate devices (WaisMed, Ltd., B.I.G. - Bone Injection Gun (K981853) and Vidacare Corporation, Humeral Head EZ-IO (K052408)) based on similar intended uses, indications, technological characteristics, and principles of operation, rather than a detailed presentation of a de novo clinical or performance study with quantified acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022415

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-10-11

    (79 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B.I.GTM. Bone injection Gun for Pediatrics is intended to provide intraosseous access in the proximal tibia, as an alternative to IV access during emergencies. The device is for use in pediatric patients up to 12 years of age.

    Device Description

    The B.I.GTM. Bone injection Gun for Pediatrics is an instant, intravascular (IV) access device. The device comprises a housing with a spring loaded mechanism that inject a trocar needle into the bone marrow of a patient and allows for subsequent connection of a syringe or intravascular administration set.

    AI/ML Overview

    This document is a 510(k) summary for the WaisMed LTD. B.I.G.TM - Bone Injection Gun for Pediatrics. It is a premarket notification for a medical device and, as such, does not typically contain the detailed information requested regarding clinical studies with specific acceptance criteria, sample sizes for training/test sets, ground truth establishment, or expert involvement in the way a clinical trial report or a journal publication would.

    The provided document focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed de novo clinical study with quantitative performance metrics against pre-defined acceptance criteria.

    I can, however, extract related information to address some of your points based on what is available:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the 510(k) summary. The document focuses on demonstrating substantial equivalence based on technological features, principles of operation, and intended use, rather than presenting specific quantitative performance data against pre-defined acceptance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The summary states "supportive data including in vitro studies, clinical data and literature survey," but does not specify sample sizes, country of origin, or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The 510(k) summary does not describe any specific process for establishing ground truth involving experts. It relies on the clinical context of intraosseous access for emergencies.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. As an adjudication method would typically be part of a detailed clinical study, its absence is consistent with the nature of a 510(k) summary focused on substantial equivalence.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/provided. The device is a mechanical bone injection gun, not an AI-powered diagnostic tool requiring human readability or interpretation. Therefore, an MRMC study related to AI assistance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/provided. The device is a mechanical device for gaining intraosseous access, not an algorithm. Standalone algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not explicitly stated in the context of the requested detailed study. However, for a device like an intraosseous access gun, the "ground truth" for its effectiveness would inherently relate to the successful establishment of intraosseous access and its ability to deliver fluids/medications. This would typically be confirmed by direct observation during use, fluid flow, and patient physiological response, which falls under outcomes data in a broad sense (i.e., whether the device successfully achieves its intended physical function). The 510(k) summary states it provides "intraosseous access in the proximal tibia, as an alternative to IV access during emergencies." The "clinical data" mentioned would likely pertain to demonstrating this function.

    8. The sample size for the training set

    This information is not provided. There is no mention of a "training set" in the context of machine learning. The device is mechanical.

    9. How the ground truth for the training set was established

    This information is not provided/not applicable as there is no training set for a machine learning model.


    Summary of available information from the 510(k) Summary:

    The 510(k) summary primarily asserts substantial equivalence to predicate devices based on:

    • Intended Use: To provide intraosseous access in the proximal tibia as an alternative to IV access during emergencies, for pediatric patients up to 12 years of age.
    • Technological Features and Principles of Operation: Stated to be the same as the predicate Adults' B.I.G.TM device.
    • Performance Data: "Supportive data including in vitro studies, clinical data and literature survey" were used to demonstrate that minor differences from predicate devices do not raise new types of safety or effectiveness issues. However, specific details of these studies (acceptance criteria, sample sizes, ground truth, etc.) are not provided in this summary.

    This type of 510(k) summary is common for devices demonstrating substantial equivalence, where the focus is on showing similarity to already cleared devices rather than presenting a novel device with extensive new clinical trial data against pre-defined performance thresholds.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981853

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-11-17

    (175 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bone Injection Gun (B.I.G.) is intended to provide intraosseous access in the proximal tibia, as an alternative to IV access during emergencies. The device is for use in adult patients only.

    Device Description

    Bone Injection Gun (B.I.G.)

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the Bone Injection Gun (B.I.G.) device, confirming its substantial equivalence to a legally marketed device. It does not contain information about the device's acceptance criteria, reported performance, or any studies conducted to prove it meets acceptance criteria.

    Specifically, the document does not provide any of the following information requested:

    1. A table of acceptance criteria and the reported device performance
    2. Sample size used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and their qualifications
    4. Adjudication method for the test set
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or its effect size
    6. If a standalone performance (algorithm only) was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    This letter is a regulatory approval document, not a clinical study report or a technical performance specification.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1