Found 2 results

510(k) Data Aggregation

    K Number
    K170784

    Validate with FDA (Live)

    Device Name
    PNEUMOCLEAR
    Date Cleared
    2017-05-15

    (61 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device PNEUMOCLEAR™ is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, High Flow/Bariatric, Pediatric and Advanced Flow operating mode of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is indicated for pediatric laparoscopic procedures. The Vest operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The TAMIS operating mode is indicated to fill and distend the rectum and colon using CO2 gas during trans anal minimal invasive surgery.

    Device Description

    The PNEUMOCLEAR™ is a microprocessor controlled CO2 insufflator that consists of the following major components and features: (1) a casing, (2) a world power supply, (3) pressure reducers, (4) pressure sensors, (4) venting systems, (5) various safety valves. (6) a suction pump. (7) a fluid sensor and (8) a software controlled graphical user interface (GUI) touch screen with various setting keys and display elements. The proposed device offers six operating modes (i.e. Standard, High Flow/Bariatric, Pediatric, Advanced Flow, Vessel Harvest, and TAMIS. The PNEUMOCLEAR™is not intended to enter the sterile field, and cannot be sterilized. It is to be used with specially designed single-use insufflation tube sets that are delivered sterile. Specifically, five different tube sets may be used with the proposed device: (1) insufflation tube set with integrated filter, (2) insufflation tube set with integrated filter and heating wire; (3) insufflation tube set with integrated filter, heating wire and humidification media; (4) insufflation and smoke evacuation tube set with integrated filter, heating wire, and humidification media; and insufflation and smoke evacuation tube set with integrated filter. When used with a smoke evacuation tube set the PNEUMOCLEAR™ allows for removal and filtration of CO2 and surgical smoke from the abdomen, rectum or colon during laparoscopic and transanal minimally invasive procedures and at the end of a procedure.

    AI/ML Overview

    The PNEUMOCLEAR™ device is a CO2 insufflator for endoscopic procedures. The information provided outlines several performance tests but does not present a formal table of acceptance criteria with reported device performance or extensive details on study design as typically found in clinical trials for AI/ML devices.

    However, based on the provided text, we can infer some "acceptance criteria" through the comparisons made against predicate devices and adherence to relevant standards. The "study" refers to a series of bench tests, electrical safety and electromagnetic compatibility testing, software verification and validation, biocompatibility testing, and sterilization and package testing.

    Here's an attempt to structure the information as requested, with caveats regarding the absence of explicit acceptance criteria and detailed study parameters for each test:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Electrical Safety & EMCConformance to IEC 60601-1 and IEC 60601-1-2 standards."Test results demonstrate that the proposed device conforms to the below standards: IEC 60601-1 2005/(R)2012 and A1:2012. C1:2009/(R)2012 andA2:2010/(R)2012 (consolidated text); and IEC 60601-1-2 Third Edition: 2007-03."
    SoftwareDevice performs as intended, developed, tested, and verified according to FDA guidance and IEC 62304."Software verification and validation testing of the PNEUMOCLEAR™ demonstrates that the device performs as intended. It was developed, tested and verified in accordance with the above FDA guidance document and in accordance with the following standard: IEC 62304 First Edition 2006-05."
    BiocompatibilityConformance to AAMI/ANSI/ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-11 standards for insufflation tube sets."Biocompatibility testing was performed on the insufflation tube sets of the proposed device PNEOMOCLEAR™ in accordance with: AAMI/ANSI/ISO 10993-1 Fourth Edition 2009-10-15; ISO 10993-5: 2009 (R) 2014; ISO 10993-10 Third Edition 2010-08-01; and ISO 10993-11 Second Edition 2006-08-15."
    Sterilization & PackagingETO sterilization validation per ISO 11135-1, ISO 14937, ISO 10993-7, AAMI TIR 28. Residual EO < 4mg, ECH < 5mg (4 days aeration). Sterility Assurance Level (SAL) < 10^-6. Package and product integrity per ISO 11607-1, ASTM-F-1980."ETO sterilization validation for the sterile single use tube sets was performed in accordance with the below standards: ISO 11135-1 Second Edition 2014; ISO 14937 Second Edition 2009-10-15; ISO 10993-7 Second Edition 2008-10-15; and AAMI TIR 28 2009/(R)2013. Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) data shows that the limit of EO < 4 mg and ECH < 5 mg after 4 days of aeration (gas release) that remains on the tube set will not be exceeded. A sterility assurance level (SAL) < 10^-6 is achieved. Package and product integrity of the tube sets were tested in accordance with the following standards: ISO 11607-1 First Edition 2006-04-15; and ASTM-F-1980: 2002."
    Bench Testing - Pressure Regulation (High Flow/Bariatric vs. Predicate Standard)Substantially equivalent pressure regulation to predicate device (Insufflator 50L FM134) in Standard Operating Mode.Comparative bench testing performed to demonstrate substantial equivalence.
    Bench Testing - Pressure Regulation (High Flow/Bariatric vs. Predicate Bariatric)Substantially equivalent pressure regulation to predicate device (Insufflator 50L FM134) in Bariatric Operating Mode.Comparative bench testing performed to demonstrate substantial equivalence.
    Bench Testing - Advanced Flow vs. Predicate BariatricSubstantially equivalent performance to predicate device (Insufflator 50L FM134) in Bariatric Operating Mode.Comparative bench testing performed to demonstrate substantial equivalence.
    Bench Testing - Steady Control Feature (Advanced Flow)Safety and effectiveness of the Steady Control Feature demonstrated, preventing abdominal pressure rise above set value.Bench testing performed demonstrating safety and effectiveness, including plausibility checks for flow and pressure.
    Bench Testing - Advanced Flow Leakage CompensationSubstantial equivalence in large leakage compensation and abdominal pressure maintenance with Steady Control Feature active, compared to predicate Bariatric mode.Comparative bench testing performed.
    Bench Testing - TAMIS Mode Pressure RegulationSubstantially equivalent pressure regulation to predicate device (SurgiQuest AirSeal iFS System) with and without smoke evacuation activation.Comparative bench testing performed.
    Bench Testing - Stryker SDC3 HD ControlEffectiveness in controlling PNEUMOCLEAR™ and remotely displaying device settings/warnings.Bench testing performed demonstrating effectiveness.
    Bench Testing - RFID Tube Set RecognitionEffectiveness of RFID transponder technology for tube set recognition.Bench testing performed demonstrating effectiveness.
    Bench Testing - Smoke Evacuation PerformanceComparative performance of smoke evacuation function to reference device (Olympus High Flow Insufflation Unit UHI-4).Comparative bench testing performed.
    Bench Testing - Desufflation FunctionSafety and effectiveness of the desufflation function.Bench testing performed demonstrating safety and effectiveness.
    Bench Testing - Insufflation Tube Set (ST296 vs. Predicate ST260)Comparative performance of tube set with integrated filter and heating wire (ST296) to predicate tube set (ST260) in terms of safety and effectiveness.Comparative bench testing performed.
    Bench Testing - Insufflation Tube Set (ST297 vs. Reference Insuflow)Comparative performance of tube set with integrated filter, heating wire, and humidification material (ST297) to reference device (Insuflow).Comparative bench testing performed.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • The document does not specify sample sizes for any of the bench tests, electrical safety, software, biocompatibility, or sterilization tests.
    • The data provenance is not explicitly stated as country of origin, retrospective or prospective. However, the tests are stated to have been performed by "independent laboratories" implying these are controlled laboratory tests rather than clinical studies. Given the context of a 510(k) submission, these would be prospective testing protocols executed to demonstrate compliance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not applicable in the context of this device and the provided document. The PNEUMOCLEAR™ is a hardware-based insufflator, not an AI/imaging device requiring expert interpretation for ground truth establishment. The "ground truth" for its performance is determined by objective physical measurements (e.g., pressure, flow, electrical parameters) and adherence to engineering standards.

    4. Adjudication Method for the Test Set

    • This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, especially for diagnostic devices where subjective interpretation is involved. The performance of this device is assessed through objective, quantifiable measurements against design specifications and predicate device performance.

    5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, a Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices involving human interpretation (e.g., radiologists interpreting images) to assess the impact of AI assistance on reader performance. The PNEUMOCLEAR™ is a therapeutic/procedural device, not a diagnostic one of this nature. Therefore, the effect size of human readers improving with AI vs. without AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    • Beyond the software verification and validation, which assesses the algorithm's functionality, a standalone study in the context of an AI-driven diagnostic device (e.g., measuring algorithm-only performance metrics like AUC, sensitivity, specificity on a dataset) was not performed or described. The device's "algorithms" primarily refer to its control logic (e.g., for pressure regulation, "Intelligent Steady Insufflation" in Advanced Flow mode), which are evaluated through the various bench tests demonstrating the device's functional performance in a simulated environment.

    7. The Type of Ground Truth Used

    • The "ground truth" for the PNEUMOCLEAR™ device is based on objective physical measurements (e.g., pressure, flow rates, electrical parameters, chemical residuals, sterility levels) and adherence to established international and national standards (IEC, ISO, AAMI, ASTM). For comparative bench testing, the performance of the predicate devices under identical conditions serves as a benchmark for "substantial equivalence."

    8. The Sample Size for the Training Set

    • This information is not applicable. The PNEUMOCLEAR™ is a deterministic device, not an AI/ML device that typically requires a 'training set' for machine learning models. Its control logic and performance characteristics are based on engineering design and calibration, not learned from data. The software described is for controlling the device, not for making predictions or classifications based on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as there is no "training set" in the context of this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153513

    Validate with FDA (Live)

    Date Cleared
    2016-03-04

    (88 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device Insufflator 50L FM134 is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, Pediatric and Bariatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.

    Device Description

    The Insufflator 50L FM134 is a microprocessor controlled CO2 insufflator that consists of the following major components and features: a casing, a world power supply, pressure reducers, a venting system, redundant pressure measurement, a fluid sensor, a gas heater, a software controlled graphical user interface (GUI) touch screen and various setting keys and display elements. The Insufflator 50L FM134 is not intended to enter the sterile field, and cannot be sterilized. The device is to be used with specially designed single-use tube sets that are delivered sterile. Specifically, the proposed device is to be used with a single-use tube set with heating wire and integrated filter or with a single-use tube set with integrated filter but without heating wire.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Insufflator 50L FM134. However, it does not detail specific acceptance criteria or an overarching study proving the device meets acceptance criteria in the way one might expect for a diagnostic or prognostic AI/ML device (e.g., sensitivity, specificity, AUC thresholds).

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through various tests and comparisons. The "acceptance criteria" can be inferred from the standards and comparisons performed.

    Here's a breakdown based on the provided text, addressing the points where information is available and noting where it is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a medical device (CO2 insufflator) and not an AI/ML diagnostic, the acceptance criteria are not in terms of traditional metrics like sensitivity/specificity but rather compliance with safety, performance, and biocompatibility standards, and comparability to predicate devices.

    Acceptance Criterion (Inferred from tests performed)Reported Device Performance (Summary from study)
    Electrical Safety (IEC 60601-1:2005)- Conforms to the standard.
    Electromagnetic Compatibility (IEC 60601-1-2:2007)- Conforms to the standard.
    Software Development & Verification- Developed, tested, and verified in accordance with "Software Contained in Medical Devices" guidance and IEC 62304:2006. - Design verification demonstrates the device performs as intended and does not raise new questions of safety and effectiveness.
    Comparative Bench Testing (vs. 45L Core Insufflator F114)- Reaching set pressure: Comparable to predicate. - Compensation of small and large leakages: Comparable to predicate. - Overpressure scenarios: Comparable to predicate. - Maximum flow rate (50 l/min vs. 45 l/min): Leads on average to a slightly better insufflation performance compared to the predicate device 45L Core Insufflator. Also substantially equivalent to Nebulae™ I 50 LPM Insufflator for this characteristic.
    Biocompatibility (ISO 10993 series)- Cytotoxicity: Conforms to ISO 10993-5:2009. - Irritation and Skin Sensitization: Conforms to ISO 10993-10:2010. - Systemic Toxicity: Conforms to ISO 10993-11:2006. - Chemical Characterization (for heated tube set): Conforms to ISO 10993 - 18:2005.
    ETO Sterilization Validation (Tube Sets)- EO & ECH Residuals: Limit of EO < 4 mg and ECH < 5 mg (after 10 days aeration) not exceeded. - Sterility Assurance Level (SAL): < 10^-6. - Conforms to: ISO 11135-1:2007, ISO 14937:2009, ISO 10993-7:2008, AAMI TIR 28:2009.
    Package & Product Integrity (Tube Sets)- Conforms to ISO 11607-1 and ASTM-F-1980:2002.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: The document does not specify a "sample size" in terms of patient data or case numbers for the performance tests. The tests performed are primarily laboratory-based bench tests (e.g., electrical safety, EMC, software verification, comparative performance of insufflation characteristics, biocompatibility of materials, sterilization validation).
    • Data Provenance: The tests were performed by "independent laboratories" (for electrical safety and EMC) and implicitly by the manufacturer or their designated testing facilities for others. The nature of the device (insufflator) means there wouldn't typically be "country of origin of the data" in the sense of clinical patient data, nor is it described as retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable and not provided. The "ground truth" for this device's evaluation is based on established engineering standards, physical performance measurements, and biological compatibility standards, rather than expert interpretation of medical images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical evaluations where there might be disagreement among experts. The tests performed for this insufflator are objective measurements against predefined standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was it done?: No. An MRMC study is not relevant for this type of medical device (an insufflator). This type of study is typically performed for AI/ML diagnostic tools where human readers (e.g., radiologists) interpret images with and without AI assistance.
    • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable.

    6. Standalone Performance

    • Was it done?: Yes, to the extent that an insufflator operates as a standalone device. The device's performance (e.g., flow rate, pressure regulation, leakage compensation) was tested independently against specifications and compared to predicate devices in bench testing. There is no "human-in-the-loop" performance as the device directly performs its function (insufflation).

    7. Type of Ground Truth Used

    The "ground truth" for the tests described is based on a combination of:

    • International Standards: e.g., IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993 series, ISO 11135-1, ISO 14937, ISO 11607-1, ASTM-F-1980.
    • Predicate Device Performance: The characteristics of the legally marketed predicate devices (45L Core Insufflator F114 and Nebulae™ I 50 LPM Insufflator) served as a benchmark for comparison to demonstrate substantial equivalence.

    8. Sample Size for the Training Set

    This information is not applicable and not provided. This device is an electro-mechanical insufflator, not an AI/ML system that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable.

    Ask a Question

    Ask a specific question about this device