510(k) Data Aggregation

    K Number
    K193197

    Validate with FDA (Live)

    Date Cleared
    2020-04-20

    (152 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuickClear Mechanical Thrombectomy System is intended for removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.

    Device Description

    The QuickClear Mechanical Thrombectomy System is an aspiration thrombectomy system designed for removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. The QuickClear Mechanical Thrombectomy System removes thrombus from the peripheral vasculature using continuous vacuum aspiration, through a single lumen aspiration catheter. The system components are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The QuickClear Mechanical Thrombectomy System comprises the following devices and accessories:

    • QuickClear Aspiration Pump
    • A 60cc Syringe
    • QuickClear Aspiration Catheters: Catheters are available in the following sizes: 6F straight tip, 8F straight tip, 8F shaped tip and 10F shaped tip.
    • An Obturator is supplied with the 8F and 10F catheters.
    • A Waste Collection Bag
    AI/ML Overview

    This document is a 510(k) summary for the QuickClear Mechanical Thrombectomy System, seeking substantial equivalence to a predicate device. It primarily details the device's design, intended use, and a comparison to existing devices. It does not describe the acceptance criteria or a study proving the device meets performance criteria for a diagnostic tool (e.g., an AI-powered one).

    Instead, this device is a mechanical thrombectomy system, meaning it physically removes thrombus (blood clots). The testing described is for the physical performance and safety of a medical device, not the diagnostic performance of an AI algorithm.

    Therefore, I cannot extract the requested information (acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert qualifications for ground truth, MRMC study, etc.) from this document. This document pertains to the regulatory clearance of a mechanical medical device, not a diagnostic AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182972

    Validate with FDA (Live)

    Date Cleared
    2018-12-20

    (55 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle. The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle. The Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the Phoenix® Atherectomy System. It describes the device, its intended use, and the testing conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain information related to software performance, acceptance criteria for an AI/ML device, or any study involving human readers or expert consensus on image interpretation.

    The document is about a mechanical atherectomy system used for clearing peripheral vasculature. The "testing summary" section lists various engineering and functional tests to demonstrate the device's performance and safety (e.g., dimensional inspection, simulated use, torque tests, temperature rise, trackability, etc.). These are typical for a physical medical device, not for a software-based AI/ML product.

    Therefore, I cannot provide a response that directly answers your request using the provided text. The questions you posed are specifically designed for evaluating an AI/ML-based medical device, particularly one that involves image interpretation and human expert involvement in ground truth establishment.

    If you have a different document describing an AI/ML medical device, please provide it, and I would be happy to analyze it according to your criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181877

    Validate with FDA (Live)

    Date Cleared
    2018-12-17

    (157 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. The system is not intended for use in the coronary, carotid, iliac, pulmonary, or renal vasculature.

    Phoenix Atherectomy Plus System:

    When used with the Phoenix Aspiration Pump as the vacuum source, the Phoenix 2.4mm Deflecting Atherectomy System is indicated for the removal of thrombus from vessels of the peripheral arterial vasculature.

    Device Description

    The Phoenix 2.4mm Atherectomy Plus System comprises the currently cleared Phoenix 2.4mm Deflecting Atherectomy System (K172386) and a new Phoenix Aspiration Pump. which when used in conjunction with each other are designed for removal of thrombus from vessels of the peripheral arterial vasculature.

    The Phoenix 2.4mm Deflecting Atherectomy System and the Phoenix Aspiration Pump are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The Phoenix 2.4mm Deflecting Atherectomy System comprises the Phoenix 2.4mm Deflecting Atherectomy Catheter and the Phoenix Atherectomy Handle with Wire Support Clip.

    The Phoenix Catheter is an over-the-wire (OTW), multi-lumen Catheter with a cutter at the distal tip that continuously captures and clears debulked (excised) material proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectorny Catheter at its specified rotational speed, and is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Phoenix Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System. The Phoenix Catheter is compatible with commercially available 0.014" exchange length (260 cm or greater) guidewires.

    For the purpose of this 510(k), the design of the Phoenix 2.4mm Deflecting Atherectomy System remains unchanged.

    The Phoenix Aspiration Pump connects to the disposal outlet of the Phoenix Catheter via the pump's aspiration tubing. The exit end of the aspiration tubing is connected to a waste disposal bag to collect aspirated material. A 60 ml syringe is provided as part of the Aspiration Pump Assembly to assist in priming the pump aspiration tubing and purging the aspiration system of air. The Aspiration Pump is battery-operated and is operated by an ON/OFF switch. It serves as a vacuum source for the Aspiration System for aspirating the thrombus from the target vessel out of the Phoenix Catheter via the pump aspiration tubing and into the disposal bag. The Aspiration Pump is non-patient contacting.

    AI/ML Overview

    The provided text is a 510(k) summary for the Phoenix 2.4mm Atherectomy Plus System. It describes the device, its intended use, and a comparison to predicate and reference devices, as well as a summary of testing performed to demonstrate substantial equivalence. However, it does not contain the specific information requested about acceptance criteria and a study that proves the device meets those criteria in the context of an AI/algorithm-driven device performance study.

    The document describes the device as a mechanical atherectomy system with an aspiration pump for removing thrombus, which is a physical device, not an AI or algorithm. The "Testing Summary" section lists various engineering and preclinical tests (e.g., visual inspection, vacuum and leak test, flow rate test, system simulated use test, preclinical animal testing), which are typical for medical devices of this nature. There is no mention of an algorithm, AI, or any data-driven performance study comparing human readers with and without AI assistance, or standalone algorithm performance, or ground truth establishment based on expert consensus for image analysis.

    Therefore, I cannot provide the requested information from the given text. The text does not detail an AI/algorithm-driven study with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172386

    Validate with FDA (Live)

    Date Cleared
    2017-08-24

    (16 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle.

    The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle. The Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System.

    There are multiple models of the Phoenix Catheter. The Phoenix Tracking Catheter models track directly over the guidewire with no tip deflection capability. These models are available in 1.8 mm and 2.2 mm tip diameter sizes; a 2.4 mm tip diameter size (identical in design to the currently marketed 2.4 mm Phoenix Deflecting Catheter) has been added under this 510(k). The controls for rotation are housed in the Phoenix Handle when the Catheter is inserted into the Handle. All Phoenix Catheter models are compatible with commercially available 0.014" exchange length (260 cm or greater) guidewires, and all use the same Phoenix Handle.

    This 510(k) includes modifications to the 1.8 mm and 2.2 mm tip diameter 149 cm Phoenix Tracking Catheter models to improve the overall catheter pushability and internal debris maceration of the longer length catheter models and (as just noted previously) to also add a 2.4 mm tip diameter 130 cm Phoenix Tracking Catheter to the Phoenix Catheter product family. Tables 9-1 and 9-2 summarize the subject modifications relative to the predicate devices.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Phoenix Atherectomy System. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study proving the device meets specific performance acceptance criteria from an AI/reader study perspective.

    Therefore, many of the requested categories for AI/reader study evaluation cannot be fulfilled from this document.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not detail specific acceptance criteria with numerical targets. Instead, it states that the modified Phoenix Atherectomy System's performance and technological characteristics were evaluated by a series of tests to demonstrate equivalence to the predicate device. The general acceptance is that the device "meet pre-defined design requirements and that the modified Phoenix Atherectomy System performs in a manner equivalent to the predicate Phoenix Atherectomy System with the identical intended use."

    The tests performed include:

    • Dimensional and Visual Inspection
    • Simulated Use
    • Cutter Torque / Chain Torque-to-Failure Test
    • Functional Outer Shaft Torque Test
    • Knob to Shaft Testing
    • Catheter Drive Train Stress Test
    • Cutter Stall Test
    • Temperature Rise of Catheter During Simulated Use
    • Kink Bend Radius Test
    • Catheter Trackability in Below-the-Knee Anatomy
    • Sheath Flow Rate (2.4mm X 130 cm Phoenix Catheter only)
    • Sheath Compatibility (2.4mm X 130 cm Phoenix Catheter only)

    The document asserts that "The results from this testing demonstrate that the performance and technological characteristics of the modified Phoenix Atherectomy System meet pre-defined design requirements and that the modified Phoenix Atherectomy System performs in a manner equivalent to the predicate Phoenix Atherectomy System with the identical intended use."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described appears to be bench testing and simulated use, not clinical or human-data-driven studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document does not describe an AI/reader study or the establishment of ground truth by experts in that context. The "ground truth" here would relate to the engineering specifications and performance capabilities verified through the various tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable as this is not an AI/reader study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study or AI assistance is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a manual atherectomy system, not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance would be derived from engineering specifications, material properties, and the results of the described bench and simulated use tests. It is based on the device's physical and functional capabilities as measured against design requirements for safety and effectiveness, and comparison to the predicate device.

    8. The sample size for the training set

    Not applicable. There is no AI model or training set described.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI model or training set described.