510(k) Data Aggregation

    K Number
    K243940

    Validate with FDA (Live)

    Date Cleared
    2025-02-21

    (63 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VITA YZ MULTI TRANSLUCENT is indicated for:

    • · fully anatomical anterior and posterior crowns
    • · fully anatomical 4 unit anterior and posterior bridges
    • · fully and partially veneered single tooth and up to 4-unit bridge substructures in the anterior and posterior tooth regions
    • · inlays
    • · Onlays
    • · veneers
    Device Description

    VITA YZ MULTI TRANSLUCENT are zirconia blanks for reliable shade reproduction. They can be used for the production of fully and partially veneered reconstructions and monolithic bridge restorations in the anterior and posterior tooth regions. VITA YZ MULTI TRANSLUCENT is part of the VITA YZ SOLUTIONS product group.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a dental material, VITA YZ MULTI TRANSLUCENT. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing extensive details on a comprehensive standalone study with specific acceptance criteria, ground truth, and human reader performance that would be typical for an AI/ML medical device.

    Therefore, many of the requested details about acceptance criteria, study design (e.g., sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance), and ground truth establishment are not directly available in this document as it pertains to a material rather than an AI/ML algorithm.

    However, I can extract the acceptance criteria and performance data available for the material's properties, as this document focuses on non-clinical testing of the material itself.

    Here's the information that can be extracted, interpreted in the context of the material's performance:

    1. Table of Acceptance Criteria and Reported Device Performance (for the material's properties):

    Test ParameterAcceptance CriteriaReported Device Performance (VITA YZ MT)
    Flexural Strength> 800 MPa922 ± 95 MPa
    Linear Coeff. of Thermal Expansion (CTE)Within manufacturer's specified range (9.8 - 10.8 x 10-6 K-1 for VITA YZ ST Multicolor)10.4 × 10-6 K-1
    Chemical Solubility (Weight Loss)< 100 µg/cm²0 µg/cm²
    Radioactivity (Uranium-238)≤ 1.0 Bq/g≤ 1.0 Bq/g
    Packaging, Marking, LabelingCompliance with EN ISO 6872:2019Compliant (Visual inspection)
    UniformityEven distribution, no segregation of pigmentsCompliant (Visual inspection)
    Freedom from Extraneous MaterialsAbsence of foreign materialCompliant (Visual inspection)
    BiocompatibilityNo adverse biological effectsVerified (According to DIN EN ISO 10993-1 and DIN EN ISO 7405)

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated for each test, but standard material testing typically involves multiple samples for statistically significant results. The summary mentions "tested samples" without providing the exact number for each parameter (e.g., number of flexural strength specimens).
    • Data Provenance: The tests were conducted according to EN ISO 6872 standards for dental ceramic materials. The radioactivity test was performed at Jülich Forschungszentrum. These are typically laboratory-based, controlled material property tests, not based on patient data provenance.
      • Retrospective/Prospective: Not applicable, as these are material property tests, not clinical studies on patient outcomes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Experts: Not applicable. For material property tests, "ground truth" is established through standardized physical and chemical measurements (e.g., using calibrated instruments for flexural strength, spectrometers for chemical composition, etc.) as defined by international standards like EN ISO 6872. These are objective measurements rather than subjective expert interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. Material property tests rely on objective measurements and established scientific protocols for data collection and analysis, not on expert consensus or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This is a material, not an AI/ML device, so MRMC studies involving human readers and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable. This device is a material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: For these material tests, the "ground truth" is defined by the physical and chemical properties as measured by validated testing methods adhering to international standards (e.g., EN ISO 6872). It is based on objective laboratory measurements, not expert consensus, pathology, or patient outcomes data.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable. This is a manufactured material, not an AI/ML algorithm that requires a training set. The "development" of the material would involve iterative formulation and testing, but not a "training set" in the AI/ML sense.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable, for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180703

    Validate with FDA (Live)

    Date Cleared
    2018-08-01

    (135 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VITA YZ® ST and VITA YZ® XT materials are blanks used for fabricating dental restorations.

    VITA YZ® ST is indicated for fully anatomical anterior crowns, multi-unit fully anatomical anterior and posterior bridges, fully anatomical crowns and multi-unit bridges on direct screw-retained restorations, and single-tooth and multi-unit bridge frameworks of anterior and posterior crowns and bridges.

    VITA YZ® XT is indicated for fully anatomical anterior and posterior crowns, fully anatomical 3-unit anterior and posterior bridges, and anterior and posterior single-tooth and 3-unit bridge frameworks.

    Device Description

    VITA YZ ST and VITA YZ XT discs are pre-sintered zirconia materials, partially stabilized with yttrium oxide (Y-TZP, yttria stabilized tetragonal zirconia polycrystal). VITA YZ ST is made of Super Translucent Zirconia and VITA YZ XT is made of Extra Translucent Zirconia. They offer higher translucency to better match natural dentition.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for two dental materials, VITA YZ ST and VITA YZ XT (zirconia blanks for fabricating dental restorations).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative sense for all attributes. Instead, it refers to meeting specific international standards and demonstrating "similar" properties to predicate devices. For mechanical strength, a direct comparison is given.

    Acceptance Criteria / Standard MetReported VITA YZ ST PerformanceReported VITA YZ XT PerformancePredicate (VITA IN-CERAM YZ) PerformancePredicate (priti® multidisc Zr02 HT) Performance
    Material Composition (similar to predicate)YesYesIdentical Yttrium-Stabilized ZirconiaSimilar Zirconia
    Physical Properties (similar to predicate)YesYesYesYes
    Indications for Use (similar to predicate)YesYesYesYes
    Meet ISO 6872:2015 – Dentistry, Ceramic materials (requirements)YesYesYesYes
    Meet EN ISO 9693-2:2016 – Dentistry, Compatibility Testing, Ceramic-ceramic systems (requirements)YesYesNot explicitly stated for predicateNot explicitly stated for predicate
    Meet Biocompatibility requirements per ISO 10933-1:2009YesYesYesYes
    Meet Biocompatibility requirements per ISO 7405:2008YesYesYesYes
    Flexural Strength (MPa)934 ± 104 MPa678 ± 62 MPa1165 ± 56 MPa> 650 MPa

    Note: For many criteria, the "performance" is stated as meeting a standard or being "similar" rather than a specific numerical result. The flexural strength is the most quantitative comparison provided. The conclusion explicitly states that "The differences between VITA YZ ST & VITA YZ XT and the predicate devices do not impact safety and effectiveness, as VITA YZ ST & VITA YZ XT share same indications as the predicate devices, meet the same biocompatibility, and ISO 6872 test criteria."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data or human-derived data. All performance testing described is non-clinical (bench testing) and biocompatibility testing.

    • Sample Size for Bench Tests: Not explicitly stated for flexural strength or other mechanical/physical properties. Standard testing methods (e.g., ISO 6872) typically define minimum sample sizes for such tests.
    • Data Provenance: The data is generated from non-clinical bench testing. The country of origin for the testing is not specified, but the applicant (VITA Zahnfabrik H.Rauter GmbH Co.) is located in Germany.
    • Retrospective or Prospective: Not applicable as no clinical data or human studies were conducted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. "Ground truth" in the context of clinical diagnosis or interpretation by experts is not relevant here, as this is a materials science evaluation and not a diagnostic device. The "truth" is established by conformance to international material standards and laboratory measurements.

    4. Adjudication Method for the Test Set

    Not applicable, as no expert adjudication involving human readers/interpreters was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The device is a dental material, not a diagnostic or AI-assisted diagnostic device. The study focused on material properties and biocompatibility.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a material for fabricating dental restorations, not an algorithm.

    7. The Type of Ground Truth Used

    • For material properties (e.g., flexural strength, compliance with ISO 6872): The "ground truth" is defined by the requirements of the international standards (e.g., ISO 6872) and measurable physical properties determined through laboratory testing using standardized methodologies.
    • For biocompatibility: The "ground truth" is established by conformance to the requirements of the international standards for biocompatibility (ISO 10993-1 and ISO 7405) through specific biological evaluation tests.

    8. The Sample Size for the Training Set

    Not applicable. No "training set" in the context of machine learning or AI development was used, as this is a material science product.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. No training set was used.