510(k) Data Aggregation

    K Number
    K213357

    Validate with FDA (Live)

    Date Cleared
    2022-07-19

    (280 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Study Watch with Irregular Pulse Monitor is indicated for use by adult patients (22 years and older) who have been diagnosed with, or are susceptible to developing, atrial fibrillation enabling them to monitor and record their heart rhythms. Study Watch is also intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms.

    The Study Watch is intended for prescription use only.

    Device Description

    The Study Watch with Irregular Pulse Monitor is a miniaturized physiological data monitoring device that is intended to record, store, transfer and display single-channel electrocardiogram (ECG) rhythms. The Study Watch with Irregular Pulse Monitor is intended to notify the user in the event of an irregular pulse, such as atrial fibrillation (AF), and recommend acquisition of an ECG. The device utilizes a band and proprietary watch that incorporates two dedicated sensing electrodes to obtain a single-channel ECG measurement. In practice, a healthcare professional (HCP) may prescribe the Study Watch with Irregular Pulse Monitor to a patient and recommend the capture and transmission of one or more ECG rhythms on a daily basis for analysis by the HCP. The patient uses prompts provided on the graphical user interface (GUI) of the watch to collect a real-time 45-second ECG measurement. The patient is able to capture an ECG rhythm on-demand by navigation to the ECG menu, upon receipt of an irregular pulse notification. While an animated ECG waveform is briefly displayed on the GUI of the watch, the patient will not have direct access to the collected waveform. Using the provided charging dock (Cradle) and Verily's Study Hub or Watch App, the ECG measurements are securely transferred to Verily's cloud server and are viewed by the HCP using a web portal. Additionally, data may be securely transferred to interoperable devices for further analysis. The web portal is solely intended for use by the HCP to view time-stamped ECG waveforms collected by the patient and does not include any analysis features. Collectively, the Study Watch consists of the wearable watch and band, the Study Hub or Watch app (used to transfer data from the watch to the cloud), a Cradle, and the web portal.

    AI/ML Overview

    Below is a detailed description of the acceptance criteria and the study proving the device meets these criteria, based on the provided document.

    Device Name: Study Watch with Irregular Pulse Monitor

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Threshold)Reported Device Performance (Home Use)
    Per-interval SensitivityNot explicitly stated (only "exceeds pre-specified thresholds")96.1% (95% CI: 92.7 - 98.0)
    Per-interval SpecificityNot explicitly stated (only "exceeds pre-specified thresholds")98.1% (95% CI: 97.2 - 99.1)

    Note: The document states the algorithm "exceeds the pre-specified per-interval performance thresholds for sensitivity and specificity," but the exact numerical thresholds are not provided within the given text.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document mentions "thousands of ECG data collected from iRhythm Technologies' Zio devices" for algorithm training but does not explicitly state the sample size for the validation (test) set. However, it describes the "Study Watch AF Detection At Home (NCT04546763) data" as consisting of "free-living, multi-day PPG recordings obtained from the Study Watch, along with ECG-based rhythm labels obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously."
    • Data Provenance: The study was conducted in the USA, with regional demographics provided: Midwest (8%), Mountain (32%), West (32%), Northeast (14%), South (13%). The data was collected prospectively in a "free-living (home use) setting."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not specified.
    • Qualifications of Experts: The algorithm was trained using ECG data that had undergone "Certified Cardiographic Technician (CCT) review." It can be inferred that similar qualified personnel were involved in establishing the ground truth for the validation set, as the reference standard (Zio XT Patches) typically requires expert interpretation.

    4. Adjudication Method for the Test Set

    • The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It only mentions that ECG-based rhythm labels were "obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously" and that the training data underwent "Certified Cardiographic Technician (CCT) review."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • No, an MRMC comparative effectiveness study was not done in the context of human readers improving with AI vs. without AI assistance. This study focused on the standalone performance of the irregular pulse monitoring algorithm.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone performance study was done. The described "Clinical Performance Data" section focuses solely on the algorithm's performance (sensitivity and specificity) in detecting irregular pulses based on PPG data, using a reference ECG as ground truth, without human-in-the-loop evaluation.

    7. The Type of Ground Truth Used

    • Clinical Ground Truth: The ground truth for validation was established using ECG-based rhythm labels obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously with the Study Watch. This is a robust clinical ground truth for cardiac rhythm analysis.

    8. The Sample Size for the Training Set

    • The algorithm was trained using "thousands of ECG data collected from iRhythm Technologies' Zio devices." A specific number is not provided beyond "thousands."

    9. How the Ground Truth for the Training Set Was Established

    • The ground truth for the training set was established through "ECG data collected from iRhythm Technologies' Zio devices, which have undergone Certified Cardiographic Technician (CCT) review." This indicates human expert review by qualified professionals.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192415

    Validate with FDA (Live)

    Date Cleared
    2020-01-17

    (135 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Study Watch with Irregular Pulse Monitor is indicated for use by adult patients (22 years and older) who have been diagnosed with, or are susceptible to developing, atrial fibrillation enabling them to monitor and record their heart rhythms. Study Watch is also intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms. The Irregular Pulse Monitor is indicated for use in professional healthcare facility environments, while worn as a wrist watch.

    Device Description

    The Study Watch with Irregular Pulse Monitor is a miniaturized physiological data monitoring device that is intended to record, store, transfer and display single-channel electrocardiogram (ECG) rhythms. The Study Watch with Irregular Pulse Monitor is intended to notify the user in the event of irregular pulse, such as atrial fibrillation (AF), and recommend acquisition of an ECG. The device utilizes a band and proprietary watch that incorporates two dedicated sensing electrodes to obtain a single-channel ECG measurement. In practice, a healthcare professional (HCP) may prescribe the Study Watch with Irregular Pulse Monitor to a patient and recommend the capture and transmission of one or more ECG rhythms on a daily basis for analysis by the HCP. The patient uses prompts provided on the graphical user interface (GUI) of the watch to collect a realtime 60-second ECG measurement. The patient is able to capture an ECG rhythm on-demand by navigation to the ECG menu, in response to a daily ECG reminder, or upon receipt of an irregular pulse notification. While an animated ECG waveform is briefly displayed on the GUI of the watch, the patient will not have direct access to the collected wave form. Using the provided charging dock (Cradle) and Verily's Study Hub device, the ECG measurements are securely transferred to Verily's cloud server and are viewed by the HCP using a web portal. The web portal is solely intended for use by the HCP to view time-stamped ECG waveforms collected by the patient and does not include any analysis features. Collectively, the Study Watch consists of the wearable watch and band, the Study Hub (used to transfer data from the watch to the cloud), a Cradle, and the web portal.

    AI/ML Overview

    The document describes the Study Watch with Irregular Pulse Monitor, focusing on its ability to detect irregular pulses, particularly atrial fibrillation (AF). The performance data presented primarily relates to the Irregular Pulse Monitor algorithm, which uses photoplethysmogram (PPG) data as input.

    Here's a breakdown of the acceptance criteria and study findings:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaRequirement (Implicit)Reported Device Performance (PPG-based Irregular Pulse Monitor algorithm)
    Sensitivity (Per-interval)Not explicitly stated as a numerical acceptance criterion, but the text states the algorithm "exceeds the pre-specified per-interval performance thresholds for sensitivity."0.85 (85%) with a confidence interval of (0.79-0.90)
    Specificity (Per-interval)Not explicitly stated as a numerical acceptance criterion, but the text states the algorithm "exceeds the pre-specified per-interval performance thresholds for specificity."0.96 (96%) with a confidence interval of (0.93-0.99)

    Note: The document does not provide explicit numerical acceptance criteria values, only that the reported performance exceeds them. These reported values are for the PPG-based irregular pulse monitor algorithm, not necessarily the ECG functionality of the watch which is for recording, storing, transfer, and display.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state a numerical sample size for the test set of the clinical study. It mentions the study "collected Study Watch PPG waveform (Test) data in clinic from patients with a history of AF."
    • Data Provenance: The document does not specify the country of origin. It indicates the data was collected "in clinic," implying a prospective clinical study specifically for this device's performance validation.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The document states that "contemporaneous Holter ECG (Reference) data" was used as "ground truth." It does not mention the use of human experts to establish this ground truth for the test set. Holter ECG data is considered an objective gold standard for rhythm analysis.

    4. Adjudication Method for the Test Set

    • The document implies that the ground truth was established by "contemporaneous Holter ECG (Reference) data," which would not require a human adjudication method in the traditional sense (e.g., 2+1, 3+1). The Holter ECG itself serves as the reference.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document does not describe an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance. The study focuses purely on the standalone performance of the irregular pulse monitor algorithm against a clinical reference standard (Holter ECG).

    6. Standalone (Algorithm Only) Performance

    • Yes, a standalone performance evaluation was done. The reported sensitivity and specificity values of 85% and 96%, respectively, are for the "PPG-based Irregular Pulse Monitor algorithm" operating on the Study Watch PPG data, with Holter ECG as the ground truth. This indicates algorithm-only performance.

    7. Type of Ground Truth Used

    • The ground truth used for the clinical study was contemporaneous Holter ECG data. This is a clinical gold standard for continuous heart rhythm monitoring.

    8. Sample Size for the Training Set

    • The document does not provide information regarding the sample size used for the training set of the algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not provide information on how the ground truth for the training set was established. Given the nature of a machine learning algorithm for pulse monitoring, it's highly probable that ground truth for training data would also have been established using reference ECGs or similar clinical gold standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182456

    Validate with FDA (Live)

    Device Name
    Study Watch
    Date Cleared
    2019-01-17

    (132 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Study Watch is intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms. The Study Watch is intended for use by healthcare professionals, adult patients (22 years or older) with known or suspected heart conditions, and health conscious individuals.

    Device Description

    The Study Watch is a miniaturized physiological data monitoring device that is intended to record, store, transfer and display single-channel electrocardiogram (ECG) rhythms. The device utilizes a band and proprietary watch that incorporates two dedicated sensing electrodes to obtain a single-channel ECG measurement. In practice, a healthcare professional (HCP) may prescribe the Study Watch to a patient and recommend the capture and transmission of one or more ECG rhythms on a daily basis for analysis by the HCP. The patient uses prompts provided on the graphical user interface (GUI) of the watch to collect a real-time 60-second ECG measurement. While an animated ECG waveform is briefly displayed on the GUI of the watch, the patient will not have direct access to the collected waveform. Using the provided charging dock (Cradle) and Verily's Study Hub device, the ECG measurements are securely transferred to Verily's cloud server and are viewed by the HCP using a web portal. The web portal is solely intended for use by the HCP to view time-stamped ECG waveforms collected by the patient and does not include any analysis features. Collectively, the Study Watch consists of the wearable watch and band, the Study Hub (used to transfer data from the watch to the cloud), a Cradle, and the web portal.

    AI/ML Overview

    The device in question is the Study Watch, intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of specific acceptance criteria. However, it states that "All tests confirmed that the product met the predetermined acceptance criteria." It also details the clinical performance finding:

    Acceptance Criteria / Performance AspectReported Device Performance
    Non-Clinical PerformanceMeeting Predetermined Acceptance Criteria: "All tests confirmed that the product met the predetermined acceptance criteria." This included validation of specifications related to software development, sensor performance, stability, and material properties.
    Clinical Performance (Qualitative)Qualitative Clinical Equivalence: The Study Watch generates data that meets the clinical quality requirements for an accurate ECG waveform display. The study sufficiently drew a conclusion of substantial equivalence based on meeting qualitative acceptance criteria.
    Clinical Performance (Quantitative)Inter-beat Interval (IBI) Bland-Altman Analysis: Bias of 1.2ms and 95% limits-of-agreement <20ms when compared to an industry-standard 12-lead ECG device. The study sufficiently drew a conclusion of substantial equivalence based on meeting quantitative acceptance criteria.
    Human Factors AssessmentSafety and Effectiveness: "This study demonstrated that the Study Watch is safe and effective for the specified intended use" by identifying errors associated with use that could lead to patient harm and assessing effectiveness of risk management measures.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size for the clinical performance test set, nor does it specify the country of origin of the data or whether it was retrospective or prospective. It only generically mentions "clinical performance testing was conducted."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number of experts used or their qualifications to establish ground truth for the test set. It only mentions comparison to an "industry-standard 12-lead ECG device."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not specify any adjudication method used for the test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The Study Watch, unlike its predicate, does not include any analysis features for detecting conditions like atrial fibrillation or normal sinus rhythm and therefore does not involve AI assistance for human readers in that capacity. Its purpose is to record, store, transfer, and display single-channel ECG rhythms for review by a healthcare professional.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done to some extent for the device's core function of ECG recording and display, as indicated by the "Clinical performance testing was conducted to demonstrate that the Study Watch generates data that meets the clinical quality requirements for an accurate ECG waveform display." The quantitative comparison (Bland-Altman analysis) was directly comparing the Study Watch's output to a standard ECG device, implying evaluation of the device's intrinsic data generation capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical performance testing was established by comparing the data generated by the Study Watch to an "industry-standard 12-lead ECG device." This implies the 12-lead ECG device served as the reference standard for accurate ECG waveform measurement, particularly for the inter-beat interval analysis.

    8. The sample size for the training set

    The document does not specify any sample size for a training set. The Study Watch does not appear to employ an AI algorithm that would typically require a training set in the context of diagnostic interpretation features (as these features are explicitly absent).

    9. How the ground truth for the training set was established

    Not applicable, as no training set for an AI algorithm is mentioned or implied for diagnostic interpretation. The device's primary function is data acquisition and display, not AI-driven analysis of the ECG.