Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K200439

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-02-23

    (365 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Velano ExT™ Extension Set with needle-free connector is for single use only. Each port of the Velano ExT™ Extension Set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. The pressure rated port of the Velano ExT™ Extension Set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

    Device Description

    The Velano ExT™ Extension Set is a sterile, single use, non-pyrogenic intravenous administration extension set comprised of an approximate 5" inch tube bonded to a T-connection port on one end and a female luer on the other end with a clamp in between. The T-connection port has a swabable, needle-free port on one side and a locking male luer on the other side. The device contains and integrated stabilizer under the swabable needle-free port. This component (along with IV tape / adhesive) is intended to help anchor the extension set to the patient. This device is not made with the plasticizer Diethylhexylphthalate (DEHP). The Velano ExT™ Extension Set is compatible with Velano Vascular PIVOT™ devices.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device called the "Velano ExT™ Extension Set". This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving a device meets specific acceptance criteria based on a clinical trial or algorithm performance study as you might find for an AI/ML medical device.

    Therefore, the document does not contain the information required to answer questions 1 through 9, as it does not describe an acceptance criteria table, a study proving the device meets these criteria, or details regarding algorithm performance, ground truth establishment, or expert reviews.

    The document details the following regarding the device and its testing:

    • Device Name: Velano ExT™ Extension Set
    • Intended Use: Administration of intravenous fluids, and use with low-pressure power injectors up to 325 psi and a maximum flow rate of 10 mL/second.
    • Predicate Device: Velano Vascular Q2® Low Power Injector Extension Set (K182897)
    • Functional and Safety Testing Performed: This section lists various tests conducted to verify the device's design meets functional and performance requirements. These tests are physical and mechanical in nature, not related to an algorithm's diagnostic performance.
      • Visual inspection
      • Simulated shipping
      • Priming volume
      • Microbial ingress
      • Particulate
      • Backpressure leak under normal use and power injection
      • Flow rate for normal use and power injection
      • Spin collar height and spin collar angle
      • Simulated use
      • Tubing bond strength
      • Multiple Engagement
      • Continuous Engagement
      • Activation Force
      • Tubing kink resistance
    • Standards Applied: The document lists several ISO, USP, ASTM, and AAMI standards related to medical device sterilization, biological evaluation, particulate matter, packaging, and accelerated aging.

    In summary, this document is a regulatory submission for a physical medical device (an extension set), not an AI/ML-driven device. Thus, the concepts of acceptance criteria for algorithm performance, test set size, expert adjudication, MRMC studies, standalone algorithm performance, or training set details are not applicable or present in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193569

    Validate with FDA (Live)

    Device Name
    PIVO(TM)
    Manufacturer
    Date Cleared
    2020-01-21

    (29 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PIVO™ device is attached to a peripheral IV catheter for use as a direct blood draw device into a vacuum tube or a syringe.

    Device Description

    The PIVO™ device is a sterile, single use device. It is a needlefree blood collection device that attaches to a peripheral IV (PIV) catheter system. The device is comprised of an inner tube with a pusher/slider, proximal flexible tube with female luer, outer housing and clip-to-connect distal end. The clip-to-connect attaches to the PIV system. The female luer attaches to a blood transfer device or syringe. The inner tube is then advanced to collect a blood sample. Once complete, the inner tube is retracted, and the device is removed from the PIV. The device is available in three sizes, 20, 22, and 24 gauge.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device called PIVO™. This document focuses on demonstrating substantial equivalence to a previously cleared device, rather than a study proving performance against acceptance criteria in the context of AI/ML.

    Here's why it's not applicable to the requested information and what information it does contain:

    • No AI/ML Component: The PIVO™ device is a "blood specimen collection device" (a needle-free blood collection device that attaches to a peripheral IV catheter). There is no mention of any AI or machine learning component in its functionality or development. Therefore, questions related to AI performance, such as sample size for test sets, ground truth establishment, MRMC studies, or standalone algorithm performance, are not relevant.

    • Substantial Equivalence, Not De Novo Performance: The document's primary purpose is to justify that the modified PIVO™ device is "substantially equivalent" to an already legally marketed predicate device (K190604). This means the focus is on showing that the changes (alternate colorant material, second contract manufacturer, alternate packaging material) do not raise new questions of safety or effectiveness. It's not about proving inherent performance against new, specific acceptance criteria for a novel device.

    • Performance Testing for Equivalence: The "Performance Testing" section lists tests conducted to demonstrate that the modifications do not negatively impact the device's performance compared to the predicate. These are:

      • Leak testing
      • Dimensional testing
      • Joint strength testing
      • Flow rate testing
      • Packaging testing (per ASTM D4169-16, ASTM F1980-16, ISO 11607-1:2019)
      • Biocompatibility testing (per ISO 10993-1)

      The document states: "All testing was performed on sterilized product." And, "The changes made were tested using the same acceptance criteria as the predicate device and demonstrated that there are no new risks and the device is substantially equivalent."

    Since it's not an AI/ML device and the submission is for substantial equivalence rather than original performance evaluation, I cannot fill in the requested table and study details related to an AI/ML device. The document does not provide information on:

    1. Acceptance Criteria and Reported Device Performance (as typically defined for AI): It states "same acceptance criteria as the predicate device" for the performance tests listed, but doesn't quantify them.
    2. Sample sizes for test set and data provenance.
    3. Number of experts used to establish ground truth and qualifications.
    4. Adjudication method.
    5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
    6. Standalone (algorithm-only) performance.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    8. Sample size for the training set.
    9. How ground truth for the training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190604

    Validate with FDA (Live)

    Device Name
    PIVO
    Manufacturer
    Date Cleared
    2019-05-13

    (66 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PIVO™ device attaches to a peripheral IV catheter system for use as a direct blood draw device into a vacuum tube or a syringe.

    Device Description

    The PIVO™ device is a sterile, single use device. It is a needle- free blood collection device that attaches to a peripheral IV (PIV) catheter system. The device is comprised of an inner tube with a pusher/slider, proximal flexible tube with female luer, outer housing and clip-to-connect distal end. The clip-to-connect attaches to the PIV system. The female luer attaches to a blood transfer device or syringe. The inner tube is then advanced to collect a blood sample. Once complete, the inner tube is retracted, and the device is removed from the PIV. The device is available in three sizes, 20, 22, and 24 gauge.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the PIVO™ device, a blood specimen collection device. The submission is for a line extension to include a 24G compatible size device, along with some dimensional and material changes to existing 20G and 22G devices.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The text does not explicitly provide a table of acceptance criteria with corresponding performance metrics. However, it lists several performance tests conducted and states that "All testing was performed on sterilized product." The overarching acceptance criterion for the subject device (PIVO™ 24G and modified 20G/22G) is that it "demonstrated that there are no new risks and the device is substantially equivalent" to the predicate devices. It also states that testing used "the same acceptance criteria as the predicate device."

    Here's an inferred table based on the provided information:

    Performance TestImplied Acceptance Criteria (based on "no new risks" and "same acceptance criteria as predicate")Reported Device Performance
    Leak testingNo leaksDemonstrated no new risks, substantially equivalent to predicate.
    Dimensional testingMeets specified dimensions for compatibilityDifferences do not create additional risk to safety and effectiveness.
    Joint strength testingMeets specified joint strength requirementsDemonstrated no new risks, substantially equivalent to predicate.
    PIV compatibility testingCompatible with PIV gauges (24G for subject device)Compatible with smaller (24G) PIV systems; no additional risk.
    Flow rate testingMeets specified flow rate requirements (comparable to predicate)Demonstrated no new risks, substantially equivalent to predicate.
    Hemolysis testingNo significant hemolysisDemonstrated no new risks, substantially equivalent to predicate.
    Shelf life testingMaintains performance over shelf lifeDemonstrated no new risks, substantially equivalent to predicate.
    Biocompatibility testingMeets ISO 10993-1 requirementsDemonstrated no new risks, substantially equivalent to predicate.
    Proximal Tube Material(for material change) Performance (leak testing, joint strength, hemolysis) is not affected.Material change does not affect the effectiveness of the subject device.
    Inner Tubing Wall Thickness(for wall thickness changes) Performance (hemolysis, flow rate, kink resistance) is not affected.No additional risks to safety and effectiveness.

    The subsequent information is either not applicable (N/A), not explicitly stated (NSE), or can be inferred as follows:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: NSE. The document states that testing was performed, but does not specify the number of devices or samples used for each test.
    • Data Provenance: NSE regarding country of origin or whether it was retrospective/prospective. The testing was conducted by the manufacturer, Velano Vascular (based in San Francisco, California, though this doesn't specify test location).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. This device is a blood collection device, not an AI/diagnostic imaging device requiring expert interpretation of results for ground truth. Performance testing would rely on objective physical and chemical measurements rather than expert consensus on a 'ground truth' image or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, which is not the case for the performance testing described for this physical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This device is not an AI-assisted diagnostic tool, so an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This device is not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Physiological standards and engineering specifications. For a blood collection device, "ground truth" would be established by predefined engineering specifications (e.g., specific dimensions, flow rates), established biochemical limits (e.g., for hemolysis), and relevant international standards (e.g., ISO 10993-1 for biocompatibility). The claim of "no new risks" and "substantially equivalent" implies meeting these objective standards relevant to the device's function.

    8. The sample size for the training set

    • N/A. This device is a physical medical device, not an AI/machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    • N/A. As there is no training set for a physical device, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182897

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-26

    (161 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pressure Rated: The Q2® Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

    Non Pressure Rated: The Q2® Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.

    Device Description

    Sterile, single use non-pyrogenic intravenous administration extension set comprised of 5.75 inch tubing bonded to a male luer on one end and a T-connector on the other end with a clamp in between. The T-connector has a female luer on one side and a swabable, needle-free port on the other side. This device is not made with the plasticizer Diethylhexylphthalate (DEHP). The Velano Vascular Q2® Extension Set is compatible with PIVO™ devices.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device (Velano Vascular Q2 Extension Set), which focuses on demonstrating substantial equivalence to a predicate device, rather than proving that an AI/ML device meets specific performance criteria through a rigorous study. Therefore, the information typically required to answer your prompt about AI acceptance criteria and study details (such as sample size for training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance) for an AI/ML device is not present in this document.

    This document describes a traditional medical device (an intravascular administration set) and performs mechanical and functional testing to demonstrate its safety and effectiveness relative to a legally marketed predicate device.

    However, I can extract the relevant information regarding the device's functional and safety testing, despite it not being an AI-driven device.

    Here's a breakdown of the acceptance criteria and study information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a table of "acceptance criteria" against "reported performance" in a quantitative manner for each test. Instead, it lists the types of functional and safety tests performed and states that these tests ensured the device design meets its functional and performance requirements. The conclusion explicitly states that "Velano Vascular considers the Velano Vascular Q2® Extension Set device to be substantially equivalent to the predicate device listed above. This conclusion is based upon the intended use, design specifications, patient contacting materials, manufacturing and sterilization processes." This implies that the device met the implicit acceptance criteria for each test to achieve substantial equivalence.

    Implicit Acceptance Criteria and Demonstrated Performance (Inferred from the document):

    Test PerformedImplicit Acceptance CriterionReported Device Performance (Implied)
    Visual inspectionDevice must be free from visible defects.Met requirements (ensured device design meets functional and performance requirements).
    Simulated shippingDevice must withstand shipping stresses without damage or compromise to function.Met requirements.
    Priming volumePriming volume must be consistent with specifications and safe for intended use.< 1 mL (consistent with predicate and acceptable).
    Backpressure leak (normal use & power injection)No leakage under normal operating pressures and specified power injection pressures (max 325 psi).Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device compared to the predicate, which also withstands 325 psi).
    Flow rate (normal use & power injection)Flow rate must meet specified values for normal and power injection (10ml/sec at 325 psi).Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device, and the shorter length of the new device still results in acceptable flow rate and kink resistance).
    Simulated useDevice must perform as intended during simulated clinical use.Met requirements.
    Tubing bond strengthBonds must withstand specified forces without separation.Met requirements.
    Multiple EngagementNeedleless connector must function accurately after multiple engagements.Met requirements.
    Continuous EngagementNeedleless connector must maintain integrity during continuous engagement.Met requirements.
    Activation ForceForce required to activate the device must be within acceptable limits.Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device).
    Tubing kink resistanceTubing must resist kinking under normal manipulation.Met requirements (demonstrates the change in length does not risk safety or effectiveness).
    Multiple clampingClamp must function effectively after multiple closures.Met requirements.
    Prolonged clampingClamp must maintain occlusion over prolonged periods.Met requirements.
    PIVO compatibility testingDevice must be compatible with PIVO™ devices as claimed.Met requirements (demonstrates compatibility with PIVO™ devices does not affect safety or effectiveness).
    Sterilization (ISO 11135:2014)Sterilization process (EtO) achieves a sterility assurance level (SAL) of 1 x 10^-6.All patient contacting materials, sterilization, and design specifications are identical and unchanged from the predicate device. Additional sterilization assessments and testing were performed in accordance with ISO 11135:2014, achieving 1 x 10^-6 SAL.

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: The document repeatedly refers to "representative samples" being used for mechanical testing (Page 7). However, it does not specify the exact number of samples (sample size) used for each test.
    • Data Provenance: Not applicable in the context of "data" for an AI/ML device. For a physical device, the data provenance is the location where the manufacturing and testing occurred. This information is not detailed beyond the manufacturer's address (San Francisco, CA). The tests are presumably performed in a controlled laboratory environment. The testing is prospective in the sense that it's performed as part of the submission process for the new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as this is not an AI/ML device requiring human expert ground truth for classification or interpretation. The performance is assessed through objective physical and functional tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as this is not an AI/ML device requiring human expert adjudication. The test results are based on objective measurements and established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not applicable as this is not an AI/ML device. The device's "standalone" performance is its physical and functional performance during the mechanical tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For physical/mechanical devices, "ground truth" is established by engineering specifications, industry standards (e.g., ISO), and physical measurements against known values and benchmarks (like the predicate device's performance). For example, a leak test's "ground truth" is "no leak detected under specified pressure." This is not based on expert consensus in the diagnostic sense, but rather adherence to predefined engineering limits and physical laws.

    8. The sample size for the training set:

    • This is not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • This is not applicable as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163508

    Validate with FDA (Live)

    Device Name
    PIVO
    Manufacturer
    Date Cleared
    2017-02-09

    (57 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PIVO™ device is attached to a peripheral IV catheter for use as a direct blood draw device into a vacuum tube or a syringe.

    Device Description

    The PIVO™ device is a sterile, single use device. It is a needle- free blood collection device that attaches to a peripheral IV system (PIV). The device is comprised of an inner tube with a pusher/slider, proximal flexible tube with female luer, outer housing and clip-to-connect distal end. The clip-to-connect attaches to the PIV system. The female luer attaches to a blood transfer device or syringe. The device is then advanced to collect a blood sample. Once complete, the device is retracted and removed from the PIV. The device comes in two sizes compatible with 20 and 22 gage PIV's, respectively.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called PIVO™, a blood specimen collection device. It details the device's characteristics, indications for use, and a comparison to its predicate device, TIVA™.

    However, the document does not contain information related to a study proving the device meets acceptance criteria for an AI-powered medical device, or any acceptance criteria related to AI/algorithm performance. The acceptance criteria mentioned are for a physical medical device, PIVO™, comparing its performance characteristics to a predicate device (TIVA™) after material and component changes.

    Therefore, I cannot extract the information required by your prompt regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/algorithm.

    The document discusses:

    • Device Performance Testing:
      • Clamp functional testing
      • Leak testing
      • Dimensional testing
      • Joint strength testing
      • Performance testing
      • Flow rate testing
      • Biocompatibility testing (per ISO 10993-1, -4, -5, -10, -11)
    • Summary of Substantial Equivalence: It states that the changes made to PIVO™ (formerly TIVA™) do not raise new questions regarding safety and efficacy, and that PIVO™ is equivalent to the predicate device based on identical indications for use, principles of operation, fundamental scientific technology, and performance specifications. It explicitly states: "The changes made were tested using the same acceptance criteria as the predicate device and provide objective evidence that there are no new risks and the device is substantially equivalent."

    Based on the provided text, the device is a physical blood collection device, not an AI or algorithm-driven device. Thus, your request for an AI acceptance criteria table or details about an AI study (sample size, experts, adjudication, MRMC, standalone, ground truth, training set) cannot be fulfilled from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152924

    Validate with FDA (Live)

    Device Name
    TIVA¿
    Manufacturer
    Date Cleared
    2016-01-07

    (94 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TIVA™ device is attached to a peripheral IV catheter for use as a direct blood draw device into a vacuum tube or a syringe.

    Device Description

    The TIVA™ device is a sterile, single use device. It is a needle- free blood collection device that attaches to a peripheral IV system (PIV). The device is comprised of an inner tube with plunger, proximal flexible tube with female luer, and outer barrel with male luer. The male luer attaches to the PIV system. The female luer attaches to a blood transfer device or syringe. The device is then advanced to collect a blood sample. Once complete, the device is retracted and removed from the PIV. The device comes in two sizes compatible with 20 and 22 gauge PIV's, respectively.

    AI/ML Overview

    The provided document is a 510(k) summary for the TIVA™ blood specimen collection device (K152924). It compares the subject device to a predicate device (K142946). The document does not include information about AI/ML models or clinical studies involving human readers or comparative effectiveness studies of AI vs. human readers. Therefore, I can only address aspects related to the device's functional and safety testing as described.

    Based on the provided information, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details performance testing for both the predicate and subject devices. The "acceptance criteria" are implied by successful completion of these tests, demonstrating the device meets its functional and safety requirements.

    Acceptance Criteria (Implied by Test)Reported Device Performance
    Clamp functionalitySuccessful completion of "Clamp functional testing." The subject device's added clamp is an "additional blood control mechanism" and "performs as intended."
    Leak integritySuccessful completion of "Leak testing." For the predicate, this was part of "Performance testing." For the subject device, it was specifically re-evaluated due to modifications ("Additional bench testing (leak testing) demonstrates the modified TIVA device performs as intended").
    Joint strengthSuccessful completion of "Joint strength testing" (performed on the predicate device and unchanged for the subject device). Implies the joints maintain structural integrity during use.
    Flow rate functionalitySuccessful completion of "Flow rate testing" (performed on the predicate device and unchanged for the subject device). Implies the device facilitates appropriate blood draw speed.
    Biocompatibility (Hemolysis, Intracutaneous, Cytotoxicity, Sensitization, Material Mediated Pyrogen, Systemic Toxicity)Successful completion of "Biocompatibility testing per ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-11" (performed on the predicate device and unchanged for the subject device). Implies the device is bio-compatible and does not cause adverse biological reactions.
    Sterilization effectivenessSuccessful completion of "Sterilization validation testing per ISO-11137-1" (performed on the predicate device and unchanged for the subject device). Implies the device is terminally sterilized and maintains sterility until use.
    Performance as intended (general)"Performance testing provides objective evidence that the subject TIVA device does not introduce any new questions of safety or effectiveness compared to the predicate device and performs as intended."
    Substantial Equivalence"The TIVA device is substantially equivalent to the predicate device listed above. This conclusion is based upon the identical intended use, design specifications, patient contacting materials, manufacturing and sterilization processes. Additional bench testing (leak testing) demonstrates the modified TIVA device performs as intended and is substantially equivalent to the predicate TIVA." This is the ultimate acceptance criterion for 510(k) clearance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices tested) for the functional and safety tests. It generally refers to "representative samples of the device." The data provenance is internal testing conducted by Velano Vascular, Inc. (the manufacturer). The testing is retrospective in the sense that it's performed on manufactured devices as part of product verification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not provided. The ground truth for functional and safety testing is typically established by engineering specifications, recognized industry standards (like ISO standards), and internal quality control protocols. The "experts" would be the engineers and technicians performing and validating these tests, ensuring they follow established procedures and standards.

    4. Adjudication Method for the Test Set

    Not applicable in the context of functional and safety testing of a medical device. The tests have predefined pass/fail criteria based on engineering specifications and industry standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The TIVATM device is a mechanical blood collection device, not an AI/ML-driven diagnostic or interpretative tool. Therefore, no MRMC study or AI-related comparative effectiveness study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as the device is a mechanical blood collection tool and does not involve any algorithm or AI.

    7. The Type of Ground Truth Used

    For the functional and safety testing described, the ground truth is based on:

    • Engineering Specifications: Defined parameters for device performance (e.g., specific thresholds for leak testing, joint strength).
    • Recognized Standards: International standards like ISO 10993 (biocompatibility) and ISO 11137-1 (sterilization).
    • User Feedback: The document explicitly states the addition of the clamp was based on "user feedback that a clamp is preferred when using blood control devices with a syringe." This indirectly influences the "ground truth" for a desirable functional feature.

    8. The Sample Size for the Training Set

    Not applicable, as the device is a mechanical one and does not involve AI/ML requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set for AI/ML was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142946

    Validate with FDA (Live)

    Device Name
    TIVA
    Date Cleared
    2015-01-08

    (90 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TIVA™ device is attached to a peripheral IV catheter at the time of IV catheter placement for use as a direct blood draw device into a vacuum tube or a syringe.

    Device Description

    The TIVATM device is a sterile, single use device. It is a needle-free blood collection device that attaches to a peripheral IV system (PIV). The device is comprised of an inner tube with plunger, proximal flexible tube with female luer, and outer barrel with male luer. The male luer attaches to the PIV system. The female luer attaches to a blood transfer device or syringe. The device is then advanced to collect a blood sample. Once complete, the device is retracted and removed from the PIV. The device comes in two sizes 20 and 22 gauge.

    AI/ML Overview

    The provided text is a 510(k) summary for the TIVA™ blood specimen collection device. It describes the device, its intended use, and its substantial equivalence to a predicate device based on functional and performance requirements and a comparison of design specifications.

    However, the document does not contain specific acceptance criteria, a detailed study proving performance against those criteria, or information regarding sample sizes for training/test sets, ground truth establishment, expert qualifications, or comparative effectiveness studies (MRMC).

    The section titled "Functional and Safety Testing" briefly mentions that "representative samples of the device underwent biocompatibility, sterilization, and mechanical testing in accordance with the following industry standards," but it does not provide the results of these tests or specific acceptance criteria for them.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor can I provide information on the study specifics as the input document does not contain this level of detail.

    In summary, the document states that functional and safety testing was performed according to industry standards, but it does not report the acceptance criteria or the specific performance results for the TIVA™ device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1