Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-10

    (111 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sphere and Asphere:
    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens in the spherical and aspheric design are indicated for monthly disposable wear for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less that does not interfere with visual acuity.

    Toric:
    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens in the toric design is indicated for monthly disposable wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -2.25 diopters.

    Multifocal:
    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens in the multifocal design is indicated for monthly disposable wear for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    The lenses are to be prescribed for daily wear, with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, or disposal, as recommended by the eye care professional. Lenses should be discarded and replaced with a new pair each month, or more often, if recommended by the eye care professional.

    Device Description

    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens is a hemispherical shell with a molded base curve and a molded front surface. The hydrophilic characteristics allow aqueous solutions to enter the lens. The lens is fabricated from somofilcon A, which is a random co-polymer of silicone containing monomers and hydrophilic monomers. The lens consists of 44.0% somofilcon A and 56.0% water by weight when immersed in saline solution. The somofilcon A name has been adopted by the United States Adopted Names Council (USAN).

    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens contains phthalocyanine blue (21 CFR Part 74.3045) for visibility and handling. The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens incorporates a benzotriazole UV blocking monomer to help protect against transmission of harmful UV radiation. The lens blocks >95% in the UVB range (280nm - 315nm), and >80% in the UVA range (316nm - 380nm).

    The VizionFocus (somofilcon A) Silicone Hydrogel Soft (hydrophilic) Contact Lens is manufactured in spherical/aspheric, toric, multifocal design configurations. The material properties and available parameters of the finished lenses are as follows:

    ParameterRangeTolerance
    Chord Diameter11.00 mm to 15.00 mm±0.20 mm
    Center Thickness0.050 mm to 0.200 mmWhen ≤ 0.10 mm ±0.010 mm + 10% When > 0.10 mm ±0.015 mm + 5%
    Base Curve7.0 mm to 10.0 mm±0.20 mm
    Back Vertex Power (F'v)-20.00 D to +20.00 D (in 0.25 D steps)When 0.00 <
    Cylinder Power (F'c)-0.25 D to -2.25 D (in 0.25 D steps)When 0.00 < F'c ≤ 2.00 D ±0.25 D When 2.00 < F'c ≤4.00 D±0.37 D
    Cylinder Axis10° to 180° in 10° steps (in 10° steps)When 0.00 < F'c ≤ 1.50 D ± 8° When F'c > 1.50 D± 5°
    Multifocal Power+0.25 D to +4.00 D (in 0.25 D steps)±0.25D
    Oxygen Permeability (x 10-11 (cm2•ml O2)/(sec•ml•mmHg))60±20%
    Visible Light Transmittance>95%±5%
    Ultraviolet radiation Transmittance< 5% TUVB <20% TUVATUVB (280 to 315nm) < 0.05 TV TUVA (316 to 380nm) < 0.05 TV
    Water Content56%±2%
    Refractive Index1.398±0.005
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243953

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-10

    (108 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SPHERICAL/ASPHERE Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.

    The TORIC Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses are indicated for the correction of refractive error in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 5.00 diopters. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.

    The MULTIFOCAL Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.

    Daily wear replacement schedules may vary from patient to patient and should be decided by eye care practitioners in consultation with their patients.

    Eye care practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

    Eye care practitioners may prescribe any of the above lenses for single use daily disposable wear. When prescribed for daily disposable wear the lens is to be discarded after each removal.

    Device Description

    The Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are hemispherical shells with molded spherical base curves and molded front surfaces. The hydrophilic characteristics allow aqueous solutions to enter the lens. The lenses are fabricated from ocufilcon D, which is a hydrophilic co-polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and methacrylic acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% ocufilcon D and 55% water by weight when immersed in saline solution with polymeric wetting agents. The color additive (Pathalocyaninto(2-))Copper, Iron Oxide and Reactive Yellow 15 are added to the lens material to create a light yellow-green edge-to-edge color to make it easier to see when handling, and additionally, reduce transmittance of short wavelength light in the range of 380nm to 460nm. In addition, lenses contain a benzotriazole monomer to filter UVA and UVB radiation. The (ocufilcon D) name has been adopted by the United States Adopted Names Council (USAN).

    The Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are available tinted for visibility, and tinted to enhance or alter the apparent color of the eye. The lenses are processed to incorporate the 'listed' color additives and contain only the amount of the additive needed to accomplish the intended coloring effect. The lenses contain one or a combination of one or more of the following 'listed' color additives: Reactive Yellow 15, Rutile TiO2, Iron Oxide, (Pathalocyaninto(2-))Copper, Carbazole Violet, Phthalocyanine Green.

    When producing the color lenses, the manufacturing process changes the specifications to the light yellow-green contact lens by pad-printing the color pigment(s)—entrapping the colorants in the interpenetrating network of the contact lens material in a location that corresponds to the iris. The color pigments used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The tinting pattern has a clear pupil diameter of 6.0 mm.

    The Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) incorporate a UV absorbing monomer. The lenses filter >95% in the UVB range (280nm - 315nm), and >80% in the UVA range (315nm - 380nm).

    The Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are manufactured in the sphere/asphere, toric, and multifocal design configurations.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated summary pertain to a contact lens device, the Aurora (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lens (Tinted, Color). It's important to note that this document describes the device as a "Soft (Hydrophilic) Contact Lens," which is a device type, not an AI/ML algorithm. Therefore, many of the typical acceptance criteria and study parameters related to AI/ML device performance (such as sensitivity, specificity, MRMC studies, ground truth establishment for AI/ML, etc.) are not applicable to this type of medical device submission.

    The "acceptance criteria" for a contact lens device primarily revolve around its physical, chemical, and biological properties, as well as its safety and functionality for human use, established through non-clinical (laboratory) testing and comparisons to predicate devices. The study proving these criteria are met is typically a combination of non-clinical bench testing and demonstrating substantial equivalence to an already marketed predicate device.

    Here's the information based on the provided document, restructured to address your points where applicable:


    1. A table of acceptance criteria and the reported device performance

    For a contact lens, explicit "acceptance criteria" are often derived from recognized standards (like ANSI Z80.20 mentioned) and the performance of predicate devices. The reported device performance is compared against these and is generally expected to be within acceptable tolerances or equivalent to the predicate.

    ParameterAcceptance Criteria (Implicitly from Standard/Predicate)Reported Device Performance (Aurora)
    Chord Diameter11.00 mm to 15.00 mm (±0.20 mm tolerance)11.00 mm to 15.00 mm (±0.20 mm tolerance)
    Center Thickness0.05 mm to 0.15 mm (Specific tolerances)0.05 mm to 0.15 mm (Specific tolerances)
    Base Curve7.00 mm to 10.0 mm (±0.20 mm tolerance)7.00 mm to 10.0 mm (±0.20 mm tolerance)
    Back Vertex Power (F'v)+20.00D to -20.00D (Specific tolerances)+20.00D to -20.00D (Specific tolerances)
    Cylinder Power (F'c)-0.25D to -4.00D (Specific tolerances)-0.25D to -4.00D (Specific tolerances)
    Cylinder Axis10° to 180° (Specific tolerances)10° to 180° (Specific tolerances)
    Multifocal Add Power+0.25D to +4.00D (±0.37 D tolerance)+0.25D to +4.00D (±0.37 D tolerance)
    Surface AppearanceClear with no surface defectClear with no surface defect
    Oxygen Permeability (Dk)19.6 (±20% tolerance)19.6 (±20% tolerance)
    Light Transmission (380-780nm)95% (±5% tolerance)95% (±5% tolerance)
    Light Transmission (380-460nm)>75% (>75% tolerance)>75% (>75% tolerance)
    UV-B Transmittance< 5 %< 5 %
    UV-A Transmittance< 20 %< 20 %
    Water Content55% (±2% tolerance)55% (±2% tolerance)
    Refractive Index (hydrated)1.410 (±0.005 tolerance)1.410 (±0.005 tolerance)
    In-Vitro CytotoxicityNot cytotoxicNot cytotoxic
    Systemic ToxicityNo acute systemic toxicityNo acute systemic toxicity
    Acute Ocular IrritationNo ocular irritationNo ocular irritation
    Shelf LifeStability, sterility, and package integrity maintainedStability, sterility, and package integrity maintained over labeled expiration date

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of an AI/ML algorithm for image analysis or diagnostics. Instead, the "testing" refers to a series of non-clinical (laboratory) studies:

    • Toxicology: "All non-clinical toxicology tests were conducted in accordance with the GLP regulation." This includes In-Vitro Cytotoxicity, Systemic Toxicity, and Acute Ocular Irritation tests. The sample sizes for these biological tests would be defined by the relevant ISO standards (ISO 10993-5, ISO 10993-11, ISO 10993-23) but are not explicitly reported in this summary.
    • Shelf Life: Testing was performed to evaluate stability, sterility, and package integrity. Sample sizes are not specified for this summary.
    • Physicochemical & Mechanical Properties: Tests for refractive index, water content, Dk, % transmission, tensile strength, modulus, % elongation to break, specific gravity, and polymerization residuals. Sample sizes are not specified.

    Data Provenance: The studies were non-clinical, implying laboratory-based testing rather than patient data from a specific country. They were conducted under "Good Laboratory Practice (GLP) regulations," indicating a controlled, high-quality laboratory environment. There is no mention of retrospective or prospective patient data from a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the device is a contact lens, not an AI/ML diagnostic or image analysis device requiring expert-established ground truth on a test set of images or clinical cases. The ground truth for device properties is established by scientific measurement and adherence to defined standards.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical or image-based studies, which are not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study is relevant for AI/ML diagnostic devices where human readers (e.g., radiologists) interpret cases with or without AI assistance. This device is a contact lens. The document explicitly states: "Clinical testing is not required. The clinical performance of soft (hydrophilic) contact lenses manufactured from (ocufilcon D) materials has been demonstrated previously."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This device is a contact lens, not an AI/ML algorithm.

    7. The type of ground truth used

    For this contact lens device, the "ground truth" (or basis for evaluation) comes from:

    • Established scientific standards and methodologies: For physicochemical properties (e.g., water content, Dk, light transmission) and mechanical properties (e.g., tensile strength, modulus).
    • Biological safety standards: Adherence to ISO 10993 series for biocompatibility (cytotoxicity, systemic toxicity, ocular irritation).
    • Comparison to a legally marketed predicate device: "Substantial equivalence" is demonstrated by showing that the new device has the same fundamental scientific technology, materials, and performs comparably to a predicate device, which itself has established safety and effectiveness.

    8. The sample size for the training set

    This information is not applicable. This device is a contact lens, not an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sphere/Asphere:
    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Daily Disposable Soft (hydrophilic) Contact Lenses in the spherical/aspheric design are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less that does not interfere with visual acuity.

    Toric:
    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Daily Disposable Soft (hydrophilic) Contact Lenses in the toric design are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -2.25 diopters.

    Multifocal:
    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Daily Disposable Soft (hydrophilic) Contact Lenses in the multifocal design are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Daily Disposable Soft (hydrophilic) Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use. The VizionFocus Color (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lenses are tinted to enhance or alter the apparent color of the eye.

    Device Description

    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lenses are hemispherical shells with molded base curves and molded front surfaces. The hydrophilic characteristics allow aqueous solutions to enter the lenses are fabricated from mififilcon A. which is a random co-polymer of silicone containing monomers and hydrophilic monomers. The lenses consist of 40.0% mififilcon A and 60.0% water by weight when immersed in saline solution. The mififilcon A name has been adopted by the United States Adopted Names Council (USAN).

    The VizionFocus (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lens contains phthalocyanine blue (21 CFR Part 74.3045) for visibility and handling. The VizionFocus Color (miffilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lenses are tinted to enhance or alter the apparent color of the eye. The lenses are processed to incorporate the 'listed' color additives and contain only the amount of the additive needed to accomplish the intended coloring effect. The lenses contain one or a combination of one or more of the following 'listed' color additives: Rutile TiO2, Iron Oxide, (Pathalocyaninto(2-))Copper, Carbazole Violet, Phthalocyanine Green, Mica.

    When producing the color lenses, the manufacturing process changes the specifications to the clear contact lens by pad-printing the color pigment(s)-entrapping the colorants in the interpenetrating network of the contact lens material in a location that corresponds to the iris. The color pigments used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The tinting pattern has a clear pupil diameter of 6.0 mm.

    The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lenses incorporate a benzotriazole UV blocking monomer to help protect against transmission of harmful UV radiation. The lenses block >95% in the UVB range (280mm - 315nm), and >80% in the UVA range (316mm - 380mm). The VizionFocus and VizionFocus Color (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lenses are manufactured in spherical/aspheric, toric, and multifocal design configurations.

    AI/ML Overview

    The provided text describes the 510(k) submission for VizionFocus contact lenses, demonstrating their substantial equivalence to predicate devices. However, this submission does not contain acceptance criteria or study details for an AI/ML powered medical device. The document pertains to contact lenses, which are physical medical devices, not software-based AI/ML devices. Therefore, a table of acceptance criteria and study proving an AI device meets acceptance criteria cannot be extracted from this text.

    The information provided describes:

    • Device Type: VizionFocus (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lens; VizionFocus Color (mififilcon A) Silicone Hydrogel Soft (hydrophilic) Daily Disposable Contact Lens. These are physical contact lenses, not an AI/ML powered device.
    • Study Purpose: To demonstrate the safety and effectiveness of the contact lenses by comparing them to legally marketed predicate devices, establishing substantial equivalence. This is a typical regulatory pathway for physical medical devices.
    • Studies Conducted: Non-clinical performance testing (biocompatibility/toxicology, shelf life, bench tests like refractive index, water content, Dk, % transmission, tensile strength, modulus, % elongation to break, specific gravity, and polymerization residuals) and a clinical study.
    • Clinical Study Design: Multi-center, single-masked (examiner), bilateral, randomized concurrent-control study with 91-day treatment follow-up.
    • Clinical Study Sample Size: 120 subjects enrolled (60 test, 60 control), 117 completed.
    • Clinical Study Data Provenance: Not explicitly stated, but typically clinical trials for FDA submissions are conducted in the US or in compliance with international standards that are accepted by the FDA. The indication of a multi-center study implies diverse geographic recruitment within the study's operational regions.
    • Ground Truth: For the clinical study, the ground truth was established through biomicroscopy findings, symptoms, adverse reactions, and vision safety measures (keratometry changes, refractive changes, best corrected visual acuity) observed and recorded by examiners. There is no mention of expert consensus, pathology, or outcomes data in the context of establishing ground truth for an AI model.

    The prompt specifically asks for details around an AI/ML powered medical device's acceptance criteria and study, which are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1