Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K043604

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-04-29

    (121 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LifeShirt Real-Time is intended for use during daily activities of living and sleep, for the purpose of recording physiological data for later analysis by a physician. Respiration, ECG, pulse oximetry, blood pressure, temperature, and body position data may be collected. The system is intended to provide analysis of breathing patterns as an aid in classifying apneas as well as displaying heart rate changes from electrocardiographic waveforms in the wake and sleeping states as well as activities of daily living. The LifeShirt Real-Time is indicated for reuse by multiple adult patients in applications that may include pharmaceutical studies in which respiratory information is a useful indicator, or the general healthcare market where patients may be monitored at home and the data provided to their physicians as an aid to diagnosis and treatment.

    Device Description

    The LifeShirt Real-Time consists of the following components: the LifeShirt garment, LifeShirt Recorder/Transmitter, a wireless flash card, a data Flash Card, VivoMonitor™ real-time software and VivoLogic® analysis and reporting software. A cleared pulse oximeter, blood pressure cuff, and a temperature sensor are optional accessories provided for use with the device.

    AI/ML Overview

    This 510(k) summary focuses on modifications to an existing device, the LifeShirt, rather than a standalone study of a new medical device. As such, the information provided primarily addresses the equivalence of the modified device to its predicate, with limited detail on specific performance acceptance criteria and study data in the format requested for a new device's comprehensive performance evaluation.

    Here's an analysis based on the provided text, highlighting the limitations due to the nature of this 510(k):

    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Accurate data transmission and reception (for newly added wireless capabilities)"Performance testing included verification that data recorded by the LifeShirt recorder is accurately transmitted and received by the PC computer system." (Details on specific metrics or thresholds are not provided.)
    Electromagnetic Compatibility (EMC)"In addition, testing was performed to demonstrate electromagnetic compatibility." (Details on specific standards or results are not provided.)
    Safety and Effectiveness equivalence with predicate device"Performance data demonstrates that the LifeShirt Real-Time is as safe and effective as the predicate Multiple Patient Use LifeShirt." (Specific quantitative data is not provided.)

    Explanation: The document describes a "Special 510(k)" submission, which is used for modifications to a legally marketed device that do not significantly alter its fundamental scientific technology or intended use. For such submissions, the FDA typically focuses on demonstrating that the changes do not raise new questions of safety or effectiveness and that the modified device remains substantially equivalent to its predicate. Therefore, the "acceptance criteria" here are largely implied to be maintaining the performance of the predicate device, with specific verification of the new features (wireless communication and temperature probe).

    2. Sample size used for the test set and the data provenance

    The document does not specify a separate "test set" sample size in the context of clinical data for performance evaluation. The "performance testing" mentioned refers to engineering verification and validation activities for the new features (wireless data transmission, EMC). There is no mention of a pre-defined subject cohort used for clinical performance testing for the modified features.

    Data Provenance: Not explicitly stated for specific clinical tests, as the focus is on engineering verification of new features and equivalence to an already cleared device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable in this Special 510(k). The document describes engineering verification testing, not a clinical study requiring expert-established ground truth for a diagnostic outcome.

    4. Adjudication method for the test set

    Not applicable. There is no mention of a clinical test set requiring expert adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study is mentioned. The device records physiological data for later analysis by a physician.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This Special 510(k) is about a hardware device with software components, not a standalone algorithm in the typical sense of a diagnostic AI. The "VivoLogic analysis and reporting software" is part of the system for "later analysis by a physician," implying a human-in-the-loop workflow. The performance testing focuses on the accuracy of data transmission and fundamental device capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not detail the type of ground truth used as it relates to clinical diagnostic accuracy, as this was not a primary focus of this Special 510(k). The "ground truth" for the performance testing mentioned (e.g., accuracy of data transmission) would be engineering standards and successful transfer of data files.

    8. The sample size for the training set

    Not applicable. This document does not describe a machine learning algorithm that requires a training set for its development. The "VivoLogic processing software" likely uses established algorithms for physiological data analysis, not a learning-based approach that requires a training set in the AI sense.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031550

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-07-29

    (74 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reusable Multiple-Patient LifeShirt is intended for use during daily activities of living and sleep, for the purpose of recording physiological data for later analysis by a physician. Respiration, ECG, pulse oximetry, blood pressure, and body position data may be collected. The system is intended to provide analysis of breathing patterns as an aid in classifying apneas as well as displaying heart rate changes from electrocardiographic waveforms in the wake and sleeping states as well as activities of daily living. The Reusable Multiple-Patient LifeShirt is indicated for reuse by multiple adult patients in applications that may include pharmaceutical studies in which respiratory information is a useful indicator, or the general healthcare market where patients may be monitored at home and the data provided to their physicians as an aid to diagnosis and treatment.

    Device Description

    The Reusable Multiple-Patient LifeShirt consists of the following components:

    1. LifeShirt data acquisition system including the vest with inductive plethysmograph sensors, ECG sensors, a data cable and a connected Handspring Visor® device with software for data collection and storage onto a flash memory card;
    2. LifeShirt data transfer software to transmit the data from the user's personal computer to the VivoMetrics Data Center;
    3. VivoMetrics Data Center, a database operated by VivoMetrics personnel which stores and provides an audit trail for collected data; and
    4. VivoLogic data analysis software for viewing and analyzing patient data.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the VivoMetric's Reusable Multiple-Patient LifeShirt:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Device can be cleaned for multiple adult patient use.Cleaning validation performed and demonstrated. The ability of the cleaning process recommended in the Reusable Multiple-Patient LifeShirt User Guide to clean the device was validated.
    Device performs as intended following multiple washings.Functional validation performed and demonstrated. The device performs as intended following multiple washings.
    Maintains the same intended use as the predicate device.The device has the "same intended use" as the predicate device (recording physiological data for later analysis by physician; analysis of breathing patterns as an aid in classifying apneas; displaying heart rate changes; indicated for reuse by multiple adult patients in pharmaceutical studies or general healthcare market).
    Maintains the same principles of operation as the predicate device.The device has "same principles of operation" as the predicate device.
    Maintains the same technological characteristics as the predicate device such that the change in use does not affect diagnostic effect.The device has "same technological characteristics" as the predicate device. The only difference is reuse by multiple vs. single patients, which "does not affect the device's diagnostic effect," which monitors the same physiological parameters (respiration, ECG, pulse oximetry, blood pressure, body position) as the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state a specific sample size for a 'test set' in the context of evaluating diagnostic performance or clinical effectiveness. The validation focused on cleaning and functionality. No information on data provenance (country of origin, retrospective/prospective) related to a diagnostic test set is available, as the study was not primarily a diagnostic performance study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not mentioned in the provided 510(k) summary. The submission focuses on showing substantial equivalence concerning changes for multi-patient reuse and the associated cleaning/functional validation, not on evaluating diagnostic accuracy against a ground truth established by experts.

    4. Adjudication Method for the Test Set

    This information is not mentioned as there was no diagnostic performance study involving interpreting a test set of data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done based on the provided text. The submission focuses on substantial equivalence based on technical characteristics and validation of cleaning/functionality, not on evaluating human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone performance study of the VivoLogic Analysis Software in terms of its diagnostic accuracy (e.g., classifying apneas) was not explicitly detailed in the context of this 510(k) submission. The submission states the system is "intended to provide analysis of breathing patterns as an aid in classifying apneas," but it focuses on the hardware modification for multi-patient reuse, not the re-validation of the analytical software's standalone performance. The previous 510(k) for the single-patient device might have contained such data, but it is not included here.

    7. Type of Ground Truth Used

    For the cleaning validation, the "ground truth" would be the successful removal of biological contaminants to a specified level, verifiable through laboratory assays (e.g., microbial counts, protein detection). For functional validation, the "ground truth" would be that the device accurately continues to record physiological data within specified tolerances after multiple washings, which could be measured against a known signal input or by comparison to another validated device. Specific details of these ground truths (e.g., specific assays, tolerance levels) are not provided in this summary. There is no mention of ground truth in the context of diagnostic accuracy from expert consensus, pathology, or outcomes data, as this was not a diagnostic performance study.

    8. Sample Size for the Training Set

    This information is not applicable/not mentioned for this submission. The 510(k) specifically addresses a modification to allow multi-patient reuse, not the de novo development or re-training of an AI algorithm. The VivoLogic analysis software's training would have occurred prior to the original single-patient device's submission.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not mentioned for this submission. As above, the submission focuses on hardware modification, not AI algorithm training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011903

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-04-12

    (298 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Life Shirt system is intended for use by one adult during daily activities of The Life Shirt System is michaea cerading physiological data for later analysis by a physician. Respiration, ECG, pulse oximetry, blood pressure, and body position a physician. The system is intended to provide analysis of breathing data may be conected. Thic tyot appeas as well as displaying heart rate changes patterns as an all in claboring aparis the wake and sleeping states as well as trom electrocardidographic wavelorms in all was were aceutical studies in which acuvities of daily fiving. Trippiesul indicator, or the general healthcare market respiratory information is a userel manue and the data provided to their physicians as an aid to diagnosis and treatment.

    Device Description

    The LifeShirt is a noninvasive ambulatory monitoring device that continuously monitors and stores respiration and ECG onto a CompactFlash™ memory card within a portable battery powered Handspring Visor® worn on a belt or within a pocket. The patient may be located at home or in an alternate care setting. The monitored subject can enter symptoms, activities, and medications into the Handspring Visor® that becomes part of the digital data stream. An optional NONIN pulse oximeter and Critikon noninvasive blood pressure monitor can be connected into the Visor for recordings during sleep or defined activities. The purpose of this system is to collect and store cardiac, respiratory, and blood pressure data as well as body position, various types of activities annotated with symptoms and a medication diary for subsequent analysis and archival at a data center utilizing the VivoLogic 1.0 software program. Data is transferred to the data center generally after the end of a patient data recording session.

    AI/ML Overview

    I am sorry, but the provided text does not contain the acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) summary for the VivoMetrics LifeShirt System, detailing its components, intended use, and substantial equivalence to predicate devices. It includes a letter from the FDA confirming clearance and the indications for use. However, it does not present any performance data, acceptance criteria, or details of a validation study for the device's accuracy or effectiveness.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1