Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K241879

    Validate with FDA (Live)

    Device Name
    vRad Viewer
    Date Cleared
    2024-07-18

    (20 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The vRad Viewer software is used with general purpose computing hardware which meets or exceeds minimum specifications. The vRad Viewer is intended to receive, display, and transmit images for clinical purposes. The vRad Viewer is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications and networked with storage components. The vRad Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad Viewer can process medical images from DICOM modalities, such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM-compliant modalities.

    The vRad Viewer may be used for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM "For Presentation" format and may only be used on monitors determined to be appropriate for the desired task.

    Device Description

    The vRad Viewer is a Software as a Medical Device (SaMD) that imports medical data in the Digital Imaging and Communications and Medicine (DICOM) standard from local storage or remote Picture Archiving and Communications System (PACS) sources. The vRad Viewer is a modified version of the Viewer component cleared as part of vRad PACS with Mammography (K162145). Like the Viewer component of the vRad PACS with Mammography device, the vRad Viewer is intended to allow users to visualize, manipulate, and interpret medical images from X-ray radiography, X-ray computed tomography, MRI, ultrasound, nuclear medicine, and other DICOM-compliant modalities on compatible PC workstations. The vRad Viewer is designed to interface with radiologist workflow applications, such as PACS or caching applications, as well as operate as a standalone image viewer.

    The vRad Viewer allows users to perform operations related to the manipulation, enhancement, compression, and quantification of medical images at the command of the operator and based on user preferences. Other workflow tools (e.g., Color Map, Window Level Presets, Overlay Display Schemes) exist within the Viewer to allow for operator customization of display and usage preferences.

    The Viewer is a Microsoft Windows-based software application that runs on "off-the-shelf" computers that meet the specified minimum basic computer configuration for the Viewer and are running Microsoft Windows 10 Operating System (OS). For operator input and control of the software, standard Microsoft Windows-compatible input devices are used, including keyboards of various configurations, multi-button mice, grips, foot pedal controls, the Contour ShuttlePRO (Jog-Dial), etc. The vRad Viewer is not a hardware device; however, a monitor is required to display images and to interface with the software. Image quality may vary depending on the display monitor, and users must utilize a compatible monitor that meets technical specifications required for the tasks intended to be performed by the user.

    AI/ML Overview

    The provided text does not contain detailed information about the acceptance criteria and a specific study proving the device meets those criteria. The document is primarily an FDA 510(k) clearance letter and a summary of the device, its intended use, and its equivalence to a predicate device.

    The document states that "Software verification and validation testing were conducted" and "Cybersecurity testing for the vRad Viewer was conducted." However, it does not provide specifics on:

    • A table of acceptance criteria and reported device performance.
    • Sample size and data provenance of any test set.
    • Number and qualifications of experts for ground truth establishment.
    • Adjudication method for a test set.
    • Results of a multi-reader, multi-case (MRMC) comparative effectiveness study.
    • Results of a standalone (algorithm only) performance study.
    • Type of ground truth used (e.g., pathology, expert consensus, outcomes data).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document primarily focuses on establishing "substantial equivalence" to a predicate device (vRad PACS with Mammography, K162145) based on intended use, principles of operation, and technological characteristics. It mentions that "Any minor differences in technological characteristics do not raise new or different questions regarding its safety and effectiveness when used as labeled. Software and cybersecurity testing demonstrate that the device performs as expected." This indicates that the software was tested to ensure it functions as intended and securely, but the specific performance metrics and studies you've asked for are not detailed in this documentation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162145

    Validate with FDA (Live)

    Date Cleared
    2016-10-25

    (85 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The vRad PACS with Mammography software is used with general purpose computing hardware which meets or exceeds minimum specifications. vRad PACS with Mammography is intended to receive, transmit, store, and display images for clinical purposes and is comprised of three components: Viewer, Storage, and Cache. The vRad PACS Viewer component is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications and networked with vRad PACS Storage component. The vRad PACS Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad PACS Viewer can process medical images from DICOM modalities such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM-compliant modalities. The vRad Storage component is intended to handle the DICOM protocol and store images as DICOM files to a location where they can be read by vRad PACS Cache and transmitted to radiologists' workstations for viewing by the vRad PACS Viewer.

    vRad PACS with Mammography may be used for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM "For Presentation" format and displayed on a monitor that meets technical specifications reviewed and cleared by FDA.

    Device Description

    vRad PACS with Mammography is a device which consists solely of software that allows electronic transmission of radiological patient images from one location to another. The device is capable of accepting, storing, digitally processing, and displaying medical images for the purposes of providing digital diagnostic image interpretation services by trained radiologists on PC workstations. The software provides functions for performing operations related to manipulation, enhancement, compression, and quantification of medical images.

    vRad PACS with Mammography is a modified version of vRad PACS (K090649) that will now allow display of presentation-quality digital mammography images.

    AI/ML Overview

    The provided text describes a PACS system (Picture Archiving and Communications System) with mammography capabilities, not an AI or CAD (Computer-Aided Detection) device. Therefore, the concept of "acceptance criteria" as it relates to statistical performance metrics (like sensitivity, specificity, AUC) for disease detection, or the effect size of AI assistance on human readers, is not directly applicable in the same way it would be for an AI-powered diagnostic tool.

    This 510(k) submission is for a PACS system, which is essentially software for managing and displaying medical images. The "acceptance criteria" here are primarily functional, performance, and safety requirements to demonstrate that the device is substantially equivalent to a predicate device.

    However, I can extract information related to the device's performance claims and the testing conducted, aligning it to the closest relevant categories from your request.

    Here's a breakdown based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't present a formal table of quantitative acceptance criteria with reported performance in the way an AI diagnostic device would (e.g., "Sensitivity > X%"). Instead, the acceptance criteria are implicitly met by successfully passing a comprehensive set of tests and showing substantial equivalence to the predicate device.

    Acceptance Criteria CategoryReported Device Performance / Assessment
    Functional EquivalenceShown through a detailed comparison demonstrating vRad PACS with Mammography has features and functionalities (e.g., display of digital mammography images, on-demand access to database, viewing study lists, decompression, display images, viewing reports, hanging protocol, spine labeling, reference line display, key images identification, link multiple series, measurement tools, annotation tools, standard image manipulation tools) that are equivalent to the predicate device (Synapse Workstation).
    Technical EquivalenceDemonstrated through comparison of technical characteristics such as product availability (software only), operating systems (Windows 7), web browser (Internet Explorer), image and data processing (client side), technology platforms (Windows .Net for client, Windows Server for server), and programming languages (C++, C# .Net for client).
    Safety and Efficacy"vRad PACS with Mammography introduces no new safety or efficacy issues other than those already identified with the cleared vRad PACS (K090649)." Hazard analysis conducted, and appropriate mitigations taken, indicating the device is of moderate concern. Labeling contains instructions, cautions, warnings, and notes for safe and effective use. Concluded to be "as safe and effective as the predicate device."
    Verification & Validation"vRad PACS with Mammography tested successfully with reference to its product requirements, as well as design verification and validation document and traceability matrix document." This included system-level functionality test, component testing, verification testing, integration testing, usability testing, labeling testing, and risk mitigation testing. "Pass/fail criteria were based on the requirements and intended use of the product. Test result results showed that all tests successfully passed." Established performance, functionality, and reliability characteristics.
    Compliance with StandardsThe device is in compliance with NEMA PS 3.1 3.20 (DICOM), IEC 62304 Ed. 1.1 (Medical Device Software Life Cycle), ISO 14971 2nd Ed. (Risk Management), and ISO 16142-1 (Essential principles of safety and performance). The submission also considered various FDA guidance documents.
    Substantial EquivalenceThe overall conclusion is that the device "does not introduce any new significant potential safety risks and is substantially equivalent to the predicate device in terms of performance, safety and effectiveness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set" in the context of a dataset of medical images used to evaluate diagnostic performance (like sensitivity/specificity for disease detection). The testing described is software verification and validation. Therefore, there's no mention of sample size for a clinical test set or data provenance in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. As this is not an AI/CAD diagnostic device, there's no mention of expert-established ground truth for a test set of medical images. The "ground truth" for the software's functionality would be its adherence to design specifications and user requirements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, for the same reasons as above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device, and no MRMC study is mentioned. The device's purpose is to display, store, and transmit images, not to provide diagnostic assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a PACS system, which is a tool for human radiologists. It does not perform diagnostic tasks independently.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the conventional sense of diagnostic performance for a disease. The "ground truth" for this device's testing relates to its correct functioning against design specifications and user requirements, not medical conditions.

    8. The sample size for the training set

    Not applicable. The vRad PACS with Mammography is not a machine learning model that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090649

    Validate with FDA (Live)

    Date Cleared
    2009-03-27

    (16 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The vRad™ PACS software is used with general purpose computing hardware, which meets or exceeds minimum specifications. vRad™ PACS software is intended to receive, transmit, store and display images for clinical purposes. The vRad™ PACS Viewer component is intended for installation on an off-the-shelf PC, meeting or exceeding minimum specifications and networked with vRad™ PACS Storage component. The vRad™ PACS Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad™ PACS Viewer can process medical images from DICOM modalities such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM compliant modalities.

    The vRad™ PACS system should not be used for Mammography primary image diagnosis.

    Device Description

    vRad™ PACS is a device which consists solely of software and allows electronic transmission of radiological patient images from one location to another. The device has the capability to accept, transfer, display, store, and digitally process medical images to trained and qualified radiologists for the purposes of providing digital diaging The software provides functions for performing operations interpretation services. related to image manipulation, enhancement, compression, and quantification of medical images (except mammography images). The software enables the user to display 3D maximum intensity projection (MIP) and Multi-Planar Reformatting (MPR) visualization of study images.

    AI/ML Overview

    The provided text describes a 510(k) summary for the vRad™ Picture Archiving and Communications System (PACS) Software. However, it does not include information about specific acceptance criteria or a study that proves the device meets such criteria.

    The document is a regulatory submission for a device, not a report on its performance against specific acceptance criteria. It focuses on demonstrating substantial equivalence to predicate devices based on design, technical requirements, and intended use, which is the core of a 510(k) submission.

    Therefore, I cannot extract the requested information as it is not present in the provided text. The document does not describe performance metrics, sample sizes for test or training sets, expert qualifications, adjudication methods, or results of comparative effectiveness studies or standalone performance studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1