Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110355

    Validate with FDA (Live)

    Device Name
    VENX REVOLUTION
    Manufacturer
    Date Cleared
    2011-04-22

    (74 days)

    Product Code
    Regulation Number
    878.4800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VENX REVOLUTION™ is a manual surgical instrument intended for use in the destruction of spider veins in the lower extremities of the body.

    Device Description

    The VENX REVOLUTION™ is a sterile, disposable, single patient use surgical instrument for the disruption and destruction of spider veins located in the lower extremities of the body. The device is made available in several sizes to accommodate a range of veins that form in the lower body. The VENX REVOLUTION™ consists of a thermoplastic handle containing a lever actuated rack and gear mechanism for rotating a shaft extending from the handle. At the end of the shaft are stainless steel, needle-like tines sharpened and spaced appropriately to straddle veins located just below the surface of the skin. Once the tines are in place straddling the vein, the levers are actuated causing the shaft to rotate approximately 360° or more. Actuating the levers causes internal springs to compress. Once the levers have rotated the shaft 360° and the tines have disrupted the vein, releasing the levers allows the compressed springs to return the tines to their original position. The device is then withdrawn and re-inserted at another position within the clinician-selected area to repeat the operation on the same vein or on another vein. The procedure is repeated until all veins within the selected area have been treated.

    AI/ML Overview

    The VENX REVOLUTION is a manual surgical instrument for the disruption and destruction of spider veins in the lower extremities. The device incorporates a lever-actuated mechanism to rotate tines a minimum of 360 degrees.

    Here's an analysis of the provided information regarding acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Supporting Performance
    Functional EquivalenceAbility to disrupt and destruct spider veinsPerformed as intended and equivalent to the predicate deviceBench testing under simulated actual use conditions
    Device Integrity/DurabilityTine deformationNot explicitly detailed as a separate criterion, but included in "Testing"Bench testing
    Device FunctionalityCycle lifeNot explicitly detailed as a separate criterion, but included in "Testing"Bench testing
    Mechanical OperationRotating mechanism operationNot explicitly detailed as a separate criterion, but included in "Testing"Bench testing

    2. Sample size used for the test set and the data provenance:

    The provided document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). The testing described is "Bench testing... using simulated actual use conditions," which suggests in-vitro testing rather than human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not mention the use of experts to establish ground truth for the test set. The evaluation was done through "Bench testing," likely involving engineering and technical assessment rather than clinical expert consensus.

    4. Adjudication method for the test set:

    No adjudication method is mentioned as there were no human subjects or expert assessments involved in the "Bench testing."

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No, an MRMC comparative effectiveness study was not done. The testing described is bench testing, not a clinical study involving human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The concept of "standalone performance" typically applies to AI algorithms. Since this is a manual surgical instrument, this question is not directly applicable. The device itself is a "standalone" mechanical tool, and its performance was assessed in isolation through bench testing.

    7. The type of ground truth used:

    Given that the study involved "Bench testing using simulated actual use conditions," the "ground truth" would likely be based on physical measurements and engineering specifications related to the device's ability to disrupt simulated veins and its mechanical integrity, rather than expert consensus, pathology, or outcomes data from human subjects. For example, successful disruption of a simulated vein might have been a measurable outcome against a predefined physical standard.

    8. The sample size for the training set:

    There is no training set as this is a mechanical device undergoing bench testing, not an AI or algorithmic system that requires training data.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device and study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1