Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K252316

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-08-19

    (25 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venclose System (Venclose digiRF Generator with EVSRF Catheter) is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

    Device Description

    The Venclose™ digiRF Generator is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter to be used with the generator (time, temperature, etc.), which is connected via a triaxial catheter connector port. The Venclose™ digiRF Generator is intended to be used with Venclose™ RF Catheter(s) (either the Venclose™ EVSRF Catheter or the Venclose™ Maven Catheter) as a system. The Venclose™ RF System uses resistive radiofrequency ablation via energy delivery to heat the wall of an incompetent vein with temperature-controlled RF energy to cause irreversible luminal occlusion, followed by fibrosis and ultimately resorption of the vein.

    The scope of this 510(k) is only the Venclose™ digiRF Generator as the generator software has been modified. The Venclose™ Maven Catheter is not in the scope of this submission as the changes discussed within this submission are only applicable to the Venclose™ digiRF Generator as used with the Venclose™ EVSRF Catheter. There is no change to the Venclose™ EVSRF Catheter, as previously cleared via K160754.

    AI/ML Overview

    Based on the provided 510(k) Clearance Letter, the device in question is the "Venclose digiRF Generator" and the modifications are related to its software when used with the "Venclose EVSRF Catheter." The clearance letter states that the scope of this 510(k) is only the Venclose digiRF Generator as the generator software has been modified, and there is no change to the Venclose EVSRF Catheter.

    The document explicitly states that the device is an "Electrosurgical Cutting And Coagulation Device And Accessories" and the testing performed was "Software Verification and Validation." There is no mention of a study involving human subjects, interpretation of medical images by experts, or any kind of diagnostic performance evaluation typically seen with AI/ML-based diagnostic devices.

    Therefore, many of the requested criteria regarding acceptance criteria for diagnostic performance, ground truth establishment, expert adjudication, MRMC studies, and training/test set details for AI/ML models are not applicable to this submission. This is a clearance for a software modification to an electrosurgical generator, not a medical imaging AI/ML diagnostic aid.

    Here's the breakdown of what can be gathered from the provided text, addressing the points where information is available and indicating where it is not applicable or not provided.


    Device: Venclose digiRF Generator (VCRFG1) with Venclose EVSRF Catheter (VC10A256F60, VC10A256F100)
    Type of Modification: Software Modification to the Venclose digiRF Generator.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a software modification to an electrosurgical generator, the acceptance criteria are not related to diagnostic performance metrics like sensitivity, specificity, or AUC, but rather to software functionality, safety, and effectiveness. The document states that internal risk assessments procedures were used.

    Acceptance Criterion (Software)Reported Device Performance (Summary)
    Functional Performance (e.g., proper execution of treatment parameters, temperature control, time management, connectivity with catheter)"The results demonstrate that the technological characteristics and performance criteria of the modified Venclose™ digiRF Generator is comparable to the predicate devices and that it performs as safely and as effectively as the legally marketed predicate devices." (Implied successful completion of software verification and validation, meeting defined specifications)
    Safety (e.g., absence of new hazards, proper error handling, electrical safety, EMC compliance)"The results demonstrate that...it performs as safely...as the legally marketed predicate devices." (Implied successful safety testing)
    Effectiveness (e.g., maintaining intended use and performance characteristics)"The results demonstrate that...it performs as...effectively as the legally marketed predicate devices." (Implied successful effectiveness testing, maintaining intended function for endovascular coagulation)
    Software Verification & Validation (e.g., adherence to software requirements, robust and reliable operation)"Software Verification and Validation" was performed. Results demonstrated comparability to predicate devices.

    Note: Specific numerical acceptance values are not detailed in this public 510(k) summary, as they are typically proprietary and part of detailed engineering and software validation reports submitted to the FDA.


    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable in the context of an AI/ML diagnostic performance test set. The "test set" here refers to the software verification and validation activities. The document does not specify exact "sample sizes" (e.g., number of test cases or iterations) for the software testing. Data provenance is also not applicable in the context of clinical data for AI/ML, as the testing relates to engineering and software validation.

    • Software Verification and Validation: This typically involves rigorous testing against defined requirements, including unit testing, integration testing, system testing, and perhaps regression testing. The "sample size" would relate to the number of test cases executed, input parameters varied, and error conditions simulated. Specific numbers are not provided in this summary.
    • Data Provenance: Not applicable as no clinical data for diagnostic performance was used in this clearance for a software modification to an electrosurgical device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth, in the sense of expert annotation of medical data, is not established for an electrosurgical generator's software modification. The "ground truth" for this device's performance would be its adherence to engineering specifications and its ability to safely and effectively deliver RF energy for its intended purpose, as measured by calibrated equipment and functional tests.


    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are for consensus building among human experts for ground truth label generation in diagnostic studies. This process is not part of software verification and validation for an electrosurgical device.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No. An MRMC study is relevant for evaluating the impact of an AI/ML diagnostic device on human reader performance. This 510(k) is for a software modification to an electrosurgical generator, not a diagnostic AI/ML device. Therefore, no MRMC study was performed or reported.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. "Standalone performance" refers to the diagnostic accuracy of an AI/ML algorithm by itself. This device is an electrosurgical generator. Its performance is measured by its ability to generate and deliver RF energy according to specifications, not by its diagnostic capabilities. The software's "performance" was evaluated through verification and validation activities.


    7. The Type of Ground Truth Used

    Not applicable in the AI/ML diagnostic sense. For this device, the "ground truth" for software validation would be derived from:

    • Design Specifications: The documented requirements and expected behavior of the software and the device.
    • Predicate Device Performance: The existing performance characteristics of the previously cleared predicate devices, to which the modified device is being compared for substantial equivalence.
    • Engineering Standards and Measurements: Data from calibrated test equipment, electrical measurements, temperature readings, and time controls.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The software was likely developed using traditional software engineering methodologies.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI/ML training set, there is no ground truth to establish for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250068

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-04

    (25 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venclose System (Venclose digiRF Generator with EVSRF Catheter) is intended for endovascular coaqulation of blood vessels in patients with superficial vein reflux.

    The Venclose Maven System (digiRF Generator and Maven Catheter) is intended for endovascular coagulation of blood vessels in patients with perforator and tributary vein reflux.

    Device Description

    The Venclose™ digiRF Generator is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter to be used with the generator (time, temperature, etc.), which is connected via a triaxial catheter connector port. The Venclose™ digiRF Generator is intended to be used with Venclose™ RF Catheter(s) (either the Venclose™ EVSRF Catheter or the Venclose™ Maven Catheter) as a system. The Venclose™ RF System uses resistive radiofrequency ablation via energy delivery to heat the wall of an incompetent vein with temperature-controlled RF energy to cause irreversible luminal occlusion, followed by fibrosis and ultimately resorption of the vein.

    AI/ML Overview

    This FDA 510(k) summary for the Venclose digiRF Generator (K250068) focuses on substantial equivalence based on software modifications to an existing device.

    Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document does not provide a table of specific acceptance criteria with corresponding performance metrics for the software modifications. It generally states that "the results demonstrate that the technological characteristics and performance criteria of the modified Venclose™ digiRF Generator is comparable to the predicate devices and that it performs as safely and as effectively as the legally marketed predicate devices."

    The acceptance criteria implied are that the modified software continues to allow the device to perform its intended function (endovascular coagulation of blood vessels) safely and effectively, with no adverse effects introduced by the software changes.

    2. Sample size used for the test set and the data provenance:

    The document mentions "Software Verification and Validation" as a performed non-clinical test. However, it does not specify the sample size used for this test set (e.g., number of test cases, number of simulated scenarios). It also does not detail the data provenance (e.g., retrospective or prospective tests, or country of origin of data if real-world data was used, which is unlikely for software verification of this type of device).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For software verification and validation, ground truth might be defined by expected outputs for given inputs, established through engineering specifications and design documents rather than clinical expert consensus. Clinical experts would likely be involved in defining the original device's performance requirements, but not necessarily in the technical software verification changes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    The document does not specify any adjudication method. It's highly probable that for software verification and validation, the "ground truth" (expected behavior or output) is defined by the device's design specifications and tested against these specifications, rather than through a human adjudication process for subjective interpretation as would be common in image analysis AI.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done or mentioned. This type of study is typically performed for AI-powered diagnostic or assistive devices where human interpretation is involved. The Venclose digiRF Generator is an electrosurgical device; the software modification is presumably to its control and operational parameters, not its interpretation of patient data for a human.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The non-clinical test mentioned is "Software Verification and Validation." This type of testing is inherently a standalone (algorithm only) performance evaluation against predefined specifications. The document doesn't explicitly label it as such, but it's the nature of verifying software changes in a medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for "Software Verification and Validation" tests would typically be the design specifications and expected behavior of the software as defined by the engineering and quality teams. This is not clinical ground truth like pathology or outcomes data, but rather technical correctness and adherence to specified functional and non-functional requirements.

    8. The sample size for the training set:

    Not applicable and not provided. This device is not an AI/ML device that requires training data in the conventional sense. The "software modifications" refer to changes in the control software of the electrosurgical generator, not a learning algorithm.

    9. How the ground truth for the training set was established:

    Not applicable and not provided, as there is no training set for this type of device modification.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211806

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-03

    (53 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venclose Maven System (digiRF Generator and Maven Catheter) is intended for endovascular coagulation of blood vessels in patients with perforator and tributary vein reflux.

    Device Description

    Venclose Maven System consists of two (2) main components:

      1. Maven Catheter with integrated cable connector and,
    • digiRF Generator which delivers radiofrequency (RF) energy to the Maven catheter. 2)
      The Maven catheter is a sterile single-use disposable medical device for endovenous radiofrequency ablation. The primary components of the catheter include the shaft, handle and integrated connector cable. The catheter shaft is 6 French (6F = 2.0 mm in diameter) in profile, with an insertable length of 40 cm and a 0.5 cm heating coil. The catheter is energized by the digiRF Generator which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter.
      The Venclose Maven System (digiRF Generator and Maven Catheter) uses resistive radiofrequency ablation (via energy delivered to heat the wall of an incompetent vein) with temperature-controlled RF energy, and an already widely accepted procedure to cause irreversible luminal occlusion. This is followed by fibrosis and ultimately resorption of the vein.
      The Venclose Maven System (digiRF Generator and Maven Catheter) supports an optional foot pedal.
    AI/ML Overview

    The acceptance criteria and device performance are described below, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceStudy Type
    Safe and effective for treatment of incompetent perforator veins without significant adverse outcomesObserved to be safe and effective for treatment of incompetent perforator veins.Clinical Study
    Occlusion rate at 30-day follow-up80%Clinical Study
    Reflux-free rate at 30-day follow-up93%Clinical Study

    Note: The text does not explicitly state pre-defined acceptance thresholds for occlusion and reflux-free rates; it only reports the achieved performance.

    Study Information

    The information provided relates to a clinical study.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 27 subjects
    • Data Provenance: Prospective, single-investigator study. The country of origin is not explicitly stated, but the submission is to the U.S. FDA by a company based in San Jose, California, suggesting a U.S. study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the text. The study is described as a "single-investigator study," but it does not specify how ground truth (e.g., incompetence of perforator veins, occlusion, reflux status) was established or adjudicated, nor does it detail the qualifications of any experts involved in assessing these outcomes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided in the text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This device is for endovascular coagulation, not an imaging-based diagnostic AI device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable as the device is a medical instrument (catheter and generator) for a surgical procedure, not an algorithm or AI system.

    7. The type of ground truth used:

    • The ground truth for effectiveness (occlusion rate, reflux-free rate) was established through clinical assessment during the 30-day follow-up, likely using imaging techniques (e.g., ultrasound) to determine vein patency and reflux. The basis for determining initial "incompetent perforator veins" would also be clinical and imaging findings. For safety, ground truth was based on the absence of "significant adverse outcomes."

    8. The sample size for the training set:

    • This question is not applicable as the device is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • This question is not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160754

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-09

    (175 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venclose™ Radiofrequency System (Generator and Catheters) is intended for use in endovascular coagulation of blood vessels in patients with superficial vein reflux.

    Device Description

    The Venclose RF System is composed of two basic components: an energy console or generator (digiRF) and catheters (EVSRF) designed specifically to be used with the generator. The Venclose RF System uses resistive radiofrequency ablation via energy delivery to heat the wall of an incompetent vein with temperature-controlled RF energy to cause irreversible luminal occlusion, followed by fibrosis and ultimately resorption of the vein. The blood then naturally reroutes to healthy veins.

    The technique involves percutaneous access and insertion of the EVSRF Catheter into the great saphenous vein (GSV) or other superficial vein under ultrasound guidance, injection of local anesthesia, and thermal energy from a radiofrequency generator (digiRF) applied into the target vein. As the EVSRF Catheter is withdrawn stepwise down the treated length of the vein, thermal damage is inflicted upon the venous endothelium and through the vein wall, resulting in contraction and ultimately destruction of the vessel. The procedure can be performed in an outpatient setting, without the need for general anesthesia, allowing for a walkin/walk-out procedure with minimal postoperative recovery time.

    AI/ML Overview

    I am sorry, but the provided text does not contain detailed acceptance criteria or a specific study that proves the device meets those criteria in the format requested. The document is a 510(k) summary for the Venclose™ Radiofrequency System, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance metrics against specific acceptance criteria.

    While it mentions non-clinical bench testing and pre-clinical animal testing, it does not provide:

    • A table of acceptance criteria with reported device performance.
    • Sample sizes for a test set or data provenance for specific performance evaluation.
    • Number of experts or their qualifications for establishing ground truth.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness study results.
    • Standalone (algorithm-only) performance.
    • Specific ground truth types used for detailed performance evaluation.
    • Sample size for a training set or how its ground truth was established, as this is not an AI/algorithm-based device in the context of typical AI performance studies.

    The document primarily focuses on:

    • Device Description: What the Venclose RF System is and how it works.
    • Materials: Composition of the console and catheters.
    • Comparative Summary of Technological Characteristics: A table comparing the subject device with predicate devices in terms of features like energy/frequency, maximum output power, dimensions, etc., to establish similarity.
    • Non-Clinical Testing: A general statement that testing was conducted to evaluate conformance to product specifications and applicable standards (e.g., ISO 10555-1, IEC 60601-1). It lists types of bench tests (Catheter Simulation, Temperature, Pressure, Tensile, Fluid Test) but does not provide specific acceptance values or results for these tests.
    • Pre-clinical testing (Animal Study): Describes the purpose (demonstrate proper functionality), the model used, and objectives (insertion, navigation, reaching/holding temperature, vein diameter reduction). It concludes that "All of the objectives were met resulting in a satisfactory study conclusion," but again, it doesn't provide quantitative acceptance criteria or specific performance data from this study.
    • Rationale for Substantial Equivalence: Concludes that the non-clinical and animal testing demonstrates the Venclose System is as safe, effective, and performs at least as safely and effectively as the predicate devices.

    In summary, the provided text establishes substantial equivalence for regulatory purposes but does not offer the detailed performance data against explicit acceptance criteria that your request asks for, especially in the context of an AI-driven device performance study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1