510(k) Data Aggregation

    K Number
    K252304

    Validate with FDA (Live)

    Date Cleared
    2026-03-18

    (237 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    syngo.via RT Image Suite VC10 is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).

    It provides tools to view existing contours, create, edit, modify, copy contours of regions of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create simple geometric treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System.

    The software combines the following digital image processing and visualization tools (not all of them might be available at each customer):

    • Multimodality viewing and contouring of anatomical, functional, and multi parametric images such as but not limited to CT, PET, PET/CT, MRI, Linac CBCT images
    • Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
    • Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
    • Automated Contouring on CT and MR images, including known (diagnosed) brain metastases
    • Creation of contours on images supported by the application without prior assignment of a planning CT or planning MR
    • Manual and semi-automatic registration using rigid and deformable registration
    • Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points
    • Supports multi-modality image fusion
    • Visualization and contouring of moving tumors and organs
    • Management of points of interest including but not limited to the isocenter
    • Creation of simple geometric treatment plans
    • Generation of a synthetic CT based on pre-defined MR acquisitions
    Device Description

    syngo.via RT Image Suite VC10 is an image analysis and radiation therapy preparation software that provides multimodality image viewing, registration, segmentation, synthetic CT generation, and patient marking workflows. Within the Medical device syngo.via RT Image Suite VC10 the name 'CT Sim&Go' is used. CT Sim&GO is the name used for syngo.via RT Image Suite VC10 when it is deployed in the CT scanner workflow. CT Sim&GO is not a device of its own. CT Sim&GO is a subset of syngo.via RT Image Suite VC10 functionalities that can be accessed from the CT scanner workplace. More precisely it comprises the Patient Marking and Beam Placement modules of syngo.via RT Image Suite VC10. The current submission includes modifications affecting the following functionalities:

    MR Autocontouring: The AI‑based MR autocontouring functionality has been introduced to include segmentation of previously diagnosed brain metastases, in addition to MR‑based segmentation of brain OARs and male pelvis OARs. The metastasis‑specific model is restricted to identification of metastases already diagnosed by a clinician, and the software does not provide diagnostic capability or detect new or unknown lesions. These contours remain fully editable by the user.

    CT Autocontouring: syngo.via RT Image Suite VC10 includes 29 new organs and structures for deep‑learning–based CT autocontouring. The underlying DL architecture is unchanged; however, additional segmentation guidelines and organ‑coverage expansion were incorporated.

    Improvements to Isocenter Definition & Patient Marking: The patient‑marking workflow includes semi‑automated isocenter estimation for breast (originally cleared under K192065) and vertebral regions (originally cleared under K220783). New capability to send the coordinates from a movable laser positioning system to the syngo.via RT Image Suite VC10, enabling a workflow of setting isocenter directly on the skin and transferring it to the planning images.

    Synthetic CT: The synthetic CT feature was updated to include a new 3D deep‑learning–based algorithm for brain and pelvis, replacing the prior 2D model and improving HU accuracy and geometric fidelity.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the syngo.via RT Image Suite VC10 device, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance for syngo.via RT Image Suite VC10

    1. Table of Acceptance Criteria and Reported Device Performance

    The device includes several AI-based auto-contouring functionalities and a synthetic CT feature. The acceptance criteria and reported performance vary by feature.

    1.1 CT Auto-Contouring (Existing Organs)

    Structure GroupAcceptance Criterion (DICE Score)Reported Device Performance (Mean DICE)
    Head and NeckStatistical non-inferiority of the Dice score compared with the reference predicate (lower 95th percentile confidence bound > mean reference performance - 10% margin).0.37 (Optic Chiasm) - 0.99 (Body)
    Thorax and AbdomenStatistical non-inferiority of the Dice score compared with the reference predicate (lower 95th percentile confidence bound > mean reference performance - 10% margin).0.42 (LAD) - 0.96 (Liver)
    PelvisStatistical non-inferiority of the Dice score compared with the reference predicate (lower 95th percentile confidence bound > mean reference performance - 10% margin).0.68 (LN Presacral) - 0.95 (Bladder, Proximal Femur Left)

    1.2 CT Auto-Contouring (New Organs)

    | Structure Group | Acceptance Criteria | Reported Device Performance (Mean DICE / Mean ASSD) |
    | :------------------------ | :------------------------------- |
    | New Head and Neck OARs | 1. Statistical non-inferiority of the Dice score compared with a reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with a reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings). | DICE: 0.68 (Lacrimal Gland Right) - 0.94 (Humeral Head Right) ASSD (mm): 0.7 (Humeral Head Left) - 1.15 (Lacrimal Gland Right) |
    | New Thorax OARs | 1. Statistical non-inferiority of the Dice score compared with a reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with a reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings). | DICE: 0.28 (CA Left Circumflex) - 0.75 (N2 Station 3A) ASSD (mm): 1.19 (N1 Station 10 Left) - 5.17 (N2 Station 9 Left) |
    | New Pelvis OARs | 1. Statistical non-inferiority of the Dice score compared with a reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with a reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings). | DICE: 0.92 (Sacrum) - 0.95 (Femoral Head Left, Hip Bone Right) ASSD (mm): 0.4 (Hip Bone Right) - 2.92 (Bowel Bag) |

    1.3 CT Auto-Contouring (Subgroup Analysis for Photon-Counting CT)

    Body RegionAcceptance Criterion (Mean DICE)Reported Device Performance (Mean DICE)
    Image ContrastMean Dice >= reference mean Dice (between reference and variable image impression).0.90 (Head and Neck) - 0.97 (Abdomen, Body)
    Image ResolutionMean Dice >= reference mean Dice (between reference and variable image impression).0.93 (Head and Neck) - 0.99 (Abdomen, Body)

    1.4 MR Auto-Contouring (Brain Metastasis)

    MetricAcceptance CriterionReported Device Performance (Mean)
    Lesionwise DICEStatistical non-inferiority compared with the reference device (lower 95th percentile confidence bound > mean reference dice coefficient - 10% margin).0.74
    Lesionwise SensitivityStatistical non-inferiority compared with the reference device (lower 95th percentile confidence bound > mean reference sensitivity - 10% margin).92.5%

    1.5 MR Auto-Contouring (Brain and Pelvis OARs)

    Structure GroupAcceptance CriteriaReported Device Performance (Mean DICE / Mean ASSD)
    Brain OARs1. Statistical non-inferiority of the DICE score compared with the reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with the reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings).DICE: 0.49 (Cochlea Right) - 0.93 (Brainstem) ASSD (mm): 0.36 (Eye Right) - 1.95 (Lacrimal Gland Left)
    Pelvis OARs (T1)1. Statistical non-inferiority of the DICE score compared with the reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with the reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings).DICE: 0.93 (Femur head Left) - 0.98 (Body) ASSD (mm): 0.96 (Femur head Right) - 1.82 (Body)
    Pelvis OARs (T2)1. Statistical non-inferiority of the DICE score compared with the reference device (lower 95th percentile confidence bound > mean reference performance - 10% margin). 2. Statistical non-inferiority of the ASSD score compared with the reference device (upper 95th percentile confidence bound < mean reference performance + Std.Dev). 3. Average user evaluation of 3 or higher (on a 4-point scale for time savings).DICE: 0.72 (Seminal Vesicles) - 0.9 (Bladder) ASSD (mm): 0.91 (Penile Bulb) - 2.47 (Anus)

    1.6 Synthetic CT (Pelvis and Brain)

    Acceptance criteria for Geometric Fidelity and HU Accuracy are mentioned but not specifically detailed in the provided text. The study states "The testing ensures the quantitative performance of the resulting synthetic CT. Analysis was performed on geometric fidelity and HU accuracy."

    1.7 Semi-automated isocenter estimates of vertebrae

    Acceptance criterion is "percentage of cases which need corrections" but the specific threshold is not detailed. The study mentions "Analysis was performed on percentage of cases which need corrections." and that it was cleared under K220450.


    2. Sample Size Used for the Test Set and Data Provenance

    Feature/ModelNumber of Subjects (Test Set)Data Provenance (Country of Origin, Retrospective/Prospective)
    CT Auto-Contouring469Europe (IT, PT, CH, UK, NL, DE), North America (US, CA), South America (BR), Australia, Asia (JP, IN); Retrospective (implied, as it's from existing data)
    MR Brain Metastasis Model30USA, EU; Retrospective (implied)
    MR Brain OAR Model81 (# Test data sets)USA, EU; Retrospective (implied)
    MR Pelvis OAR Model153North America (US), Europe (Germany, Romania, France, Switzerland, Spain), Australia; Retrospective (implied)
    Synthetic CT Model51US, Europe; Retrospective (implied)
    Semi-automated isocenter estimation/laser-based (breast)10Not specified beyond general statement: "All test datasets were independent of training datasets."
    Semi-automated isocenter estimation/laser-based (vertebra)10Not specified beyond general statement: "All test datasets were independent of training datasets."
    Subgroup analysis for photon-counting CT199 oncological patient casesNot explicitly stated but mentions "various CT image contrasts... and image resolutions" and "oncological patient cases with and without metal implants"

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not explicitly stated as a single number. The document mentions "an expert team" for most ground truth annotations.
    • Qualifications: "expert clinicians using well-established international contouring guidelines" and "rigorous independent quality assurance reviews." For spine landmarks, "an expert team" was used. Specific specialties (e.g., radiologist, radiation oncologist) or years of experience are not provided.

    4. Adjudication Method for the Test Set

    The document consistently states that ground truth contours created by expert teams underwent "rigorous independent quality assessment" or "rigorous independent quality assurance reviews." This implies an adjudication process where an independent expert or team reviewed and likely corrected or confirmed the initial expert annotations. However, the specific method (e.g., 2+1, 3+1, etc.) is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size

    • MRMC Study: No, a formal MRMC comparative effectiveness study comparing human readers with AI vs. without AI assistance was not reported in this summary.
      • The user evaluation for new organs in auto-contouring (CT and MR OARs) included a "four-point scale to evaluate each contour in the context of time savings compared to contouring from scratch," which assesses user perception of usefulness but is not an MRMC comparative effectiveness study.
    • Effect Size: Therefore, no effect size for human reader improvement with AI assistance is provided.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, extensive standalone algorithm performance testing was done for all AI-based features. The metrics reported (DICE, ASSD, Hausdorff Distance, Sensitivity, False Positive Rate, HU accuracy, geometric fidelity) are all measures of the algorithm's performance without human intervention after the contour is generated. The contours are stated to be "fully editable by the user," but the reported metrics reflect the raw output of the algorithms.

    7. The Type of Ground Truth Used

    • CT auto-contouring, MR brain metastasis, MR brain OAR, MR pelvis OAR: "Manual ground-truth segmentations were annotated by an expert team based on well accepted international contouring guidelines, followed by a rigorous independent quality assessment." This indicates expert consensus (or at least expert-generated and quality-assured) ground truth.
    • Synthetic CT: Ground truth would likely be the actual (e.g., diagnostic-quality) CT images used for comparison, with analysis performed on "geometric fidelity and HU accuracy."
    • Semi-automated isocenter estimates of vertebrae: "Manual ground-truth of spine landmarks were annotated by an expert team, followed by a rigorous independent quality assessment." This is also expert consensus.

    8. The Sample Size for the Training Set

    The document mentions "Training datasets consisted of curated, multicenter CT and MR image collections with expert-annotated reference standards and standardized preprocessing." However, the exact sample size for the training set is not explicitly provided in this summary for any of the features. It only states that "All test datasets were independent of training datasets."

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set was established similarly to the test set: "curated, multicenter CT and MR image collections with expert-annotated reference standards and standardized preprocessing." This implies expert annotation, consistent with how the test set ground truth was created.

    Ask a Question

    Ask a specific question about this device

    K Number
    K252977

    Validate with FDA (Live)

    Date Cleared
    2026-01-21

    (126 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    0 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Halcyon and Ethos radiotherapy system are indicated for the delivery of stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation is indicated for adults and pediatric patients.

    Halcyon and Ethos radiotherapy system with the HyperSight imaging feature produce kV CBCT anatomical images that can be used in the simulation and planning of radiation therapy.

    Device Description

    Halcyon and Ethos Radiotherapy System are single energy medical linear accelerators (linacs) designed to deliver Image Guided Radiation Therapy and radiosurgery, using Intensity Modulated and Volumetric Modulated Arc Therapy techniques. They consist of the accelerator and patient support within a radiation shielded treatment room and a control console outside the treatment room.

    An electron gun generates electrons which are accelerated by radio frequency (RF) power from a magnetron. The electrons strike a tungsten target producing photons (X-rays) for treatment and MV Imaging. The photons produced by the target are monitored and controlled by a pressurized ion chamber.

    A beam collimation subsystem consisting of a primary and secondary collimator and two stacked multileaf collimators (MLCs) shapes the photon beam to define the treatment area.

    X-Ray images of the patient are used by the treater to verify the correct treatment location. MV Imaging uses the treatment beam and a flat panel imager whereas kV imaging uses a high-capacity kV X-ray tube, a kV collimation system with full fan bowtie filter with movable y-blades to define the imaging beam size and to capture the image, a kV imager.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252919

    Validate with FDA (Live)

    Device Name
    IDENTIFY (5.0)
    Date Cleared
    2025-12-16

    (95 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    18 - 999
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IDENTIFY is indicated for adult patients undergoing radiotherapy treatment simulation and/or delivery. IDENTIFY is indicated for positioning of patients, and for monitoring patient motion including respiratory patterns. It allows for data output to radiotherapy devices to synchronize image acquisition or treatment delivery with the acquired motion information.

    Device Description

    IDENTIFY uses surface guidance technology to monitor patient motion during radiotherapy treatment simulation and delivery. Its high precision SGRT cameras support:

    • Positioning of the patient for treatment delivery
    • Monitoring of the patient position during treatment delivery
    • Respiratory motion management during simulation and treatment delivery
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252884

    Validate with FDA (Live)

    Date Cleared
    2025-11-19

    (70 days)

    Product Code
    Regulation Number
    892.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Guided Aarhus Applicator Set is indicated for use for cancer treatment of the vagina, vaginal stump, cervix, uterus, and endometrium using HDR brachytherapy.

    Device Description

    The Guided Aarhus Applicator Set is an applicator for intracavitary/interstitial brachytherapy.

    Brachytherapy is a form of radiotherapy using Gamma rays from a radioactive source placed at locations close to or within a tumor or other treatment area to a predefined treatment plan. The treatment plan defines the positions and times for the source to ensure the correct dose for the treatment area. The applicator acts to guide the radioactive source to the correct location or locations for treatment.

    Brachytherapy is a form of radiotherapy using Gamma rays from a radioactive source placed at locations close to or within a tumor or other treatment area to a predefined treatment plan. The treatment plan defines the positions and times for the source to ensure the correct dose for the treatment area. The applicator acts to guide the radioactive source to the correct location or locations for treatment.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252948

    Validate with FDA (Live)

    Date Cleared
    2025-10-07

    (21 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam radiotherapy system can produce CBCT images that can be used in image guided radiation therapy, and the simulation and planning for adaptive radiation therapy.

    The TrueBeam, TrueBeam STx and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, hemangiomas, and medically refractory essential tremor (indicated for adults only)), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis.

    VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

    VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis.

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription.

    The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250099

    Validate with FDA (Live)

    Device Name
    Mobius3D (4.1)
    Date Cleared
    2025-05-16

    (122 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).

    Mobius3D is not a treatment planning system. It is to be used only by trained radiation oncology personnel as a quality assurance tool.

    Device Description

    Mobius3D is a software product used within a radiation therapy clinic for quality assurance and treatment plan verification. It is important to note that while Mobius3D operates in the field of radiation therapy, it is neither a radiation delivery device (e.g. a linear accelerator), nor is it a Treatment Planning System (TPS). Mobius3D cannot design or transmit instructions to a delivery device, nor does it control any other medical device. Mobius3D is an analysis tool meant solely for quality assurance (QA) purposes when used by trained medical professionals. Being a software only QA tool, Mobius3D never comes into contact with patients.

    AI/ML Overview

    It appears there's a misunderstanding based on the provided document. The request asks for acceptance criteria and a study that proves the device meets those criteria, including specifics like sample sizes, expert qualifications, and ground truth establishment.

    However, the provided FDA 510(k) clearance letter for Mobius3D (4.1) does not contain the detailed performance study results that would prove the device meets specific acceptance criteria.

    The 510(k) summary (pages 5-7) primarily discusses:

    • Device Description and Intended Use: What Mobius3D is and what it's used for (QA, treatment plan verification, patient alignment).
    • Comparison to Predicate Device: How Mobius3D 4.1 differs from 4.0.
    • Summary of Performance Testing (Non-Clinical):
      • Mentions software verification and validation, including unit, integration, and end-to-end testing.
      • Highlights MLC Modelling Accuracy testing comparing different Mobius3D versions, measurements, and a Treatment Planning System (Eclipse TPS 16.1).
      • States that "studies and reviews have been performed to assess the accuracy of newly introduced features and modifications" for Rapid Arc Dynamic Support and MLC Tongue and Groove Modelling.
      • Notes conformance to cybersecurity and interoperability requirements.
      • Crucially, it explicitly states: "No animal studies or clinical tests have been included in this pre-market submission." This means there isn't a human-in-the-loop study or a study directly demonstrating clinical performance against ground truth in a clinical setting.
    • Use of Consensus Standards: A list of standards the device's design and evaluation conform to.
    • Determination of Substantial Equivalence: Varian's conclusion that the device is substantially equivalent to the predicate.

    Therefore, many of the specific details requested (Table of acceptance criteria, sample sizes for test sets, number/qualifications of experts for ground truth, adjudication methods, MRMC study, standalone performance, type of ground truth, training set sample size/ground truth establishment) are NOT present in this 510(k) clearance letter.

    The letter focuses on the regulatory submission process and the FDA's determination of substantial equivalence based on the provided non-clinical testing and comparison to a predicate device. It doesn't typically include the full, detailed study reports with precise performance metrics and ground truth methodologies. Such details would typically be found in the more extensive technical documentation submitted by the manufacturer to the FDA, but they are summarized at a high level in the public 510(k) summary.

    In summary, based only on the provided text, I cannot provide the detailed information requested regarding the specific acceptance criteria and the study that proves the device meets those criteria in the format you've requested. The document indicates non-clinical software testing and accuracy assessments were performed but does not provide the specific metrics or study design details for clinical performance proof.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250299

    Validate with FDA (Live)

    Date Cleared
    2025-05-09

    (98 days)

    Product Code
    Regulation Number
    892.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CT/MR Tandem and Ovoid Set is indicated for use for cancer treatment of the uterus, cervix, paracervix, endometrium, and vagina using HDR Brachytherapy.

    Device Description

    {{overview}}

    AI/ML Overview

    This FDA clearance letter describes a new device, the "CT/MR Tandem and Ovoid Set," but it does not describe a study involving AI or software performance evaluation. The document focuses on the substantial equivalence of this physical medical device (an applicator for brachytherapy) to a predicate device, primarily through mechanical, material, and compatibility testing.

    Therefore, I cannot extract information related to acceptance criteria for AI model performance or a study proving that an AI device meets acceptance criteria, as the provided text doesn't contain such a study.

    The document discusses:

    • Biocompatibility Testing: Ensuring the materials are safe for patient contact.
    • Magnetic Resonance Testing (MR): Confirming the device is MR safe/compatible.
    • Cleaning, Disinfection, and Sterilization Testing: Verifying the device can be properly cleaned and sterilized for repeated use.
    • Human Factors Validation Study: To ensure the device is user-friendly and performs as intended for its users and environment.
    • Mechanical and Acoustic Testing: To demonstrate the device performs as intended and meets design specifications.

    These tests are for the physical device's safety and effectiveness, not for an AI algorithm's diagnostic or predictive performance.

    Therefore, I cannot fill in the requested table and details for AI-related performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250289

    Validate with FDA (Live)

    Date Cleared
    2025-04-25

    (84 days)

    Product Code
    Regulation Number
    892.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intracavitary/Interstitial System is intended for cancer treatment of the uterus, cervix, and vagina using HDR or PDR brachytherapy.

    Device Description

    The Intracavitary/Interstitial System is a device with two (2), hollow, open-ended catheters which connect to applicator sets with ovoids, ring caps or templates on the distal end.

    The guide tubes extend outside of the patient. An interstitial needle is guided through the guide tube from outside the patient to an interstitial hole in the applicator. The proximal end of the guide tube contains a fixation mechanism for fixing the needle insertion depth between needle insertion and removal of the implant.

    The Intracavitary/Interstitial System is intended to be used with the following required accessories: Plastic Needles (K120341) and Leak Stop Channel Markers (K152018).

    AI/ML Overview

    I am sorry, but based on the provided FDA 510(k) clearance letter for the Intracavitary/Interstitial System (K250289), there is no information provided regarding acceptance criteria for device performance, nor is there a study described that proves the device meets such criteria in terms of clinical or algorithmic performance.

    The document primarily focuses on:

    • Administrative details of the 510(k) clearance: Device name, regulation number, product code, date of clearance, contact information.
    • Device description and intended use: What the device is and what it's used for (brachytherapy for uterus, cervix, vagina).
    • Comparison to a predicate device: Highlighting similarities in intended use, technological characteristics, materials, anatomical sites, MR compatibility, and use environment.
    • Non-clinical performance testing: This section details engineering and safety tests like biocompatibility, cleaning/disinfection/sterilization, human factors, mechanical/acoustic, MR compatibility (physical properties under MR, not clinical performance), and electromagnetic compatibility (EMC). It also lists various consensus standards followed during design and evaluation.

    Specifically, the document does NOT contain any of the following information requested in your prompt:

    1. A table of acceptance criteria and reported device performance related to a clinical outcome or algorithmic performance metric. The "performance data" mentioned is exclusively for non-clinical, engineering, and safety assessments.
    2. Sample size used for a test set or data provenance for any clinical performance evaluation. No clinical test set is described.
    3. Number of experts, qualifications, or adjudication methods for establishing ground truth. This is relevant for studies involving human interpretation or AI performance assessment, which are not detailed here.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study information or effect sizes.
    5. Standalone (algorithm only) performance.
    6. Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for a clinical or AI performance study.
    7. Sample size for a training set or how ground truth was established for a training set. These are relevant for AI/ML device submissions, but this document describes a physical medical device, not an AI algorithm for diagnosis or treatment planning.

    Conclusion:

    The provided 510(k) letter demonstrates that the Intracavitary/Interstitial System (a physical medical device) has undergone various engineering, safety, and compatibility tests to show it is as safe and effective as its predicate. However, it does not detail any clinical performance study, nor does it refer to an AI/ML component where acceptance criteria for diagnostic/prognostic accuracy or human-in-the-loop performance would be relevant. The performance data presented are purely non-clinical validation steps for a hardware device.

    Therefore, I cannot populate the requested tables or provide details on the study proving the device meets performance criteria in the way you've outlined, as that information is not present in the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243939

    Validate with FDA (Live)

    Date Cleared
    2025-03-03

    (73 days)

    Product Code
    Regulation Number
    892.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Titanium Flexible Geometry FSD Applicator Set with the optional interstitial ovoids is intended for cancer treatment of the uterus, cervix, paracervix, endometrium and vaqina, with intracavitary or with a combination of intracavitary and interstitial HDR brachytherapy.

    Device Description

    The Titanium Flexible Geometry FSD Applicator Set with the optional interstitial ovoids is intended for cancer treatment of the uterus, cervix, paracervix, endometrium and vagina, with intracavitary or with a combination of intracavitary and interstitial HDR brachytherapy.

    The device does not contain or consist of software/firmware. The device does not contain any biologics or drug components. The device has patient-contacting materials. The device is designed for repeated use. No parts of the system are provided sterile. The device can be steam sterilized with common parameters using pre-vacuum sterilization.

    The intended patient population includes adult female patients whose cancer can be treated using HDR brachytherapy, as determined by the prescribing physician. The subject device is intended to be used in a healthcare or treatment facility by trained and qualified personnel.

    The key performance characteristics of this applicator set are as follows:

    • . The ability to customize the relative depth of the tandem and ovoids, as well as adapt to different anatomies with three tandem angles (15°, 30°, 45°) and a movable cervical stop.
    • 3.5 mm diameter intrauterine tandems allow for easy insertion into the cervical channel helping provide maximum patient comfort.
    • The tandems and the pivot assembly are made from strong, lightweight titanium.
    • Four different ovoid sizes to allow for varying anatomy.
    • . Channel marking on the intrauterine tandems and colpostat tandems for easy identification.
    • . Ability to image the patient using CT imaging.
    • Ability to image the patient using MR imaging.
    • Optional interstitial ovoids for more complex cases with guide tubes to allow for the needles to be preloaded outside of the patient and locked into place.
    • . Compatible with cervical sleeves to allow easy multiple fraction and help provide maximum patient comfort and eliminate the risk of perforation.
    • Can be manually or machine cleaned, disinfected and steam sterilizable.
    • Can be securely connected to the BRAVOS Afterloader System and GammaMedeplus iX series afterloader.
    • Suitable for patient contact for a period of less than 30 days
    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Titanium Flexible Geometry FSD Applicator Set (GM11013400). It does not describe an AI/ML powered device. The information provided focuses on the substantial equivalence to a predicate device through non-clinical performance testing. Therefore, many of the requested categories related to AI/ML device studies (such as MRMC studies, standalone algorithm performance, AI training/test sets, expert ground truth establishment, etc.) are not applicable to this submission.

    Here is the information that can be extracted relevant to the performance and acceptance criteria for this medical device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Tests ConductedReported Device Performance (Summary)
    BiocompatibilityISO 10993-1 and applicable partsCompatibility of skin-contact component material meets biocompatibility requirements.
    Magnetic Resonance (MR) CompatibilityASTM F2503-23 (Marking)MR Conditional; complies with standards for marking, RF induced heating, magnetically induced displacement force, torque, and image artifacts.
    ASTM F2182-19e2 (RF Heating)
    ASTM F2052-21 (Displacement Force)
    ASTM F2213-17 (Torque)
    ASTM F2119-07 (2013) (Image Artifacts)
    Cleaning, Disinfection, SterilizationDemonstrated effectiveness for components and specified number of usesTesting conducted to demonstrate effective sterilization and multi-use capability.
    Human Factors/UsabilityIEC 62366 (Human Factors Validation Study)Device performs well as intended for intended users, uses, and use environments.
    Mechanical & Acoustic TestingFormal Design Verification & Validation Testing (21 CFR §820, ISO 13485, ISO 14971)Device performs as intended; conformance to applicable requirements and hazard safeguards.
    Safety & EfficacyConformance to FDA-recognized consensus standards (various ISO and EN ISO standards, e.g., ISO 13485, ISO 14971, ISO 10993, IEC 60601)Demonstrated through verification and validation, meeting safety and performance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable. This submission concerns a physical medical device (applicator set), not an AI/ML model for data-driven analysis. Performance testing likely involved physical prototypes or units, but specific "sample sizes" in the context of data analysis are not presented.
    • Data Provenance: Not applicable in the context of clinical data used for AI/ML development. The testing described is non-clinical (biocompatibility, MR compatibility, cleaning/sterilization, mechanical, human factors).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. Ground truth as typically defined for AI/ML (e.g., expert labels on medical images) is not relevant for the type of device and testing described. The "ground truth" for this device's performance is established by objective engineering and safety standards.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication is typically for resolving discrepancies in expert labels for AI/ML ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI-powered device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an AI-powered device.

    7. The Type of Ground Truth Used

    • The ground truth for this device's performance is based on objective engineering standards, regulatory requirements, and validated test methods. This includes:
      • Compliance with ISO 10993 for biocompatibility.
      • Compliance with ASTM standards for MR compatibility.
      • Demonstrated effectiveness for cleaning, disinfection, and sterilization.
      • Usability verification through human factors studies (IEC 62366).
      • Verification and validation against design specifications and risk management (21 CFR §820, ISO 13485, ISO 14971).

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242957

    Validate with FDA (Live)

    Device Name
    Identify (4.0)
    Date Cleared
    2025-02-07

    (135 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IDENTIFY is indicated for adult patients undergoing radiotherapy treatment simulation and/or delivery. IDENTIFY is indicated for positioning of patients, and for monitoring patient motion including respiratory patterns. It allows for data output to radiotherapy devices to synchronize image acquisition or treatment delivery with the acquired motion information.

    Device Description

    IDENTIFY is a system for motion monitoring during radiotherapy treatment simulation and delivery. It incorporates patient safety, quality, and workflow efficiency. Its high precision SGRT cameras support proper patient positioning and enable to monitor the patient's respiratory motion and to detect intra-fraction patient position changes during the treatment.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria or a dedicated study proving the device meets these criteria with reported performance metrics. The document is primarily a 510(k) summary for the IDENTIFY (4.0) device, outlining its intended use, description, and non-clinical testing for substantial equivalence to a predicate device.

    Therefore, I cannot fulfill all parts of your request. I can, however, extract information about what non-clinical testing was conducted, which suggests certain underlying acceptance criteria related to safety, effectiveness, and performance against standards.

    Here's what can be inferred and stated based on the provided text:

    Acceptance Criteria and Study Information (Based on Inferred Information from Non-Clinical Testing Section):

    The document does not present a table of explicit acceptance criteria with reported device performance in the format requested. Instead, it describes general compliance with standards and internal testing to ensure safety, effectiveness, and performance.

    Inferred Acceptance Criteria from Non-Clinical Testing:

    While not explicitly listed as a table of "acceptance criteria," the non-clinical testing section implies the device needed to meet the following:

    • Conformance to Applicable Requirements Specifications: The device must meet its defined functional and performance specifications.
    • Hazard Safeguards Functioning Properly: Safety mechanisms must work as intended.
    • Software Compliance: Adherence to FDA's "Content of Premarket Submissions for Device Software Functions" guidance, specifically for a "major" level of concern.
    • Electrical Safety: Compliance with IEC 60601-1 standards.
    • Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1-2 standard.
    • Quality Management System: Adherence to ISO 13485.
    • Risk Management: Adherence to ISO 14971.
    • No Unresolved Anomalies: No Discrepancy Reports (DRs) with "Safety Intolerable" or "Customer Intolerable" priority remaining.
    • Performance at least as well as Predicate Device: The device performs comparably in terms of safety and effectiveness to the IDENTIFY (K230576).

    Reported Device Performance:

    The document states: "Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly." And "The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports/Unresolved Anomalies) remaining which had a priority of Safety Intolerable or Customer Intolerable (applicable to the US)."

    Missing Information:

    The document does not provide the following details that would be typically found in a clinical study report or a more detailed performance evaluation:

    • A specific table of quantitative acceptance criteria and corresponding numerical performance results.
    • Details on a specific "study" with a test set, sample sizes for test or training sets, data provenance, expert qualifications, or ground truth establishment methods for a clinical or performance evaluation.
    • Information on MRMC studies or a human-in-the-loop effect size.
    • Information on standalone algorithm performance.

    Summary of Available Information from the Text:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria (Inferred): Conformance to requirements, proper functioning of hazard safeguards, compliance with specific software guidances (FDA Software Functions guidance, "major" level of concern), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), Quality Management (ISO 13485), Risk Management (ISO 14971), and absence of critical unresolved anomalies (Safety Intolerable or Customer Intolerable DRs). Performance at least as well as the predicate device.
      • Reported Device Performance: "Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly." "The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports/Unresolved Anomalies) remaining which had a priority of Safety Intolerable or Customer Intolerable (applicable to the US)."
    2. Sample sizes used for the test set and the data provenance:

      • Not provided. The document refers to "hardware and software verification and validation testing" but does not specify sample sizes for a test set of patient data or data provenance (e.g., country of origin, retrospective/prospective). This often implies bench testing and software verification without a dedicated clinical performance study.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not provided. This information would be relevant for a clinical performance study using expert labels, which isn't detailed here.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not provided.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not provided. The document focuses on the device's own performance and substantial equivalence, not an MRMC study with human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "Non-clinical Testing" section refers to "hardware and software verification and validation testing," implying performance of the device's functions, which would be "standalone" in nature. However, specific metrics or a dedicated "standalone study" in terms of clinical performance are not detailed. It's more about technical compliance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not specified as a formal ground truth for a clinical dataset. The "ground truth" for the non-clinical testing likely refers to engineering specifications, established safety standards, and validated software requirements.
    8. The sample size for the training set:

      • Not provided. Training set information is relevant for AI/ML devices, but IDENTIFY is described as a "system for motion monitoring," not explicitly an AI/ML diagnostic or predictive device in the traditional sense that would require a large training dataset with labeled ground truth of patient outcomes.
    9. How the ground truth for the training set was established:

      • Not applicable/Not provided. As above, a specific training set with associated ground truth is not detailed, as this appears to be a traditional medical device verification and validation rather than an AI/ML model for clinical decision support.