Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K240097

    Validate with FDA (Live)

    Date Cleared
    2024-05-22

    (131 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Polysaccharide Hemostatic Powder (HP-2, HP-3, HP-5) is indicated for external temporary use as a topical dressing on minor bleeding wounds such as minor cuts, minor lacerations and minor abrasions.

    Device Description

    The Polysaccharide Hemostatic Powder (HP-2, HP-5) are sterile, topical wound dressings comprised of plant base polysaccharides. The hemostatic particles contained in powder absorb the water in the blood through physical adsorption, so that the blood components are concentrated to form a stable clot and achieve the purpose of hemostasis.

    AI/ML Overview

    The information provided pertains to the regulatory submission for a medical device (Polysaccharide Hemostatic Powder) and does not describe acceptance criteria for a study showing device performance in the way typically associated with AI/ML-based diagnostic devices. The document focuses on demonstrating substantial equivalence to a predicate device for regulatory clearance, primarily through non-clinical testing.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, training set information) are not applicable or not explicitly detailed in this type of submission.

    However, I can extract the non-clinical testing performed and the conclusions drawn to fulfill the spirit of the request regarding "acceptance criteria" and "study proving the device meets the acceptance criteria" in the context of this regulatory document.

    Here's a breakdown based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Demonstrated through Non-Clinical Testing)Reported Device Performance (Summary of Non-Clinical Testing)
    Sterilization: Sterilization procedure validated per ISO 11135:2014.Sterilization procedure validated in accordance with ISO 11135:2014, for ETO sterilization.
    Shelf-Life: Demonstrate essential performance is achieved before and after shelf-life test at 2 years, validated per ASTM F1980-21 and packaging requirements per ISO 11607-1:2006.Shelf-life for two years validated via accelerated testing (ASTM F1980-21) and packaging requirements (ISO 11607-1:2006) met. Essential performance achieved before and after shelf-life test.
    Biocompatibility: Meet biocompatibility requirements per 2020 FDA Guidance (Use of International Standard ISO-10993-1) for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, and pyrogenicity.Biocompatibility evaluations conducted per FDA Guidance document 'Use of International Standard ISO-10993-1'. Tests performed for cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, and material-mediated pyrogenicity.
    Physical & Functional Requirements: Design verification testing demonstrates physical and functional requirements are met.Design verification testing performed to demonstrate physical and functional requirements were met.
    Safety and Effectiveness (Substantial Equivalence): Performed substantially equivalent to the predicate device in an animal model.Safety and effectiveness study performed on a porcine animal model demonstrated substantial equivalence to the predicate device.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in terms of numerical size for the animal model or other non-clinical tests. The document only mentions "a porcine animal model."
    • Data Provenance: The animal study was performed (implied as part of the manufacturer's testing). No country of origin for the data is specified, nor is it classified as retrospective or prospective (animal studies are typically prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This information is typically required for studies involving human clinicians interpreting data or images. The current submission describes non-clinical and animal testing.

    4. Adjudication method for the test set

    • Not Applicable. Again, this is relevant for studies involving human interpretation and disagreement resolution, which is not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No MRMC study was conducted, as this device is a physical hemostatic powder, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or software device.

    7. The type of ground truth used

    • For the animal study ("safety and effectiveness study on a porcine animal model"), the "ground truth" would likely be derived from direct observation of hemostasis and potentially histological examination of the wound site, compared to the performance of the predicate device. For other non-clinical tests (sterilization, shelf-life, biocompatibility, design verification), the ground truth is established by meeting predefined scientific and regulatory standards (e.g., bioassay results meeting specified limits, physical tests meeting design specifications).

    8. The sample size for the training set

    • Not Applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/ML device requiring a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191472

    Validate with FDA (Live)

    Device Name
    Biopsy Needle
    Date Cleared
    2020-01-29

    (240 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.

    The Biopsy Needle is also indicated to provide breast tissue sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

    The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

    Device Description

    The Biopsy Needle are used to remove, by cutting, a specimen of tissue for microscopic evaluation. The organs in which the device may be used include but are not limited to breast, kidney, liver, prostate, spleen and lymph nodes plus various soft tissue masses. As the device is single use device, which is individually packaged sterile devices. In the package, there is a whole device with the structure shown in the picture above without any accessories. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified bacterial film to maintain the sterilization condition of the device.

    AI/ML Overview

    This document describes the Ushare Medical Inc. Biopsy Needle (K191472) and its substantial equivalence to a predicate device. It primarily focuses on comparing technical characteristics and performance testing. However, it does not contain information about the device's diagnostic performance or any studies related to human-in-the-loop performance, as it is a medical instrument (biopsy needle) and not a diagnostic AI/ML device.

    Here's a breakdown of the requested information based on the provided text, with clear indications where the information is not available for this type of device submission:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify formal "acceptance criteria" for diagnostic performance, as the device is a biopsy needle and not a diagnostic tool. Instead, it outlines performance testing to demonstrate substantial equivalence to the predicate device in terms of mechanical, biological, and sterilization characteristics.

    CharacteristicAcceptance Criteria (Implied by testing to standards/predicate)Reported Device Performance (Summary from text)
    BiocompatibilityMeet ISO 10993 requirementsMet ISO10993-1, ISO10993-4, ISO10993-5, ISO10993-10, ISO10993-11
    SterilizationMeet ISO 11135, ISO 11138, ASTM F1980, ISO 11737-1, ISO 10993-7, ISO 11607 requirementsMet ISO11135, ISO11138, ASTM F1980, ISO11737-1, ISO10993-7, ISO11607
    Device Shelf-life (5 years simulated)Maintain seal strength, dye penetration, vacuum leak, bacterial performance after accelerated agingAged subject devices (60°C, 162 days, <50% RH) tested for seal Strength, Dye Penetration, Vacuum Leak, and Packaging resistance Bacterial performance (implied to have met requirements as supporting substantial equivalence)
    Performance (Mechanical/Operational)Equivalent to predicate device for specific testsPerformance tests conducted on "Pressing parts, Pressing parts firing force, Cannula firing force, Safety switch, Sampling switch, Scale marks firmness, Total heavy metal content, Scale mark identification, Sampling structure, Sampling method, Penetration force." Comparison of results between proposed and predicate devices to prove equivalency.
    Biopsy Sample QualityEquivalent to predicate deviceBiopsy Sample Testing – Comparison of samples obtained by predicate and proposed devices to prove equivalency.
    Ultrasound VisibilityEnsure safety and effectivenessUltrasound Visibility Testing - Verification of the proposed device ultrasound visibility to ensure safety and effectiveness.
    PyrogenicityNon-pyrogenicResults from device testing indicate that subject devices are non-pyrogenic.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not available in the provided text. The document refers to "Performance Testing" which includes comparison testing with a predicate device and specific tests for biocompatibility, sterility, shelf-life, and mechanical function. However, details about the sample sizes of devices tested, the origin of the samples (e.g., if tissue samples were used), or whether the data was retrospective or prospective are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/available. As a biopsy needle, the device itself does not establish a "ground truth" in a diagnostic sense (e.g., identifying disease from an image). Its function is to obtain tissue samples. The "ground truth" for the effectiveness of the biopsy sample, once acquired, would typically be established by a pathologist examining the tissue. This document does not detail any studies involving expert interpretation of collected biopsy samples.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/available. Adjudication methods are typically relevant for diagnostic studies where expert consensus is needed to establish a ground truth for imaging or other diagnostic interpretations. For a physical device like a biopsy needle, this concept does not directly apply to the reported performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is a biopsy needle, an instrument used for tissue sampling, not an AI/ML-driven diagnostic tool. Therefore, an MRMC study or an assessment of human reader improvement with AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a medical instrument (biopsy needle), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is partially applicable but not explicitly detailed for the performance testing.

    • For Biopsy Sample Testing, the "ground truth" would implicitly be the histological examination of the tissue samples. The text states "Comparison of samples obtained by predicate and proposed devices to prove equivalency." This suggests that the quality or characteristics of the obtained tissue (which would be assessed via pathology) were compared. However, details on how this comparison was quantified or what specific metrics were used are not provided.
    • For other tests (biocompatibility, sterility, mechanical performance), the "ground truth" is adherence to established international standards and comparable performance to the predicate device.

    8. The sample size for the training set

    This information is not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable. As above, this device does not utilize a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1