Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K250393

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-12

    (89 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ZircaGlow & ZircaGlow HT Zirconia blanks are indicated for use by dental technicians for the production of full contour and substructures restorations up to a full arch.

    Device Description

    ZircaGlow & ZircaGlow HT Zirconia are disc and block shaped dental porcelain zirconia oxide blanks that come in various sizes that are used in custom restorations by the dental laboratory. The dental laboratory will further process the blank by milling the blank based upon the anatomically rendering of the patient's teeth (done at the dental office) through "Computer Aided Drafting/ Computer Aided Machining (CAD/CAM). Once the custom rendered blank is milled the product is fully sintered and colored (if required) and fitted to the patient's teeth as a crown, bridge or coping.

    AI/ML Overview

    The provided FDA 510(k) clearance letter details a dental product (zirconia blanks) and its substantial equivalence to a predicate device, rather than a medical device that utilizes artificial intelligence or machine learning. Therefore, many of the requested criteria related to AI/ML device performance (e.g., ground truth, expert adjudication, MRMC studies, effect size of AI assistance, training set details) are not applicable or cannot be extracted from this document.

    However, I can provide information based on the physical and chemical properties and the nonclinical testing performed to establish substantial equivalence.

    Acceptance Criteria and Device Performance for ZircaGlow & ZircaGlow HT Zirconia

    The device is a dental material (zirconia blanks) and its performance is evaluated against material properties and compliance with industry standards, comparing it to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for ZircaGlow & ZircaGlow HT Zirconia are primarily based on meeting or exceeding the requirements of established international standards (ISO 6872:2024 and ISO 13356:2015) for dental ceramic materials, and demonstrating similar performance characteristics to a predicate device.

    PropertyAcceptance Criteria (ISO Standard Requirement / Predicate Similarity)ZircaGlow Zirconia PerformanceZircaGlow HT Zirconia PerformancePredicate Device (ArgenZ HT+) Performance
    Sintered Density≥ 6.0 g/cm³ (ISO 13356:2024 Sec 4.1)≥ 6.04 g/cm³≥ 6.04 g/cm³6.08 g/cm³
    Bending/Flexural Strength> 800 MPa (ISO 6872:2024, Class 5)White: ~1138 MPaA4: ~1085 MPaC4: ~1058 MPa (>800 MPa)White: ~963 MPaA4: ~961 MPaC4: ~956 MPa (>800 MPa)1,348 MPa
    Fracture Toughness (KIC)≥ 5.0 MPa√m (ISO 6872:2024 Annex A, minimum for Class 5)White: 8.5 MPa√mA4: 8.1 MPa√mC4: 8.3 MPa√m (>5.0 MPa√m)White: 8.7 MPa√mA4: 8.5 MPa√mC4: 8.4 MPa√m (>5.0 MPa√m)Not supplied (but meets >5.0 MPa√m implicitly for Class 5)
    Chemical Solubility< 100 µg/cm² (ISO 6872:2024)< 20 µg/cm²< 20 µg/cm²66.11 µg/cm²
    Radioactivity (of ²³⁸U)≤ 0.1 Bq·g⁻¹ of ²³⁸U (ISO 6872:2024 Section 5.2 active conc. of not more than 1.0 Bq g-1)≤ 0.1 Bq·g⁻¹≤ 0.1 Bq·g⁻¹Not supplied
    Thermal Expansion Coefficient (20-500°C)No specific requirement in ISO 6872:2024; similar to predicate10.5 µm/m °C10.5 µm/m °C10.3 µm/m °C
    Material Composition (ZrO2+HfO2+Y2O3)> 99 wt% (Similar to predicate)> 99.9 wt%> 99.9 wt%> 99 wt%
    BiocompatibilityCompliant with ISO 10993-1:2018Assured through use of same materials and manufacturing processes as predicate devicesAssured through use of same materials and manufacturing processes as predicate devicesTested for Cytotoxicity on shaded material, no adverse reactions identified

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes for the mechanical and chemical property tests. The data provenance is derived from nonclinical testing performed by United Zirconia on their ZircaGlow and ZircaGlow HT Zirconia blanks. The location of the testing is not specified, but the applicant's address is Cairo, Egypt. This testing is prospective in the sense that the manufacturer specifically conducted these tests to demonstrate compliance for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the tests are for material properties, not for diagnostic or predictive performance requiring human expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for material properties testing. The values are determined by standardized laboratory methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a dental material, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a dental material, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device is based on objective measurements of physical and chemical properties determined through standardized laboratory testing methods (e.g., ISO 6872:2024, ISO 13356:2015). This is not equivalent to expert consensus, pathology, or outcomes data used for diagnostic devices.

    8. The sample size for the training set

    Not applicable. This is a dental material, not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. This is a dental material, not a machine learning model. The relevant "truth" for the manufacturing process would be adherence to quality system regulations (21 CFR Part 820).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1