Found 3 results

510(k) Data Aggregation

    K Number
    K192278

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-01

    (71 days)

    Product Code
    Regulation Number
    884.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hysteroscopy: The LUMINELLE DTx Hysteroscopy System is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    NOTE: Hysteroscopes are used as tools to access the uterine cavity and are not, in and of themselves, a method of surgery. When LUMINELLE DTx Hysteroscopy System is specifically used with the LUMINELLE Dx 360 Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only.

    Generally recognized indications for diagnostic hysteroscopy include:

    • Abnormal bleeding
    • Infertility and pregnancy wastage
    • Evaluation of abnormal hysterosalpingogram
    • Intrauterine foreign body
    • Amenorrhea
    • Pelvic Pain

    Generally recognized indications for use for operative hysteroscopy include:

    • Directed endometrial biopsy
    • Polypectomy
    • Submucous myomectomy
    • Transection of intrauterine adhesions
    • Transection of intrauterine septa
    • Endometrial ablation

    Cystoscopy: The LUMINELLE DTx Hysteroscopy System is intended for use in endoscopic access to and examination of the lower urinary tract, including the blader. When combined with accessory instruments, the user to perform various diagnostic and therapeutic procedures.

    NOTE: When LUMINELLE DTx Hysteroscopy System is specifically used with the LUMINELLE Dx 360 Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only.

    Device Description

    The LUMINELLE DTx Hysteroscopy System originally received 510(k) clearance in 2018 as a hvsteroscopic and cvstoscopic svstem (K181909). In addition, LUMINELLE DTx Hysteroscopy System received 510(k) clearance in 2019 for the addition of a sheath component, LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid to the system (K190827).

    Based on user feedback for a thinner option for the disposable sheath for only diagnostic use, UVision360, Inc. (hereafter UVision) proposes the addition of a new accessory, the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), as part of an engineering performance change.

    The LUMINELLE DTx Hysteroscopy System indications for use is identical to the previously cleared indications for use for the system. However, when the system is used with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), the indications for use are limited to diagnostic use only.

    No changes are proposed for the previously cleared components of the LUMINELLE DTx Hysteroscopy System. The new LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) accessory is a modification of the LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid component of the previously cleared LUMINELLE DTx Hysteroscopy System (K190827) with the following main changes:

    • . The new accessory has one fluid line rather than the two fluid lines of the previously cleared component.
    • . The new accessory has a combined channel for the scope and the fluid line rather than the separate channels of the previously cleared component.
    • The new accessory has no operative channel available unlike the previously cleared . component, due to the fact that it is for diagnostic use only.
    AI/ML Overview

    The provided text describes a 510(k) submission for the LUMINELLE DTx Hysteroscopy System, which includes the addition of a new accessory, the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic).

    The study aims to demonstrate substantial equivalence to a previously cleared predicate device (LUMINELLE DTx Hysteroscopy System, K190827).

    Summary of Acceptance Criteria and Device Performance:

    The document primarily focuses on verifying the safety and effectiveness of the new accessory by demonstrating that the modified device remains as safe and effective as the predicate. It does not provide specific quantitative acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity for detecting specific conditions) for the new diagnostic sheath, as the intent of this 510(k) is an engineering performance change and limitation of use for the new accessory rather than a new diagnostic claim. Instead, the "acceptance criteria" are implied through the successful completion of various engineering, biocompatibility, and usability tests.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Acceptance Criteria (Implied)Reported Device Performance
    Sterilization adoption for the new sheath in accordance with AAMI TIR28:2016.Product adoption and Process Equivalence for Ethylene Oxide Sterilization was conducted. Results demonstrated that the LUMINELLE Dx 360° Rotatable Disposable Sheath sterilization may be adopted into the existing validation protocol after microorganism characterization and EO residuals testing.
    13-month shelf life for the new sheath, confirmed by accelerated aging.An accelerated aging shelf life study was performed for packaged LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ASTM F1980-16. Visual inspection and functional design verification confirmed that all samples met specifications after aging.
    Packaging maintaining a sterile barrier throughout normal transportation and distribution.Distribution simulation testing in accordance with ASTM D4169-16 (Distribution Cycle 13: Assurance Level I) was performed. This included seal strength testing (ASTM F88-15) and bubble leak testing (ASTM F2096-11). Packages were confirmed to maintain integrity following simulated distribution.
    New sheath performing according to its intended use in in vivo models; user able to operate the system as intended; product conforming to user needs.Usability and design validation testing was repeated for the LUMINELLE DTx Hysteroscopy System with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic). This testing was conducted "in in vivo models." Devices met predefined acceptance criteria.
    New sheath not inducing sensitization or irritation.Sensitization and irritation testing was performed for the final finished LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ISO 10993-10:2016. Results demonstrate the device is not irritating or sensitizing.
    New sheath not inducing cytotoxicity.Cytotoxicity testing was performed for the final finished LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ISO 10993-5:2009. Results demonstrate the device is not cytotoxic.

    Detailed Information about the Study:

    1. A table of acceptance criteria and the reported device performance: Refer to the table above.

      • Note: The document does not specify quantitative diagnostic performance metrics because the submission is for an engineering change and a limitation of use for an accessory, not a new diagnostic claim that would require clinical performance study. The "acceptance criteria" are primarily related to safety, sterility, shelf life, and basic functionality as a medical device to ensure equivalence to the predicate.
    2. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: The document does not explicitly state specific sample sizes for the various tests (e.g., how many sheaths were tested for accelerated aging, how many "in vivo models" were used for usability). It refers to "all samples met specifications" or "packages were confirmed."
      • Data Provenance: The information provided is retrospective in nature, as it's a summary of testing already performed for a 510(k) submission. The country of origin for the data is implied to be within the scope of UVision360 Inc.'s operations (Raleigh, NC, USA), but not explicitly stated for the testing sites.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for this type of submission. The "ground truth" for these engineering, biocompatibility, and usability tests is established by adherence to recognized standards (e.g., AAMI TIR28, ASTM F1980-16, ASTM D4169-16, ASTM F88-15, ASTM F2096-11, ISO 10993-10:2016, ISO 10993-5:2009) and predefined acceptance criteria set during the device's development and validation. Experts in relevant engineering, toxicology, and quality assurance fields would have designed and overseen these tests, but their specific number and qualifications are not detailed in this regulatory summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This document describes technical device performance testing (e.g., sterility, shelf life, biocompatibility, mechanical function), not clinical diagnostic performance requiring expert adjudication of patient cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML-enabled device and no MRMC study was performed or needed for this type of submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML-enabled device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for the tests described is the adherence to specified technical and safety standards (e.g., ISO, ASTM standards for biocompatibility, packaging integrity, and accelerated aging) and predefined functional specifications for the device. For the usability and design validation, it implicitly relies on objective performance measures and user feedback in "in vivo models" compared against "predefined acceptance criteria."
    8. The sample size for the training set: Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190827

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-06-06

    (66 days)

    Product Code
    Regulation Number
    884.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hysteroscopy: The LUMINELLE DTx Hysteroscopy System is used to permit direct viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
    Note: Hysteroscopes are used as tools to access the uterine cavity and of themselves, a method of surgery. Generally recognized indications for diagnostic hysteroscopy include:

    • Abnormal bleeding
    • Infertility and pregnancy wastage
    • Evaluation of abnormal hysterosalpingogram
    • Intrauterine foreign body
    • Amenorrhea
    • Pelvic pain
      Generally recognized indications for operative hysteroscopy include:
    • Directed endometrial biopsy
    • Polypectomy
    • Submucous myomectomy
    • Transection of intrauterine adhesions
    • Transection of intrauterine septa
    • Endometrial ablation
      Cystoscopy: The LUMINELLE DTx Hysteroscopy System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the user to perform various diagnostic and therapeutic procedures.
    Device Description

    The LUMINELLE DTx Hysteroscopy System previously received 510(k) clearance (K181909) in 2018 as a hysteroscopic and cystoscopic system. To provide the user with a more rigid option for the RotoSheath and Introducer, UVision360, Inc. (hereafter UVision) proposes the addition of a new component to the system, the LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid in addition to the currently available flexible Rotatable Disposable Sheath option cleared during the previous submission (K181909). No changes are proposed for the previously cleared components or the principles of operation of the LUMINELLE DTx Hysteroscopy System. The proposed change includes only the addition of a new component which is a modification to the previously cleared flexible LUMINELLE DTx 360° Rotatable Disposable Sheath.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the LUMINELLE DTx Hysteroscopy System. This submission is for the addition of a new component, a rigid sheath, to an already cleared system. Therefore, the focus of the documentation is on demonstrating the substantial equivalence of the modified device, particularly the new rigid sheath, to the previous cleared flexible sheath version.

    Based on the provided text, a formal study demonstrating the device meets a specific set of acceptance criteria in the conventional sense (e.g., accuracy, sensitivity, specificity for an AI/diagnostic device) is not explicitly detailed. Instead, the document describes design validation testing and other engineering tests to ensure the new component performs as intended and does not raise new questions of safety or effectiveness compared to the predicate device.

    Here's a breakdown of the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics (e.g., sensitivity, specificity, accuracy). Instead, it states that:

    Acceptance Criteria CategoryReported Device Performance
    Design Validation"The LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid met all predefined acceptance criteria" for confirming the sheath performs to its intended use in in vivo models, the user is able to operate the system as intended, and the product conforms to user needs.
    Sterility and Shelf-lifePrevious sterilization, simulated distribution, and packaging shelf-life test documentation from the predicate device (K181909) were leveraged. The 1-year shelf-life was demonstrated through pull testing on the stainless steel hypotube bond, maintaining specifications after accelerated aging.
    BiocompatibilityBiocompatibility data from the predicate device (K181909) was leveraged because there are no differences in patient-contacting materials, manufacturing, or processing between the new rigid sheath and the predicate flexible sheath.
    Electrical SafetyDielectric strength testing was conducted for the rigid sheath to confirm it can withstand voltages in accordance with IEC 60601-1:2005/(R)2012 and A1:2012. Results were "acceptable per IEC 60601-2-2 [...] which is specified by IEC 60601-2-18."
    Substantial Equivalence"Performance testing shows that the LUMINELLE DTx Hysteroscopy System is as safe and effective as the previously cleared LUMINELLE DTx Hysteroscopy System (K181909)." The intended use and indications for use are identical to the predicate. The only technological difference (rigid vs. flexible sheath) "does not raise different questions of safety and effectiveness." The rigid sheath is also noted to be similar to a component of a reference device (K150158).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size for the "in vivo models" used in design validation testing.
    • Data Provenance: Not specified. The study is described as "design validation testing" using "in vivo models," which typically refers to animal or cadaveric models for medical devices, but specific details are absent. No information on country of origin or retrospective/prospective nature is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the type of testing described. The design validation testing focuses on confirming the sheath's functional performance and user operability, rather than diagnostic accuracy against a ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The testing described is functional and safety testing of a medical device component, not a diagnostic performance study requiring expert adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a hysteroscopy system, which is an imaging device for direct viewing and performing procedures. It does not appear to incorporate AI or machine learning for diagnostic interpretation that would warrant an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" in the diagnostic sense (expert consensus, pathology, outcomes) is not applicable here. The "ground truth" for the design validation testing would be the predefined functional and user performance specifications of the device component.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181909

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-16

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hysteroscopy: The Luminelle DTx Hysteroscopy System is used to permit direct viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    Note: Hysteroscopes are used as tools to access the uterine cavity and of themselves, a method of surgery.

    Generally recognized indications for diagnostic hysteroscopy include:

    • Abnormal bleeding
    • Infertility and pregnancy wastage
    • · Evaluation of abnormal hysterosalpingogram
    • · Intrauterine foreign body
    • · Amenorrhea
    • · Pelvic pain

    Generally recognized indications for operative hysteroscopy include:

    • · Directed endometrial biopsy
    • · Polypectomy
    • Submucous myomectomy
    • · Transection of intrauterine adhesions
    • · Transection of intrauterine septa
    • · Endometrial ablation

    Cystoscopy: The Luminelle DTx Hysteroscopy System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the System allows the user to perform various diagnostic and therapeutic procedures.

    Device Description

    The Luminelle DTx Hysteroscopy System is comprised of four components:

    1. Luminelle DTx Hysteroscope - a non-sterile, reusable flexible hysteroscope with an integrated light source, light fiber bundle and CMOS sensor.
    2. The Luminelle DTx 360° Rotatable Disposable Sheath - a sterile, two-part sheath installed over the insertion tube of the Hysteroscope and comprised of the Operative Introducer and the 360° RotoSheath. The Operative Introducer contains an operative channel for inserting surgical accessories, two fluid management lines (inflow/outflow), and a channel to protect the scope insertion tube. The 360° RotoSheath allows for rotational positioning of the camera and provides rigidity to the Scope/Introducer assembly.
    3. Luminelle Communication Cable - a cable that provides both the power to the hysteroscope and transmits the image.
    4. Luminelle Control Hub - the main power supply, image converter and visualization connection. The Control Hub has both HDMI and USB connectors for connection to a monitor and/or a PC. USB and HDMI cables are provided in the package.

    The Luminelle DTx Hysteroscope, Communication Cable, Control Hub, USB and HDMI cables and power cord are provided together in a single package. The 360° Rotatable Disposable Sheath is provided in a separate sterile package. The 360º Rotatable Disposable Sheath is required for use with the Hysteroscope, therefore it is a component of the system and not an accessory to the system.

    The CMOS sensor and fiber optic illumination bundles are contained in the Hysteroscope Insertion Tube. The insertion tube length is 29.8 cm (11.73 in.) and the outside diameter is 1.95 mm (0.077 in.). The Luminelle DTx Hysteroscope is packaged in a white form fit tray with a clear slide-on cover. The tray is made of high-impact polystyrene (HIPS). The clear cover is made of polyethylene terephthalate with glycol (PETG).

    The Luminelle 360° RotoSheath has a working length of 24.0 cm (9.45 in.) with an outside diameter of 5.7 mm (0.244 in.). The Luminelle Operative Introducer has an outside diameter of 4.8 mm (0.189 in.) and the inside diameter of the operative channel is 1.85 mm (0.073 in.).

    Both the Luminelle 360º RotoSheath and Luminelle Operative Introducer are supplied as a single assembled unit packaged in a bottom tray with retaining lid made of polyethylene terephthalate with glycol (PETG) and sealed with Tyvek. The assembled package is sterilized using ethylene oxide (EO).

    The Control Hub supplies power to the Hysteroscope via a reusable Communication Cable. The CMOS sensors have a processor that converts the visual image into data to be transferred. A processor in the Control Hub converts this image data into streaming HDMI language, takes still images when the button is pushed, and also controls the light exposure during changing light conditions.

    Visualization on a video monitor is provided via a HDMI cable, and image transfer to a PC is provided via a Type A USB cable. A 12V DC, 3A power adapter is also provided. Technical specifications for these reusable components are provided in the Instructions for Use.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Luminelle DTx Hysteroscopy System. It focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study to prove the device meets specific performance acceptance criteria in the context of an AI/ML-driven medical device.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/ML performance (e.g., sample sizes for test/training sets, expert ground truth, MRMC studies, effect sizes) is not present in this document.

    However, based on the provided text, I can infer and extract information related to performance testing that was conducted to support the substantial equivalence claim.

    Here's the breakdown of what can be gleaned:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present a table of specific quantitative acceptance criteria with corresponding performance metrics for image quality in a way that would be typical for an AI/ML device validating diagnostic accuracy. Instead, it refers to compliance with established standards for medical devices and compares features to predicates.

    Acceptance Criteria (Implied/Standard Complied)Reported Device Performance (as stated in the document)
    Optical Resolution (per ISO 8600)Performed, but specific resolution metrics (e.g., lp/mm) and comparison to predicate are not explicitly stated in the summary. The "Image Resolution" comparison mentions "CMOS chip is 400 x 400 pixels. USAF 1951 bar code Group-Element: 1-5" for the proposed device, and "CCD chip specifications not stated in product literature" for the predicate. Conclusion states "Different chip technologies are used, but differences do not raise different questions on safety or effectiveness."
    Biocompatibility (per ISO 10993; cytotoxicity, sensitization, irritation)Compliant. "Biocompatibility testing per ISO 10993" for the proposed device and "Biocompatibility testing per ISO 10993" for the predicates resulted in the conclusion "All three devices have patient contacting materials which have been tested per ISO 10993 for biocompatibility."
    Electrical Safety (per IEC 60601-1; IEC 60601-2-18)Compliant. "Compliant with requirements of IEC 60601-1; IEC 60601-2-18" for both proposed and primary predicate. Conclusion: "Both the proposed and primary predicate are compliant with the requirements of IEC 60601-1 and IEC 60601-2-18 for electrical safety."
    Thermal Safety (per IEC 60601-1; IEC 60601-2-18)Compliant. "Compliant with requirements of IEC 60601-1; IEC 60601-2-18." Conclusion: "Both the proposed and primary predicate are compliant with the requirements of IEC 60601-1 and IEC 60601-2-18 for thermal safety."
    Electromagnetic Compatibility (EMC) (per IEC 60601-1-2)Performed (stated simply as "electromagnetic compatibility per IEC 60601-1-2"). No specific results.
    Endoscope Specific Safety and Performance (per IEC 60601-2-18)Performed, no specific results given beyond general compliance.
    UsabilityPerformed. No specific results or metrics given.
    Reprocessing Validation (per FDA reprocessing guidance)Performed. Conclusion: "Both the proposed device and primary predicate are reprocessed the same way between each use, depending on whether it was used as a hysteroscope or cystoscope."
    Sterilization Validation (per ISO 11135)Performed. The 360° Rotatable Disposable Sheath is sterilized using ethylene oxide (EO).
    Shelf Life and Packaging ValidationPerformed. No specific results given.
    Software Validation (per FDA software guidance for moderate level of concern)Performed. No specific results given. The Control Hub has a processor that converts image data into streaming HDMI language, takes still images, and controls light exposure.
    Degree of Protection Against Electrical ShockType BF. Conclusion: "Both the proposed and primary predicate have the same degree of protection against electrical shock."
    Degree of Protection Against Invasion of LiquidsIPX7 for proposed device. "Fully immersible (per reprocessing instructions)" for predicate. Conclusion: "Both the proposed and primary predicate can be immersed in liquids for reprocessing (the primary predicate does not have an IP rating)."
    Objective Lens Characteristics (Focal Length, Field of View, Direction of View)Proposed: Focal Length: 5 – 50 mm, Field of View: 120° in air, Direction of View: Forward (0°). Predicate: Focal length: 3 – 50 mm, Field of view: 110° in air, Direction of View: Forward (0°). Conclusion: "Both the proposed and primary predicate have similar objective lens characteristics. The differences do not raise different questions on safety or effectiveness of the device."
    Mode of OperationContinuous. Conclusion: "Both the proposed and primary predicate are used in a continuous mode of operation."

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable in the context of an AI/ML test set with patient data. The performance testing described is primarily engineering and bench testing, as well as validation of reprocessing and sterilization.
    • Data Provenance: Not applicable for AI/ML validation. The testing described is hardware and system functionality testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable as this is not an AI/ML device requiring clinical diagnostic ground truth. The "ground truth" for the tests performed would be adherence to engineering specifications and regulatory standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable as this is not an AI/ML device nor a clinical study involving human readers and image interpretation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was performed as this is not an AI-assisted diagnostic device. It is an endoscope for direct visualization.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • Not applicable, as this device does not contain an AI algorithm. It is an imaging system (hysteroscope/cystoscope).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance validation is adherence to established international and FDA standards for medical devices (e.g., ISO, IEC). For example, electrical safety is validated against IEC 60601-1; biocompatibility against ISO 10993; optical resolution against ISO 8600.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is not an AI/ML device.