Found 2 results

510(k) Data Aggregation

    K Number
    K050653

    Validate with FDA (Live)

    Date Cleared
    2005-07-01

    (112 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TransEAR™ Bone Conduction Hearing Aid is indicated for persons with single sided deafness to provide the perception of sound from the deaf ear. The amplified signal is received into the deaf ear and transferred through the bones of the skull to the better cochlea.

    Device Description

    TransEAR™ is a bone conduction hearing aid. Unlike conventional air-conduction hearing aids, which depend on acoustic coupling through the air, TransEAR™ utilizes bone conduction technology to transmit sound vibrations through the bones of the skull to the cochlea.

    TransEAR™ consists of a processor and a transducer (oscillator) connected via a wire cable. TransEAR's processor uses conventional digital hearing aid technology housed in a conventional behind-the-ear aid housing. The transducer is a mechanical oscillator housed in a conventional ear mold made of standard ear mold material. Each ear mold is custom made from an impression of the wearer's ear. A standard Zinc-Air hearing aid battery is utilized for power.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the TransEAR™ Bone Conduction Hearing Aid, asserting substantial equivalence to predicate devices rather than providing a detailed study with acceptance criteria and device performance metrics. Such a submission focuses on demonstrating that a new device is as safe and effective as an already legally marketed device (predicate device) and does not typically include novel clinical studies with specific performance targets.

    Therefore, many of the requested details about acceptance criteria, detailed study design, sample sizes, expert qualifications, and specific performance metrics are not present in this type of regulatory document. The "study" here is essentially functional testing comparing the new device to existing ones.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of quantitative acceptance criteria or specific discrete performance metrics. Instead, "substantial equivalence" is the overarching acceptance criterion, evaluated based on intended use, function, design, materials, and performance compared to predicate devices. The "reported device performance" is a qualitative statement that "Results of this testing indicate that the TransEAR™ is substantially equivalent to the predicate devices."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    Not explicitly stated. The document refers to "Functional testing" but does not detail the size of any test set (e.g., number of participants, amount of data) or the data provenance. In a substantial equivalence claim for a hearing aid, functional testing would likely involve technical performance measurements against industry standards or predicate device specifications, rather than a clinical trial with a large human test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable and not provided. "Ground truth" in the context of this 510(k) is about demonstrating engineering and functional equivalence rather than clinical diagnostic accuracy requiring expert panel review.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable and not provided. Adjudication methods are typically used in studies where there is subjective interpretation of data (e.g., medical image analysis) to establish a consensus ground truth. This is not described for the functional testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a hearing aid, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or is relevant here.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable in the AI sense. The device itself is a standalone functional device. The "functional testing" refers to the device's technical specifications and operation, not an algorithm's performance in isolation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically discussed in diagnostic accuracy studies (e.g., pathology for imaging algorithms) is not directly applicable here. For a functional hearing aid, "ground truth" in the context of its performance would be established by objective measurements against established engineering and audiological standards. The document states that "Functional testing... was conducted. Results of this testing indicate that the TransEAR™ is substantially equivalent to the predicate devices." This implies comparison to the known performance of predicate devices and relevant standards.

    8. The sample size for the training set

    Not applicable. This is a conventional medical device, not an AI/machine learning algorithm requiring a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no AI training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970828

    Validate with FDA (Live)

    Date Cleared
    1997-05-28

    (83 days)

    Product Code
    Regulation Number
    874.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indication for use of the air conduction hearing aids in this submission is to amplify sound for individuals with impaired hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)): Severity: x 1. Slight x 2. Mild x 3. Moderate Configuration: x 1. High Frequency -Precipitously Sloping x 2. Gradually Sloping x 3. Reverse Slope x 4. Flat Other x 1. Low tolerance To Loudness

    Device Description

    UHS MicroNET Programmable Air Conduction Hearing Aids and Programmer Model Nos .; This hearing aid comes in your choice of models, as follows: Full Concha --P901 Demi Concha --P902 Custom Canal --P903 Mini Canal --P903m Low Profile --P906 CIC CanalMate --P909 All the above models are available in a choice of one of two circuits, the Etymotic DSD K-AMP circuit and the Gennum DSD-DynamEQ II circuit. The hearing aids are programmed using the UHS microNET Programmer, a handheld device. The aids can be fit to hearing impaired individuals with mild to moderate hearing losses. Because of the range of the programming, the aid can fit most all audiogram configurations and has provisions for patients with low tolerance to loud sounds.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically hearing aids, and as such, it does not contain the detailed study information typically found in a clinical trial report or a scientific publication that would include acceptance criteria and specific performance studies.

    However, based on the information provided in the document, here's what can be extracted and inferred regarding the "acceptance criteria" for the device to be cleared by the FDA as "substantially equivalent" to a predicate device:

    1. A table of acceptance criteria and the reported device performance

    The document states:

    • "Performance characteristics are Included with each aid. The performance characteristics were obtained using the test procedures outlined in ANSI S.22 (1987)."
    • The FDA's letter states: "We have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976...".

    Therefore, the acceptance criteria by the FDA for this 510(k) submission was achieving substantial equivalence to a predicate device, demonstrated through the device meeting the ANSI S.22 (1987) standard for performance characteristics and aligning with the stated indications for use.

    Since the document is a summary and an FDA clearance letter, it does not provide a table with specific numerical acceptance criteria (e.g., SNR, frequency response ranges) that would be detailed in a full testing report. Instead, it broadly states that the device adheres to recognized industry standards.

    Acceptance Criteria (Inferred from 510(k) Process)Reported Device Performance
    Substantial Equivalence: Device's technological characteristics and performance are comparable to a legally marketed predicate device."determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976..."
    Compliance with ANSI S.22 (1987): Performance characteristics meet the standards outlined in ANSI S.22 (1987) for hearing aids."Performance characteristics were obtained using the test procedures outlined in ANSI S.22 (1987)."
    Indications for Use Alignment: Device is suitable for individuals with specified types and severities of hearing loss.Indications for Use: "amplify sound for individuals with impaired hearing... indicated for individuals with losses in the following category(ies): Slight, Mild, Moderate; High Frequency - Precipitously Sloping, Gradually Sloping, Reverse Slope, Flat; Low tolerance To Loudness."
    Compliance with Federal Hearing Aid Labeling Regulations: Adherence to 21 CFR 801.420."Each aid comes with a user's brochure and other labeling which will ensure that the hearing aids will be marketed in compliance with Federal hearing aid labeling regulations (21 CFR 801.420)."
    Good Manufacturing Practice (GMP) Compliance: Adherence to 21 CFR Part 820."A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820)..."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set" in terms of subject sample size for a clinical study. The performance characteristics were "obtained using the test procedures outlined in ANSI S.22 (1987)." This standard typically involves objective electroacoustic measurements of the device itself rather than clinical trials with human subjects for 510(k) submissions of this nature, especially for devices similar to those already on the market.

    Therefore, there is no stated sample size for a human test set, nor is there information on data provenance (country of origin, retrospective/prospective). The "test set" in this context refers to the physical devices undergoing testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. As this concerns performance testing against an ANSI standard, rather than diagnostic interpretation or complex clinical outcomes, the ground truth would typically be established by calibrated measurement equipment and engineers/technicians following the standard's protocol.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document, as it likely refers to objective measurements against an ANSI standard rather than expert adjudication of clinical outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done or reported in this document. This type of study is typically associated with AI/CAD systems that assist human operators in interpreting medical images or data, which is not the nature of this hearing aid device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable to this device. Hearing aids are inherently "human-in-the-loop" devices designed to be worn and used by individuals. The performance reported (compliance with ANSI S.22) is for the device's electroacoustic properties, which is a standalone assessment of the device's functional characteristics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance, as per the submission, is compliance with the ANSI S.22 (1987) standard. This standard dictates the objective electroacoustic performance characteristics (e.g., gain, frequency response, output, distortion) that a hearing aid should meet. The ground truth is therefore based on standardized, objective physical measurements of the device's output and characteristics.

    8. The sample size for the training set

    No training set is referenced or applicable in this document. This device is a traditional hearing aid, not an AI or machine learning algorithm that requires a training set. The "programming" mentioned refers to adjusting acoustic parameters by a professional, not machine learning model training.

    9. How the ground truth for the training set was established

    As there is no training set for this device, this question is not applicable.