510(k) Data Aggregation

    K Number
    K110447

    Validate with FDA (Live)

    Device Name
    ECHOINSIGHT
    Date Cleared
    2011-05-27

    (100 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by or on the order of a qualified physician for analysis of ultrasound imaging of the human heart. Cardiac (adult and pediatric) applications using B-mode (including harmonic) imaging are supported. The system provides image data analysis applications that supply information on cardiac structure and motion.

    Device Description

    The Echolnsight™ software system enables the production and visualization of 2D tissue motion measurements (including tissue velocities, strain rates) and cardiac structural measurement information derived from tracking speckle in tissue regions visualized in any Bmode (including harmonic) imagery loops as captured by most commercial ultrasound systems.

    The Echolnsight™ software system has been designed to ingest and process ultrasound imagery stored in files with content organized and encoded in conformance to the following the following standards:

    • NEMA PS3 (3.3, 3.5, 3.6, 3.10): Digital Imaging and Communications in Medicine . (DICOM), 2008.
    • ISO/IEC 10918-1:1994-02: Information technology Digital compression and coding of . continuous-tone still images – Requirements and guidelines (JPEG standard), 1994.
    • ISO/IEC 15444-1:2004: Information technology JPEG 2000 image coding system: . Core coding system, 2004.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the EchoInsight™ device. However, it does not contain the acceptance criteria or details of a study proving the device meets acceptance criteria. The document primarily focuses on:

    • Device Description: What EchoInsight™ does (processing 2D tissue motion measurements, including velocities, strain rates, and cardiac structural measurements from ultrasound B-mode imagery using speckle tracking).
    • Intended Use: For analysis of ultrasound imaging of the human heart (adult and pediatric), providing information on cardiac structure and motion.
    • Technological Comparison to Predicate Device: Stating substantial equivalence to the Siemens VVI product (K091286) based on similar speckle tracking techniques for tissue motion, strain, and strain rate, and similar visualization capabilities.
    • FDA Clearance Letter: Confirming the substantial equivalence determination.

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria, device performance, or study details, as these are not present in the provided text.

    To illustrate what the requested information would look like if it were present, here's a hypothetical structure:


    Hypothetical Acceptance Criteria and Study Details (Based on typical medical device submissions, NOT on the provided text):

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (e.g., Predicate Equivalence, Clinical Threshold)Reported Device PerformanceOutcome
    Accuracy of Strain MeasurementMean absolute difference (MAD) ≤ 5% compared to predicate device or gold standard.MAD = 3.8% vs. predicateMet
    Precision of Velocity MeasurementCoefficient of Variation (CV) ≤ 10% for repeated measurements.CV = 7.2%Met
    Processing TimeMean processing time per image < 5 seconds.Mean processing time = 3.5 secondsMet
    Inter-operator Variability (Qualitative Agreement)Kappa statistic > 0.8 for visual assessment of global and regional cardiac function.Kappa = 0.85Met
    Intra-operator Variability (Quantitative Reproducibility)Intraclass Correlation Coefficient (ICC) > 0.9 for key cardiac parameters (e.g., Global Longitudinal Strain).ICC = 0.92Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: For quantitative accuracy/precision, 150 cardiac ultrasound studies from 100 distinct patients. For qualitative assessment, 200 distinct cardiac ultrasound studies.
    • Data Provenance: Retrospective, de-identified cardiac ultrasound studies collected from three university hospitals in the United States, Germany, and Japan.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    • Number of Experts: 3 independent expert cardiologists.
    • Qualifications: Each expert had over 10 years of experience in echocardiography interpretation, board-certified in cardiology, and specialized in advanced cardiac imaging techniques including speckle tracking.

    4. Adjudication Method for the Test Set

    • Adjudication Method: 2+1 (Two experts' consensus was taken as ground truth; if they disagreed, a third expert provided a tie-breaking decision. All three experts were blinded to the device's output during ground truth establishment).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect Size of Human Readers Improvement with AI vs. without AI assistance:
      • ROC AUC Improvement: Average increase in Area Under the Receiver Operating Characteristic Curve (ROC AUC) of 0.08 (e.g., from 0.82 to 0.90) for diagnosing early-stage cardiomyopathy.
      • Reading Time Reduction: Average reduction in reading time per case by 25% (e.g., from 120 seconds to 90 seconds).
      • Diagnostic Accuracy Increase: Increase in diagnostic accuracy by 10% (e.g., from 80% to 90%) in identifying regional wall motion abnormalities.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes.
    • Performance Metrics:
      • Sensitivity: 92% for detecting global longitudinal strain abnormalities.
      • Specificity: 88% for detecting global longitudinal strain abnormalities.
      • Accuracy: 90% for classifying normal vs. abnormal cardiac function based on strain.

    7. Type of Ground Truth Used

    • Ground Truth Type: Expert consensus from the 3 cardiologists, validated against subsequent patient outcomes data (e.g., pathology reports, follow-up clinical events over 12 months for cases of suspected disease).

    8. Sample Size for the Training Set

    • Training Set Sample Size: 15,000 cardiac ultrasound studies from 10,000 distinct patients.

    9. How the Ground Truth for the Training Set was Established

    • Training Set Ground Truth: Established through a combination of:
      • Automated extraction of quantitative parameters from existing clinical reports (e.g., ejection fraction, global longitudinal strain measured by predicate devices or manual methods).
      • Annotation by a team of trained, certified echocardiography technicians and junior cardiologists under the supervision of senior cardiologists.
      • Randomized review of 10% of annotations by a senior cardiologist to ensure quality and consistency.

    Crucially, the K110447 document is a 510(k) summary focused on substantial equivalence to a predicate device, not on detailed performance studies against specific acceptance criteria for a novel algorithm. For such details, one would typically look at the full 510(k) submission, associated peer-reviewed publications, or the manufacturer's technical documentation if it were a newer device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100363

    Validate with FDA (Live)

    Date Cleared
    2010-05-21

    (98 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultra DCI 5000 System is intended for use by, or on the order of, a qualified physician for ultrasound imaging of the human heart. The cardiac (adult and pediatric) application using Bmode imaging is supported.

    Device Description

    The Ultra DC1 Model 5000 is a general purpose, mobile, software-controlled, diagnostic ultrasound system designed to provide the user with the ability to collect standard 2D brightness mode (B-mode) echocardiographic views of the heart, visualize these views and store the collected data for later visualization. The device includes a 4-lead electrocardiograpic (ECG) module and patient connections to collect and display a single ECG trace (PQRST wave) along with the B-mode imagery.

    AI/ML Overview

    While the provided text describes the "Ultra DCI Model 5000" diagnostic ultrasound system and its intended use, it does not contain information about acceptance criteria or a specific study proving the device meets those criteria.

    The document is a 510(k) summary for premarket notification, which focuses on demonstrating substantial equivalence to predicate devices rather than presenting detailed performance studies with acceptance criteria.

    Here's what can be inferred from the document regarding the information requested, and what is explicitly missing:

    1. A table of acceptance criteria and the reported device performance

    • Missing. The document does not provide a table for acceptance criteria for clinical performance or any specific reported performance metrics (e.g., sensitivity, specificity, accuracy). It only states that the device is "substantially equivalent" to predicate devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Missing. No information about a test set sample size or data provenance is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Missing. No information about ground truth establishment or expert qualifications is provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Missing. No information about an adjudication method for a test set is provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Missing. No MRMC study is mentioned. The device described does not appear to be an AI-assisted device; it is a general-purpose diagnostic ultrasound system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Missing. As it's a diagnostic ultrasound system operated by a physician, a standalone algorithm-only performance is not applicable in the context described.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Missing. No information about the type of ground truth is provided.

    8. The sample size for the training set

    • Missing. No mention of a training set or its sample size is made.

    9. How the ground truth for the training set was established

    • Missing. No information about a training set or its ground truth establishment is provided.

    What the document does include that is relevant to device safety and performance:

    • Device Description: The Ultra DCI Model 5000 is a general-purpose, mobile, software-controlled, diagnostic ultrasound system for 2D B-mode echocardiographic views of the heart, with a 4-lead ECG module.
    • Intended Uses: For use by, or on the order of, a qualified physician for ultrasound imaging of the human heart, supporting cardiac (adult and pediatric) application using B-mode imaging.
    • Predicate Devices: K022567 (Sequoia Diagnostic Ultrasound System Signature II) and K024236 (Ultrasound Scanner Type 2400). The current device is considered "substantially equivalent" to these.
    • Compliance with Safety Standards: The device has been designed to meet NEMA UD-2 Measurement Standard (1998), IEC 60601-1:1988+A1:1991+A2:1995, IEC 60601-1-2:2001+A1:2004, and IEC 60601-2-37 (2004), (2005) Amendment 2. These are safety and electromagnetic compatibility standards, not specific clinical performance metrics.
    • Modes of Operation: Cardiac Adult and Cardiac Pediatric applications use B-mode and B-mode Harmonic imaging (marked as "N" for new indication in the tables). This indicates the specific imaging modalities available.

    In summary, the provided text describes a 510(k) submission for a diagnostic ultrasound system seeking substantial equivalence based on its technological similarities and adherence to general safety standards. It does not include the detailed performance study information, acceptance criteria, or ground truth establishment relevant to the specific questions asked.