Found 14 results

510(k) Data Aggregation

    K Number
    K152908

    Validate with FDA (Live)

    Date Cleared
    2016-05-03

    (215 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UCP Compact Drug Tests Cards and UCP Compact Drug Tests Cups are rapid, qualitative, competitive binding immunoassays for qualitatively the detection of the following drugs and their metabolites in human urine at the following cut-off concentrations:

    TestCalibratorCut-off
    Marijuana:Delta-9-THC-COOH50 ng/mL
    Cocaine:Benzoylecgonine300 ng/mL
    Amphetamine:D-Amphetamine1000 ng/mL
    Methamphetamine:D-Methamphetamine1000 ng/mL
    Opiates:Morphine2000 ng/mL
    Morphine:Morphine300 ng/mL
    Phencyclidine:Phencyclidine25 ng/mL
    Barbiturates:Secobarbital300 ng/mL
    Benzodiazepines:Oxazepam300 ng/mL
    BuprenorphineBuprenorphine10 ng/mL
    Methadone:Methadone300 ng/mL
    EDDP:EDDP300 ng/mL
    Oxycodone:Oxycodone100 ng/mL
    MDMA:MDMA500 ng/mL
    PropoxyphenePropoxyphene300 ng/mL
    Tricyclic Antidepressants:Nortriptyline1000 ng/mL

    The tests are intended for over-the-counter (OTC) consumer use as the first step process to provide the consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory test in the second step of the two-step process, is provided in the package labeling.

    The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The tests will yield preliminary positive results when the prescription drugs Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidentessants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Tricyclic Antidepressant in urine. The multi-drug of abuse urine test device shows the drug was or was not present at the cutoff level. The tests provide only preliminary test results, which must be confirmed by other methods such as gas chromatography/mass spectrometry (GCMS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring the drug levels.

    Device Description

    UCP Compact Drug Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, EDDP (Methadone metabolite), Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Morphine, Oxycodone, Phencyclidine, Propoxyphene. Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    The document describes the performance of UCP Compact Drug Tests, which are rapid, qualitative, competitive binding immunoassays for detecting various drugs and their metabolites in human urine. The study presented aims to demonstrate the substantial equivalence of the device to legally marketed predicate devices.

    Here's an analysis of the acceptance criteria and study data based on the provided text, structured to answer your specific questions:

    Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state numerical "acceptance criteria" in a table format for each drug test. Instead, it reports the accuracy in comparison to a reference method (GC/MS for most drugs, HPLC for TCA). The key performance metric mentioned is accuracy, particularly for the EDDP test. For consumer studies, the "agreement rate" with GC/MS results and understanding of instructions are the criteria.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Performance MetricImplied Acceptance Criteria (Based on reported success)Reported Device Performance (as stated in the document)
    Accuracy (EDDP test)≥ 98% accuracy (based on reported performance)"UCP EDDP tests demonstrated performance of ≥ 98% accuracy when compared to the reference method GC/MS."
    Lay User Agreement Rate (Consumer Study)≥ 97% agreement rate (based on reported success)"The test results performed by the lay users showed 97% or above agreement rate with GC/MS results"
    Lay User Understanding of Instructions/InterpretationSufficient understanding for satisfactory use"Participants responses support that the lay users can understand how to run the test, interpret the test results, the importance of confirmatory test, and some issues concerning certain prescription medicines and foods may affect the test results."

    Note: For the other drug tests (Marijuana, Cocaine, Amphetamine, etc.), the document states that their accuracy studies "can be found in the following 510(k) submissions: K061457, K091588, K091612, K110515, K122419, K123863, K130463, K131811, K132812." This implies that the accuracy for these drugs was established in prior submissions and accepted previously, but the specific numerical accuracy values are not reiterated in this document for each drug.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

    • EDDP Test: 80 clinical urine specimens per drug (EDDP test specific, implies 80 specimens for EDDP). The document specifies "containing approximately 10% of the specimens containing one type drug at concentrations between -50% cutoff ranges, another 10% of the specimens containing one type drug at concentrations between cutoffs to +50% cutoff ranges."
    • Consumer Study: 230 lay persons participated.
    • Data Provenance:
      • Country of Origin: Not explicitly stated for either clinical specimens or consumer study participants.
      • Retrospective or Prospective: Not explicitly stated. The description of preparing urine samples for the consumer study ("spiking pure drugs or drug metabolites into drug free human urine") suggests a prospective, controlled laboratory setting rather than purely retrospective clinical samples. For the EDDP test, "clinical urine specimens" suggests real-world samples, but it's not specified if they were collected prospectively for the study or retrospectively.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • EDDP Test & Consumer Study: The ground truth was established by laboratory reference methods (GC/MS or HPLC). There is no mention of human "experts" establishing the ground truth by consensus for the core performance metrics.

    3. Adjudication Method for the Test Set:

    • Since the ground truth was established by reference laboratory methods (GC/MS/HPLC), no human adjudication method (e.g., 2+1, 3+1) was applicable or used.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size:

    • No, an MRMC comparative effectiveness study involving human readers assisting AI or vice-versa was not done. This device is a rapid diagnostic test for drug detection, where the output is directly observed (e.g., lines on a test card), not an AI algorithm assisting human interpretation of complex images. The "consumer study" evaluates lay user interpretation of the device's results, but it's not a comparative effectiveness study of human readers with vs. without AI assistance.

    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    • This question is not directly applicable in the context of this device. The device itself is a standalone diagnostic test (a chemical immunoassay) that produces a visual result. There is no "algorithm" in the sense of software processing images or complex data. The performance of the device itself (its ability to detect drugs at given cut-offs) is assessed independently of human interpretation in the accuracy studies (e.g., EDDP test vs. GC/MS). The "consumer study" then assesses the human-in-the-loop performance (lay users interpreting the device's results).

    6. The Type of Ground Truth Used:

    • Laboratory Reference Methods:
      • For the EDDP accuracy study: Gas Chromatography/Mass Spectrometry (GC/MS).
      • For the consumer study: GC/MS for most drugs, and HPLC for Tricyclic Antidepressants (TCA).
    • This falls under "outcomes data" in a sense, but more precisely, it's highly accurate chemical analysis acting as the gold standard.

    7. The Sample Size for the Training Set:

    • This is not applicable as the device is a chemical immunoassay, not a machine learning or AI model that requires a training set. The "development" of the test involves chemical and manufacturing processes, not data-driven model training.

    8. How the Ground Truth for the Training Set was Established:

    • Not applicable (see point 7).
    Ask a Question

    Ask a specific question about this device

    K Number
    K151213

    Validate with FDA (Live)

    Date Cleared
    2015-08-21

    (107 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Drug Test Mini Cups are rapid, qualitative, competitive binding immunoassays for the following drugs and their metabolites in human urine:

    TestCalibratorCut-off
    AmphetamineD-Amphetamine1000 ng/mL
    BarbituratesSecobarbital300 ng/mL
    BenzodiazepinesOxazepam300 ng/mL
    BuprenorphineBuprenorphine10 ng/mL
    CocaineBenzoylecgonine300 ng/mL
    MarijuanaDelta-9-THC-COOH50 ng/mL
    MethadoneMethadone300 ng/mL
    MethamphetamineD-Methamphetamine1000 ng/mL
    MDMAMDMA500 ng/mL
    MorphineMorphine300 ng/mL
    Opiates20002000 ng/mL
    OxycodoneOxycodone100 ng/mL
    PhencyclidinePhencyclidine25 ng/mL
    PropoxyphenePropoxyphene300 ng/mL
    Tricyclic AntidepressantNortiptyline1000 ng/mL

    The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding - the second step in the process, along with the materials for shipping the urine speciment to the laboratory, is provided. The test is also intended for prescription use

    The tests will yield preliminary positive results when the prescription drugs Barbiturates, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    Device Description

    UCP Drug Test Mini Cups is competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Morphine, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    The provided document is a 510(k) summary for the UCP Drug Test Mini Cups, a rapid, qualitative, competitive binding immunoassay for detecting various drugs and their metabolites in human urine. It describes the device, its intended use, and performance characteristics compared to a predicate device.

    Here's an analysis based on your requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present "acceptance criteria" in a table format with corresponding "reported device performance" in a manner typical for a standalone AI/device study. Instead, it states that the performance was established in a predicate submission and that verification studies were conducted for the modified device (smaller cup size). These verification studies are described as having "demonstrated that UCP Drug Test Mini Cups performs satisfactorily when used according to the package inserts."

    However, we can infer the acceptance criteria from the nature of the device (a qualitative drug test) and the fact that its performance was deemed satisfactory by the FDA for substantial equivalence. The key performance characteristics for such a device typically revolve around accuracy (sensitivity and specificity relative to a gold standard like GC/MS at specific cut-off levels), precision, and reproducibility. While specific numerical criteria are not listed, the document states:

    "The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Tricyclic Antidepressant in urine. The multi-drug of abuse urine test device shows the drug was or was not present at the cutoff level."

    Performance CharacteristicAcceptance Criteria (Inferred)Reported Device Performance
    AccuracyDetect presence/absence of drugs/metabolites at specified cut-off levels (e.g., 50% positive at cut-off, <50% below, >50% above). Preliminary positive results for certain prescription drugs at therapeutic doses."UCP Drug Test Mini Cups performs satisfactorily when used according to the package inserts." "The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses." "The multi-drug of abuse urine test device shows the drug was or was not present at the cutoff level."
    PrecisionConsistent results across replicates and different levels of analyte.Mentioned as a "precision study" conducted to support the modification, with satisfactory results. Specific quantitative results are not provided.
    ReproducibilityConsistent results across different manufacturing lots.Mentioned as an "inter lots reproducibility study" conducted to support the modification, with satisfactory results. Specific quantitative results are not provided.
    SensitivityAbility to correctly identify positive samples at or above the cut-off.Mentioned as a "sensitivity study" conducted to support the modification, with satisfactory results. Specific quantitative results are not provided. The device aims to detect drugs at the specified cut-off concentrations (e.g., Marijuana: 50 ng/mL, Cocaine: 300 ng/mL, etc.).
    pH and Specific GravityPerformance maintained across various pH and specific gravity levels of urine.Mentioned as a "pH and specific gravity study" conducted to support the modification, with satisfactory results. Specific quantitative results are not provided.
    Lay User PerformanceAbility for OTC users to correctly perform the test and interpret results according to instructions.Mentioned as a "lay users study" conducted; "satisfactory" results are implied by the overall conclusion of substantial equivalence for OTC use. The package insert provides information for consumers on the two-step process (initial screening, then confirmatory lab test). The summary emphasizes the preliminary nature of the results and the need for professional judgment.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "Analytical performance was established in the predicate submission. In addition, verification studies were conducted in support of the modification - a smaller size test cup, including the precision study, inter lots reproducibility study, sensitivity study, pH and specific gravity study, accuracy study and the lay users study."

    • Test Set Sample Size: Not explicitly stated in this summary. The summary refers to the predicate submission (K130463) for the primary analytical performance data. For the verification studies related to the cup size modification, specific sample sizes are not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be laboratory-based analytical studies and a lay user study, which are typically prospective in nature for performance verification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a drug test kit, not an imaging device typically requiring expert interpretation for ground truth.

    • Number of Experts & Qualifications: Not applicable in the same way as an imaging study. The "ground truth" for drug detection is typically established through confirmatory laboratory methods, such as Gas Chromatography/Mass Spectrometry (GC/MS). The document explicitly states: "The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)." This implies that GC/MS would be the reference method for determining the true presence or absence of drugs/metabolites at the cut-off levels in the urine samples used for validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable in the context of this device. Adjudication methods like 2+1 or 3+1 refer to agreement among multiple human readers for an imaging study. For a diagnostic test like this, performance is typically assessed against a definitive laboratory reference method (GC/MS).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a standalone diagnostic test kit (an immunoassay), not an AI system intended to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, this is essentially a standalone device. The UCP Drug Test Mini Cups is a qualitative, competitive binding immunoassay that provides a visual result (line appears or doesn't appear). While a human interprets the visual result, the device itself is a chemical assay, and its performance is evaluated based on its ability to produce correct results given a sample, independent of ongoing human intervention in the assay process. The "lay users study" assesses the human-in-the-loop aspect of over-the-counter use, but the analytical performance (sensitivity, specificity, precision) is intrinsic to the device's chemical reactions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The type of ground truth used for such drug tests is confirmatory laboratory methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS). The document explicitly states: "The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)." This is the gold standard for drug detection in urine.

    8. The sample size for the training set

    This device is not an AI/machine learning algorithm, so there is no "training set" in the conventional sense of AI development. The device's design and parameters (e.g., antibodies, cut-off levels) are established through research and development, but not through an iterative learning process on a large training dataset.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for an AI model. The parameters of the immunoassay are determined through biochemical principles and established analytical chemistry.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132812

    Validate with FDA (Live)

    Date Cleared
    2014-03-06

    (178 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Multi-Drug Test Key Cups are rapid tests for preliminary detection of the following drugs in human urine: Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressant. The test configuration comes with a single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers with information concerning the presence or absence of the above stated drugs in a urine sample. The second step is to send preliminary positive samples for confirmation testing by GCMS. The test is not intended to distinguish between prescription use or abuse of the following drugs: Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants. There are no uniformly recognized cutoff concentration levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxyodone, Propxyphene, Tricyclic Antidepressant in urine. Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. For Over-The-Counter (OTC) use For In Vitro Diagnostics only

    Device Description

    UCP Multi-Drug Test Key Cups are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Opiates 300, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the UCP Multi-Drug Test Key Cups, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the UCP Multi-Drug Test Key Cups are defined by the agreement rates with reference methods (GC/MS or HPLC) and the predicate device. The performance of the device is then measured against these criteria.

    Test TypeAcceptance CriteriaReported Device Performance
    Accuracy Studies (Clinical Samples):
    Candidate vs. Predicate Device100% agreement required (implied by "demonstrated 100% agreements")100% agreement between the candidate device and the predicate device.
    Candidate vs. Reference Method (GC/MS or HPLC)> 97.5% agreement (stated explicitly)Over 97.5% agreement between the candidate devices and the reference method (GC/MS or HPLC).
    Consumer Studies (Lay Users):
    Lay User vs. Reference Method (GC/MS or HPLC)97% or above agreement rate (stated explicitly)97% or above agreement rate with GC/MS results.

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Samples for Accuracy Studies):
      • Sample Size: Total 80 clinical urine samples per one drug test. (Since there are 15 distinct drug tests listed, the total number of clinical samples used across all drug types would be 15 * 80 = 1200 samples).
      • Data Provenance: The clinical samples were obtained from "reference laboratories." It is not explicitly stated which country, but it is implied to be relevant to the US regulatory context (given the FDA submission). The data is retrospective, as the samples were "obtained" and then tested.
    • Test Set (Consumer Studies):
      • Sample Size: 115 lay persons participated. The number of urine samples used per person or per drug is not specified, but the samples were prepared to contain various drug concentrations (strong negative, weak negatives, weak positives, high positive).
      • Data Provenance: Not explicitly stated, but the study was conducted "among 115 lay persons in three geographic regions." This suggests a prospective study specifically designed for the consumer evaluation.

    3. Number of Experts and Qualifications for Ground Truth

    • Clinical Sample Ground Truth: The ground truth for the clinical samples was established by "reference laboratories." The text states that "all clinical urine samples including drug negative urine samples and drug positive urine samples were tested by the reference method GC/MS, except TCA. The TCA positive urine samples were tested by HPLC method." This indicates that GC/MS and HPLC are the expert-level reference methods providing the ground truth, rather than human experts adjudicating individual cases. The qualifications of the personnel performing these reference methods are not specified, but it's implied they are qualified laboratory professionals.
    • Consumer Study Ground Truth: The ground truth for the urine samples used in the consumer study was also established by objective analytical methods: "the final drug concentrations in each urine sample were confirmed by GC/MS but TCA, TCA concentrations in the urine samples was confirmed by HPLC."

    4. Adjudication Method for the Test Set

    • Clinical Sample Studies: There was no human adjudication method described. The comparison was directly between the device's results and the objective results from the reference methods (GC/MS and HPLC).
    • Consumer Studies: There was no human adjudication method described. The comparison was directly between the lay users' interpretations of the device's results and the objective results from the reference methods (GC/MS and HPLC).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not explicitly mentioned or described in the provided text. The studies focused on device accuracy against reference methods and lay user performance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    • Yes, a standalone performance study was conducted. The "Accuracy Studies" directly compare the UCP Multi-Drug Test Key Cups (device only, without human interpretation as part of the accuracy measure) against established reference methods. The "Consumer Studies" also have a standalone component, as the device itself is producing a result that the lay users then interpret. The reported agreement rates of the device with GC/MS/HPLC in both types of studies represent its standalone performance in terms of detecting the target substances.

    7. Type of Ground Truth Used

    • The primary type of ground truth used was objective analytical methods/pathology, specifically:
      • Gas Chromatography/Mass Spectrometry (GC/MS) for most drug classes.
      • High-Performance Liquid Chromatography (HPLC) for Tricyclic Antidepressants (TCA).

    8. Sample Size for the Training Set

    • The text does not explicitly state a sample size for a training set. This is common for lateral flow immunochromatographic assays like the UCP Multi-Drug Test Key Cups, which are typically developed and validated rather than "trained" in the machine learning sense. The performance characteristics were established through precision, sensitivity, specificity, cross-reactivity, interference, and stability studies, but these do not typically involve a "training set" in the same way an AI/ML model would.

    9. How the Ground Truth for the Training Set Was Established

    • As a training set is not explicitly referred to in the context of this device's development or the provided studies, the method for establishing its ground truth is not applicable from the given information. The ground truth for the test sets (clinical and consumer samples) was established using GC/MS and HPLC, as detailed in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131236

    Validate with FDA (Live)

    Date Cleared
    2013-11-07

    (190 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UCP Pregnancy Cassette Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The device is intended for Prescription Use Only including at point of care sites.

    UCP Pregnancy Dip Card/Strip is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The device is intended for Prescription Use Only including at point of care sites.

    UCP Home Pregnancy Cassette Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy.

    UCP Home Pregnancy Midstream Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy.

    UCP Home Pregnancy Dip Card/Strip is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy.

    Device Description

    U

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study proving the device meets them:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance Criteria (from Predicate Device)Reported Device Performance (UCP Home/Pregnancy Test)
    Sensitivity25 mIU/mL25 mIU/mL
    SpecificityNo interferences when tested with LH, FSH and TSHNo interferences when tested with LH, FSH and TSH
    Agreement (Professional vs. Professional)Not explicitly stated, but predicate performance is the benchmark.100% agreement with predicate device results
    Agreement (Lay User vs. Professional)Not explicitly stated, but high agreement is implied for OTC use.100% agreement between lay users and professionals for UCP Home Pregnancy Tests
    Ease of Use/Interpretation (Lay User)Not explicitly stated, but essential for OTC.Participant responses support easy understanding of instructions and interpretation of results.

    Study Proving Device Meets Acceptance Criteria:

    The document describes several studies to demonstrate the safety and effectiveness of the UCP Home Pregnancy Tests/UCP Pregnancy Tests, thereby showing it meets the acceptance criteria (primarily by demonstrating substantial equivalence to the predicate device).

    2. Sample Size and Data Provenance:

    • Test Set Sample Size: 100 urine specimens.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates that the specimens were collected from "100 women who fit into the following categories: childbearing age, suspected pregnant women, women early in pregnancy, and the first trimester of pregnancy." The study appears to be prospective in the sense that the collection of specimens and testing were part of this specific comparison study.

    3. Number and Qualifications of Experts for Ground Truth:

    • Number of Experts: Not explicitly stated, but "lab professionals" are mentioned as performing testing for both the candidate and predicate devices. It is implied that these professionals are the "experts" in this context.
    • Qualifications of Experts: "Lab professionals" are mentioned. No further specific qualifications (e.g., years of experience, specific certifications) are provided.

    4. Adjudication Method for the Test Set:

    • The document implies a direct comparison, stating "There was 100% agreement between UCP Home Pregnancy Test results by lay users and by the professionals. The results were also 100% in agreement between UCP Home Pregnancy Tests/UCP Pregnancy Tests and the predicate device results tested by the professionals." This suggests a direct comparison method rather than a formal adjudication process (like 2+1 or 3+1). The predicate device's results, as interpreted by lab professionals, served as the primary benchmark.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done in the sense of evaluating how much human readers improve with AI assistance. This device is a rapid chromatographic immunoassay (a "stick test") for qualitative detection of hCG, not an AI-powered diagnostic imaging or analysis system that would typically be evaluated with MRMC studies.
    • However, there was a comparison of "readers" – specifically, lay users versus lab professionals – to assess the ease of use and interpretation for the Over-The-Counter (OTC) versions of the device. This is a different type of "reader" evaluation.

    6. Standalone (Algorithm Only) Performance Study:

    • The concept of a "standalone (algorithm only)" study is not applicable here because the device is a physical, chemical immunoassay, not an algorithm. The device itself is the "standalone" component in its intended use. The performance characteristics (sensitivity, specificity, precision, etc.) were established for the device itself.

    7. Type of Ground Truth Used:

    • The primary ground truth appears to be expert consensus / established clinical practice as demonstrated by the predicate device's results when interpreted by "lab professionals."
    • Implicitly, the presence of hCG in the urine specimens (from "suspected pregnant women, women early in pregnancy, and the first trimester of pregnancy") serves as the underlying clinical truth.

    8. Sample Size for the Training Set:

    • The concepts of "training set" and "test set" are typically used for machine learning algorithms. For this type of immunoassay, there isn't a "training set" in the machine learning sense. The "training" or development of the device would involve biochemical formulation and optimization, not data-driven algorithm training.
    • The document describes "comparison studies" using 100 urine specimens as the "test set" for performance evaluation against the predicate.

    9. How Ground Truth for the Training Set Was Established:

    • As there is no "training set" in the AI/ML context for this device, this question is not applicable. The development and calibration of the assay would rely on established biochemical principles and standards (e.g., WHO Fourth International Standard for hCG).
    Ask a Question

    Ask a specific question about this device

    K Number
    K131811

    Validate with FDA (Live)

    Date Cleared
    2013-07-19

    (30 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Compact Drug Test Cards and UCP Compact Drug Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users. The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes the UCP Compact Drug Test Cards and Cups, rapid, qualitative, competitive binding immunoassays for detecting various drugs and their metabolites in human urine. The information relevant to acceptance criteria and supporting studies is extracted below.

    Acceptance Criteria and Device Performance

    The acceptance criteria for each drug are defined by their respective cut-off concentrations. The device's performance is demonstrated by its qualitative detection of these drugs at or above these cut-off levels. The document indicates that the tests will yield preliminary positive results when the prescription drugs [listed] are ingested, even at or above therapeutic doses. This implies the device is performing as intended when it detects the presence of the drug at or above the specified cut-off.

    A table summarizing the acceptance criteria (Cut-off) for each drug is provided below. The "Reported Device Performance" column indicates the qualitative detection capabilities as implied by the "Indications for Use" statement.

    TestCalibratorAcceptance Criteria (Cut-off)Reported Device Performance
    AmphetamineD-Amphetamine1000 ng/mLDetects at or above cut-off
    BarbituratesSecobarbital300 ng/mLDetects at or above cut-off
    BenzodiazepinesOxazepam300 ng/mLDetects at or above cut-off
    BuprenorphineBuprenorphine10 ng/mLDetects at or above cut-off
    CocaineBenzoylecgonine300 ng/mLDetects at or above cut-off
    MarijuanaDelta-9-THC-COOH50 ng/mLDetects at or above cut-off
    MethadoneMethadone300 ng/mLDetects at or above cut-off
    MethamphetamineD-Methamphetamine1000 ng/mLDetects at or above cut-off
    MDMAMDMA500 ng/mLDetects at or above cut-off
    MorphineMorphine300 ng/mLDetects at or above cut-off
    Opiates 2000Morphine2000 ng/mLDetects at or above cut-off
    OxycodoneOxycodone100 ng/mLDetects at or above cut-off
    PhencyclidinePhencyclidine25 ng/mLDetects at or above cut-off
    PropoxyphenePropoxyphene300 ng/mLDetects at or above cut-off
    Tricyclic AntidepressantsNortiptyline1000 ng/mLDetects at or above cut-off

    Study Information

    The provided document is an FDA 510(k) clearance letter and an "Indications for Use" form. While it states that the device is "substantially equivalent" to legally marketed predicate devices, it does not include details of specific studies, test sets, or validation methodologies that demonstrate the device meets these acceptance criteria. The information below is not explicitly available in the provided text.

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not specified in the provided text.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not specified. However, the document mentions that results "should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)," implying that GC/MS would serve as a confirmatory ground truth method.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not specified.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a rapid diagnostic test (immunoassay), not an AI-assisted diagnostic tool requiring human reader interpretation in an MRMC study context.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The device itself is designed for standalone performance, providing preliminary positive/negative results. However, the document states it's the "first step in a two step process" where confirmatory testing (e.g., GC/MS) is the second step. The performance statistics for this standalone step are not detailed.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "Indications for Use" form states that "preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)." This implies that GC/MS is the intended "gold standard" or ground truth for confirmation.

    7. The sample size for the training set: Not applicable, as this is an immunoassay device, not a machine learning algorithm that requires a "training set."

    8. How the ground truth for the training set was established: Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130463

    Validate with FDA (Live)

    Date Cleared
    2013-05-15

    (82 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Home Drug Screening Test Cards, UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressant. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The tests are intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for prescription use. The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels. For Over-The-Counter (OTC) use and prescription use For In Vitro Diagnostics only.

    Device Description

    UCP Home Drug Screening Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Opiates 300, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    The UCP Home™ Drug Screening Test Cards and Cups are rapid, qualitative, competitive binding immunoassays for the detection of drugs and their metabolites in human urine. The acceptance criteria for this device are based on its ability to correctly identify the presence or absence of various drugs at specific cut-off levels, with a high agreement rate with GC/MS results.

    Here's an breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is intended to provide preliminary positive results when drug concentrations are at or above the specified cut-off levels. The performance is measured by the agreement rate with confirmatory laboratory methods.

    Drug TestCalibratorCut-off (ng/mL)Acceptance Criteria (Agreement with GC/MS)Reported Device Performance (Agreement with GC/MS)
    AmphetamineD-Amphetamine1000≥ 97%≥ 97%
    BarbituratesSecobarbital300≥ 97%≥ 97%
    BenzodiazepinesOxazepam300≥ 97%≥ 97%
    BuprenorphineBuprenorphine10≥ 97%≥ 97%
    CocaineBenzoylecgonine300≥ 97%≥ 97%
    MarijuanaDelta-9-THC-COOH50≥ 97%≥ 97%
    MethadoneMethadone300≥ 97%≥ 97%
    MethamphetamineD-Methamphetamine1000≥ 97%≥ 97%
    MDMAMDMA500≥ 97%≥ 97%
    MorphineMorphine300≥ 97%≥ 97%
    Opiates 2000Morphine2000≥ 97%≥ 97%
    OxycodoneOxycodone100≥ 97%≥ 97%
    PhencyclidinePhencyclidine25≥ 97%≥ 97%
    PropoxyphenePropoxyphene300≥ 97%≥ 97%
    Tricyclic AntidepressantNortriptyline1000≥ 97%≥ 97%

    2. Sample Size Used for the Test Set and Data Provenance

    The consumer study, which evaluates the device's performance when used by lay persons, involved 115 lay persons.
    The urine samples for this study were prepared with various drug concentrations (strong negative, very weak negative, weak negative, very weak positive, weak positive, high positive) by spiking pure drugs or metabolites into drug-free human urine. This suggests the data provenance is prospective as samples were specifically prepared for the study. The country of origin of the data is not explicitly stated, but given the US FDA submission, it is likely the study was conducted in the United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the test set (urine samples with various drug concentrations) was established by Gas Chromatography/Mass Spectrometry (GC/MS) for all drugs except Tricyclic Antidepressants (TCA), for which High-Performance Liquid Chromatography (HPLC) was used. This indicates laboratory methods, not human visual experts, established the ground truth for drug concentrations.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method for the comparison between the device results and the GC/MS/HPLC ground truth. The comparison appears to be a direct assessment of agreement between the device's qualitative results and the quantitative ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The UCP Home™ Drug Screening Test Cards and Cups are in-vitro diagnostic devices designed for direct user interpretation (either by lay users or healthcare professionals). There is no "AI assistance" component to improve human reader performance in this type of qualitative chemical assay. The consumer study evaluated the human user's ability to interpret the test and instructions, not their diagnostic performance with or without AI.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not explicitly applicable in the typical sense of a "standalone algorithm" for image analysis or similar AI applications. However, the device itself, when used by a lay person, functions as a standalone qualitative test. The "performance studies" mentioned (method comparison, specificity, cut-off study) assess the inherent chemical performance of the device itself, independent of user interpretation, through comparison with GC/MS/HPLC. The results for these studies demonstrated the device "performs satisfactorily when used according to the package inserts."

    7. The Type of Ground Truth Used

    The ground truth used was laboratory gold standard methods:

    • Gas Chromatography/Mass Spectrometry (GC/MS) for Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene.
    • High-Performance Liquid Chromatography (HPLC) for Tricyclic Antidepressants (TCA).

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. This device is an immunoassay, not a software algorithm that requires a training set. The performance studies used prepared urine samples to evaluate the device's analytical performance across different concentrations.

    9. How the Ground Truth for the Training Set Was Established

    As there is no distinct "training set" in the AI sense, this question is not applicable. The chemical properties and performance of the immunoassay are inherent to its design and manufacturing. The cut-off levels are predefined. The "ground truth" for evaluating its performance (as described in point 7) was established through established laboratory analytical techniques (GC/MS, HPLC).

    Ask a Question

    Ask a specific question about this device

    K Number
    K123863

    Validate with FDA (Live)

    Date Cleared
    2013-01-09

    (26 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Compact Drug Test Cards, UCP Compact Drug Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiate 2000, Oxycodone, Phencyclidine, Tricyclic Antidepressant. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text regarding the UCP Compact Drug Test Cards and Cups, focusing on the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity, accuracy) but rather defines the calibrator and cut-off levels for each drug the device is intended to detect. The "reported device performance" in this context refers to the device's design to detect these substances at or above the specified cut-off concentrations.

    TestCalibratorCut-off (Acceptance Criteria)Reported Device Performance (as per intended design)
    AmphetamineD-Amphetamine1000 ng/mLDetects Amphetamine at or above 1000 ng/mL
    BarbituratesSecobarbital300 ng/mLDetects Barbiturates at or above 300 ng/mL
    BenzodiazepinesOxazepam300 ng/mLDetects Benzodiazepines at or above 300 ng/mL
    CocaineBenzoylecgonine300 ng/mLDetects Cocaine at or above 300 ng/mL
    MarijuanaDelta-9-THC-COOH50 ng/mLDetects Marijuana at or above 50 ng/mL
    MethadoneMethadone300 ng/mLDetects Methadone at or above 300 ng/mL
    MethamphetamineD-Methamphetamine1000 ng/mLDetects Methamphetamine at or above 1000 ng/mL
    MDMAMDMA500 ng/mLDetects MDMA at or above 500 ng/mL
    MorphineMorphine300 ng/mLDetects Morphine at or above 300 ng/mL
    Opiate 2000Morphine2000 ng/mLDetects Opiates at or above 2000 ng/mL
    OxycodoneOxycodone100 ng/mLDetects Oxycodone at or above 100 ng/mL
    PhencyclidinePhencyclidine25 ng/mLDetects Phencyclidine at or above 25 ng/mL
    Tricyclic AntidepressantNortriptyline1000 ng/mLDetects Tricyclic Antidepressants at or above 1000 ng/mL

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not contain any information about the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The provided text does not contain any information about the number of experts used or their qualifications for establishing ground truth.

    4. Adjudication Method for the Test Set

    The provided text does not contain any information about the adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.

    The provided text does not indicate that an MRMC comparative effectiveness study was conducted. This device is a rapid qualitative immunoassay, not an AI or imaging device that would typically involve human readers for interpretation in the same way. The language specifies "preliminary positive results... should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)," implying a two-step process, not an aided human reading.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.

    The device itself, being a rapid immunoassay card/cup, is inherently a standalone "algorithm" (in the sense of a chemical reaction interpreted visually). It is designed to provide a result without human interpretation of complex images or data. However, the text does not explicitly describe a "standalone study" in the context of a software algorithm in the way typically discussed for AI devices. Its performance is based on its chemical reactivity and visual output.

    7. The Type of Ground Truth Used

    The ground truth for this type of drug test is typically established by confirmatory methods such as Gas Chromatography/Mass Spectrometry (GC/MS). The document explicitly states: "The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS)." This indicates that GC/MS is the gold standard used for confirmation and, by extension, would be the ground truth for evaluating the device's accuracy.

    8. The Sample Size for the Training Set

    The provided text does not contain any information about the sample size for a training set. This is not an AI/machine learning device that would typically have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Since this is not an AI/machine learning device, there is no "training set" in that context, and therefore, no information on how ground truth for a training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122419

    Validate with FDA (Live)

    Date Cleared
    2012-11-30

    (113 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Tricyclic Antidepressant. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users. The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Oxycodone, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Oxycodone, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    Device Description

    UCP Home Drug Screening Test Cups are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Opiates 300, Oxycodone, Phencyclidine, Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    Acceptance Criteria and Study Summary for UCP Home™ Drug Screening Test Cups

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the UCP Home™ Drug Screening Test Cups are implicitly defined by the agreement rates obtained in the consumer study against GC/MS (or HPLC for TCA) reference method. While explicit numerical acceptance criteria (e.g., "must achieve >95% agreement") are not stated, the reported performance of "97% or above agreement rate with GC/MS results" indicates that this level of performance was deemed acceptable for market clearance.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance (Agreement Rate with Reference Method)
    Consumer Study (Lay Users)≥97% Agreement≥97% agreement with GC/MS or HPLC results

    Note: An "Accuracy Study" comparing the device to predicate devices and reference methods (GC/MS/HPLC) also demonstrated "100% agreements between the candidate device and the predicate device" and "over 97.5% agreement between the candidate devices and the reference method GC/MS." This suggests that the internal accuracy criteria were met for healthcare professional use. However, for OTC clearance, the consumer study is the primary focus for usability and interpretability by lay users.

    2. Sample Size and Data Provenance

    Test Set (Consumer Study):

    • Sample Size: 115 lay persons participated.
    • Data Provenance: The study was conducted in "three geographic regions" within the United States. The data is prospective as participants performed the tests during the study.

    Test Set (Accuracy Study):

    • Sample Size: "Total 120 clinical urine samples per one drug test were included." (e.g., 120 samples for Marijuana, 120 for Cocaine, etc., implying multiple sets of 120 samples for each drug tested).
    • Data Provenance: Clinical urine samples were obtained from "reference laboratories." The document does not specify the country of origin of these laboratories or whether the collection was retrospective or prospective, but the context of a 510(k) submission generally implies samples are relevant to the US population.

    3. Number of Experts and Qualifications for Ground Truth

    • Accuracy Study (Reference Method Confirmation): The reference method for confirming drug concentrations was GC/MS (Gas Chromatography/Mass Spectrometry), except for TCA which used HPLC (High-Performance Liquid Chromatography). These are instrumental methods that do not rely on human "experts" in the traditional sense of clinical interpretations. The "experts" would be the skilled technicians or analysts operating and interpreting the results from these analytical instruments. No specific number or qualification of these individuals is provided.
    • Consumer Study (Reference Method Confirmation): Similar to the accuracy study, GC/MS (or HPLC for TCA) was used to confirm the final drug concentrations in the prepared urine samples.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for reconciling discrepancies between interpretations by multiple human readers (e.g., 2+1, 3+1). Instead:

    • Accuracy Study: The device's results were directly compared against the objective reference methods (GC/MS/HPLC).
    • Consumer Study: The lay users' interpretations of the device results were compared against the confirmed GC/MS/HPLC concentrations of the spiked urine samples.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The study mainly focused on:

    1. The device's accuracy against a gold standard (GC/MS/HPLC), simulating healthcare professional use.
    2. Lay user's ability to correctly perform and interpret the test according to instructions (consumer study).
      There is no mention of human readers improving with or without AI assistance, as this is a qualitative immunoassay device, not an AI-powered diagnostic tool requiring human interpretation of complex images or data.

    6. Standalone Performance Study

    Yes, a standalone performance study was done for the device itself.

    • Accuracy Study: This study directly assessed the device's performance (results from the test cups as interpreted by trained professionals or objectively read) against the reference methods (GC/MS/HPLC). This represents the algorithm's (or device's intrinsic chemistry's) performance.
    • Consumer Study: While involving lay users, the core assessment was whether the device's output could be correctly interpreted by them, thereby confirming the standalone performance as read by its intended (lay) user.

    7. Type of Ground Truth Used for the Test Set

    The ground truth used for both the accuracy study and the consumer study was objective chemical analysis:

    • Gas Chromatography/Mass Spectrometry (GC/MS) for most drugs.
    • High-Performance Liquid Chromatography (HPLC) for Tricyclic Antidepressants (TCA).
      This provides a highly accurate and quantifiable measure of the drug and metabolite concentrations in the urine samples.

    8. Sample Size for the Training Set

    The document does not provide information on a specific "training set" or sample size for training. This type of immunoassay device typically relies on established chemical and biological principles, and its performance characteristics are determined through analytical validation rather than machine learning training. The "training" in this context refers to the development and optimization of the assay components and parameters, which is not usually quantified by a "training set sample size" in regulatory submissions for such devices.

    9. How Ground Truth for the Training Set Was Established

    As noted above, a distinct "training set" in the context of machine learning is not applicable here. The ground truth for the development and validation of the assay (analogous to how a training set informs an algorithm) would have been established through well-controlled laboratory experiments using known concentrations of drugs and their metabolites, confirmed by highly accurate analytical methods like GC/MS or HPLC. This iterative development and refinement process ensures the device's antibodies and reagents react appropriately to the target analytes at specified cut-off levels.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110515

    Validate with FDA (Live)

    Date Cleared
    2011-04-20

    (56 days)

    Product Code
    Regulation Number
    862.3150
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection the following drugs and their metabolites in human urine:

    TestCalibratorCut-off
    AmphetamineD-Amphetamine1000 ng/mL
    BarbituratesSecobarbital300 ng/mL
    BenzodiazepinesOxazepam300 ng/mL
    CocaineBenzoylecgonine300 ng/mL
    MarijuanaDelta-9-THC-COOH50 ng/mL
    MethadoneMethadone300 ng/mL
    MethamphetamineD-Methamphetamine1000 ng/mL
    MDMAMDMA500 ng/mL
    MorphineMorphine300 ng/mL
    Opiate 2000Morphine2000 ng/mL
    OxycodoneOxycodone100 ng/mL
    PhencyclidinePhencylidine25 ng/mL
    Tricyclic AntidepressantNortriptyline1000 ng/mL

    The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users.

    The tests will yield preliminary positive results when the prescription drugs Barbiturates, Oxycodone, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Oxycodone, Tricyclic Antidepressant in urine. The tests only provide a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method for most drugs (HPLC is the preferred confirmatory method for Tri-cyclic Antidepressants). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    Device Description

    The UCP Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection the following drugs and their metabolites in human urine:

    TestCalibratorCut-off
    AmphetamineD-Amphetamine1000 ng/mL
    BarbituratesSecobarbital300 ng/mL
    BenzodiazepinesOxazepam300 ng/mL
    CocaineBenzoylecgonine300 ng/mL
    MarijuanaDelta-9-THC-COOH50 ng/mL
    MethadoneMethadone300 ng/mL
    MethamphetamineD-Methamphetamine1000 ng/mL
    MDMAMDMA500 ng/mL
    MorphineMorphine300 ng/mL
    Opiate 2000Morphine2000 ng/mL
    OxycodoneOxycodone100 ng/mL
    PhencyclidinePhencylidine25 ng/mL
    Tricyclic AntidepressantNortriptyline1000 ng/mL

    The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users.

    The tests will yield preliminary positive results when the prescription drugs Barbiturates, Oxycodone, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Oxycodone, Tricyclic Antidepressant in urine. The tests only provide a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method for most drugs (HPLC is the preferred confirmatory method for Tri-cyclic Antidepressants). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    AI/ML Overview

    The document is a 510(k) premarket notification approval letter for the UCP Drug Screening Test Cups. It provides information about the device's intended use and lists the drugs it screens for, along with their calibrators and cut-off levels. However, it does not include details about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement in a study.

    Therefore, I cannot provide the requested information from this document. The document is an FDA approval letter, not a study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091612

    Validate with FDA (Live)

    Date Cleared
    2010-02-24

    (266 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UCP Drug Screening Buprenorphine, Amphetamine 300, Methamphetamine 500, Cocaine 150 Tests are rapid, qualitative, competitive binding immunoassays for the detection the following drug in human urine:

    TestCalibratorCut-off
    BuprenorphineBuprenorphine10 ng/mL
    AmphetamineD-Amphetamine300 ng/mL
    MethamphetamineD-Methamphetamine500 ng/mL
    CocaineBenzoylecgonine150 ng/mL

    The tests contain three formats: 1) Test Card/Strip; 2) Test Device, 3) Test Cup. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for in vitro diagnostics use.

    The tests only provide a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or Liguid chromatography/mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    Device Description

    UCP Drug Screening Buprenorphine, Amphetamine 300, Methamphetamine 500, Cocaine 150 Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Buprenorphine, Amphetamine, Methamphetamine, Cocaine and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a numerical or percentage format beyond the overarching statement regarding performance. However, based on the accuracy study, the implied acceptance criterion for accuracy against both a legally marketed device and GC/MS or LC/MS is essentially that the device performs at a high level.

    Criterion TypeAcceptance Criteria (Implied)Reported Device Performance
    AccuracyHigh performance (e.g., ≥95% agreement/correlation with predicate and confirmatory methods)≥ 97% for all drugs when compared to a legally marketed device and GC/MS or LC/MS
    Qualitative DetectionDetection of Buprenorphine, Amphetamine, Methamphetamine, Cocaine, and their metabolites at specified cut-off levelsSuccessfully detects Buprenorphine (10 ng/mL), Amphetamine (300 ng/mL), Methamphetamine (500 ng/mL), Cocaine (150 ng/mL)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 80 clinical urine specimens per drug (Buprenorphine, Amphetamine, Methamphetamine, Cocaine), totaling 320 specimens (80 specimens * 4 drugs).
      • Approximately 10% (8 specimens per drug) were at concentrations between -50% cutoff and cutoff ranges.
      • Approximately 10% (8 specimens per drug) were at concentrations between cutoff and +50% cutoff ranges.
    • Data Provenance: Clinical urine specimens. The country of origin is not specified, but the study was conducted at "point of care sites." The data is retrospective in the sense that these were "clinical urine specimens," implying they were collected prior to this specific study for testing. However, the study itself is designed to prospectively test the new device against established methods.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the "number of experts" or their qualifications for establishing ground truth. The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) analysis, which are analytical laboratory methods considered the gold standard for confirmatory drug testing. These methods are performed by trained laboratory personnel, rather than medical "experts" in the clinical sense (e.g., radiologists).

    4. Adjudication Method for the Test Set

    The document describes a comparison study where the UCP device's results were compared to both GC/MS or LC/MS results and the predicate devices. It doesn't detail a specific "adjudication method" involving human consensus for discrepancies. The GC/MS or LC/MS results are considered the definitive ground truth, and device performance is measured against them.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described. The study focuses on the device's accuracy against laboratory gold standards and predicate devices, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study described is a standalone performance study of the device (UCP Drug Screening Buprenorphine, Amphetamine 300, Methamphetamine 500, Cocaine 150 Tests). These are rapid, qualitative, competitive binding immunoassays that can be performed "without the use of an instrument," implying visual interpretation of results. While human observation is part of reading the test, the performance being evaluated is that of the assay itself in detecting the target analytes in urine, not for an algorithm in an AI system.

    7. The Type of Ground Truth Used

    The type of ground truth used was confirmatory analytical methods: Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS).

    8. The Sample Size for the Training Set

    The document does not provide any information regarding a training set or its sample size. This type of device (rapid, qualitative immunoassay) typically does not involve machine learning or AI models that require specific training sets in the same way. The "training" for such devices is usually in the manufacturing process and quality control to ensure consistent chemical reactivity.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned for this type of immunoassay device, the information on how its ground truth was established is not applicable or provided in the document.

    Ask a Question

    Ask a specific question about this device