510(k) Data Aggregation

    K Number
    K251750

    Validate with FDA (Live)

    Device Name
    Trax EX Anchor
    Date Cleared
    2026-03-12

    (276 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. Specific indications are listed below:

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

    Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

    Device Description

    The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. The anchors are manufactured from 6AL-4V ELI Titanium. The Trax EX Anchor is a 2-component anchor compromised of the Trax EX Screw-In and the Trax EX Push-In Anchor with diameter sizes ranging from 3.5mm to 5.0mm. The Trax EX Screw-In Anchor is a hollow 6AL-4V ELI titanium anchor that is threaded, preloaded with suture and designed to fixate into bone by way of threaded engagement. The Trax EX Push-In Anchor component is a hollow 6AL-4V ELI titanium anchor that is ribbed, preloaded with suture, and designed to fixate into bone by way of press-fit engagement. Both anchors will be offered in multiple diameters, lengths, and variations in configuration of the distal end of the implant to allow further functional options to the surgeon. The Trax EX Screw-In Anchor is designed to be used alone or in conjunction with the Trax EX Push-In Anchor. The Trax EX Push-In Anchor component is designed to be used only in conjunction with the Trax EX Screw-In Anchor and is not sold separately. The sutures utilized in all implants are high-strength, Ultra High Molecular Weight Polyethylene (UHMWPE). The Trax EX Anchor Delivery Instruments are intended to prepare the site and fixate the anchors. All kits are intended to be sterilized by ethylene oxide (EO) and provided in a blister tray inside of a Tyvek pouch.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250712

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-14

    (65 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Linkt™ Compression Staple System is intended to be used for fracture and osteotomy fixation and joint arthrodesis of the hand and foot.

    Device Description

    The Linkt Compression Staple System consists of individually packaged, sterile, bone staple implants and a separate single use, sterile instrument set. The implants are manufactured from a nickel titanium alloy (Nitinol) which meets the requirements of ASTM F2063. The staples utilize chemical etching and passivation to form a protective oxidation layer on the outer surface. The staples are available in sizes ranging from 9mm x 9mm to 20mm x 20mm to allow the surgeon to select the appropriate device for the patient's anatomy.

    The single use instruments are intended to prepare the site and deploy the staple. The instrument kit includes a drill, drill guide, locating pins and an insertion tool.

    AI/ML Overview

    This FDA 510(k) clearance letter is for a physical medical device, the Linkt Compression Staple System, not a software or AI-based diagnostic device. Therefore, many of the typical acceptance criteria and study details relevant to AI/software (like sample size for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) are not applicable.

    The clearance is based on non-clinical performance testing to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Performance Area)Reported Device Performance (as demonstrated by testing)
    Pull-out StrengthDemonstrated substantial equivalence to predicates via ASTM F564 testing.
    Static 4-Point BendDemonstrated substantial equivalence to predicates via ASTM F564 testing.
    Dynamic 4-Point BendDemonstrated substantial equivalence to predicates via ASTM F564 testing.
    Corrosion SusceptibilityDemonstrated substantial equivalence to predicates via ASTM F2129 testing.
    Radio-Frequency Induced HeatingDemonstrated substantial equivalence to predicates via ASTM F2182 testing.
    Material (Nitinol)Meets requirements of ASTM F2063.
    Design and Indications for UseSimilar to predicates.
    Principles of OperationSame as predicates for bone fixation.
    Sterilization MethodGamma irradiation (same as primary predicate).

    2. Sample size used for the test set and the data provenance:
    This information is not provided as the evaluation was based on non-clinical, benchtop testing, not a dataset of patient cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not applicable as there was no test set requiring expert ground truth for diagnosis/interpretation. The device's performance was evaluated through standardized mechanical and material testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable, as there was no test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a surgical staple, not a diagnostic AI tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    Not applicable. The "ground truth" for this device's performance is established by its ability to meet specific engineering and material standards through validated testing methods (ASTM standards) and demonstrate mechanical properties equivalent to legally marketed predicate devices.

    8. The sample size for the training set:
    Not applicable. There is no "training set" for this physical device.

    9. How the ground truth for the training set was established:
    Not applicable.

    Study/Evidence that Proves the Device Meets Acceptance Criteria:

    The device claims to meet its acceptance criteria through non-clinical performance testing that demonstrates substantial equivalence to legally marketed predicate devices. The studies conducted were:

    • Pull out Strength Testing according to ASTM F564.
    • Static 4 Point Bend Testing according to ASTM F564.
    • Dynamic 4 Point Bend Testing according to ASTM F564.
    • Corrosion Susceptibility Testing according to ASTM F2129.
    • Radio-Frequency Induced Heating according to ASTM F2182.

    These tests are standard methods for evaluating the mechanical and material properties of bone fixation devices. The conclusion reached by Trax Surgical and accepted by the FDA is that based on the results of these tests, the Linkt Compression Staple System is substantially equivalent to the predicate devices and does not raise new questions of safety or effectiveness.

    Key takeaway for this specific device: The FDA clearance is based on the device conforming to established engineering standards and demonstrating equivalent physical performance to existing devices, rather than clinical efficacy or diagnostic accuracy, which would involve different types of studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230946

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-15

    (164 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accelerate Compression Screw System is indust and skeletally mature pediatic patients (aged 12-21 years), for fracture repair and fixation, osteotomy, joint fusion, reconstruction and arthrodesis of bones appropriate for the size of the device.

    Device Description

    The Trax Accelerate Compression Screw System is intended to be used to treat fracture and reconstruction of the small bones of the extremities (hands and feet). The screws are manufactured from 6AL-4V titanium and are anodized in various colors for easy size recognition. Washers will be provided for some screw sizes as appropriate. The fixation screws come in three (3) different configurations, headed, headless and break away. The headed and headless screws are cannulated. The breakaway screws are solid. They vary in diameter and length to allow the surgeon to select the appropriate device for the patient's anatomy. All devices are intended to be provided in an autoclavable instrument tray that includes a set of Class I instruments. The set will be provided non-sterile to the end user. The surgical instruments are intended to prepare the site and fixate the screws. The screws and instruments shall be designed to be maintained in an autoclavable tray. Replacement implants and instruments will be available separately as required.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Trax Surgical Accelerate Compression Screw System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies or AI algorithm evaluation.

    Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, reported device performance, sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or how ground truth was established.

    Instead, it addresses the following for the Accelerate Compression Screw System:

    • Indications for Use: For fracture repair and fixation, osteotomy, joint fusion, reconstruction, and arthrodesis of bones appropriate for the device size, in adult and skeletally mature pediatric patients (12-21 years).
    • Performance Data (Non-Clinical): Evaluated according to ASTM F543-17 and FDA guidance for Orthopedic Non-Spinal Metallic Bone Screws and Washers. Static torsion (torque to failure), Insertion/Removal torque, and Axial pullout testing were performed to demonstrate substantial equivalence to predicate devices.
    • Biocompatibility: Demonstrated using materials that meet applicable standards and/or are used in 510(k) cleared devices (K172383).
    • Technological Characteristics: Similar to predicate devices in indications for use, materials and coatings, function, and range of sizes.
    • Predicate Devices: Captivate Compression Screws (K222409), GEO Bone Screw System (K202817), Treace Medical Concepts Snap-off Screw System (K183363). A reference device is Arthrosurface Bone Screws (K172383).

    In summary, this document is a 510(k) submission and primarily relies on non-clinical performance data and technological characteristics to assert substantial equivalence to existing devices, not on human clinical trials or AI algorithm performance against acceptance criteria.