Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K233553

    Validate with FDA (Live)

    Date Cleared
    2024-06-18

    (225 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Model number 1024 is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1024 is intended for use only when connected to a 22 mm fitting.

    Model number 1024A is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1024A is intended for use only when connected to a 22 mm fitting.

    Model number 1025 is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1025 is intended for use only when connected to a 22 mm fitting.

    Model number 1025A is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1025A is intended for use only when connected to a 22 mm fitting.

    Model number 1543 is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1543 is intended for use only when connected to a 15 mm fitting.

    Model number 1543A is a single patient, disposable device for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner. The device is indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments. REF 1543A is intended for use only when connected to a 15 mm fitting.

    Device Description

    Model numbers 1024, 1025, 1543, 1024A, 1025A and 1543A are breathing circuit connectors that function as general purpose actuators for dispensing prescribed aerosolized medication from a pressurized metered dose inhaler (hereafter referred to as "pMDI") canister into a breathing circuit.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding a medical device called the MiniSpacer Dual Spray MDI Adapter. This document is a regulatory submission for a device, which typically describes an engineering verification and validation study rather than a clinical study for AI software.

    Therefore, the requested information regarding acceptance criteria and performance against those criteria, especially details such as sample size for test sets, data provenance, expert ground truth adjudication, MRMC studies, standalone algorithm performance, and training set information (which are all highly relevant to AI/ML device evaluations), is not present in this document. The document primarily focuses on demonstrating substantial equivalence to predicate devices based on design, materials, intended use, and general performance, rather than AI-specific metrics.

    However, I can extract the following relevant information based on the request, reinterpreting some categories in the context of a non-AI medical device submission:

    Acceptance Criteria and Device Performance

    • Product: MiniSpacer® 1024, 1025, 1543 and 1024A, 1025A, 1543A Dual Spray MDI Adapter
    • Study Type: This document describes a premarket notification (510(k)) to establish substantial equivalence to predicate devices, not a clinical trial or AI performance study. The "study" here refers to the engineering verification and biocompatibility assessment for the physical device.
    Acceptance Criterion (Standard or Predicate Equivalence)Reported Device Performance
    Material Equivalence:
    - Body Material: Modified SBC performance equivalent toToxicological assessment supports substantial equivalence and
    Legacy SBC.biological safety for the intended device population compared to
    - Other Materials (Actuator, Cap): ABS, SEBS equivalentpredicate devices.
    to predicate.
    Functional Equivalence:The subject device shares the same intended use, indications for
    - Dispensing pMDI medication into breathing circuit.use, common name, classification, and regulatory status as the
    - Single patient, disposable.predicate devices. It also maintains the same dimensions and dual
    - Prescription device.spray orifice nozzle design.
    - Non-sterile.
    Biocompatibility:
    - Meeting biocompatibility standards for externalA thorough toxicological assessment was performed, and the data
    communicating, tissue, long-term (> 30 days) contact.supports substantial equivalence and biological safety for the
    intended device population, despite a change in the body resin.

    Additional Requested Information (as applicable to this document):

    1. Sample sizes used for the test set and data provenance:

      • Test Set Sample Size: Not specified in the document. The testing described appears to be bench testing for material and functional equivalence.
      • Data Provenance: Not explicitly stated, but the toxicological assessment and comparisons against predicate devices would be based on internal testing and regulatory standards. It's a regulatory submission, not a clinical trial, so "country of origin" of data or "retrospective/prospective" does not apply in the typical sense for patient data.
    2. Number of experts used to establish the ground truth for the test set and their qualifications:

      • Not applicable in the context of this device. Ground truth is established through engineering specifications, material standards, and regulatory requirements (e.g., biocompatibility testing interpreted by toxicologists) rather than expert consensus on medical images or patient outcomes.
    3. Adjudication method for the test set:

      • Not applicable. This is not a study requiring adjudication of expert interpretations. Compliance is determined by engineering testing and regulatory assessment.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size:

      • No, an MRMC study was not done. This device is a physical medical component (an actuator for MDI medication), not an AI diagnostic or treatment planning system that would involve human readers interpreting data.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device does not involve an algorithm or AI component.
    6. The type of ground truth used:

      • For material and functional performance, the "ground truth" is defined by established engineering specifications, performance standards, mechanical testing results, and toxicological assessment standards. For regulatory approval, the ultimate "ground truth" is established by demonstrating substantial equivalence to legally marketed predicate devices through comparative analysis and testing.
    7. The sample size for the training set:

      • Not applicable. This device does not involve machine learning or a training set.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210558

    Validate with FDA (Live)

    Date Cleared
    2022-04-01

    (400 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LiteAire is a collapsible, disposable dual-valved holding chamber designed to aid in the delivery of aerosolized medications delivered via a pressurized metered dose inhaler (MDI).

    The LiteAire features a standard port designed for compatibility with standard MDI mouthpieces. It is a non-sterile device for single-patient use.

    The LiteAire is intended to be used by adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healtheare professional.

    Device Description

    The LiteAire® BASIC Dual Valved, Collapsible MDI Holding Chamber (also referred to as the LiteAire® BASIC) and the modified LiteAire® Dual Valved, Collapsible MDI Holding Chamber (also referred to as the modified LiteAire®), are intended for use in the inhalation of medications delivered via a pressurized metered dose inhaler (pMDI). The subject devices feature a universal port designed for compatibility with most MDI medications. The subject devices are intended for use by a single patient and, when properly cared for, are reusable for up to one week. The devices consist of a collapsible paperboard housing and two one-way valves to control the direction of air flow when the patient inhales and exhales through the devices are popped-up by the user prior to use by pressing against the sides of the devices, can be collapsed flat between uses, and are anti-static. The intended environments of use include the home, hospitals and clinics. Note that the only differences between the modified LiteAire® and LiteAire® BASIC configurations is the removal of the window.

    AI/ML Overview

    The provided text describes modifications to an existing medical device, the LiteAire Dual-Valved, Collapsible MDI Holding Chamber, and presents testing to demonstrate its substantial equivalence to the predicate device. It does not describe an AI/ML powered device, therefore the response below is based on the available information and identifies when specific AI/ML related questions are not applicable.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the modified devices (LiteAire BASIC and Modified LiteAire) is to demonstrate substantial equivalence to the predicate device (LiteAire K160109) in terms of aerosol characterization (particle size distributions) and other performance aspects. The tables below summarize the key aerosol characterization data.

    Table 1: Acceptance Criteria and Reported Device Performance (Aerosol Characterization at Adult Flow Rates - 28.3 L/min)

    Parameter (MDI Medication)Acceptance Criteria (Implicit: No statistically significant difference from Predicate)Predicate LiteAire (N=9) Mean ± SD, 95% CILiteAire BASIC (N=9) Mean ± SD, 95% CIModified LiteAire (N=9) Mean ± SD, 95% CITest Result/Conclusion
    Albuterol Sulfate
    Total Emitted Dose (µg)Statistically similar to predicate56.06 ± 6.94, 50.73–61.4054.75 ± 4.96, 50.94–58.4553.85 ± 2.66, 51.81–55.90Deemed substantially equivalent
    Fine Particle Dose (µg)Statistically similar to predicate56.06 ± 6.94, 50.73–61.4054.66 ± 4.99, 50.83–58.5053.85 ± 2.57, 51.72–55.67Deemed substantially equivalent
    MMAD (µm)Statistically similar to predicate2.15 ± 0.05, 2.12–2.192.17 ± 0.05, 2.14–2.212.24 ± 0.06, 2.19–2.28Deemed substantially equivalent
    GSD (µm)Statistically similar to predicate1.44 ± 0.07, 1.39–1.491.43 ± 0.08, 1.37–1.491.41 ± 0.02, 1.39–1.43Deemed substantially equivalent
    Ipratropium Bromide
    Total Emitted Dose (µg)Statistically similar to predicate7.39 ± 0.55, 6.96–7.817.46 ± 0.61, 6.99–7.927.71 ± 0.62, 7.10–8.25Deemed substantially equivalent
    Fine Particle Dose (µg)Statistically similar to predicate7.39 ± 0.55, 6.96–7.817.46 ± 0.61, 6.99–7.927.71 ± 0.62, 7.10–8.25Deemed substantially equivalent
    MMAD (µm)Statistically similar to predicate0.51 ± 0.06, 0.46–0.550.57 ± 0.14, 0.46–0.670.60 ± 0.09, 0.53–0.66Deemed substantially equivalent
    GSD (µm)Statistically similar to predicate4.79 ± 0.70, 4.26–5.333.90 ± 1.81, 2.51–5.293.44 ± 0.95, 2.71–4.17Deemed substantially equivalent
    Fluticasone Propionate
    Total Emitted Dose (µg)Statistically similar to predicate40.96 ± 5.23, 36.94–44.9841.27 ± 6.33, 36.41–46.1438.43 ± 11.26, 28.72–44.55Deemed substantially equivalent
    Fine Particle Dose (µg)Statistically similar to predicate40.96 ± 5.23, 36.94–44.9841.27 ± 6.33, 36.41–46.1438.43 ± 11.26, 29.78–47.09Deemed substantially equivalent
    MMAD (µm)Statistically similar to predicate2.49 ± 0.17, 2.358–2.6192.52 ± 0.03, 2.50–2.542.54 ± 0.06, 2.50–2.59Deemed substantially equivalent
    GSD (µm)Statistically similar to predicate1.39 ± 0.02, 1.38–1.411.41 ± 0.03, 1.39–1.431.46 ± 0.05, 1.426–1.499Deemed substantially equivalent

    (Note: "Deemed substantially equivalent" is concluded from the 510(k) summary statement that "the modification in design of the subject devices does not negatively impact the device performance and the modified (subject) devices continues to perform with substantial equivalence to the cleared (predicate) device." Statistical methods for "statistically similar" are not explicitly detailed but are implied by the comparison of means and confidence intervals.)

    Table 2: Other Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceTest Result/Conclusion
    Biocompatibility (per ISO 10993-1)Confirmed using FDA guidance document, no new risks identified due to material or process changes. Materials, manufacturing processes, finished device geometry, and body/fluid contact characterization are the same as the predicate (K160109).Met
    Visual InspectionAll tests passedPassed
    First Article InspectionAll tests passedPassed
    Accelerated AgingAll tests passedPassed
    Pop/Collapse functionalityAll tests passedPassed
    Functional Equivalence (purpose, function, scientific technology, method of operation)"The subject devices are substantially equivalent to the predicate device (K160109) in purpose, function, scientific technology and method of operation."Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Aerosol Characterization: N=9 for each device (Predicate LiteAire, LiteAire BASIC, Modified LiteAire) for each MDI medication and flow rate tested (adult and pediatric). This means 9 units of each device type were tested for each specific drug and inhalation profile combination.
    • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective. Given that it's a bench performance test for a medical device's physical properties, the concept of "retrospective or prospective" as applied to clinical studies is not directly applicable. These are laboratory-based, controlled experiments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is Not Applicable as the device is a medical device (MDI Holding Chamber), not an Artificial Intelligence/Machine Learning (AI/ML) powered device that relies on expert interpretation for ground truth establishment. The performance is measured through objective physical and chemical tests (e.g., aerosol particle size, emitted dose).

    4. Adjudication Method for the Test Set

    This section is Not Applicable as adjudication methods (like 2+1, 3+1) are typically used in clinical studies or for establishing ground truth in AI/ML performance evaluations involving human readers. The tests performed here are objective bench tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This section is Not Applicable as the device is a medical device (MDI Holding Chamber), not an AI/ML powered device, and no human-in-the-loop performance or reader studies were conducted or described.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is Not Applicable as the device is a physical medical device, not an algorithm or AI/ML product.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance is established through objective physical and chemical measurements obtained from standardized laboratory tests, specifically:

    • Aerosol Characterization: Particle size distributions (MMAD, GSD), Total Emitted Dose, and Fine Particle Dose measured using an Andersen Cascade Impactor (ACI) as per FDA guidance.
    • Biocompatibility: Conformance to ISO 10993-1, confirmed by material composition and manufacturing process identical to a previously cleared device.
    • Bench Performance: Visual inspection, first article inspection, accelerated aging, and pop/collapse functionality.

    The performance values obtained from the predicate device served as the reference for establishing substantial equivalence.

    8. The Sample Size for the Training Set

    This section is Not Applicable as the device is a traditional medical device, not an AI/ML powered device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This section is Not Applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160109

    Validate with FDA (Live)

    Date Cleared
    2017-05-19

    (486 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LiteAire is a collapsible, disposable dual-valved holding chamber designed to aid in the delivery of aerosolized medications delivered via a pressurized metered dose inhaler (MDI).

    The LiteAire features a standard port designed for compatibility with standard MDI mouthpieces. It is a non-sterile device for single-patient use.

    The LiteAire is intended to be used by adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healthcare professional.

    Device Description

    The Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding Chamber is intended for use in the inhalation of medications delivered via a pressurized metered dose inhaler (pMDI). The device consists of a collapsible paperboard housing and 2 one-way valves to control the direction of air flow when the patient inhales and exhales through the device. The LiteAire is not sterile, but clean and ready to use right out of the package. The LiteAire can be assembled by gently pushing in the edges of the device. The holding chamber can also be collapsed flat between uses and is anti-static.

    AI/ML Overview

    The provided text describes modifications to a medical device (LiteAire Dual-Valved, Collapsible MDI Holding Chamber) and evaluates its substantial equivalence to a predicate device. The information presented focuses on non-clinical performance testing rather than a clinical study involving human readers or AI.

    Here's an analysis of the provided text in the context of your questions:

    1. A table of acceptance criteria and the reported device performance.

    The document discusses performance in the context of non-clinical testing, specifically regarding the aerosolization efficiency of the device with different MDI products. The "acceptance criteria" are implied by the comparison to the predicate and the general expectation for an MDI holding chamber to deliver appropriate doses. The "reported device performance" is given in tables for Total Delivered Dose, Coarse Particle Dose, Fine Particle Dose, MMAD, and GSD.

    Acceptance Criteria (Implied / Functional Requirements):

    • Aerosol Delivery Performance: The ability to effectively deliver the prescribed dose from an MDI, with appropriate particle size distribution, comparable to or better than predicate devices. This is assessed via "Total Emitted Dose, Course Particle Dose or Fine Particle Dose."
    • Mechanical Integrity: Maintain structural integrity and functionality (pop-ability, collapsibility, valve resistance) over expected shelf-life and under environmental stresses.
    • Biocompatibility: No adverse biological reactions when in contact with lung tissue, skin, and mucosal membranes.
    • Anti-static Properties: Exhibit anti-static properties to minimize drug adherence to the chamber walls.

    Reported Device Performance:

    The document provides specific data in Tables 2 and 3 for the subject device (K160109) with three different MDI products (Proventil HFA, Atrovent HFA, QVAR 80 mcg) at adult (28 L/min) and pediatric (12 L/min) flow rates. The values reported are 95% confidence intervals.

    Table 1: Device Performance (Excerpts from Tables 2 & 3 Combined)

    Performance MetricMDI Product Tested (Example Range)Flow Rate
    Total Delivered Dose (µg/actuation)39.77 – 77.62Adult/Pediatric
    Coarse Particle Dose (>4.7 µm) (µg/actuation)0 – 3.42Adult/Pediatric
    Fine Particle Dose (<4.7 µm) (µg/actuation)39.81 – 75.34Adult/Pediatric
    MMAD (µm)0.48 – 2.44Adult/Pediatric
    GSD (µm)0 – 6.08Adult/Pediatric

    Note: The specific ranges reflect the minimum and maximum observed across all MDI products and flow rates for each metric in the tables.

    Additional Performance Points:

    • Shelf-life testing: "There were no differences in appearance, pop-ability and collapsibility, Total Emitted Dose, Course Particle Dose or Fine Particle Dose between the new and aged devices. Furthermore, the inhalation and exhalation valve resistances of both devices were found to be acceptably low."
    • Environmental and mechanical testing (Temperature and Humidity): "There were no differences in Total Emitted Dose, Course Particle Dose, Fine Particle Dose and breakability upon dropping between the new devices and devices exposed to high temperature and humidity. Furthermore, the inhalation and exhalation valve resistances of 'exposed' devices were found to be acceptably low."
    • Anti-static surface resistivity testing: "The LiteAire has a surface resistivity of 1.0 x 10^10 Ω/sq (at 23°C/ 50% RH). This meets the requirement for being 'dissipative' or 'anti-static'."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective).

    • Sample Size: For Particle Characterization testing (Tables 2 & 3), the sample size was:
      • 3 LiteAire devices / 3 ACI replicates each = 9 ACI runs/drug for each MDI product (Proventil HFA, Atrovent HFA, QVAR 80 mcg) and each flow rate (Adult 28 L/min, Pediatric 12 L/min).
    • Data Provenance: The document does not specify the country of origin for the data. The data appears to be prospective wet-lab testing conducted for the purpose of the 510(k) submission, not data collected from a clinical patient population.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience).

    This question is not applicable to the provided document. The "ground truth" for this device's performance is established through physical and chemical measurements (e.g., cascade impaction for particle size, resistance measurements for valves, surface resistivity) and biocompatibility testing, not through expert human interpretation of medical images or patient outcomes. Therefore, no experts in the sense of medical diagnosticians were used to establish "ground truth" for this type of performance study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set.

    This question is not applicable. Adjudication methods are typically used in studies involving human interpretation or subjective assessments to resolve discrepancies. The performance tests described here are objective, quantifiable measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    This question is not applicable. The study described is a non-clinical performance evaluation of a mechanical medical device (MDI holding chamber), not a study involving AI assistance to human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.

    This question is not applicable. There is no algorithm or AI involved in the performance evaluation described. The device is a physical, non-electronic medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc).

    The "ground truth" for this device performance evaluation is based on objective physical and chemical measurements performed in a laboratory setting, in accordance with recognized standards (e.g., ISO 10993 for biocompatibility, ASTM-D-257 for anti-static properties, and implicit standards for aerosol characterization via cascade impaction). This is not a clinical ground truth derived from patient outcomes, pathology, or expert consensus. It's a "ground truth" of the device's functional characteristics.

    8. The sample size for the training set.

    This question is not applicable. This is a performance study of a physical device, not an AI model. Therefore, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established.

    This question is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111570

    Validate with FDA (Live)

    Date Cleared
    2011-09-29

    (115 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MINISPACER® 1024A, 1025A and 1543A adapters are single patient, disposable devices for dispensing pressurized metered dose inhaler (pMDI) medication into a breathing circuit, as prescribed by a physician or other licensed health care practitioner.

    The MINISPACER® 1024A, 1025A and 1543A adapters are indicated for patients on a breathing circuit, for whom aerosol medication has been prescribed, in short and long term critical care environments.

    Device Description

    Not Found

    AI/ML Overview

    My deepest apologies, but the provided text does not contain specific acceptance criteria or the details of a study that proves a device meets those criteria. The document is an FDA 510(k) clearance letter for the MiniSpacer® 1024A, 1025A, and 1543A MDI Adapters.

    This type of document is a regulatory approval, stating that the device is "substantially equivalent" to legally marketed predicate devices. It confirms that the FDA has reviewed the manufacturer's submission and found it acceptable for market, but it does not typically include:

    • A table of acceptance criteria and reported device performance.
    • Details about sample size, data provenance, expert qualifications, or adjudication methods for a test set.
    • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or standalone algorithm performance.
    • Specifics about the type of ground truth used or the size and establishment of ground truth for a training set.

    These details would normally be found in the V&V (Verification and Validation) reports or clinical study summaries submitted by the manufacturer to the FDA as part of their 510(k) application, but they are not typically included in the public-facing clearance letter itself.

    Therefore, I cannot fulfill your request for this specific document. If you have a different document that details performance studies, I would be happy to analyze it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1