510(k) Data Aggregation
(140 days)
The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. It may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors.
The Midline Catheter is a peripherally inserted catheter composed of a radiopaque polyurethane catheter, a catheter slider, a catheter hub, needle hub, and 3D-Shin™ needle featuring an engineered three-dimensional geometry of the needle tip, with a passive safety mechanism. After observing the blood return, the 3D-Shin™ needle is withdrawn, and the needle tip is fully covered by the safety mechanism, reducing the risk of accidental needlestick injury.
N/A
Ask a specific question about this device
(139 days)
SURFLO Hybria Closed System Safety IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. The device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. When the device is used with a power injector, the needle-free connector should be removed, and the maximum pressure limit is 325 psi (2.24MPa).
SURFLO Hybria Closed System Safety IV Catheter is an over-the-needle intravascular catheter. It primarily includes a catheter, a 3D-Shin™ needle with a safety shield, double wings, an extension tube, a luer adapter, vent fitting(s), and a pre-attached SURPLUG™ AD needle-free connector (for some configurations).
The double wings of the SURFLO Hybria Closed System Safety IV Catheter can be folded and pinched together for puncture and catheter advancement. The distance from the grip position to the needle tip is short, which facilitates vascular puncture. The 3D-Shin™ needle features an engineered three-dimensional geometry of the needle tip. The needle tip is completely covered by the safety shield when it is withdrawn after visualization of blood return, which minimizes the risk of accidental needlestick injury. The closed design can prevent blood leakage throughout the entire puncture process.
N/A
Ask a specific question about this device
(403 days)
The Surflo Winged Infusion Set is indicated for venipuncture and intravenous administration of fluids for up to 24 hours using compatible/appropriate disposable infusion devices for medical purposes. It may be used for any patient population with consideration given to patient size.
The Surflo Winged Infusion Set consists of a protector, needle, wing, tubing, connector, and connector cap. The connector is 6% female Luer. It is primarily made from medical grade PVC (the whole device uses TOTM as plasticizer instead of DEHP) and medical grade stainless steel. The Surflo Winged Infusion Set is operated manually. The device is intended to be used by trained healthcare professionals in a healthcare facility. The Surflo Winged Infusion Set is available in different gauge sizes and tubing lengths. The Surflo Winged Infusion Set is a disposable device intended for single use only. The device is packaged and sterilized by ethylene oxide gas.
This document is a 510(k) Premarket Notification for the Surflo Winged Infusion Set, seeking substantial equivalence to a legally marketed predicate device. As such, it describes performance testing conducted to demonstrate product safety and effectiveness, rather than a study proving a device meets acceptance criteria for an AI/ML product. Therefore, several points in your requested output, such as those related to AI/ML specific criteria (e.g., sample size for training set, number of experts for ground truth establishment, MRMC study, standalone performance) are not applicable to this type of medical device submission.
However, I can extract and present the relevant information regarding acceptance criteria and performance testing for this traditional medical device:
Acceptance Criteria and Device Performance for Surflo Winged Infusion Set
The Surflo Winged Infusion Set underwent non-clinical performance testing to ensure it met predetermined specifications, conformed to applicable ISO and non-ISO standards, and demonstrated substantial equivalence to its predicate device.
1. Table of Acceptance Criteria (Standard Compliance) and Reported Device Performance
For tests conducted against specific standards, the document implies that compliance with the standard is the acceptance criterion. For additional performance tests, the acceptance criterion was "Meets acceptance criteria," established based on the predicate device's performance.
| Test | Standard / Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Flow rate | GB 18671-2009 Intravenous needles for single use | Complies with standard |
| Leak test | JIS T 3222:2011 Sterile winged intravenous devices | Complies with standard |
| Wing to tubing connection strength (static pull force) | ISO 8536-4:2019 Infusion equipment for medical use | Complies with standard |
| Tubing to connector connection strength (static pull force) | ISO 8536-4:2019 Infusion equipment for medical use | Complies with standard |
| Needle to wing connection strength (static pull force) | ISO 8536-4:2019 Infusion equipment for medical use | Complies with standard |
| Needle curvature | JIS T 3222:2011 Sterile winged intravenous devices | Complies with standard |
| Wing to tubing connection strength (dynamic pull force) | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Tubing to connector connection strength (dynamic pull force) | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Needle to wing connection strength (dynamic pull force) | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Wing-needle protector fit | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Needle penetration resistance | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Blockage test | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
| Priming volume | Meets acceptance criteria (based on predicate) | Meets acceptance criteria |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each performance test. It only mentions that "Performance testing was conducted to ensure the Surflo Winged Infusion Set met the predetermined specifications." The data provenance is implied to be from the manufacturer's internal testing facilities in Hangzhou, China, as it is manufactured by "Terumo Medical Products (Hangzhou) Co., Ltd." The testing is described as non-clinical, implying a prospective testing approach conducted specifically for this 510(k) submission.
3. Number of Experts Used and Qualifications
Not applicable. This is a non-clinical performance test for a traditional medical device, not a study requiring expert readers for ground truth establishment.
4. Adjudication Method for the Test Set
Not applicable. This is a non-clinical performance test for a traditional medical device, not a study requiring adjudication of expert readings.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is a non-clinical performance test for a traditional medical device, not a study evaluating human reader performance with or without AI assistance.
6. Standalone Performance (Algorithm Only)
Not applicable. This is a traditional medical device, not an AI/ML algorithm.
7. Type of Ground Truth Used
The "ground truth" for the performance tests were adherence to established industry standards (ISO, JIS, GB) and internal company specifications derived from the predicate device's performance.
8. Sample Size for the Training Set
Not applicable. This is a traditional medical device, not an AI/ML product requiring a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. This is a traditional medical device, not an AI/ML product requiring a training set.
Additional Test Information:
- Biocompatibility: Tests were conducted on non-aged and accelerated-aged devices according to ISO 10993-1, with results demonstrating biocompatibility throughout the product's shelf life.
- Sterilization: The device is sterilized to a Sterility Assurance Level (SAL) of 10^-6, validated in accordance with ISO 11135:2014.
- Clinical Tests: The 510(k) submission states that it "does not include data from clinical tests."
Ask a specific question about this device
(90 days)
RADIFOCUS Torque Device is intended to facilitate guidewire manipulation during interventional procedures.
RADIFOCUS Torque Device is a supportive device intended to facilitate guide wire manipulation during interventional procedures. This product consists of a cap that screws onto the main body of the device (handle) to apply a clamping force on the guide wire. The cap is made of polypropylene and the handle is made of polyoxymethylene. The torque device can be used for the guide wires with diameters of 0.010" to 0.038". RADIFOCUS Torque Device is sterilized with ethylene oxide.
The provided text describes a 510(k) submission for the RADIFOCUS Torque Device, a medical device intended to facilitate guidewire manipulation. It specifically states that no clinical tests were performed or deemed necessary for this submission. The evaluation relies on non-clinical performance testing and comparison to a predicate device.
Therefore, the requested information regarding acceptance criteria and a study proving device performance cannot be fully provided in the context of a clinical study with human subjects, given that no such study was conducted or required.
However, based on the provided text, I can infer the "acceptance criteria" and "study" refers to the non-clinical performance testing conducted.
Here's the breakdown of what can be extracted from the provided text regarding the non-clinical performance testing:
1. A table of acceptance criteria and the reported device performance:
The document states: "The RADIFOCUS Torque Device tested met the predetermined acceptance criteria." While the specific numerical or qualitative acceptance criteria for each test are not detailed, the results are uniformly positive.
| Performance Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Appearance | Meets predetermined criteria | Meets acceptance criteria |
| Fixed strength of guide wire | Meets predetermined criteria | Meets acceptance criteria |
| Fixed strength of cap | Meets predetermined criteria | Meets acceptance criteria |
| Fixable dimension | Meets predetermined criteria | Meets acceptance criteria |
| Guide wire damage | Meets predetermined criteria | Meets acceptance criteria |
| Collet release | Meets predetermined criteria | Meets acceptance criteria |
| Torque slip force | Meets predetermined criteria | Meets acceptance criteria |
2. Sample size used for the test set and the data provenance:
- Sample size: Not specified. The document only mentions "Performance testing was conducted."
- Data provenance: Not specified. It's implied to be internal testing by the manufacturer (Terumo Medical Products (Hangzhou) Co., Ltd.) given the context of a 510(k) submission and the lack of external study details. The testing is retrospective in the sense that it was conducted before the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This is not applicable as there was no clinical study involving human subjects or expert assessment of clinical outcomes. The "ground truth" for the non-clinical performance tests would be the established engineering specifications and measurement standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This is not applicable for non-clinical performance testing of a medical device. Adjudication methods are typically used in clinical studies for interpretation of imaging or clinical events.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is not applicable. The device is a physical torque device, not an AI-assisted diagnostic tool. No MRMC study was mentioned or implied.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This is not applicable as the device is not an algorithm or AI. The performance testing was for the physical device.
7. The type of ground truth used:
- For the non-clinical tests, the "ground truth" was established by predetermined specifications and engineering standards for mechanical performance (e.g., strength, dimensions, torque).
8. The sample size for the training set:
- This is not applicable as there was no AI/algorithm that required a training set. This is a physical medical device.
9. How the ground truth for the training set was established:
- This is not applicable for the same reason as above.
Ask a specific question about this device