510(k) Data Aggregation

    K Number
    K181732

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-27

    (299 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InterSpace Knee is indicated for temporary use (maximum 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.

    InterSpace Knee is applied on the femoral condyles and on the tibial plate following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection).

    InterSpace Knee is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethylmethacrylate), the is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period

    InterSpace Knee ATS is indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process, and where a large tibial defect is present.

    InterSpace Knee ATS is applied on the tibial plate following removal of the existing implant and radical debridement. The device must be combined with the tibial component of InterSpace Knee. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).

    InterSpace Knee ATS is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethacrylate), the Interspace Knee ATS is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.

    Device Description

    The InterSpace Knee, cleared by FDA via K101356 and K062274, is a temporary spacer device available in three sizes (S-M-L). With this 510(k) submission, Tecres is adding the additional XL size to the FDA cleared range of InterSpace Knee sizes.

    The InterSpace Knee ATS is a temporary spacer device available in four sizes, intended to be combined with the tibial component of InterSpace Knee. Specific sizes of the InterSpace Knee ATS are compatible with certain InterSpace Knee sizes.

    The devices, as single device (InterSpace Knee, K101356 and K062274) or once InterSpace Knee is combined with the InterSpace Knee ATS, provide patients with a temporary, complete knee implant that allows for a natural range of motion and partial weight-bearing during treatment of the infection, and preserves the soft tissue to prevent further complications, such as muscular contraction, to facilitate the subsequent joint replacement procedure after systemic treatment of the underlying infection.

    The InterSpace Knee XL size and InterSpace Knee ATS components are sterile, single-use device intended for temporary use (maximum 180 days) as a partial joint replacement. The devices are made of fully formed polymethylmethacrylate (PMMA), which is radio-opaque and contains gentamicin. The mass used to fill the molds (the unformed PMMA resin) is prepared from powder and liquid components. The liquid component consists of methylmethacrylate (MMA), N,N- dimethyl-p-toluidine and hydroquinone; the powder component consists of PMMA, barium sulphate, benzoyl peroxide and gentamicin sulphate.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA, and it describes a medical device called "InterSpace Knee Extra-Large Size" and "InterSpace Knee ATS." This document details the device's indications for use, its characteristics, and how it compares to predicate devices to establish substantial equivalence.

    Crucially, this document focuses on establishing the substantial equivalence of new sizes/components of an existing knee spacer device based on engineering and material performance criteria, NOT on clinical diagnostic performance criteria for an AI/ML-driven device.

    Therefore, most of the requested information regarding acceptance criteria and studies for AI/ML device performance (e.g., sample sizes for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) is not applicable to this document. This submission is for a physical medical implant, not a diagnostic or AI-powered device.

    However, I can extract the relevant "acceptance criteria" (in the context of engineering performance) and reported performance from the document.

    Acceptance Criteria and Reported Device Performance (Engineering/Material Focus):

    The "acceptance criteria" in this context refer to passing various engineering and material tests, often by meeting the requirements of recognized standards (ISO, ASTM) or by demonstrating equivalence to a predicate device that has previously met these standards. The "reported device performance" indicates that the device did meet these criteria.

    Acceptance Criteria (Test/Standard)Reported Device Performance (Met/Compared To)
    InterSpace Knee (XL size):
    Static performances of the resin (ISO 5833)Resin was tested and shown to meet the minimum requirements of ISO 5833.
    Fatigue performances of the resin (ASTM F2118)Resin was tested and shown to meet the acceptance criteria according to ASTM F2118.
    Fatigue test (comparison with other sizes)The bigger size (XL) has been demonstrated to be a better case compared to the other InterSpace Knee sizes (S-M-L). (This implies it performed as well or better, serving as a worst-case scenario validation for the existing predicate's fatigue performance).
    Wear test (debris evaluation)An evaluation of the debris has been compared with another temporary spacer already cleared by FDA. (Implies comparable wear characteristics).
    Surface Roughness (comparison with other sizes)Surface roughness is the same of the other sizes already approved.
    Antibiotic (gentamicin) elution testing (comparison)The gentamicin elution profile demonstrated substantial equivalence with gentamicin elution of another temporary spacer already cleared by FDA.
    Sterilization validation (e.g., UNI EN ISO 11135:2014)Complies with UNI EN ISO 11135:2014; EC 1-2011 UNI EN ISO 11737-1:2006, UNI EN ISO 11737-2:2010, UNI EN ISO 10993-7:2009, UNI EN 556-1:2002, EP current Edition. Validated to a sterility assurance level of 10-6.
    Shelf Life (e.g., real-time and accelerated aging)Established through real-time and accelerated aging studies. All demonstrate a shelf life of 5 years.
    Biocompatibility (ISO 10993 and FDA Draft Guidance)Evaluation performed to show device materials are safe, biocompatible, and suitable for intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" have been taken into account. (Implies compliance/satisfaction of these guidelines).
    InterSpace Knee ATS (additional criteria where applicable):
    Static performances of the resin (ISO 5833)Resin was tested and shown to meet the minimum requirements of ISO 5833.
    Fatigue performances of the resin (ASTM F2118)Resin was tested and shown to meet the acceptance criteria according to ASTM F2118.
    Fatigue test on the device (ASTM F1800)Has been performed according to ASTM F1800. (Presumably, it passed, or the comparison results were favorable for substantial equivalence).
    Antibiotic (gentamicin) elution testing (comparison)The gentamicin elution profile demonstrated substantial equivalence with gentamicin elution of another temporary spacer already cleared by FDA.
    Sterilization validation & Shelf Life & Biocompatibility (same as XL)Met the same criteria as the XL size, given the shared materials and manufacturing processes.

    Since this document is for a physical orthopedic implant and not an AI/ML diagnostic tool, the following points are not applicable and cannot be answered from the provided text:

    1. Sample sized used for the test set and the data provenance: Not applicable to a physical device performance study of this nature. Performance tests are typically conducted on a minimal number of samples to demonstrate compliance with standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML diagnostic is typically based on clinical expert consensus or pathology; for this device, "ground truth" refers to physical and chemical properties confirmed by laboratory testing.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable in the context of diagnostic performance. "Ground truth" here is the adherence to material and mechanical specifications.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122175

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-28

    (248 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Mendec Spine HV / Mendec Spine HV System is indicated for the treatment of pathological fractures of the vertebral body using a vertebroplasty or kyphoplasty procedure. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

    Device Description

    Mendec Spine HV and Mendec Spine HV System are highly viscous, radioopaque acrylic resins (PMMA based) for percutaneous vertebroplasty or kyphoplasty. These devices are made with the same raw material, but are supplied in different ways: . Mendec Spine HV is a traditional bone cement product: the liquid is contained in a vial and the powder in a sachet; these components are packaged in unitary PVC-blister with Tyvek lid, which is placed in an aluminum bag. . Mendec Spine HV System holds the powder and liquid components separately within a closed syringe-like device haa serves as a mixing chamber. The device is packaged in unitary PVC-blister with tray, sealed with Tyvek lid, which is placed in an aluminum bag. The devices are sold disposable and sterile.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Tecres Mendec Spine HV and Mendec Spine HV System, which are polymethylmethacrylate (PMMA) bone cements for vertebroplasty/kyphoplasty. This type of submission relies on demonstrating substantial equivalence to predicate devices, rather than a novel study proving a device meets specific clinical acceptance criteria in the way an AI/ML device would.

    Therefore, many of the requested categories (e.g., sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable to this type of device submission.

    Here's how the available information maps to your request:

    1. A table of acceptance criteria and the reported device performance

    For a PMMA bone cement, acceptance criteria are typically defined by demonstrating that the device has comparable material and mechanical properties to legally marketed predicate devices.

    Acceptance Criteria CategoryReported Device Performance
    Mechanical PropertiesTested (static and fatigue properties, viscosity) and found substantially equivalent to predicate devices.
    Chemical-Physical PropertiesTested (setting-time, MMA release) and found substantially equivalent to predicate devices.
    Intended Use EquivalenceThe intended use is the same as the predicate devices: treatment of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty.
    Material EquivalenceComposed of the same raw material as predicate devices.

    2. Sample size used for the test set and the data provenance

    Not applicable for this type of device. The "test set" here refers to the actual physical devices undergoing laboratory testing for their material and mechanical properties, not a clinical data set. The provenance of the data is from Tecres S.P.A.'s own performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth, in the context of device performance, relates to laboratory measurements and conformity to recognized standards or comparison to predicate device specifications, not expert interpretation of clinical data.

    4. Adjudication method for the test set

    Not applicable. This is not a clinical study involving human readers or interpretations needing adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device (bone cement), not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device is based on established material science and engineering principles, and comparison to the specifications and performance of legally marketed predicate devices. This includes:

    • Physical measurements of mechanical properties (e.g., strength, viscosity).
    • Chemical analysis of material composition and properties (e.g., setting time, monomer release).
    • Verification that the device meets safety and performance standards equivalent to the predicates.

    8. The sample size for the training set

    Not applicable. There is no AI/ML model, hence no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI/ML model, hence no training set or ground truth establishment method for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112983

    Validate with FDA (Live)

    Device Name
    SPACER-S
    Manufacturer
    Date Cleared
    2011-12-12

    (67 days)

    Product Code
    Regulation Number
    888.3690
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Interspace Shoulder is indicated for temporary use (maximum of 180 days) as a shoulder replacement (SR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. The head and stem are inserted into the glenoideal cavity and the humeral medullary canal, respectively, following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). Interspace Shoulder is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.).

    Device Description

    The Interspace Shoulder devices is a temporary device composed of fully formed PMMA bone cement with gentamicin and an inner stainless steel metal core. The design mimics a hemi-shoulder prosthesis.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Tecres Interspace Shoulder Temporary Shoulder Prosthesis:

    Based on the provided 510(k) summary (K112983), this submission is for a modified medical device, not a new AI/software-as-a-medical-device (SaMD) or diagnostic device. As such, the typical "acceptance criteria" and "study" questions you'd ask for an AI model's performance metrics (like sensitivity, specificity, AUC) are not directly applicable in this context.

    Instead, the "acceptance criteria" here refer to production specifications and performance adequate for in vivo application under temporary conditions of use, and the "study" is a performance testing conducted to demonstrate substantial equivalence to a previously cleared device.

    Here's the breakdown of the information as requested, translated into the context of this medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Material PropertiesMechanical properties matching predicate device and suitable for use.Evaluated and found to support substantial equivalence.
    Drug Release ProfileGentamicin release consistent with intended therapeutic effect.Evaluated and found to support substantial equivalence.
    StabilityDevice stability for temporary use (up to 180 days).Evaluated and found to support substantial equivalence.
    Manufacturing StandardsMeeting production specifications.Evaluated and found to support substantial equivalence.
    BiocompatibilitySuitable for in vivo application.Implicitly covered by substantial equivalence to predicate.
    Indications for UseSame as predicate device (temporary shoulder replacement for septic process, up to 180 days).Confirmed as the same, thus meeting this "performance" criterion.

    Note: The document states that the performance testing was to verify that the implant performance "continues to meet the production specifications and be adequate for in vivo application under the temporary conditions of use." The primary "acceptance criteria" seem to be demonstrating that the modified device is substantially equivalent to its predicate (K060535) despite the addition of a metal reinforcing structure and a new material supplier.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated. The "performance testing" described is likely mechanical and chemical testing of the device itself and its components, rather than a clinical trial with patient data. Therefore, the "test set" would refer to the number of devices or material samples subjected to laboratory tests (e.g., fatigue testing, tensile strength, gentamicin elution studies). Without further documentation, the exact number cannot be determined from this summary.
    • Data Provenance: Not applicable in the sense of patient data. The testing would have been conducted in a laboratory setting by the manufacturer (Tecres S.p.A.).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. For material and mechanical performance testing, "ground truth" is established through standardized engineering and chemical testing methods, not expert consensus on clinical cases.
    • Qualifications of Experts: Not applicable. The testing would be performed by qualified engineers, chemists, and technicians following established protocols.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This type of testing involves objective measurements against predefined specifications, not human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI diagnostic tool on human reader performance, which is not the nature of this device.

    • Effect Size: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Standalone Performance: Not applicable. This device is an implanted prosthesis, not an algorithm or a software device.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For the performance testing, the "ground truth" would be established by:
      • Engineering specifications and standards: For mechanical properties (e.g., ISO standards for implant materials, internal design specifications for strength and durability).
      • Chemical analysis standards: For gentamicin release and stability (e.g., pharmacopeial methods, validated analytical chemistry techniques).
      • Comparison to predicate device's established performance: To demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This is not an AI/machine learning device. The device itself is the product. The term "training set" is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set."

    Summary of the 510(k) Submission's Core Argument:

    The K112983 submission for the Interspace Shoulder Temporary Shoulder Prosthesis argues for substantial equivalence to its own previously cleared version (K060535). The key modifications are the introduction of a metal reinforcing structure and a new material supplier. The manufacturer conducted performance testing (mechanical properties, gentamicin release, and stability data) to demonstrate that these changes did not negatively impact the device's ability to meet its production specifications and remain adequate for its temporary in vivo application. The FDA's clearance indicates that they found this demonstration of substantial equivalence sufficient without requiring complex clinical trials or new performance metrics beyond what was presented for the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101356

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-20

    (494 days)

    Product Code
    Regulation Number
    888.3360
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spacer-K is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.

    Spacer-K is applied on the femoral condyles and on the tibial plate following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).

    Spacer-K is not intended for use for more than 180 days, at which time it must be explanted and permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethylmethacrylate), Spacer-K is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.

    Spacer-G is indicated for temporary use (maximum of 180 days) as a total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.

    The device is inserted into the femoral medullary canal and acetabular cavity following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).

    Spacer-G is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).

    Device Description

    Spacers are temporary joint prostheses. Spacer-G is a single piece device that mimics a hemi-hip prosthesis, and is available in six sizes. Spacer-K includes a femoral and tibial component, and is available in three sizes. Spacer-G and Spacer-K are made of fully formed polymethylmethacrylate (radiopaque PMMA with gentamicin). Spacer-G contains an inner stainless steel (AISI 316L stainless steel) reinforcing structure. The mass used in the filling of the molds (the PMMA unformed resin) is prepared from a powder component and a liquid component. The liquid component consists of methylmethacrylate, N,N-dimethyl-p-toluidine, hydroquinone; the powder component consists of polymethymethacrylate, barium sulphate, benzovl peroxide, and gentamicin sulphate.

    The Spacer devices provide patients, undergoing a two-stage revision procedure for an infected total joint, a temporary implant to 1) allow for partial weight bearing and 2) provide a natural range of motion. The devices also maintain a patient's soft tissue and joint space, preventing further complications such as muscular contraction. The gentamicin protects the device from bacterial colonization.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Tecres Spacer-G and Spacer-K, based on the provided documents:

    Acceptance Criteria and Device Performance

    The provided 510(k) summary (K101356) does not explicitly state specific quantitative acceptance criteria or detailed performance data in a table format. Instead, it relies on substantiating that the modified devices (Spacer-G and Spacer-K) are substantially equivalent to their previously cleared versions (K062274 for Spacer-K and K062273 for Spacer-G).

    The core acceptance criterion implicitly stated is:

    • The modified Spacer devices must perform equivalently to their previously cleared versions, ensuring continued safety and effectiveness for their intended temporary use.

    The reported device performance is described as:

    • "Performance testing was conducted to verify that implant performance continues to meet the production specifications and be adequate for in vivo applications under the temporary conditions of use."
    • "Mechanical properties, gentamicin release and stability data were evaluated and found to support the substantial equivalence of the devices."

    Without quantitative metrics provided in these documents, a table cannot be fully populated. However, we can infer the categories of assessment:

    Acceptance Criteria CategoryReported Device Performance (Qualitative)
    Mechanical PropertiesEvaluated and found to support substantial equivalence to previously cleared devices.
    Gentamicin ReleaseEvaluated and found to support substantial equivalence to previously cleared devices (for antibacterial function).
    StabilityEvaluated and found to support substantial equivalence to previously cleared devices.
    Production SpecificationsImplant performance continues to meet production specifications.
    In Vivo AdequacyAdequate for in vivo applications under temporary conditions of use.

    Study Information

    The 510(k) summary for K101356 describes an evaluation for "substantial equivalence" rather than a de novo clinical study with patients. The study primarily involves performance testing to compare a modified device (minor material change) to an established, previously cleared device.

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the sample size for any test set in terms of clinical data or patient cohorts. The performance testing appears to be conducted on the devices themselves (e.g., in vitro mechanical testing, elution studies).
      • Data provenance: Not explicitly stated as retrospective or prospective clinical data. Given the nature of a 510(k) for a minor material change to a previously cleared device, the testing would likely be bench testing and possibly some in vitro or ex vivo studies. No specific country of origin for clinical data is mentioned as such data doesn't appear to be the primary basis for this particular submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the context of this 510(k) submission. The evaluation is based on engineering and material science performance testing, not on interpretation of patient data by medical experts. "Ground truth" here refers to established material properties and performance benchmarks.
    3. Adjudication method for the test set:

      • Not applicable as there is no mention of a human expert review or adjudication process for a test set of clinical cases. The evaluation is against engineering specifications and pre-established performance of the predicate device.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a temporary orthopedic implant, not an AI-powered diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical medical device.
    6. The type of ground truth used:

      • Engineering specifications and performance characteristics of the previously cleared predicate devices (K062274 and K062273). The "ground truth" for the modified device's performance is that it matches or exceeds the established performance of its predecessors, as demonstrated through mechanical, gentamicin release, and stability testing.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/machine learning study, so there is no concept of a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092773

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-24

    (76 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cemex Genta/Cemex Genta System/Cemex Genta System Fast bone cement is indicated for the fixation of prostheses to living bone in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.

    Device Description

    All Cemex Genta bone cements contain the same individual chemical constituents. The liquid components contain methylmethacrylate, N-N dimethy] p-toluidine and hydroquinone. The dry powder component contains polymethylmethacrylate, barium sulphate, benzoyl peroxide and gentamicin sulphate.

    AI/ML Overview

    The provided text does not contain information about specific acceptance criteria with numerical performance targets for the device (Cemex Genta bone cement) or a detailed study proving its direct performance against such criteria. Instead, it focuses on demonstrating substantial equivalence to predicate devices through performance testing based on established standards.

    Here's a breakdown of what can be extracted and what is not available in the given document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not explicitly provided in the format requested. The document states that "Performance testing was conducted to verify that the modified bone cement performance continue to be adequate for in vivo applications and meet the requirements of ISO5833 and ASTM 451-99." However, it does not list specific numerical acceptance criteria (e.g., tensile strength > X MPa) or the precise performance values achieved by the Cemex Genta devices against those criteria. It only states that the properties were "evaluated and found to support the substantial equivalence of the device."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not specify the sample size used for performance testing.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only mentions "in vivo applications," which implies testing relevant to biological systems.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable as the document describes performance testing of a physical medical device (bone cement) against engineering standards (ISO and ASTM), not a diagnostic or AI-driven system requiring expert-established ground truth from a test set of images or clinical cases.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    MRMC studies are relevant for diagnostic devices that involve human readers interpreting data, often with AI assistance. This document describes a bone cement, which is a therapeutic device. Therefore, an MRMC study was not done, and the concept of human readers improving with AI assistance is not relevant here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    This concept is also not applicable to a bone cement. The "device" itself is the bone cement, not an algorithm. Its performance is evaluated through material and mechanical testing, not as a standalone algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for evaluating the bone cement's performance was based on meeting the requirements of ISO 5833 and ASTM 451-99. These are international and American standards for "Implants for surgery – Acrylic resin cements" and "Standard Specification for Acrylic Bone Cement," respectively. These standards define the physical, chemical, and mechanical properties expected of such cements.

    8. The Sample Size for the Training Set:

    This information is not applicable. Bone cement is a physical product, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reasons as point 8.


    Summary of available information regarding the study:

    • Study Type: Performance testing of chemical-physical and mechanical properties, gentamicin release, and stability data.
    • Purpose: To verify that the modified bone cement's performance remains adequate for in vivo applications and meets the requirements of ISO 5833 and ASTM 451-99.
    • Conclusion: The properties were "evaluated and found to support the substantial equivalence of the device."
    • Comparison: The modified Cemex Genta cements were found substantially equivalent to predicate devices (Cemex Genta and Cemex Genta System cleared via K033596; Cemex Genta System Fast cleared via K043403).
    Ask a Question

    Ask a specific question about this device

    K Number
    K062273

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-05-22

    (654 days)

    Product Code
    Regulation Number
    888.3360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spacer-G is indicated for temporary use (maximum 180 days) as a total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. The device is inserted into the femoral medullary canal and acetabular cavity following removal of the existing femoral and acetabular implants and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). Spacer-G is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.).

    Device Description

    Spacer-G is composed of fully formed gentamicin/polymethylmethacrylate (PMMA) bone cement. The one piece design mimics a hemi-hip prosthesis.

    AI/ML Overview

    This 510(k) summary for the Spacer-G device primarily focuses on demonstrating substantial equivalence to existing predicate devices, rather than presenting a study with specific acceptance criteria and performance data for the device itself.

    Therefore, many of the requested sections about acceptance criteria, study design, and performance metrics are not explicitly detailed in this document. The core of this submission is to show that the modified Spacer-G is just as safe and effective as already approved devices by having similar design, materials, performance characteristics, and gentamicin release profile.

    Here is an attempt to address your requests based on the provided text, highlighting where information is absent:


    Acceptance Criteria and Study to Prove Device Meets Criteria: Spacer-G Modification

    The provided 510(k) summary for the Spacer-G device demonstrates substantial equivalence to predicate devices rather than presenting a study with specific acceptance criteria directly linked to a novel performance claim for the modified device. The "acceptance criteria" can be inferred as the standard for substantial equivalence, meaning the device must perform comparably to legally marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Substantial Equivalence Goal)Reported Device Performance
    Material Composition: Same as predicate Spacer-G"incorporates the same materials"
    Design: Same as predicate Spacer-G"has the same design"
    Mechanical Characteristics: Equivalent to predicate Spacer-G"has equivalent performance and mechanical characteristics"
    Shelf Life and Packaging: Same as predicate Spacer-G"has the same shelf and packaging"
    Gentamicin Release Profile: Similar to predicate Biomet Stage One Disposable Cement Spacer Mold (K052990) when used with predicate Biomet Cobalt G HV Bone Cement (K051532)"has a similar gentamicin release profile as that of the predicate Biomet Stage One Disposable Cement Spacer Mold... when used with predicate Biomet Cobalt G HV Bone Cement"
    Intended Use: Identical to predicate deviceSame Indications For Use as described.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the 510(k) summary. Given the nature of a substantial equivalence submission relying on comparison to existing devices, it is unlikely a large "test set" in the sense of a clinical trial was used for this specific modification. The "performance" assessment appears to be based on engineering principles and potentially laboratory testing rather than human clinical outcomes data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable/provided. The submission focuses on substantial equivalence based on technical and material characteristics, not on a clinical ground truth established by experts.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. There is no mention of a clinical test set requiring adjudication in this 510(k) summary.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is a temporary hip prosthesis, not an AI or imaging diagnostic device that would involve human readers or AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This device is a medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the performance characteristics of the legally marketed predicate devices. The modified Spacer-G is evaluated against these established benchmarks. No pathology, outcomes data, or expert consensus specific to the modified device's novel performance is referenced as a "ground truth" to meet new performance criteria.

    8. The Sample Size for the Training Set

    This information is not applicable. This device is not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable. This device is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062274

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-05-22

    (654 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spacer-K is indicated for temporary use (maximum 180 days) as an adjunct in total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. Spacer-K is only indicated for an implantation period of 180 days or less. Because of the inherent mechanical limitations of the device material (gentamicin/polymethilmethacrylate), Spacer-K is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes ) throughout the implantation period.

    Device Description

    The Spacer-K is a temporary device that mimics a "total knee prosthesis". The two-component unconstrained design incorporates a femur and tibial component, it's fully formed by gentamicin/polymethyImethacrylate (PMMA) bone cement.

    AI/ML Overview

    This 510(k) summary for the Spacer-K knee prosthesis modification does not include the detailed information required to describe acceptance criteria and a study proving device performance in the way you've outlined. This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than presenting a performance study with acceptance criteria.

    However, I can extract information related to the device's characteristics and the basis for its clearance:

    Here's what can be gathered, and what is explicitly missing from the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Missing from the document. The 510(k) summary states that the modified Spacer-K has "equivalent performance and mechanical characteristics" to the predicate Spacer-K (K032522) and a "similar gentamicin release profile" to another predicate (K050210). However, it does not provide specific acceptance criteria (e.g., minimum tensile strength, specific drug release rates) or reported numerical performance data from a study.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Missing from the document. There is no description of a specific test set or clinical study with patient data. The equivalence is based on design, materials, and mechanical characteristics (likely bench testing or literature review, but details are not provided).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Missing from the document. This type of information is relevant for studies involving subjective assessment (e.g., image interpretation), which is not the primary focus of this 510(k) application.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Missing from the document. Not applicable given the type of submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Missing from the document. Not applicable. The Spacer-K is a physical medical device (a temporary knee prosthesis), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Missing from the document. Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Missing from the document. For a device like this, "ground truth" would typically refer to established engineering standards for materials and mechanical properties, or validated methods for drug release testing. The document refers to "equivalent performance and mechanical characteristics" and "similar gentamicin release profile," implying that these aspects were compared to predicate devices, but the specific methodologies for determining this "truth" are not detailed.

    8. The sample size for the training set

    • Missing from the document. Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Missing from the document. Not applicable.

    Summary of what the document DOES provide regarding clearance:

    The device (Spacer-K Modification) received 510(k) clearance (K062274) based on a demonstration of substantial equivalence to legally marketed predicate devices.

    • Predicate Devices:
      • Spacer-K device (K032522): The modified Spacer-K has the "same design, incorporates the same materials, has equivalent performance and mechanical characteristics, and has the same shelf and packaging" as this predicate.
      • Biomet Stage One Disposable Cement Spacer Mold for Temporary Knee Prosthesis with Reinforcement Stem (K050210): The modified Spacer-K has a "similar gentamicin release profile" when used with Biomet Cobalt G HV Bone Cement (K051532).

    Key Takeaway: This 510(k) summary is typical for a device modification, where the primary goal is to show that the new version is as safe and effective as a previously cleared device. It does not contain the kind of detailed performance study and acceptance criteria you would expect from a de novo submission or a product requiring extensive clinical trials. The "study" here is essentially the comparison against the predicate devices across defined parameters (design, materials, mechanical characteristics, drug release).

    Ask a Question

    Ask a specific question about this device

    K Number
    K063438

    Validate with FDA (Live)

    Device Name
    VSPSPN
    Manufacturer
    Date Cleared
    2007-01-17

    (64 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vspspn is indicated for the treatment of pathological fractures of the vertebral body using a vertebroplasty or kyphoplasty procedure. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

    Device Description

    Vspspn is a two-part (powder and liquid) radiopaque, polymethyl methacrylate acrylic resin. The powder and liquid components are mixed together at the point of use to form the resin. The product is supplied sterile, for single use.

    AI/ML Overview

    This 510(k) submission (K063438) for Vspspn, an acrylic resin for treating pathological vertebral body fractures, does not include information about acceptance criteria or a study proving device performance against such criteria.

    The submission focuses entirely on demonstrating substantial equivalence to a predicate device, Mendec Spine (K042415). It asserts that Vspspn has the "same design, incorporate the same materials, have the same performance and mechanical characteristics, have the same shelf life and equivalent packaging, and are sterilized by the same method" as the predicate.

    Therefore, I cannot provide the requested information regarding:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size for the test set and data provenance.
    3. Number and qualifications of experts for ground truth.
    4. Adjudication method.
    5. MRMC comparative effectiveness study results or effect sizes.
    6. Standalone algorithm performance.
    7. Type of ground truth used for performance evaluation.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    This type of submission relies on demonstrating similarity to an already approved device rather than presenting new performance data against specific criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062452

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-09-21

    (30 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kit Mendec Spine and delivery system is indicated for the treatment of pathological fractures of the vertebral body using a vertebroplasty procedure. Painful vertebral compression fractures may result from osteoporosis, benign (hemangioma), and malignant lesions (metastatic cancers, myeloma).

    Device Description

    Kit Mendec Spine and delivery system contains an acrylic resin for vertebroplasty and a delivery system. The acrylic resin consists of exportion of acryic Tosh 101 vercent. The delivery system consists of a syringe-like device with a "gun-system", an Extension tube, an Extension tube and may include, depending on the variant, a Needle.

    AI/ML Overview

    This 510(k) premarket notification (K062452) describes the "Kit Mendec Spine and Delivery System," a convenience kit for percutaneous vertebroplasty. This submission is a "Special 510(k)," indicating that it likely relies on its substantial equivalence to a previously cleared predicate device, the "Mendec Spine" (K042415), and that any modifications are minor and do not raise new questions of safety or effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided 510(k) summary (and the full submission, by its nature as a Special 510(k)) does not specify quantifiable acceptance criteria or detailed device performance metrics in the format typically seen with clinical studies for novel devices or significant modifications. Instead, the document focuses on demonstrating substantial equivalence to a predicate device.

    For a Special 510(k), the acceptance criteria are implicitly that the new device performs "as intended" and is "substantially equivalent" to the predicate, meaning it has the same intended use, technological characteristics, and raises no new safety or effectiveness concerns. Performance is generally demonstrated through comparisons to the predicate, often involving:

    • Materials comparison: Demonstrating that the materials used in the new device are the same as or equivalent to those in the predicate, or are well-established for this application.
    • Design comparison: Showing that the design features are either identical or that any changes do not adversely affect performance or safety.
    • Performance testing (if applicable to changes): For a Special 510(k), this typically involves bench testing to ensure modified components meet established standards or perform comparably to the predicate. Clinical studies are rare for Special 510(k)s unless a significant change warranted it.

    Based on the provided information, typical "acceptance criteria" and "reported device performance" in the sense of clinical accuracy or outcome rates are not present, as the submission relies on substantial equivalence and non-clinical demonstrations.

    Acceptance Criteria (Implied for Special 510(k))Reported Device Performance (Summary of Submission)
    Intended Use: Same as the predicate device (treatment of pathological fractures of the vertebral body using a vertebroplasty procedure).Intended Use: "The Kit Mendec Spine and delivery system is indicated for the treatment of pathological fractures of the vertebral body using a vertebroplasty procedures for the nodinall of painting may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (may myeloma)." (This is identical to the predicate's intended use, as verified by the FDA's clearance letter and the Indications for Use statement in Attachment 2).
    Technological Characteristics: Materials, design, and operating principles are the same as or equivalent to the predicate, or differences do not raise new safety/effectiveness questions.Device Description: "Kit Mendec Spine and delivery system contains an acrylic resin for vertebroplasty and a delivery system. The acrylic resin consists of exportion of acryic Tosh 101 vercent. The delivery system consists of a syringe-like device with a 'gun-system', an Extension tube, an Extension tube and may include, depending on the variant, a Needle." The submission states: "The components of the Kit Mendec Spine and delivery system are either exempt from premarket notification or have been found to be substantially equivalent through the premarket notification process for the use for which the sit is to be intended. Any difference that may exist do not significantly affect the substantial equivalence of the device."
    Safety and Effectiveness: No new questions of safety or effectiveness are raised compared to the predicate device.Demonstrated by comparison to the predicate device (Mendec Spine, K042415) and by providing adequate documentation (likely including bench testing, biocompatibility, and sterilization validation, though not detailed in this summary) to show that any proposed changes do not adversely impact safety or effectiveness. The FDA's clearance letter confirms: "We have reviewed your Section 510(k) premarket notification of intent to market the device [...] and found the device is substantially equivalent..."

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. For a Special 510(k) demonstrating substantial equivalence, a traditional "test set" with patient data (as one would use for AI performance evaluation) is not typically required or described. The evaluation relies on demonstrating equivalence to the predicate through materials, design, and performance testing data (e.g., mechanical, biocompatibility, sterilization studies), which are not clinical test sets.
    • Data Provenance: Not applicable for a clinical test set. The data supporting this submission would primarily be non-clinical (bench testing, material characterization, design specifications) provided by the manufacturer (Tecres S.p.A., Italy).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is a medical device (acrylic resin and delivery system), not an AI diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study assessing diagnostic performance against a ground truth established by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a vertebroplasty kit, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. This is a physical medical device. The "ground truth" for its safety and effectiveness is established through adherence to standards, biocompatibility testing, mechanical testing, and comparison to the predicate device's established performance, rather than an expert consensus on patient data.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that uses a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K060535

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-05-24

    (85 days)

    Product Code
    Regulation Number
    888.3690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spacer-S is intended for use as a temporary (maximum 180 days) shoulder replacement in patients undergoing a two-stage procedure due to a septic process.

    The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection) as defined by physician and/or surgeon.

    Spacer-S is indicated for temporary use (maximum 180 days) as an adjunct to total shoulder replacement and hemi shoulder replacement procedures in skeletally mature patients undergoing a two-stage revision procedure due to a septic process. Spacer-S is only indicated for an implantation period of 180 days or less.

    Device Description

    Spacer-S is composed of fully formed gentamicin/polymethylmethacrylate (PMMA) bone cement. The one piece design mimics a hemi-shoulder prosthesis.

    AI/ML Overview

    The K060535 filing describes the Tecres Spacer-S, a temporary shoulder prosthesis. The information provided primarily focuses on its device description, intended use, and a demonstration of substantial equivalence to a predicate device through performance testing, rather than an AI/ML device and its associated acceptance criteria and study data.

    Therefore, many of the requested categories (e.g., sample size for test/training sets, experts for ground truth, MRMC studies, standalone performance, etc.) are not applicable or cannot be extracted from the provided text, as this is a medical device approval for a physical implant, not a software or AI/ML diagnostic tool.

    However, I can extract the relevant performance data that was used to establish substantial equivalence.

    Here's the information that can be extracted or inferred from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Fatigue strengthEvaluated and found adequate for anticipated in vivo load applications under temporary conditions of use.
    Static strengthEvaluated and found adequate for anticipated in vivo load applications under temporary conditions of use.
    Antibiotic releaseEvaluated and found to support substantial equivalence.
    Overall Goal:Substantial equivalence to the predicate Equinoxe Shoulder System device (#K042021).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The performance testing refers to mechanical and material evaluations, not clinical trials with human subjects in the typical sense of test sets for AI/ML.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. Ground truth, in the context of this traditional medical device, is established through engineering and laboratory testing (e.g., measuring physical properties like strength and antibiotic release), not through expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. Adjudication methods are relevant for studies involving human interpretation or clinical endpoints, which is not the focus of the performance testing described for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a physical medical implant, not an AI or imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a physical medical implant, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the performance data, the "ground truth" would be established through engineering standards and laboratory measurements of the device's mechanical properties (fatigue strength, static strength) and its antibiotic release characteristics. There is no mention of clinical pathology or outcomes data being used to define the "ground truth" for these specific performance tests.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" in the context of this traditional medical device approval based on performance testing of the device itself.

    9. How the ground truth for the training set was established

    This information is not applicable. As there is no "training set," there is no ground truth to establish for it.