    K Number
    K243425

    Validate with FDA (Live)

    Date Cleared
    2025-07-22

    (260 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Guided DAS Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Guided DAS Surgical Kit is intended to allow sterilization of the enclosed medical devices.

    Guided DAS Surgical Kit requires the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

    The kit is to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: dynamic air removal steam sterilization- Exposure at 132 °C for 4 minutes and 30 minutes dry time.

    Guided DAS Surgical Kit is intended for sterilization of non-porous loads.

    Guided DAS Surgical Kit is recommended not to be stacked during sterilization.

    The combined weight of the Guided DAS Surgical Kit and the associated instruments is 760 grams.

    The weight of the empty Guided DAS Surgical Kit is 600 grams.

    Device Description

    This tray is designed to hold various dental surgical drills and tools to organize and protect the instruments that are sterilized in the trays by the healthcare provider.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for the "Guided DAS Surgical Kit," which is a sterilization wrap/container. It outlines the device's indications for use, classification, and a comparison to a predicate device.

    However, the provided document does not contain information relevant to AI/ML device performance, such as:

    • A table of acceptance criteria and reported device performance for an AI/ML model.
    • Sample sizes for AI/ML test sets or training sets.
    • Data provenance (country of origin, retrospective/prospective).
    • Number of experts for ground truth establishment or their qualifications.
    • Adjudication methods.
    • Details of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or effect sizes of AI assistance on human readers.
    • Standalone (algorithm-only) performance metrics.
    • Type of ground truth used (expert consensus, pathology, outcome data).

    The "Summary of Nonclinical Testing Table" on page 8 focuses on validation tests typical for medical devices related to cleaning, sterilization, dry time, life cycle, and biocompatibility, which are hardware-related and not AI/ML performance metrics. The conclusion explicitly states, "No clinical data were included in this submission."

    Therefore, based solely on the provided text, it's impossible to describe the acceptance criteria and the study that proves an AI/ML device meets them, as the document is about a physical medical device (sterilization kit) and not an AI/ML algorithm.

    If the request assumes this document should or would contain information about an AI/ML component, then the answer is that the document explicitly does not provide such information.

    To answer your request within the confines of the provided text, I must state that the document does not contain the information needed to describe the acceptance criteria and study for an AI/ML device.

    The document describes the testing performed for a physical medical device (sterilization kit). The listed tests are for material and functional validation, not for an AI/ML algorithm's analytical or clinical performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243530

    Validate with FDA (Live)

    Device Name
    Dynamic TiBase
    Date Cleared
    2025-05-30

    (197 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Compatible Implant Systems:

    Compatible Implant System (Connection)Implant Body Diameter, mmImplant Platform, mm
    Osstem® TS Hiossen® ET (Internal Taper)3.5Mini
    4.0, 4.5, 5.0, 5.5, 6.0, 7.0Regular
    Neodent GM (Morse taper)3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0GM
    Nobel Active (Conical Connection)3.5NP
    4.3, 5.0RP
    5.5WP
    Straumann Bone Level (CrossFit® Morse Taper)3.3NC
    4.1, 4.8RC
    Straumann BLX (TorcFit™ Internal Hexalobular)3.5, 3.75, 4.0, 4.5RB
    5.0, 5.5, 6.5WB

    All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Talladium Medical validated milling center for manufacture.

    Device Description

    Dynamic TiBase abutments are two-piece abutments composed of a CAD-CAM fabricated zirconia superstructure and a prefabricated titanium base component where the final two-piece abutment (base and cemented superstructure) is the finished device used for the prosthetic restoration. All subject device bases are made of titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3 and ASTM F136. The Dynamic TiBase abutments are provided in engaging and non-engaging designs for single-unit and multi-unit restorations, respectively.

    For each of the compatible OEM implant lines, the prefabricated titanium base components are provided with a gingival height (in the titanium base) ranging from 0.3 mm to 4 mm, and a platform diameter ranging from 4.30 mm to 5.50 mm. Angulation and additional gingival height may be provided in the zirconia superstructure. All Dynamic TiBase prefabricated titanium base components have a post with a cut-out to accommodate a restoration with an angled channel for screw access when clinically necessary. The post height of the prefabricated titanium base component ranges from 3.8 mm to 5.40 mm, and from 2.3 mm to 3.8 mm (cut-out height). The cementable post height of the final patient-matched abutment design, measured above the total combined gingival collar, shall be no less than 4 mm.

    All zirconia superstructures (copings) used to complete the final two-piece subject device Dynamic TiBase abutment will be made at a Talladium España, SL validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.

    The design parameters for the CAD-CAM zirconia superstructure for the Dynamic TiBase vary slightly among the compatible OEM implants. The design parameters for the CAD-CAM zirconia superstructure are summarized in the following table:

    Implant Compatibility (1)Minimum Wall Thickness, mmMaximum Gingival Height, mmMinimum Gingival Height (2), mmMaximum Angulation
    Osstem® TS Hiossen® ET (Internal Taper)0.325.181.225º
    Neodent GM (Morse taper)0.325.201.230º
    Nobel Active (Conical Connection)0.356.200.30
    Straumann Bone Level (CrossFit® Morse Taper)0.325.181.130º
    Straumann BLX (TorcFit™ Internal Hexalobular)0.325.341.530º

    (1) for the compatible sizes shown in Table 1
    (2) minimum gingival height in the titanium base, not the zirconia superstructure

    The required cement for bonding the zirconia superstructure to the Dynamic TiBases to create the final two-piece abutment is Nova Resin Cement cleared in K213609.

    Also, the subject of this submission are seven (7) abutment screws for use with the subject abutments.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a dental implant abutment — the Dynamic TiBase — not an AI-powered diagnostic device or software. Therefore, the information typically requested about acceptance criteria and study designs for validating AI/ML-based medical devices (such as sample size, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) is not applicable to this document.

    The "performance data" section in this 510(k) summary refers to traditional engineering and biocompatibility testing for a physical medical device, not performance metrics for an algorithm.

    Here's how to interpret the provided document in the context of "acceptance criteria" and "proof":

    Acceptance Criteria and Reported Device Performance (as inferred for a physical device):

    For a physical device like the Dynamic TiBase, the "acceptance criteria" are generally met through demonstrating substantial equivalence to a previously cleared predicate device. This involves validating material properties, manufacturing processes, functional performance (e.g., mechanical strength, compatibility), and biocompatibility.

    The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to demonstrate substantial equivalence.

    Acceptance Criterion (Inferred for a Dental Abutment)Reported Device Performance / Proof
    Biocompatibility (Device materials are safe for biological contact)Referenced from K232151 (provided in K212108): Biocompatibility testing according to ISO 10993-5 (cytotoxicity) for abutment materials ASTM F136 (titanium alloy) and ISO 13356 (zirconia). This indicates the materials did not show cytotoxic effects when tested according to the standard.
    Mechanical Performance/Strength (Abutment can withstand oral forces without failure)Provided in this submission: Mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments in conjunction with the compatible OEM implants. This standard specifically evaluates the fatigue strength of endosseous dental implants. The successful completion of this testing, validated against the predicate device, implies the device meets required load-bearing capacities.
    MRI Safety (Device does not pose undue risk in MRI environment)Provided in this submission: Non-clinical worst-case MRI review to evaluate subject device components in the MR environment using scientific rationale and published literature. This included addressing magnetically induced displacement force and torque parameters as per FDA guidance. The review concluded that the device is safe for MRI under specified conditions.
    Sterilization Efficacy (Device supplied non-sterile can be effectively sterilized by user)Referenced from K232151 (provided in K212108): Moist heat sterilization validated to a sterility assurance level (SAL) of 10-6 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2. This demonstrates that the specified sterilization method for end-users achieves the required sterility.
    Dimensional Compatibility & Fit (Device correctly interfaces with specified implants)Provided in this submission: Reverse engineering dimensional analysis (of OEM implant bodies, OEM abutments, and OEM abutment screws) to demonstrate that the subject device abutments are compatible with listed OEM Implant Systems (Osstem® TS/ Hiossen® ET, Neodent GM, NobelActive, Straumann Bone Level, and Straumann BLX Implant Systems). This confirms the physical design correctly matches the intended implant systems.
    Material Conformance (Materials meet specified standards)Materials confirmed to conform to: Ti-6Al-4V (ASTM F136, ISO 5832-3) for titanium components; Zirconia (ISO 13356) for superstructure; Nova Resin Cement (cleared in K213609) for bonding. This demonstrates the use of medical-grade, standard-compliant materials.
    Manufacturing Quality (Device is produced under controlled conditions)Stated that zirconia superstructures will be made at a Talladium España, SL validated milling center under FDA quality system regulations. This implies adherence to Good Manufacturing Practices (GMP).

    Regarding the specific questions about an AI/ML context:

    1. A table of acceptance criteria and the reported device performance: Provided above, adapted for a physical medical device.
    2. Sample size used for the test set and the data provenance: Not applicable. The "test set" for this physical device refers to the number of physical abutment samples or material samples subjected to mechanical, biocompatibility, and MRI testing. The document does not specify exact sample numbers for these engineering tests, only the standards used (e.g., ISO 14801 typically specifies minimum sample sizes). Data provenance is "non-clinical data" generated from laboratory testing.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a dental abutment's performance is established by engineering standards, material specifications, and physical testing, not by expert human interpretation of images or clinical outcomes in the same way as an AI diagnostic.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This is a concept used in evaluating human reader performance in AI studies.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI-assisted diagnostic devices.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This refers to an AI algorithm's performance.
    7. The type of ground truth used: For a physical device, "ground truth" is defined by adherence to established engineering standards (e.g., ISO 14801 for mechanical strength), material properties, and biocompatibility standards. "Reverse engineering dimensional analysis" served as a form of "ground truth" for compatibility. No pathology or outcomes data was used for this premarket notification.
    8. The sample size for the training set: Not applicable. There is no "training set" as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) clearance is for a physical medical device (dental abutment), and the "performance data" section details the engineering and material testing conducted to demonstrate its safety and effectiveness, primarily through substantial equivalence to previously cleared devices. It does not involve AI/ML validation methodologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241170

    Validate with FDA (Live)

    Date Cleared
    2024-10-21

    (178 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
    TRI Screws are intended to secure TRI-matrix® Crown Abutments to TRI-matrix® implants.

    Device Description

    Dynamic TiBase abutments are designed for retention of a CAD-CAM fabricated zirconia superstructure where the final two-piece abutment (base and cemented superstructure) is the finished device used for the prosthetic restoration. All subject device bases are made of titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3 and ASTM F136. The Dynamic TiBase abutments are provided in engaging designs for single-unit restorations.
    Dynamic TiBase abutments are provided in sizes compatible with the following SGS® Dental Implants System implants:
    SGS P1D, Conical connection, 5.0 mm body diameter, 3.5 (3.0 mm) platform; and
    SGS P7N, Narrow Conical connection, 3.2 mm body diameter, 2.1 (2.5 mm) platform.
    All Dynamic TiBase abutments are provided in a straight design (no angulation in the TiBase portion). All Dynamic TiBase abutments have a prosthetic post with a cut-out to accommodate a restoration with an angled channel for screw access when clinically necessary.
    For the compatible SGS P1D Concal connection implant, the Dynamic TiBase abutments are provided with a gingival height (in the titanium base) of 0.5 mm, or 3.0 mm, and a prosthetic platform diameter of 4.3 mm. Angulation and additional gingival height may be provided in the zirconia superstructure. For the compatible SGS P1D Conical connection implant, the abutments have a prosthetic post height (length above the gingival height) of 4.0 mm, and a cut-out height of 2.5 mm.
    For the SGS P7N Narrow Conical connection implant, the gingival height (in the titanium base) is 0.5 mm or 1.5 mm. and a prosthetic platform diameter of 3.8 mm. Angulation and additional gingival height may be provided in the zirconia superstructure. For the compatible P7N Narrow Conical connection implant the abutments have a prosthetic post height (length above the gingival height) of 4.5 mm, and a cut-out height of 2.5 mm. Dynamic TiBase abutments with a 4.5 mm maximum post height may be shortened to no less than 4 mm for a single-unit restoration.
    The subject device TRI Screws are to be used only to attach TRI-matrix® Crown Abutments. The TRI-matrix® Crown Abutments and TRI-matrix® implants are manufactured by TRI Dental Implants Int. AG and were cleared in K203660. The TRI screws have a hexalobular instrument interface, a screw head diameter of 2.6 mm, 1-72 UNF thread, and an overall length of 5.15 mm or 7.6 mm. The TRI Screws are anodized to a gold color and are only intended to be used with straight abutments. The compatible abutments are made from zirconia and are for direct final restorations.

    AI/ML Overview

    This is a substantially equivalent (SE) determination letter from the FDA, not a study report. It states that the device is substantially equivalent to legally marketed predicate devices, meaning it has the same intended use and similar technological characteristics. Therefore, it does not contain the detailed information about acceptance criteria and study results you requested for a novel device.

    The document states that a non-clinical mechanical testing study was performed to support the performance of the abutments and zirconia superstructure. However, the specific acceptance criteria and detailed results of this study are not provided in this regulatory document.

    Here's a breakdown of what can be extracted or inferred based on your request, and what is explicitly not available:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria for Substantial Equivalence: The primary acceptance criterion for this FDA clearance is "substantial equivalence" to predicate devices. This means the device has the same intended use and similar technological characteristics (materials, design, manufacturing, sterile processing, etc.), and any differences do not raise new questions of safety or effectiveness.
    • Reported Device Performance: The document states that mechanical testing conducted according to ISO 14801 was performed to support the performance of the subject device abutments in conjunction with compatible OEM implants. It also mentions a "non-clinical worst-case MRI review" using scientific rationale and published literature. However, the specific numerical acceptance criteria for these tests (e.g., specific fatigue limits, displacement values) and the actual quantitative results are not provided in this summary. It only states that these tests were performed to mitigate risks.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • This information is not provided in the document. The mechanical testing would have involved a specific number of samples (e.g., implant-abutment assemblies) tested according to ISO 14801. The document only states that such testing was performed to demonstrate performance. The country of origin for the testing data is not mentioned, but the manufacturer is Talladium España, SL (Spain).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This kind of information is not applicable to this type of regulatory submission (Class II dental implant components). There is no "ground truth" established by experts in the context of diagnostic accuracy for this device. The assessment is based on engineering principles, materials science, and mechanical testing against recognized standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This is not applicable as it pertains to expert consensus in diagnostic studies, not mechanical testing of medical devices.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable. This is a physical dental implant component, not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable. This is a physical dental implant component, not an AI diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the mechanical testing, the "ground truth" would be the physical properties and mechanical integrity of the device measured against predetermined performance specifications outlined in relevant ISO standards (e.g., ISO 14801 for fatigue testing of dental implants). For biocompatibility, it's testing against ISO 10993 standards.

    8. The sample size for the training set

    • This is not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • This is not applicable.

    Summary of available and missing information regarding acceptance criteria and study:

    CategoryInformation Provided in Document
    Acceptance Criteria (General)Substantial Equivalence to predicate devices (K232151, K182219, K203660) in intended use and technological characteristics. Any differences do not raise new questions of safety or effectiveness. This is the overarching "acceptance criterion" for FDA clearance.
    Specific Performance CriteriaMechanical testing conducted according to ISO 14801. Non-clinical worst-case MRI review per FDA guidance (magnetically induced displacement force and torque). Sterility assurance level (SAL) of 10^-6 by overkill method (ANSI/AAMI/ISO 17665-1 and TIR 17665-2). Biocompatibility testing according to ISO 10993-5 (cytotoxicity) for materials ASTM F136 and ISO 13356. Specific numerical acceptance values for these tests are NOT provided.
    Reported Device PerformanceThe document states these tests were performed and used to demonstrate substantial equivalence and mitigate risks. Specific numerical results are NOT provided.
    Study Type(s)Non-clinical mechanical testing (ISO 14801). Non-clinical MRI safety review. Sterilization validation. Biocompatibility testing.
    Sample Size (Test Set)Not provided.
    Data Provenance (Country, Retrospective/Prospective)Manufacturer is from Spain (Talladium España, SL). Data provenance details (e.g., where testing was performed, whether data was prospective or retrospective) are not explicitly provided for the non-clinical tests.
    Number/Qualifications of Experts (Ground Truth)Not applicable for this type of device.
    Adjudication MethodNot applicable.
    MRMC Comparative Effectiveness StudyNot applicable.
    Standalone Performance Study (AI)Not applicable.
    Type of Ground Truth UsedFor mechanical testing: adherence to performance specifications in ISO 14801. For biocompatibility: adherence to ISO 10993 standards. For sterilization: achievement of SAL 10^-6 per ANSI/AAMI/ISO 17665.
    Training Set SizeNot applicable.
    How Ground Truth for Training Set EstablishedNot applicable.

    This document is a regulatory communication affirming substantial equivalence based on non-clinical data, not a detailed scientific study report. For the specific acceptance criteria and quantitative performance results, one would typically need to refer to the full 510(k) submission, which is not publicly available in this level of detail.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231559

    Validate with FDA (Live)

    Date Cleared
    2023-11-17

    (171 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multi-Unit DAS System abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.

    Device Description

    Multi-Unit DAS System abutments are designed for retention of single-unit and multi-unit restorations. The multi-unit abutments are provided in a straight design (no angulation in the base portion) that threads directly to the OEM implant. For each of the compatible OEM implant lines, the multi-unit abutments are provided with gingival heights ranging from 1 mm to 5 mm, a prosthetic platform diameter of 4 mm, and a prosthetic post height of 1.4 mm. The multi-unit abutments are the base of a two-piece abutment. The second piece is a metal coping, called a Ti-Base in this submission, that attaches to the multi-unit abutment (not directly to an implant). For permanent restorations a zirconia superstructure is attached to the Ti-Base, and additional gingival height and angulation may be provided in the zirconia superstructure.

    The subject device metal copings (Ti-Bases) include: a straight, prepable design with an additional gingival height of 1.5 mm and a prepable 9 mm prosthetic post; and Dynamic Ti-Bases in three (3) designs, each with an additional gingival height of 0.5 mm and a cut-out in the prosthetic post to accommodate a restoration with an angled screw channel when clinically necessary. The Dynamic Ti-Base prosthetic post heights are 4.5 mm (maximum height) / 3.0 mm (cut-out height), and 9.0 mm/3.5 mm. Multi-Unit DAS System Ti-Bases with a 9.0 mm post height may be shortened to no less than 4 mm for a single-unit restoration. The prepable Ti-Base has a platform diameter of 4 mm (platform to the multi-unit abutment) and a prosthetic platform diameter of 4 mm. The Dynamic Ti-Bases have a platform diameter of 4 mm (platform to the multi-unit abutment) and a prosthetic platform diameter of 4.15 mm.

    The compatibility between the subject device abutments and the OEM implants listed in the Indication for Use Statement was established by reverse engineering analysis of the OEM implants, OEM abutments, and OEM abutment screws.

    All subject device abutments and abutment screws are made of titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3.

    All zirconia copings (superstructures) for use with the subject device Dynamic Ti-Base will be made at a Talladium España, SL validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.

    The design parameters for the CAD-CAM zirconia superstructure for the Multi-Unit DAS System are:

    Minimum wall thickness - 0.25 mm

    Minimum post height for single-unit restorations - 4.0 mm

    Maximum gingival height in the zirconia superstructure - 5.24 mm for compatible Biomet 31 OSSEOTITE® Certain®, MegaGen AnyRidge, NobelActive®, and Zimmer Tapered Screw-Vent® implants; 5.76 mm for compatible Astra Tech OsseoSpeed TX implants

    Minimum gingival height - 0.5 mm (in the Ti-Base)

    Maximum angulation - 30° for compatible Biomet 3i OSSEOTITE® Certain®, MegaGen AnyRidge, NobelActive®, and Zimmer Tapered Screw-Vent® implants; 25° for compatible Astra Tech OsseoSpeed TX implants

    The recommended cement for bonding the zirconia superstructure to the Dynamic TiBases to create the final two-piece abutment is G-CEM LinkAce™, cleared as GAM-200 in K120243.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Multi-Unit DAS System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device PerformanceStudy that Proves Device Meet Criteria
    Mechanical PerformanceAccording to ISO 14801 (Fatigue Test)Mitigated by mechanical testing.Mechanical testing conducted according to ISO 14801.
    BiocompatibilityAccording to ISO 10993-5 (Cytotoxicity)Conforming to standards.Referenced from K221966 (provided in K212108), which included biocompatibility testing.
    Material CompositionTitanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3.All abutments and abutment screws are made of Ti-6Al-4V conforming to ASTM F136 and ISO 5832-3.N/A (Material specification and sourcing)
    Zirconia Superstructure MaterialConforming to ISO 13356.Material will conform to ISO 13356.N/A (Material specification and sourcing)
    SterilizationMoist heat sterilization to a Sterility Assurance Level (SAL) of 10^-5 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2 for non-sterile devices.Moist heat sterilization for non-sterile devices validated to a SAL of 10^-5 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2.Validation study conducted for moist heat sterilization.
    MR SafetyAssessment of magnetically induced displacement force and torque.No adverse effect on safety.Non-clinical analysis performed using scientific rationale and published literature (TO Woods et al. 2021) to evaluate subject devices in the MR environment.
    Compatibility with OEM ImplantsEstablished by reverse engineering analysis.Compatibility established.Reverse engineering analysis (of OEM implants, OEM abutments, and OEM abutment screws) for compatible OEM implants.
    Zirconia Superstructure Design ParametersMinimum wall thickness - 0.25 mmMinimum post height for single-unit restorations - 4.0 mmMaximum gingival height (in the zirconia superstructure) - 5.24 mm (Biomet, MegaGen, NobelActive, Zimmer), 5.76 mm (Astra Tech)Minimum gingival height - 0.5 mm (in the Ti-Base)Maximum angulation - 30° (Biomet, MegaGen, NobelActive, Zimmer), 25° (Astra Tech)All parameters met.Technical specification and design validation.
    Recommended CementG-CEM LinkAce™ (cleared as GAM-200 in K120243).G-CEM LinkAce™ is the recommended cement.Adherence to existing cleared product.

    2. Sample size used for the test set and the data provenance

    • Mechanical Testing (ISO 14801) & Sterilization Validation: The document does not explicitly state the sample sizes used for these tests. For ISO 14801, typically a statistically significant number of samples are tested to demonstrate fatigue resistance, but the exact number isn't provided. For sterilization, typically multiple runs with biological indicators are performed.
    • Data Provenance: The document does not specify the country of origin for the data for specific tests, but the manufacturer, Talladium España, SL, is located in Lleida, Spain. The data is non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable (N/A) as the submission details non-clinical data (mechanical testing, material specifications, biocompatibility, sterilization, MR safety analysis) rather than studies involving human expert assessment of images or clinical outcomes.

    4. Adjudication method for the test set

    • N/A, as no expert adjudication for medical diagnoses/interpretations was conducted due to the non-clinical nature of the studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This is a medical device for dental implants, not an AI software intended for diagnostic assistance. The document explicitly states: "No clinical data were included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No standalone algorithm performance study was done. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Mechanical Testing: Ground truth is defined by the ISO 14801 standard and the manufacturer's performance specifications.
    • Biocompatibility: Ground truth is established by the cellular responses and lack of toxicity as defined by ISO 10993-5.
    • Material Composition: Ground truth is defined by the ASTM F136 and ISO 5832-3 standards.
    • Sterilization: Ground truth is defined by the Sterility Assurance Level (SAL) of 10^-5, validated against microbiological indicators as per ANSI/AAMI/ISO 17665.
    • MR Safety: Ground truth is established by scientific rationale and published literature, determining the device's behavior in an MR environment.
    • Compatibility with OEM Implants: Ground truth is established by the precise measurements and design characteristics obtained through reverse engineering analysis of the OEM implants and their components.

    8. The sample size for the training set

    • N/A. There is no AI component or machine learning model that requires a training set for this device.

    9. How the ground truth for the training set was established

    • N/A. There is no AI component or machine learning model that requires a training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232151

    Validate with FDA (Live)

    Device Name
    Dynamic TiBase
    Date Cleared
    2023-10-17

    (90 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.

    Device Description

    Dynamic TiBase abutments are designed for retention of a CAD-CAM fabricated zirconia superstructure where the final two-piece abutment (base and cemented superstructure) is the finished device used for the prosthetic restoration. Dynamic TiBase abutments are provided in engaging and non-engaging designs for single-unit and multi-unit restorations, respectively. For each of the compatible OEM implant lines, the Dynamic TiBase abutments are provided with a gingival height (in the titanium base) ranging from 0.3 mm to 5 mm, and a prosthetic platform diameter ranging from 4.3 mm to 5.9 mm. Additional gingival height and angulation may be provided in the zirconia superstructure. All Dynamic TiBase abutments have a prosthetic post with a cut-out to accommodate a restoration with an angled channel for screw access when clinically necessary. The prosthetic post heights range from 3.8 mm to 4.5 mm (maximum height), and from 2.3 mm to 2.7 mm (cut-out height). Dynamic TiBase abutments with a 4.5 mm maximum post height may be shortened to no less than 4 mm for a single-unit restoration. All subject device abutments and abutment screws are made of titanium alloy (Ti-6A1-4V) conforming to ASTM F136 and ISO 5832-3. All zirconia copings (superstructures) for use with the subject device Dynamic TiBase will be made at a Talladium España, SL validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Dynamic TiBase dental abutments). It describes the device, its intended use, and its substantial equivalence to previously cleared devices. However, it does not contain any information about acceptance criteria, the study design, or performance metrics typically associated with AI/ML-based medical devices.

    The document explicitly states: "No clinical data were included in this submission." and focuses on non-clinical data to demonstrate substantial equivalence, primarily through:

    • Mechanical testing: "mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments in conjunction with the compatible OEM implants."
    • Reverse engineering analysis: "reverse engineering analysis (of OEM implants, OEM abutments, and OEM abutment screws) demonstrating compatibility between the subject device abutments and the OEM implants listed in the Indication for Use Statement."
    • Material compatibility and sterilization: referencing previous 510(k) submissions for biocompatibility and sterilization validation.

    Therefore, I cannot fulfill your request for information regarding acceptance criteria and a study proving device performance in the context of an AI/ML-based device (e.g., sample size, ground truth, expert adjudication, MRMC studies), as this information is not present in the provided text. The device described is a physical dental abutment, not an AI/ML diagnostic or assistive tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212108

    Validate with FDA (Live)

    Device Name
    Dynamic TiBase
    Date Cleared
    2021-09-24

    (80 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.

    Device Description

    The purpose of this submission is to obtain marketing clearance for Dynamic TiBase, a series of twopiece titanium base abutments that require the fabrication of patient-specific custom superstructures using CAD-CAM technology. In final, finished form, the subject device abutments are intended to be used as a two-piece abutment composed of the base bottom-half (titanium base) with a bonded CAD-CAM zirconia top-half. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Talladium España, SL Dynamic TiBase, which is an endosseous dental implant abutment. The primary purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove clinical effectiveness or safety in the same way clinical trials do. Therefore, the information provided focuses on non-clinical data and comparisons to existing devices, rather than clinical study results with human subjects.

    Based on the provided document, here's a breakdown of the acceptance criteria and the study that proves the device meets them, specifically in the context of a 510(k) submission:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission for a device like this, the "acceptance criteria" are not clinical performance metrics in the traditional sense (e.g., sensitivity, specificity for a diagnostic device). Instead, they revolve around demonstrating substantial equivalence to predicate devices in terms of:

    • Intended Use: The device performs the same function for the same patient population.
    • Technological Characteristics: The device has similar design, materials, and operating principles.
    • Performance (Non-Clinical): The device performs similarly to or better than the predicate in engineering/bench tests relevant to its function and safety. Any differences in technological characteristics must not raise new questions of safety or effectiveness.

    The reported device performance is primarily based on non-clinical (bench) testing and comparison tables, not clinical studies in human subjects.

    Here's a table summarizing the "acceptance criteria" (or areas of equivalence demonstration) and the reported device performance as per the document:

    Acceptance Criterion (Area of Equivalence)Reported Device Performance (as demonstrated in 510(k))
    Intended UseDynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient. Demonstrated Equivalence: Stated to be substantially equivalent to primary predicate (K202026) and reference devices (K182246, K191919, K162021) in indications for use. Differences in wording do not affect intended use.
    Technological CharacteristicsDevice Type: CAD-CAM Titanium Base Abutments (two-piece, titanium base with bonded CAD-CAM zirconia superstructure). Materials: Titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3 for abutments and screws. Zirconia (ISO 13356) for copings. Design/Features: Straight design, 0.7 mm gingival height, cut-out for angled screw channel, specific post heights. Compatible with Thommen SPI® CONTACT dental implant fixtures.
    BiocompatibilityAll subject device abutments are biocompatible. Performance: Biocompatibility testing performed according to ISO 10993-12 (referenced) and confirmatory testing for finished devices (ISO 10993-5 and ISO 10993-12) was done.
    Sterilization ValidationThe device can be effectively sterilized by the end-user. Performance: Moist heat sterilization validation performed according to ISO 17665-1 and ISO 11737-2 (referenced). Device provided non-sterile for end-user sterilization.
    Mechanical Performance (Strength/Fatigue)The device is mechanically robust enough for its intended use. Performance: Static compression and compression fatigue testing performed according to ISO 14801 of worst-case constructs (abutments, zirconia superstructures, abutment screws, compatible implants). The fatigue limit data demonstrated sufficient strength. Differences in superstructure design parameters (wall thickness, gingival height) were mitigated by mechanical testing.
    Manufacturing ProcessThe manufacturing process is consistent and yields safe devices. Performance: Titanium alloy components manufactured from identical materials in identical facilities using identical processes as previously cleared products (K162021). Anodization process is also identical.
    Compatible Implant InterfaceThe device properly interfaces with specified implants. Performance: Established compatibility with Thommen SPI® CONTACT dental implant fixtures (cleared in K034014 and K072933) via business agreement and identical interface connections/platforms.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document explicitly states "No clinical data were included in this submission." The "studies" were non-clinical (bench) tests. While the ISO 14801 standard specifies the number of samples for fatigue and static testing (typically 5-10 for static, 10-12 for fatigue), the exact sample sizes used for each specific test (e.g., number of abutments tested at different loads) are not detailed in this summary. It only says "worst-case constructs" were tested.
    • Data Provenance: The data is non-clinical (bench testing) performed by the manufacturer or a contracted lab. The manufacturer is Talladium España, SL, located in Lleida, Spain. This indicates the testing was likely done in Spain or a European lab adhering to ISO standards. The data is retrospective in the sense that the tests were completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For a 510(k) submission based on non-clinical data, there are no "experts" establishing a clinical "ground truth" through consensus on patient cases. The "ground truth" for non-clinical testing is compliance with established engineering and biocompatibility standards (e.g., ISO 14801, ISO 10993). The "experts" here would be the engineers and scientists who designed and conducted the tests and analyzed the results according to these international standards. Their qualifications are inherent in their ability to perform such testing and demonstrate compliance with these standards.

    4. Adjudication method for the test set

    Not applicable. There's no clinical test set requiring adjudication. The non-clinical tests (biocompatibility, sterilization, mechanical) are evaluated against predefined acceptance criteria established by the relevant ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental implant abutment, not an AI-powered diagnostic imaging tool. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this 510(k) submission is compliance with established international engineering standards (e.g., ISO 14801 for mechanical performance, ISO 10993 for biocompatibility, ISO 17665-1 and ISO 11737-2 for sterilization) and the demonstration of substantial equivalence to predicate devices. This is achieved through:

    • Bench Test Results: Quantitative data from static compression and compression fatigue tests.
    • Material Specifications: Conformance to ASTM and ISO material standards.
    • Biocompatibility Test Reports: Results showing no adverse biological reactions.
    • Sterilization Validation Data: Proof of effective sterilization.
    • Direct Comparison: Tables and textual analysis comparing the subject device's features and performance to those of legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical device.