Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K070118

    Validate with FDA (Live)

    Device Name
    OTOGRAM, CA3350
    Manufacturer
    Date Cleared
    2007-03-07

    (54 days)

    Product Code
    Regulation Number
    874.1050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Otogram is an auditory diagnostic instrument intended to administer, under supervision by a trained healthcare professional, a battery of diagnostic and screening procedures that include the following:

    1. Pure tone air and bone conduction audiometry with automated masking

    2. Speech recognition threshold with automated masking

    3. Speech discrimination with automated masking

    4. Tympanometry

    5. Acoustic reflex threshold, both ipsilateral and contralateral

    6. Distortion product otoacoustic emissions

    7. Pure tone Stenger

    8. Patient survey

    The Otogram is indicated for use by trained healthcare professionals on both adult and pediatric subjects for measurement of audiometric parameters to identify and supply data to help diagnose hearing loss and ear disorders.

    Device Description

    The Otogram CA3350 is a computer-controlled, audiometric instrument combining the functions of an Audiometer, Distortion Product Otoacoustic Emissions Analyzer and Auditory Acoustic Impedance/Admittance Tester (i.e. Tympanometer). The device is controlled through the use of ASCII commands transmitted over a standard RS-232 communication port. Control software is installed on the supplied PC.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary for the Otogram CA3350, focusing on the acceptance criteria and the study that demonstrates its conformity:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (K070118) is a 510(k) premarket notification for a new device, the Otogram CA3350, claiming substantial equivalence to predicate devices. In such submissions, the acceptance criteria are generally established by demonstrating that the new device performs at least as well as, or is equivalent to, the legally marketed predicate devices. The "reported device performance" is primarily based on this comparison and adherence to relevant standards.

    Since this is a submission focused on substantial equivalence to existing devices, the "acceptance criteria" are implied to be the performance characteristics of the predicate devices themselves, and the "reported device performance" is the demonstration that the Otogram CA3350 meets or is equivalent to these.

    CharacteristicAcceptance Criteria (Predicate Devices' Performance)Reported Device Performance (Otogram CA3350)
    Intended UseTo diagnose various middle ear disorders (tympanometry, acoustic reflex), test cochlear function (otoacoustic emissions), and diagnose hearing/otologic disorders using audiometry. (Based on combined predicate device functions)Administer diagnostic and screening procedures: pure tone air/bone conduction audiometry with automated masking, speech recognition threshold/discrimination with automated masking, tympanometry, acoustic reflex threshold (ipsilateral/contralateral), distortion product otoacoustic emissions, pure tone Stenger, patient survey. Indicated for trained healthcare professionals, adult/pediatric subjects. (Broader, but encompasses predicate uses)
    Computer InterfaceRS 232C (Madsen Zodiac 901 & Bio-Logic AuDX)Computer included, RS 232C
    Internal PrinterThermal (112mm or 59mm) (Madsen Zodiac 901 & Bio-Logic AuDX)NA (Implies no internal printer, or relies on external printing)
    Printer InterfaceParallel port (Madsen Zodiac 901), RS 232C thermal printer (Bio-Logic AuDX)Parallel port or USB (Utilizes modern computer connectivity)
    DisplayGraphic supertwist LCD (Madsen Zodiac 901), LCD 1" x 3" (Bio-Logic AuDX)Color, resistive touchscreen (Improved display technology)
    Control InterfaceKeyboard (Madsen Zodiac 901), Touchpad (Bio-Logic AuDX)Keyboard, touchscreen (Combines and improves control)
    Size/WeightVarious (e.g., Madsen: 370 x 385 x 120 mm, 7.6 kg; Bio-Logic: 98 x 41 x 197 mm, 0.45 kg)356 x 348 x 493 mm, 21.1 kg (Within a comparable range for combined devices)
    Energy SourceAC 50/60 Hz, 100-240 V (Various)AC 47-63 Hz, 90-250 VAC (Standard AC power compatibility)
    Hardcopy OutputThermal paper or external printer paper (Madsen Zodiac 901), External thermal printer (Bio-Logic AuDX)External printer paper (Relies on external computer printing capabilities)
    Performance/Calibration StandardsIEC 1027, ANSI S3.39 (Madsen Zodiac 901), NA (Bio-Logic AuDX - implies compliance via general safety)IEC 1027, ANSI S3.39, IEC 60645-5 (Compliance with relevant audiometric and otoacoustic emissions standards)
    Electrical Safety StandardsEN 60601-1, class I, type B (Madsen Zodiac 901), EN 55011: 1991 group 1 Class B (Bio-Logic AuDX)IEC 60601-1 and 60601-1-1, class I, type B and BF and IEC 60601-1-2 (EMC) (Compliance with comprehensive medical electrical safety standards)
    Safety and EffectivenessAs safe and effective as predicate devices."The results of bench, user, and laboratory testing indicate that the new device is as safe and effective as the predicate devices." (Conclusion of the submission)

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Comparison testing with predicate devices was performed by a Hearing Instrument Specialist Intern," and that "Test reports are contained in Exhibit 12 of the 510(k) submission." However, the specific sample size for the test set is not provided in the publicly available summary. There is no information on the number of subjects or tests performed for this comparison.

    The data provenance is implied to be prospective testing conducted specifically for this 510(k) submission. The country of origin of the data is not explicitly stated but would presumably be the United States, given Tympany, Inc.'s location.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The summary does not explicitly describe the establishment of a "ground truth" through expert consensus for the test set. Instead, the testing likely involved comparing the results generated by the Otogram CA3350 directly against the results of the predicate devices. The performance of the predicate devices is considered to be the "truth" against which the new device is measured.

    The testing for the Otogram CA3350 was performed by a "Hearing Instrument Specialist Intern." While this individual is trained on equipment use, their specific qualifications are not detailed beyond "Intern," and they are explicitly stated not to be an audiologist. This suggests the testing focused on operational equivalence and consistency with established devices rather than independent expert clinical diagnosis for ground truth.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set. Since the evaluation appears to be a direct comparison of the new device's output against predicate device outputs (rather than independent diagnostic assessments with potential disagreements), an adjudication process would not be necessary in the conventional sense. The "ground truth" is considered to be the established performance of the predicate devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned or implied. The submission focuses on device equivalence, not on the impact of AI or assisted reading on human performance. The device is an audiometric instrument, not an AI-powered diagnostic imaging tool that typically involves MRMC studies to evaluate reader improvement.

    6. Standalone Performance Study

    Yes, a form of standalone performance was likely done. The "comparison testing with predicate devices" and "bench, user, and laboratory testing" evaluate the Otogram CA3350's performance as a standalone device against the established performance of the predicate devices. The device is intended to perform its functions independently, similar to how the predicate devices operate. The "algorithm only" aspect isn't directly applicable as it's a hardware/software system, but its output is evaluated independently.

    7. Type of Ground Truth Used

    The primary "ground truth" used for this submission is the established performance and specifications of the legally marketed predicate devices (Madsen Zodiac 901 and Bio-Logic AuDX). The Otogram CA3350 claims substantial equivalence by demonstrating that its technical characteristics, performance, and intended use are as safe and effective as these predicates, or improve upon them without raising new questions of safety or effectiveness.

    8. Sample Size for the Training Set

    The document does not mention a training set sample size. This type of device (audiometer, otoacoustic emissions analyzer, tympanometer) does not typically rely on complex machine learning models that require large labeled training sets in the same way an AI-powered image analysis system would. The device's functionality is based on established acoustic and electrical engineering principles, standardized measurement techniques, and calibrated hardware. Any "training" would be in the sense of calibrating the device to known standards, rather than training a machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since there's no mention of a traditional machine learning "training set" for the device, the concept of establishing ground truth for it is not applicable in that context. If the question implicitly refers to the accuracy and calibration of the device's measurements, then the ground truth would be established through:

    • Adherence to recognized national and international standards: IEC 1027, ANSI S3.39, IEC 60645-5 for performance and calibration, and IEC 60601 series for electrical safety. These standards define the expected accurate behavior and output of such devices.
    • Calibration with certified reference equipment: During manufacturing and maintenance, the device would be calibrated against instruments that meet primary metrological standards to ensure its measurements are accurate.

    In summary, the K070118 submission demonstrates substantial equivalence by comparing the Otogram CA3350's features, intended use, technical specifications, and general performance against legally marketed predicate devices. The "ground truth" for this process is primarily derived from the well-established safety and effectiveness of those predicate devices and compliance with relevant industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041853

    Validate with FDA (Live)

    Device Name
    OTOGRAM
    Manufacturer
    Date Cleared
    2004-09-03

    (57 days)

    Product Code
    Regulation Number
    874.3310
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    "Otogram" can perform the functions of:
    Pure tone and speech audiometer, 1.
    Impedance audiometer (tympanometry, acoustic reflex), and 2.
    Distortion product otoacoustic emissions analyzer. 3.
    The unit is intended for use by a qualified/trained audiologist on both adult and pediatric subjects for measurement of acoustic impedance.

    Device Description

    The Otogram is a computer-controlled, audiometric instrument combining the functions of an Audiometer, a Impedance Instrument, and a DPOAE Instrument. The device is controlled through the use of ASCII commands transmitted over a standard RS-232 communication port. Control software is installed on the supplied PC. The device performs comprehensive audiometry; tympanometry, acoustic reflex; and otoacoustic emissions. Additional features include manual audiometer and NOAH v.3 compliant (hearing aid fitting software). Available in multiple languages, including English, Spanish, Russian, Mandarin, Vietnamese, and Korean.

    AI/ML Overview

    The provided text describes the regulatory submission for the Tympany Otogram device and its comparison to predicate devices, focusing on substantial equivalence rather than a study proving the device meets specific acceptance criteria in a clinical performance sense.

    Therefore, many of the requested elements (like sample size for test/training sets, number and qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and type of ground truth) either do not apply directly to this type of submission or are not detailed in the provided information. This document is a 510(k) summary, which focuses on demonstrating equivalence to existing legally marketed devices based on technical characteristics and intended use, often supported by bench and user testing rather than extensive clinical trials with human subjects.

    Here's an attempt to answer the questions based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state "acceptance criteria" in the format of specific quantitative thresholds to be met. Instead, it demonstrates "substantial equivalence" to predicate devices by comparing technical characteristics. The "reported device performance" is essentially the detailed technical specifications of the Otogram and its comparison to the predicate devices.

    The table below reconstructs the comparison from the provided Substantial Equivalence Chart as effectively serving the role of demonstrating performance relative to established benchmarks (the predicate devices). The implied "acceptance criterion" for each characteristic is that the Otogram's performance/specification is comparable to or within acceptable ranges of the predicate devices.

    CharacteristicAcceptance Criterion (Predicate Devices)Reported Device Performance (Tympany Otogram)
    Probe tone:
    Probe tone frequencyMadsen Zodiac 901: 226 Hz226 Hz, +/- 3%
    Probe tone levelMadsen Zodiac 901: 85 dB SPL70 dB SPL, +/- 3 dB
    Admittance measurement:
    Total rangeMadsen Zodiac 901: 0.1 ml to 8.0 ml0.2 ml to 6.0 ml
    Reflex rangeMadsen Zodiac 901: 0 ul to 375 ul0 ul to 375 ul
    CalibrationMadsen Zodiac 901: 2 cc cavity0.5, 2, 5 cc cavity
    Pressure system:
    Pump systemMadsen Zodiac 901: Plunger type; stepper motor, digitally controlledPeristaltic pump; stepper motor, digitally controlled
    Pressure range - normalMadsen Zodiac 901: +200 to -400 daPa+200 to -400 daPa
    Pressure range - extendedMadsen Zodiac 901: +400 to -600 daPaNA
    AccuracyMadsen Zodiac 901: +/- 10% or +/- 10 daPa (whichever is greater)+/- 15% or +/- 10 daPa (whichever is greater)
    Pump speedMadsen Zodiac 901: As fast as possible: 400 daPa/sec, 200 daPa/sec, 100 daPa/sec, 50 daPa/sec50 daPa/sec to 150 daPa/sec
    Air releaseMadsen Zodiac 901: Mechanical safety release valve set to +600 and -800 daPa. Manual & automatic air releaseMechanical safety release valve set to +600 and -800 daPa. Manual & automatic air release
    Ipsi/contralateral stimuli:
    Pure toneMadsen Zodiac 901: 0.5, 1, 2, 4 kHz0.5, 1, 2, 4 kHz
    Step sizeMadsen Zodiac 901: 1, 2, 5, 10 dB1 to 10 dB
    Attenuator range, ipsi lateral stimulusMadsen Zodiac 901: 0.5, 1.0, 2.0, kHz 50-115 dB HL, 4.0 kHz 50-90 dB HL0.5, 1.0, 2.0, 4.0 kHz 80-105 dB HL
    Attenuator range, contra lateral stimulusMadsen Zodiac 901: 0.5, 1.0, 2.0, 4.0 kHz 50-100 dB HL, white noise0.5, 1.0, 2.0, 4.0 kHz 80-100 dB HL
    Frequency accuracyMadsen Zodiac 901: +/- 0.5% (also Bio-Logic AuDX: +/- 0.5%)+/-0.5%
    Distortion Product Otoacoustic Emissions:
    F2 test frequencyBio-Logic AuDX: 2.0, 3.0, 4.0, 5.0 kHz1.0, 1.5, 2.0, 3.0, 4.0 kHz
    Stimulus level rangeBio-Logic AuDX: 55 dB SPL and 65 dB SPL45-70 dB SPL
    Standard and safety characteristics:
    Performance and calibrationIEC 1027, ANSI S3.39 (for Madsen Zodiac 901)IEC 1027, ANSI S3.39
    Electrical safetyEN 60601-1, class I, type B (for Madsen Zodiac 901); EN 55011: 1991 Group 1 Class B (for Bio-Logic AuDX)EN 60601-1, class I, type B

    The study proving the device meets these criteria is described as "bench, user, and laboratory testing" (Section 5. and 7.). These tests indicate that the new device is "as safe and effective as the predicate device," thus demonstrating substantial equivalence.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document mentions "bench, user, and laboratory testing" but does not quantify the sample size (e.g., number of devices tested, number of users, number of measurements).
    • Data Provenance: Not specified. It's internal company testing ("Tympany Inc."). There is no indication of country of origin or whether the testing was retrospective or prospective in detail, although "user testing" implies prospective interaction with the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not specified. This is a technical medical device, and the "ground truth" for its performance is typically established by comparing its physical outputs (e.g., sound frequencies, pressure measurements) against calibrated standards and predicate devices, rather than expert consensus on a diagnosis for a test set. The submission focuses on technical performance and safety.

    4. Adjudication method for the test set

    Not applicable/Not specified. Adjudication methods like 2+1 or 3+1 are typically used in studies where human readers evaluate cases and their agreement needs to be resolved. This document describes bench and user testing of a medical device's technical specifications and safety, not an evaluation of diagnostic accuracy by human readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This type of study (MRMC, AI assistance) is not mentioned and is typically associated with diagnostic algorithms or imaging interpretation, which is not the primary focus of this device's regulatory submission. The Otogram is a measurement device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable for this device in the context of "algorithm only." The performance of the device itself (its hardware and embedded software) was tested, implying a "standalone" evaluation of its technical capabilities against its specifications and predicate devices. However, this is not in the context of an "AI algorithm." The device's operation is "computer-controlled" (Section 4), but the submission focuses on its measurement functions, not on a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's technical performance would be established by calibrated standards and the specifications of predicate devices. For example, to verify "Probe tone frequency: 226 Hz, +/- 3%", the device's output would be measured against a known, highly accurate frequency standard. For safety, compliance with standards like EN 60601-1 is the ground truth. There's no mention of patient-specific "outcomes data" or "pathology" as ground truth for this type of device submission.

    8. The sample size for the training set

    Not applicable. The device described (Otogram) is an audiometer, impedance instrument, and DPOAE instrument. It is a measurement device, not an AI/machine learning model that requires a "training set" in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable (as established in point 8).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1