510(k) Data Aggregation

    K Number
    K190864

    Validate with FDA (Live)

    Date Cleared
    2020-08-27

    (512 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KONIX Anti-Fog solution is a sterile, disposable, functional, anti-fog solution. The antifog solution is intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens. KONIX Antifog Solution is used to prevent the lenses of the imaging device from fogging due to the difference in body temperature before and during endoscopic and laparoscopic procedures.

    Device Description

    Konix Anti-Fog Solution is a single-use, sterile, and biocompatible laparoscopic accessory device. The product is released to the market as sterile and disposable with one polyurethane foam pad in a 6 ml polyethylene bottle. Secondary packaging is composed of PET film and Medical Kraft Sealing Paper.

    Konix Anti-Fog Solution is intended to be used to prevent the lenses of the imaging device from fogging due to the difference in body temperature before and during endoscopic and laparoscopic procedures.

    The mechanism of defogging or anti-fogging action of Konix Anti-Fog Solution, works by physically changing the water droplets that form on the lens surface from round droplets, to a flat transparent sheet of water, a phenomenon known as 'wetting'.

    AI/ML Overview

    The provided text describes the Konix Anti-Fog Solution, a device intended to prevent fogging of endoscope and laparoscope lenses. The document is a 510(k) premarket notification summary submitted to the FDA. While it highlights safety and effectiveness, it does not include acceptance criteria or a "study that proves the device meets the acceptance criteria" in the way one might expect for a diagnostic AI/ML device.

    Instead, the performance data presented focuses on biocompatibility, sterilization validation, shelf-life, and a general "fog resistance test." The "acceptance criteria" for these tests are implied by "Pass" results against established standards.

    Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided text for certain categories:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/Criteria CategoryAcceptance Criteria (Implied by Standards)Reported Device Performance
    BiocompatibilityMeet requirements of ISO 10993 series (Cytotoxicity, Sensitization, Intracutaneous Reactivity)All tests passed the requirements as indicated in the applicable standards.
    Cytotoxicity TestISO 10993-5:2009 standardsPass
    Sensitization TestISO 10993-10:2010 standardsPass
    Intracutaneous Reactivity TestISO 10993-10:2010 standardsPass
    Sterilization ValidationAchieve Sterility Assurance Level (SAL) < 10-6 according to ISO 11137-2, ISO 11737-1, ISO 11737-2Controlled at the dose of no less than 13.9 kGy, meeting SAL < 10-6.
    Shelf-Life (3 years accelerated aging)Maintain package integrity, seal strength, sterility, and product stability (fog resistance) according to ASTM F1980-16, ASTM F1929-15, ASTM F88/F88M-15, ISO 11737-2.All aging test data showed the product to be safe and effective for 3 years.
    Package Integrity TestASTM F1929-15 standardsPass
    Seal Peel Strength TestASTM F88/F88M-15 standardsPass
    Product Sterility TestISO 11737-2:2019 standardsPass
    Product Stability Studies (Fog Resistance Test)Internal testing protocol met (specific criteria not detailed)Pass (stated that "Konix Antifog Solution is effective" based on this test)

    Regarding the other requested information that is typically relevant for AI/ML devices, this document does not contain that information as it describes an anti-fog solution, not an AI/ML diagnostic device. Therefore, the following points cannot be answered from the provided text:

    1. Sample size used for the test set and the data provenance: Not applicable. The performance tests are laboratory-based for physical and biological properties.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. "Ground truth" for this device refers to meeting established physical, chemical, and biological safety and efficacy standards.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140506

    Validate with FDA (Live)

    Date Cleared
    2015-04-17

    (414 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JoyDrops® Natural Personal Lubricant Gel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.

    Device Description

    JoyDrops® Natural Personal Lubricant Gel is a water-based personal lubricant gel that is contained in a 3.38 fluid ounces (100 mL) polymer container with a pump dispenser.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: Joy Drops™ Natural Personal Lubricant Gel. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving the device meets acceptance criteria through clinical studies involving human readers or AI.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for a diagnostic device utilizing AI is not applicable to this document.

    The document primarily focuses on:

    • Biocompatibility testing: To ensure the product is safe for contact with human tissue.
    • Condom compatibility testing: To ensure it doesn't degrade various types of condoms.
    • Comparison to a predicate device: To demonstrate substantial equivalence based on intended use, technological characteristics, and safety/performance data.

    Here's an attempt to answer the questions based only on the information available for this specific product, acknowledging that many questions are not directly relevant to a lubricant's 510(k) submission:


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (from text)Reported Device Performance (from text)
    Condom Compatibility:
    Compatible with latex condoms (per ASTM D7 6661-10 and ARDL method)Found to be compatible with latex condoms.
    Compatible with polyisoprene condoms (per ASTM D7 6661-10 and ARDL method)Found to be compatible with polyisoprene condoms.
    Compatible with polyurethane condoms (per ASTM D7 6661-10 and ARDL method)Found to be compatible with polyurethane condoms.
    Biocompatibility (Safety):
    No cellular toxicity (per ISO 10993 methodology)No cellular toxicity was found.
    No systemic toxicity (per NAMSA protocol, FDA approved)No systemic toxicity was found. (Table: No Mortality or Evidence of Systemic Toxicity)
    Non-irritant to vaginal tissue (per ISO 10993-10)The test article was considered a non-irritant to vaginal tissue. (Table: No Irritation Observed)
    No skin sensitization (per ISO 10993-10/AI standard practices)No skin sensitization was reported. (Table: No Skin Sensitization was Observed for both animal and human patch tests)
    No skin irritation (Human Patch Test, 30 min and 24 hours)No Skin Irritation was Observed (for both 30 minutes and 24 hours).
    Stability:
    Shelf-life of three yearsA shelf-life of three years was established based on data from temperature control cabinets (22°C and 60°C).
    Microbiological specifications <10 cfu/mL-gMicrobiological specifications were set at <10 cfu/mL-g based on stability data.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the condom compatibility or biocompatibility tests (e.g., number of condoms tested, number of animals, or human subjects for patch tests). It refers to "standard practices" and ISO methodologies.

    • Data Provenance: The document does not specify the country of origin of the data for the tests. It refers to "an independent laboratory" for stability testing. The submitter is from Turkey (Turkuaz Sağlik Hizmetleri Medikal Temizlik Kimyasal Urünler San. ve Tic. LTD ŞTI). The studies appear to be prospective, laboratory-based tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The device is a personal lubricant, not a diagnostic device requiring expert interpretation of results to establish ground truth for a test set. The tests performed (biocompatibility, condom compatibility) are objective laboratory measurements, not subjective expert evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 are used for diagnostic interpretations by experts, which is not relevant for this product's testing. Laboratory tests have defined protocols and result interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable, as the device is a personal lubricant and does not involve AI, human readers, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable, as the device is a personal lubricant and does not involve an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This question is not applicable in the context of diagnostic "ground truth." The "ground truth" for this product's performance refers to the objective results of standardized laboratory tests for:

    • Condom integrity: Measured by established ASTM and ARDL methods.
    • Biocompatibility: Measured by ISO 10993 standards for cytotoxicity, irritation, sensitization, and systemic toxicity.
    • Shelf-life: Measured by stability studies under controlled temperature conditions.

    8. The sample size for the training set

    This question is not applicable, as the device does not employ machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as the device does not employ machine learning or AI, and therefore does not have a "training set" or "ground truth for a training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K130581

    Validate with FDA (Live)

    Date Cleared
    2014-06-18

    (470 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As stated on Label: Contact medium for ultrasonic and electrical transmission gel.

    Konix® Sterile Gel is sterilized by gamma irradiation and intended for use in all diagnostic ultrasound procedures that currently use an ultrasound coupling gel or other fluid, alone or in combination with a transducer cover, where sterlility and in vivo biocompatibility are required. It can be used on the skin where the risk for infections is especially high, especially with oven tissue Doppler applications; and can be used for cystoscopic and ultrasound examinations.

    Konix® Serile Gel can be used on the skin or in open tissue where there is a risk for infections, such as in open tissue Doppler applications. It can be used during surgery; for cystoscopic and vaginal ultrasound and Doppler examinations; for cardiac surgery and ultrasound procedures; and as a lubricant in biopsy applications

    Device Description

    Konix® Sterile Gel is an in vivo biocompatible sterile couplant intended for use as a scanning gel in surgical procedures, biopsies and similar sterile applications. It is used to couple sound waves between the patient and medical imaging electronic transducers during intraoperative and intracavitary medical diagnostic ultrasound imaging procedures such as transcutaneous ultrasound image guided biopsy and aspiration, doppler examinations of organ and tissue intraoperative ultrasound imaging, intra-cavity ultrasound imaging, and gel infusion sonography.

    Major characteristics include:

    • No toxic effects; produced as a completely harmiess material .
    • Hypoallergenic; non-irritating; no cytotoxicity .
    • Water soluble, non-staining and easily cleanable
    • Does not contain any lipid components .
    • Free from formaldehyde and salt .
    • Odorless .
    • Gamma radiated to assure sterility .
    • Will not damage the probe .
    • Does not contain air bubbles .
    • pH = 7 .
    AI/ML Overview

    The provided document is a 510(k) Summary for the Konix® Sterile Gel, an ultrasound coupling gel. The submission is not for an AI/ML device, but for a medical device that establishes substantial equivalence to a predicate device. Therefore, many of the typical acceptance criteria and study design details for AI/ML performance evaluation (like sample size for test sets, data provenance, ground truth establishment by experts, MRMC studies, standalone performance, etc.) are not applicable here.

    Instead, the submission focuses on demonstrating the physical, chemical, and biological properties of the Konix® Sterile Gel are substantially equivalent to a legally marketed predicate device (Sterile Aquasonic® 100 Ultrasound Gel).

    Here's an analysis of the provided information, framed as closely as possible to your requested structure, while acknowledging the nature of this submission:


    Acceptance Criteria and Device Performance for Konix® Sterile Gel

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for Konix® Sterile Gel are implicitly defined by its physical, chemical, and performance characteristics being comparable to, or within acceptable ranges relative to, the predicate device and established standards for sterile ultrasound gels.

    CharacteristicAcceptance Criteria (Predicate/Standard)Reported Konix® Sterile Gel PerformanceMeets Criteria?
    General PropertiesSterile, Hypoallergenic, Non-cytotoxic, Non-sensitizing, Water soluble, Non-staining, Easily cleanable, No lipid components, Free from formaldehyde and salt, Odorless, Will not damage probe, No air bubblesAll reported as met/presentYes
    pHPredicate: 6.5 - 6.95; Release Specification: 6.50-7.506.5 ± 0.75Yes
    Density (g/cm³)Predicate: 1.03; Release Specification: 0.683-1.2830.983Yes
    Acoustic Impedance (10⁵gm/cm²)Predicate: 1.60; Close to human dermis (1.6 kg/m²s)1.45 (1.49 kg/m²s in performance test)Yes (similar)
    Speed of Sound (m/s)Close to skin (1518 m/s)1516Yes (similar)
    Viscosity (cp)High viscosity (Predicate is viscous); Performance: 100,000-200,000100,000-200,000Yes
    SterilityGamma radiated to assure sterility; Zero CFU/GmConfirmed sterile post-radiationYes
    Shelf-LifeStable for three yearsStable for three yearsYes
    CytotoxicityNon-cytotoxicProved negativeYes
    SensitizationNon-sensitizingProved negativeYes
    IrritationNon-irritatingProved negativeYes
    Systemic ToxicityNot toxicProved negativeYes

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set: Not applicable in the context of an AI/ML performance evaluation with a "test set" of cases. The "testing" here refers to laboratory and biocompatibility tests of the gel's properties.
    • Data Provenance: The document does not specify the country of origin for the reported test results, but the submitter is from Turkey. The data appears to be from laboratory testing related to the product's physical, chemical, and biological characteristics, performed prospectively as part of product development and regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable. The "ground truth" for this device relates to established scientific and medical standards for gel properties (e.g., pH, density, sterility) and biocompatibility testing interpretation, not expert consensus on medical image interpretations. The "ground truth" is determined by validated laboratory methods and standard toxicology protocols.

    4. Adjudication Method for the Test Set

    • Not applicable. This concept relates to resolving disagreements among multiple expert readers in diagnostic classification tasks, which is not relevant for evaluating the physical and biological properties of an ultrasound gel.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • No. This is not an AI/ML device. MRMC studies are specific to evaluating human reader performance with and without AI assistance in diagnostic tasks.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This term is specific to AI/ML algorithm performance. The Konix® Sterile Gel's "standalone" performance is its intrinsic physical, chemical, and biological properties, which were characterized through various laboratory tests as detailed in the "Performance Test," "Sterility and Shelf-Life," and "Additional Tests" sections.

    7. The Type of Ground Truth Used

    The ground truth used for evaluating Konix® Sterile Gel consists of:

    • Laboratory Measurements: Quantifiable physical and chemical properties (pH, density, viscosity, acoustic impedance, speed of sound).
    • Microbiological Standards: Demonstrated absence of microbial growth (sterility) confirmed by standard culture methods.
    • Biocompatibility Protocols: Results from in vitro cytotoxicity, in vivo sensitization, irritation tests, and mouse systemic toxicity tests, which are assessed against established pass/fail criteria from standard biological evaluation frameworks (e.g., ISO 10993).
    • Shelf-Life Stability: Demonstrated stability over time for specified properties.

    8. The Sample Size for the Training Set

    • Not applicable. There is no "training set" in the context of an AI/ML model for this device. The development of the gel involved formulation and testing, not machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI/ML model, there is no ground truth establishment in that sense. The "ground truth" for the gel's formulation and development would involve standard chemical engineering and pharmaceutical development processes, ensuring ingredients meet quality standards, and the final product has the desired characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101952

    Validate with FDA (Live)

    Date Cleared
    2011-05-27

    (319 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Konix® Ultrasound Gel is intended for general use as a nonsterile transmission media for acoustically coupling a transducer to a human body surface during external, diagnostic ultrasound imaging procedures. It is placed on the patient's skin prior to initiating an ultrasound examination. It is indicated for prescription use only.

    Device Description

    Konix® Ultrasound Gel consists of deionized water, carbomer, triethanolamine, monopropylene glycol, 5-chloro-2-methyl-4-isothiazolin-3-one and 2-methyl-4-isothiazolin-3 one and is a type of conductive medium, i.e., scanning gel, used in ultrasound diagnostic techniques. A scanning gel acts as a couplant that provides an acoustic pathway between the transducer and the skin. In addition, the gel eliminates air (a disruptive influence) from the interface and adapts the contours of the probe to the skin.

    The major characteristics of Konix® Ultrasound Gel include:

    • . Hypoallergenic, non-irritating
    • Water soluble, non-staining and easily cleanable .
    • Does not contain oil and fatty matter .
    • Free from formaldehyde and salt .
    • No toxic effects .
    • Produced as a completely harmless material .
    • No smell ◆
    • Vacuum treated production .
    • Not-damageable to the probe .
    • . Does not contain air bubbles
    • pH level is 7 .
    AI/ML Overview

    The Konix® Ultrasound Gel is a medical device designed for general use as a non-sterile transmission medium for acoustically coupling a transducer to a human body surface during external, diagnostic ultrasound imaging procedures. The device's acceptance criteria and studies are primarily focused on demonstrating its acoustic performance and equivalence to existing predicate devices.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for Konix® Ultrasound Gel were based on its acoustic properties being virtually identical to human skin and other commonly used coupling gels in the United States.

    Acceptance Criteria / CharacteristicTarget Value (based on human skin/optimal average)Reported Device Performance (Konix® Ultrasound Gel)
    Sound velocity (m/sec) at 30°C1,518 (human skin), 1,520 (optimal average)1,516
    Density (kg/m³) at 30°CNot specified as clear target, but comparable0.98 x 10-3
    Acoustic impedance (kg/m² sec) at 30°C1.6 (human skin)1.49
    Attenuation coefficient as a function of frequency, a/f (dB/(cm-MHz))Not explicitly stated, but "low" is desired< 0.05

    2. Sample Size and Data Provenance

    The document does not specify a separate "test set" and corresponding sample size for the acoustic performance evaluation. The acoustic properties (Sound velocity, Density, Acoustic impedance, Attenuation coefficient) were measured for the Konix® Ultrasound Gel.

    The data provenance for the comparison values (e.g., human skin sound velocity and acoustic impedance) is from published scientific literature:

    • Ogura, I., Kidikoro, T., linuma, K., Tanaka, K. and Matsuda, A. "Measurement of Acoustic Impedance of Skin" Ultrasound in Medicine, Vol. 4, RC 78.7, U4 A 5a, page 535 (1978).

    The study for Konix® Ultrasound Gel itself appears to be a prospective measurement of its physical properties. The country of origin of the device manufacturer is Turkey.

    3. Number of Experts and Qualifications for Ground Truth

    This type of device (ultrasound gel) does not typically involve expert review for establishing ground truth in the way medical imaging algorithms do. The "ground truth" for comparison for acoustic properties comes from established scientific literature regarding human skin properties. Therefore, there were no "experts" in the traditional sense involved in establishing ground truth for a test set of images or clinical outcomes.

    4. Adjudication Method

    Not applicable. The study involved direct measurements of the physical properties of the gel and comparison to published scientific data. There was no adjudication of expert opinions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This type of study is suitable for AI algorithms that assist human readers in interpreting medical images, which is not applicable to an ultrasound gel.

    6. Standalone Performance Study

    Yes, a standalone study was performed. The non-clinical performance evaluation focused on directly measuring the acoustic properties of the Konix® Ultrasound Gel itself, rather than its performance in conjunction with human interpretation. The study states: "Konix® Ultrasound Gel was evaluated for its acoustic performance."

    7. Type of Ground Truth Used

    The ground truth used was established scientific literature and physical properties of human skin and other commonly used ultrasound coupling gels. For example, the reported longitudinal velocity of skin (1,518 m/sec) and acoustic impedance of skin (1.6 kg/m²s) were referenced from a scientific publication (Ogura, 1978).

    8. Sample Size for the Training Set

    Not applicable. This device is an ultrasound gel, not an AI algorithm that requires a training set. The "study" here involves physical property measurements and comparison, not machine learning model training.

    9. How Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI algorithm, there was no training set or ground truth established for such a set.