510(k) Data Aggregation

    K Number
    K181817

    Validate with FDA (Live)

    Date Cleared
    2019-03-26

    (260 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The transGlideXT Expandable Introducer is intended to be inserted into the femoral artery, over a guidewire, and once expanded, to provide a guide for catheters and/or devices introduced into the femoral iliac arteries.

    Device Description

    The transGlideXT Expandable Introducer (also referred to as "transGlideXT") is the next generation expandable Mesh Introducer consisting of a polymer braid which expands to accommodate catheters and/or devices with profiles up to 22F. The Mesh Introducer, with an indwelling Dilator (also referred to as "Mesh Assembly"), is inserted into the femoral artery, over a 0.035" guidewire (or smaller). The device is introduced at a small diameter of 12F inner diameter (ID) and is designed with a hydrophilic coating on the outer diameter (OD) and ID along the 30 cm usable length, thus facilitating passage through the femoral artery and passage of the commercial catheter and/or device. The proximal 11 cm of the Mesh usable length is sealed to prevent blood loss at the access site. Once at the target location, the Dilator is removed and the Mesh Introducer provides a bearing surface for the commercial catheter and/ or device, which is designed to reduce the axial forces applied to the artery wall while the commercial catheter and/or device is being inserted.

    The transGlideXT Expandable Introducer is a sterile, non-pyrogenic, single-use prescription device. The transGlideXT Expandable Introducer does not supply but recommends use with commercially available 0.035" (or smaller) Guidewires.

    AI/ML Overview

    The TransAortic Medical, Inc. transGlideXT Expandable Introducer (K181817) aims to establish substantial equivalence to its predicate device, the transGlide Expandable Introducer (K152194). The device is intended to be inserted into the femoral artery, over a guidewire, and once expanded, to provide a guide for catheters and/or devices introduced into the femoral iliac arteries.

    Here's an analysis of the acceptance criteria and the study conducted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting specific numerical acceptance criteria and performance data in a direct table format. However, it outlines various performance tests conducted and states that the results confirmed safety and functionality comparable to the predicate.

    Acceptance Criteria CategorySpecific Test/EvaluationReported Device Performance
    BiocompatibilityCytotoxicity (MEM Elution), Sensitization (Magusson-Kligman), Irritation (Intracutaneous Toxicity), Systemic Toxicity (Systemic Injection), Hemocompatibility (Thrombogenicity, Complement Activation, Partial Thromboplastin Time, Hemolysis), Pyrogenicity (Material Mediated Pyrogen, Bacterial Endotoxins-LAL)All biocompatibility tests (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity) were performed. The implicit acceptance criterion is that the device demonstrates biocompatibility comparable to or better than the predicate device. The results are not explicitly stated, but the conclusion of substantial equivalence implies these criteria were met.
    Bench TestingVisual Inspection and Dimensional VerificationPerformed. Implied compliance with design specifications.
    Flush TestingPerformed. Implied proper fluid dynamics.
    Simulated Use (Advancement, Dilator Removal, Retraction & Inspection)Performed. Implied successful device operation during simulated clinical scenarios.
    Leak Testing (BS EN ISO 11070:2014)Performed. Implied structural integrity and prevention of leakage.
    Bend/Kink Resistance TestingPerformed. Implied resistance to kinking during use.
    Radiopacity TestingPerformed. Implied visibility under fluoroscopy.
    Interventional Device Advancement and RemovalPerformed. Implied compatibility with other interventional devices.
    Hydrophilic Coating LubricityPerformed. Implied smooth advancement through vessels.
    Hydrophilic Coating DurabilityPerformed. Implied coating integrity over time.
    Hydrophilic Coating Particulate CharacterizationPerformed. Implied minimal particulate shedding.
    Hub to Sheath RotationPerformed. Implied proper rotational functionality of components.
    Tensile Tests (BS EN ISO 11070:2014)Performed. Implied mechanical strength.
    Corrosion Testing (BS EN ISO 11070:2014)Performed. Implied resistance to degradation in biological environment.
    Packaging Validation (BS EN ISO 11607-:2009 + A1:2014)Performed. Implied maintenance of sterility and integrity during storage and transport.
    Sterilization Validation (ANSI/AAMI/ISO 11135:2014, ANSI/AAMI/ISO 10993-7:2008/(R) 2012)Performed. Implied effective sterilization to a specified sterility assurance level (SAL).
    Shelf LifePerformed. Implied maintenance of device properties over its specified shelf life.
    Animal StudiesIn vivo chronic and acute studies in ovine modelSafety Acceptance Criteria: Met based on pathology and histopathology results. Performance Observations: Functional and safe for intended use, comparable to predicate. The specific numerical acceptance criteria for these studies are not detailed but are stated as "consistent with the animal testing performed and submitted for the predicate (K152194)."

    2. Sample Sizes and Data Provenance:

    • Test Set (Animal Studies): The document states "A series of in vivo evaluations were performed." The exact number of animals in the ovine model is not specified, but it implies multiple animals were used for both acute and chronic studies.
    • Data Provenance: Prospective animal study conducted in an ovine (sheep) model. The country of origin is not explicitly stated.

    3. Number of Experts and their Qualifications for Ground Truth - Test Set:

    • The document mentions "pathology and histopathology results" were used for safety acceptance criteria in the animal study. This implies that veterinary pathologists were involved in establishing the ground truth.
    • The exact number and specific qualifications (e.g., years of experience) of these experts are not provided.

    4. Adjudication Method for the Test Set:

    • The document does not specify an adjudication method for the animal study (e.g., use of multiple independent evaluators with a specific tie-breaking rule). It simply states that pathology and histopathology results were used to determine if safety criteria were met. This suggests a standard pathology review process rather than a complex adjudication method for comparing interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is typically used for diagnostic imaging devices to evaluate human reader performance with and without AI assistance. The transGlideXT is an introducer device, and its evaluation focuses on physical performance, biocompatibility, and procedural safety/effectiveness rather than diagnostic accuracy.
    • Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.

    6. Standalone Performance Study:

    • Yes, the performance testing described (biocompatibility, bench testing, and animal studies) represents a standalone evaluation of the device's performance. The device's functionality, safety, and performance characteristics were assessed independently, comparing them against established standards and the predicate device, rather than only in conjunction with human-in-the-loop performance.

    7. Type of Ground Truth Used:

    • For the animal studies, the ground truth was established through pathology and histopathology results. This involves microscopic examination of tissues to identify and characterize any device-related adverse effects or tissue reactions.
    • For bench testing, the ground truth is based on engineering specifications and established test standards (e.g., BS EN ISO 11070:2014, ANSI/AAMI/ISO 11135:2014).

    8. Sample Size for the Training Set:

    • This information is not applicable to the type of device and study described. The transGlideXT Expandable Introducer is a physical medical device, not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of machine learning. The studies described are physical performance validations.

    9. How Ground Truth for the Training Set Was Established:

    • This information is not applicable as there is no training set for an AI/machine learning algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152194

    Validate with FDA (Live)

    Date Cleared
    2016-04-07

    (246 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The transGlide Expandable Introducer is intended to be inserted into the femoral artery, over a guidewire, and once expanded, to provide a guide for catheters and/or devices introduced into the femoral iliac arteries.

    Device Description

    The transGlide Expandable Introducer consists of an expandable Mesh Assembly and a flexible Sheath Assembly. The Mesh Assembly, with an indwelling Dilator, is inserted into the femoral artery over a 0.035" (or smaller) guidewire. The Mesh Assembly is introduced at a small diameter of 13F inner diameter (ID) and is designed with a hydrophilic coating on the outer diameter (OD) of the usable length (effective length), thus facilitating passage through the femoral artery. The proximal 11cm of the Mesh is sealed to prevent blood loss at the access site. Once at the target location, the Mesh Dilator is removed and the Sheath Assembly is inserted, with an indwelling Dilator, through the Mesh Assembly over the guidewire.

    Prior to removal, the Sheath Assembly is withdrawn through the Mesh Assembly until the Sheath Removal Indicator is visible just proximal to the Docking Port, leaving 8cm of Sheath usable length in place to prevent blood loss while the entire assembly is withdrawn and removed from the patient.

    The Mesh consists of a polymer braid, which expands to accommodate the profile of the Sheath. The Mesh provides a bearing surface for the Sheath, which is designed to reduce the axial forces applied to the artery wall while the Sheath is being inserted. The Sheath Assembly Dilator is removed leaving a large (16F, 18F or 20F) central lumen extending from the proximal end to the distal end of the Sheath with a usable length of 30cm.

    The transGlide Expandable Introducer is a sterile, non-pyrogenic, single-use prescription device. The transGlide Expandable Introducer does not supply but recommends use with commercially available 0.035" (or smaller) Guidewires.

    AI/ML Overview

    The provided text describes the transGlide Expandable Introducer and its substantial equivalence to a predicate device, the Terumo SoloPath® Balloon Expandable TransFemoral Introducer (K100819). However, the document does not contain the specific acceptance criteria and detailed device performance results in a table format as requested in point 1 of your prompt. It lists the types of tests performed but not the quantitative pass/fail criteria or the actual results.

    Therefore, the following information will be provided based on what is available in the document, and will explicitly state what is not present.


    1. Table of Acceptance Criteria and Reported Device Performance

    As noted above, the document does not provide a table of acceptance criteria with specific quantitative pass/fail values, nor does it present detailed numerical device performance results for most of the listed tests. It generally states that "the safety acceptance criteria for the study were met" for the animal study, and that non-clinical testing "demonstrate[s] that the device is substantially equivalent... and functions as intended and meets design specifications."

    Here's a summary of the types of performance evaluated:

    Test CategorySpecific Tests PerformedAcceptance Criteria / Reported Performance
    Biocompatibility- Cytotoxicity: MEM Elution (L-929) - Sensitization: Magusson-Kligman Method - Irritation: Intracutaneous Toxicity (ISO) - Systemic Toxicity: Systemic Injection (ISO) - Hemocompatibility (Thrombogenicity, Complement Activation C3a and SC5b-9, Partial Thromboplastin Time, Hemolysis - Direct and Extract) - Pyrogenicity (Material Mediated Pyrogen, Bacterial Endotoxins-Limulus Amebocyte Lysate (LAL))The implication is that the device met the biocompatibility requirements for substantial equivalence. Specific criteria and results are not provided.
    Bench Testing- Visual Inspection and Dimensional Verification - Flush Testing - Simulated Use: Advancement, Dilator Removal, Retraction & Inspection - Leak Testing (BS EN 11070:1999) - Bend/Kink Resistance Testing - Radiopacity Testing - Interventional Device Advancement and Removal - Hydrophilic Coating Lubricity - Hydrophilic Coating Durability - Hydrophilic Coating Particulate Characterization - Hub to Sheath Rotation - Tensile Tests (BS EN 11070:1999) - Corrosion Testing (BS EN 11070:1999) - Packaging Validation (BS EN ISO 11607-111607-112009 + A1:2014) - Sterilization Validation (ANSI/AAMI/ISO 11137-2:2013) - Shelf LifeThe document states these tests were performed to "verify and validate the performance of the device and ensure the transGlide Expandable Introducer functions as intended and meets design specifications." Specific criteria and results are not provided.
    Animal StudiesGLP animal study in an ovine model, comparing transGlide against the predicate (Terumo SoloPath® K100819). Evaluated safety and performance."Based on pathology and histopathology results, the safety acceptance criteria for the study were met. Performance observations were made based on detailed characteristics of the device. No untoward observations were found by the clinician."

    2. Sample Size Used for the Test Set and the Data Provenance

    • Test Set Sample Size:
      • Animal Study: The document states "A GLP animal study was performed... in an ovine model." It does not specify the number of animals (sample size) used in this study.
      • Bench Testing: For individual bench tests (e.g., tensile tests, leak tests), the sample sizes are not provided.
    • Data Provenance: The document does not specify the country of origin for the data for the animal study or bench testing, but the context implies it was likely conducted in the US or under US regulatory standards (given the FDA submission). It is not stated whether the animal study was purely prospective or if any retrospective data was included. Bench testing is typically prospective, performed specifically for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Animal Study: The document mentions "pathology and histopathology results" and "observations were made by the clinician." This implies input from veterinary pathologists and clinicians, but it does not specify the number of experts or their detailed qualifications (e.g., specific years of experience, board certifications).
    • Bench Testing: Ground truth for bench testing is generally based on engineering standards and measurement accuracy, not typically human expert consensus in the same way clinical data is.

    4. Adjudication Method for the Test Set

    • Animal Study: The document does not specify an adjudication method (e.g., 2+1, 3+1) for the pathology, histopathology, or clinical observations. It simply states the results met safety acceptance criteria and no untoward observations were found.
    • Bench Testing: Adjudication methods are not typically applicable to the objective measurements derived from bench testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study typically involves multiple human readers interpreting medical images or data, with and without an AI algorithm's assistance, to measure the AI's impact on human performance. The device described (transGlide Expandable Introducer) is a physical medical device, not an AI/software as a medical device (SaMD), so an MRMC study is not relevant here.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not done. As mentioned above, this device is a physical medical device, not an AI algorithm. Therefore, an "algorithm only" performance study is not applicable.

    7. Type of Ground Truth Used

    • Animal Study: The ground truth for the animal study seems to be based on a combination of histopathology results and clinical observations by experts (veterinary pathologists and clinicians) following GLP guidelines. This is a form of expert assessment integrated with objective pathological findings.
    • Bench Testing: The ground truth for bench testing is based on established engineering standards and measurements.

    8. Sample Size for the Training Set

    • This question is not applicable as the transGlide Expandable Introducer is a physical medical device, not an AI algorithm that requires a "training set" in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable for the same reason as point 8; there is no AI training set.