Found 2 results

510(k) Data Aggregation

    K Number
    K251727

    Validate with FDA (Live)

    Date Cleared
    2025-08-29

    (85 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Red Light is intended to treat full face wrinkles.
    The Blue Light is intended to treat mild to moderate inflammatory acne.
    The Yellow Light is intended to treat wrinkles.

    Device Description

    The GLAM LED light therapy mask is a home use wearable light emitting diode phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment of full face wrinkles and mild to moderate acne of the face. The outer shell of the mask is manufactured from Polyethylene terephthalate (PET). The inner shell is a clear Polycarbonate (PC). The Light emitting diodes are mounted behind the clear Polycarbonate. The LEDs generate the light. The ear hooks are made of Acrylonitrile butadiene styrene (ABS) and the silicone goggle protect the eyes from LED lights. Unfold the ear hooks and place the mask on your face, the mask will automatically activate the light therapy mode. The LEDs produce blue, red and yellow light in the visible spectrum (Blue:415nm +/- 10nm, Yellow: 590nm +/-10nm, Red: 625nm +/-10nm.). The device works by emitting the specified wavelengths to treat full face of wrinkles or to treat mild to moderate inflammatory acne. Press the touch switch on the right ear hook twice to select the light therapy mode you want to use. Each mode operates in a 15-minute cycle. After 15 minutes, the device automatically deactivates the light mode and enters the standby mode.

    AI/ML Overview

    This FDA 510(k) clearance letter for the GLAM LED Facial Mask (TB-2386F) does not include any information regarding clinical testing, acceptance criteria for device performance related to efficacy (e.g., wrinkle reduction or acne treatment), or studies comparing the device's performance against such criteria.

    The letter focuses on the substantial equivalence based on:

    • Technological Comparison: Comparing light wavelengths, power density, and treatment time to predicates.
    • Safety Standards Compliance: Verification through non-clinical tests (electrical safety, EMC, biocompatibility, photobiological safety, software verification).

    Therefore, I cannot provide details on the specific acceptance criteria for efficacy or a study proving the device meets those criteria from the provided document. The 510(k) summary explicitly states: "No animal or clinical study is included in this submission."

    However, if we were to hypothesize what acceptance criteria and a study might look like for a device with these indications, and then illustrate what would be missing from this document:

    Hypothetical Acceptance Criteria and Performance Study (Not Found in Provided Document):

    Since the provided document explicitly states no clinical study was included, the following tables and sections represent what would typically be expected for a device claiming therapeutic efficacy, but none of this information is present in the provided 510(k) clearance letter.

    Hypothetical Acceptance Criteria and Reported Device Performance (If Clinical Data Existed)

    Acceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
    Red Light (Wrinkle Treatment):
    Primary Endpoint: A statistically significant reduction (e.g., >20%) in the appearance of fine lines and wrinkles (e.g., Fitzpatrick Wrinkle Scale, or qualitative dermatological grading) across the full face after 8-12 weeks of treatment, compared to baseline or a control group.Not reported in 510(k) letter. If available, this section would state the measured percentage reduction in wrinkles, p-values, confidence intervals, and the specific grading scale used. Example: "Mean reduction of 32% in Fitzpatrick Wrinkle Scale scores for fine lines after 10 weeks (p < 0.001, 95% CI: 28-36%)."
    Secondary Endpoint (e.g., Patient Satisfaction): A statistically significant improvement (e.g., >70% of subjects reporting improvement) in overall skin appearance, texture, or satisfaction as reported by subjects via a validated questionnaire.Not reported in 510(k) letter. If available, this would include the percentage of satisfied subjects and questionnaire scores. Example: "85% of subjects reported 'moderate' or 'significant' improvement in overall skin appearance (p < 0.001)."
    Blue Light (Mild to Moderate Inflammatory Acne Treatment):
    Primary Endpoint: A statistically significant reduction (e.g., >40%) in the mean number of inflammatory acne lesions (papules, pustules, nodules) on the face after 8-12 weeks of treatment, compared to baseline or a control group.Not reported in 510(k) letter. If available, this would state the measured percentage reduction in lesion count, p-values, and confidence intervals. Example: "Mean reduction of 55% in inflammatory lesion count after 10 weeks (p < 0.001, 95% CI: 50-60%)."
    Secondary Endpoint (Investigator Global Assessment - IGA): A statistically significant improvement (e.g., at least a 2-point reduction on a 5-point IGA scale) in at least 70% of subjects.Not reported in 510(k) letter. If available, this would specify the percentage of subjects achieving the IGA improvement. Example: "75% of subjects achieved at least a 2-point reduction on the IGA scale (p < 0.001)."
    Yellow Light (Wrinkle Treatment):
    Primary Endpoint & Secondary Endpoint: Mirroring the Red Light criteria for wrinkle treatment, demonstrating similar statistically significant reductions or improvements.Not reported in 510(k) letter. Similar to Red Light, specific efficacy metrics for wrinkle reduction with Yellow Light would be presented here. Example: "Mean reduction of 28% in Fitzpatrick Wrinkle Scale scores for fine lines after 10 weeks (p < 0.001, 95% CI: 24-32%)."
    Safety Endpoints (Common to all lights):
    No serious adverse events (SAEs) related to device use. Incidence of treatment-related adverse events (AEs) such as erythema, dryness, or irritation below a pre-defined threshold (e.g., <10% for mild, <1% for moderate/severe).Only general safety compliance (IEC standards, biocompatibility) is reported in the 510(k) letter. If clinical data existed, this would list observed AEs. Example: "No SAEs reported. Mild, transient erythema observed in 5% of subjects, resolving within 24 hours. No moderate or severe AEs."

    Details of the Hypothetical Study (Assuming Clinical Data Existed)

    As stated, the provided 510(k) summary explicitly states "No animal or clinical study is included in this submission." Therefore, the following are hypothetical details of what such a study would entail, not actual data from this submission.

    1. Sample Size and Data Provenance (Hypothetical):

      • Sample Size: (e.g., N=150 subjects for efficacy assessment, randomly allocated to device treatment and control groups)
      • Data Provenance: (e.g., Prospective, multicenter clinical study conducted in the USA.)
    2. Number of Experts and Qualifications for Ground Truth (Hypothetical):

      • Number of Experts: (e.g., 3 independent, board-certified dermatologists)
      • Qualifications: (e.g., Each dermatologist with at least 10 years of experience in aesthetic dermatology and clinical trials, trained on the specific grading scales used.)
    3. Adjudication Method (Hypothetical):

      • (e.g., 2+1 consensus: Two experts independently assessed each subject's condition. If their scores differed, a third, senior dermatologist reviewed the case to reach a consensus score. If 2 of 3 agreed, that was the final consensus.)
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical):

      • Not Applicable: For a direct-to-consumer LED mask, an MRMC study comparing human readers (e.g., radiologists interpreting images) is not relevant. The device directly treats the skin, and efficacy is assessed via clinical observation/photography and patient self-assessment.
    5. Standalone Performance (Algorithm Only) (Hypothetical):

      • Not Applicable: This device is a physical light therapy device, not an AI software algorithm (unless there was a sophisticated integrated diagnostic AI, which is not indicated here). Its "performance" would be its direct therapeutic effect, not an algorithmic output that requires standalone evaluation.
    6. Type of Ground Truth Used (Hypothetical):

      • (e.g., Expert Clinical Assessment: Standardized clinical photography, physical examination by dermatologists using validated scales (e.g., Fitzpatrick Wrinkle Scale for wrinkles, inflammatory lesion counts for acne, Investigator Global Assessment (IGA) for overall severity), and patient-reported outcomes via validated questionnaires.)
    7. Sample Size for the Training Set (Hypothetical):

      • Not Applicable: Since no AI/ML algorithm requiring a training set is mentioned or implied for efficacy assessment in the 510(k) summary, this concept does not apply here. The device's "training" for efficacy would be its clinical development and demonstration through clinical trials.
    8. How Ground Truth for the Training Set Was Established (Hypothetical):

      • Not Applicable: As above, no training set for an algorithm is relevant to the information provided. If it were a predictive AI device, the ground truth would typically be established by expert consensus (e.g., multiple dermatologists labeling images) or by linking imaging data to ground truth data (e.g., biopsy results for skin conditions).

    In summary, the provided 510(k) clearance documentation for the GLAM LED Facial Mask focuses solely on the device's substantial equivalence in design, technology, and safety compliance according to international standards for electrical safety, EMC, biocompatibility, and photobiological safety. It explicitly states that no clinical or animal studies were included in the submission to demonstrate efficacy for its intended uses (wrinkle reduction and acne treatment). The FDA cleared this device based on its similarity to previously cleared predicate devices and its adherence to safety standards, implying that its efficacy is presumed to be similar to those predicates without requiring new clinical proof for this specific submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183217

    Validate with FDA (Live)

    Date Cleared
    2019-02-15

    (88 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IPL Hair Removal Device (Model: TB-1755) is an Over the Counter device intended for the removal of unwanted body hair.

    Device Description

    IPL Hair Removal Device Model: TB-1755, a small over-the-counter device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. A personal Light-Based Hair Removal System. Emission activation is by finger switch. Device includes a treatment window head, an adaptor and goggles. The size of the device is about 185 x 72.4 x 69.2mm (H x W x D). The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows. The device contains a Xenon Lamp and a skin proximity sensor to detect appropriate skin contact. If the IPL Hair Removal Device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

    AI/ML Overview

    The provided text is a 510(k) Summary for an IPL Hair Removal Device (Model: TB-1755). This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than presenting a detailed study with specific acceptance criteria and performance data for the subject device in the way one might expect for a diagnostic or AI-driven device.

    Therefore, the requested information, specifically regarding acceptance criteria, reported performance from a study, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, is largely not present in this type of regulatory submission.

    Here's an explanation based on the provided text, highlighting what is available and what is not:


    Request for Acceptance Criteria and Device Performance Study

    The provided document, a 510(k) Summary, describes a process for market clearance based on "substantial equivalence" to legally marketed predicate devices, not on proving absolute efficacy against predetermined acceptance criteria for a novel device through a clinical study. As such, the specific information requested in your prompt detailing acceptance criteria and performance statistics of a study is not explicitly available in this document.

    The "study" mentioned mainly refers to bench testing against recognized safety and performance standards for medical electrical equipment and biological evaluation (e.g., IEC and ISO standards), and a comparison of the subject device's specifications and indications for use against predicate devices.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Implicit): The implicit acceptance criteria are that the device adheres to the relevant IEC and ISO standards listed for safety (electrical, EMC, biological evaluation) and that its performance specifications (e.g., wavelength, fluence, pulse duration) are similar to or within the ranges of the predicate devices, and do not raise new questions of safety or effectiveness.
    • Reported Device Performance: No explicit performance data (e.g., hair reduction percentage, adverse event rates in a clinical trial) from a specific study on the subject device is reported in this summary. The summary focuses on showing that the device meets standards and is comparable to already cleared devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not provided. This document does not describe a clinical test set or participant sample size for evaluating hair removal efficacy. The "tests" mentioned are primarily laboratory bench tests for compliance with electrical, electromagnetic, and biocompatibility standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. This type of information is pertinent to studies involving expert review for diagnostic accuracy or clinical outcomes, which is not the focus of this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided. The testing mentioned is for technical compliance and safety, not for clinical adjudication of outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not provided. This device is an IPL hair removal device, not a diagnostic imaging device involving human readers or AI assistance in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not provided. This is not an AI-driven device or an algorithm for standalone performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Implicit Ground Truth for Bench Tests: The "ground truth" for the bench tests is adherence to the specifications defined by the referenced IEC and ISO standards and the performance characteristics observed in the laboratory environment. For the substantial equivalence argument, the "ground truth" is that the predicate devices are legally marketed and considered safe and effective for their intended use.
    • No clinical outcomes data as ground truth is described for the subject device.

    8. The sample size for the training set

    • Not applicable/Not provided. This applies to machine learning models, which is not relevant to this device.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. As above, this is not a machine learning device.

    Summary of available information related to "testing" and "performance":

    The "Test Summary" section (Section 7) on page 5 details that the IPL Hair Removal Device's safety and performance were evaluated by lab bench testing according to several standards:

    • Safety Standards (Electrical, EMC, Home Healthcare, Specific to non-laser sources):
      • IEC 60601-1 (General Safety)
      • IEC 60601-1-2 (Electromagnetic Compatibility - EMC)
      • IEC 60601-1-11 (Home Healthcare Environment)
      • IEC 60601-2-57 (Non-laser source equipment for therapeutic/cosmetic use)
    • Biological Evaluation Standards:
      • ISO 10993-5 (In Vitro Cytotoxicity)
      • ISO 10993-10 (Irritation and Skin Sensitization)
    • Software Guidance:
      • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Indicates software control is present, as seen in the comparison table)

    The "Comparison to Predicate Device" section (Section 8) on pages 5-6 indicates that the subject device's design principles, intended use, indications for use, functions, material, and applicable standards are "very similar" to the predicate devices. Differences in specifications like wavelength range, spot size, max fluence, pulse duration, and output energy are noted but deemed not to affect safety and effectiveness because they are "similar to the predict devices', or under their ranges" and the "safety and effectiveness of the subject device is verified via tests" (referring to the bench tests mentioned above).

    Conclusion: The document confirms that the device underwent specific technical and biological safety tests to demonstrate compliance with recognized standards. However, it does not include data from a clinical effectiveness study with human participants, specific acceptance criteria for hair removal efficacy, or the other detailed study parameters requested in your prompt. The clearance is based on substantial equivalence, relying on the established safety and effectiveness of the predicate devices and the subject device's compliance with safety standards.