510(k) Data Aggregation

    K Number
    K080771

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-10

    (236 days)

    Product Code
    Regulation Number
    882.4725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Top Neuropole Needles are injection needles which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.

    Device Description

    Top Neuropole Needles are injection needles which are used for temperature controlled radiofrequency lesioning. A nerve is localized either by using electrical stimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. After injection of anesthetics the nerve may then be blocked by radiofrequency lesion. Top Neuropole Needles are supplied sterile and are labeled for Single Use Only.

    AI/ML Overview

    This submission is for a Traditional 510(k) for the Top Neuropole Needles. This device is an injection needle for nerve blocks and radiofrequency lesioning. This submission relies on substantial equivalence to predicate devices, and therefore does not contain acceptance criteria or a study to prove the device meets acceptance criteria. A summary of the device information available in the provided text is included below.

    Device Description:
    Top Neuropole Needles are injection needles used for temperature-controlled radiofrequency lesioning. They can localize a nerve using electrical stimulation through the needle or by injecting contrast medium and using radiography. After anesthetic injection, the nerve can be blocked by a radiofrequency lesion. The needles are supplied sterile and are for single use only.

    Intended Use:
    The Top Neuropole Needles are injection needles for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. They localize a nerve by electrostimulation through the needle or by injecting contrast medium with radiography. The nerve can then be blocked by local anesthetic or a radiofrequency lesion.

    Technological Characteristics:
    The Top Neuropole Needles have the same device characteristics, materials, dimensions, and intended use as their predicate devices. They have been tested to comply with applicable industry standards and US regulations.

    Conclusion:
    The intended use and performance characteristics of the Top Neuropole Needles are the same as the predicate devices, raising no new questions of safety and effectiveness. The device is substantially equivalent to the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062946

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-02-15

    (140 days)

    Product Code
    Regulation Number
    882.4725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Top Neuropole Needles are injection needles used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.

    Device Description

    Top Neuropole Needles are teflon insulated injection needles for safe thermolesion procedures and regional anesthesia. There are 5 models offered, the ST, X, RC, XE and TL. The Top Neuropole Needle models TL, XE and RC are for pain management, while models ST and X are for local anesthesia. The needles are available in multiple sizes. The needle gauge ranges from 20 to 24 and the length from 30 to 200 mm. They are for single use only and supplied sterile.

    They may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. The nerve is localized either through the needle or by injecting contrast medium through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic or a radiofrequency lesion may be made.

    AI/ML Overview

    The provided text describes a 510(k) summary for the "Top Neuropole Needles," which seeks substantial equivalence to a predicate device. This type of submission focuses on demonstrating that a new device is as safe and effective as an already legally marketed device, rather than proving performance against specific acceptance criteria through a clinical study in the way a novel AI or diagnostic device might.

    Therefore, many of the typical questions for AI/diagnostic device studies (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance) are not applicable in this context. The "study" mentioned is one that demonstrates compliance with applicable industry standards and US regulations, rather than a performance study measuring clinical endpoints.

    Here's the breakdown based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Demonstrated Equivalency to Predicate)Reported Device Performance
    Same device characteristics as predicate device (Radionics K870028)Top Neuropole Needles have the same device characteristics as the predicate device.
    Same materials as predicate deviceTop Neuropole Needles have the same materials as the predicate device.
    Same dimensions as predicate deviceTop Neuropole Needles have the same dimensions as the predicate device.
    Same intended use as predicate deviceTop Neuropole Needles have the same intended use as the predicate device: "used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made."
    Complies with applicable industry standards and US regulations regarding safety and effectivenessTop Neuropole Needles have been tested to ensure the devices comply with applicable industry standards and US regulations. The submission concludes that "The intended use and performance characteristics of the Top Neuropole Needles are the same as the predicate and raise no new questions of safety and effectiveness. The Top Neuropole Needles are substantially equivalent to the predicate device." The FDA concurred with substantial equivalence in their letter.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as the submission is a 510(k) for substantial equivalence, not a clinical performance study with a test set of patient data. The "tests" mentioned refer to engineering and biocompatibility tests to meet standards, rather than tests on patient samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth, in the sense of clinical diagnoses or interpretations, is not established for this type of device submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. There is no "test set" in the context of clinical images or data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. No MRMC study was conducted as this device is a physical medical needle, not an AI or imaging assistance tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's acceptance is based on demonstrating that its technological characteristics (materials, dimensions, intended use) are identical to a legally marketed predicate device, and that it complies with applicable industry standards and US regulations for performance and safety (e.g., sterilization, biocompatibility, mechanical integrity). This is primarily established through design specification comparisons and engineering testing, not clinical "outcomes data" or expert consensus on clinical cases.

    8. The sample size for the training set

    This information is not applicable. This device does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable. There is no training set for this device.