Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K083135

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-10-31

    (8 days)

    Product Code
    Regulation Number
    870.5900
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ChillBuster® Model 8002 has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition. Use is limited to whole-body warming in adult humans, free of skin conditions or other impairments where distributed heat application is deemed contraindicated by the responsible physician.

    Device Description

    This device, as now cleared, permits use with either a sterile or a non-sterile blanket cover, as necessary, upon the determination by a health care practitioner. Most intended use does not involve surgery. Also, even during surgery, only those interventions where the blanket cover would threaten a sterile operating field actually require a sterile blanket cover to achieve intended function. The present design modification involves no change to intended use, indications for use, technological basis, device design, or substantial equivalence. And both the ChiliBuster® Predicate and Modified devices require use of a blanket cover with every therapy session. For ThermoGear's ChillBuster® product, the only Predicate/Modified device difference lies in labeling that addresses when to use a sterile versus non-sterile blanket cover. While the Predicate 8002 requires use of a sterile blanket cover for every therapy, the Modified 8002 "requires the decision to use a sterile or a non-sterile blanket the woulded by a medical practitioner, familiar with the patient's condition and circumstance."

    AI/ML Overview

    This 510(k) summary (K083135) describes a modification to the ChillBuster® Model 8002 Portable Electric Blanket, specifically regarding the labeling for when to use sterile versus non-sterile blanket covers. It is not a study demonstrating device performance against specific acceptance criteria. Instead, it is a statement affirming that the modified device is substantially equivalent to its predicate device and that the change is a labeling update, not a change requiring new performance data.

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable as this submission explicitly states no performance data was required for this particular change.

    Here's why each point is not applicable based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Not Applicable. The document explicitly states: "Clearance of the subject of this submittal, requirements of when to use a sterile or non-sterile blanket cover, does not require performance data. The two cover types are the same in every way, save the state of sterility." Since no performance data was collected or required, there are no acceptance criteria or reported device performance for this specific modification.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not Applicable. No performance testing was conducted for this specific labeling modification, so there is no test set or associated data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not Applicable. As no performance testing was performed, there was no need for experts to establish ground truth for a test set. The change is related to the decision of a medical practitioner (implied expert in clinical practice) regarding sterility, not a diagnostic or therapeutic performance evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. No performance testing was performed, so no adjudication method was used for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This device is a portable electric blanket, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study or evaluation of human reader improvement with AI is irrelevant and was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This is a hardware device (electric blanket) and not an algorithm. Standalone algorithm performance is not relevant to this submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not Applicable. No performance testing was performed, so no ground truth was established for the purpose of evaluating the modified device's performance. The "ground truth" for the submission is the regulatory and clinical understanding that practitioners can determine when a sterile vs. non-sterile cover is needed.

    8. The sample size for the training set

    Not Applicable. This device uses a predicate device model; it doesn't involve machine learning or AI that would require a training set.

    9. How the ground truth for the training set was established

    Not Applicable. As there is no training set, there is no ground truth to establish for it.

    In summary, the provided document describes a regulatory submission for a labeling change to a medical device, not a study evaluating its performance against specific technical criteria. The core argument is that the change itself (allowing practitioner discretion for blanket cover sterility) does not impact safety or effectiveness and therefore does not require new performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071918

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-11-09

    (120 days)

    Product Code
    Regulation Number
    870.5900
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ChillBuster® Model 8002 has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition. Use is limited to whole-body warming in adult humans, free of skin conditions or other impairments where distributed heat application is deemed contraindicated by the responsible physician.

    Device Description

    The ChillBuster® Model 8002 Portable Electric Blanket System features a special electrical resistively heated Blanket, a Controller to manage power and device function, a rechargeable Battery, special cabling, and means to connect to external DC or AC power for recharging the Battery and, at the option of the user, operating the Blanket. During therapy, the Blanket is encased in a Sterile, Single-Use Blanket Cover to prevent cross-contamination of successive patients and to prevent soiling. The device can perform therapy at virtually any interval in the following activities or activity chains: a) initial patient preparation and transport in the field, and at any stage of ER, surgery, and ICU or post-op care for injury victims from the field; b) pre-op, surgery, ICU or post-op care of in-facility patients; and c) special situations involving potential or actual hypothermia (e.g. hemodialysis; peritoneal dialysis; plasmapheresis; patient warming during transport from point to point, etc.).

    AI/ML Overview

    This is a 510(k) summary for a portable electric blanket, not an AI device. As such, the requested information regarding AI device acceptance criteria, study details, ground truth, and expert involvement is not applicable or available in the provided text.

    Here's an overview of the non-clinical performance data provided, which serves as the acceptance criteria for this non-AI medical device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standards Met/Performance Goal)Reported Device Performance
    Compliance with Consensus Standards:
    IEC/EN 60601-1 (General Medical Safety)Device has been tested to currently valid versions of this standard.
    EN/JEC 60601-1-2 (Electromagnetic Compatibility)Device has been tested to currently valid versions of this standard.
    IEC 60601-2-35 (Heating Pads and Blankets)Device has been tested to currently valid versions of this standard. The safety cutoff temperature is 41 ± 1℃, which is the same as specified in IEC 60601-2-35 for maximum safety.
    ISO 14971 (Application of Risk Management)Device has been tested to currently valid versions of this standard.
    CAN/CSA C22.2 No. 601.1-M90 (Canadian General Medical Safety)Device has been tested to currently valid versions of this standard.
    Durability (Washing/Drying Cycles):Blanket subjected to 50 washing/drying cycles using household detergent and machines. No sign of wear or deterioration. Function demonstrated to be normal in five sessions of heating performance tests.
    Water Resistance/Functionality:Demonstrated to function normally even when completely submerged (except for the Blanket Connector).
    Heat Density Comparison:Analytical calculations showed higher heat density at the actively heated surface compared to the ChillBuster® predicate, but a small percentage of the Life-Air predicate.
    Temperature Safety Limit:Blanket limit is 41 ± 1℃. This is within the maximum safety cutoff temperature specified in International Standard IEC 60601-2-35 and lower than the Life-Air predicate's 43 ± 3℃.
    Operational Modes:Can operate the Blanket while connected to the AC line (a significant difference from the predicate).
    Power Output:Higher power output to the Blanket compared to the predicate (but still low compared to the Life-Air Predicate).
    Temperature Monitoring/Distribution:Improved Blanket temperature monitoring and assurance of even heat distribution (compared to the predicate).

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size:
      • Washing/Drying: 1 blanket (subjected to 50 cycles).
      • Heating Performance: 1 blanket (tested across 5 sessions).
      • Submersion: 1 blanket.
    • Data Provenance: Not explicitly stated, but the tests were conducted as non-clinical performance data studies, likely by the manufacturer. No country of origin for "data" in the context of patient data is relevant here, as this is a physical device testing. The studies are prospective in nature, as they involve actively testing the new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is a medical device (electric blanket) undergoing engineering and safety testing against established standards, not an AI algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" here is the pass/fail criteria of the tests and compliance with regulatory standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for human interpretation tasks in AI studies. For this device, performance is evaluated against predefined engineering standards and test protocols.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI device, and therefore, no MRMC study involving human readers and AI assistance was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm. The "standalone performance" refers to the device's functional and safety performance as described in point 1 above.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance is established by international and national consensus standards for medical device safety and performance (e.g., IEC/EN 60601-1, IEC 60601-2-35, ISO 14971, CAN/CSA C22.2 No. 601.1-M90) and the physical attributes and intended function of the device. For example, the maximum temperature cutoff of 41 ± 1℃ is a safety standard specified in IEC 60601-2-35.

    8. The sample size for the training set:

    • Not applicable. This is not an AI device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no training set for an AI algorithm, no ground truth needed to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K991684

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-03-01

    (289 days)

    Product Code
    Regulation Number
    870.5900
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ChillBuster® Model 8001 Portable Electric Blanket System has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition.

    Device Description

    The ChillBuster® Model 8001 Portable Electric Blanket System is made up of six major components: Blanket, Control Module, Battery, Cabling, AC Adapter, and Carrying Bag. The Blanket features a waterproof nylon outer layer, an inner layer of soft polar fleece, and a unique pattern of special thermal wire and a wire-adjacent thermistor. The Control Module contains an electronic circuit board, a User Interface Panel, Interconnect Wiring, and a Molded Chassis with a compartment for the Battery. The System uses a sealed lead acid, rechargeable Battery. Cabling includes a standard Blanket Interconnect Cable and an optional cable for connection to a vehicle cigarette lighter socket. All Systems include a standard AC-to-DC Adapter. The Carrying Bag conveys no impact on system safety.

    AI/ML Overview

    The provided text does not contain acceptance criteria or study details for a medical device in the context of the requested information (AI/algorithm performance, ground truth, expert review, MRMC studies, etc.).

    Instead, the document is a 510(k) summary for a "ChillBuster® Model 8001 Portable Electric Blanket System," which is a physical medical device, not a software algorithm or AI-based system.

    Therefore, I cannot extract the information required to populate the table or answer the specific questions about AI/algorithm performance. The non-clinical performance data section describes tests conducted on the physical electric blanket, focusing on safety standards (IEC, UL), general operation, warming performance, and comparison to a predicate device in terms of heat output and distribution.

    To directly address your request given the provided input:

    1. Table of acceptance criteria and reported device performance:
    The document does not explicitly state "acceptance criteria" for an algorithm. It describes safety standards and functional verification for a physical electric blanket, and "product specifications" were met.

    Acceptance Criteria (Not applicable for AI/Algorithm)Reported Device Performance (Physical device - ChillBuster Model 8001)
    NAMet IEC 601-1 (1998; Amd. 1: 1991; Amd. 2; 1995) Safety Standard
    NAMet UL 2601-1 Safety Standard
    NAMet IEC 601-1-2 EMC Standard
    NAGenerated distributed, low-level active heat throughout intended operating range
    NADemonstrated progressive lengthening of functional blanket heat delivery as temperature setting varied from highest to lowest
    NAMet all product specifications under simulated in-use test conditions
    NACompared favorably to predicate device (Life-Air 1000) in heat production and distribution (Model 8001 showed more even heat distribution while predicate had higher max output)
    NAInformal tests by >50 subjects reported rapid rise to moderate, comfortable temperature, then constant
    NANo adverse equipment or performance encountered in informal testing
    NAMaintained function after 10 machine wash/dry cycles without degradation of temperature rise time or heat delivery

    2. Sample size used for the test set and the data provenance: Not applicable for an AI/Algorithm test set. For the physical device, informal tests involved ">50 individuals" (independent subjects). The formal tests were conducted by Northwest EMC, Inc., an independent commercial firm. The data provenance is primarily from independent testing by Northwest EMC, Inc. and informal testing by ThermoGear™ in the US (Tigard, OR).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/algorithm is not discussed. For the physical device, compliance with standards and functional performance was assessed by an independent testing firm and informal user feedback.

    4. Adjudication method for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a physical electric blanket, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used: Not applicable for an AI/algorithm. For the physical device, "ground truth" was based on adherence to recognized safety and EMC standards (IEC 601-1, UL 2601-1, IEC 601-1-2) and the observed functional characteristics during testing (e.g., heat generation, battery life, washing performance).

    8. The sample size for the training set: Not applicable.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1