    Intended Use

    The X-FLO Fluid Management System is intended to:

    · Warm and pump fluid, via intravenous (IV) bags or fluid bottles, to provide fluid distention and flushing of the uterus for gynecology procedures. The device can also measure fluid deficit of the surgical procedure.

    · Warm and pump fluid, via IV bags or fluid bottles, to provide fluid distention and flushing of the bladder and kidneys, or general distention and flushing, for urological procedures. The device can also measure fluid deficit of the surgical procedure.

    · Warm and pump fluid, via IV bags or fluid bottles, to provide fluid distention and flushing of the shoulder, knee, or other small joints for orthopedic procedures. The device can also measure fluid deficit of the surgical procedure.

    Device Description

    The X-FLO Fluid Management System ("X-FLO") (P/N 01910), with an optional deficit module (P/N 01601) is intended for fluid distention, fluid warming, and fluid volume/deficit measurements, as well as external suction regulation pertaining to deficit measurements for diagnostic and/or operative endoscopic procedures within gynecology, urology, and orthopedic disciplines.

    The X-FLO utilizes a touchscreen user interface and performs fluid pressurization, warming, and volume/deficit monitoring functions. More specifically, the device employs a peristaltic pump to pressurize and deliver fluid to the surgical site for distending and continually flushing the surgical site for visualization purposes, utilizes infrared lamps to optionally warm the fluid to body temperature, and uses fluid bag weight to monitor fluid inflow.

    Regarding the optional deficit monitoring function, the device utilizes external suction to return fluid from the surgical site. The X-FLO monitors and adjusts the vacuum level to the operator selected setpoint by opening the suction line to atmosphere if necessary. Once the fluid is returned from the surgical site, the single-use deficit cartridge measures the fluid volume collected, prior to transfer to fluid collection equipment and/or the hospital's waste disposal system.

    The X-FLO is controlled via a Graphical User Interface (GUI), wherein the user is guided to select a surgical discipline and a procedure type. Once selected, the user can input the desired setpoint within the established minimum and maximum parameters for that procedure type. As set forth below, the X-FLO can operate in Pressure Control (the default control, or X-Control. Each control view has defaults and adjustment ranges for fluid pressure, flow, temperature, deficit alarm, and external suction regulation.

    In addition to adjusting the fluid pressure setpoint via the GUI, the user may optionally utilize a foot pedal included with the X-FLO to temporarily increase the fluid pressure. By pressing the foot pedal, the user initiates an increase in fluid pressure. The increase is user configurable but can never exceed the maximum allowable fluid pressure for the procedure. When the user desires to have the fluid pressure return to the setpoint, the foot pedal is released.

    AI/ML Overview

    The provided text details the 510(k) submission for the Thermedx X-FLO Fluid Management System, which is a medical device intended for fluid distention, warming, and volume/deficit measurements during various endoscopic procedures.

    Here's an analysis of the acceptance criteria and study data based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document presents a comparison table between the subject device (X-FLO Fluid Management System) and its predicate device (FluidSmart), which implicitly defines the performance criteria. The relevant criteria mainly relate to fluid warming, fluid delivery (pressure/flow), and deficit monitoring.

    Table of Acceptance Criteria and Reported Device Performance:

    Feature/CriterionAcceptance Criteria (from comparison with predicate/stated capability)Reported Device Performance (X-FLO Subject Device K210628)
    I. Imaging/AI Specifics(Not Applicable - This is not an AI/imaging device)(Not Applicable - This is not an AI/imaging device)
    II. Device Performance
    Fluid Warming
    Heating TechnologyInfrared lampsInfrared lamps
    Fluid Warming RateIncrease inlet fluid temp to 40°C.Capable of increasing inlet fluid temperature from 18 °C ± 1 °C to 40 °C ± 3 °C for flow rates < 800 mL/min; Capable of heating 18 ±1 °C inlet fluid to a 15 ±3 ℃ temperature rise at flow rates of >800 mL/min.
    Maximum Setpoint40°C40 °C
    Temperature MeasurementMeasured at exit from cartridge component of tubing setIrrigation fluid temperature measured at exit from the cartridge component of the tubing set
    Temperature DisplayedDisplayed at GUIIrrigation fluid temperature displayed at GUI.
    Over-temperature SafetyDisable pumping if fluid temp exceeds set-point + 3°CThe device will disable the lamps if the fluid temperature reaches 41.5 ± 1 °C. The device will disable pumping if the actual fluid temperature reaches 46 ± 1 °C.
    Fluid Delivery
    Pumping MechanismPeristaltic pumpPeristaltic pump with ability to reverse to relieve pressure
    Pressure Measurement(Implied accurate measurement)Dual transducers measuring fluid pressure
    Maximum Flow Rate1200 mL/min (stated for predicate)1200 mL/min
    Fluid Pressure30-300 mmHg (predicate)30-350 mmHg
    Max Fluid Vessel Size(2) 5-Liter (predicate)(4) 5-Liter (Note: The document states this "does not raise different questions of safety and effectiveness")
    Fluid Weight Accuracy± 10% or 250 mL (whichever is greater) (predicate)± 5 mL (Note: This is significantly more precise than the predicate's reported accuracy)
    Temporary Pressure IncreaseUser interface (predicate)Foot pedal or User interface
    Over-pressure SafetyAudible/visual notifications, suspend fluid flow if outside ±10% or 12mmHgAudible and visual notifications and disable pumping if the fluid pressure exceeds the setpoint by the greater of 10% or 12 mmHg. Disable pumping prior to irrigation pressure exceeding the maximum set point by more than the higher of 10% or 12 mmHg.
    Deficit Monitoring
    Fluid Deficit MonitoringYesYes
    Monitoring Accuracy SpecThe greater of 250ml or 10% of the volume pumped (predicate)1 – 1000 mL: ± 75 mL; 1000 – 2500 mL: ± 5 %; 2500 – 5000 mL: ± 3 %; ≥ 5000 mL: ± 2.2 % (Note: This is a different and more precise specification than the predicate's reported accuracy)
    Measurement MeansMeasurement by weight (predicate)Measurement by flow (Note: This is a different method than the predicate, justified by performance testing)
    Fluid Deficit DisplayDisplayed on GUI (predicate implied)Displayed on GUI
    External Suction RegulationNone (predicate)0-400 mmHg Suction Setpoint Range (Note: New feature, justified as not raising new questions of safety/effectiveness as it can be evaluated by performance testing)
    Over-deficit Safety (Isotonic)Audible/visible notification, suspend fluid flow if deficit reachedThe device will disable pumping when the default setpoint or maximum deficit volume (2,500 mL) is reached
    Over-deficit Safety (Hypotonic)(Not applicable for predicate)The device will disable pumping and provide audible notification when the deficit level reaches the deficit alarm setpoint inputted by the user. The device will disable pumping and provide audible notification when the maximum allowable deficit level of 2,500 mL is reached if the fluids used are a combination of isotonic and hypotonic. The device will disable pumping when the maximum deficit of 1,000 mL is reached. (Note: New feature due to the new deficit measurement method)
    III. Other Performance
    BiocompatibilityCompliant with ISO 10993-1, 5, 10, 11Conducted per 2020 FDA guidance. Tests include: Cytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), Irritation (ISO 10993-10:2010), Acute Systemic Toxicity (ISO 10993-11:2017)
    Electrical Safety & EMCCompliant with IEC 60601-1, 1-2, 1-6Complies with IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012), IEC 60601-1-2:2014, and IEC 60601-6:2010 + A1:2013.
    Software V&V TestingCompliant with FDA guidance for "major" level of concernConducted as recommended by 2005 FDA guidance document. Software considered "major" level of concern.
    Sterility & Shelf-LifeSterile, compliant with ISO 11135, ASTM F198/F1929/F88Sterilized via Ethylene Oxide per ISO 11135:2014. Shelf life of 6 months. Shelf-life testing per accelerated aging (ASTM F198-16) with visual inspection (ASTM F1929-15) and seal strength (ASTM F88).
    Bench TestingSpecifications met in simulated use for various parametersEvaluated alarm sound levels, tubing/cartridge leak testing, tubing mechanical strength, fluid sensing, fluid compatibility (sterile water, lactated ringers, mannitol, glycine, saline, sorbitol), irrigation pressure controls (with three hysteroscope models), over-temperature, over-pressure, over-deficit testing, tissue trap capacity, deficit calculation accuracy, empty fluid bag detection, temperature accuracy, and pressure and flowrate accuracy.

    Study Proving Device Meets Acceptance Criteria:

    The document describes non-clinical performance testing conducted to support the substantial equivalence determination.

    1. Sample Size Used for Test Set and Data Provenance:

      • The document implies that testing was performed on representative units of the device and its accessories (tubing sets, deficit cartridge, etc.).
      • Sample Size: Not explicitly stated as a number of devices or data points for each bench test. For example, "Fluid compatibility with sterile water, lactated ringers solution, 5 % mannitol solution, 1.5 % glycine solution, 0.9% saline solution, and sorbitol solution" implies multiple tests, but the quantity of tests, runs, or devices tested for each parameter is not specified. Similarly, "Irrigation pressure controls with three different hysteroscope models" specifies the models but not the number of runs or devices.
      • Data Provenance: The document does not specify the country of origin of the data. It is a 510(k) submission to the US FDA, so the testing was presumably conducted in a manner acceptable for US regulatory submission. The studies are bench/laboratory studies focused on device performance, not patient data studies, so "retrospective or prospective" is not an applicable distinction in the sense of patient data collection.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This is a technical performance study of a medical device, not a study involving human interpretation of clinical data (like AI for imaging). Therefore, the concept of "experts establishing ground truth" in the clinical sense (e.g., radiologists) is not applicable. The "ground truth" for these tests is the physical measurement of device performance against engineering specifications using calibrated equipment and standard test methods.
    3. Adjudication Method for the Test Set:

      • Not Applicable. As this involves objective bench testing, there is no "adjudication" in the sense of reconciling subjective expert opinions. Device performance measurements are compared directly against pre-defined engineering and regulatory specifications.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No. This is not an AI/imaging device or a device whose performance is dependent on human reader interpretation. Therefore, an MRMC study is not relevant.
    5. If a Standalone (algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. While the device has software ("Software controlled"), its performance is measured as an integrated system (hardware and software) in bench testing, not as a standalone algorithm evaluating clinical data. The "stand-alone" performance is implicitly the results of the bench testing.
    6. The Type of Ground Truth Used:

      • The "ground truth" for the performance studies is the pre-defined engineering specifications and internationally recognized standards (e.g., IEC 60601 series for electrical safety, ISO 10993 for biocompatibility) that the device must meet. Performance metrics (e.g., temperature, pressure, flow rate, deficit volume) are objectively measured using calibrated instruments during bench testing.
    7. The Sample Size for the Training Set:

      • Not Applicable. This is a hardware medical device with embedded software; it does not involve machine learning or AI models that require "training sets" in the conventional sense of data-driven model training. The software is developed and verified via traditional software engineering principles.
    8. How the Ground Truth for the Training Set Was Established:

      • Not Applicable. (As explained above, no traditional training set exists for this type of device.) The "ground truth" for software functionality testing (verification and validation) would be the software requirements specifications derived from the device's intended use and design.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172048

    Validate with FDA (Live)

    Device Name
    FluidSmart
    Manufacturer
    Date Cleared
    2017-09-05

    (61 days)

    Product Code
    Regulation Number
    884.1700
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, and orthopedic disciplines.

    Device Description

    FluidSmart® is intended for irrigation, distention, fluid warming, and fluid volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines. Fluids commonly used include 0.9% saline, lactated ringer's solution, 5% mannitol, 1.5% glycine, and sterile water.

    The main unit is mounted on a rolling stand which houses the technology used to warm the fluids and the peristaltic pump used to pressurize the fluids. A touch-screen operated processor provides closed loop controls for fluid warming and pumping. The processor monitors temperature and pressure sensors to adjust warming power and pump speed as required. Load cells attached to the fluid bag hooks and suction canister ring allow the processor to monitor fluid levels pumped to and suctioned from the surgical site.

    AI/ML Overview

    This document is a 510(k) summary for the FluidSmart device, which is a hysteroscopic insufflator. Such summaries typically focus on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and study results in the manner requested for AI/ML devices. Therefore, much of the requested information, which is common in regulatory submissions for AI/ML or novel devices, is not present here.

    Here's a breakdown of the available information and what is not present:

    1. Table of acceptance criteria and the reported device performance

    The document provides a comparative table of technological characteristics to predicate devices, but it does not specify quantitative acceptance criteria for performance benchmarks, nor does it present device performance against such criteria directly in a table format as might be expected for software performance metrics.

    Technological CharacteristicAcceptance Criteria (Implicit from Predicate/Study Purpose)FluidSmart® Reported Performance
    Fluid Warming MethodSame as Predicate (Infrared heating of fluid)Infrared heating of fluid
    Pressure Generation MethodSame as Predicate (Peristaltic pump)Peristaltic pump
    User InterfaceSame as Predicate (Touch Screen)Touch Screen
    Device TypeSame as Predicate (Electro-Mechanical)Electro-Mechanical
    Maximum Pressure (ureteroscopy procedures)300 mmHg (matching secondary predicate at 300-500 mmHg)300 mmHg

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    Not applicable. This device is a hardware device (insufflator, fluid warmer) with associated software for control. The testing involved non-clinical design verification, not analysis of a "test set" of patient data in the context of an AI/ML device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as this is not an AI/ML device requiring expert ground truth for a test set.

    4. Adjudication method for the test set

    Not applicable, as this is not an AI/ML device requiring expert adjudication of a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI/ML device and thus an MRMC study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML algorithm. The device itself (FluidSmart®) is a standalone system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as this is not an AI/ML device. The "ground truth" for a device like this would involve engineering specifications, safety standards, and functional performance benchmarks (e.g., temperature accuracy, pressure regulation accuracy, leak tests), which are evaluated through non-clinical testing.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device with a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device with a training set.


    Summary of the Study Mentioned (Non-Clinical Testing):

    The document states:

    • Non-clinical testing performed including the following design verification testing:
      1. Software verification testing.
      2. Tubing set verification testing.

    This indicates that Thermedx conducted internal testing to ensure the software functions as designed and the disposable tubing sets meet their specifications (e.g., integrity, pressure relief valve operation, biocompatibility, and expiration dating, as mentioned elsewhere). No clinical testing was conducted or submitted. The "study" here is essentially internal design verification and validation against engineering specifications and regulatory standards for safety and effectiveness for a medical device.

    Key takeaway: This 510(k) submission primarily focuses on demonstrating substantial equivalence to existing predicate devices by comparing technological characteristics and asserting that the minor proposed changes (specifically, increasing maximum pressure for ureteroscopy procedures from 200 mmHg to 300 mmHg) do not raise new questions of safety or effectiveness. It is not an AI/ML device, and therefore, the typical AI/ML study parameters are not applicable or provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102275

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-03-25

    (226 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Thermedx 37~5 electrocautery probes are indicated for electrosurgical cutting and coagulation during a variety of urological, gynecological, laparoscopic and general surgical procedures.

    Device Description

    The Thermedx 37~5 electrode probes are electrosurgical devices which enable the user to have the functions of cutting and coagulation along with suction and irrigation in a single hand held device. These probes are intended to be used in conjunction with the Trumpet valve disposable assembly (model number TV0002), The probes come in a variety of tip configurations. Monopolar electrodes include a J Hook. L Hook. Spatula and Needle tip. . All electrodes have working dimensions of 33 cm in length and 5 mm diameter. Electrodes are manufactured of biocompatible materials per ISO 10993 and are safe, effective and durable for their intended use. All electrodes have an ABS coupling to connect to the trumpet valve disposable assembly and are connected to an electrosurgical unit with a cable of PVC insulation and copper wires.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    This submission does not provide specific quantitative acceptance criteria or detailed device performance metrics in the format of a table. The document focuses on demonstrating substantial equivalence to predicate devices. The "performance" assessment is qualitative, asserting that the new device has "similar technological characteristics" and "minor differences are not significant with respect to the safety or effectiveness."

    The key performance claims are:

    • Electrosurgical cutting and coagulation capability.
    • Suction and irrigation capabilities (for the integrated hand-held device).
    • Compliance with electrical safety (IEC 601-2-2) and biocompatibility (ISO 10993-1) standards.

    Since no specific numerical acceptance criteria or test results are provided in the summary, a table cannot be constructed with performance values. The primary acceptance criterion for this 510(k) submission was the demonstration of substantial equivalence to predicate devices, which was achieved through:

    • Same intended use.
    • Similar technological characteristics.
    • Compliance with relevant safety standards (IEC 601-2-2 and ISO 10993-1).

    2. Sample Size Used for the Test Set and Data Provenance

    No specific test set or data provenance (e.g., country of origin, retrospective/prospective) is described for demonstrating the performance of the device. The submission explicitly states: "No animal or clinical test results were required to demonstrate substantial equivalence." This indicates that no human or animal test data was used for this 510(k) submission from the perspective of a "test set." The evaluation relied on technical comparisons and compliance with recognized standards.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Since no animal or clinical test results were required for substantial equivalence, there was no "ground truth" to be established by experts in the context of clinical or performance data for this submission. The ground truth for regulatory acceptance was based on a comparison to predicate devices and compliance with established standards.

    4. Adjudication Method for the Test Set

    Not applicable. As no test set data was presented or required for this 510(k) submission, there was no adjudication method used for such data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No animal or clinical test results were required to demonstrate substantial equivalence." This type of study would involve human readers and clinical data, which was not part of this submission.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is an electrosurgical probe, a physical medical instrument, not an algorithm or AI software. Therefore, there is no concept of "standalone (algorithm only)" performance in this context.

    7. Type of Ground Truth Used

    The "ground truth" for this 510(k) submission was based on:

    • Predicate Device Equivalence: The primary ground truth was the established safety and effectiveness of the identified predicate devices (Modulap Disposable Electrosurgical Probes, Genicon Re-Usable Monopolar Probes, Stryker Stiletto Electrosurgical Probe, Conmed Electrosurgical Probe). The new device was deemed substantially equivalent to these devices.
    • Compliance with Standards: Compliance with recognized international standards for electrical safety (IEC 601-2-2) and biocompatibility (ISO 10993-1) served as a "ground truth" for basic safety and performance characteristics.

    No clinical expert consensus, pathology, or outcomes data from the new device was used to establish ground truth for this submission, as such data was not required.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm requiring a training set. The "training" for the device's design and manufacture would involve engineering principles, materials science, and adherence to quality system regulations, rather than data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI/ML algorithm and thus no "training set," the concept of establishing ground truth for a training set is not relevant to this device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091939

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-23

    (388 days)

    Product Code
    Regulation Number
    884.1700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 37-5 is a multi-purpose fluid management system indicated for the following:

    • Laparoscopy: warming fluids for irrigation and aspiration .
    • Hysteroscopy: uterine distention during diagnostic and operative procedures .
    • Cystoscopy: distention of the bladder during diagnostic and operative procedures (with optional . fluid warming)
    • Arthroscopy: distention and irrigation of the knee, shoulder, and small joints, during diagnostic . and operative procedures

    During hysteroscopic, cystoscopic and arthroscopic procedures, the 37~5 can be used to monitor fluid deficit.

    Device Description

    The Thermedx 37-5 is a general purpose fluid management system intended for standard surgical fluid management applications. The device consists of a Base Unit and Disposable Tubing Sets. A rolling standmounted device houses infrared fluid heating technology and a peristaltic fluid pump with closed-loop controls. The unit operates in 2 primary modes with fluid heating; flow control and pressure control. The Disposable Tubing Sets include IV bag spike, heating cartridge, and trumpet valve or luck connectors for endoscopes or other devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Thermedx 37-5 Fluid Management System, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not provide a table with specific numerical acceptance criteria (e.g., target temperature ranges, flow rates with specific tolerances, pressure ranges with specific accuracies) or detailed numerical performance results for the Thermedx 37-5. Instead, it makes a general statement about meeting specifications.

    Acceptance CriterionReported Device Performance
    Simulated Laparoscopy Performance: Ensure safe and effective fluid management during simulated laparoscopic procedures."demonstrate the device meets all performance specifications"
    Simulated Uterine Distension Performance: Ensure safe and effective uterine distension during simulated hysteroscopic procedures."demonstrate the device meets all performance specifications"
    Over-temperature Protection: Mitigate risks associated with excessively heated fluids."Over-temperature...testing. These tests demonstrate the device meets all performance specifications"
    Over-pressure Protection: Mitigate risks associated with excessive fluid pressure."Over-pressure...testing. These tests demonstrate the device meets all performance specifications"
    Over-Flow Protection: Mitigate risks associated with excessive fluid flow rates."Over-Flow...testing. These tests demonstrate the device meets all performance specifications"
    Over-Deficit Protection: Mitigate risks associated with excessive fluid deficit measurement errors."Over-Deficit testing. These tests demonstrate the device meets all performance specifications"
    Electrical Safety (IEC 60601-1): Conform to international standards for electrical safety in medical devices."Electrical Safety Testing in accordance to IEC 60601-1 electrical safety standards. These tests demonstrate the device meets all performance specifications"

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document mentions "non-clinical performance testing" which typically involves a series of tests, but not a "sample size" in the statistical sense of a patient cohort. The tests are simulated, indicating no human or animal subjects were used.
    • Data Provenance: Not applicable, as the testing was "non-clinical" and "simulated." This would imply in-house laboratory testing rather than data from a country of origin or retrospective/prospective studies on patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts being used to establish a "ground truth" in the context of this 510(k) summary. The testing conducted was non-clinical and simulated, implying engineering and technical validation against pre-defined specifications rather than expert clinical judgment.

    4. Adjudication Method for the Test Set

    Not applicable. As no expert ground truth or clinical data requiring adjudication is mentioned, no adjudication method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is not relevant for this type of medical device (fluid management system) which does not involve image interpretation or diagnostic performance that would require human readers. The document details performance testing of the device's physical functions.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This question is not directly applicable in its typical sense for an AI/algorithm-based diagnostic device. However, the "non-clinical performance testing" described (simulated laparoscopy, uterine distension, over-temperature, etc.) is standalone device performance testing, as it assesses the device's functions without direct human intervention in its operation for the purpose of the test. The device itself is designed to be used with human operators in a clinical setting.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the defined performance specifications and safety standards. For instance:

    • Temperature accuracy (e.g., maintaining fluid at approximately body temperature).
    • Pressure limits (e.g., maintaining pressure within safe ranges).
    • Flow rates (e.g., meeting specified flow capabilities).
    • Electrical safety standards (IEC 60601-1).

    The device was tested against these pre-defined engineering and safety specifications.

    8. The Sample Size for the Training Set

    Not applicable. This device is a mechanical/electronic fluid management system, not an AI/machine learning algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.