Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K232235

    Validate with FDA (Live)

    Date Cleared
    2023-10-25

    (90 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

    Device Description

    The Horizon® 3.0 TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation. ("TMS").

    The Horizon® 3.0 TMS Therapy System is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

    The Horizon® 3.0 TMS Therapy System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient settings, including physicians' offices, clinics, and hospitals.

    Horizon® 3.0 TMS Therapy System is an integrated system consisting of a combination of hardware, software, and accessories.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the Magstim®Horizon® 3.0 TMS Therapy System. It describes the device, its intended use, and the non-clinical testing performed to demonstrate its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Key Takeaway: The device discussed is the Magstim®Horizon® 3.0 TMS Therapy System, and the submission is a 510(k) to demonstrate substantial equivalence to a previously cleared version (K222171). The testing described is primarily non-clinical bench testing focused on changes made to the device's camera tracking system and EMG amplifier, ensuring these changes do not raise new questions of safety or effectiveness and that the performance is equivalent or better than the predicate device. No clinical studies or human reader studies (MRMC) were conducted or mentioned as a requirement for this particular submission.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on demonstrating performance equivalent to or better than the predicate device, especially for the modified components (camera tracking and EMG). The performance metrics are largely quantitative measurements from bench testing.

    Acceptance Criteria (Expected Result)Reported Device Performance (Actual Results)
    Stress Testing - Sample Rate: Tracking remains stable under load to support sampling rates of 20Hz and 50Hz. (These rates are the highest for FDA cleared protocols - OCD Protocol and iTBS Protocol).Actual Results: Maintains a constant sample rate above 50Hz. 50Hz chosen as that is the fastest rate of delivery out of cleared protocols. Under Abnormal PC Load: 52Hz min recorded. Average Capture Rate in Normal Conditions: 120Hz. Equivalent to Predicate Device.
    Stress Testing - Data Storage: 1000 rMT positions can be stored and 10,000 stimuli positions can be stored.Actual Results: Supports recording of 1000 Simultaneous MT Locations for Motor Hotspot procedures and recording of up to 12,000 Stimuli which is sufficient for treatment (Maximum stimuli from FDA protocol is 3000) – Equivalent to Predicate Device.
    Positional and Volumetric Accuracy - Camera Accuracy: <= 0.25mm Root Mean Squared (RMS) Positional Accuracy.Actual Results: 0.16mm RMS Positional Accuracy – Equivalent to Predicate Device.
    Landmark Registration Accuracy (Rigid Body Testing): <= 1mm RMS Rigid-body Positional Accuracy.Actual Results: 0.45mm RMS Rigid-body Positional Accuracy - Equivalent to Predicate Device.
    Orientational Accuracy: <= 5.5 degrees RMS Orientational Accuracy.Actual Results: 1.8 degrees RMS Orientational Accuracy. Equivalent to Predicate Device, provides an equivalent level of rotational accuracy.
    Treatment Delivery Performance on Phantom Head: Pickup coil should detect a reading of the same voltage induced at the MT hotspot and Treatment Target (+/- 5%).Actual Results: Over 20 samples (repeat of 10 tests for MDD and 10 tests for OCD) there was an average difference of 1.02% for MDD and 1.92% for OCD of the induced voltage when measured at the expected treatment target following the principles of operation. Resulting in a combined average difference of 1.47% from the voltage induced in a search coil at the MT location compared to when it is placed at the treatment location following the Navigation software. The treatment location was determined to be the location presented by Navigation software and when the software indicated it was in the position (green). Equivalent to Predicate Device. When following the same principles of operation as the predicate device – the new tracking system results in equivalent output.
    Direct Accuracy Comparison to Manual Method and Predicate Device - Overall System Accuracy: The mean difference between the manual method and the subject device navigation method is equivalent or better than the predicate device navigation method compared to the same manual method (< 4.0mm).Actual Results: MDD Mean Difference to Manual Method (5.5cm forward from MT Hotspot): 1.51mm (Std. Dev 1.14mm)OCD Mean Difference to Manual Method (4cm forward from MT Hotspot): 1.66mm (Std. Dev 0.76mm)Compared to the results of the predicate device:Predicate – 3.0mm Mean Delta (Std. Dev 1.8mm)Subject – 1.5mm (Std. Dev 1.1mm)Equivalent to Predicate Device – subject device offers better approximation/guidance to the treatment target in accordance with the principles of operation.
    EMG Performance - Timing Jitter: < 5ms Jitter (TMS Pulse to EMG Capture).Actual Results: Timing Jitter Measured: Maximum of 3.25ms and mean of 1.68ms over samples. Equivalent to Predicate Device. All measured values of the system are within tolerance.
    EMG Performance - Noise Level: < 5uV RMS.Actual Results: Noise Level Measured: 3.03uV RMS. Equivalent to Predicate Device. All measured values of the system are within tolerance.
    EMG Performance - Waveform Accuracy: error of < 10% in Amplitude and Time.Actual Results: Waveform Accuracy Measured: Maximum Amplitude Error - 6.5%, Maximum Pulse Width Error - 3%, Maximum Frequency Error - 3.16%. Equivalent to Predicate Device. All measured values of the system are within tolerance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Positional and Volumetric Accuracy - Camera Accuracy: "A reflective marker is stepped through 1000 different positions (ground truths) with 30 samples taken at each position."
      • Treatment Delivery Performance on Phantom Head: "Over 20 samples (repeat of 10 tests for MDD and 10 tests for OCD)."
      • Other tests like "Stress Testing," "Landmark Registration Accuracy," "Orientational Accuracy," "Direct Accuracy Comparison," and "EMG Performance" indicate "samples" but do not specify a precise number for each measurement run, though the methods imply multiple measurements were taken (e.g., "moved along rotational axes by known amounts," "known waveforms are input," which suggests a range of conditions).
    • Data Provenance: The tests are described as non-clinical testing using simulated settings and phantom heads. The data is thus synthetic/experimental, not from human patients. The company is based in the United Kingdom.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The ground truth for the non-clinical bench tests appears to be established by the physical and measurable setup of the experiments themselves (e.g., "known positions," "known waveforms," "precisely positioned and secured pickup coil"). Expert consensus is not relevant for this type of engineering performance testing.


    4. Adjudication Method for the Test Set

    This is not applicable as the tests are non-clinical, quantitative bench tests with direct measurements against pre-defined engineering specifications. There is no human interpretative element requiring adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or required. The submission explicitly states: "There is no clinical testing required to support this submission." The device modifications are to the tracking camera, EMG, and associated software, and the testing focuses on ensuring their fundamental performance characteristics are equivalent or better than the predicate. This is not an AI diagnostic algorithm that assists human readers in interpreting medical images.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The "standalone" performance closest to what might be inferred is the performance of the camera tracking system and EMG amplifier in isolation, as tested in the bench assessments. The results detailed in the table (e.g., positional accuracy, orientational accuracy, EMG jitter, noise, waveform accuracy) are essentially "algorithm only" or component-level performance metrics, demonstrating the technical capabilities of the new components. However, this is not a diagnostic AI algorithm in the traditional sense that "stands alone" to provide a clinical output.


    7. The type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on engineering measurements and pre-defined physical parameters.

    • For positional accuracy tests: reflective markers are stepped through "1000 different positions (ground truths)".
    • For treatment delivery performance: "known position on the phantom head" and "precisely positioned and secured pickup coil at the ideal location."
    • For EMG performance: "known waveforms are input into the EMG interface using a function generator."

    8. The Sample Size for the Training Set

    This information is not applicable / not provided. This submission is for a medical device (TMS system) with updated hardware and software components, not for a deep learning or AI model trained on a large dataset of patient data. The "software" updates mentioned are related to integrating the new hardware and improving user interaction, not for training a predictive model.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable / not provided for the same reasons as #8. No "training set" in the context of an AI/ML model for clinical tasks is mentioned or implied.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130403

    Validate with FDA (Live)

    Date Cleared
    2013-08-23

    (185 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magstim D702 coil is a stimulating coil intended for use with the Magstim® 2002 Stimulating Unit or the Magstim® Rapid2 Stimulating Unit for the purpose of peripheral nerve stimulation for diagnostic purposes. The coil is an accessory to the Magstim ® 2002 Stimulating Unit, and the coil is an accessory to the Magstim® Rapid2 Stimulating Unit.

    Device Description

    The Magstim D702 coil is designed for use with the Magstim 2002 stimulator or the Magstim Rapid2 stimulator. The Magstim D70 coil provides a method of non-invasively delivering its output to a target such as the peripheral nerves. Discharge current through the coil generates a brief, high intensity magnetic pulse. When the coil is applied to a conductive medium such as the human body, eddy currents are produced in the medium by electromagnetic induction. This stimulation method enables deep and otherwise inaccessible nerves to be stimulated easily and relatively painlessly. In addition, no skin preparation is required and stimulation can be achieved through clothing. The coil is a non-sterile reusable device intended for multipatient use. The coil is transportable and is intended to be used in a laboratory, consulting room or equivalent environment. The coil is intended for short-term patient contact where single or cumulative contact does not exceed 24 hours.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Magstim D702 Coil, a medical device for peripheral nerve stimulation. It details how the device is considered substantially equivalent to predicate devices, primarily through engineering and performance testing, rather than studies involving human readers or extensive clinical trials.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (Magstim D702 Coil)
    Intended UsePeripheral nerve stimulation for diagnostic purposesPeripheral nerve stimulation for diagnostic purposes
    Indications for UsePeripheral nerve stimulation for diagnostic purposesPeripheral nerve stimulation for diagnostic purposes
    Average Inductance15.50 µH13 µH
    Number of Turns911
    Number of Windings22
    Wire Cross Sectional Area9 mm²12.5 mm²
    Coil Cable Length2.0m1.8m
    Manner of StimulationPeripheralPeripheral
    Peak Magnetic Field (Magstim 2002)1.7T (or 1.2T for Rapid2)2.1T
    Peak Magnetic Field (Magstim Rapid2)1.2T1.2T
    Number of pulses before exceeding max operating temp (Magstim 2002)110>290
    Number of pulses before exceeding max operating temp (Magstim Rapid2)>600 (Rapid1) / 110>600 (Rapid1) - Note: Discrepancy in table for Rapid2 vs Rapid1
    CoolingNoNo
    Multiple UseYesYes
    SterileNoNo
    Weight1.8kg1.95kg
    Integral AdapterNoYes

    Interpretation of "Acceptance Criteria": In this context, the "acceptance criteria" are not explicitly stated as numerical thresholds for success (e.g., "must achieve X% accuracy"). Instead, the acceptance criteria are implicitly defined by demonstrating that the modified device (Magstim D702 Coil) is substantially equivalent to the predicate devices. This means that:

    • Its intended use and indications for use are the same.
    • Its technological characteristics are similar, and any differences do not raise new questions of safety or effectiveness.
    • Testing shows performance is comparable or improved and meets applicable safety standards.

    The specific performance metrics for inductance, peak magnetic field, and number of pulses before overheating are compared directly to the predicate device's characteristics, and the device is deemed acceptable if these differences are understood and do not negatively impact safety or effectiveness, or in some cases, show improvement (e.g., higher pulse count due to better heat dissipation).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe "test sets" in the context of patient data or clinical imaging. The "testing activities" mentioned involved:

    • Safety testing from risk analysis.
    • Laboratory studies for electrical safety testing.
    • EMC testing.
    • Usability testing for consumer accuracy (this might involve human interaction but isn't elaborated as a clinical study).
    • Magnetic field measurement testing.

    Therefore, there is no stated sample size for a patient test set or data provenance because the evaluation was primarily engineering and performance-based, comparing device specifications to a predicate. It was not a study involving patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not Applicable. The evaluation was engineering-focused. There was no "ground truth" derived from expert consensus on medical images or clinical outcomes from a patient test set.

    4. Adjudication Method for the Test Set

    Not Applicable. As there was no patient test set requiring expert interpretation or consensus, there was no adjudication method used.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No. The document does not describe an MRMC study. The comparison is between the new device and predicate devices based on engineering specifications and laboratory testing, not human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This device is a hardware component (a stimulating coil) for a medical stimulator, not an algorithm. Therefore, "standalone algorithm performance" is not relevant. The device's 'performance' is its ability to generate magnetic fields and deliver pulses according to its specifications.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the established performance characteristics and safety profiles of the legally marketed predicate devices, along with relevant international standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-1). The new device's performance was measured against these benchmarks through laboratory testing.

    8. The Sample Size for the Training Set

    Not Applicable. As this is a hardware device and not an AI/ML algorithm that requires training data, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083242

    Validate with FDA (Live)

    Date Cleared
    2009-07-29

    (268 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurosign Avalanche nerve monitor is intended to help surgeons locate, identify, and preserve cranial motor nerves during surgery.

    The intended uses of the Neurosign Avalanche Motor Nerve Monitor is:

    2 channel thyroidectomy and parathyroidectomy, mastoidectomy

    parotidectomy, mastoidectomy, skull-base 4 channel

    Device Description

    The Neurosign Avalanche Motor Nerve Monitor is a 2 or 4 channel nerve monitor for intraoperative use in general surgery, skull-base surgery and ENT. The Neurosign Avalanche Motor Nerve Monitor allows the user to detect motor nerves and to check their integrity by acquiring the compound muscle action potentials at the target muscles. The acquired signals are processed by the monitor and displayed as waveforms on the monitor screen and played back acoustically via the built in audio amplifier and loudspeaker.

    The Neurosign Avalanche Motor Nerve Monitor provides additional information to the user, helping the surgeon to locate nerves by using an electrical stimulator. The sound is independent from the waveform display on the monitor's screen.

    The displayed waveforms can be printed along with the patient information via the built in thermal array printer, or as a generated report in Word format.

    AI/ML Overview

    The provided document describes the Neurosign Avalanche Motor Nerve Monitor and establishes its substantial equivalence to a predicate device, the Neurosign 400. However, it does not contain a comprehensive study with acceptance criteria and detailed device performance results in the format requested. The document focuses on comparing the features and intended uses of the new device to the predicate device in a non-clinical setting.

    Here's an analysis of the provided text with respect to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state quantitative acceptance criteria or report device performance in terms of metrics like sensitivity, specificity, or accuracy. Instead, it relies on a comparison of features and intended use to establish substantial equivalence. The "Non-Clinical Tests Undertaken" section (Section 5.4) lists features that were found to be "similar."

    FeaturePredicate Device (Neurosign 400)Subject Device (Neurosign Avalanche)Reported Similarity/Performance
    Intended useCranial motor nerve monitoringCranial motor nerve monitoringSimilar, Thyroid version restricted to simpler applications
    Method of operationEMG monitor and stimulatorEMG monitor and stimulatorSimilar (detects EMG, amplifies, graphical display)
    Stimulation frequency3 or 30Hz3 or 30HzAll machines use the same frequencies, pulse width, and rectangular waveform
    Method of displayAmplitude triggered response for continuous monitoring, stimulator triggered responseScreen divided into 2 sections (amplitude triggered, stimulator triggered)Information displayed is the same, but presented differently
    Audio14W rms amplifier10W rms amplifierAll machines amplify the EMG signal for audible response. Proportional to input.
    Signal detectionNeedle and Laryngeal electrodeNeedle and Laryngeal electrodeThe same electrodes and stimulating probes are used
    Bandwidth10Hz - 5kHz ±3dB8Hz - 8kHz ±3dB(Not directly stated as similar in the table, but the Subject Device has a broader range compared to the Predicate. This is a technical specification, not performance.)
    Signal Gain477500(A technical specification, Subject Device has slightly higher gain.)
    Stimulation Range0.05mA - 5mA0.05mA - 10mA(A technical specification, Subject Device has a wider range.)
    Software/ScreenC++ proprietary code, 6" electroluminescent displayWindows XP Embedded, 15" color touchscreen(Differences noted, but no performance impact assessed in the summary.)
    Method of ControlDedicated buttons or software via menusTouchscreen (all controls via software except power ON/OFF)(Differences noted, but no performance impact assessed in the summary.)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states that the assessment was conducted in a "non-clinical setting" (Section 5.4). This means there was no patient data, and therefore no "test set" in the traditional sense of a clinical study. The comparison was based on the functional and technical specifications of the devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    Since no clinical study or test set with patient data was involved, there was no need for experts to establish ground truth or interpret results. The comparison was based on engineering and design specifications.

    4. Adjudication Method for the Test Set:

    Not applicable, as there was no test set or clinical data requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    No, an MRMC comparative effectiveness study was not done. The submission is a 510(k) for substantial equivalence, which primarily relies on technological comparison and intended use, not clinical efficacy studies demonstrating improvement for human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    The device is a medical instrument (nerve monitor) used in real-time by a surgeon "with human-in-the-loop," meaning a standalone algorithmic performance isn't applicable in the context of this device. The information it provides is directly interpreted and used by the surgeon.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is the manufacturer's own documentation of the technical specifications, design, and intended use of both the predicate device and the new device. There is no external "ground truth" such as pathology, expert consensus, or outcomes data mentioned for this substantial equivalence submission.

    8. The Sample Size for the Training Set:

    Not applicable. This is a medical device, not an AI/machine learning algorithm that requires a "training set" of data in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there was no training set.

    In summary:

    This 510(k) submission (K083242) for the Neurosign Avalanche Motor Nerve Monitor relies on a non-clinical comparison of technical specifications and intended use against a previously cleared predicate device (Neurosign 400). It establishes "substantial equivalence" based on these comparisons rather than presenting clinical study data with acceptance criteria, sample sizes, ground truth establishment, or human-in-the-loop performance metrics. The document explicitly states that both devices were "assessed in a non-clinical setting and were determined to be similar."

    Ask a Question

    Ask a specific question about this device

    K Number
    K080499

    Validate with FDA (Live)

    Date Cleared
    2008-12-08

    (287 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magstim® Double 70mm Air Film™ Coil is a stimulating coil solely intended for use with the Magstim® Rapid2 Stimulating Unit for the purposes of peripheral nerve stimulation. The coil is an accessory of the Magstim® Rapid2 Unit.

    Device Description

    The Magstim Double 70mm Air Film Coil is an air-cooled coil intended for peripheral stimulation of the nerves. This coil enables deep, and otherwise inaccessible, nerves to be stimulated easily and relatively painlessly. Additionally, no skin preparation is required and stimulation can be achieved through clothing. The Magstim Double 70mm Air Film Coil is for use by, or under the supervision of, a medical practitioner only.

    AI/ML Overview

    The provided text is a 510(k) submission for the Magstim® Double 70mm Air Film™ Coil. This document primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a performance study with acceptance criteria in the context of an AI/ML device.

    Therefore, many of the requested items (e.g., acceptance criteria for reported device performance, sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set details) are not applicable (N/A) because the submission's purpose is not to evaluate the diagnostic or treatment effectiveness of an AI/ML model.

    This submission is for a physical medical device (a stimulating coil) that is an accessory to another device. Its clearance is based on demonstrating that it is as safe and effective as existing legally marketed predicate devices.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied from 510(k) comparison)Reported Device Performance (Magstim® Double 70mm Air Film™ Coil)
    Average Inductance (Comparable to predicates)12 µH (Predicate 1: 16.35 µH, Predicate 2: 15.50 µH) - Note: Lower value, considered acceptable if overall safety/effectiveness is maintained.
    Manner of Stimulation (Peripheral)Peripheral
    Cooling (Feature comparison)Yes (Predicates: No) - Note: This is a distinguishing feature, not an acceptance criterion in the traditional sense, but part of the comparison.
    Multiple Use (Yes)Yes
    Sterile (No)No
    Weight (Comparable to predicates)3kg (Predicate 1: 1.6kg, Predicate 2: 1.8kg) - Note: Higher value, considered acceptable if overall safety/effectiveness is maintained.
    Integral Adapter (Feature comparison)Yes (Predicates: No) - Note: Distinguishing feature, not an acceptance criterion.
    Safety and Effectiveness (Demonstrated as similar to predicates)"The Magstim Double 70mm Air Film Coil is both safe and effective and is similar in its risks and benefits, as well as its manner of performance, to the predicate devices listed above."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • N/A. This submission does not involve a "test set" in the context of evaluating an AI/ML model's performance on clinical data. The comparison is based on device specifications and intended use.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. No ground truth established by experts for a test set in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Not applicable, as there is no test set for clinical performance evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. No MRMC study was conducted. This is a physical device, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. Not applicable, as this is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. Not applicable, as there is no ground truth generation in the context of clinical performance evaluation for this device. The "ground truth" for the submission is the established safety and effectiveness of the predicate devices.

    8. The sample size for the training set

    • N/A. Not applicable, as this is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • N/A. Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071349

    Validate with FDA (Live)

    Date Cleared
    2007-12-12

    (212 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Non-invasively monitoring of the laryngeal nerves during thyroid surgery, and of the Xth cranial nerve during skull-base surgery.

    Device Description

    The Neurosign® Laryngeal Electrode, which was specially developed for use with the Neurosign® 100, 400 and 800 models, consists of: a single-use electrode constructed from medical-grade inks suspended in a polyester substrate; a connector made of polypropylene; and a cable assembly. The electrode is available in two sizes to suit different sexes and age groups. The 6/7mm electrode should be used with a 6 or 7mm endotracheal tube; the 8/9mm electrode should be used with an 8 or 9mm tube. The cable assembly is available separately and will have a long life if it is not abused and is stored appropriately.

    Using the laryngeal electrode lowers the risk of damage to the laryngeal or Xth cranial nerve (the Vagus nerve) during thyroidectomy or parathyroidectomy and, since it is noninvasive, it also lessens the risk of infection during the monitoring procedure.

    AI/ML Overview

    The Neurosign® Laryngeal Electrode is a device intended for intraoperative monitoring of laryngeal nerves during thyroid surgery and the Xth cranial nerve (Vagus nerve) during skull-base surgery. The device consists of a single-use electrode attached to an endotracheal tube, which then connects to a Neurosign® machine for EMG readout.

    The submission does not contain specific acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) of the device. Instead, the FDA's clearance is based on a determination of substantial equivalence to predicate devices. This means that the Neurosign® Laryngeal Electrode is considered as safe and effective as existing legally marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a submission based on substantial equivalence, there are no predefined numerical acceptance criteria or reported performance metrics in the traditional sense (e.g., sensitivity, specificity, accuracy) as would be found in a clinical study evaluating diagnostic performance. The "acceptance" is based on the device having similar characteristics, intended use, and safety/effectiveness profiles to the predicate devices.

    FeatureAcceptance Criteria (based on predicate equivalence)Reported Device Performance (Neurosign® Laryngeal Electrode)
    Laryngeal surface electrodeYesYes
    Monitoring siteTrachea/larynx or Larynx/OesophagusTrachea/larynx
    Monitoring typeContinuous EMG monitoringContinuous EMG monitoring
    CompatibilityMay be used with commercial EMG unitsYes (implicitly, for Neurosign® 100, 400, 800 models)
    Electrode attachmentAttached to surface or embedded within endotracheal tubeAttached to the surface of the endotracheal tube
    Number of electrodes2 or 42
    Number of channels1 or 22
    Surface of electrodeConductive materials (Stainless steel wire, Carbon w/Ag)Conductive ink on a polyester substrate
    Single-use onlyYesYes
    Safety characteristicsNon-invasiveNon-invasive

    2. Sample size used for the test set and the data provenance:

    This document does not describe a clinical study with a "test set" for performance evaluation in the typical sense of diagnostic or AI device studies. The clearance is based on substantial equivalence to existing devices. Therefore, there is no sample size for an independent test set or information on data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as no clinical study with a "test set" requiring expert-established ground truth is described in this submission.

    4. Adjudication method for the test set:

    Not applicable, as no clinical study with a "test set" and adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a laryngeal electrode for nerve monitoring, not an AI-powered diagnostic tool requiring human reader interpretation or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is a physical electrode used in conjunction with a nerve monitor, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not applicable. The basis for clearance is substantial equivalence based on technical characteristics and intended use, not a clinical trial with a ground truth established by expert consensus or other clinical endpoints.

    8. The sample size for the training set:

    Not applicable. This device is a physical electrode, not a machine learning algorithm.

    9. How the ground truth for the training set was established:

    Not applicable. This device is a physical electrode, not a machine learning algorithm.

    Summary of the "Study" (Basis for Acceptance):

    The "study" or basis for acceptance in this 510(k) submission is a comparison to predicate devices to demonstrate substantial equivalence. The manufacturer, The Magstim Company Ltd., compared the Neurosign® Laryngeal Electrode to two predicate devices:

    • Neurovision LSE500 Laryngeal Surface Electrodes (K003745)
    • Medtronic NIM™ EMG Endotracheal Tube (K925640)

    The comparison focused on technical characteristics, intended use, and safety features. The conclusion drawn by the manufacturer (and accepted by the FDA for clearance) is that the Neurosign® Laryngeal Electrode is "both safe and is similar in its risks and benefits, as well as its manner of performance, to the predicate devices." This substantial equivalence determination means the new device is considered as safe and effective as the predicate devices that are already legally marketed. No new clinical trials demonstrating specific performance metrics for the Neurosign® Laryngeal Electrode were necessary for this type of submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051864

    Validate with FDA (Live)

    Date Cleared
    2005-12-21

    (166 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stimulation of Peripheral Nerves.

    Device Description

    This submission is being made because the Magstim Company, referred to as Magsimil, has mounted the device marked under K992911 to replace the external laptop interfaced with the main system via a computer with an embedded interest. Previously, a laptop interfaced with the main bespoke touch-screen display moduler. The new system architecture means that unnt via the I CNCD I V or of also ove controlling the system, resulting in it no longer there is dedicated, in-nouse, so Microsoft Windows operating environment. The requiring all external laptop of Interesses wo and four booster modules, which has been replaced with single and dual module power supplies in the Rapid'. 2001 has been replaced with single are of coils are also introduced with this 510(k), generation contral Magstim Rapid coil family in K992911 can be used with the Rapid2, and the differenced between the 2nd generation and the original coils do not Rapid , and the unrelemed between ss. The intended use of the Magstim Rapid2 and Talse Issues of Salery and Chrostrents in K992911, and the frequency output has been Magstin Super Rapid '15 the Same as also as a power levels, which is consistent with the Changed High Speed Magnetic Stimulator in K905059.

    AI/ML Overview

    The provided document is a 510(k) summary for the Magstim Rapid2 and Magstim Super Rapid2 devices. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than presenting a study designed to establish new acceptance criteria or device performance through a standalone clinical trial with statistically significant outcomes against predefined acceptance criteria.

    Therefore, many of the requested details about acceptance criteria, specific study design, sample sizes, ground truth establishment, expert qualifications, and comparative effectiveness studies are not available in this document.

    Here's a breakdown of the available information:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a clinical study. Instead, it compares the technical specifications and performance characteristics of the new devices (Magstim Rapid2 & Magstim Super Rapid2) against predicate devices (Magstim Rapid K992911 and Cadwell High Speed Magnetic Stimulator K905059) to demonstrate "substantial equivalence." The "performance" reported is essentially the device's technical specifications.

    ParameterMagstim Rapid2 & Magstim Super Rapid2 (New Device)K992911 (Predicate)Cadwell K905059 (Predicate for 60Hz Comparison)"Acceptance Criteria" (Implied by Predicate Performance)
    Field Strength1.2 Tesla for Double 70mm Coil supplied with System1.2 Tesla (depends on coil type and power level)(Not specified for Cadwell)Comparable to 1.2 Tesla
    Super Frequency60Hz (Magstim Super Rapid2)50Hz (Super Rapid)(Not specified for Cadwell in same context)Comparable to 50Hz (Super Rapid)
    Frequency36Hz (Magstim Rapid2)25Hz (Rapid)(Not specified for Cadwell in same context)Comparable to 25Hz (Rapid)
    Frequency Resolution1Hz Steps (user controllable)1Hz Steps (User controllable)(Not specified for Cadwell)1Hz Steps (User controllable)
    Train DurationUp to 10 Seconds depending on frequency and coilUp to 10 seconds depending on frequency and coil(Not specified for Cadwell)Up to 10 seconds depending on frequency and coil
    Train Resolution0.1 seconds0.1 seconds(Not specified for Cadwell)0.1 seconds
    Single Stimulus ModeYesYes(Not specified for Cadwell)Yes
    Interpulse SpacingUp to 250 stimuli per stimulus trainUp to 250 stimuli per stimulus train(Not specified for Cadwell)Up to 250 stimuli per stimulus train
    Number of StimuliUp to 250 stimuli per stimulus trainUp to 250 stimuli per stimulus train(Not specified for Cadwell)Up to 250 stimuli per stimulus train
    Coil Temperature5-40°C Operating Range5-40°C Operating Range(Not specified for Cadwell)5-40°C Operating Range
    36Hz/60HzMax 30% of power / Max 37% of powerUnknown / Max 40% of power (for 60Hz in Cadwell)Max 40% of power (for 60Hz)
    100% Power11Hz/22Hz25Hz25Hz

    The "acceptance criteria" here is implicitly demonstrating that the new devices' specifications meet or exceed the performance of legally marketed predicate devices, thus ensuring they are equally safe and effective for their intended use.

    Specific Study Information:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable / Not provided. This document does not describe a clinical study with a "test set" of patients or data in the context of a diagnostic or treatment effectiveness study where such metrics are typical. The evaluation is based on technical specifications and engineering tests. Data provenance for such engineering tests is not specified, but it would have been conducted by the manufacturer, Magstim Company Limited (UK).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / Not provided. No "ground truth" in the clinical sense (e.g., disease diagnosis) was established by experts for a test set in this 510(k) summary. The "ground truth" for the device's technical performance would be objective measurements and engineering standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable / Not provided. No adjudication of clinical data was performed as no clinical "test set" was described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable / Not provided. This is not an AI-assisted diagnostic device, and no MRMC study was conducted or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable / Not provided. This is an electromagnetic stimulator, not an algorithm, and no standalone performance study in this context was described. The "standalone" performance here relates to the technical specifications of the device itself.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable / Not provided for clinical ground truth. The "ground truth" for this submission are the technical specifications and performance characteristics of the predicate devices and the new devices, as measured and verified through engineering and bench testing. These include parameters like field strength, frequency, train duration, etc., as shown in the tables.
    7. The sample size for the training set:

      • Not applicable / Not provided. As this is not an AI/machine learning device, there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable / Not provided. No training set exists for this type of device.

    Overall Conclusion from the document:

    The submission focuses entirely on substantial equivalence to predicate devices through technical comparisons and engineering performance data. The FDA agreed that "Software documentation and testing, environmental and EMC testing, as well as on measurements where provided to demonstrate safety and performance" were sufficient to establish this. No clinical studies are mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053141

    Validate with FDA (Live)

    Date Cleared
    2005-12-14

    (35 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To locate and identify cranial motor nerves during ENT and intra-cranial procedures.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for a medical device called "Neurosign Nerve Monitor." This document focuses on regulatory approval based on substantial equivalence to a predicate device and does not contain information about acceptance criteria, detailed study results, sample sizes, expert qualifications, or ground truth establishment that would allow me to answer your request.

    The letter explicitly states: "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This indicates that the approval is based on comparison to existing devices, not necessarily a new, detailed performance study with the specific metrics you requested.

    Therefore, I cannot provide the information you asked for based on the input text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991583

    Validate with FDA (Live)

    Date Cleared
    1999-05-19

    (12 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurosign 400 is intended to monitor and to stimulate cranial motor nerves. Cranial motor nerves are monitored by detecting EMG activity in the muscles they innervate.
    To locate and identify cranial motor nerves during ENT and intra-cranial procedures.

    Device Description

    The Neurosign 400 is an EMG machine and nerve stimulator. The machine consists of a central control unit to which a pre-amplifier pod and intramuscular, or surface, electrodes are attached for EMG recordings. Additionally, connected to the main unit is a stimulator pod to which a stimulating probe, which delivers current to nerve tissue, may be connected. The Neurosign 400 is intended to monitor and to stimulate cranial motor nerves. Cranial motor nerves are monitored by detecting EMG activity in the muscles they innervate.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Neurosign 400 Motor Nerve Monitor, focusing on acceptance criteria and the study proving its effectiveness:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document extensively describes the Neurosign 400 as being "similar" to previously cleared devices (Neurosign 800 and Neurosign 100), and relies heavily on the safety and effectiveness history of these predicate devices. The acceptance criteria aren't explicitly quantified with specific performance metrics (e.g., minimum accuracy percentages, sensitivity, specificity). Instead, the criteria are qualitative and based on equivalence and lack of adverse events from predicate devices.

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Safety: No complaints related to safety for similar products."The history of the similar products listed above lacks complaints related to safety and effectiveness, and as far as we are aware there have been no reports of adverse effects or reportable incidents connected with them, or in any published papers on nerve monitoring." The Neurosign 400 "has met all the specifications and expectations for its intended use."
    Effectiveness: No complaints related to effectiveness for similar products."The history of the similar products listed above lacks complaints related to safety and effectiveness..." The Neurosign 400 "has met all the specifications and expectations for its intended use based on the evaluations conducted thus far."
    Equivalence to Predicate Devices: Performance similar to existing cleared devices.The Neurosign 400 is marketed as "similar to the following products which are currently in commercial distribution in the U.S.A.: Neurosign 800 Nerve Monitor, The Magstim Company, K964869; Neurosign 100 Nerve Monitor, The Magstim Company, K923056/A." The FDA's 510(k) clearance explicitly states the device is "substantially equivalent" to predicate devices.
    Meeting Specifications for Intended Use: Ability to monitor and stimulate cranial motor nerves."The Neurosign 400 has met all the specifications and expectations for its intended use based on the evaluations conducted thus far."
    Manufacturing Quality: Materials and methods are consistent with current Magstim Company practices."The materials and methods used to manufacture the Neurosign 400 do not vary from the current materials and methods used by the Magstim Company in the manufacture of the Neurosign 800."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a test set sample size or describe any specific clinical study involving a "test set" in the traditional sense of a new, independent dataset for evaluation.

    Instead, the safety and effectiveness are established through:

    • Historical data of predicate devices: Review of complaints and adverse events for the Neurosign 800 and Neurosign 100. The provenance is implied to be historical market data from the U.S. and potentially other regions where these devices were distributed. This is retrospective data.
    • In-house and on-site clinical evaluation: Performed by The Magstim Company for "all its products, both prior to, and after product release for distribution." The data provenance for these internal evaluations is not detailed (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    No specific "test set" with ground truth established by external experts is described. The acceptance relies on:

    • Market feedback on predicate devices: The absence of reported adverse events or complaints is implicitly the "ground truth" for proving safety and effectiveness of the existing technology.
    • Internal "team participation" for evaluations: "Team members include, but are not limited to Marketing, Manufacturing, Engineering and Quality Assurance." While these individuals are experts in their respective fields, they are internal to the company and not independent clinical experts establishing ground truth for a defined test set. Their qualifications are not specified beyond their departmental roles.

    4. Adjudication Method for the Test Set:

    No specific test set or adjudication method (e.g., 2+1, 3+1) is mentioned or implied, as there was no formal independent clinical study with a predefined test set and expert review for this 510(k) submission. Acceptance is based on the lack of complaints for predicate devices and internal company evaluations.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study is presented or referenced. The submission does not involve comparing human reader performance with or without AI assistance, as the device is a nerve monitor/stimulator and not an AI-driven image analysis tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This detail is not applicable in the context of the Neurosign 400. The device is an EMG machine and nerve stimulator designed for direct use by a clinician during surgical procedures. Its functionality is not based on a standalone algorithm or AI performance that would be evaluated without human interaction. The "performance" is its ability to accurately detect EMG activity and stimulate nerves, which is inherently part of a human-in-the-loop diagnostic/monitoring process.

    7. The Type of Ground Truth Used:

    The ground truth mainly relies on:

    • Historical safety and effectiveness data of predicate devices: Absence of adverse events and complaints for Neurosign 800 and Neurosign 100. This is essentially "outcomes data" in a broad sense, demonstrating safe and effective use in real-world clinical practice over time.
    • Internal company specifications and expectations: The device passed internal evaluations against established product specifications.

    8. The Sample Size for the Training Set:

    The document does not describe a "training set" in the context of machine learning or algorithm development. The device's functionality is based on established electrophysiological principles and hardware design, not on a data-driven learning model. Therefore, sample sizes for training sets are not applicable or mentioned.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no described training set for an algorithm, the concept of establishing ground truth for a training set is not relevant to this 510(k) submission. The safety and effectiveness are established through substantial equivalence based on the history of predicate devices and internal validation processes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K980148

    Validate with FDA (Live)

    Date Cleared
    1998-08-05

    (202 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To locate and identify cranial and peripheral motor nerves during surgery, including spinal nerve roots.

    Device Description

    Neurosign 800. 8 Channel Motor Nerve Monitor

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the Neurosign 800, 8 Channel Motor Nerve Monitor, dated August 5, 1998. The letter states that the device is substantially equivalent to legally marketed predicate devices. However, this type of regulatory correspondence does not contain the detailed information necessary to answer your specific questions about acceptance criteria and a study proving device performance.

    Specifically, the document does not include:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes used for test sets or data provenance.
    3. Details on experts used to establish ground truth or their qualifications.
    4. Adjudication methods.
    5. Information about multi-reader multi-case (MRMC) comparative effectiveness studies.
    6. Description of a standalone algorithm performance study.
    7. The type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    This document is primarily for regulatory clearance and focuses on the substantial equivalence determination. To find the information you're looking for, you would typically need to refer to the 510(k) premarket notification submission itself or potentially peer-reviewed publications related to the device, which are not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964869

    Validate with FDA (Live)

    Date Cleared
    1997-05-14

    (161 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurosign 800 is intended to monitor and to stimulate cranial motor nerves. The motor nerves are monitored by detecting EMG activity in the muscles they innervate.

    Device Description

    The Neurosign 800 is an EMG machine and nerve stimulator. The machine consists of a central control unit to which a pre-amplifier pod and intramuscular or surface electrodes are attached for EMG recordings. Additionally connected to the main unit is a stimulator pod to which a stimulating probe which delivers current to nerve tissue may be connected.

    AI/ML Overview

    This 510(k) submission for the Neurosign 800 Motor Nerve Monitor does not contain the kind of detailed information about acceptance criteria and a specific study proving it meets those criteria that is typically found in submissions for AI/ML-driven medical devices.

    The provided text focuses on:

    • Device Description: What the Neurosign 800 is and how it works.
    • Intended Use: Monitoring and stimulating cranial motor nerves.
    • Substantial Equivalence: Listing similar predicate devices (Neurosign 100, Brackmann II, NIM-2XL).
    • Safety and Effectiveness Justification (General):
      • Lack of complaints/adverse effects for predicate devices.
      • "Extensive in-house and on-site clinical evaluation" of all products by The Magstim Company.
      • Statement that the Neurosign 800 "has met all specifications and expectations for its intended use based on the evaluations conducted thus far."
      • Consistency in manufacturing methods with a predicate device.
      • Conclusion that it is "safe and effective for its intended use."

    It explicitly states: "The Magstim Company performs extensive in-house and on-site clinical evaluation of all its products both prior to and after product release for distribution. Our methods and procedures for acceptance of products as well as our techniques for clinical evaluation are performed using team participation. ... The Neurosign 800 has met all specifications and expectations for its intended use based on the evaluations conducted thus far."

    This is a very general statement and does not provide the specific details requested in your prompt, which are more geared towards validating the performance of an algorithm or an AI model against quantifiable metrics.

    Therefore, for your specific questions, based only on the provided text, the answer is that this information is not available or not described in the given 510(k) summary.

    Here's a breakdown of how the information you requested maps (or doesn't map) to the provided text:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in detail. The text generally states "met all specifications and expectations for its intended use." This is not a quantifiable criterion.Not specified in detail. The text does not provide specific performance metrics, e.g., sensitivity, specificity, accuracy, latency, signal-to-noise ratio, or specific EMG output characteristics.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size (Test Set): Not specified.
    • Data Provenance: Not specified. The evaluations are described as "in-house and on-site clinical evaluation." Whether these were formal studies with defined test sets, controlled environments, or specific data collection methodologies (prospective/retrospective) is not detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not specified. The "team participation" mentioned includes Marketing, Manufacturing, Engineering, Clinical Engineering, Manufacturing Engineering, and Quality Assurance, which are involved in "methods and procedures for acceptance of products" and "techniques for clinical evaluation." This is a product development team, not necessarily a panel of independent clinical experts establishing ground truth for a test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not specified. This device is a nerve monitor and stimulator, not an AI-driven diagnostic tool that assists human readers in interpreting images or data in the way an AI algorithm might. The summary does not mention any AI component or human-in-the-loop assistance for interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not specified. This is a hardware device with associated software for signal acquisition and display, not an algorithm performing a diagnostic or predictive function independently.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified. Given the nature of an EMG and stimulator, typical "ground truth" might relate to direct observation of nerve stimulation response, correlation with surgical findings, or clinical neurological assessment, but this is not detailed in the summary.

    8. The sample size for the training set

    • Not applicable/Not specified. This device does not describe an AI/ML component that uses a "training set."

    9. How the ground truth for the training set was established

    • Not applicable/Not specified.

    In summary: The provided 510(k) refers to general internal company evaluations and clinical assessments that led to the conclusion of safety and effectiveness, drawing heavily on substantial equivalence to predicate devices and their unblemished safety record. It does not contain the detailed, quantitative study data or acceptance criteria typically provided for AI/ML device submissions, or even for non-AI devices where specific performance claims (e.g., accuracy, precision) need to be validated in a formal study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1