510(k) Data Aggregation

    K Number
    K211851

    Validate with FDA (Live)

    Device Name
    Ora-Aid
    Manufacturer
    Date Cleared
    2022-01-11

    (210 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intra-oral Wound Dressing is intended for the management of all types of oral wounds, injuries and ulcerations of the gingival and oral mucosa, including stomatitis, minor chafing and traumatic ulcers, abrasions caused by braces and illfitting dentures, and lesions associated with oral surgery.

    Intra-oral Wound Dressing operates to relieve pain by adhering to and protecting affected tissues from further irritation, thereby allowing healing.

    Device Description

    Intra-oral Wound Dressing is a device comprised of an oral mucosa adhesive side and a protection side. The oral mucosa adhesive side is composed of a water soluble polymer that when exposed to moisture in the oral cavity changes into a gel state to achieve adhesion to the wound area. The protection side consists of a waterinsoluble polymer which covers the wound to protect the applicable area from the environment in the oral cavity. Intra-oral Wound Dressing is simply a non-sterile bandage to protect oral wounds.

    AI/ML Overview

    This FDA 510(k) summary does not contain information about a clinical study with acceptance criteria and device performance. Instead, it describes a non-clinical evaluation of the "Ora-Aid Intra-oral Wound Dressing" (subject device) for substantial equivalence to a predicate device, "PerioPatch."

    The document concludes that the subject device is substantially equivalent to the predicate device based on non-clinical safety and performance tests, not a clinical study demonstrating performance against acceptance criteria in human subjects.

    Here's a breakdown of the requested information based on the provided text, highlighting what is not present as well:

    1. A table of acceptance criteria and the reported device performance

    This information is not available from the provided text for device performance against specific clinical acceptance criteria. The document focuses on demonstrating substantial equivalence through non-clinical testing against standards.

    Acceptance Criteria (Not Explicitly Stated for Clinical Performance)Reported Device Performance (Non-Clinical)
    Biocompatibility requirementsMet ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Sensitization, Intracutaneous Reactivity Test), ISO 10993-11:2016 (Acute Systemic Toxicity Test, Material-Mediated Pyrogen Test)
    Microbial limitsMet KP11th standard (same test method as USP 40NF35 <61> and USP 40NF35 <62>); no growth observed.
    Shelf-life stabilityMet ASTM F 1980 (Accelerated Aging Test)
    Absorbency (functional performance)Met EN 13726-1:2002
    Adhesion time (functional performance)Subject device: 4 hours; Predicate device: 30-45 minutes. (Subject device's longer adhesion time did not raise concerns regarding safety and effectiveness)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: This information is not specified as a clinical test set was not described. The non-clinical tests would have their own sample sizes (e.g., for biological materials, microbial samples, physical property tests), but these are not detailed in the summary.
    • Data provenance: The submission is from TBM Corporation in Gwangju, Korea, and the consultant KMC, Inc. is in Seoul, Republic of Korea. The standards used for non-clinical testing are international (ISO, EN, ASTM) and US Pharmacopeia (USP).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as there was no clinical study described that required expert-established ground truth. The evaluation was based on non-clinical laboratory testing against established standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there was no clinical study described that required adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an "Intra-oral Wound Dressing" and not an AI-powered diagnostic or assistive device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical wound dressing and not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This is not applicable as no clinical ground truth was established for a performance study. The evaluation relied on adherence to international standards and pharmacopeia for non-clinical safety and performance.

    8. The sample size for the training set

    This is not applicable as this is a physical medical device, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    This is not applicable as this is a physical medical device, not an AI/ML algorithm requiring a training set or ground truth for training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183436

    Validate with FDA (Live)

    Device Name
    OraCure
    Manufacturer
    Date Cleared
    2019-09-18

    (281 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OraCure® is intended for use as an oral wound dressing to protect ulcer tissue by forming a physical barrier on the wound tissue to avoid further irritation and thus to relieve pain. It is indicated for use with all types of ulcers and small wounds of the oral mucosa including canker sores, aphthous ulcers, and injuries such as traumatic ulcers caused by self-biting, braces, and ill-fitting dentures.

    Device Description

    OraCure® is an oral wound dressing comprising a mucous-adhesive side on a protection side, wherein the adhesive side has a strong adhesion when it is applied and hydrated on mucus. Adhesive side is water-soluble polymer that reacts with moisture (water and saliva) in the mouth within a short time and changes into a gel state. The protection side for protection of the mucous-adhesive side consists of water-insoluble polymer on the protection side is insoluble in the water and covers the wound to protect the applicable area from the environment (bacteria, saliva, food, etc.) in the mouth. OraCure® does not contain any biological additives or drugs with respect to wound treatment, but it is simple non-sterile bandage to protect ulcer tissue.

    AI/ML Overview

    The provided text describes the OraCure® oral wound dressing and its substantial equivalence to a predicate device, Ulceloocin™ Oral Ulcer Patch. The document focuses on non-clinical data to support this claim, rather than a clinical study involving human patients. Therefore, information regarding acceptance criteria, device performance in terms of clinical outcomes, sample sizes for test and training sets, number of experts for ground truth, adjudication methods, or MRMC comparative effectiveness studies are not present.

    However, the document does list the acceptance criteria in terms of safety and performance tests performed and indicates that the device meets these criteria through the successful completion of these tests.

    Here's a summary of the available information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The device performance is reported as meeting the standards for each test listed. The document concludes that "the safety and performance test reports are supported to the safety and effectiveness of the subject device."

    Test ItemStandards UsedReported Device Performance
    Biocompatibility TestsPassed, demonstrating no undue risks from biological hazards.
    CytotoxicityISO 10993-5:2009Passed
    SensitizationISO 10993-10:2010Passed
    Intracutaneous Reactivity TestISO 10993-10:2010Passed
    Acute Systemic Toxicity TestISO 10993-11:2016Passed
    Material-Mediated Pyrogen TestISO 10993-11:2016Passed
    Subacute/Subchronic Toxicity TestISO 10993-11:2016Passed
    GenetoxicityISO 10993-3:2014Passed
    ImplantationISO 10993-6:2016Passed
    Microbial Limit Test (MLT)Passed, indicating acceptable levels of microorganisms in the non-sterile product.
    Microbial Limit TestUSP 40NF35 <61>, <62>, <1111>Passed
    Shelf-life TestPassed, establishing the expiration date and stability of physical properties.
    Accelerated Aging TestASTM F 1980Passed
    Performance TestDemonstrated effectiveness of product performance.
    AbsorbencyEN 13726-1:2002Passed
    Moisture Vapour Transmission RateEN 13726-2:2002Passed
    ConformabilityEN 13726-4:2003Passed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The study described is a series of laboratory-based non-clinical tests rather than a clinical trial with human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document describes non-clinical laboratory testing, not a study requiring expert interpretation of results for ground truth establishment. The "ground truth" here would be defined by the Pass/Fail criteria of the respective international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as there were no subjective assessments requiring adjudication described in the non-clinical tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The document describes a medical device (wound dressing), not an AI-powered diagnostic or assistive technology. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical wound dressing, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical tests is based on the objective criteria and analytical methods defined by the international standards listed (e.g., ISO, USP, ASTM, EN standards). These standards outline specific methodologies and acceptance ranges for each test item.

    8. The sample size for the training set

    This information is not provided and is not applicable as the document refers to non-clinical testing of a physical medical device, not a machine learning model that would require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as no training set was used.