510(k) Data Aggregation

    K Number
    K120968

    Validate with FDA (Live)

    Date Cleared
    2012-09-19

    (173 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

    AI/ML Overview

    The provided text describes a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving performance against acceptance criteria in the typical sense of AI/software device evaluation.

    Therefore, many of the requested points related to AI/software device studies (like sample size for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, training sets, etc.) are not applicable to this type of medical device submission. The device is a physical product (gloves), not a software or AI tool.

    However, I can extract the relevant acceptance criteria and reported performance based on the information provided, which refers to conformance with ASTM standards.

    Here's the breakdown of the information as it applies to this specific submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (from ASTM standards and FDA)Reported Device Performance (by manufacturer)Result of Comparison
    Device Description & SpecificationsMeet all current specifications listed under ASTM Specification D5250 (If vinyl)Meets ASTM D5250Substantially equivalent
    MaterialIf the glove is made of a polymer or other type of material, identify the material.PVCSubstantially equivalent
    Dusting or Donning Powder / LubricantIf a donning lubricant is used, state the composition and include biocompatibility data for the lubricant; also state name, manufacturer, and address. (For "powder-free" gloves, meet residual powder limits per ASTM D6124).PU (Polyurethane) as surface coating agent. (The document implies it is powder-free, aligned with the device name and ASTM D6124 conformance)Substantially equivalent
    Product SpecificationsCertify that finished "powder-free" gloves meet ASTM D5250.Meets ASTM D5250Substantially equivalent
    Performance Data (Holes Detection)Meet ASTM D5151 (Detection of Holes in Medical Gloves)Meets ASTM D5151Substantially equivalent
    Performance Data (Residual Powder)Meet ASTM D6124 (Residual Powder on Medical Gloves)Meets ASTM D6124Substantially equivalent
    BiocompatibilitySkin Irritation: Dermal and Sensitization studies per ISO 10993-10SKIN IRRITATION DERMAL and SENSITIZATION STUDIES Meets ISO 10993-10Substantially equivalent
    Water Leak Test on Pinhole AQLMeet FDA requirements for AQL (Implicitly covered by ASTM D5151 for holes detection)The conclusion states: "Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL..." (specific AQL not stated in table, but overall claim made)-
    LabelingRequirements for labeling (e.g., "Powder-Free", "Patient Examination Glove", "Single Use Only", "Manufactured For:", "Lot")Labeling includes: -Powder-Free -Patient Examination Glove -Single Use Only - Manufactured For: - LotSubstantially equivalent

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is not described as a single, centralized study in the document. Instead, it refers to the device's conformance with recognized consensus standards (ASTM D5250-06, ASTM D6124-06, ASTM D5151, and ISO 10993-10) and FDA requirements (specifically 21 CFR 800.20 for water leak test on pinhole AQL).

    The manufacturer asserts that their "Powder free vinyl patient examination gloves, Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10." This implies that the device successfully passed testing according to the methodologies outlined in these standards.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated in the summary. For physical products tested against ASTM standards, the sample sizes are typically defined within the specific standard's test methods (e.g., for AQL testing for holes, specific sample plans are used).
    • Data Provenance: Not explicitly stated beyond "nonclinical submitted". Given the submitter's address in China, it's presumed the testing was conducted in China. It is a retrospective evaluation against established standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This is a physical device submission. Ground truth in the context of expert consensus for AI models is not relevant here. Conformance is determined by physical and chemical testing against objective standards.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As per point 3, this is a physical device tested against standards; an adjudication method in the context of expert review is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, Not Applicable. MRMC studies are for evaluating diagnostic performance of imaging systems or AI software. This is a physical glove.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • No, Not Applicable. This is a physical device.

    7. The Type of Ground Truth Used:

    • For this device, "ground truth" equates to the objective measurements and results obtained from testing the gloves against the criteria and methodologies defined in the referenced ASTM standards (D5250, D5151, D6124) and ISO standard (10993-10), as well as FDA regulations for AQL. These are direct physical properties and performance characteristics.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is a physical device; there is no "training set" in the context of an AI/software algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. As per point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120970

    Validate with FDA (Live)

    Date Cleared
    2012-07-19

    (111 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Patient Examination Gloves, Coloured (Blue) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Nitrile Patient Examination Gloves , Blue Color that meets all of the requirements of ASTM standard D 6139-10

    AI/ML Overview

    This document is a 510(k) summary for a medical device called "Powder Free Nitrile Patient Examination Gloves, Blue Color." It aims to demonstrate that this new device is substantially equivalent to a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Device Performance
    DimensionASTM standard D 6139-10Meets
    Physical PropertiesASTM standard D 6139-10Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-06 and D6124-06<2mg/glove
    Biocompatibility: Primary Skin IrritationPrimary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    Biocompatibility: Dermal SensitizationDermal sensitization in the guinea pigPasses (Not a Dermal sensitization)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample sizes used for the various tests.
    The data provenance is not explicitly stated as retrospective or prospective, nor does it specify the country of origin of the data. However, the manufacturer is from China (Tangshan Zhonghong Pulin Plastic Co., Ltd., China), suggesting the testing might have been conducted there or through certified labs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this device. The testing involves adherence to established technical standards (ASTM, CFR, ISO) for physical properties and biocompatibility, not expert interpretation of diagnostic images or clinical outcomes. The "ground truth" is defined by the objective pass/fail criteria of these standards.

    4. Adjudication Method for the Test Set

    This information is not applicable to this device. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or clinical events, not for standardized physical and chemical property testing of a product like examination gloves. The testing follows predefined methodologies within the cited standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not conducted and is not relevant for this device. Such studies are typically performed for diagnostic or screening devices where human readers interpret results, often with AI assistance. This device is a patient examination glove, and its effectiveness is determined by its physical and biological properties.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone performance study was not conducted and is not relevant for this device. This concept applies to AI/software algorithms. The "performance" of these gloves is their intrinsic material and functional characteristics, which are tested through laboratory methods.

    7. Type of Ground Truth Used

    The ground truth used for this device is based on established technical standards and regulatory requirements. Specifically:

    • Physical properties (Dimension, Physical Properties, Freedom from pinholes, Powder Residual): Defined by the specified ASTM standards (D 6139-10, D 5250-06, D6124-06) and FDA regulation (21 CFR 800.20). The "ground truth" is whether the glove's measured properties fall within the acceptable ranges or thresholds defined by these standards.
    • Biocompatibility: Defined by the results of specific animal tests (Primary Skin Irritation in rabbits, Dermal Sensitization in guinea pigs) which have clear "Pass" criteria for demonstrating absence of irritation or sensitization. The ISO 10993-10 standard would outline the methodology and criteria for these tests.

    8. Sample Size for the Training Set

    This information is not applicable to this device. The concept of a "training set" belongs to machine learning and AI algorithms. The manufacturing process of gloves does not involve a training set in this context. The "training" for manufacturing would be the established quality control and production procedures.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as point 8.