Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K082597

    Validate with FDA (Live)

    Date Cleared
    2008-10-24

    (46 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the "Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)" device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance CriteriaDevice Performance
    DimensionASTM standard D 5250-06Meets
    Physical PropertiesASTM standard D 5250-06Meets
    Freedom from pinholes21 CFR 800.20 (Water leak test on pinhole AQL)Meets
    Powder ResidualASTM standard D 5250-06 and D6124-01 (< 2mg/glove implied from predicate)<2mg/glove
    Biocompatibility (Primary Skin Irritation)Primary Skin Irritation in rabbits (ISO10993-10)Passes (Not a Primary Skin Irritation)
    Biocompatibility (Dermal Sensitization)Dermal sensitization in the guinea pig (ISO10993-10)Passes (Not a Dermal sensitization)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for testing each characteristic. The data provenance is not explicitly stated in terms of country of origin for the individual tests, but the device manufacturer is based in China. The studies appear to be prospective as they are conducted to demonstrate the device meets specific standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This information is not provided in the document. For device testing (like these gloves), the "ground truth" is typically defined by the objective measurement adhering to the specified ASTM standards and CFR regulations, rather than expert consensus on subjective interpretations. For biocompatibility, the interpretation of results would be by qualified laboratory personnel following established protocols.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods like "2+1" or "3+1" are typically used in scenarios where multiple human readers are assessing subjective data (e.g., medical images) and their disagreements need to be resolved. For objective tests on physical properties of gloves, the results are typically directly measured against a standard, not adjudicated by a panel of experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not applicable to the assessment of physical properties or biocompatibility of examination gloves.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical product (gloves), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on objective measurements against established industry standards and regulatory requirements.

    • ASTM standards (D 5250-06, D6124-01): These standards define the acceptable range for dimensions, physical properties, and powder residue. The ground truth is whether the measured values fall within these defined ranges.
    • 21 CFR 800.20: This regulation defines the acceptable level of freedom from pinholes based on a water leak test. The ground truth is the pass/fail outcome of this test.
    • ISO10993-10: This standard governs biocompatibility testing. The ground truth is the biological response observed (e.g., presence or absence of irritation/sensitization) as interpreted by standard toxicological protocols.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. The development and testing of these gloves do not involve a "training set" in the context of machine learning or AI models. The gloves are manufactured and then tested against specified standards.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1