510(k) Data Aggregation

    K Number
    K181523

    Validate with FDA (Live)

    Date Cleared
    2018-07-11

    (30 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CO2RE is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).

    The CO-RE is indicated for dermatological procedures requiring ablation, coagulation and resurfacing of soft tissue, including skin.

    The Standard Handpiece of the CO2RE System is indicated for the following:

    Dermatology & Plastic Surgery
    The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

    • laser skin resurfacing
    • laser dermabrasion
    • laser burn debridement.

    Laser skin resurfacing (ablation and/or vaporization) for the treatment of:

    • wrinkles, rhytids, and furrows (including fines lines and texture irregularities)
      Clinical literature demonstrates that skin resurfacing of wrinkles, rhytids, and furrows with CO2 laser increases the amount of sub-epidermal collagen.

    Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:

    • keratoses, including actinic and sebortheic keratosis, sebortheic wart, and verruca sebortheica;
    • vermillionectomy of the lip;
    • cutaneous horns;
    • solar/actinic elastosis;
    • chelitis, including actinic chelitis;
    • lentigines, including lentigo maligna or Hutchinson's malignant freckle
    • uneven pigmentation/ dyschromia;
    • acne scars;
    • surgical scars:
    • keloids including acne keloidalis nuchae;
    • hemangiomas (including Buccal, port wine and pyogenic granuloma pyogenicum/granuloma telagiectaticum):
    • tattoos;
    • telangiectasia:
    • removal of small skin tumors, including periungual (Koenen) and subungual fibromas;
    • superficial pigmented lesions;
    • adenosebaceous hypertrophy or sebaceous hyperplasia;
    • rhinophyma reduction;
    • cutaneous papilloma (skin tags);
    • milia;
    • debridement of eczematous or infected skin;
    • basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions;
    • nevi, including spider, epidermal and protruding;
    • neurofibromas:
    • laser de-epithelialization;
    • tricoepitheliomas;
    • xanthelasma palpebrarum;
    • syringoma

    Laser ablation, vaporization and /or excision for complete or partial nail matrixectomy. Vaporization of:

    • benign/malignant vascular/avascular skin lesions;
    • Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts.

    Podiatry
    Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of:

    • verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts;
    • fungal nail treatment;
    • porokeratoma ablation;
    • ingrown nail treatment;
    • neuromas/fibromas, including Morton's neuroma;
    • debridement of ulcers;
    • other soft tissue lesions.

    Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.

    Otolaryngology (ENT)
    Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:

    • leukoplakia, including oral, larynx, uvula, palatal, upper latera pharyngeal tissue;
    • nasal obstruction;
    • adult and juvenile papillomatosis polyps;
    • polypectomy of nose and nasal passages; lymphangioma removal;
    • removal of recurrent papillomas in the oral cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords;
    • superficial lesions of the ear, including chondrodermatitis nodularis chronica helices/Winkler's disease;
    • rhinophyma;
    • verrucae vulgares (warts);

    Gynecology (GYN)
    Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:

    • conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia (VIN, VAIN);
    • condyloma acuminate, including cervical, vulvar, perineal, and Bowen's disease, (Erythroplasia of Quevrat) and Bowenoid papulosa (BP) lesions;
    • leukoplakia (vulvar dystrophies);
    • herpes vaporization;
    • urethral caruncle vaporization;
    • cervical dysplasia:
    • benign and malignant tumors;

    Dental/Oral Surgery
    Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:

    • treatment of ulcerous lesions, including aphthous ulcers;
    • excision and ablation of benign and malignant lesions;
    • leukoplakia:

    Genitourinary
    Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include:

    • benign and malignant lesions of external genitalia;
    • condyloma;
    • erythroplasia

    The Fractional Handpiece of the CO2RE System is indicated for dermatological procedures requiring ablation, coagulation and resurfacing of soft tissue, including skin.

    The Surgical Handpiece of the CO2RE System is indicated for the following:

    Dermatology & Plastic Surgery
    Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:

    • vermillionectomy of the lip;
    • cutaneous horns;
    • removal of small skin tumors, including periungual (Koenen) and subungual fibromas;
    • rhinophyma reduction;
    • basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions;
    • neurofibromas;
      Laser ablation, vaporization and /or excision for complete or partial nail matrixectomy. Vaporization/coagulation of:
    • benign/malignant vascular/avascular skin lesions;
    • Moh's Surgery;
    • Lipectomy;
    • Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts.

    Laser incision and /or excision of soft tissue for the performance of upper and lower evelid blepharoplasty.

    Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.

    Podiatry
    Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of:

    • verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts;
      Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.

    Otolaryngology (ENT)
    Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:

    • choanal atresia,
    • nasal obstruction:
    • adult and juvenile papillomatosis polypectomy of nose and nasal passages; lymphangioma removal;
    • removal of vocal cord/fold nodules, polyps and cysts;
    • removal of recurrent papillomas in the oral cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords;
    • laser/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue;
    • Zenker's Diverticulum/pharyngoesophageal diverticulum [endoscopic laser-assisted esophagodivertuculostomy (ELAED)];
    • stenosis, including subglottic stenosis;
    • tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ablation/tonsillotomy;
    • pulmonary bronchial and tracheal lesion removal;
    • benign and malignant nodules, tumors and fibromas (larynx, trachea, tracheobronchial/endobronchial);
    • benign and malignant lesions and fibromas (nose and nasal passages);
    • benign and malignant tumors and fibromas (oral);
    • stapedotomy/stapedectomy; acoustic neuroma in the ear;
    • superficial lesions of the ear, including chondrodermatitis nodularis chronica helices/Winkler's disease:
    • telangiectasia/hemangioma of larynx and trachea (includes uvula, palatal or upper lateral pharyngeal tissue);
    • cordectomy, cordotomy (for the treatment of vocal fold motion impairment), and cordal lesions of larynx, pharynx and trachea;
    • myringotomy/tympanostomy (tympanic membrane fenestration);
    • uvulopalatoplasty (LAUP, laser UPPP);
    • turbinectomy and turbinate reduction; septal spur ablation/reduction and septoplasty; partial glossectomy;
    • tumor resection on oral, subfacial and neck tissues;
    • rhinophyma;
    • verrucae vulgares (warts):
    • gingivoplasty/gingivoctomy.

    Gynecology (GYN)
    Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:

    • conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia (VIN, VAIN);
    • condyloma acuminate, including cervical, genital, and Bowen's disease, (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions;
    • incision and drainage (1&D) of Bartholin's and nubuthian cysts;
    • benign and malignant tumors;
    • hemangiomas.

    GYN Laparoscopy
    Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic surgery, including GYN laparoscopy, for treatment of:

    • endometrial lesions, including ablation of endometriosis;
    • excision/lysis of adhesions;
    • salpingostomy
    • oophorectomy/ovariectomy; fimbrioplasty;
    • metroplasty:
    • microsurgery (tubal);
    • uterine myomas and fibroids;
    • ovarian fibromas and follicle cysts;
    • uterosacral ligament ablation;
    • hysterectomy.

    Neurosurgery
    Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of: Cranial

    • posterior fossa tumors;
    • peripheral neurectomy;
    • benign and malignant tumors and cysts (e.g. gliomas (including basal tumors), acoustic neuromas, lipomas and large tumors);
    • arteriovenous malformation;
    • pituitary gland tumors (transphenoidal approach).

    Spinal Cord

    • incision/excision and vaporization of benign and malignant tumors and cysts;
    • intra- and extradural lesions;
    • laminectomy/ laminotomy/ microdiscectomy.

    Orthopedics
    Incision/excision and vaporization and of soft tissue in orthopedic surgery. Applications include: Arthroscopy

    • menisectomy;
    • chondromalacia:
    • chondroplasty;
    • ligament release (lateral and other);
    • excision of plica;
    • partial synovectomy.

    General

    • debridement of traumatic wounds;
    • debridement of decubitus and diabetic ulcers;
    • microsurgery:
    • artificial joint revision;
    • PMMA removal.

    General/Thoracic Surgery
    Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:

    • debridement of decubitus ulcers, stasis, diabetic and other ulcers;
    • mastectomy;
    • debridement of bums;
    • rectal and anal hemorrhoidectomy;
    • breast biopsy;
    • reduction mammoplasty;
    • cytoreduction for metastatic disease;
    • laparotomy and laparoscopic applications;
    • mediastinal and thoracic lesions and abnormalities;
    • skin tag vaporization;
    • atheroma;
    • cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips;
    • pilonidal cyst removal and repair;
    • abscesses:
    • other soft tissue applications.

    Dental/Oral Surgery

    • Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
    • gingivectomy- removal of hyperplasias;
    • gingivoplasty:
    • incisional and excisional biopsy;
    • incision of infection when used with antibiotic therapy;
    • frenectomy (frenum release);
    • excision and ablation of benign and malignant lesions;
    • homeostasis;
    • operculectomy;
    • crown lengthening;
    • removal of soft tissue, cysts and tumors;
    • oral cavity tumors and hemangiomas;
    • abscesses;
    • extraction site hemostasis;
    • salivary gland pathologies;
    • preprosthetic gum preparation;
    • partial glossectomy;
    • periodontal gum resection.

    Genitourinary
    Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include:

    • benign and malignant lesions of external genitalia;
    • condyloma;
    • Phimosis;
    • Erythroplasia.
    Device Description

    The CO₂RE System is a computer controlled radio-frequency (RF) excited carbon dioxide (CO2) laser system that delivers CO2 energy to the target treatment area at 10,600 nm. The CO2RE Laser System is comprised of multiple components, including the control unit and six handpieces. Laser is transmitted to the tissue via a series of lenses integrated into the articulated arm. The CO2 laser energy is absorbed by water in tissue to achieve its intended treatment effects.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) summary relating to the CO2RE Laser System, focusing on the requested acceptance criteria and study details.

    Important Disclaimer: The provided document is a 510(k) Summary, not a full 510(k) submission. A 510(k) Summary focuses on demonstrating substantial equivalence to a predicate device, and often relies on existing knowledge of the technology rather than extensive novel clinical studies. Therefore, many of the requested details about acceptance criteria, human reader studies, and detailed ground truth establishment might not be explicitly present or required for a 510(k) submission for this type of device. The information provided below is extracted directly from the given text.


    Acceptance Criteria and Device Performance for CO2RE Laser System (K181523)

    1. Table of Acceptance Criteria and Reported Device Performance

    Criterion / AspectAcceptance Criteria (Implicit for 510(k) based on predicate)Reported Device Performance (Summary of Study Findings)
    Intended UseSame as predicate device (CO₂RE Laser System K151655) for surgical applications requiring ablation, vaporization, excision, incision, and coagulation of soft tissue."The CO₂RE System has the same intended use... as its predicate devices."
    Technological CharacteristicsSimilar to predicate device (CO₂RE Laser System K151655) including wavelength, pulse duration, maximum laser power, and maximum energy."The CO₂RE System technological characteristics are very similar to those of the predicate devices. The CO₂RE System and its predicate devices are all RF excited CO₂ laser systems that operate at the same or similar parameters... The systems are comprised of the same or similar components for the same intended use."
    Performance (General)Performs as intended for a CO2 laser surgical instrument."The CO₂RE System performs as intended based on the performance data provided in the submission."
    Software Verification & ValidationSoftware is appropriate for release."Verification and validation testing of the software confirm that the software version is appropriate for release."
    Electrical Safety & Electromagnetic Compatibility (EMC)Meets applicable recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22)."The CO₂RE System has been tested for electromagnetic compatibility and electrical safety per the applicable recognized consensus standards... Results were passing."
    BiocompatibilityEstablished per FDA's guidelines for patient-contacting components."Biocompatibility of the patient contacting components of the device has been established per FDA's guidelines."
    SterilizationSterilization parameters are validated."Sterilization parameters have been validated."
    Safety and Effectiveness QuestionsTechnological differences do not raise new safety or effectiveness questions compared to predicate."The technological differences between the CO₂RE System and its predicate devices do not raise any new types of safety or effectiveness questions given that the key treatment parameters remain similar..."
    Overall ConclusionSubstantially equivalent to predicate."The CO₂RE System is substantially equivalent to its predicate."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of typical AI/software performance validation involving patient data. For this laser device, the "testing" refers to engineering and quality assurance tests on the device itself.

    • Sample Size: Not applicable in the traditional sense of a clinical test set for diagnostic accuracy. Testing was conducted on the device's components and system.
    • Data Provenance: Not applicable. The "data" here refers to measurements from electrical safety, EMC, software, biocompatibility, and sterilization tests. These are internal lab tests, not patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts: Not applicable. Ground truth, in the context of a medical laser device for surgical and dermatological procedures, typically relies on established medical knowledge, clinical outcomes, and the device's physical/electrical parameters meeting specified standards. It doesn't involve "experts" labeling a test set of images or patient data for an algorithm.
    • Qualifications of Experts: Not applicable for the type of testing described.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. The device's performance is demonstrated through adherence to engineering standards and technical specifications, not through a diagnostic assessment requiring expert adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • MRMC Study: No. This is a laser surgical instrument, not an AI-assisted diagnostic tool where human readers compare performance with and without AI. The submission focuses on demonstrating the device's technical specifications and safety/effectiveness equivalence to a predicate, not improving human reader performance.
    • Effect Size: Not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

    • Standalone Performance: Not applicable. This device is a directly operated laser system, not an artificial intelligence algorithm that operates independently or provides diagnostic output. It does not have an "algorithm only" mode in the sense of AI performance evaluation. The "software" mentioned refers to the operating software of the physical device.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For this type of device, "ground truth" is established by:
      • Engineering Specifications and Benchmarks: The device's output (e.g., laser power, wavelength, pulse duration) must meet design specifications.
      • Industry Consensus Standards: Compliance with recognized standards like IEC 60601-1 for electrical safety and electromagnetic compatibility.
      • Predicate Device Performance: The primary ground truth for a 510(k) is the demonstration of substantial equivalence to a legally marketed predicate device, implying that if the predicate is safe and effective, the new device with similar characteristics and performance also is.
      • Biocompatibility Data: Established guidelines for material safety.
      • Sterilization Validation: Demonstrated ability to be safely sterilized.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device does not use machine learning with a "training set" of data in the typical sense. Its functionality is based on established physical principles of laser operation.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set Establishment: Not applicable, as there is no "training set" for an AI algorithm.

    Summary of 510(k) Specifics for CO2RE Laser System:

    This 510(k) submission for the CO2RE Laser System (K181523) is an example of a substantial equivalence determination for a medical device that is not an AI/ML-driven diagnostic or therapeutic tool. Its 'acceptance criteria' and 'study' are primarily centered around:

    • Predicate Equivalence: Demonstrating that the new device has the same intended use and similar technological characteristics, indications, and principles of operation as a previously cleared device (K151655).
    • Performance Testing: Laboratory and engineering tests to ensure the device meets declared technical specifications, electrical safety, EMC, biocompatibility, and sterilization requirements.
    • Software Validation: Ensuring the embedded software controls the device appropriately and safely.

    The framework for evaluating AI/ML devices with detailed criteria like MRMC studies, expert adjudication, and distinct training/test sets is not applicable to a device like the CO2RE Laser System.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180593

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-06-05

    (91 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    1,064nm:

    The Vbeam Prima Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lakes, leg veins, and poikiloderma of civatte and treatment of benign cutaneous lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots), cafe au lait manules, sebortheic keratoses, nevi, chloasma, verrucea, skin tags, and keratoses. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

    595 nm:

    General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions.

    Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia. rosacea, port wine stains, hemangioma, spider angioma, poikiloderma of civatte, and benign cutaneous lesions, such as warts, scars, striae, psoriasis and the treatment of benign epidermal pigmented lesions. Treatment of inflammatory acne vulgaris.

    Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology.

    Podiatry: Treatment of benign cutaneous lesions, such as warts.

    Device Description

    The Vbeam Prima Laser system is designed as a transportable device that delivers laser energy at wavelengths of 595 nm (pulse-dye) or 1064 nm (Nd: YAG). The device is comprised of a main console, power supply, controlled electronics, firmware, software, cooling system, and handpieces. There is also an internal calibration port with an internal meter located in the system. The handpiece offers a set of distance gauges available in the following spot ranges: 3-7 mm, 7-11 mm, 11-15 mm, and 3 X 10 mm. The user selects treatment settings, such as fluence, number of pulses, and spot size selection via a color touch screen. The handpiece incorporates a Dynamic Cooling Device (DCD) into its handpiece that is designed to cool the skin surface during treatment, which consists of an electrically controlled spray nozzle, cryogen reservoir canister and an electronic control circuitry. The Vbeam Prima Laser System also offers an optional EverCool Delivery System Handpiece for skin cooling during treatment.

    AI/ML Overview

    The Vbeam Prima Laser System's acceptance criteria and the study proving it meets these criteria are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Electrical Safety and EMCCompliance with recognized standards:The device passed the required testing and is in full compliance with the following standards:
    - IEC 60825-1- IEC 60825-1: Safety of Laser Products – Part 1: Equipment classification and requirements
    - IEC 60601-1-2- IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for safety - collateral standard: electromagnetic compatibility (EMC)requirements and test
    - IEC 60601-1- IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
    - IEC 60601-2-22- IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
    BiocompatibilityEstablished based on predicate devices and ISO standards.Biocompatibility was established based on predicate devices and the results of ISO 10993-5 (in vitro cytotoxicity) and ISO 10993-10 (irritation and skin sensitization) series of testing.
    Software Verification & ValidationAcceptable for software release according to FDA guidance.Software verification and validation testing was conducted, and the results demonstrated that testing results were found acceptable for software release. Software testing was performed per FDA's guidance document "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices."
    Bench Testing (Cooling Performance)DCD handpiece and Evercool handpiece produce similar cooling results.A bench study was conducted to confirm that the DCD handpiece and Evercool handpiece can produce similar cooling results. (The document does not provide specific performance metrics, but states "confirm that the DCD handpiece and Evercool handpiece can produce similar cooling results," implying the criterion was met.)
    Clinical TestingNot needed due to similarity to predicate devices.Based on the similarities of the device specifications, intended use, and indications for use between the Vbeam Prima Laser System and its predicate devices, no clinical studies were needed to support this 510(k) Premarket Notification. The safety and effectiveness are established by substantial equivalence to previously cleared devices.

    2. Sample Size Used for the Test Set and Data Provenance

    Since no clinical studies were conducted for the Vbeam Prima Laser System due to its substantial equivalence to predicate devices, there isn't a "test set" in the traditional sense for clinical performance.

    • Electrical Safety and EMC: Testing was conducted on the device itself (Epic Pro 940, implicitly the Vbeam Prima or a prototype). This involved laboratory testing, not human subjects or patient data.
    • Biocompatibility: Testing was on materials of the device, according to ISO 10993 standards. This is laboratory testing.
    • Software Verification & Validation: Testing was on the software of the device. This is internal software testing.
    • Bench Testing (Cooling Performance): Testing was on the DCD and Evercool handpieces. This is laboratory testing.

    Therefore, there is no human subject data provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable, as no clinical studies were performed. The "ground truth" for the device's safety and effectiveness was established by its compliance with recognized standards and its technological similarity to legally marketed predicate devices, which had previously demonstrated safety and effectiveness through their own studies and clinical use.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical studies were performed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done, as no clinical studies were performed. The Vbeam Prima Laser System is a physical medical device (laser system), not an AI algorithm that assists human readers.

    6. Standalone (Algorithm Only) Performance

    Not applicable, as the Vbeam Prima Laser System is a physical medical device, not an AI algorithm. Its performance is evaluated through engineering and biocompatibility testing, not AI algorithm standalone performance.

    7. Type of Ground Truth Used

    The "ground truth" for the Vbeam Prima Laser System's safety and effectiveness relies on:

    • Compliance with recognized international standards for electrical safety, electromagnetic compatibility, and biocompatibility.
    • Demonstrated equivalence of technological characteristics and intended use to predicate devices that have an established history of safe and effective use.

    8. Sample Size for the Training Set

    Not applicable, as this is a physical medical device, not an AI algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is a physical medical device, not an AI algorithm.