Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K233900

    Validate with FDA (Live)

    Date Cleared
    2024-09-06

    (270 days)

    Product Code
    Regulation Number
    870.1220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nordica PV Cryo Mapping Catheter is indicated for multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart.

    Device Description

    The Nordica PV Cryo Mapping Catheter is a 3F sterile, single use, multi-electrode, diagnostic catheter designed to map cardiac signals during ablation procedures. The catheter is provided with a fixed loop diameter of 15mm or 20mm with either 5 or 4 evenly spaced, radiopaque pairs of electrodes. The proximal end of the catheter contains an electrical connection for the EP electrical cable that integrates with electrophysiology lab recording systems. Once deployed in the vasculature through the central lumen of a catheter, a circular or spiral shape is established such that the electrodes contact the endocardial surface. This allows for recording, pacing, and interrogation of electrical conduction between the left atrium and the pulmonary veins.

    AI/ML Overview

    The provided text describes the submission for a 510(k) premarket notification for the "Nordica PV Cryo Mapping Catheter" and focuses on demonstrating substantial equivalence to a predicate device.

    Crucially, this document focuses on the safety and performance of a medical device (a catheter) through various engineering, sterilization, biocompatibility, and animal studies. It explicitly states: "This section is not applicable as the subject and predicate devices do not contain any software components or utilize software for their intended use." and "Human Clinical Performance Testing: Clinical testing was not required to demonstrate substantial equivalence to the predicate device."

    Therefore, the request for acceptance criteria and a study proving a device meets the acceptance criteria in the context of AI/software performance, human readers, ground truth, and training/test sets is not applicable to this document.

    The document details the following types of testing for the physical medical device:

    1. A table of acceptance criteria and the reported device performance:

    The document provides a "Table 1: Comparison of Technological Characteristics with the Predicate Device" which implicitly serves as a comparison of the subject device's features against the predicate's known characteristics, thereby establishing the "acceptance criteria" through equivalence. The "reported device performance" is then demonstrated through various specific tests.

    Feature / Acceptance Criteria CategoryPredicate Device (Implied Acceptance)Subject Device (Reported Performance)Rationale for Substantial Equivalence
    Product CodeDRFDRFSame Product Code as Predicate Device.
    Regulation21 CFR 870.122021 CFR 870.1220Same Regulation Number as Predicate.
    Indications for Use"multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The Achieve Advance mapping catheter is designed to obtain electrograms in the atrial regions of the heart.""multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart."Same Indications for Use as Predicate Device.
    Intended Use"to map intracardiac structures of the heart.""to map intracardiac structures of the heart."Same Intended Use as Predicate Device.
    SterilizationEthylene Oxide SterilizedEthylene Oxide Sterilized (Validated to ISO 11135:2014+A1:2019, SAL 10^-6 or better)Same Sterilization Method as Predicate Device (with updated standard compliance).
    Single UseYesYesSame Single Use as Predicate Device.
    BiocompatibilityComplies with ISO 10993-1 (as specified in K153139)Complies with ISO 10993-1 (2018) and FDA Guidance. Tested endpoints: ISO MEM Elution, ASTM Hemolysis, SC5b-9 Complement Activation Assay, ASTM Partial Thromboplastin Time (PTT), In-Vivo Thrombogenicity Evaluation [GLP Animal Study], ISO Guinea Pig Maximization Sensitization Test, ISO Intracutaneous Study in Rabbits, ISO Material Mediated Pyrogen Study (GLP), ISO Acute Systemic Toxicity Study in Mice. All met acceptance criteria.Same Biocompatibility compliance as Predicate Device - subject device complies to current revision of the standard.
    Performance (General Standard)Per ISO 10555-1Per ISO 10555-1:2013/AMD 1:2017 (for general catheter performance)Same Performance standards as Predicate Device - subject device complies to current revision of the standard.
    Performance Testing (Specific)Joint Strength, Corrosion Resistance, Radiopacity, Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Kink Resistance, Stiffness, Life Testing.Dimensional, Atraumatic Tip & Surface, Joint Strength / Peak Tensile Force, Corrosion Resistance, Radiopacity (verified in GLP animal study), Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Device Functionality Kink Resistance Stiffness Life Testing. All met design input acceptance criteria.Same as predicate device - testing for subject device performed in accordance with latest versions of standards. Requirements established for these tests were validated for their intended use through V&V testing.
    Outer Diameter3F (1mm)3F (1mm)Same Outer Diameter size as predicate device.
    Effective Length146 cm149 cmEquivalent effective length; demonstrated compatibility, safe and effective use through V&V studies.
    Distal End ShapeCircular LoopCircular LoopSame Distal End Shape as predicate device.
    Loop Diameter15mm & 20mm (K162892), 25 mm (K153139)15mm & 20mmSame Loop Diameters as predicate device.
    Number of Electrodes8 (15mm loop), 10 (25mm loop)8 (15mm loop), 10 (20mm loop)Same number of electrodes as predicate device. (Note: Predicate has 25mm loop option not present in subject device, but the 15mm and 20mm subject options match predicate's corresponding electrode counts).
    Electrode SpacingUnpaired, Spacing Between Electrodes: 4.1mm (15mm loop), 5.8mm (20mm loop), 5.8mm (25mm loop)Paired Electrodes, Spacing Between Pairs: 7.3 mm (15mm loop/8 electrodes), 7.3mm (20mm loop/10 electrodes); Spacing Between Electrodes: 2 mm (15mm and 20mm loops)Similar electrode spacing to predicate device. Differences due to paired electrodes; demonstrated safety and performance through V&V testing.
    Delivery MethodDelivered Through a Delivery Catheter: YesDelivered Through a Delivery Catheter: YesSame delivery method as predicate device.
    MaterialsCatheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel / Polyimide (Proximal). Loop Material: Nitinol insulated with PET (Pebax Covered). Electrode Material: Not Specified. Tip: Not Specified.Catheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel (Proximal). Loop Material: Nitinol covered with Pebax. Electrode Material: 90% Platinum / 10% Iridium. Tip: Loctite 4311 Medical Grade Adhesive.Subject and predicate devices constructed of equivalent materials. Differences qualified through biological safety and V&V testing.
    Sterilization Residuals(Implicitly meets standards based on predicate's clearance)Met acceptance criteria below limits specified in EN ISO 10993-7:2008 + A1:2022.Compliance with updated standard.
    Shelf-Life(Implicitly demonstrated for predicate)Accelerated aging per ASTM F1980-21. Package Integrity (label, bubble leak ASTM F2096-11(2019), seal visual, seal strength ASTM F88/F88M-21) and Product V&V performed after aging.Demonstrated through testing.
    Electrical Safety / EMC(Implied compliance for predicate)Complies with IEC 60601-1-2 Ed. 4.1 (2020-09) for Class A Emissions; IEC 60601-1:2005 (and amendments); IEC 60601-1-6:2010 (and amendments) for usability.Compliance with updated standards.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test set for Biocompatibility: "All tested subject devices met the acceptance criteria for each biological endpoint test". The exact number of devices tested for each biological endpoint (e.g., ISO MEM Elution, Hemolysis, etc.) is not specified, but it implies a sufficient sample size was used for each test to provide robust results.
    • Test set for Benchtop Studies: "The results of benchtop testing demonstrated the design outputs of the subject device met the design input acceptance criteria required to demonstrate substantial equivalence with the predicate device." Similarly, exact sample sizes for each benchtop test (e.g., buckle force, catheter deflection, etc.) are not provided.
    • Pre-Clinical GLP Animal Study: A "chronic GLP animal study" was performed on a "canine animal model". The exact sample size (number of animals) is not specified.
    • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective, beyond stating it was a "chronic GLP animal study" which implies a prospective, controlled study in a laboratory setting. All testing appears to be internal (Synaptic Medical Corporation) or contracted to ISO/ASTM compliant labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the studies are engineering, biological, and animal model evaluations, not human-read AI evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device has no software components or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the various tests, the "ground truth" is defined by the established scientific and engineering standards and methods:

    • Biocompatibility: Defined by the acceptance criteria of ISO 10993 series standards (e.g., no cytotoxicity, no hemolysis, no sensitization, no pyrogenicity, etc.).
    • Sterilization: Defined by ISO 11135:2014+A1:2019 (achieving SAL of 10^-6 or better) and EN ISO 10993-7:2008 + A1:2022 (residual limits).
    • Shelf-Life: Defined by ASTM F1980-21 for accelerated aging and specific packaging integrity tests (ASTM F2096-11(2019), ASTM F88/F88M-21).
    • Benchtop Performance: Defined by ISO 10555-1:2013/AMD 1:2017 and internal "design input acceptance criteria" for various physical properties (e.g., buckle force, tensile strength, electrical continuity).
    • Pre-Clinical GLP Animal Study: Evaluation of "chronic safety, performance, and usability" in a canine model, presumably against pre-defined physiological endpoints and observational criteria.

    8. The sample size for the training set

    Not applicable, as this device does not involve AI/machine learning.

    9. How the ground truth for the training set was established

    Not applicable, as this device does not involve AI/machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240246

    Validate with FDA (Live)

    Date Cleared
    2024-07-26

    (178 days)

    Product Code
    Regulation Number
    870.1390
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K233708

    Validate with FDA (Live)

    Date Cleared
    2024-02-16

    (88 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit is intended to introduce various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

    Device Description

    The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit consists of a 1) Steerable Sheath, 2) Dilator and 3) Guidewire, and is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. When the left side of the heart is accessed, the device is used with a compatible Transseptal Needle to puncture the interatrial septum for transseptal catheterization procedures. The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit is a sterile, single use device, with asymmetrical bi-directional steerable introducer with usable length of 65cm. The Steerable Sheath is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the Steerable Sheath tip clockwise ≥ 180° and counterclockwise ≥ 90°. The steerable introducer features distal vent holes to facilitate aspiration and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    The provided text describes the NaviGo 12F Steerable Intracardiac Catheter Introducer Kit and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document describes various performance tests conducted, but it does not explicitly list quantitative acceptance criteria for each test in a table format with corresponding reported performance values. Instead, it generally states that the device "met the acceptance criteria" or "demonstrated robust biocompatibility."

    However, based on the descriptions, we can infer some general criteria and outcomes:

    Performance Test CategoryAcceptance CriteriaReported Device Performance
    SterilizationSuccessful validation per ISO 11135:2014+A1:2019. Bioburden testing per ISO 11737-1. LAL testing per ANSI/AAMI ST72:2019 and FDA Guidance. EO and ECH residuals below limits per EN ISO 10993-7:2008+A1:2022.Sterilization process developed and validated. Bioburden testing performed. LAL testing performed. EO and ECH residual testing demonstrated results below specified limits.
    Shelf-LifePackage integrity (label, bubble leak, seal visual, seal strength) maintained after accelerated aging (ASTM F1980-21). Product V&V (safety and performance) confirmed after accelerated aging.Shelf-life testing performed via accelerated aging. Package integrity testing (label, bubble leak, seal visual, seal strength) performed and confirmed. Product V&V performed and confirmed. (Note: 6-month shelf-life initially, with potential for extension).
    BiocompatibilityCompliance with ISO 10993-1:2018 for "Externally Communicating Device, Circulating Blood with Limited Exposure" (e.g., passing results for cytotoxicity, hemolysis, complement activation, thrombogenicity, sensitization, irritation, pyrogenicity, acute systemic toxicity).All tested subject devices met the acceptance criteria for each biological endpoint test and demonstrated robust biocompatibility in accordance with ISO 10993-1.
    Benchtop StudiesDesign outputs met design input acceptance criteria for each specific test (Visual & Dimensional, Buckle Force, Catheter Deflection, Shaft Bending & Kink Resistance, Simulated Use & Compatibility, Freedom from Leakage, Peak Tensile Force & Bond Strength).Results of benchtop testing demonstrated the design outputs of the subject device met the design input acceptance criteria.
    Pre-Clinical (GLP Animal Study)Primary Objective: Overall safety demonstrated. Secondary Objective: Performance and usability assessed with "Acceptable" and "Above Average" ratings. Thrombogenicity assessment completed.Overall safety demonstrated. Performance and usability evaluation met acceptance criteria with "predominant rating of Acceptable and Above Average." Thrombogenicity assessment completed.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sterilization Testing: Not explicitly stated, but bioburden testing, LAL testing, and EO/ECH residual testing were performed on "the subject device."
    • Shelf-Life Testing: Not explicitly stated, but accelerated aging studies and various package integrity tests were performed on "the subject device."
    • Biocompatibility Testing: Not explicitly stated, but various in-vitro and in-vivo tests were conducted. The "In-Vivo Thrombogenicity Evaluation [GLP Animal Study]" refers to the animal study below. Materials were from the device itself.
    • Benchtop Studies: Not explicitly stated, but various devices would have been tested for each category.
    • Pre-Clinical (GLP Animal Study):
      • Sample Size: Six (6) healthy canines.
      • Data Provenance: Prospective animal study conducted in a GLP (Good Laboratory Practice) environment. The location/country of the GLP lab is not specified but it's an animal model, not human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Pre-Clinical (GLP Animal Study):
      • Number of Experts: Two (2)
      • Qualifications: "Board Certified Electrophysiologists specializing in ablation of pulmonary veins in the management of atrial fibrillation."

    4. Adjudication Method for the Test Set

    • For the Pre-Clinical (GLP Animal Study), the two electrophysiologists "performed the procedures and rated the overall performance and usability of the subject device." The document does not specify a formal adjudication method (e.g., 2+1, 3+1). It states "predominant rating of Acceptable and Above Average," suggesting a consensus or combined assessment was reached, but the process is not detailed.
    • For other tests (benchtop, sterilization, biocompatibility), the ground truth is based on standard test methods and established scientific criteria, not expert adjudication in the same sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This device is a physical medical device (catheter introducer kit), not an AI/software device that would typically involve human readers interpreting images with or without AI assistance. The study described is a pre-clinical animal study to assess safety, performance, and usability of the physical device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    No, a standalone algorithm performance study was not done. This is not an AI/software device.

    7. The Type of Ground Truth Used

    • Sterilization: Regulatory standards (ISO, ANSI/AAMI, EN ISO) for sterilization efficacy, bioburden levels, endotoxin levels, and residual limits.
    • Shelf-Life: Regulatory standards (ASTM F1980-21, ASTM F2096-11(2019), ASTM F88/F88M-21) and product V&V testing.
    • Biocompatibility: Regulatory standards (ISO 10993-1:2018, ISO 10993-4, -5, -7, -10, -11, -23) and FDA guidance for biological evaluation endpoints.
    • Benchtop Studies: Design input acceptance criteria based on engineering specifications and industry standards relevant to catheter function (e.g., buckle force, deflection, kink resistance, leakage, bond strength).
    • Pre-Clinical (GLP Animal Study): Direct observation, clinical assessment, and expert rating by Board Certified Electrophysiologists in a canine model for safety, performance, and usability. This includes observations for adverse events, device maneuverability, and overall handling, as well as thrombogenicity assessment.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device. The "training set" concept is typically relevant to AI/machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203793

    Validate with FDA (Live)

    Date Cleared
    2022-02-14

    (413 days)

    Product Code
    Regulation Number
    870.1220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RithmID-SD Steerable Diagnostic Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

    Device Description

    The RithmID-SD Steerable Diagnostic Electrophysiology Catheter are biocompatible, flexible, radiopaque electrophysiology catheters that are available in a variety of diameters, lengths, curve shapes, and electrode number and spacing configurations, with a high-torque shaft with an array of platinum iridium alloy electrodes at the distal tip that can be used for recording electrical signals. The catheter is designed to facilitate the electrophysiological mapping of the heart.

    The RithmID-SD Steerable Diagnostic Electrophysiology Catheter consist of a handle, a shaft and a steerable diagnostic tip. The catheter is introduced through the sheath and into the femoral vein, from the inferior vena cava into the heart and coronary sinus. The RithmID-SD Steerable Diagnostic Electrophysiology Catheter are available in 6F in various curves, including B, D, Y, and R curves.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "RithmID-SD Steerable Diagnostic Electrophysiology Catheter". This document focuses on demonstrating substantial equivalence to predicate devices, primarily through bench testing, biocompatibility testing, and comparisons of technical characteristics.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list acceptance criteria for each test in a separate column. Instead, for most tests, it states "All test samples met the acceptance criteria," indicating that the criteria were satisfied. The performance results are qualitative; for a few tests, comparative statements are made relative to the reference device.

    Here's a compilation of the tests and their reported performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Corrosion ResistanceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Particulate Matter EvaluationNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria and is equivalent to the Reference Device.
    Radiopacity DetectabilityNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Signal AcquisitionNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    UsabilityNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria. The performance of the Subject Device is better or equivalent to the Reference Device.
    Electrical SafetyNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Visual Inspection (Bench Testing)Not explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Dimensional VerificationNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Simulated UseNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Connection Plug ForceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Tip Fatigue ToleranceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Shaft Fatigue ToleranceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Flexural Fatigue ToleranceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Torsional ForceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Electrode Conductor ResistanceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria.
    Buckling ForceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria. The buckling force of the Subject Device is less than or equal to the Reference Device.
    Peak Tensile ForceNot explicitly stated, implied "met criteria"All test samples met the acceptance criteria. The peak tensile force of the Subject Device was compared to the Reference Device.
    Torque StrengthData comparison, not a pass/fail criterionThe average number of 360° rotations when permanent mechanical deformation occurs was recorded for the Subject Device and Reference Device. The torque strength of the Subject device is compared to the Reference Device for reference only.
    Packaging Integrity (Distribution Sim.)ASTM D4169-16All test samples met the acceptance criteria per ASTM D4169-16.
    Packaging Integrity (Visual Insp.)ASTM F1886/F1886M-16All test samples met the acceptance criteria per ASTM F1886/F1886M-16.
    Packaging Integrity (Bubble Leak)ASTM F2096-11All test samples met the acceptance criteria per ASTM F2096-11.
    Packaging Integrity (Tray Seal Strength)F88/F88M-15All test samples met the acceptance criteria per F88/F88M-15.
    CytotoxicityNon-cytotoxic (ISO 10993-5)Non-cytotoxic
    SensitizationNon-sensitizer (ISO 10993-10)Non-sensitizer
    Intracutaneous ReactivityNon-irritant (ISO 10993-10)Non-irritant
    Acute Systemic ToxicityNon-Toxic (ISO 10993-11)Non-Toxic
    Systemic Toxicity (Pyrogenicity)Non-Pyrogenic (ISO 10993-11)Non-Pyrogenic
    Hemocompatibility (Hemolysis)Non-Hemolytic (ISO 10993-4)Non-Hemolytic
    Hemocompatibility (Thrombogenicity)Passed (ISO 10993-4)Passed
    Hemocompatibility (Complement Act.)Non-Activator (ISO 10993-4)Non-Activator
    Hemocompatibility (PTT)Passed (ISO 10993-4)Passed
    Shelf Life (Accelerated Aging)Meets design specification at 1 yearAll acceptance criteria met; device performs as intended to Design Specification.
    Sterilization (SAL)SAL of 10^-6Achieved a SAL of 10^-6.
    EO and ECH ResidualsBelow limits specified in ISO 10993-7Below the limits specified in ISO 10993-7.
    Bacterial Endotoxin Levels<20.0 EU/Device<20.0 EU/Device

    2. Sample Size Used for the Test Set and Data Provenance

    The document consistently states "All test samples met the acceptance criteria" for various bench tests and indicates "All test samples" for packaging and sterilization validations. However, it does not specify the exact numerical sample size (e.g., "n=5" or "n=10") for any of these tests.

    Regarding data provenance:

    • Country of Origin: Not specified in the provided text.
    • Retrospective or Prospective: The studies are described as "Performance Bench Testing and Animal Testing," "Biocompatibility Testing," "Accelerated Shelf-Life Testing," "Packaging validation," and "Sterilization validation." These are all standard types of prospective pre-market validation studies conducted specifically for regulatory submission, rather than retrospective data analysis.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable for this device. This document describes the 510(k) submission for an electrophysiology catheter, not an AI/ML-driven diagnostic device. The performance data presented are primarily from physical, chemical, and biological bench tests and animal studies, not clinical studies involving human observers or interpretative tasks where "ground truth" would be established by experts like radiologists.

    4. Adjudication Method for the Test Set

    • Not applicable. As stated above, this is not a study involving human interpretation or subjective assessments that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC study was not done. MRMC studies are typically conducted for imaging devices or AI algorithms where human readers' performance (with or without AI assistance) is evaluated. The RithmID-SD is a physical diagnostic catheter. The "Usability" test mentioned is likely a bench test or simulated use to assess the device's functional handling, not a human reader study.

    6. Standalone (Algorithm Only) Performance

    • Not applicable. This device is a physical catheter, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the device's performance is established by:

    • Engineering specifications and standards: For most bench tests (e.g., dimensional verification, electrical safety, fatigue tolerance, tensile force, buckling force), the "acceptance criteria" are derived from the device's design specifications and often relevant international standards for medical devices (though the specific standards beyond ISO 10993 for biocompatibility and ASTM for packaging aren't explicitly listed for every bench test).
    • Biocompatibility standards: ISO 10993 series for the biocompatibility tests.
    • Sterilization standards: For bacterial endotoxin levels and residuals (ISO 10993-7).
    • Animal testing: The document mentions "Animal Testing" and specifically for hemocompatibility, "Thrombogenicity in Canine (GLP)" was performed. This forms part of the "ground truth" for biocompatibility.
    • Physical measurements and objective criteria: For tests like torque strength, specific physical properties are measured.

    8. Sample Size for the Training Set

    • Not applicable. This is a physical device, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1