Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K173272

    Validate with FDA (Live)

    Device Name
    Ergonomic Handle
    Date Cleared
    2018-02-09

    (120 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Laparoscopic surgical manual instruments are designed to be used endoscopically through cannula to perform cutting, grasping, dissecting, retracting, and manipulating functions. The Ergonomic handle is intended to be used with 5mm inserts available from K&W and previously cleared under K973259.

    Device Description

    Symmetry® laparoscopic surgical manual instruments consist of a handle and a 5mm shaft, which hold various 5 mm inserts cleared under K973259. The shaft comes in two lengths: 33 cm and 45 cm. One handle can be used in combination with a variety of different 5 mm inserts. The handles have monopolar posts which allow them to be used for electrosurgery.

    The ergonomic laparoscopic handles have an adjustable grip. They may be used in one of the three handle positions: pistol grip, hybrid, and inline. To change the handle position, retract the position adjustment button and lock the grip in the chosen position. The handle comes in two sizes.

    AI/ML Overview

    This document describes the Ergonomic Handle (K173272), which is an accessory to laparoscopic surgical manual instruments. The device allows for adjustable grip angles to improve surgeon comfort. The information provided is for regulatory clearance (510(k)) and focuses on demonstrating substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and supporting studies, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-11 (for External Communicating Devices: Tissue/Bone/Dentin with limited contact duration <24 hours)Cytotoxicity, Intracutaneous Reactivity, Sensitization, and Acute Systemic Toxicity tested. All acceptance criteria of the standards were met through testing or justified through risk assessment.
    Sterilization ValidationISO 17665-1 (for Pre-vacuum steam sterilization cycles of 132°C for 4 minutes and 134°C for 3 minutes)Testing was done to validate common cycles to achieve a Sterility Assurance Level of 10⁻⁶.
    Electrical SafetyANSI/AAMI ES60601-1 (General Requirements for Basic Safety and Essential Performance) and IEC 60601-2-2 (Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories)The proposed devices comply with these standards for safety.
    Mechanical TestingNot explicitly stated, but included testing for shaft deflection and electrical safety after simulated use.Reliability testing was conducted, including shaft deflection and electrical safety after simulated use. (Implied to have met performance needs for substantial equivalence).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify sample sizes for any of the testing conducted (biocompatibility, sterilization, electrical safety, mechanical testing).

    The data provenance is not explicitly stated beyond the testing being performed in support of the substantial equivalence determination for the device. Given the context of a 510(k) submission, this typically refers to testing conducted by the manufacturer or a contracted lab under their direction, likely within the US, but this is an inference. It is retrospective in the sense that the testing was performed on completed device designs to demonstrate compliance, rather than in a prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes pre-market testing for a medical device (laparoscopic handle). These types of tests (biocompatibility, sterilization, electrical safety, mechanical) do not typically involve human experts establishing "ground truth" in the way it's understood for AI/diagnostic studies. The "ground truth" here is defined by compliance with established scientific standards and regulatory requirements. The experts involved would be the testers and engineers conducting the studies, and potentially external auditors or consultants specializing in those regulatory standards. Their qualifications would be in their respective fields (e.g., microbiology for sterilization, toxicology for biocompatibility, electrical engineering for safety, mechanical engineering for mechanical testing). No specific number or qualifications of such experts are provided in this document.

    4. Adjudication Method for the Test Set

    As mentioned above, the tests are primarily objective measurements against established standards, not subjective assessments requiring adjudication. Therefore, no adjudication method like 2+1 or 3+1 is applicable or mentioned for the described testing. The assessments are based on whether predefined metrics (e.g., bacterial kill rate, material toxicity, electrical leakage) meet specified limits.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done or is mentioned. This type of study is typically performed for diagnostic or AI-assisted interpretation devices to compare human performance with and without AI. This device is a surgical instrument handle, not an interpretive medical device.

    6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done

    N/A. This device is a manual surgical instrument handle, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance assessment of this ergonomic handle is defined by:

    • Compliance with recognized international and national standards: ISO 10993 series (biocompatibility), ISO 17665-1 (sterilization), ANSI/AAMI ES60601-1 and IEC 60601-2-2 (electrical safety).
    • Engineering specifications and test protocols: for mechanical reliability (shaft deflection, electrical safety after simulated use).
    • Demonstration of substantial equivalence to a predicate device (K150127) through comparing technological characteristics and ensuring no new issues of safety or effectiveness are raised.

    8. The Sample Size for the Training Set

    N/A. This device is a physical surgical instrument, not an AI/machine learning system, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    N/A. Since there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161744

    Validate with FDA (Live)

    Date Cleared
    2016-09-07

    (75 days)

    Product Code
    Regulation Number
    882.4840
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symmetry Sharp Kerrison® Rongeur is indicated for cutting bony tissue of the spine. The Symmetry Sharp Kerrison® Rongeur is a manually operated instrument that consists of a Handle and Tips (available in single use or reusable).

    Device Description

    The Symmetry Sharp Kerrison Rongeur is designed to have detachable tips that are disposable. The detachable tips are provided as either sterile single use or as nonsterile reusable tips. The reusable tips must be cleaned and sterilized before use. The sterile, disposable tips are available in 4 cup sizes: (1, 2, 3, and 4 mm); which are identifiable by etchings on each device. The Reusable Tips for the Symmetry Sharp Kerrison® Rongeur Handle are available in 3 cup sizes: (2, 3, and 4 mm); which are identifiable by etchings on each device. The reusable tips have a mating code etched on each pair to ensure that only matched pairs are used. Each tip is designed with a shaft key that fits into its mating slot on the detachable instrument handle. The tip has cups, one within the stationary footplate and the other within the upper sliding shaft, performing the cutting of the bony tissue in the spine area. The various cup sizes within the working tips provide the surgeon various cutting configurations needed for the different characteristics of the bony tissue within the spine. The Symmetry Sharp Kerrison® Rongeur Handle operates with a cup and ball assembly for smooth motion of the footplate when the handle is squeezed. The detachable, reusable instrument handles of this device are provided non-sterile and must be cleaned and sterilized prior to use. The instrument handles are available in three shaft lengths: (5, 6, and 7 inches).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text, focusing on the device "Symmetry Sharp Kerrison® Rongeur":

    This document describes a medical device called the "Symmetry Sharp Kerrison® Rongeur," which is a manual rongeur used for cutting bony tissue of the spine. The information provided heavily emphasizes comparison to a predicate device (K130541 Kerrison Disposable Tip Rongeur) to establish substantial equivalence for FDA 510(k) clearance. The testing described primarily focuses on validating the safety and performance characteristics for a mechanical surgical instrument, not an AI/ML-driven device. Therefore, many standard questions related to AI/ML device studies (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance) are not applicable or directly addressed in this type of submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestAcceptance CriteriaReported Device Performance (Results)
    Cleaning Validation for Reusable ComponentsResidual proteins, residual hemoglobin, and residual carbohydrates must meet specified acceptance criteria."For substantial equivalence the device must be able to be adequately cleaned for reuse. Pass" (Indicates criteria were met)
    Steam Sterilization Validation for Reusable ComponentsMust meet requirements in applicable ISO Standard for the overkill method to achieve an SAL (Sterility Assurance Level) of 10-6."For substantial equivalence the reusable components must be able to be effectively sterilized with steam sterilization. Pass" (Indicates criteria were met)
    Gamma Sterilization Validation for Single-Use ComponentsMust meet requirements in applicable ISO standard for VDMax to achieve an SAL of 10-6."For substantial equivalence the single use component must be able to be sterilized using gamma sterilization. Pass" (Indicates criteria were met)
    Confirmation of Multiple Use for Reusable Components (Durability)Detachable tips must maintain acceptable cutting performance for up to 100 cuts. Reusable handle must maintain performance for 50 reuse cycles."For substantial equivalence the reusable must be sufficiently robust to support reuse. Pass" (Indicates criteria were met for both tips and handle as per the criteria)
    Resistance to Corrosion for Reusable ComponentsMust not corrode when exposed to a corrosion challenge."For Substantial Equivalence device must be resistant to corrosion. Pass" (Indicates criteria were met)
    Packaging Validation/Shelf Life for Single-Use ComponentsPackage and product must demonstrate package and product stability over the claimed shelf life period."Package and product performance evaluations conducted on the single use component for shelf life confirm a two year shelf life. Pass" (Indicates a 2-year shelf life was confirmed, meeting stability criteria)

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test (e.g., number of rongeurs tested for cleaning, sterilization, or durability). It refers to standard validation activities but doesn't quantify the number of units or cycles performed for each test.

    • Data Provenance: The studies are described as "Design verification and validation testing" which are prospective tests conducted by the manufacturer specifically for this submission. The origin (e.g., country) is not specified beyond the manufacturer's location (Antioch, TN, USA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided. The "ground truth" for a mechanical device like a rongeur is based on objective performance metrics (e.g., sterilization efficacy, cutting performance, corrosion resistance) measured against defined standards or engineering specifications, not expert consensus on interpretations.

    4. Adjudication Method for the Test Set

    Not applicable. The tests are objective measurements against defined standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a mechanical surgical instrument, not an AI/ML diagnostic or assistive device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a mechanical surgical instrument.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the performance tests relies on:

    • Established Standards: Primarily ISO standards for sterilization (SAL of 10-6), ASTM standards for shelf life.
    • Engineering Specifications: Criteria for residual proteins/hemoglobin/carbohydrates, number of cuts/reuse cycles for durability, and absence of corrosion. These are objective, measurable parameters.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152111

    Validate with FDA (Live)

    Date Cleared
    2016-04-14

    (260 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symmetry Surgical Quad-Lock® Container System is indicated for use by hospitals and by health care facilities to: · organize and protect stainless steel, aluminum, plastic* and silicone surgical instruments that will be sterilized · allow sterilization of the contained instruments by prevacuum steam, ethylene oxide, and low temperature hydrogen peroxide sterilization cycles (validated modalities and parameters are shown in Table 1) · When sterilized by prevacuum steam sterilization, the Quad-Lock® Containers maintain the sterility of the contents for up to 180 days during storage and transport within the healthcare facility, as long as the integrity of the container has not been compromised · When sterilized by ethylene oxide or low temperature hydrogen peroxide cycle types the Quad-Lock® Containers maintain the sterility of the contents when sterilized by any of the above for up to 30 days during storage and transport within the healthcare facility, as long as the integrity of the container has not been compromised The system has not been tested for maintenance of sterility after transportation outside the health care facility. The system is intended for use with stainless steel, aluminum, plastic and silicone surgical instruments. *The use of plastic instruments has only been confirmed for steam sterilization. For effective sterilization and drying of any size Symmetry Surgical Quad-Lock® Sterilization Container, the recommended maximum combined weight of the single container, lid. basket/tray contents is defined in Table 1. For each cvcle type the qualified lumen size is the minimum length that has been qualified for use are shown in Table 1. Container System Compatibility by Cycle Type details are shown in Table 2.

    Device Description

    The Symmetry Surgical Quad-Lock® Container System provides a means to organize and protect stainless steel, aluminum, titanium, plastic and silicone surgical instruments during sterilization and storage. The filter lids, security lids, and container bottoms are made from anodized aluminum. The security lids help ensure sterility during storage and protect the container lid when the containers are stacked. The Container System includes the container base, filter lid, security lid, filters and other accessories.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study to demonstrate the performance of the Symmetry Surgical Quad-Lock™ Container System.

    Here's an organized breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Sterilization EfficacyAchieve a sterilization assurance level (SAL) of 10-6 for prevacuum steam, ethylene oxide (EtO), and low-temperature hydrogen peroxide (H2O2) sterilization cycles.Prevacuum Steam: - Temperature: 132°C - Exposure Time: 4 minutes - Drying Time: 30 minutes - Stackable: Yes, up to three containers - Qualified Lumen Size (Full/Three Quarter): 3mm x 400mm - Qualified Lumen Size (Half Size): 3mm x 250mm - Maximum Quantity: 4 lumen per container - Load Requirements: 25 lbs (all sizes) Ethylene Oxide: - Concentration: 725mg/L - Temperature: 55°C - Exposure Time: 60 minutes - Humidity: 50-70% - Aeration Time: 8 hours at 110°F (43°C) - Stackable: No - Qualified Lumen Size (Full/Three Quarter): 3mm x 400mm - Qualified Lumen Size (Half Size): 2mm x 250mm - Maximum Quantity: 4 Lumen per container - Load Requirements: 25 lbs (all sizes) Low Temperature H2O2 Sterilizers: (Parameters are set by sterilizer manufacturer) - STERRAD 100NX Standard/DUO: Fixed Cycle Type, Stackable: No - Qualified Lumen Size (Full Size): 2mm x 400mm - Qualified Lumen Size (Three Quarter/Half Sizes): 2mm x 250mm - Maximum Quantity: 5 Lumen per container - Load Requirements (Full): 21.4 lbs - Load Requirements (Three Quarter/Half): 13.85 lbs (DUO: 13.2 lbs) - STERRAD 100NX FLEX: Fixed Cycle Type, Stackable: No - Qualified Lumen Size (All Sizes): One 1mm x 850mm Flexible Endoscope per container - Load Requirements (Full): 16.65 lbs - Load Requirements (Three Quarter): 12.40 lbs - Load Requirements (Half): 10.75 lbs - STERRAD NX Standard/Advanced: Fixed Cycle Type, Stackable: No - Qualified Lumen Size (Full Size): 1mm x 400mm (Standard) / 2mm x 250mm (Advanced) - Qualified Lumen Size (Three Quarter/Half Sizes): 2mm x 250mm - Maximum Quantity: 5 Lumen per container - Load Requirements (Full): 21.4 lbs - Load Requirements (Three Quarter/Half): 13.85 lbs - STERRAD 100S Standard: Fixed Cycle Type, Stackable: No - Qualified Lumen Size (Full Size): 3mm x 400mm - Qualified Lumen Size (Three Quarter/Half Sizes): 2mm x 250mm - Maximum Quantity: 5 Lumen per container - Load Requirements (All Sizes): 18.2 lbs - STERIS V-PRO™ 1 PLUS Lumen/No Lumen: Cycle time = 55 minutes, Stackable: No - Qualified Lumen Size (Full Size): 3mm x 400mm - Qualified Lumen Size (Three Quarter/Half Sizes): 2mm x 250mm - Maximum Quantity: 10 lumen per container - Load Requirements (All Sizes): 19.65 lbs All sterilization efficacy and simulated use testing to achieve SAL of 10-6 passed.
    Maintenance of Sterility (Shelf Life)Maintain sterility for up to 180 days with prevacuum steam sterilization. Maintain sterility for up to 30 days with ethylene oxide and low-temperature hydrogen peroxide sterilization cycles.Prevacuum Steam: Sterility maintained for up to 180 days during storage and transport within the healthcare facility, as long as container integrity is not compromised. Ethylene Oxide & Low Temperature H2O2: Sterility maintained for up to 30 days during storage and transport within the healthcare facility, as long as container integrity is not compromised. The shelf life testing for 30 days of storage passed.
    Reuse TestingDemonstrate microbial challenge following multiple sterilization cycles for up to 100 sterilization cycles.Microbial Challenge following multiple sterilization cycles verified for up to 100 sterilization cycles passed.
    BiocompatibilityIn vitro evaluations for biocompatibility, extractables, and leachables must have acceptable results.In vitro evaluations for these tests had acceptable results.
    Functional Strength of HandlesTesting performed in accordance with ST77 (presumably ANSI/AAMI ST77, a standard for sterile packaging).Testing performed in accordance with ST77 had acceptable results.

    2. Sample size used for the test set and the data provenance:
    The document does not specify numerical sample sizes for the test set for its performance studies (sterilization efficacy, shelf life, reuse testing, biocompatibility, or functional strength).
    The data provenance is not explicitly mentioned as a country of origin or whether it's retrospective or prospective; however, the submission is from "Symmetry Surgical, Inc." which is based in Antioch, TN, USA, suggesting the studies likely occurred in the US. The studies are described as "performance testing" and "non-clinical tests," implying they were conducted specifically for this submission, making them prospective in nature for this application.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document. The studies are described as laboratory-based performance testing rather than studies requiring expert interpretation of results for ground truth establishment (e.g., medical image analysis).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    This information is not applicable and therefore not provided in the document. The studies described are objective performance tests (e.g., sterilization assurance levels, material properties) that do not typically involve human adjudication in the way medical diagnostic studies do.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable and therefore not provided in the document. The device is a sterilization container system, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    This information is not applicable and therefore not provided in the document. The device is a physical medical device (sterilization container), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The ground truth for the performance studies appears to be based on objective, measurable criteria derived from scientific and engineering principles for sterilization, material integrity, and functionality. For instance, "sterilization efficacy" ground truth is defined by achieving a specified Sterilization Assurance Level (SAL of 10-6), which is typically confirmed through validated microbiological testing (e.g., biological indicators). Shelf life is determined by maintaining sterility over time, also verified through microbiological methods. Biocompatibility relies on established in vitro evaluation standards.

    8. The sample size for the training set:
    This information is not applicable and therefore not provided in the document. The product is a physical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established:
    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1