Found 6 results

510(k) Data Aggregation

    K Number
    K193563

    Validate with FDA (Live)

    Date Cleared
    2020-08-19

    (240 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Harmonic ACE+7 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gyneoologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter using the Advanced Hemostasis hand control button.

    Device Description

    The Medline ReNewal Reprocessed Harmonic ACE+7 Shears (models HARH23, HARH36, and HARH45) are used for coagulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices allow the surgeon to grasp, coagulate, and transect soft tissue with a single instrument. The devices are hand-actuated with a shaft and tissue effector that can be rotated. The energy delivery can be activated with hand activation or with an optional generator foot switch.

    AI/ML Overview

    This document is a 510(k) summary for the Medline ReNewal Reprocessed Harmonic ACE+7 Shears, which is a reprocessed medical device. It aims to demonstrate substantial equivalence to a legally marketed predicate device.

    Acceptance Criteria and Device Performance (Based on the document):

    The document primarily focuses on demonstrating substantial equivalence to the predicate device (Ethicon Harmonic ACE+7 Shears, K132612) through comparisons of various characteristics and performance testing. The "acceptance criteria" here implicitly refer to meeting the performance and safety profiles of the original device after reprocessing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines performance testing conducted to ensure equivalency after reprocessing. The acceptance criteria are implied to be achieving results comparable to or within acceptable ranges of the original, new device.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Summary from "Performance Testing" section)
    Electrical Safety & EMCCompliance with IEC 60601-1 and 60601-1-2"Evaluated and found to be equivalent to the predicate devices in accordance with IEC 60601-1 and 60601-1-2"
    Simulated UseEquivalent performance during simulated surgical use"Evaluated and found to be equivalent to the predicate devices"
    Device IntegrityMaintenance of structural and functional integrity"Evaluated and found to be equivalent to the predicate devices"
    Handle OperationProper and equivalent handle functionality"Evaluated and found to be equivalent to the predicate devices"
    Shaft Knob RotationSmooth and effective shaft rotation"Evaluated and found to be equivalent to the predicate devices"
    Device RecognitionProper recognition by the associated generator (Ethicon GEN11)"Evaluated and found to be equivalent to the predicate devices" (Specifically, the ReNewal Key-integrated flex circuitry enables additional use, implying successful recognition for subsequent use).
    Cutting TimeComparable cutting efficiency/speed"Evaluated and found to be equivalent to the predicate devices"
    Thermal Analysis CharacterizationComparable thermal profiles to minimize thermal injury"Evaluated and found to be equivalent to the predicate devices"
    Tissue StickingComparable minimal tissue sticking properties"Evaluated and found to be equivalent to the predicate devices"
    Burst PressureMaintenance of vessel sealing integrity/strength"Evaluated and found to be equivalent to the predicate devices"
    HistopathologyComparable tissue effects related to coagulation and transection"Evaluated and found to be equivalent to the predicate devices"
    Seal QualityEffective and consistent tissue sealing"Evaluated and found to be equivalent to the predicate devices"
    Cleaning (Protein, Carbohydrates, Endotoxin)Demonstration of effective decontamination/sterilization from biological residueAll tests passed; specific levels not provided but implied to meet acceptable limits for reprocessing.
    Biocompatibility (Cytotoxicity, Sensitization, Irritation, Materials-mediated Pyrogen, Acute Systemic Toxicity)No unacceptable biological reactions from reprocessed materialsAll tests passed; implied to meet acceptable standards for patient contact.
    Performance QualificationOverall device performance consistently meeting specifications"Evaluated and found to be equivalent to the predicate devices"
    Sterilization ValidationDevice can be effectively sterilized for patient safetyPassed and achieved required sterility assurance levels.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document broadly states that the functional characteristics were "evaluated." It does not specify the exact sample sizes (N) for each of the performance tests listed. The data provenance is implied to be from prospective testing conducted by Medline ReNewal as part of their 510(k) submission process for the reprocessed device. There is no indication of country of origin for the testing, other than the submitting company being based in the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This document describes technical and functional performance testing of a physical medical device, not an AI/algorithm-based diagnostic device. Therefore, the concept of "experts used to establish ground truth" (e.g., radiologists, pathologists) for interpreting test results is not applicable in the same way as for diagnostic AI. The "ground truth" for the device's performance is established by the direct measurements and results of the physical and biological tests against established engineering and safety standards. The individuals conducting and analyzing these tests would be qualified engineers, technicians, and potentially toxicologists/biologists. The document doesn't specify how many or their exact qualifications.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this device's performance testing. Adjudication methods like 2+1 or 3+1 are typical for human interpretation of medical images or data, not for direct physical performance measurements. The results of the tests would be objectively measured and compared to predefined criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for evaluating the impact of an AI system on human reader performance in diagnostic tasks, which is not the subject of this 510(k) submission for a reprocessed surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical surgical instrument (Harmonic ACE+7 Shears), not an AI algorithm. Its performance is inherent to the reprocessed device itself.

    7. The type of ground truth used:

    The "ground truth" for this reprocessed device is established through various physical, electrical, mechanical, and biological measurements against established engineering specifications and safety standards for the original device and similar devices. This includes:

    • Direct measurements of cutting time, burst pressure, and other functional parameters.
    • Results from standardized cleaning and sterilization validation protocols.
    • Outcomes of biocompatibility testing.
    • Comparisons against the performance data of the original predicate device (implicitly serving as a benchmark for acceptable "ground truth" performance).

    8. The sample size for the training set:

    Not applicable. This device is a physical reprocessed medical instrument and does not involve AI or machine learning models that require training sets.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181738

    Validate with FDA (Live)

    Date Cleared
    2019-03-22

    (263 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors models MAXAL, MAXP, and MAXI are indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for patients in the sizes indicated in the respective sensor directions for use. These devices are for prescription use only.

    Device Description

    The Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors, models MAXA, MAXAL, MAXP, and MAXI are designed for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in conjunction with a Nellcor Pulse Oximeter. The Reprocessed Nellcor OxiMax SpO2 Sensors, models MAXA, MAXAL, MAXP, and MAXI are intended for prescription use with adult, pediatric and infant patients in hospitals, hospital-type facilities, and intra-hospital transport. The proposed device is not provided sterile.

    AI/ML Overview

    This document describes the validation of Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors (models MAXA, MAXAL, MAXP, and MAXI), which are reprocessed pulse oximetry sensors. The study aims to demonstrate that the reprocessed devices are substantially equivalent to the original, legally marketed predicate device (Nellcor OxiMax Pulse Oximeter Sensors, K052186) and a reference device (Hygia Health Services Reprocessed OxiMax Sensors, K041867).

    Due to the nature of this medical device (a reprocessed pulse oximetry sensor that measures physiological parameters, not an AI/ML powered device), many of the requested criteria related to AI/ML software validation (e.g., sample size for test set/training set, number of experts for ground truth, adjudication method, MRMC studies) are not applicable (N/A) in this context. The study focuses on demonstrating physical and performance equivalence through non-clinical and clinical testing, rather than evaluating the accuracy of a machine learning algorithm against a human expert's interpretation of images or other complex data.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on achieving performance comparable to the predicate device, particularly in terms of SpO2 and pulse rate accuracy, and satisfying relevant standards and guidance documents.

    Acceptance Criteria CategorySpecific Criteria (Target)Reported Device Performance (Achieved)
    SpO2 AccuracySpO2 accuracy of 70% - 100% ± 2 digits (as per predicate device, under steady state and non-motion conditions)Achieved "an accuracy of 2% for 70% - 100% SpO2." "The study concluded that the SpO2 accuracy performance of the proposed devices passed the Arms specification of 3% under steady state and non-motion conditions for the range of 70% to 100%." (Note: The predicate device listed 2 digits, while the study concludes it passed 3% Arms, which is generally acceptable for pulse oximeters within the 70-100% range, often assessed using Arms and bias. The 2% claimed for 70-100% SpO2 seems to match the predicate's ±2 digits.)
    Pulse Rate Accuracy20 - 250 bpm ± 3 digits (as per predicate device)20 - 250 bpm ± 3 digits (explicitly stated in the comparison table as "Medline ReNewal is same as predicate")
    BiocompatibilityNo cytotoxicity, sensitization, irritation; acceptable acute systemic toxicity.Tested and deemed acceptable: "Biocompatibility: cytotoxicity, sensitization, irritation; acute systemic toxicity"
    DisinfectionEffective disinfection.Tested and deemed acceptable.
    Shelf LifeAppropriate shelf life validation.Tested and deemed acceptable.
    Electrical SafetyCompliance with electrical safety standards.Tested and deemed acceptable.
    Performance TestingAcceptable tissue heating, pulse rate accuracy, active element assessment, adhesive peel, and environmental (extreme heat and operating conditions) performance.Tested and deemed acceptable: "tissue heating," "pulse rate accuracy," "active element assessment," "adhesive peel," and "environment (extreme heat and operating conditions)."
    Cleaning EfficacyEffective cleaning with minimal residual protein and hemoglobin.Tested and deemed acceptable: "Cleaning: visual inspection; cleaning efficacy (residual protein and residual hemoglobin)."
    Intended Use EquivalenceMedline ReNewal will not make claims for neonatal use or for motion or low perfusion performance. Otherwise, same as predicate.Conformed to this restriction. "Medline ReNewal will only reprocess pulse oximeter sensors. Medline ReNewal will not make claims for neonatal use or for motion or low perfusion performance; otherwise same as written." The patient population for the proposed device is "Adult, Pediatric, Infant," excluding neonates compared to the predicate/reference which included neonates.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Test Set: 10 healthy adult volunteer subjects.
    • Data Provenance: The clinical study was conducted from May 9 to May 10, 2018. While the specific country of origin is not explicitly stated, the FDA 510(k) submission format suggests it's likely a study conducted in the U.S. or in accordance with U.S. regulatory standards (CFR for Non-significant Risk Investigational Studies, ISO 14155:2011, and ISO 80601-2-61:2011 are referenced). It was a prospective clinical study specifically designed to evaluate the reprocessed device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • N/A. For this type of device (pulse oximetry sensor), "ground truth" for SpO2 accuracy is not typically established by human expert consensus or radiologists' interpretations. Instead, it's established by a recognized clinical methodology, such as arterial blood gas (ABG) analysis (co-oximetry), which provides a direct and objective measurement of arterial oxygen saturation (SaO2) and is considered the gold standard for validating SpO2 accuracy. The study mentions it followed "ISO 80601-2-61:2011 Procedure for invasive laboratory testing on healthy volunteers," which describes how to perform such comparisons using co-oximetry. There are no "experts" in the sense of image readers or diagnosticians determining a "ground truth" for each measurement.

    4. Adjudication Method for the Test Set

    • N/A. As the ground truth is established by objective, invasive laboratory testing (e.g., co-oximetry), there is no need for human adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study is relevant for evaluating diagnostic imaging systems where human readers interpret cases with and without AI assistance. This study pertains to a reprocessed medical sensor for physiological measurement, not an imaging or AI diagnostic device. Therefore, no MRMC study was performed, and no effect size regarding human reader improvement with AI assistance is applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance

    • Yes, in essence. The clinical study evaluates the reprocessed sensor's performance (SpO2 accuracy) on subjects by comparing its readings to a recognized gold standard (invasive laboratory testing). While the device outputs a reading for a human to interpret, its fundamental performance (accuracy of the SpO2 measurement) is assessed as a standalone measurement, irrespective of specific human interpretation biases. There isn't an "algorithm" in the AI/ML sense, but the device's sensing and processing of physiological signals are evaluated directly.

    7. Type of Ground Truth Used

    • Clinical Gold Standard / Objective Measurement: The ground truth for SpO2 accuracy was established through invasive laboratory testing, specifically by comparing the device's SpO2 readings to reference SaO2 values obtained from arterial blood samples (co-oximetry), as per the procedures outlined in ISO 80601-2-61:2011. This method is considered the gold standard for pulse oximeter accuracy validation.

    8. Sample Size for the Training Set

    • N/A. This device is a reprocessed sensor, not a machine learning or AI-powered system that requires a "training set." Its performance is based on its physical and functional characteristics after reprocessing, which are validated against established standards and the predicate device, not by training an algorithm on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • N/A. As there is no training set for an AI/ML algorithm, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182588

    Validate with FDA (Live)

    Date Cleared
    2018-10-31

    (41 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Reprocessed LigaSure Maryland Jaw Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, adhesiolysis, hysterectomy, oophorectomy, etc.

    The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

    Device Description

    The Medline ReNewal Reprocessed LigaSure Maryland Jaw Sealer/Dividers (models LF1723, LF1737 and LF1744) are sterile, single-use, hand-held bipolar electrosurgical instruments designed for use with the Covidien electrosurgical generators to ligate (seal) and divide (cut) vessels, tissue bundles and lymphatics during open and minimally invasive general surgical procedures using radio frequency (RF) energy. A hand actuated mechanism allows the user to open and close the instrument jaws. When the instrument jaws are correctly placed over tissue or vessel to be sealed, the user operates a second control to initiate delivery of bipolar energy, which seals the tissue. When the sealing cycle is complete, the user operates a separate control to activate a knife, which divides the tissue along the seal line.

    AI/ML Overview

    The request is for information on the acceptance criteria and study proving device performance for the Medline ReNewal Reprocessed LigaSure Maryland Jaw Sealer/Dividers (models LF1723, LF1737, and LF1744), as detailed in the FDA 510(k) summary K182588.

    Based on the provided document, the device is a reprocessed version of a legally marketed predicate device. The submission primarily focuses on demonstrating that the reprocessed device is substantially equivalent to the original predicate device (Covidien LigaSure Maryland Jaw One Step Sealer/Dividers with 510(k) K141153) through performance testing.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the functional characteristics of the proposed reprocessed devices have been evaluated and found to be equivalent to the predicate devices. The acceptance criteria are implied to be that the reprocessed device performs comparably to the new predicate device across various tests.

    Acceptance Criterion (Implied)Reported Device Performance
    Electrical Safety & Electromagnetic Compatibility (EMC)In accordance with IEC 60601-2-2 and 60601-1-2. (The reprocessed device meets these standards, implying equivalence to the predicate which would also meet these standards).
    Simulated UseEvaluated and found equivalent to the predicate devices. This implies the reprocessed devices perform as expected during simulated surgical procedures.
    Device IntegrityEvaluated and found equivalent to the predicate devices. This suggests the physical and structural integrity of the reprocessed devices is maintained.
    Blade Trigger Advance/ReturnEvaluated and found equivalent to the predicate devices. The cutting mechanism functions correctly.
    Activation Button Device RecognitionEvaluated and found equivalent to the predicate devices. The device properly interacts with the associated electrosurgical generators.
    Thermal Analysis CharacterizationEvaluated and found equivalent to the predicate devices. The reprocessed device's thermal performance (e.g., heat generation during use) is comparable to the predicate.
    Tissue StickingEvaluated and found equivalent to the predicate devices. The reprocessed device minimizes tissue sticking during use, similar to the predicate.
    Burst PressureEvaluated and found equivalent to the predicate devices. This likely refers to the strength of the seal created by the device on vessels, demonstrating its ability to occlude vessels effectively, comparable to the predicate.
    HistopathologyEvaluated and found equivalent to the predicate devices. This suggests the tissue effects (e.g., thermal damage, seal quality) produced by the reprocessed device are similar to those of the predicate.
    Seal QualityEvaluated and found equivalent to the predicate devices. This is a direct measure of the effectiveness of the device in sealing vessels, confirming it performs comparably to the predicate.
    Cleaning: Protein, and Carbohydrates (Residuals After Reprocessing)Evaluated and found equivalent to the predicate devices. This confirms the reprocessing methods effectively remove biological contaminants to a safe level, meeting the standards necessary for a reprocessed device. (Though "equivalent to the predicate" is stated, for cleaning, the acceptance is typically passing a defined threshold for residuals, implying the reprocessed devices meet these thresholds).
    Biocompatibility: Cytotoxicity, Sensitization, Irritation, Pyrogenicity, and Acute Systemic ToxicityEvaluated and found equivalent to the predicate devices. This indicates that the materials and reprocessing do not induce adverse biological responses, meeting established biocompatibility standards. (Similar to cleaning, "equivalent to the predicate" here means meeting the same biological safety standards of a new device).
    Performance QualificationEvaluated and found equivalent to the predicate devices. This generally refers to verification that the manufacturing process consistently produces device that meets specifications.
    Sterilization ValidationEvaluated and found equivalent to the predicate devices. This confirms the sterilization process effectively renders the reprocessed devices sterile to an acceptable sterility assurance level (SAL).
    Product StabilityEvaluated and found equivalent to the predicate devices. This assesses the device's ability to maintain its safety and performance characteristics over its shelf life.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each of the performance tests. It broadly states "The functional characteristics of the proposed devices have been evaluated..." without providing numerical details for the test sets.

    The data provenance is not explicitly stated in terms of country of origin. Given it is an FDA submission for a device manufactured by "Surgical Instrument Service and Savings Inc (dba Medline ReNewal)" and the predicate is from "Covidien," it can be inferred that the testing and data would be primarily from the United States, following U.S. regulatory guidelines. The data is retrospective in the sense that it relies on established standards and comparisons to an already legally marketed predicate device, rather than a novel prospective clinical trial for the reprocessed device's primary effectiveness.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for this type of device comparison study. The tests listed are primarily engineering and laboratory-based performance evaluations (e.g., electrical, mechanical, material, and in-vitro or ex-vivo tissue property tests). For biocompatibility and histopathology, qualified laboratories and pathologists would perform the evaluations, but their specific number and qualifications are not detailed in this summary.

    4. Adjudication method for the test set

    Not applicable. This is not a study requiring adjudication of clinical outcomes or interpretations by multiple readers/experts. The testing involves objective measurements against established engineering and biological standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving AI or human readers for diagnostic interpretation. It is a performance evaluation of a reprocessed surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance evaluation of this reprocessed device is established by:

    • Predicate Device Performance: The primary ground truth is the established safety and effectiveness of the legally marketed predicate device (Covidien LigaSure Maryland Jaw One Step Sealer/Dividers, K141153). The reprocessed device is deemed substantially equivalent if its performance matches or falls within acceptable ranges compared to the predicate.
    • Industry Standards: Compliance with international and national standards (e.g., IEC 60601 series for electrical safety, ISO standards for biocompatibility and sterilization, and internal performance specifications), which represent a form of accepted "ground truth" for device safety and performance.
    • Laboratory-based measurements: Objective measurements of physical, chemical, and biological properties through various tests (e.g., burst pressure, protein residuals, cytotoxicity).
    • Histopathology: Evaluation of tissue effects, performed by qualified personnel (likely pathologists), forms a 'ground truth' for tissue interaction.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning or AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170955

    Validate with FDA (Live)

    Date Cleared
    2017-07-03

    (94 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Reprocessed Sonicision Cordless Ultrasonic Dissection Devices models SCD 391 and SCD 396 are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The devices can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures. The Sonicision Cordless Ultrasonic Dissection Devices can be used to coagulate isolated vessels up to 5 mm diameter.

    Device Description

    The Sonicision Cordless Ultrasonic Dissector Device, models SCD391 and SCD396, are sterile, single-use components to which the Sonicision Reusable Generator and Reusable Battery Pack attach. (The Sonicision Reusable Generator and Reusable Battery Pack are not included in this submission and will not be reprocessed by Medline ReNewal.) This component provides control for device functions such as selecting power levels, blade placement and position, grasping, coagulating and dissecting tissue. The Cordless Ultrasonic Dissection Devices can coagulate vessels up to 5 mm in diameter. Furthermore, they are designed to be inserted and extracted through a compatible 5 mm trocar when used endoscopically.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    The document is a 510(k) premarket notification for a reprocessed medical device, specifically the Medline ReNewal Reprocessed Sonicision Cordless Ultrasonic Dissector, Models SCD 391 and SCD 396. The core of a 510(k) submission is to demonstrate "substantial equivalence" to a legally marketed predicate device, rather than proving novel clinical effectiveness or superiority. Therefore, the testing described focuses on demonstrating that the reprocessed device performs as intended and is as safe and effective as the original, rather than establishing new performance benchmarks for all surgical devices.

    1. A table of acceptance criteria and the reported device performance

    The document doesn't provide a direct table of specific numerical acceptance criteria alongside reported performance for each test. Instead, it lists types of performance tests conducted to establish equivalence. The general acceptance criterion for all the tests is that the reprocessed device must demonstrate equivalence to the predicate device in terms of safety and effectiveness.

    Acceptance Criteria (General Description)Reported Device Performance (Summary from document)
    Equivalence to predicate device in functional characteristicsFound to be equivalent to the predicate devices based on various tests.
    Grasping/Pulling ForceEvaluated (Found equivalent)
    Cutting Effectiveness/FunctionalityEvaluated (Found equivalent)
    Drop TestEvaluated (Found equivalent)
    Device IntegrityEvaluated (Found equivalent)
    Cutting/Coagulation Evaluation for Prolonged PeriodEvaluated (Found equivalent)
    Thermal Analysis CharacterizationEvaluated (Found equivalent)
    Tissue StickingEvaluated (Found equivalent)
    Cleaning (Effectiveness of reprocessing)Evaluated (Found equivalent)
    Protein, Carbohydrates, and Endotoxins (Residuals after cleaning)Evaluated (Found equivalent)
    Biocompatibility (Cytotoxicity, Sensitization, Irritation)Evaluated (Found equivalent)
    Pyrogenicity, and Acute Systemic ToxicityEvaluated (Found equivalent)
    Performance QualificationEvaluated (Found equivalent)
    Sterilization ValidationEvaluated (Found equivalent)
    Product StabilityEvaluated (Found equivalent)

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify the exact sample sizes (N numbers) used for each individual performance test. It only states that the functional characteristics "have been evaluated."
    • Data Provenance: The data provenance is not explicitly stated. However, since this is a submission for a reprocessed device by Medline ReNewal, the testing would have been conducted by Medline ReNewal or their contracted testing facilities. Given the nature of performance testing for medical devices, it's generally done in a controlled laboratory environment. The origin of the reprocessed devices themselves would be from various healthcare facilities where they were originally used before being sent for reprocessing. The testing itself is prospective in the sense that Medline ReNewal performed these tests on their reprocessed devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information (expert consensus, qualifications) is typically relevant for studies involving subjective interpretation, like imaging diagnostics or clinical endpoints, where human experts establish ground truth. For this submission, which is about the functional and safety equivalence of a reprocessed surgical instrument, the "ground truth" is established by direct physical and chemical testing against defined specifications or comparison to the performance of the new predicate device. Therefore, information about "experts establishing ground truth" in the clinical sense is not applicable or provided in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods (like 2+1, 3+1) are typically used in clinical trials or diagnostic accuracy studies where multiple readers interpret cases and disagreements are resolved. This document pertains to performance testing of a reprocessed surgical instrument, which involves objective laboratory measurements and comparisons to predetermined specifications or predicate device performance. Therefore, an adjudication method in this context is not applicable and not mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic aids, especially those involving AI interpretation, where human readers interact with a system to make diagnoses. The Medline ReNewal device is a surgical instrument, not an AI diagnostic tool.
    • Effect Size of AI assistance: Not applicable, as this is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This is a physical surgical instrument, not an algorithm or software. Its performance is inherent to its mechanical, electrical, and material properties when used by a surgeon.

    7. The type of ground truth used

    The ground truth for evaluating the safety and effectiveness of the reprocessed device is established through:

    • Comparison to the predicate device's known performance: The fundamental scientific technology and intended use are considered identical to the predicate.
    • Pre-defined specifications and standards: Each performance test (e.g., grasping/pulling force, cutting effectiveness, cleaning, biocompatibility, sterilization) would have acceptance criteria based on established engineering standards, regulatory requirements, and the expected performance of a new device.
    • Objective laboratory measurements: Physical, chemical, and biological tests are performed to directly measure different aspects of the device's function and safety.

    Essentially, the "ground truth" is objective measurement against established requirements and comparison to the original device's performance.

    8. The sample size for the training set

    This concept of a "training set" is relevant for machine learning or AI models. Since this is a physical medical device and not an AI or software product, there is no training set in the traditional sense. The development and validation of the reprocessing procedure would involve internal process development and verification, but it's not a "training set" for an algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153745

    Validate with FDA (Live)

    Date Cleared
    2016-05-02

    (125 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reprocessed Medline ReNewal LigaSure Impact LF4318 Curved Large Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The LF4318 is intended to be used with the ForceTriad energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

    The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The LigaSure Impact 4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles.

    The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

    Device Description

    The reprocessed Medline ReNewal LigaSure Impact, Curved, Large Jaw, Open Sealer/Divider (LF4318) is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad Energy Platform (cleared under K051644, K070162) to seal and divide vessels (including pulmonary) up to and including 7 mm in diameter, tissue bundles, and lymphatics during open general surgical procedures. The ForceTriad's' tissue-fusion (LigaSure) mode is designed to deliver precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion.
    The device combines the benefits of a long, curved jaw and a shaft-based jaw mechanism for improved access and visibility to critical structures. The pistol-grip-style handle was designed for improved comfort and usability, with an integrated hand switch and cutter for the subsequent sealing and dividing of tissue. Hand controls have been symmetrically placed to facilitate handling by both left- and right-handed users. The LF4318 device connects to the ForceTriad energy platform with a 10-foot cord containing a LigaSure cable connector. This connector functions as a unique product identifier for device -specific recognition by the generator. The 10-foot cord, cable connector, and ForceTriad are not reprocessed by Medline ReNewal.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a reprocessed medical device, the Medline ReNewal Reprocessed LigaSure Impact Open Sealer/Divider (Model LF4318). This submission aims to demonstrate substantial equivalence to a legally marketed predicate device (Covidien LigaSure Impact LF4318).

    Here's an analysis of the acceptance criteria and the study information, extracting what's available from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" with specific pass/fail thresholds for each test, nor does it provide a direct numerical "reported device performance" against such criteria. Instead, it states that the functional characteristics "have been evaluated and found to be equivalent to the predicate devices based on the following tests."

    However, we can infer the type of performance evaluation conducted:

    Test TypeImplied Acceptance Criteria (Inferred from "found to be equivalent")Reported Device Performance (Inferred from "found to be equivalent")
    Electrical Safety and EMC (IEC 60601-1, 60601-1-2)Compliance with international standards for electrical safety and electromagnetic compatibility.Compliant (equivalent to predicate).
    Simulated UsePerformance under simulated surgical conditions comparable to the predicate device.Equivalent to predicate.
    Device IntegrityStructural and functional soundness comparable to the predicate device.Equivalent to predicate.
    Blade Trigger Advance/ReturnProper functioning of the cutting blade mechanism.Equivalent to predicate.
    Shaft Knob RotationProper functioning of the shaft rotation.Equivalent to predicate.
    Device RecognitionCorrect recognition by the associated energy platform (ForceTriad).Equivalent to predicate.
    Thermal Analysis CharacterizationThermal profile during use comparable to the predicate device.Equivalent to predicate.
    Handle LockingProper functioning of the handle locking mechanism.Equivalent to predicate.
    Burst PressureVessel sealing strength (burst pressure) comparable to the predicate device.Equivalent to predicate.
    HistopathologyTissue effect (e.g., seal quality, thermal spread) comparable to the predicate device.Equivalent to predicate.
    Seal QualityQuality of vessel seals comparable to the predicate device.Equivalent to predicate.
    Tissue StickingDegree of tissue sticking to the device comparable to the predicate device.Equivalent to predicate.
    Cleaning: Protein, CarbohydratesEffectiveness of cleaning process to remove biological contaminants to accepted levels.Met (implied by "equivalent").
    Biocompatibility: Sensitization, IrritationNo adverse biological reactions (sensitization, irritation) when in contact with tissues, comparable to predicate.Met (implied by "equivalent").
    Biocompatibility: PyrogenicityAbsence of fever-inducing substances (pyrogens) to accepted levels.Met (implied by "equivalent").
    Biocompatibility: Acute Systemic ToxicityAbsence of acute toxic effects, comparable to predicate.Met (implied by "equivalent").
    Performance QualificationOverall functional performance meeting specified requirements.Met (implied by "equivalent").
    Sterilization ValidationSterility assurance level (SAL) achieved and maintained.Met (implied by "equivalent").
    Product StabilityDevice maintaining its characteristics over shelf life.Met (implied by "equivalent").

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for any of the tests conducted.
    The document does not specify the data provenance (e.g., country of origin, retrospective or prospective nature of the testing). The nature of the tests (e.g., simulated use, burst pressure, histopathology, cleaning, biocompatibility) suggests they are prospective bench and lab-based studies, rather than clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth, as this type of preclinical testing typically relies on predefined material properties and functional parameters rather than expert interpretation of complex clinical data. For histopathology, it would typically be a pathologist, but no specifics are given.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method. This is generally not relevant for the types of bench and lab tests listed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed or described. The device is a reprocessed electrosurgical instrument, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This section is for AI/software devices. The device is a physical electrosurgical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests would be established through:

    • Physical and electrical measurements: For electrical safety, device integrity, blade function, shaft rotation, handle locking, device recognition.
    • Material and mechanical properties standards: For burst pressure, product stability.
    • Microscopic evaluation/laboratory analysis: For histopathology (tissue effect), seal quality, tissue sticking, cleaning (protein, carbohydrates).
    • Standardized biological assays: For biocompatibility (sensitization, irritation, pyrogenicity, acute systemic toxicity).
    • Sterility testing: For sterilization validation.

    These are objective measurements and standardized tests designed to demonstrate equivalence to the predicate device's established performance specifications.

    8. The sample size for the training set

    Not applicable. This device is a physical instrument, not a machine learning model that requires a training set. The term "training set" would not apply here.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151617

    Validate with FDA (Live)

    Date Cleared
    2016-01-22

    (221 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

    Device Description

    Medline ReNewal Reprocessed St. Jude Livewire Electrophysiology Catheters are commonly placed at the high right atrium, right ventricular apex and His bundle, and in the coronary sinus, and are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

    AI/ML Overview

    The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter (K151617) underwent performance testing to demonstrate substantial equivalence to its predicate device, the St. Jude Medical Livewire Electrophysiology Catheter (K022380).

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a list of functional performance studies, cleaning validation, biocompatibility testing, packaging/sterilization validation, and product stability tests performed. However, it does not explicitly define specific numerical acceptance criteria for each test nor provides detailed quantitative results (e.g., "bond strength was X N, meeting the criteria of >Y N"). Instead, it states that the functional characteristics were "evaluated and found to be substantially equivalent to the predicate device."

    Here’s a summary of the tests performed, which implicitly form the basis for acceptance:

    Acceptance Criteria (Implicitly: Performance comparable to predicate)Reported Device Performance (Implicitly: Met comparable performance)
    Functional Performance Studies:
    Bond strength (tensile testing)Tested and found comparable
    Torsional strengthTested and found comparable
    Leakage currentTested and found comparable
    Steerable deflection and continuityTested and found comparable
    Catheter/handle joint flexibility and continuity (flex fatigue)Tested and found comparable
    Electrode adherenceTested and found comparable
    Applied force deflection (tip flexibility)Tested and found comparable
    Shaft flexibilityTested and found comparable
    Tip bucklingTested and found comparable
    Direct current resistanceTested and found comparable
    Impedance at 5 kHzTested and found comparable
    Corrosion resistanceTested and found comparable
    Cleaning:
    Protein, total organic carbon, and endotoxinsTested and cleaning performance qualified
    Visual inspection under magnificationTested and cleaning performance qualified
    Cleaning performance qualificationPerformed and qualified
    Biocompatibility:
    CytotoxicityTested and found acceptable
    SensitizationTested and found acceptable
    IrritationTested and found acceptable
    Acute systemic toxicityTested and found acceptable
    PyrogenicityTested and found acceptable
    Hemocompatibility (hemolysis, thrombogenicity, complement activation)Tested and found acceptable
    Packaging and sterilization validation:
    Bioburden testingTested
    Ethylene oxide and ethylene chlorohydrin residuals testingTested
    Product stabilityTested and found acceptable

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for each of the performance tests. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth:

    This information is not applicable and not provided in the context of this 510(k) summary, as it describes a reprocessed medical device's physical and functional performance, not an AI/algorithmic medical device requiring expert ground truth for classification or diagnosis.

    4. Adjudication Method for the Test Set:

    This information is not applicable and not provided for the reasons stated in point 3. The testing involves laboratory and benchtop evaluations of device properties.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not performed, as it is not relevant for the type of device (reprocessed catheter) described. This type of study is typically used for diagnostic imaging or AI devices that interact with human readers/clinicians.

    6. Standalone (Algorithm Only) Performance:

    Standalone performance (algorithm only) was not performed, as it is not relevant for this reprocessed physical medical device. This concept applies to AI/software as a medical device.

    7. Type of Ground Truth Used:

    The "ground truth" for this device's performance evaluation is based on established engineering and biological standards relevant to medical devices, as well as comparison to the performance characteristics of the original, new device (predicate device). For instance, for bond strength, the "ground truth" would be the required mechanical integrity, and for cleaning, it would be the absence of specified contaminants below acceptable thresholds.

    8. Sample Size for the Training Set:

    This information is not applicable and not provided. The listed tests are for verifying the performance of the reprocessed device itself, not for training an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable and not provided for the reasons stated in point 8.