Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K212326

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-01

    (128 days)

    Product Code
    Regulation Number
    882.1275
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AC Cream - Conductive paste is intended for use in clinical and research EEG/EP recordings from humans. It is used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin.

    Device Description

    AC Cream - Conductive paste is intended for use in clinical and research EEG/EP recordings from humans. It is used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin.
    It's characterized by pale yellow colour, no crystallization, no flocculation, bright.
    AC Cream - Conductive paste function is of conductor between the electrode used and the patient's skin and of getting the impedance lower for a better recording of the signal. AC Cream - Conductive paste is for external use with recording electrodes only.
    AC Cream - Conductive paste is made of powders, Potassium Chloride and Sodium Chloride as conductors, combined with thickening agents and humectants, all in an aqueous solvent.
    The composition is the following: Water, Ceteareth-20, Glycerol, Propylene Glycol, Bentonite, Sodium chloride, Potassium chloride, Calcium carbonate, Polysorbate 20, Phenoxyethanol, Ethylhexylglycerin
    The pH range is 6÷8, and Impedance at 10Hz is 120± 25 Ohm.
    The mean impedance after 7 days use, considering the worst values detected, is around 480 ± 10 Ohm.
    The Conductivity is 20 mS/cm
    Shelf life of AC Cream - Conductive paste is 3 years if stored properly in the closed packaging, kept away from the sunlight and within the limit temperature

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for AC Cream - Conductive paste:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Set by Standard)Reported Device Performance (AC Cream - Conductive paste)Pass/Fail?
    Biocompatibility:Pass
    ISO 10993-5 (Cytotoxicity)YesPass
    ISO 10993-10 (Irritation)YesPass
    ISO 10993-10 (Skin Sensitization)YesPass
    Shelf Life (Accelerated Aging - 3 years):Pass
    Pale yellow color, no flocculation, brightExhibits pale yellow color, no flocculation, bright after 84 days (accelerated aging equivalent to 3 years)Pass
    Instant Impedance < 2000 Ω (according to ANSI/AAMI EC12:2000/(R)2015)Max Instant Impedance measured was 144 Ohm (PE Jar) and 118 Ohm (Aluminum tube) after 84 days (accelerated aging equivalent to 3 years)Pass (Significantly Below Limit)
    Long-Term Conductivity (7 days):Pass
    DC Offset Voltage < 100 mV (according to ANSI/AAMI EC12:2000/(R)2015)Complies with limit even after 7 days testingPass
    AC Impedance < 2000 Ohm (according to ANSI/AAMI EC12:2000/(R)2015)Complies with limit even after 7 days testingPass

    2. Sample size used for the test set and the data provenance:

    • Test Set Description: The summary describes performance testing for biocompatibility, accelerated aging (shelf life), and long-term conductivity.
    • Sample Size: The exact sample size for each test (e.g., number of units tested for aging, number of samples for biocompatibility) is not explicitly stated in the provided text.
    • Data Provenance: The document indicates that the performance testing was "conducted on AC Cream - Conductive paste." This implies the testing was performed internally by Spes Medica S.r.l. or by a contract lab on their behalf. There is no information regarding the country of origin of the data or whether the tests were retrospective or prospective since it concerns product qualification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to this type of device (electroconductive paste) or the described studies. The studies are bench tests and biocompatibility tests against established standards. Ground truth, in the sense of expert consensus on diagnostic interpretations, is not relevant here. The "ground truth" for these tests is the quantitative measurement against predefined physical and biological acceptance limits from international standards (ISO, ANSI/AAMI).

    4. Adjudication method for the test set:

    This information is not applicable as the tests performed are objective physical and biological measurements against specific numerical or qualitative (e.g., "no flocculation") criteria. There is no subjective interpretation that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This device is an electroconductive paste, not an AI or imaging diagnostic device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm performance study was not done. This device is a physical product (conductive paste), not an algorithm or software.

    7. The type of ground truth used:

    The ground truth used for the performance testing is based on:

    • International Standards: ISO 10993-5, ISO 10993-10, and ANSI/AAMI EC12:2000/(R)2015. These standards define the acceptable limits and methodologies for testing biocompatibility and electrical performance parameters for electroconductive media.
    • Physical/Chemical Properties: Measured values for color, crystallization, flocculation, brightness, pH, impedance, and conductivity.

    8. The sample size for the training set:

    This information is not applicable. This device is not an AI/machine learning product and therefore does not have a "training set" in that context. The device's formulation was likely developed through R&D, but this is not a "training set" in the context of typical AI device submissions.

    9. How the ground truth for the training set was established:

    This information is not applicable, as there is no "training set" in the AI/ML context for this device. The formulation and properties were established through product development, material science principles, and internal quality control, with the goal of meeting the performance requirements eventually tested against the standards mentioned above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192606

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-17

    (119 days)

    Product Code
    Regulation Number
    882.1275
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAC2 is intended for use in clinical and research EEG/EP recordings from humans. It is used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin.

    Device Description

    SAC2 is intended for use in clinical and research EEG/EP recordings from humans. It is used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin. SAC2 is provided in an aluminum tube of 100g. It's characterized by light grey colour, no crystallization, no flocculation, no adverse smell, opaque. SAC2 function is of conductor between the electrode used and the patient's skin and of getting the impedance lower for a better recording of the signal. SAC2 is for external use with recording electrodes only. SAC2 is made of powders, Potassium Chloride and Sodium Chloride as conductors, combined with thickening agents and humectants, all in an aqueous solvent. The composition is the following: Water, Talc, Celite, Glycerol, CarboxyMethylCellulose, Sodium chloride, Potassium chloride, Phenoxyethanol, Ehylhexylglycerin The pH range is 8÷10, and Impedance at 10Hz is 50 ± 10 Ohm. The Conductivity is 20 mS/cm Shelf life of SAC2 is 3 years if stored properly in the closed aluminum tube kept away from the sunlight.

    AI/ML Overview

    The provided text describes the 510(k) summary for the SAC2 - Electrode cream. It does not contain information about a study involving human readers or AI assistance. The performance testing section focuses on bench tests (aging, long-term conductivity) and biocompatibility, not on a clinical effectiveness study with human subjects.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceStandard/Methodology
    Aging TestASTM F1980-16
    Light grey colorComplies (light grey color)Visual inspection
    No crystallizationComplies (no crystallization)Visual inspection
    No flocculationComplies (no flocculation)Visual inspection
    OpaqueComplies (opaque)Visual inspection
    pH range8-10Internal testing
    Impedance at 10Hz50 ± 10 OhmInternal testing, complies with ANSI/AAMIEC12:2000
    Long Term Conductivity TestANSI/AAMIEC12:2000/(R)2015
    DC Offset voltageDoes not exceed 100mV (after 7 days testing)Electrical testing
    AC ImpedanceDoes not exceed 2000 Ohm (after 7 days testing)Electrical testing
    Biocompatibility
    CytotoxicityPassedISO 10993-5
    IrritationPassedISO 10993-10
    SensitizationPassedISO 10993-10

    2. Sample size used for the test set and the data provenance

    Not applicable for this type of product/study. The performance tests described are bench tests on the product itself, not on data from human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth for biocompatibility and bench testing is established through standardized laboratory procedures and instrument readings, not expert consensus.

    4. Adjudication method for the test set

    Not applicable. Adjudication is typically used in studies involving subjective interpretation (e.g., medical imaging), which is not the case for these performance tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The document pertains to an electrode cream, not an AI software or system that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. The device is a physical electrode cream, not an algorithm.

    7. The type of ground truth used

    The ground truth for the performance tests (aging, long-term conductivity) was established by objective measurements against defined parameters (e.g., color, pH, impedance, DC offset voltage) and the biological responses observed in standardized biocompatibility tests (cytotoxicity, irritation, sensitization). This is based on established industry standards like ASTM F1980-16, ANSI/AAMIEC12:2000, ISO 10993-5, and ISO 10993-10.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. As noted above, this device does not involve a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192603

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-22

    (63 days)

    Product Code
    Regulation Number
    882.1350
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spes Medica Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG) and Nerve potential signals. The electrodes are sterile and for single patient use only.

    Device Description

    Spes Medica Subdermal Needle Electrodes are monopolar needles intented to use for Electromyography (EMG), Electroencephalograph (EEG) and Nerve potential signals in example and they are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. The Subdermal Needle Electrodes are used for recording and stimulation (micro stimulation), this stimulation has no therapeutic purpose but has the purpose of activating an electrical response.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance testing of the Spes Medica Subdermal Needle Electrodes, but it largely focuses on biocompatibility, sterilization/packaging, and electrical safety standards rather than a comparative effectiveness study involving human readers or AI.

    Therefore, many of the requested fields related to AI performance, human reader studies, and large-scale clinical data will not be present in this document.

    Here's a breakdown of the available information based on your request:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test Description/Standard)Reported Device Performance (Results)
    Biocompatibility
    Cytotoxicity (ISO 10993-5)No reactivity
    Skin Irritation (ISO 10993-10)Negligible irritation response
    Hemolysis test (ISO 10993-4)Does not cause hemolysis
    Skin Sensitization (ISO 10993-10)Not sensitizing
    Systemic Toxicity (ISO 10993-11)No toxic symptoms, satisfies requirements
    Pyrogenicity test (ISO 10993-11)Meets requirements for absence of pyrogens
    Sterilization/Packaging
    Packaging Validation Protocol (5 years shelf-life)No substantial alterations after aging process, devices remain sterile (Tested against ASTM D4332, ASTM D4169, ISO 11135-1, 11737-1, 11737-2)
    Packaging validation test (Sterility test)All samples met seal strength requirements per ASTM F88
    Electrical Safety
    Electrical Safety - Isolation (IEC 60601-1: 3rd Ed + CORR 2:2007 + A1:2012 clause 8.5.2.3)Isolated and in compliance with requirements for electrically safety
    Impedance test and dielectric strengthStimulation doesn't affect or reduce the needle and its performance
    Physical Performance
    Cable tensile testingVery high strength force (average 40.1 N) during pull-out test, ensuring manufacturing process safety and strength of use
    Shipment Tests (Handling, Vehicle Stacking, Loose Load Vibration, Vehicle Vibration, Concentrated Impact)No damage to the product observed
    Climatic Tests (Changing temperature and relative humidity)No damage to the product observed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify general "test sets" or data provenance in the way one would for an AI or clinical study. The performance tests are laboratory-based and conducted on the device components or prototypes. The sample sizes for each specific test (e.g., number of units tested for tensile strength, number of samples for biocompatibility) are not explicitly stated, nor is the country of origin or whether they were retrospective/prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document describes performance testing for a medical device to establish its safety and effectiveness based on international standards, not the establishment of ground truth by clinical experts for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic study requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document does not describe an AI device or an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the various performance tests, the "ground truth" is defined by the requirements of the international standards (e.g., ISO, ASTM, IEC) that the device must meet. For biocompatibility tests, it's the biological response against established thresholds; for electrical safety, it's adherence to voltage and isolation limits; for mechanical tests, it's resistance to forces.

    8. The sample size for the training set

    Not applicable. This is not an AI device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190050

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-07-19

    (190 days)

    Product Code
    Regulation Number
    882.1275
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tech Dots are intended for use in clinical and research EEG/EP recordings from humans. They are used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin

    Device Description

    Tech Dots are conductive gel dots to be used with external electrodes as the conductor between skin and electrode and to reduce impedance between the electrode surface and the skin. A single Tech Dot has a 11 ± 1 mm diameter, 2.5 ± 0.5 mm high, and weights 0.14 ± 0.01 g. It's characterized by clear colour, no crystallization, no flocculation, no adverse smell, brightness. Tech Dots function is of conductor between the electrode used and the patient's skin and of getting the impedance lower for a better recording of the signal. Tech Dots are for use with external electrodes only. Tech Dot is made of Potassium Chloride as conductor, combined with thickening agents and humectants, all in an aqueous solvent. The composition is the following: Water, Glycerol (vegetable origin), Polyacrylate co-polymer (proprietary), Potassium chloride. The pH range is 4.1 ± 0.1, and Impedance at 10Hz is 80 ± 10 Ohm. The Conductivity is 2 mS/cm. Shelf life of TechDots is 3 years if stored properly in sealed aluminum pouch and at the limits of temperature indicated on the labeling.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Tech Dots - Conductive gel," submitted by Spes Medica S.r.l. to the FDA. The submission aims to demonstrate that Tech Dots are substantially equivalent to a legally marketed predicate device (Electro-Gel, K111717).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for "Tech Dots - Conductive gel" are primarily established through demonstrating substantial equivalence to the predicate device and meeting relevant performance standards for electroconductive media. The reported performance is based on internal testing.

    Characteristic / Acceptance CriteriaReported Device Performance (Tech Dots)Comparison to Predicate (Electro-Gel)
    Intended Use (Clinical and research EEG/EP recordings from humans, as conductor between scalp/skin and recessed electrodes to reduce impedance)Intended for use in clinical and research EEG/EP recordings from humans. Used with external electrodes as the conductor between the scalp and recessed electrodes to reduce impedance between the electrode surface and the skin.Same as predicate device.
    Regulation Name (Media, Electroconductive)Media, ElectroconductiveSame as predicate device.
    Regulation Number (882.1275)882.1275Same as predicate device.
    Environment of Use (Electrophysiological)ElectrophysiologicalSame as predicate device.
    Intended User (Neurologists)NeurologistsSame as predicate device.
    Target Patient (Adult and children)Adult and childrenSame as predicate device.
    Where Used (Topically on intact skin)Topically on intact skinSame as predicate device.
    Conductive Material (Salt)Potassium Chloride (a salt)Predicate uses NaCl (Sodium Chloride). The document states "Salt (NaCl)" for Tech Dots here, but later specifies "Potassium Chloride." The discussion indicates "Same as predicate device" for "Conductive material: Salt (NaCl)", but this appears to be a slight inconsistency as the detailed composition lists Potassium Chloride. However, the overall characteristic of "Salt Base" is maintained and considered equivalent.
    Thickening AgentSodium Acrylates Copolymers, GlycerinPredicate uses Aragum, Glycerin. Considered "Equivalent to predicate device."
    Sterilization Method (Provide non sterile)Provide non sterileSame as predicate device.
    Shelf-life3 yearsMore than predicate device (1 year). This is a difference, but not one that raises new questions of safety or effectiveness, as demonstrated by the aging test.
    Chemical Safety (No OSHA PEL)No OSHA PELSame as predicate device.
    PreservativeNo preservativePredicate uses Methylparaben and Propylparaben. Considered "Equivalent to predicate device." This difference (absence of preservatives) is presented as not raising new safety/effectiveness concerns.
    Biocompatibility Testing (Test in accordance with ISO 10993)Test in accordance with ISO 10993 (ISO 10993-5 for cytotoxicity, ISO 10993-10 for irritation and skin sensitization)Same as predicate device. Pass/Fail criteria met for cytotoxicity, irritation, and sensitization.
    CytotoxicityYes (positive result meaning "Pass")Same as predicate device.
    IrritationYes (positive result meaning "Pass")Same as predicate device.
    SensitizationYes (positive result meaning "Pass")Same as predicate device.
    Single UseYesSame as predicate device.
    pH4 ± 0.1 (detailed description on page 3) / 4-5 (summary table)Predicate: 4.5-6.0. Considered "Comparable to predicate device." Internal evaluation by Spes Medica shows 4-5, which overlaps with the lower end of the predicate's range.
    Impedance80 ± 10 Ohm (at 10Hz)Predicate: 0.5 K Ohm (500 Ohm). Tech Dots impedance (80 ± 10 Ohm) is "Less than predicate device," which is usually a desirable characteristic for conductive gel, indicating better conductivity. This doesn't raise new safety/effectiveness concerns. Tested according to ANSI/AAMIEC12:2000/(R)2015, not exceeding 2000 Ohm.
    Weight (per dot where applicable)0.14 ± 0.01 g per DotPredicate: 16, 32 or 128 ounces (presumably in jars). This is a difference in "Different packaging and shape," but not related to inherent performance for substantial equivalence.
    Characteristics (Salt Base, Non-irritating, Non toxic)Salt Base, Non-irritating, Non toxicEquivalent to predicate device.
    Packaging MaterialAluminum/PET/PEPredicate: PE. "Different ways of packaging. Both materials are validated."
    Aging Test (Shelf Life)Clear color, no crystallization, no flocculation, no adverse smell, brightness after accelerated aging. Impedance complies with ANSI/AAMIEC12:2000/(R)2015.Demonstrates meeting 3-year shelf life.
    Long Term Conductivity (Electrical performance over time)DC Offset voltage ≤ 100mV (pass). AC Impedance ≤ 2000 Ohm (pass). Parameters comply with ANSI/AAMIEC12:2000/(R)2015 limits even after 7 days testing.Bench testing successfully demonstrates the product's ability to maintain electrical performance within specified standards over time.

    2. Sample Size Used for the Test Set and Data Provenance

    The document details performance testing (Biocompatibility and Bench Testing) for the Tech Dots device itself, rather than a clinical study involving human subjects or image data for AI.

    • Sample Size for Testing: The specific sample sizes for tests like aging, pH, impedance, and long-term conductivity are not explicitly stated in the provided text (e.g., "how many units were tested"). It mentions "Spes Medica Tech Dots are tested internally for pH, impedance on a regular basis," implying ongoing quality control, but not a specific sample size for a formal study dataset. For biocompatibility, it indicates the gel "Gel" as 'Component Name' being tested.
    • Data Provenance: The studies are described as "Performance Testing was performed on device characteristics of Spes Medica Tech Dots," and "Spes Medica Tech Dots are tested internally." This implies the data is prospective (generated for this submission/validation) and produced by the manufacturer (Spes Medica S.r.l.) in Italy.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This document describes the validation of a conductive gel, not an AI-powered diagnostic device that requires expert-established ground truth from medical images. Therefore, the concept of "experts" establishing "ground truth" for a test set (e.g., radiologists for disease presence) is not applicable here. The "ground truth" for the device's performance is established by technical specifications and recognized industry standards (e.g., ISO 10993, ANSI/AAMIEC12). The "experts" involved would be the engineers and quality assurance personnel at Spes Medica S.r.l. who conducted the tests and verified compliance with these standards. Their specific qualifications are not detailed beyond "Quality Assurance and Regulatory Affairs" for the Official Correspondent.

    4. Adjudication Method for the Test Set

    Not applicable. As this is not an AI device or a clinical study requiring human interpretation of medical data, there is no need for an adjudication method. The test results are objective measurements against defined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device, and therefore no MRMC study or assessment of human reader improvement with AI assistance was conducted or needed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or software device. Its performance is inherent to its physical and chemical properties as a conductive gel.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for validating "Tech Dots - Conductive gel" is based on:

    • Performance Standards: Recognized international and national standards for medical devices and electroconductive media (e.g., ISO 10993 for biocompatibility, ANSI/AAMIEC12 for electrical performance/impedance).
    • Predicate Device Characteristics: The established characteristics and cleared performance of the legally marketed predicate device (Electro-Gel, K111717) serve as a comparative ground truth for demonstrating substantial equivalence of intended use and technological characteristics.
    • Defined Chemical and Physical Specifications: Internal specifications for pH, conductivity, viscosity, appearance, etc., which are then verified through laboratory testing.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI model, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133348

    Validate with FDA (Live)

    Device Name
    STIM PROBE
    Manufacturer
    Date Cleared
    2014-11-21

    (387 days)

    Product Code
    Regulation Number
    874.1820
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identify nerves and spinal nerve roots and to assess nerve function. IOM Stimulator Probes disposable surgical stimulators are indicated for tissue dissection and stimulation of cranial and peripheral nerves for location and identification during surgery, including spinal nerve roots.

    Device Description

    IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identify nerves and spinal nerve roots and to assess nerve function. The probes are supplied sterile and are for single use only. The probes are connected to an electrical stimulator using a flexible lead wire(s) and a 'touch-proof' safety connector(s) on the distal end.

    All IOM Stimulator Probes are sterilized by Ethylene Oxide and single use only.

    The IOM Stimulator Probes are composed by the part in contact with the patient that is medical grade stainless steels, then there is the handle part and cable in biocompatible plastic; finally the cable has a 'touch-proof' safety connectors.

    AI/ML Overview

    The provided document is a 510(k) summary for the "IOM Stimulator Probes" and describes a comparison study against predicate devices to establish substantial equivalence. It does not contain information about a study that proves the device meets specific acceptance criteria in the manner typically associated with clinical performance or algorithm evaluation (e.g., sensitivity, specificity, accuracy). Instead, it focuses on demonstrating that the device has similar technological characteristics to legally marketed predicate devices, and that differences do not raise new questions of safety or effectiveness.

    Therefore, many of the requested details, such as sample size for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, and training set information, are not applicable to the type of study presented in this document.

    Here's an analysis of what is available in the document related to acceptance criteria and the "study" (which is a comparison for substantial equivalence):

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" for performance metrics like sensitivity or specificity. Instead, the acceptance criterion for regulatory clearance is establishing substantial equivalence to predicate devices in terms of intended use, technological characteristics, and safety/effectiveness. The "reported device performance" is demonstrated through a comparison of technological characteristics with predicate devices.

    The document provides multiple comparison tables, one for each type of IOM Stimulator Probe. Below is a summarized example for one probe type, to illustrate how the comparison functions as "performance" in this context.

    Table: Comparison of IOM Stimulator Probes (Example for Neural stimulator probe insulated monopolar)

    SpecificationAcceptance Criteria (Implied by Predicate Performance)Proposed Spes Medical Probe PerformanceDiscussion on Equivalence (Meeting Criteria)
    Stimulating Surface Area (mm²)~4.7 mm² (from Rhythmlink PD)7.5 mm²"The stimulating area is larger than the PD, this doesn't change in dangerous way current densities. So these two differences not affect the safety and effectiveness as compared to the predicate."
    Tip Diameter (mm)0.75 mm (from Rhythmlink PD)1.2 mm"The tip diameter is bigger in Spes Medica device... So these two differences not affect the safety and effectiveness as compared to the predicate."
    Shaft Length (mm)100 mm (from Rhythmlink PD)45 mm"The differences between the two devices are shaft length... Shaft is insulated and the length is only an ease of use... So these two differences not affect the safety and effectiveness as compared to the predicate."
    Lead Length (cm)150 cm and 250 cm (from Rhythmlink PD)250 cm"The cable length is the same 250 cm"
    Bend Angle (degrees)0° (from Rhythmlink PD)The bend angle is the same.
    Material In Contact with PatientStainless steel (from PDs)Stainless steel AISI 316LSame material.
    Touch Proof ConnectorYES (from PDs)YESSame characteristic.
    Single UseYES (from PDs)YESSame characteristic.
    SterilizationEtO (from PDs)EO sterilizedSame characteristic.

    The document includes similar comparison tables for:

    • Neural stimulator probe bipolar hook (compared to Xian Friendship Medical Electronics)
    • Neural stimulator probe tripolar hook (compared to Xian Friendship Medical Electronics)
    • Neural stimulator probe monopolar ball (compared to Xian Friendship Medical Electronics)
    • Neural stimulator probe bipolar balls (compared to Xian Friendship Medical Electronics)
    • Neural stimulator probe bipolar concentric (compared to Xian Friendship Medical Electronics)
    • Neural stimulator probe bipolar minifork (compared to Medtronic)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided. The "study" here is a comparison of product specifications and characteristics against predicate devices, not a clinical study involving patients or a performance evaluation against a labeled dataset. There is no "test set" in the context of clinical data or algorithm performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable and not provided. As above, there is no "test set" requiring expert ground truth in this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. This device is a surgical stimulator probe, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable and not provided. This device is a passive surgical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable and not provided. The "ground truth" equivalent would be the established specifications and performance of the legally marketed predicate devices, against which the new device's specifications are compared.

    8. The sample size for the training set

    This information is not applicable and not provided. This device does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable and not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120494

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-11

    (115 days)

    Product Code
    Regulation Number
    882.1320
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spes Medica Disposable Adhesive Surface Electrodes are intended for use with electrodiagnostic or neurological monitoring equipment for the recording of electrophysiological activity and for electrical stimulation. The electrodes are non-sterile and for single patient use only.

    Device Description

    Spes Medica Disposable Adhesive Surface Electrodes are single patient use, disposable devices. Electrodes are non-invasive as they are applied to the patient's skin using a selfadhesive solid-gel surface. The electrodes consist of 3 different layers of materials.

    The upper layer consists in a cotton non-woven pad, the second layer is the sensor and consists in a conductive layer. The lower layer, then the material in direct contact with the patient's skin consist in a solid Hydro-gel adhesive layer.

    Spes Medica Disposable Adhesive Surface Electrodes are supplied with leads or with snap connector.

    The electrodes which are attached to a lead wire terminate at the opposite end using different type of connectors such us:

    • DIN 42802 "touch proof" safety connectors; .
    • 0,7 mm "touch proof" pin; .
    • Concentric "touch proof" bipolar socket; .
    • 2 mm "touch proof" socket; .

    The snap connector electrode uses Ag/AgCl pellet(s) as the conductive element. The active surface of the pellet is comprised between the cotton non-woven pad and the Hydro-gel adhesive layer. The pellet is attached to a stainless steel male snap connector which is used as connection.

    AI/ML Overview

    The provided text is a 510(k) summary for Disposable Adhesive Surface Electrodes by Spes Medica s.r.l. It asserts that the device is substantially equivalent to legally marketed predicate devices. However, it does not include a study that proves the device meets specific acceptance criteria in the manner requested.

    Instead, the document details a technological comparison to establish substantial equivalence, which is a regulatory pathway to market a device based on its similarity to existing, legally marketed devices (predicates). This pathway typically relies on demonstrating that the new device has the same intended use and similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness.

    Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes for test/training sets, expert qualifications, ground truth establishment, and MRMC studies are not present in the provided document, as these are typically part of a de novo classification or premarket approval (PMA) application, not a 510(k) for substantial equivalence based on technological comparison.

    Here's a breakdown of what can be extracted and what cannot:

    Acceptance Criteria and Device Performance (Not Applicable - Substantial Equivalence Claim)

    The document does not define specific performance acceptance criteria for the new device or present a study comparing its performance against such criteria. Instead, it states:

    "Spes Medica Disposable Adhesive Surface Electrodes are similar in intended use, design, materials, packaging and other technological characteristics to the predicate devices. After analyzing performance and safety testing, it is the conclusion of Spes Medica s.r.I. that the Disposable Adhesive Surface Electrodes are safe and effective as the predicate devices and introduce no new questions concerning safety and effectiveness."

    This suggests that the "acceptance criteria" were implied to be "as safe and effective as the predicate devices," and the "study" was an analysis of the device's characteristics against those of the predicates.

    Other Requested Information (Not Present in the Document)

    The document does not contain information regarding:

    • Sample sizes used for the test set and the data provenance: No test set is described.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment for a test set is described.
    • Adjudication method for the test set: No adjudication method is described.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: No MRMC study is mentioned.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The device is a physical electrode, not software with an algorithm.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as no performance study against a ground truth is described.
    • The sample size for the training set: No training set is described.
    • How the ground truth for the training set was established: Not applicable.

    Device Description and Basis for Substantial Equivalence:

    The document's "study" consists of a comparison to the following predicate devices:

    Predicate device510(k) Holder/Applicant510(k) number
    RLI Cutaneous Disposable ElectrodeRhythmlink International, LLCK052188
    Bio-logic Disposable ElectrodeBio-logic Systems Corp.K941799
    Sunspots Pre-gelled Surface ElectrodesAxon System, Inc.K062198

    The conclusion is based on the new device being similar in:

    • Intended use
    • Design
    • Materials
    • Packaging
    • Other technological characteristics

    The document explicitly states: "No new technology or basic materials are used in these designs."

    In summary, the provided text is a 510(k) premarket notification for a Class II medical device, seeking clearance based on substantial equivalence to existing devices. It does not present a study with quantitative acceptance criteria and performance data as would be found in different types of regulatory submissions for novel devices or significant modifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093373

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-24

    (299 days)

    Product Code
    Regulation Number
    874.1820
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dragonfly Laryngeal Surface Electrode is intended to be used as disposable, self-adhesive electrode attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures. It is intended for use only by a licensed physician and in conjunction with these listed below medical grade electromyographic monitors: Neurovision Medical Products Nerveana Axon Eclipse Medtronic - NIM XLTEK EP Works Nicolet Viking Cadwell Cascade Neurosign Avalanche

    Device Description

    Spes Medica Dragonfly Laryngeal Surface Electrode are single used electrodes for evoked EMG monitoring of the larynx and it to be inserted and retained in position against the larvngeal muscle by use of an endotracheal tube. The Electrodes are intended to be used as disposable, self-adhesive electrodes attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures. Spes Medica produces two different families of electrodes: Family A and Family B. The family A and B mentioned above are the same product because, have been developed with same intended use, design, materials, packaging and other technological characteristics. The manufacturing processes are the same for both configurations (Family A and Family B) but the substrates and the shape of the electrodes are different. Indeed, for the family A the substrate for the Silver ink is a polyester film and for the family B the substrate for the Silver ink is carbon loaded vinyl. Furthermore, both families use the same connection to the EMG unit.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) premarket notification for the "Dragonfly Laryngeal Surface Electrodes," focusing on demonstrating substantial equivalence to previously marketed devices. It describes the device, its intended use, and compares it to predicate devices but does not include details on specific performance metrics, clinical study designs, sample sizes, expert qualifications, or ground truth establishment.

    Therefore, I cannot fulfill your request to provide:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size, data provenance for the test set.
    3. Number and qualifications of experts for ground truth.
    4. Adjudication method.
    5. MRMC comparative effectiveness study results.
    6. Standalone performance study details.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How training set ground truth was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K091410

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-08-11

    (90 days)

    Product Code
    Regulation Number
    890.1385
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Recording muscle activity for Electromyography (EMG) applications. For single patient use only.

    The disposable hypodermic needle is inteded to be used for iniection of the Botulinum Toxin into a muscle. while recording electromyography activity. The electrode has an open lumen and is designed for muscle stimulation, motor unit action, potential recording and Botulinum Toxin injection.

    Device Description

    A diagnostic electromyography needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle). The needle electrodes are for single patient only.

    AI/ML Overview

    This 510(k) premarket notification is for a Disposable EMG Needle Electrode, a Class II medical device. The submission focuses on demonstrating substantial equivalence to previously marketed predicate devices rather than proving novel performance against acceptance criteria through a clinical study. As such, the information typically found in acceptance criteria and efficacy studies is not present in this document.

    Here's an analysis based on the provided text, addressing your questions where possible:

    1. A table of acceptance criteria and the reported device performance

    The provided document does not contain a table of acceptance criteria or reported device performance in the way you might expect for a new, performance-based device. This 510(k) summary is for a device seeking substantial equivalence, meaning its safety and effectiveness are established by showing it is as safe and effective as a legally marketed predicate device. The focus is on demonstrating similar design, materials, packaging, and technological characteristics.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a specific test set, sample size, or data provenance because it is a 510(k) submission based on substantial equivalence. It does not present results from a clinical trial or performance study that would typically involve a test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. No "ground truth" was established for a test set as there was no study comparing the device's performance to a known truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. There was no adjudication method as there was no performance study involving a test set and human readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not performed and is not described in this document. The device is an electrode, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study was not performed and is not described. The device is a physical electrode, not a standalone algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not applicable and not provided. No "ground truth" was used in a performance study.

    8. The sample size for the training set

    This information is not applicable and not provided. This document does not describe the development of an algorithm or AI model that would require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. There was no training set for which a ground truth needed to be established.


    Summary of the K091410 Submission:

    This 510(k) premarket notification by Spes Medica s.r.l. for their "Disposable EMG Needle Electrodes" is a Class II medical device submission seeking substantial equivalence. The core argument for safety and effectiveness is that the device is "similar in design, materials, packaging and other technological characteristics" to several legally marketed predicate devices (e.g., MEDICOTEST A/S Neuroline, AMBU A/S Neuroline, MEDELEC Disposable Needle Electrode, etc.).

    The document describes the device's intended use: "Recording muscle activity for Electromyography (EMG) applications. For single patient use only." It also notes a specific use case for the hypodermic version: "injection of the Botulinum Toxin into a muscle, while recording electromyography activity."

    The FDA's letter (AUG 11 2009) confirms that they have reviewed the submission and determined the device is substantially equivalent to legally marketed predicate devices for the stated indications for use. This means that, in the context of this 510(k), the device indirectly meets "acceptance criteria" by being deemed as safe and effective as existing products, without requiring independent performance studies with acceptance criteria or ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1