Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251246

    Validate with FDA (Live)

    Device Name
    Sparrow Ascent
    Date Cleared
    2025-08-25

    (125 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

    Device Description

    The Sparrow Ascent is a transcutaneous auricular neurostimulation (tAN) system intended to provide non-invasive, transcutaneous stimulation of the nerves around the auricle (ear). The device is indicated as an aid in the reduction of opioid withdrawal symptoms in adult patients. The Sparrow Ascent is a battery operated, prescription device that delivers mild electrical stimulation to the nerves around the auricle (ear), which carry information to the central nervous system. The Sparrow Ascent is to be used in clinical environments (e.g., doctor's office, clinics, rehab centers, and hospitals) and/or at home. Users of the subject device include adults experiencing opioid withdrawal symptoms. Stimulation parameters (i.e., the strength of stimulation) are set by the user's clinician, and users can only adjust stimulation intensity (amplitude) at home. The system consists of three main components 1) a disposable Earpiece, 2) a Cable, and 3) the External Pulse Generator (EPG).

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to the Sparrow Ascent Transcutaneous Nerve Stimulator. The key aspect of this submission is a modification to the device's earpiece, specifically changing the stimulation site for the auricular branch of the vagus nerve (ABVN) from the cymba concha to the mastoid. The manufacturer aims to demonstrate that this change does not raise new questions of safety or effectiveness and that the modified device remains substantially equivalent to its predicate.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission are primarily based on demonstrating functional equivalence of the new mastoid stimulation site to the previously cleared cymba concha site, specifically in terms of brain activation patterns, and overall safety and efficacy as compared to the predicate and reference devices.

    Acceptance CriteriaReported Device Performance (from the document)
    Brain Activation Equivalence: Spatial distribution of brain activity following mastoid stimulation is equivalent to cymba concha stimulation of the ABVN.fMRI study showed activation maps for mastoid and cymba concha stimulation were significantly correlated (r = 0.61, p < 0.001). This indicates consistent central responses between the two stimulation sites and reliable stimulation-induced brain activation at the mastoid.
    Similar Current Density: The output related to nerve activation for the subject device is similar to the reference device."[T]he current density (i.e., the output related to nerve activation) is similar between the subject and the reference devices and helps demonstrate equivalent effectiveness." (No specific numerical range provided in the summary, but stated as similar.)
    Similar Power and Charge Density: The output related to stimulation site impact for the subject device is similar to the predicate and reference devices."[P]ower and charge density (i.e., the output related to the impact on the stimulation site) is similar between the subject, the predicate, and the reference devices, and helps demonstrate equivalent safety." (No specific numerical range provided in the summary, but stated as similar.)
    Non-clinical Performance: Device operates as expected, meets product requirements, delivers safe and effective stimulation."Non-clinical performance testing verified that, like the predicate, the subject device operates as expected when used as intended, and both meets and satisfies product requirements." "concluded that the subject device can safely and effectively deliver transcutaneous peripheral nerve stimulation when used as intended."
    Biocompatibility: Materials used remain biocompatible."In addition to testing performed to support the clearance of K230796 (e.g., biocompatibility), which remains applicable to the subject device..." (Implies original biocompatibility data sufficient and still valid).
    Service Life: Demonstrated 3-year service life."extended service life testing was completed for the subject device to verify a service life of 3 years."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Study): 24 healthy adults.
    • Data Provenance: The document does not explicitly state the country of origin. It describes a prospective clinical study (fMRI study) conducted to compare brain activation patterns.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • The document does not mention the use of external experts to establish ground truth for the fMRI study. The "ground truth" in this context is the objective physiological response (brain activation patterns) measured by the fMRI, which is interpreted by the study's researchers based on established neuroimaging analysis techniques (e.g., GLM).
    • The qualifications of the researchers/analysts interpreting the fMRI data are not specified in the provided text.

    4. Adjudication Method for the Test Set

    • No adjudication method is described for the fMRI study results. The analysis methods (e.g., General Linear Modeling, one-sample t-tests, correlation analysis) are standard scientific approaches for interpreting fMRI data, not an adjudication process by multiple human observers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was mentioned or performed. This submission is not evaluating human reader performance with or without AI assistance. Instead, it focuses on the equivalence of a technological characteristic (stimulation site) for a device that delivers nerve stimulation for opioid withdrawal, supported by objective physiological data (fMRI).

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance

    • This is not an AI/Software as a Medical Device (SaMD) submission in the typical sense that would require standalone algorithm performance evaluation. The device is a hardware stimulator. The fMRI study evaluates the physiological effect of the stimulation, which is a standalone effect of the device itself, not an algorithm's diagnostic performance.

    7. Type of Ground Truth Used

    • The primary "ground truth" for the clinical study was objective physiological data (brain activation patterns as measured by fMRI). This data was then analyzed statistically to demonstrate equivalence.

    8. Sample Size for the Training Set

    • Not applicable / Not explicitly mentioned. This submission describes a modification to a hardware device and an fMRI study to demonstrate functional equivalence. There is no mention of an AI algorithm or a training set for such an algorithm. The 24 healthy adults were part of a test (or validation) set for the fMRI study, not a training set for an AI model.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As no AI algorithm training set is mentioned, the method for establishing its ground truth is not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230796

    Validate with FDA (Live)

    Device Name
    Sparrow Ascent
    Date Cleared
    2023-06-20

    (90 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

    Device Description

    The Sparrow Ascent is a transcutaneous auricular neurostimulation (tAN) system intended to provide non-invasive, transcutaneous stimulation of the nerves on and/or around the auricle (ear). The device is indicated as an aid in the reduction of opioid withdrawal symptoms in adult patients. The Sparrow Ascent is a battery operated, prescription device that delivers mild electrical stimulation to the nerves on and/or around the auricle (ear), which carry information to the central nervous system. The Sparrow Ascent is to be used in clinical environments (e.g., doctor's office, clinics, rehab centers, and hospitals) and/or at home. Users of the subject device include adults experiencing opioid withdrawal symptoms. Stimulation parameters (i.e., the strength of stimulation) are set by the user's clinician, and users can only adjust stimulation intensity at home. The system consists of three main components 1) a disposable Earpiece, 2) a Cable, and 3) the External Pulse Generator (EPG).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Sparrow Ascent device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance letter does not explicitly state pre-defined acceptance criteria in terms of specific performance metrics (like sensitivity, specificity, or % reduction). Instead, the clinical study's primary efficacy endpoint served as the performance criterion for effectiveness.

    Acceptance Criteria (Primary Efficacy Endpoint)Reported Device Performance
    Successful mean percent change in COWS score (defined as a ≥15% reduction) from baseline, 60 minutes after the start of active tAN therapy.Successfully met. At the conclusion of Day 1 (120 minutes), 93.8% of participants had a clinically significant reduction in COWS score (≥15% reduction). This increased to 100% on Day 2 and was sustained across Day 5.
    Safety Endpoints: Prevalence of all adverse events (AEs), serious adverse events (SAEs), adverse device events (ADEs), serious adverse device effects (SADEs), unanticipated serious adverse device effects (USADEs), and device deficiencies.In 14 patients, who experienced a total of 943.6 hours of skin contact, there were no instances of adverse tissue reactions reported. The overall clinical study demonstrated acceptable safety profile, though specific AE rates are not detailed in this summary.
    Biocompatibility: Demonstrated compliance with ISO 10993 standards and supportive information from the RESTORE clinical study.Compliance with ISO 10993 series for cytotoxicity, skin sensitization, and irritation. RESTORE clinical study further supported biocompatibility with no adverse tissue reactions in 943.6 hours of skin contact among 14 patients.
    Software Verification: Compliance with IEC 62304 and ISO 14971.Explicitly states compliance with IEC 62304 and ISO 14971.
    Electromagnetic Compatibility and Electrical Safety: Compliance with ANSI/AAMI 60601-1, 60601-1-11, and 60601-1-2.Explicitly states compliance with ANSI/AAMI 60601-1, 60601-1-11, and 60601-1-2.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 23 participants were enrolled, and 20 received tAN therapy.
    • Data Provenance: The study was conducted at two US sites, making the data provenance United States and prospective (as it was a designed clinical study for this purpose).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of image interpretation or diagnosis. The ground truth for effectiveness was based on the Clinical Opiate Withdrawal Scale (COWS) scores, which are a standardized and objective measure of opioid withdrawal symptoms. These scores are typically assessed by trained clinical personnel. The document does not specify the number or qualifications of the personnel who administered the COWS scores in this study.

    4. Adjudication Method for the Test Set

    The document does not describe a traditional adjudication method for the test set data (e.g., 2+1, 3+1). The primary and secondary endpoints were based on changes in COWS scores, which are quantitative measures. The study was a double-blind, randomized, controlled study, which inherently helps to reduce bias in the assessment of outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a direct treatment device (nerve stimulator), not an AI-powered diagnostic tool for interpretation by human readers. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The device itself is a "standalone" therapeutic device (a transcutaneous nerve stimulator) that delivers stimulation. It does not involve an algorithm providing outputs for human interpretation. Its performance is assessed directly on patient outcomes (reduction in COWS scores). The "algorithm" in this context refers to the device's operational parameters, not an AI diagnostic algorithm. The study described assesses the device's direct therapeutic effectiveness.

    7. The Type of Ground Truth Used

    The ground truth for the effectiveness study was clinical symptom scores (COWS scores), which are a standardized and objective measure of opioid withdrawal severity. Safety ground truth was based on reported adverse events (AEs).

    8. The Sample Size for the Training Set

    The document refers to the clinical study of the predicate device (K201873) and "leverag[ing] prior clinical testing" for effectiveness, but it primarily details the study performed for the current device (Sparrow Ascent). It's not clear if there was a separate "training set" in the context of machine learning, as this is a medical device, not an AI algorithm. The study described appears to be the primary clinical evidence for the device's effectiveness.

    9. How the Ground Truth for the Training Set Was Established

    As this is a medical device and not an AI algorithm, the concept of a "training set" and establishing "ground truth for the training set" in the machine learning sense is not directly applicable. The effectiveness was established through a prospective, double-blind, randomized, controlled multi-center clinical study evaluating the impact of the nerve stimulator on COWS scores, as described in section 7 above. The summary also references the prior clearance of the predicate device (NSS-2 Bridge, DEN170018) which would have undergone its own clinical studies to establish similar effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201873

    Validate with FDA (Live)

    Date Cleared
    2021-01-02

    (179 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sparrow is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

    Device Description

    The Sparrow Therapy System is a non-invasive, battery-operated, prescription device designed to transcutaneously stimulate nerves on and/or around the auricle to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel. The system includes three components: Earpiece, Patient Controller, and the Clinician Application. Sparrow is used in clinical environments (i.e., rehab centers and hospitals) and at home. Users of the subject device include experiencing opioid withdrawal symptoms.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Key MetricAcceptance Criteria (Implied)Reported Device Performance (All Subjects, N=26)Reported Device Performance (Study Completers, N=14)
    COWS score percent reduction at 60 minutes≥15% reduction from baseline50.4%50.5%
    Percentage of patients who passed the naloxone challengeNot explicitly stated as acceptance criteria, but reported.10/26 (38.5%)10/14 (71.4%)
    Percentage of patients completing the studyNot applicable (outcome of study, not a performance metric)14/26 (53.8%)---
    Percentage of patients transitioning to MATNot explicitly stated as acceptance criteria, but reported.12/26 (46.2%)7/14 (50.0%)
    Prevalence of adverse events (AEs)Acceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis
    Prevalence of serious adverse events (SAEs)Acceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis
    Prevalence of adverse device events (ADEs)Acceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis
    Prevalence of serious adverse device effects (SADEs)Acceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis
    Prevalence of unanticipated serious adverse device effects (USADEs)Acceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis
    Prevalence of device deficienciesAcceptable safety profile (no explicit numeric threshold)Documented and considered in benefit-risk analysisDocumented and considered in benefit-risk analysis

    Primary Effectiveness Endpoint and Acceptance Criteria:
    The primary effectiveness endpoint of the clinical study was a successful mean percent change in COWS score (defined as a ≥15% reduction) from baseline to 60 minutes after the start of active tAN therapy. The reported performance of 50.4% reduction for all subjects and 50.5% for study completers demonstrates that the device met this primary acceptance criterion.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 26 participants (N=26) were enrolled in the study. 14 completed the study.
    • Data Provenance: Prospective. The study was conducted at one US site. The text indicates it was a "double-blind, randomized, prospective study."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth (COWS scores) for the test set. COWS (Clinical Opiate Withdrawal Scale) is a standardized tool, typically administered by trained clinical personnel. The study was conducted "under the supervision of trained clinical personnel," implying these professionals would have been responsible for assessing COWS scores.

    4. Adjudication Method for the Test Set:

    The document does not explicitly detail an adjudication method for the COWS scores or other assessments used in the test set. Given that COWS is a clinical assessment, it's typically performed directly by clinicians during the study visits.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device, the Sparrow Therapy System, is a transcutaneous nerve stimulator, not an AI-based diagnostic or assistive software that involves human readers interpreting images or data. Therefore, the concept of human readers improving with AI assistance is not applicable to this device type.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The Sparrow Therapy System is a physical medical device (nerve stimulator) that delivers therapy. It operates independently as intended once activated, but its use is "under the supervision of trained clinical personnel." It is not an algorithm or software-only device where "standalone" performance in the absence of human input would be a relevant metric in the same way it would be for an AI diagnostic tool. However, the effectiveness study (reduction in COWS scores) represents the "standalone" performance of the device in its intended clinical setting in achieving the therapeutic effect.

    7. The Type of Ground Truth Used:

    The primary ground truth used for effectiveness was the Clinical Opiate Withdrawal Scale (COWS) score. This is a clinical assessment tool used to quantify the severity of opioid withdrawal symptoms.

    8. The Sample Size for the Training Set:

    The document does not mention a separate "training set" in the context of device development or clinical validation for the Sparrow Therapy System. This suggests that the clinical study described (with N=26 participants) served as the primary means of evaluating the device's effectiveness. For medical devices, particularly physical ones, a "training set" in the sense of machine learning algorithms is often not applicable unless the device incorporates complex adaptive algorithms. The device's design and stimulation parameters were likely developed through engineering and preclinical testing.

    9. How the Ground Truth for the Training Set Was Established:

    As no specific "training set" is mentioned in the context of clinical data for the device, this question is not directly applicable. If the device incorporates an adaptive algorithm, the methods for establishing ground truth for any data used to train such an algorithm would need to be described, but this information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1