Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K240510

    Validate with FDA (Live)

    Date Cleared
    2024-11-08

    (260 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx INDICATIONS: Under the supervision of a healthcare professional, Microdacyn Hydrogel is intended for management of wounds associated with dermal irritation, sores, injuries and ulcers of dermal tissue. Microdacyn Hydrogel is intended for use on first and second degree burns, exuding wounds such as leg ulcers, diabetic ulcers, and for the management of mechanically or surgically debrided wounds.

    OTC INDICATIONS: Microdacyn Hydrogel is intended for use on minor skin irritations, lacerations, abrasions and minor burns. Microdacyn Hydrogel is also indicated for the management of irritation from minor sunburn.

    Device Description

    Microdacyn Hydrogel is a non-sterile, aqueous hydrogel that contains sodium hypochlorite and hypochlorous acid as preservatives to prevent the growth of microorganisms within the container during shelf-life.

    AI/ML Overview

    The provided text is a 510(k) summary for the Microdacyn Hydrogel, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a study to prove acceptance criteria in the manner you've outlined for an AI/algorithm-based device.

    The information you are asking for, such as sample size for test sets, data provenance, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set details, are typically relevant for AI/ML-driven diagnostic or prognostic devices.

    The Microdacyn Hydrogel is a wound dressing that relies on its physical and chemical properties for its intended use, not on complex data analysis or AI algorithms. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are addressed through a comparison with a predicate device and standard performance and biocompatibility testing.

    Here's an attempt to structure the available information per your request, with the understanding that many fields will be "N/A" due to the nature of the device:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from comparison with predicate)Reported Device Performance (from Performance Data section)
    Intended Use: - Rx Management of wounds associated with dermal irritation, sores, injuries, ulcers of dermal tissue.- Rx Use on first and second-degree burns, exuding wounds (leg ulcers, pressure ulcers, diabetic ulcers), and mechanically/surgically debrided wounds.- OTC Use on minor skin irritations, lacerations, abrasions, minor burns.- OTC Management of irritation from minor sunburn.Determined to have the same intended use as the predicate device (Microcyn Skin and Wound HydroGel (K093585)). The Indications for Use are almost identical, lacking only the explicit mention of "itch and pain relief" from the predicate. The FDA clearing the 510(k) indicates acceptance of this proposed use.
    Technological Characteristics:- Where Used: RX Only; OTC- Delivery System: Hydrogel- Mechanism of Action: Maintaining a moist wound environment.- Composition: Water (96.524%), Sodium Magnesium Fluorosilicate (3.000%), Sodium Phosphate (0.400%), Sodium Chloride (0.066%), Hypochlorous acid (0.008%), Sodium Hypochlorite (0.002%)- Sterilization: Non-sterile- Shelf Life: Acceptable shelf life- Specification: Viscosity- Container closure system: 50 mL, 100 mL, 200 mL PET bottles with PP Snap-top screw cap and lift 'n' peel liner.Determined to have the same technological characteristics as the predicate device.- Where Used: Same as predicate.- Delivery System: Same as predicate.- Mechanism of Action: Same as predicate.- Composition: Same as predicate.- Sterilization: Same as predicate.- Shelf Life: 24 months (longer than predicate's 18 months, which is generally acceptable if proven stable).- Specification: Viscosity: 12,000-20,000 cP (same as predicate's specification according to the table, though the predicate's value isn't explicitly stated).- Container closure system: Expanded options (50mL, 100mL, 200mL) compared to predicate's 2oz, but within acceptable parameters for hydrogel.
    Safety and Effectiveness: - Biocompatibility- Antimicrobial EffectivenessBiocompatibility Testing: - Met ISO-10993-1 Biological Evaluation of Medical Devices. Performance Testing: - Met USP Antimicrobial Effectiveness Testing <51>.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable in the context of an "AI test set." The performance data is derived from standard laboratory testing for biocompatibility and antimicrobial effectiveness, not from human subject data or a diagnostic test set.
    • Data Provenance: Not applicable in the context of clinical data provenance. The tests are laboratory-based, adhering to ISO and USP standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth, in the AI/ML sense, is not established for this type of device. The "ground truth" for the performance data comes from the established scientific standards of ISO-10993-1 and USP <51>.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are relevant for human interpretation or clinical endpoint assessments, not for laboratory performance tests of a hydrogel.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a hydrogel wound dressing, not an AI-assisted diagnostic or imaging device that would involve human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no algorithm to assess.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the AI/ML context. For this device, the "ground truth" is defined by adherence to established international and pharmacopeial standards for biocompatibility (ISO-10993-1) and antimicrobial effectiveness (USP <51>).

    8. The sample size for the training set:

    • Not applicable. This device does not involve a "training set" in the machine learning sense.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no training set for this device.

    In summary: The Microdacyn Hydrogel is a non-AI/ML device. Its acceptance criteria are met by demonstrating substantial equivalence to a legally marketed predicate device (Microcyn Skin and Wound HydroGel) through a comparison of intended use and technological characteristics, supplemented by standard laboratory-based performance (biocompatibility and antimicrobial effectiveness) testing, rather than studies involving human data sets, expert review, or AI algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233399

    Validate with FDA (Live)

    Date Cleared
    2024-09-13

    (345 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx Indications for Use: Under the supervision of a healthcare professional, Microdacyn Wound Care Solution is intended for the cleansing, irrigation, moistening, debridement and removal of foreign material and debris from exudating wounds, acute and chronic dermal lesions including stage I-IV pressure ulcers, diabetic ulcers, post-surgical wounds, first- and second-degree burns, abrasions, minor irritations of the skin, diabetic foot ulcers, ingrown toe nails, grafted/donor sites and exit sites. It is also intended for use to moisten and lubricate wound dressings and for use with devices intended to irrigate wounds.

    OTC Indications for Use: Microdacyn Wound Care Solution is intended for OTC management of minor skin abrasions, minor lacerations, minor irritations and intact skin of the face, eyelid and eyelashes.

    Device Description

    Microdacyn Wound Care Solution is a clear solution that aids in the mechanical removal of debris and foreign material from the application site. The solution contains sodium hypochlorite and hypochlorous acid as preservatives to reduce/prevent the growth of microorganisms within the solution.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification summary for a medical device called "Microdacyn Wound Care Solution." It categorizes the device as unclassified and uses predicate devices to establish substantial equivalence.

    The document does not describe acceptance criteria or a study proving the device meets the acceptance criteria in the manner one would typically expect for an AI/ML medical device submission (e.g., performance metrics like sensitivity, specificity, AUC for diagnostic accuracy). Instead, it focuses on demonstrating substantial equivalence to legally marketed predicate devices based on technological characteristics, indications for use, safety, and performance.

    Therefore, I cannot provide the information requested in the prompt (acceptance criteria table, sample size, expert ground truth, MRMC study, standalone performance, etc.) because the provided text does not contain this type of data or study design. The "performance testing" mentioned is an "Antimicrobial Effectiveness Test" (USP <51>), which is a microbiological test, not a clinical performance study measuring accuracy or efficacy in the context of disease diagnosis or treatment effectiveness in a human population.

    Here's why the requested information isn't present in this type of submission:

    • Device Type: Microdacyn Wound Care Solution is a physical chemical product (a wound cleanser), not an AI/ML diagnostic or prognostic device.
    • Regulatory Pathway: A 510(k) submission primarily demonstrates substantial equivalence to a predicate device. For devices like this, the "acceptance criteria" are based on comparing characteristics like composition, mechanism of action, intended use, and safety, rather than statistical performance metrics from a clinical trial.
    • "Study" Definition: The "studies" mentioned (Biocompatibility and Antimicrobial Effectiveness Tests) are laboratory tests to ensure safety and product integrity, not clinical trials to assess diagnostic accuracy or the impact of an AI algorithm on human reader performance.

    If this were an AI/ML device submission, the requested information would be crucial. However, for this specific 510(k) for a wound care solution, the provided text contains the relevant information for its regulatory pathway.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173896

    Validate with FDA (Live)

    Date Cleared
    2018-04-02

    (101 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx: Under the supervision of a health care professional Microcyn Antimicrobial Hydrogel is intended for the management of post non ablative laser therapy procedures, post microdermabrasion therapy and following superficial chemical peels. Microcyn Antimicrobial Hydrogel may also be used to relieve itch and pain from minor skin irritations, lacerations, abrasions and minor burns.

    OTC: Microcyn Antimicrobial Hydrogel is intended for the management of minor skin irritations following post non ablative laser therapy procedures, microdermabrasion therapy or superficial chemical peels. Microcyn Antimicrobial Hydrogel may also be used to relieve itch and pain from minor skin irritations, abrasions and minor burns.

    Device Description

    The Microcyn Antimicrobial Hydrogel is a clear hydrogel, slightly chlorinated odor, viscous gel. The product has a pH range of 6.2-7.8 and a viscosity target of >1000 cP and will be supplied in polyethylene terephthalate (PET) tube-bottles with polypropylene (PP) screw-top closure.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Microcyn Antimicrobial Hydrogel. The core of the submission is to demonstrate substantial equivalence to predicate devices, rather than establishing de novo performance criteria against a disease state or clinical outcome. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the traditional sense of clinical performance metrics are not directly applicable here.

    However, we can interpret the "acceptance criteria" as the required testing to demonstrate substantial equivalence and "reported device performance" as the results of those tests confirming equivalence to the predicates.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Required Testing for Substantial Equivalence)Reported Device Performance (Results)
    Biocompatibility Testing:
    * CytotoxicityEvaluated based on FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Implied to meet acceptance for substantial equivalence, as the conclusion states the device is substantially equivalent.)
    * Systemic toxicityEvaluated based on FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Implied to meet acceptance for substantial equivalence.)
    * IrritationEvaluated based on FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Implied to meet acceptance for substantial equivalence.)
    * SensitizationEvaluated based on FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Implied to meet acceptance for substantial equivalence.)
    * Toxicological risk assessment (for prolonged use)Provided to support prolonged use. (Implied to be acceptable for substantial equivalence.)
    Bench Testing:
    * Package integrityPerformed. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    * Visual inspectionPerformed. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    * ViscosityPerformed. Reported target: >1000 cP. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    * pHPerformed. Reported range: 6.2-7.8. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    * Free Available Chlorine (FAC)Performed. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    * Preservative Effectiveness Testing (in gel while on shelf)Performed. Conforming to USP <51>. (Microcyn Antimicrobial Hydrogel meets specification and performance characteristics and is substantially equivalent to the predicate device.)
    Comparison to Predicate Devices:
    * Indications for UseRx Indication: Proposed device's Rx indication is identical to the K153648 predicate. OTC Indication: Proposed device's OTC indication is identical to the K153648 predicate. (Differences noted with K171935 predicate, but found to be substantially equivalent overall, implying the differences are not significant enough to preclude equivalence.)
    * Sterility Claim"Non-sterile, Preserved/Conforming to USP <51>". Identical to K153648 and K171935 predicates.
    * Mechanism of Action"Wound moisturizer". Identical to K153648 and K171935 predicates.
    * Delivery System"Hydrogel". Identical to K153648 and K171935 predicates.
    * Ingredients (Qualitative and Quantitative)Proposed device and K153648 predicate have identical ingredients. K171935 predicate has additional ingredients (dimethicone, sodium bicarbonate, sodium sulfate) which are not found in the proposed device or K153648. Despite these differences, the conclusion states substantial equivalence, implying these differences do not raise new questions of safety or effectiveness.

    Since this is a 510(k) submission proving substantial equivalence for a medical device (a wound hydrogel), not a diagnostic algorithm, several of your requested points related to AI/algorithm performance studies, ground truth establishment, expert consensus, and reader studies are not applicable to this document. The document implicitly states that these tests demonstrate that the device is as safe and effective as the predicate devices, thereby meeting the "acceptance criteria" for 510(k) clearance.

    Here's why the other points are not applicable or cannot be answered from the provided text:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable / Not Provided: This device is a hydrogel, not a diagnostic algorithm tested on a dataset of patient cases. The "test sets" here would be samples of the hydrogel subject to physical, chemical, and biological testing as per standards. The document does not specify the number of samples used for each test. Data provenance (country of origin, retrospective/prospective) is not relevant for this type of product testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable: There is no "ground truth" in the context of clinical disease states or image interpretation for a wound hydrogel being cleared via substantial equivalence. The "truth" is established by adherence to recognized standards for biocompatibility and bench performance, and by comparison to the established predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic devices, to establish a definitive diagnosis or reference standard. This is not relevant for the physical, chemical, and biological testing of a hydrogel.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This product is a hydrogel, not an AI-powered diagnostic device or an AI assistant for human readers. Therefore, MRMC studies and the concept of "human readers improving with AI" are entirely irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable: As explained in point 3, there is no "ground truth" in the clinical diagnostic sense for this device. The "ground truth" for its performance is compliance with established physical, chemical, and biological specifications and equivalence to predicates already on the market.

    8. The sample size for the training set

    • Not Applicable: This device does not involve a "training set" as it is not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not Applicable: As explained in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171935

    Validate with FDA (Live)

    Date Cleared
    2017-12-01

    (156 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx Indication: Alevicyn SG Antimicrobial Gel is intended to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions. Alevicyn SG Antimicrobial Gel is indicated to manage and relieve the burning, itching, erythema, scaling, and pain experienced with various types of dermatoses, including atopic dermatitis, radiation dermatitis, seborthea and sebortheic dermal irritation, sores, injuries, and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. The Alevicyn SG Antimicrobial Gel may also be used to manage partial or full thickness wounds such as first and second degree burns, stage I - IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), and trauma wounds. Alevicyn SG Antimicrobial Gel helps to relieve dry waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process.

    OTC Indications for Use: Alevicyn SG Antimicrobial Gel is intended for OTC use to relieve itch and pain from minor skin irritations, minor cuts, exit sites, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process.

    Device Description

    The Alevicyn SG Antimicrobial Gel is an opaque gel, slightly chlorinated odor, low viscosity spray gel. The product will be supplied in polyethylene terephthalate (PET) round-bottles with polypropylene (PP) screw-top closure and a finger pump sprayer. Ingredients: Water, sodium lithium magnesium fluorosilicate, dimethicone, sodium phosphate, sodium bicarbonate, sodium chloride, sodium sulfate, with hypochlorous acid and sodium hypochlorite as gel preservatives.

    AI/ML Overview

    The document provided describes a 510(k) Premarket Notification for the Alevicyn SG Antimicrobial Gel. This type of submission aims to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, rather than proving its safety and effectiveness from scratch. Therefore, the information provided focuses on showing equivalence through comparison and specific performance data, rather than detailing a direct study to "prove acceptance criteria" in the way one might for a novel device.

    However, I can extract the acceptance criteria (specifications) and reported performance data based on the provided "Performance Data" section.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Specification)Reported Device Performance
    Biocompatibility:
    CytotoxicityMet (Evaluated in accordance with FDA guidance and ISO 10993-1)
    IrritationMet (Evaluated in accordance with FDA guidance and ISO 10993-1)
    SensitizationMet (Evaluated in accordance with FDA guidance and ISO 10993-1)
    Toxicological RiskMet (Toxicological risk assessment conducted to support prolonged use)
    Bench Testing:
    Visual InspectionMeets specification (Opaque, colorless semi-viscous gel, slightly chlorinated odor)
    pHMeets specification
    ViscosityMeets specification
    Free Available Chlorine (FAC)Meets specification
    Antimicrobial Preservative EffectivenessMeets specification (Effective against Proteus mirabilis, Serratia marcescens, antibiotic resistant Methicillin-Resistant Staphylococcus aureus (MRSA), and Acinetobacter baumannii while on the shelf, conforming to USP <51>)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for any of the biocompatibility or bench tests. It also does not specify the country of origin or whether the data was retrospective or prospective. The nature of these tests (e.g., in-vitro toxicology, chemical property testing) typically involves lab-based experiments rather than human clinical trials with "test sets" in the traditional sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided in the document. For device submissions based on predicate equivalence and bench/biocompatibility testing, the "ground truth" is established by adherence to recognized standards (e.g., ISO 10993-1, USP <51>) and internal specifications for chemical and physical properties. There are no "experts" establishing ground truth in the context of clinical interpretation, as would be the case for an AI diagnostic device.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where expert consensus is needed to resolve discrepancies in diagnoses or assessments, particularly for AI image interpretation. The studies described here are laboratory-based performance and safety evaluations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size with AI vs. Without AI Assistance

    This information is not applicable and not provided. MRMC studies are relevant for evaluating diagnostic imaging systems (often AI-powered) where human readers interpret cases. The Alevicyn SG Antimicrobial Gel is a topical medical device, not an imaging or diagnostic AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    This information is not applicable and not provided. "Standalone performance" refers to the performance of an algorithm without human intervention, again typically for AI-driven diagnostic tools. This device is not an algorithm.

    7. The Type of Ground Truth Used

    For biocompatibility, the ground truth is established by:

    • Adherence to recognized international standards like ISO 10993-1 ("Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process").
    • FDA biocompatibility guidance (issued June 16, 2016).

    For bench testing, the ground truth is established by:

    • Internal specifications defined for pH, viscosity, FAC.
    • Adherence to recognized pharmacopeial standards like USP <51> for antimicrobial preservative effectiveness.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. This device is a physical product (gel) and does not involve AI or machine learning models that require training sets.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided, for the same reason as above (no AI/ML training set).

    Ask a Question

    Ask a specific question about this device

    K Number
    K172851

    Validate with FDA (Live)

    Date Cleared
    2017-11-15

    (56 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx Indications: Under the supervision of a healthcare professional, Microcyn Plus Antimicrobial Wound Care Solution is intended for the cleansing, irrigation, moistening, debridement and removal of foreign material and debris from exudating wounds, acute and chronic dermal lesions including stage I-IV pressure ulcers, diabetic ulcers, post-surgical wounds, first- and second-degree burns, abrasions, minor irritations of the skin, diabetic foot ulcers, ingrown toe nails, grafted/donor sites and exit sites. It is also intended for use to moisten and lubricate wound dressings and for use with devices intended to irrigate wounds.

    OTC Indications for Use: Microcyn Plus Antimicrobial™ Wound Care Solution is intended for OTC use in the management of skin abrasions, lacerations, minor irritations, cuts, and intact skin.

    Device Description

    The Microcyn Plus Antimicrobial Wound Care Solution is a clear hypotonic solution that aids in the mechanical removal of debris and foreign material from the application site. It is also for the management of skin abrasions, lacerations, irritation, cuts, and wounds. The solution is supplied in glass bottles with lined caps with a pH range of 3.5 - 6.0. Ingredients: water, sodium chloride, hypochlorous acid.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Microcyn Plus Antimicrobial Wound Care Solution, which evaluates its substantial equivalence to predicate devices. This type of submission relies on demonstrating similarity to already legally marketed devices rather than conducting a full clinical study to prove efficacy, especially for Class I or II devices where such extensive studies might not be required.

    Therefore, the information regarding acceptance criteria and performance is focused on bench testing and biocompatibility to prove the device's similarity and safety, rather than a clinical trial with human subjects, AI performance metrics, or complex ground truth establishment.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" against "reported device performance" in the way one might expect for an AI/diagnostic device. Instead, the performance data provided are primarily for bench testing to ensure the new device functions similarly to the predicate devices and meets safety standards.

    Acceptance Criteria (Implicit)Reported Device Performance
    Biocompatibility:Conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Specific results are not detailed, but compliance is stated).
    Package Integrity:Bench tested. (Results are not detailed).
    Visual Inspection:Bench tested. (Results are not detailed).
    Osmolality:Bench tested. (Results are not detailed).
    pH:Bench tested. (pH range is 3.5 - 6.0, as stated in Device Description).
    Free Available Chlorine (FAC):Bench tested. (Results are not detailed).
    Preservative Effectiveness (USP <51>):Used to support preservative effectiveness with compendia organisms and specific bacteria: Proteus mirabilis, Serratia marcescens, MRSA, VRE, and Acinetobacter baumannii. (Results are not detailed, but effectiveness is supported).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable or not provided in the context of this 510(k) summary. The testing described is laboratory-based bench testing and biocompatibility assessments, not a clinical trial with a "test set" of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable or not provided. Since there isn't a "test set" of clinical data requiring expert review for ground truth, this aspect is not part of the submission. The ground truth for chemical and physical properties comes from established scientific methods and readings.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable or not provided. There is no clinical test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a wound care solution, not an AI-assisted diagnostic tool, so an MRMC study is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the tests performed is based on established laboratory standards, chemical analysis, and microbiological testing methodologies. For example:

    • Biocompatibility: Adherence to ISO standards and FDA guidelines.
    • pH, Osmolality, FAC: Instrumental measurements against known chemical standards.
    • Preservative Effectiveness: Standardized microbiological challenge tests (USP <51>) against recognized microbial strains.

    8. The sample size for the training set

    This information is not applicable or not provided. This device is not an AI/machine learning product and therefore does not have a "training set" in that context. The development and formulation of the solution would involve iterative testing, but not in the sense of a machine learning training set size.

    9. How the ground truth for the training set was established

    This information is not applicable or not provided. As above, there is no "training set" for an AI algorithm. The development of the solution would rely on scientific principles and established chemical and biological testing methods for quality control and formulation, rather than a labeled "ground truth" for training an algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1