Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K193058

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-15

    (196 days)

    Product Code
    Regulation Number
    868.5730
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SonarMed AirWave Airway Monitoring System is used to assist in verifying placement of the endotracheal tube (ETT), to assist in detecting movement of the ETT tip, to assist in detecting obstruction of the ETT, and to assist in listening to breath sounds.

    The SonarMed AirWave Airway Monitoring System is intended for use by qualified personnel to assist with artificial airway management for patients in an in-hospital setting (intensive care, operating room, and emergency department settings, as well as intra-hospital transport).

    The SonarMed AirWave Airway Monitoring System is to be used as an adjunct to normal clinical practice and is a not a stand-alone diagnostic system.

    It is intended for use with neonates, infants, children, adolescents, and adults (sizes 2.5 mm to 9.0 mm).

    Device Description

    The SonarMed AirWave is comprised of a SonarMed Monitor (Monitor) that is used in conjunction with a single-patient use SonarMed Sensor (Sensor) and software that operates the Monitor and Sensor. The Monitor is powered from an external power supply and has a battery backup. When in use, the SonarMed Sensor is placed in-line between the ventilator circuit and the proximal end of the endotracheal tube (ETT) of a patient who is connected to a ventilator.

    Using acoustic reflection technology, signals from the Sensor are displayed on the Monitor showing the clinician:

    • The baseline location of the ETT tip as established by the clinician
    • Estimation of passageway around the tipof the ETT, relative to the . ETT diameter
    • ETT movement relative to the baseline location ●
    • . ETT occlusion / obstruction information including percent obstructed and location of the obstruction
    • The clinician can choose whether to view information about the . patient's airway in either a waveform or graphic on the Monitor's LCD. Additionally, the clinician can use the microphones to listen to breath sounds. The information provided by the device is to be used in conjunction with normal clinical practice to assist with management of the artificial airway of the patient.
    AI/ML Overview

    The provided text describes the SonarMed AirWave Airway Monitoring System and its substantial equivalence determination by the FDA. However, it does not explicitly detail the acceptance criteria and a study proving the device meets these criteria in the typical quantitative manner (e.g., sensitivity, specificity, AUC values) usually associated with AI/ML device performance.

    The document focuses on demonstrating substantial equivalence to a predicate device (K143042) through various bench tests and by addressing modifications. The "acceptance criteria" appear to be implicit in the comparative performance against the predicate device, aiming for "as accurate or more accurate."

    Based on the provided text, here's an attempt to answer your questions to the extent possible, acknowledging the limitations of the information provided for a typical AI/ML device performance study:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't provide explicit quantitative acceptance criteria or corresponding reported performance metrics (like percentage accuracy thresholds) in a tabular format. Instead, it describes the conclusion of the bench testing:

    Performance AspectImplied Acceptance Criterion / ComparisonReported Device Performance
    Passageway Detection AccuracyAccurately estimates diameter of passageway around ETT tip (compared to predicate)"The Passageway Detection Study documented that the subject device accurately estimates the diameter of the passageway around the tip of the ETT for all patient populations."
    ETT Movement DetectionAccurately detects distance and direction of ETT tip movement (compared to predicate)"The Movement Study documented that the subject device accurately detects the distance and direction of the ETT tip movement both upward and downward for all patient populations."
    Obstruction DetectionAccurately detects percentage of obstruction in ETT (compared to predicate)"The Obstruction Detection Study documented that the subject device accurately detects the percentage of obstruction in the ETT for all patient populations."
    Overall PerformanceAs accurate or more accurate than the predicate.Stated under the "Performance" section of the comparison table: "As accurate or more accurate than predicate."

    2. Sample size used for the test set and the data provenance

    The document refers to "bench testing" and "studies" (Passageway Detection Study, Movement Study, Obstruction Detection Study) but does not specify the sample sizes used for these tests.
    The data provenance is not mentioned. It is implied these are pre-clinical/bench test data, not clinical patient data, given the description of testing on the device itself. Therefore, concepts like "country of origin of the data," "retrospective," or "prospective" as applied to patient cohorts are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts or their qualifications for establishing ground truth. Given that the testing is described as "bench testing" on the device's accuracy for physical measurements (ETT movement, obstruction, passageway diameter), it's more likely that the "ground truth" was established by calibrated instruments or known physical alterations to the test setup, rather than human expert interpretation of images/data.

    4. Adjudication method for the test set

    The document does not mention any adjudication method, as it does not describe a study involving multiple human readers or interpretations. The "studies" appear to be objective measurements from the device against a known physical ground truth during bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study involving human readers is not mentioned in the provided text. The evaluation focuses on the device's intrinsic measurement capabilities in a bench test setting, not on how human readers' performance with or without the device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The "bench testing" described appears to be a standalone performance evaluation of the device's sensors and algorithms, given that it aims to demonstrate the device "accurately detects" or "accurately estimates" various parameters.

    7. The type of ground truth used

    The ground truth used appears to be objective, known physical parameters established by the bench test setup for ETT tip location, obstruction percentage, and airway diameter. This is inferred from the description of "Passageway Detection Study," "Movement Study," and "Obstruction Detection Study." It is not based on expert consensus, pathology, or outcomes data in a clinical sense.

    8. The sample size for the training set

    The document does not mention a training set size. This device relies on acoustic reflection technology and embedded processing, which are described as being similar to the predicate device. It's not explicitly stated as an AI/ML device that requires a large, dedicated "training set" in the common sense of deep learning. Updates were made to software (e.g., O2 ranges, waveform screen functionality), but the underlying "technology" (acoustic reflectometry) remains the same as the predicate.

    9. How the ground truth for the training set was established

    As no training set is explicitly mentioned for a machine learning model, the method for establishing ground truth for such a set is not described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143042

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-22

    (457 days)

    Product Code
    Regulation Number
    868.5730
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SonarMed AirWave Airway Monitoring System is used to assist in verifying placement of the endotracheal tube (ETT), to assist in detecting movement of the ETT tip, to assist in detecting obstruction of the ETT, and to assist in listening to breath sounds.

    The SonarMed AirWave Airway Monitoring System is intended for use by qualified personnel to assist with artificial airway management for patients in an in-hospital setting (intensive care, operating room, and emergency department settings, as well as intra-hospital transport).

    The SonarMed AirWave Airway Monitoring System is to be used as an adjunct to normal clinical practice, and is not to be used as a stand-alone diagnostic system.

    It is intended for use with patients who use ET tube sizes 2.5, 3.0, 3.5, and for sizes 6.5, 7.0, 7.5, 8.0, 8.5, and 9.0 mm.

    Device Description

    The SonarMed AirWave is comprised of a SonarMed Monitor (Monitor) that is used in conjunction with a single-patient use SonarMed Sensor (Sensor) and software that operates the Monitor and Sensor. The Monitor is powered from an external power supply and has a battery backup. When in use, the SonarMed Sensor is placed in-line between the ventilator circuit and the proximal end of the endotracheal tube (ETT) of a patient who is connected to a ventilator.

    Using acoustic reflection technology, signals from the Sensor are displayed on the Monitor showing the clinician:
    • The baseline location of the ETT tip as established by the clinician
    • Quantification of diameter of the anatomical structure around the tip of the ETT
    • ETT movement relative to the baseline location

    • ETT occlusion / obstruction information including percent obstructed and location of the obstruction
      The clinician can choose whether to view information about the patient's airway in either a waveform or graphic on the Monitor's LCD. Additionally the clinician can use the microphones to listen to breath sounds. This information should only be used in an adjunctive manner to assist with management of the artificial airway of the patient.

    The monitor allows the user to manually adjust the sound of speed. This is to compensate for the changes in the sound of speed with high oxygen concentrations and the use of anesthesia gases. The monitor also has an auto-speed of sound mode. This mode automatically changes the speed of sound.

    The monitor allows the user to manually enter the distance to the carina from the chest x-ray. This is an optional feature that the user can elect from the setting menu. Once selected the algorithm will calculate the distance to the carina based on the movement of the tip of the tube.

    AI/ML Overview

    The provided text describes the SonarMed™ AirWave Airway Monitoring System (K143042) and its conformity with a predicate device through non-clinical testing. It focuses on demonstrating substantial equivalence to a previously cleared device (SonarMed Airway Management System K092611).

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a quantitative format typical for clinical trials or performance goals. Instead, the performance is demonstrated through comparisons to the predicate device and verification of intended function in various tests. The key performance aspects evaluated are:

    Acceptance Criteria (Implied)Reported Device Performance
    ETT Tip Movement & Passageway Size Accuracy (Infant vs. Adult Size Sensor)The new, infant sized sensors are at least as accurate as the predicate, adult sized sensors in detecting movement and passageway size. This was concluded from bench testing and a pre-clinical animal study. The pre-clinical study concluded that the tracking algorithm is accurate in determining ETT tip movement and passageway size.
    ETT Obstruction Accuracy (Infant vs. Adult Size Sensor)The new, infant sized sensors are at least as accurate as the predicate, adult sized sensors in detecting obstruction. This was concluded from bench testing.
    Functionality during Prolonged Use (Heated, Humidified, Positive Pressure Ventilation)The SonarMed AirWave Sensor functions as intended during prolonged use for over 30 days. This was demonstrated in a long-term bench study.
    BiocompatibilityTesting for cytotoxicity, sensitization, and intracutaneous reactivity all passed. Exhaustive E&L (extractable and leachable) testing determined compounds were present at levels well below health consequences for the neonatal population with large safety margins.
    Safety and Effectiveness Equivalence to PredicateThe conclusions drawn from the nonclinical testing demonstrate that the device is as safe, as effective, and performs as well as the predicate device. This is the overall conclusion of the substantial equivalence determination for non-clinical aspects.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The text mentions "Several bench tests were performed" and "A long-term bench study was conducted." It compares "movement and passageway accuracy of the infant sized sensor to the predicate, adult sized sensor" and "obstruction accuracy of the infant sized sensor to the predicate, adult sized sensor." However, specific sample sizes for the bench tests (e.g., number of ETTs, number of tests run) are not provided.
    • Pre-clinical Animal Study: "The study utilized an animal model (rabbit)." The number of rabbits used in the study is not specified.
    • Data Provenance: The studies were conducted by SonarMed, Inc. and are described as non-clinical bench and pre-clinical animal studies. The text does not refer to human data or data from a specific country of origin in terms of patient population.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of human experts to establish ground truth for the test sets. The ground truth in the bench and animal studies would have been established through controlled experimental setups (e.g., precisely measured ETT movements, intentionally created obstructions, known ETT tip positions).

    4. Adjudication Method for the Test Set

    Since human expert assessment for ground truth is not mentioned, an adjudication method for test set ground truth is not applicable or described in this context. The "adjudication" would be inherent in the controlled experimental design of the bench and animal studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or described. The device is an airway monitoring system for direct measurement, not an AI-assisted diagnostic imaging interpretation tool that would involve "human readers" interpreting "cases."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance described (accuracy in determining ETT tip movement, passageway size, and obstruction) is the standalone performance of the device's sensing and algorithm. The device is intended to provide information to a clinician, but the accuracy studies themselves evaluated the device's ability to measure these parameters directly. It is stated: "The SonarMed AirWave Airway Monitoring System is to be used as an adjunct to normal clinical practice, and is not to be used as a stand-alone diagnostic system." This statement regarding its clinical use as an adjunct does not negate that the performance evaluation of its measurements (as described in the non-clinical tests) is essentially its standalone accuracy.

    7. The Type of Ground Truth Used

    • Bench Testing: The ground truth for bench testing would be known, controlled, and precisely measured physical parameters (e.g., exact movement distances, known obstruction percentages, measured passageway sizes) created within the experimental setup.
    • Pre-clinical Animal Study: For the animal study, the ground truth for ETT movement would be based on known physical manipulations of the ETT within the rabbit trachea, bronchus, and esophagus, likely verified by external measurement or direct observation during the procedure.

    8. The Sample Size for the Training Set

    The document does not mention a concept of a "training set" as would be relevant for machine learning or AI algorithm development in the traditional sense. The device appears to be based on more traditional acoustic reflection technology and algorithms, which likely don't involve a distinct "training set" in the way a deep learning model would. If an algorithm was "trained," the details are not provided.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092611

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-28

    (245 days)

    Product Code
    Regulation Number
    868.5730
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SonarMed™ AMS is used to assist in verifying placement of the endotracheal tube (ETT), to assist in detecting movement of the ETT tip, and to assist in detecting obstruction of the ETT.

    The SonarMed AMS is intended for use by qualified personnel to assist with artificial airway management for patients in an in-hospital setting (intensive care, operating room, and emergency department settings, as well as intra-hospital transport).

    The SonarMed AMS is to be used as an adjunct to normal clinical practice, and is not to be used as a stand-alone diagnostic system.

    It is intended for use with patients who use ET tube sizes from 6.5 mm to 9.0 mm, with a body weight of >35 kg.

    Device Description

    The SonarMed AMS is comprised of a SonarMed Monitor (Monitor) that is used in conjunction with a sterile, single-use SonarMed Adapter (Adapter) and software that operates the Monitor and Adapter. The Monitor is powered from an external power supply and has a battery backup. When in use, the SonarMed Adapter is placed in-line between the ventilator circuit and the proximal end of the endotracheal tube (ETT) of a patient who is connected to a ventilator.

    Using sonic reflection technology, signals from the Adapter allow the color LCD on the Monitor to display to a clinician:
    The baseline location of the ETT tip as established by the clinician
    Quantification of passageway diameter relative to the ETT diameter (for example a reading of less than 1 may indicate the tube is in a passageway that is smaller than itself, such as the esophagus)
    ETT movement relative to the baseline location
    ETT occlusion / obstruction information including percent obstructed

    The clinician can choose whether to view information about the patient's airway in either a waveform or graphic on the Monitor's color LCD. This information should only be used in an adjunctive manner to assist with management of the artificial airway of the patient.

    AI/ML Overview

    SonarMed™ Airway Monitoring System: Acceptance Criteria and Study Details

    The provided 510(k) summary for the SonarMed™ Airway Monitoring System indicates that "The results of non-clinical (lab) performance testing demonstrate that the device is safe and effective." It explicitly states "Clinical data and conclusions were not needed for this device."

    Based on this information, the device relies on non-clinical (lab) performance testing to demonstrate its safety and effectiveness, and no clinical study was conducted or deemed necessary for this submission. Therefore, some of the requested information regarding clinical studies (like sample sizes for test/training sets, expert involvement, adjudication, MRMC studies, or specific ground truth based on patient outcomes/pathology) is not applicable or provided in this document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since the provided document only mentions "non-clinical (lab) performance testing" without detailing specific acceptance criteria or quantitative performance metrics, this table cannot be fully completed. The document broadly states the device's capabilities.

    Acceptance Criteria CategorySpecific Acceptance Criteria (from Study)Reported Device Performance (from Study)
    ETT Tip Placement VerificationNot specified in the document.Ability to display the baseline location of the ETT tip as established by the clinician.
    Passageway Diameter QuantificationNot specified in the document.Quantification of passageway diameter relative to the ETT diameter (e.g., a reading of less than 1 may indicate the tube is in a passageway that is smaller than itself, such as the esophagus).
    ETT Movement DetectionNot specified in the document.Ability to display ETT movement relative to the baseline location.
    ETT Occlusion/Obstruction DetectionNot specified in the document.Ability to display ETT occlusion/obstruction information, including percent obstructed.
    General Safety & EffectivenessNot specified in the document."The results of non-clinical (lab) performance testing demonstrate that the device is safe and effective."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable as no clinical test set was used/described. The document refers to "non-clinical (lab) performance testing."
    • Data Provenance: Not applicable as no clinical data or country of origin is mentioned for the non-clinical lab testing. The testing was reported as "non-clinical (lab) performance testing," implying an engineered or simulated environment rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set requiring expert ground truth establishment was used/described. The performance was assessed through non-clinical lab testing.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set involving human interpretation and needing adjudication was used/described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was indicated or performed. The device is an "adjunct to normal clinical practice" and is "not to be used as a stand-alone diagnostic system," implying it does not replace human interpretation but provides additional information.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes "non-clinical (lab) performance testing" which would represent a standalone evaluation of the device's ability to measure and display the stated parameters. However, the device's intended use is adjunctive, meaning it is designed to be used with human clinicians, not as a standalone diagnostic. The lab tests would have validated the technical performance of the algorithm in a controlled environment.

    7. The type of ground truth used

    For the "non-clinical (lab) performance testing," the ground truth would have been established through calibrated physical models or simulations where the true values of ETT tip location, movement, diameter, and obstruction were precisely known and controlled. This is characteristic of laboratory-based engineering validation.

    8. The sample size for the training set

    Not applicable, as no clinical training set was used/described. For non-clinical lab testing, the "training" equivalent would involve iterative testing and calibration during the device's development using controlled physical setups.

    9. How the ground truth for the training set was established

    Not applicable, as no clinical training set was used/described. For non-clinical lab testing, ground truth for development and calibration would have been established by precisely creating known conditions (e.g., ETT at a specific depth, tubes of known diameter, controlled obstructions) within a laboratory setup and using highly accurate measurement tools to verify these conditions.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1