Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222792

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-08

    (265 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SNUCONE Tissue Level Implant System is inticated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or internediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

    Device Description

    SNUCONE Implant System Fixture, also known as an endosseous implant, is surgical component that interfaces with the bone of the jaw or skull to support a dental prosthesis such as a crown, bridge, denture. Snucone's abutment and prosthetic components and tools are compatible with the Snucone's fixture only. The system includes Fixtures and Cover Screw, and Abutment and Component (Solid Abutment, InOcta Closing Screw, InOcta Healing Abutment, InOcta Abutment, InOcta Angled Abutment, InOcta Temporary Abutment, Platform Switching Abutment, and Abutment Screw).

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (SNUCONE Tissue Level Implant System). It contains details about the device, its intended use, comparison to predicate devices, and a list of non-clinical tests performed. However, it does not contain information about acceptance criteria or a study that proves the device meets those criteria, especially in the context of an FDA clearance for an AI/ML-driven device.

    The document explicitly states: "Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device." It details mechanical performance, sterilization, shelf-life, packaging, MR environment conditions, bacterial endotoxin, and biocompatibility.

    THERE IS NO MENTION OF AN AI/ML COMPONENT IN THIS DEVICE.

    Therefore, I cannot fulfill your request for information regarding:

    1. A table of acceptance criteria and the reported device performance for an AI/ML component. The document describes acceptance criteria for mechanical, material, and sterilization properties, but not AI/ML performance metrics (e.g., sensitivity, specificity, AUC).
    2. Sample size for test set and data provenance specific to an AI/ML model. The tests mentioned are laboratory bench tests, not data-driven performance studies on patient cohorts for an AI model.
    3. Number of experts and their qualifications for ground truth establishment. This is not relevant to the non-clinical tests described.
    4. Adjudication method for the test set. Not relevant.
    5. MRMC comparative effectiveness study or human reader improvement. Not relevant.
    6. Standalone (algorithm only) performance. Not relevant.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.). Not relevant to the tests performed.
    8. Sample size for the training set. Not relevant, as no AI model is described.
    9. How ground truth for the training set was established. Not relevant, as no AI model is described.

    In summary, the provided document is a 510(k) submission for a physical dental implant system and does not involve any AI/ML components or related performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210354

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-13

    (520 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SNUCONE Bone Level Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or intermediate Abutment support for fixed bridge work. SNUCONE Bone Level Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

    Device Description

    SNUCONE Implant System Fixture, also known as an endosseous implant, is surgical component that interfaces with the bone of the jaw to support a dental prosthesis such as a crown, bridge, denture. Snucone's abutment and prosthetic components and tools are compatible with the Snucone's fixture only.

    AI/ML Overview

    This document describes the SNUCONE Bone Level Implant System, a dental implant system. The information provided outlines the device, its components, and a comparison to predicate devices for substantial equivalence. It does not contain information about a study proving the device meets specific acceptance criteria in the way a clinical performance study would. Instead, it focuses on bench testing and adherence to recognized standards.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    The document details various non-clinical tests and the standards they adhere to. However, it does not provide a table with specific quantitative acceptance criteria for each test and the numerical reported performance of the device against those criteria. It generally states that "The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device."

    Here's a summary of the tests and the general statement of performance:

    Test/Criteria CategoryAcceptance Criteria (from standards)Reported Device Performance
    Mechanical Performance
    Fatigue testing (Abutments)According to "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" and "ISO 14801:2016 Dentistry -Fatigue test for endosseous dental implants"Met the criteria of the standards
    SEM/EDS analysis (Surface treatment media removal)According to "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."Confirmed removal of surface treatment media and met the criteria of the standards
    Sterilization, Shelf-life and Packaging
    End User Sterilization ValidationAccording to ISO 17665-1:2006, ISO 17665-2:2009, ISO 11737-1:2006, and ISO 11737-2:2009Met the criteria of the standards
    Gamma Sterilization ValidationAccording to ISO 11737-1:2006, and ISO 11737-2:2009Met the criteria of the standards
    Accelerated Aging Test (Shelf-life)According to ASTM F1980-16, ASTM F1140-13, ASTM F1929-15, and ASTM F2096-11Met the criteria of the standards
    MR Environment Condition
    Magnetically induced displacement force and torqueParameters per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" and published literature (Woods et al. 2019)Met the criteria based on scientific rationale and literature review
    Bacterial Endotoxin
    Endotoxin levelsAccording to ANSI/AAMI ST72:2019Met the criteria of the standards
    Biocompatibility
    Biocompatibility evaluationAccording to ISO 10993Met the criteria of the standards

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify the exact sample sizes used for each of the non-clinical bench tests (e.g., number of implants tested for fatigue). It refers to standards like ISO 14801 which would dictate sample sizes, but the numerical value is not explicitly stated.
    • Data Provenance (Country of Origin, Retrospective/Prospective): The document does not provide details on the country of origin of the data or whether the data was retrospective or prospective. Given that these are non-clinical bench tests, the concept of "retrospective" or "prospective" as applied to clinical studies is not directly applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the provided document. The document describes non-clinical bench testing of a dental implant system, not a diagnostic device requiring expert interpretation for ground truth establishment. The "ground truth" here refers to the physical and chemical properties and performance characteristics determined by the test methods outlined in the specified standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human readers or expert consensus, which is not the nature of the tests described in this document. The results of the mechanical, sterilization, MRI, endotoxin, and biocompatibility tests are determined by adherence to established, objective engineering and scientific standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This document describes a dental implant system, not an AI-powered diagnostic device, so an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests described is defined by the objective, measurable criteria within the referenced international and national standards (e.g., ISO 14801, ISO 17665, ASTM F1980, ANSI/AAMI ST72, ISO 10993). For example:

    • Fatigue testing: The "ground truth" is whether the implant withstands a specified number of cycles at a certain load without failure, as defined by ISO 14801.
    • Sterilization validation: The "ground truth" is achieving the required sterility assurance level (SAL), as defined by ISO 11737.
    • Biocompatibility: The "ground truth" is that the material does not elicit adverse biological responses, as assessed by the tests specified in ISO 10993.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" in the context of non-clinical bench testing for a physical medical device. Training sets are relevant for machine learning algorithms.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1