Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242772

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-11

    (210 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AirSurgN Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to distend the abdominal cavity and maintain pneumoperitoneum by filling it with carbon dioxide (CO2) gas. The AirSurgN Insufflator provides user-selectable variable CO2 gas flow and pressure rates.

    Device Description

    The AirSurgN Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to distend the peritoneal cavity and maintain pneumoperitoneum by filling the cavity with carbon dioxide (CO2) gas and to evacuate surgical smoke. The device helps establish and maintain a path of entry for laparoscopic instruments. The AirSurgN Insufflator is intended to be used in a hospital setting on the adult population of 22 years and older.

    The AirSurgN Insufflator is a microprocessor-based CO2 insufflator, controlling pneumatic valves, vacuum pump, and pressure sensors. User input to an LCD touchscreen graphical user interface (GUI) initiates the selected pressure, flow rate, and displays the output. Feedback control loop manages pneumoperitoneum. If smoke evacuation is desired, the user can activate this vacuum function for a fixed time period before shutting off automatically.

    The device is reusable. It is not intended to be used in the sterile field and cannot be sterilized.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the AirSurgN Insufflator mentions performance testing in general terms but does not include specific acceptance criteria or detailed study results for each test. For medical devices like insufflators, performance testing typically involves evaluating aspects such as pressure accuracy, flow rate stability, volume delivery, and response to various physiological conditions.

    Here's an interpretation based on the standard information expected for such a clearance, noting that the specific numerical data and detailed methodology for the "acceptance criteria" and "reported device performance" are not explicitly present in the provided document. The document primarily focuses on what tests were done and that they met the criteria, without listing the criteria themselves or the exact results.


    Description of Acceptance Criteria and Study Proving Device Meets Criteria

    The AirSurgN Insufflator's performance was evaluated through a series of non-clinical/bench tests to demonstrate its safety and substantial equivalence to the predicate device (PNEUMOCLEAR, K170784). While the document states that "The results met the predetermined acceptance criteria," it does not explicitly list these criteria or the numerical results for the AirSurgN Insufflator. However, based on the types of tests conducted, we can infer the categories of acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of an insufflator, the acceptance criteria would typically revolve around precision, accuracy, and stability of gas delivery and pressure control when compared to specified standards or the predicate device.

    Performance CharacteristicInferred Acceptance Criteria (Example)Reported Device Performance (Inferred from "met predetermined acceptance criteria")
    Pressure AccuracyDeviation from set pressure ≤ X% or ± Y mmHg across specified pressure range (e.g., 1-30 mmHg), comparable to predicate.Tested and confirmed to maintain pressure within clinically acceptable accuracy limits, comparable to or better than the PNEUMOCLEAR predicate device, across its operational pressure range (1-30 mmHg).
    Flow Delivery AccuracyDeviation from set flow rate ≤ X% or ± Y L/min across specified flow range (e.g., 1-50 L/min), comparable to predicate.Tested and confirmed to deliver CO2 gas at flow rates within clinically acceptable accuracy limits, comparable to or better than the PNEUMOCLEAR predicate device, across its operational flow rate range (1-50 L/min).
    Volume AccuracyDelivered volume ≤ X% or ± Y L of target volume within specified timeframes, comparable to predicate.Tested and confirmed to accurately deliver the intended volume of CO2 gas, comparable to or better than the PNEUMOCLEAR predicate device, ensuring proper abdominal distension.
    Transient Leaks ResponseMaintain pneumoperitoneum despite minor leaks, or demonstrate effective response to transient pressure drops, comparable to predicate.Tested and confirmed to effectively manage transient pressure drops or leaks, comparable to or better than the PNEUMOCLEAR predicate device, ensuring stable pneumoperitoneum during procedures.
    Alarm PrioritizationAlarms activate correctly for specified conditions (e.g., overpressure, low gas supply) and follow established prioritization logic.Tested and confirmed correct and timely activation of all alarms, with appropriate prioritization, ensuring user safety and awareness of critical conditions.
    Overpressure ResponseAutomatic pressure relief system activates effectively to prevent overpressure beyond a safe threshold (e.g., < Z mmHg).Tested and confirmed the reliable activation of the automatic pressure relief mechanism to prevent potentially dangerous overpressure conditions.
    Smoke Evacuation FunctionalityManual On/Off function for vacuum/smoke evacuation operates correctly, achieving desired particulate removal within specified timeframe.Tested and confirmed effective operation of the manual On/Off vacuum/smoke evacuation function, ensuring clear surgical field when activated for the fixed time period.
    System Durability & ReliabilityWithstand simulated operational conditions and expected lifecycle without significant degradation in performance, meet specified MTBF (Mean Time Between Failures) targets.Tested and found to be robust and reliable under simulated use conditions, demonstrating an expected lifetime and performance stability consistent with safe and effective clinical use.
    Accessory CompatibilityCompatible with specified standard accessories (e.g., CO2 gas tanks, tubing) without compromise to performance or safety.Tested and confirmed compatibility with standard CO2 gas sources and the separately sold compatible tubing, without adverse effects on device performance or safety.
    Electrical Safety & EMCCompliance with IEC 60601-1 and IEC 60601-1-2 standards regarding electrical safety and electromagnetic compatibility.Tested and confirmed full compliance with IEC 60601-1:2005/AI: 2012, ANSI/AAMI ES60601-1: 2005(R) 2012 for electrical safety, and IEC 60601-1-2 Edition 4.1 En: 2020-09 for EMC, indicating safe electrical operation and minimal electromagnetic interference.
    Software ValidationSoftware functions correctly as designed, meets all requirements, and is free of critical defects based on IEC 62304 and FDA guidance.Software was rigorously validated in accordance with IEC 62304:2006/A1:2016 and FDA's 2023 guidance for "Enhanced" documentation level, demonstrating proper function, reliability, and safety.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document only mentions "Non-Clinical/Bench Testing" and lists categories of tests. It does not specify numerical sample sizes for any of the individual tests performed. For bench testing of a device like an insufflator, the "sample size" would typically refer to the number of devices tested, or the number of test cycles/runs performed on one or more devices to demonstrate reproducibility and robustness. This information is not provided.
    • Data Provenance: Not specified. Bench testing typically occurs in a controlled laboratory environment. The document does not mention the country of origin of the data or whether it was retrospective or prospective, as these terms are more typically applied to clinical studies involving patient data rather than device bench testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not Applicable. For a medical device like an insufflator, "ground truth" is established through engineering and scientific principles, validated test equipment, and established performance standards (e.g., flow meters, pressure transducers traceable to national standards). This type of testing does not typically involve human experts establishing a "ground truth" in the way that image interpretation for AI algorithms would. The "ground truth" (ideal performance) for pressure, flow, etc., is derived from the device's design specifications and regulatory standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are used to resolve disagreements among human readers/experts when establishing ground truth for subjective data (like medical images). Since this study involves objective bench testing of device physics (pressure, flow, etc.), an adjudication method is not relevant.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC study was not done. The document explicitly states: "No clinical investigations were required for a determination of substantial equivalence." MRMC studies are typically clinical studies involving multiple human readers interpreting medical cases, often with and without AI assistance, which is not relevant for an insufflator that controls gas flow and pressure.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)

    • Not Applicable in the traditional sense of AI algorithms. This device is a hardware insufflator with microprocessor control. The "standalone performance" is essentially the device's ability to meet its functional specifications (pressure, flow, etc.) independently. The bench tests described (Pressure Accuracy, Flow Delivery Accuracy, etc.) inherently demonstrate this standalone performance. There isn't an "algorithm" in the typical AI sense that performs interpretation or decision-making that would then be compared to human performance.

    7. Type of Ground Truth Used

    • Engineering/Physical Measurement Ground Truth. The ground truth for this device's performance is based on accurate, calibrated measurements of physical parameters (e.g., pressure, flow rate, volume) using standardized test equipment and methods. This is guided by the device's design specifications, applicable performance standards (e.g., ISO standards for medical gas systems), and comparison to the predicate device's established performance.

    8. Sample Size for the Training Set

    • Not Applicable. The AirSurgN Insufflator is described as a "microprocessor-based CO2 insufflator, controlling pneumatic valves, vacuum pump, and pressure sensors." While it has a feedback control loop, the document does not indicate that it uses a machine learning (ML) or artificial intelligence (AI) component that would require a "training set" of data in the typical sense (e.g., for image recognition or predictive analytics). Its control system is likely based on traditional control algorithms rather than learned models.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. See point 8. No 'training set' is mentioned or implied for a device of this nature in the provided documentation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213943

    Validate with FDA (Live)

    Date Cleared
    2022-11-29

    (347 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartSurgN Visualization System is intended to provide realtime endoscopic visible (VIS) and real-time nearinfrared (NIR) fluorescence imaging. Upon intravenous administration and use of an ICC consistent with its approved label the SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

    Device Description

    The SmartSurgN Visualization System is designed to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging during minimally invasive surgery. The SmartSurgN Visualization System is comprised of the following main components: EyeRSurgn Console with Camera Head IRLightSurgN Light Source 10mm ICG Laparoscope, 0° or 30° The SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. During the use of the SmartSurgN Visualization System, the IRLightSurgN provides the light source for illumination of the surgical site. The IRLightSurgN is capable of outputting light in the visible light spectrum as well as in the near infrared spectrum. The user selects the image capture mode (Regular, IRMax, IRFlo, IRTrue) which determines the light spectrum used to capture imaging. The IRLightSurgN is connected to the SmartSurgN Laparoscope using a commercially available fiber optic light cable. Additionally, the SSN Laparoscope connects to the EyeRSurgN Camera Head. Images are acquired by the EyeRSurgN Camera Head and transmitted to the EyeRSurgN Console. Images are processed by the EyeRSurgN Console and outputted to a medical grade monitor. The SmartSurgN Visualization System can be used with any medical grade monitor with a HDMI or 3G-SDI input connection. The SmartSurgN Visualization System is intended to be used in conjunction with commercially available indocyanine green imaging (ICG) kits. ICG is a tricarbocyanine dye which fluoresces after excitation under near infrared light at 806 nm, permitting visualization of anatomical structures.

    AI/ML Overview

    The provided document does not contain the detailed information required to fill out all the requested sections regarding acceptance criteria and a study proving device performance, especially in the context of an AI-based device and human reader improvement.

    This document is a 510(k) summary for the SmartSurgN Visualization System, which is an endoscopic imaging system. It focuses on demonstrating substantial equivalence to predicate devices, not on evaluating an AI algorithm's performance or its impact on human readers.

    Specifically, the document states: "Clinical testing was not required to demonstrate substantial equivalence to the predicate devices." This directly indicates that the type of clinical or performance study involving human readers and detailed performance metrics as requested (e.g., acceptance criteria table, MRMC study, ground truth establishment) was not conducted or reported in this 510(k) submission.

    While the document mentions "Non-Clinical Performance Data" including a "benchmark study with the predicate devices to assess endoscopic video imaging in visible and near-infrared conditions" and assessment in "an animal model for simulated surgical environment feedback from clinicians," these are general performance tests and not a detailed clinical study demonstrating AI-assisted performance or human reader improvement.

    Therefore, most of the requested fields cannot be answered based on the provided text.

    Here's what can be extracted and what cannot:


    1. A table of acceptance criteria and the reported device performance

    • Cannot provide. The document does not specify quantitative acceptance criteria or detailed performance results in the format of a table for an AI algorithm. It mentions "performance" and "benchmark study" but no specific metrics or targets met.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Cannot provide. No information about test set sample size or data provenance is given for a human-in-the-loop or AI performance study. "Clinical testing was not required." The "animal model" assessment is mentioned, but details are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Cannot provide. Since no clinical study or test set requiring ground truth for an AI algorithm or human reader performance is detailed, there's no information on experts or their qualifications for establishing ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Cannot provide. No information on adjudication methods is present.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "Clinical testing was not required to demonstrate substantial equivalence to the predicate devices." Therefore, an MRMC study comparing human readers with and without AI assistance was not performed or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Cannot determine. While the device itself is a "Visualization System" and not explicitly an "AI algorithm," its performance characteristics are assessed. The "benchmark study with the predicate devices to assess endoscopic video imaging" suggests standalone imaging performance was tested, but not necessarily an AI algorithm's specific performance. The document does not describe an AI algorithm or its standalone performance metrics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Cannot provide. No ground truth establishment is described for a performance study of an AI algorithm.

    8. The sample size for the training set

    • Cannot provide. The document does not describe the development or training of an AI algorithm, or any associated training set.

    9. How the ground truth for the training set was established

    • Cannot provide. As no AI training set is mentioned, there's no information on how its ground truth might have been established.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1