    Intended Use

    The Disposable Trocars have applications in minimally invasive surgical procedures to establish a path of entry for endoscopic instruments in abdominal area.

    Device Description

    Disposable Trocars are medical devices that penetrate the abdominal wall during laparoscopic surgery. They provide a passageway for the surgical device insertion and removal and maintain pneumoperitoneum during surgical procedures.

    The model numbers of Disposable Trocars include AVTB 5-70, AVTB 5-100, AVTB 10-70, AVTB 10-100, AVTB 12-70, AVTB 12-100, AVTC 5-70, AVTC 5-100, AVTC 10-70, AVTC 10-100, AVTC 12-70, AVTC 12-100, AVTC 15-70, and AVTC 15-100.

    The AVTB Disposable Trocars comprise an upper sealing component, a casing component, and a puncture rod component.
    The upper sealing component is composed of a seal lid, an upper retaining ring, a wave seal ring, a conical sealing ring, a lower retaining ring, and a retaining cover for locking. The casing component is composed of choke valve, self-commissioning sealing cap, casing seat, seal ring, trocar sleeve, gas injection valve, and gas injection switch. The puncture rod component is composed of an upper fixing cover, a locking buckle, a puncture rod, and a lower fixing cover.

    AVTC Disposable Trocars comprise casing component, puncture rod component, upper sealing component, and converter component (only applicable to AVTC 15 specification).
    The casing component is composed of choke valve, self-commissioning sealing cap, casing seat, sealing ring, trocar sleeve, gas injection valve, and gas injection switch. The puncture rod component is composed of an upper fixing cover and a puncture rod. The upper sealing component (5/10/12 specification) is composed of seal lid, an upper retaining ring, a wave seal ring, a conical sealing ring, a lower retaining ring, and a retaining cover for locking. The upper sealing component (15 specifications) is composed of a seal lid, an elastic seal ring, and a retaining cover for locking. The converter component (15 specifications) is composed of converter, converter cover, sealing cap of upper converter cover, and upper converter cover.

    Disposable Trocars are divided into B and C according to different shapes. According to the inner diameter of the trocar, it is divided into four sizes: 5, 10, 12, and 15. According to the length of the trocar, it is divided into two sizes: 70 and 100.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for Sinolinks Medical Innovation, Inc.'s Disposable Trocars (K250820) do not contain information about acceptance criteria or a study proving the device meets the acceptance criteria in the context of an AI/ML-enabled device evaluation.

    The document pertains to a traditional medical device (Disposable Trocars) and its non-clinical performance testing (bench, ex vivo, biocompatibility, sterility, packaging, shelf-life, transport, etc.) to demonstrate substantial equivalence to a predicate device.

    Therefore, I cannot extract the information required to answer your prompt, which is tailored for AI/ML device evaluations. The document explicitly states: "No clinical study is included in this submission." and the "Summary of non-clinical testing" section details a range of physical, mechanical, and biological tests relevant to a non-AI/ML medical device.

    To answer your prompt for a device that is not an AI/ML device, the following would be the most relevant information:

    Acceptance Criteria and Device Performance for Disposable Trocars (K250820)

    The acceptance criteria for the Disposable Trocars are demonstrated through a series of non-clinical tests which show the device meets its design specifications and performs comparably to the predicate device. The "performance" in this context refers to the successful completion and passing of these tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a non-AI/ML device, the "acceptance criteria" are embodied in the successful completion of the various non-clinical tests to predefined standards. The "reported device performance" is that the device met these standards.

    Acceptance Criteria Category/TestReported Device Performance (Met/Passed)
    Functional Performance
    Compatible with endoscope and endoscopic staplerMet
    FlexibilityMet
    Parts compatible with each otherMet
    Airtightness and air resistanceMet
    Connection strengthMet
    Luer connector performanceMet
    Insertion and pull-out performance of puncture rodMet
    Gas leak rateMet
    Material & Biocompatibility
    Endotoxin limit (20 EU per device)Met
    CytotoxicityNo cytotoxicity detected
    SensitizationNo sensitization detected
    Skin IrritationNo irritation detected
    PyrogenicityNo pyrogen detected
    Physical & Mechanical
    Dimension verification (Shaft Diameter, Length, Trocar Width, Total Length)Met design specifications (specific values detailed in summary)
    Penetration force (ex vivo on porcine tissue)Evaluated and met expectations
    Fixation force (ex vivo on porcine tissue)Evaluated and met expectations
    Tip integrity after insertion (ex vivo on porcine tissue)Evaluated and met expectations
    Sterility & Packaging
    Sterility Assurance Level (SAL 10⁻⁶)Achieved (EO sterilized)
    Shelf-life (3 years)Supported by accelerated aging (real-time ongoing)
    Simulated transport testing (ASTM D4169)Supported product and packaging performance
    Seal Strength (ASTM F88/F88M)Tested and met
    Seal Leaks (ASTM F1929)Tested and met

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the numerical sample size for each bench or ex vivo test. It mentions "porcine tissue" for ex vivo studies. It does not provide explicit data provenance in terms of country of origin for the datasets used in these non-clinical tests, though the manufacturer is based in China. The data would be considered prospective for the purposes of these clearance tests, as they were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This concept is not applicable to the non-clinical testing of a traditional medical device like a trocar. "Ground truth" established by experts is a concept primarily relevant to diagnostic AI/ML models where human interpretation serves as the reference standard. For a trocar, the "ground truth" is determined by established engineering standards, material science principles, and biological safety evaluations.

    4. Adjudication method for the test set:

    Not applicable for non-clinical performance testing of a traditional device. Adjudication methods (e.g., 2+1 reader consensus) are used in human-in-the-loop or standalone AI/ML studies to resolve discrepancies in expert labeling or diagnoses.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No. This type of study is specifically for evaluating the impact of AI/ML assistance on human reader performance, which is irrelevant for a disposable trocar.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This is an AI/ML specific evaluation. The "device" itself performs its function (e.g., creating an entry path) without an "algorithm" in the diagnostic sense.

    7. The type of ground truth used:

    For this traditional medical device, the "ground truth" is effectively:

    • Established engineering specifications and performance standards for mechanical and physical properties (e.g., dimensions, strength, air-tightness).
    • Industry consensus standards for biocompatibility (e.g., ISO 10993 series), sterility (e.g., ISO 11135), and packaging (e.g., ASTM standards).
    • Biological responses (e.g., no cytotoxicity demonstrated on cell cultures, no irritation in animal models for biocompatibility).

    8. The sample size for the training set:

    Not applicable. The concept of a "training set" refers to data used to train an AI/ML algorithm. This device does not use AI/ML.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no AI/ML training set.