Found 2 results

510(k) Data Aggregation

    K Number
    K191050

    Validate with FDA (Live)

    Date Cleared
    2019-07-12

    (84 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

    Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

    Software syngo MR E11E with Ischemic Heart Disease (HD) Workflow, when used with a gadolinium-based contrast agent (GBCA) approved for cardiac MRI (CMRI) is indicated for the acquisition and display of images of myocardial perfusion (stress, rest) and late gadolinium enhancement (LGE) during post-contrast CMRI examination in patients with known or suspected coronary artery disease (CAD).

    Device Description

    MAGNETOM Aera and MAGNETOM Skyra with Software syngo MR E11E with IHD Workflow are the subject devices. The Software syngo MR E11E with IHD Workflow, when used with a gadolinium-based contrast agent (GBCA) approved for CMRI, extends the capability of the cleared Cardiac Dot Engine (K121434) for post-contrast CMRI exams for patients with known or suspected coronary artery disease (CAD). Software syngo MR E11E with IHD Workflow is available for MAGNETOM Aera and MAGNETOM Skyra excluding the 24-channel configuration.

    A. The cleared Cardiac Dot Engine (syngo MR D13A, K121434) helps acquisition and display of cardiac morphology and function (noncontrast CMRI).
    A comprehensive post-contrast CMRI exam includes stress/rest perfusion and late gadolinium enhanced (LGE) imaging. To accomplish post-contrast CMRI imaging, basic morphologic / functional imaging (noncontrast CMRI) is required. Therefore, the cleared Cardiac Dot Engine (syngo MR D13A, K121434) is a pre-requisite to the subject devices.

    B. The Cardiac Dot Engine together with the IHD Workflow and a GBCA approved for post-contrast CMRI provides for a complete (pre- and post-contrast) examination.

    The primary predicate devices are modified to include a new Dot Workflow named "Ischemic Heart Disease" (IHD) Workflow for a post-contrast CMRI exam using pulse sequences already cleared in the USA (syngo MR D13A, K121434). A new Dot Workflow "Ischemic Heart Disease" is added in Cardiac Dot Engine dropdown list, under the region "heart". This Dot Workflow includes the following: six new post-contrast CMRI measurement protocols and one workflow step:

    New Measurement protocols:

    1. DynamicTest (test protocol for perfusion imaging without contrast agent)
    2. DynamicStress (protocol for perfusion imaging under stress conditions)
    3. DynamicRest (protocol for perfusion imaging under rest conditions)
    4. DE overview (protocol for delayed enhancement (DE) or LGE with low spatial resolution as an overview)
    5. DE_seg_high-res_LAX (protocol for DE or LGE with high spatial resolution in long axis view)
    6. DE seg high-res SAX (protocol for DE or LGE with high spatial resolution in short axis view)

    New workflow step:

    1. Inject contrast agent. This step prompts the user to start the contrast agent injection for post-contrast CMRI exams.
    AI/ML Overview

    This document describes the regulatory approval for the Siemens MAGNETOM Aera and MAGNETOM Skyra with Software syngo MR E11E with Ischemic Heart Disease (IHD) Workflow. The IHD Workflow is an extension to the cleared Cardiac Dot Engine, intended for post-contrast Cardiac MRI (CMRI) exams for patients with known or suspected Coronary Artery Disease (CAD).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria or quantitative performance metrics for the IHD Workflow. Instead, it refers to the efficacy results of two clinical studies (GadaCAD1 and GadaCAD2) to support the device's performance. The "acceptance criteria" can therefore be inferred as the successful demonstration of the device's ability to acquire and display myocardial perfusion and late gadolinium enhancement (LGE) images that adequately detect CAD when interpreted by qualified readers.

    Acceptance Criteria (Inferred)Reported Device Performance
    Ability to acquire and display images of myocardial perfusion (stress, rest) and late gadolinium enhancement (LGE) during post-contrast CMRI examination for interpretation by trained physicians.The GadaCAD studies demonstrated that the device generated CMRI images (stress and rest perfusion and LGE) that were interpreted by qualified independent readers to adequately detect CAD.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: A total of 764 subjects were evaluated across two studies: 376 subjects in GadaCAD1 and 388 subjects in GadaCAD2. These subjects represent the test set for evaluating the post-gadobutrol CMRI capabilities.
    • Data Provenance: The GadaCAD studies were prospectively controlled, multi-national, single-arm clinical studies. The document states that they were performed by Bayer HealthCare AG.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts: The document states that the CMRI images were "interpreted by qualified independent readers". While it doesn't specify an exact number, the use of "readers" (plural) across multi-center, multi-national studies implies multiple experts were involved in the interpretation process.
    • Qualifications of Experts: The readers were described as "radiologists and cardiologists experienced in CMRI".

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1 consensus) for establishing the ground truth from the expert interpretations in the GadaCAD studies. It simply states that the images were "interpreted by qualified independent readers." The primary endpoint of the GadaCAD studies, as mentioned in the GADAVIST™ package insert, was likely the diagnostic accuracy of CMRI for CAD detection, which would have implicitly relied on these interpretations, but the specific consensus mechanism is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study designed to assess the improvement of human readers with AI vs. without AI assistance.
    • The IHD Workflow is presented as a set of modified measurement protocols and a workflow step for image acquisition and display, extending the capabilities of the existing Cardiac Dot Engine. The clinical studies evaluated the efficacy of the imaging technique itself (using the IHD Workflow components) for CAD detection, not the comparative performance of human readers with and without AI assistance for interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • The document does not present a standalone performance study for an algorithm without human-in-the-loop performance.
    • The IHD Workflow focuses on image acquisition and display protocols. The interpretation of these images for diagnosis still relies on "trained physicians." The clinical studies validate the acquisition and display capabilities to produce images sufficient for expert interpretation.

    7. Type of Ground Truth Used

    The ground truth used in the GadaCAD studies was expert consensus / clinical interpretation by qualified independent readers (radiologists and cardiologists experienced in CMRI) for the detection of CAD. The document explicitly states: "The post-gadobutrol CMRI specific acquisition protocols supported adequate detection of CAD in two multi-center, multinational clinical studies."

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. The IHD Workflow's development and validation are largely described in terms of non-clinical software verification and validation, and clinical validation using the GadaCAD studies. The GadaCAD studies served as the validation set for the device's performance, not a training set for an AI/algorithm component requiring labeled data for learning. The IHD Workflow primarily introduces new measurement protocols and a workflow step, not necessarily a machine learning algorithm that requires a distinct training set in the conventional sense.

    9. How Ground Truth for the Training Set Was Established

    Since no training set is explicitly mentioned or relevant for the described IHD Workflow (which consists of acquisition protocols and workflow steps, not an AI model requiring a training phase), the method for establishing ground truth for a training set is not applicable/not provided in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173611

    Validate with FDA (Live)

    Date Cleared
    2018-03-16

    (114 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The syngo Application Software is a medical software for real-time viewing, image manipulation, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures.

    The syngo Application Software can be deployed on independent hardware such as a stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured within a network to send and receive DICOM data.

    Furthermore, the syngo Application Software can be deployed on systems Angiography system family. It provides image guided solutions in the operating room, for image guided surgery, by Image Fusion and by navigation systems, image guided solutions in interventional cardiology and image guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology.

    The syngo Application Software can also be combined with fluoroscopy systems or Radiographic systems.

    The syngo Application Software can be configured with a variety of syngo or Windows-based software options, which are intended to assist the physician in diagnosis, treatment control. It includes commercially available post-processing techniques and OEM options.

    Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive turnor treatment.

    Device Description

    The "syngo Application Software" is medical diagnostic software for real-time viewing, diagnostic review, post-processing, image manipulation, optimization, communication, reporting and storage of medical images and data on exchange media. It provides image-quided solutions in the operating room, for imageguided surgery, by Image Fusion and by navigation systems, image-guided solutions in interventional cardiology and electrophysiology and image-guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology. It can be deployed with a variety of syngo or Windows-based software options, which are intended to assist the physician in the evaluation of digital radiographic examinations, including diagnosis and/or treatment planning.

    Siemens "syngo Application Software" (VD20B) is designed to work with digital radiographic, fluoroscopic, interventional and angiographic systems. The software platform with common software architecture, syngo application packages, and basic services is the same as used with the "syngo Application Software" (VD20B) K170747 with exception of the added software options "Instant Fusion" and "Common Patient Registration". The new optional software features are only possible in an angiography (ARTIS Family) and CT (SOMATOM CT Sliding Gantry) systems configuration known as "nexaris Angio-CT".

    "Instant Fusion" and "Common Patient Registration" is the subject of this submission. These features seamlessly integrate Angio and CT together and facilitate combined use. The "Instant Fusion" is an overlay of CT datasets to the Angiography image. The overlay is a position based registration of the CT gantry. The "Common Patient Registration" is the patient data transfer from "syngo Application Software" to the CT system using a communication interface.

    AI/ML Overview

    This document is a 510(k) summary for the syngo Application Software (VD20B) by Siemens Medical Solutions USA, Inc. The submission requests clearance for two new optional software features: "Instant Fusion" and "Common Patient Registration." The study presented is a non-clinical performance test comparing the subject device (VD20B) with its predicate device (VD20A, K170747).

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the acceptance criteria in terms of successful verification and validation testing, confirming that the new features ("Instant Fusion" and "Common Patient Registration") function as intended and do not raise new issues of safety or effectiveness. The reported device performance indicates these criteria were met.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Functional Performance- "Instant Fusion" successfully overlays CT images without additional 2D/3D registration. - "Common Patient Registration" successfully transfers patient data automatically from the Angiography system to the syngo Application Software and the CT system.- "Instant Fusion": "Measurements were performed with CT images to the interface simulator. The tests showed that CT images can be directly overlaid without additional 2D/3D registration." - "Patient Registration": "Measurements were performed with different patient images to the interface simulator. The tests showed that the Patient data is automatically transferred after Patient registration on the Angiography system."
    Safety and Effectiveness- No new issues of safety or effectiveness are raised compared to the predicate device. - Conformance to relevant medical device standards (e.g., ISO 14971, IEC 60601-1-6, IEC 62304, IEC 62366-1, IEC 80001-1).- "Results of all conducted testing were found acceptable and did not raise any new issues of safety or effectiveness." - "All software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device that is currently marketed for the same intended use." - Conformance claims made for the listed standards in section 8.
    Risk ManagementRisks identified and mitigated; testing results support meeting acceptance criteria."The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria."
    Human Factors/UsabilityNo safety-relevant functions requiring summative usability validation."The Human Factor Usability Validation showed no safety-relevant functions that need to be validated with a summative usability validation according to the IEC and FDA Guidelines."
    CybersecurityCompliance with cybersecurity requirements, including prevention of unauthorized access/misuse."Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access… Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a numerical sample size for the test set of "Instant Fusion" or "Common Patient Registration." It refers to "measurements were performed with CT images" and "measurements were performed with different patient images," implying a set of images/data was used, but the quantity is not specified.

    The data provenance is not explicitly mentioned as retrospective or prospective, nor is a country of origin provided for the data used in validation testing. However, given it's a non-clinical performance test for software features, the data would likely be generated in a controlled environment as part of product development and verification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the software's functional validation. The testing described appears to be technical verification against predefined software specifications and functional requirements rather than a clinical evaluation requiring expert consensus on diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Since the validation involved technical performance testing of software features (overlay and data transfer), rather than a diagnostic accuracy study with ambiguous cases, an adjudication method like 2+1 or 3+1 is not applicable and not mentioned. The "ground truth" here is the system's intended technical function as defined by its specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done or reported. The submission focuses on the technical functionality and safety/effectiveness equivalence to a predicate device, not on comparing human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance testing described ("Instant Fusion" overlay and "Patient Registration" data transfer) represents a standalone algorithm performance evaluation in terms of its technical functionality. The tests assessed whether the software features performed their designated tasks (image overlay, data transfer) correctly as programmed, independent of human interpretation or interaction during the measurement itself.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth for the testing of "Instant Fusion" and "Common Patient Registration" appears to be the software's predefined functional specifications and technical requirements. For "Instant Fusion," the ground truth is the successful direct overlay of CT images via position-based registration. For "Patient Registration," the ground truth is the automatic and correct transfer of patient data.

    8. The Sample Size for the Training Set

    The document is a 510(k) summary for software features within an existing "syngo Application Software." It does not mention any training set size, which suggests that these features are likely rule-based or deterministic algorithms (for image overlay and data transfer) rather than machine learning or AI models that would require a distinct training phase with a labeled dataset.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is discussed or implied for these specific software features (as they appear to be deterministic functions), the concept of establishing ground truth for a training set is not applicable here.