Search Results
Found 2 results
510(k) Data Aggregation
(23 days)
syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra.
syngo.MR Applications is a software only Medical Device consisting post-processing applications/workflows used for viewing and evaluating the designated images provided by a MR diagnostic device. The post-processing applications/workflows are integrated with the hosting application syngo.via, that enables structured evaluation of the corresponding images
The provided FDA 510(k) clearance letter and summary for syngo.MR Applications (VB80) indicate that no clinical studies or bench testing were performed to establish new performance criteria or demonstrate meeting previously established acceptance criteria. The submission focuses on software changes and enhancements from a predicate device (syngo.MR Applications VB40).
Therefore, based solely on the provided document, I cannot create the requested tables and information because the document explicitly states:
- "No clinical studies were carried out for the product, all performance testing was conducted in a non-clinical fashion as part of verification and validation activities of the medical device."
- "No bench testing was required to be carried out for the product."
The document details the following regarding performance and acceptance:
- Non-clinical Performance Testing: "Non-clinical tests were conducted for the subject device during product development. The modifications described in this Premarket Notification were supported with verification and validation testing."
- Software Verification and Validation: "The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the device Syngo.MR Applications during product development... The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence."
- Conclusion: "The predicate device was cleared based on non-clinical supportive information. The comparison of technological characteristics, device hazards, non-clinical performance data, and software validation data demonstrates that the subject device performs comparably to and is as safe and effective as the predicate device that is currently marketed for the same intended use."
This implies that the acceptance criteria are related to the functional specifications and performance of the software, as demonstrated by internal verification and validation activities, rather than a clinical performance study with specific quantitative metrics. The new component, "MR Prostate AI," is noted to be integrated without modification and had its own prior 510(k) clearance (K241770), suggesting its performance was established in that separate submission.
Without access to the actual verification and validation reports mentioned in the document, it's impossible to list precise acceptance criteria or detailed study results. The provided text only states that "all the software specifications have met the acceptance criteria."
Therefore, I can only provide an explanation of why the requested details cannot be extracted from this document:
Explanation Regarding Acceptance Criteria and Study Data:
The provided FDA 510(k) clearance letter and summary for syngo.MR Applications (VB80) explicitly state that no clinical studies or bench testing were performed for this submission. The device (syngo.MR Applications VB80) is presented as a new version of a predicate device (syngo.MR Applications VB40) with added features and enhancements, notably the integration of an existing AI algorithm, "Prostate MR AI VA10A (K241770)," which was cleared under a separate 510(k).
The basis for clearance is "non-clinical performance data" and "software validation data" demonstrating that the subject device performs comparably to and is as safe and effective as the predicate device. The document mentions that "all the software specifications have met the acceptance criteria" as part of the verification and validation (V&V) activities. However, the specific quantitative acceptance criteria, detailed performance metrics, sample sizes, ground truth establishment, or expert involvement for these V&V activities are not included in this public summary.
Therefore, the requested information cannot be precisely extracted from the provided text.
Summary of Information Available (and Not Available) from the Document:
| Information Requested | Status (Based on provided document) |
|---|---|
| 1. Table of acceptance criteria and reported performance | Not provided in the document. The document states: "The testing results support that all the software specifications have met the acceptance criteria." However, it does not specify what those acceptance criteria are or report detailed performance metrics against them. These would typically be found in the detailed V&V reports, which are not part of this summary. |
| 2. Sample size and data provenance for test set | Not provided. The document indicates "non-clinical tests were conducted as part of verification and validation activities." The sample sizes for these internal tests, the nature of the data, and its provenance (e.g., country, retrospective/prospective) are not detailed. It is implied that the data is not patient-specific clinical test data. |
| 3. Number of experts and qualifications for ground truth | Not applicable/Not provided. Since no clinical studies or specific performance evaluations against an external ground truth are described in this document, there's no mention of experts establishing ground truth for a test set. The validation appears to be against software specifications. If the "MR Prostate AI" component had such a study, those details would be in its individual 510(k) (K241770), not this submission. |
| 4. Adjudication method for test set | Not applicable/Not provided. As with the ground truth establishment, no adjudication method is mentioned because no external test set requiring such expert consensus is described within this 510(k) summary. |
| 5. MRMC comparative effectiveness study and effect size | Not performed for this submission. The document explicitly states "No clinical studies were carried out for the product." Therefore, no MRMC study or AI-assisted improvement effect size is reported here. |
| 6. Standalone (algorithm only) performance study | Partially addressed for a component. While this submission doesn't detail such a study, it notes that the "MR Prostate AI" algorithm is integrated without modification and "is classified under a different regulation in its 510(K) and this is out-of-scope from the current submission." This implies that a standalone performance study was done for the Prostate MR AI algorithm under its own 510(k) (K241770), but those details are not within this document. For the overall syngo.MR Applications (VB80) product, no standalone study is described. |
| 7. Type of ground truth used | Not provided for the overall device's V&V. The V&V activities are stated to have met "software specifications," which suggests an internal, design-based ground truth rather than clinical ground truth like pathology or outcomes data. For the integrated "MR Prostate AI" algorithm, clinical ground truth would have been established for its separate 510(k) submission. |
| 8. Sample size for the training set | Not applicable/Not provided for this submission. The document describes internal non-clinical V&V for the syngo.MR Applications software. It does not refer to a machine learning model's training set within this context. The "Prostate MR AI" algorithm, being independently cleared, would have its training set details in its specific 510(k) dossier (K241770), not here. |
| 9. How the ground truth for the training set was established | Not applicable/Not provided for this submission. As above, this document does not discuss a training set or its ground truth establishment for syngo.MR Applications. This information would pertain to the Prostate MR AI algorithm and be found in its own 510(k). |
Ask a specific question about this device
(203 days)
Syngo Carbon Clinicals is intended to provide advanced visualization tools to prepare and process the medical image for evaluation, manipulation and communication of clinical data that was acquired by the medical imaging modalities (for example, CT, MR, etc.)
OrthoMatic Spine provides the means to perform musculoskeletal measurements of the whole spine, in particular spine curve angle measurements.
The TimeLens provides the means to compare a region of interest between multiple time points.
The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired by medical imaging modalities.
An interface shall enable the connection between the Syngo Carbon Clinicals software package and the interconnected software solution for viewing, manipulation, communication, and storage of medical images.
Syngo Carbon Clinicals is a software only Medical Device, which provides dedicated advanced imaging tools for diagnostic reading. These tools can be called up using standard interfaces any native/syngo based viewing applications (hosting applications) that is part of the SYNGO medical device portfolio. These tools help prepare and process the medical image for evaluation, manipulation and communication of clinical data that was acquired by medical imaging modalities (e.g., MR, CT etc.)
Deployment Scenario: Syngo Carbon Clinicals is a plug-in that can be added to any SYNGO based hosting applications (for example: Syngo Carbon Space, syngo.via etc…). The hosting application (native/syngo Platform-based software) is not described within this 510k submission. The hosting device decides which tools are used from Syngo Carbon Clinicals. The hosting device does not need to host all tools from the Syngo Carbon Clinicals, a desired subset of the provided tools can be used. The same can be enabled or disabled thru licenses.
When preparing the radiologist's reading workflow on a dedicated workplace or workstation, Syngo Carbon Clinicals can be called to generate additional results or renderings according to the user needs using the tools available.
This document describes performance evaluation for two specific tools within Syngo Carbon Clinicals (VA41): OrthoMatic Spine and TimeLens.
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Tool | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| OrthoMatic Spine | Algorithm's measurement deviations for major spinal measurements (Cobb angles, thoracic kyphosis angle, lumbar lordosis angle, coronal balance, and sagittal vertical alignment) must fall within the range of inter-reader variability. | Cumulative Distribution Functions (CDFs) demonstrated that the algorithm's measurement deviations fell within the range of inter-reader variability for the major Cobb angle, thoracic kyphosis angle, lumbar lordosis angle, coronal balance, and sagittal vertical alignment. This indicates the algorithm replicates average rater performance and meets clinical reliability acceptance criteria. |
| TimeLens | Not specified as a reader study/bench test was not required due to its nature as a simple workflow enhancement algorithm. | No specific quantitative performance metrics are provided, as clinical performance evaluation methods (reader studies) were deemed unnecessary. The tool is described as a "simple workflow enhancement algorithm". |
2. Sample Size Used for the Test Set and Data Provenance
-
OrthoMatic Spine:
- Test Set Sample Size: 150 spine X-ray images (75 frontal views, 75 lateral views) were used in a reader study.
- Data Provenance: The document states that the main dataset for training includes data from USA, Germany, Ukraine, Austria, and Canada. While this specifies the training data provenance, the provenance of the specific 150 images used for the reader study (test set) is not explicitly segregated or stated here. The study involved US board-certified radiologists, implying the test set images are relevant to US clinical practice.
- Retrospective/Prospective: Not explicitly stated, but the description of "collected" images and patients with various spinal conditions suggests a retrospective collection of existing exams.
-
TimeLens: No specific test set details are provided as a reader study/bench test was not required.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
-
OrthoMatic Spine:
- Number of Experts: Five US board-certified radiologists.
- Qualifications: US board-certified radiologists. No specific years of experience are mentioned.
- Ground Truth for Reader Study: The "mean values obtained from the radiologists' assessments" for the 150 spine X-ray images served as the reference for comparison against the algorithm's output.
-
TimeLens: Not applicable, as no reader study was conducted.
4. Adjudication Method for the Test Set
- OrthoMatic Spine: The algorithm's output was assessed against the mean values obtained from the five radiologists' assessments. This implies a form of consensus or average from multiple readers rather than a strict 2+1 or 3+1 adjudication.
- TimeLens: Not applicable.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- OrthoMatic Spine: A reader study was performed, which is a type of MRMC study. However, this was a standalone performance evaluation of the algorithm against human reader consensus, not a comparative effectiveness study with and without AI assistance for human readers. Therefore, there is no reported "effect size of how much human readers improve with AI vs without AI assistance." The study aimed to show the algorithm replicates average human rater performance.
- TimeLens: Not applicable.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- OrthoMatic Spine: Yes, a standalone performance evaluation of the OrthoMatic Spine algorithm (without human-in-the-loop assistance) was conducted. The algorithm's measurements were compared against the mean values derived from five human radiologists.
- TimeLens: The description suggests the TimeLens tool itself is a "simple workflow enhancement algorithm" and its performance was evaluated through non-clinical verification and validation activities rather than a specific standalone clinical study with an AI algorithm providing measurements.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- OrthoMatic Spine:
- For the reader study (test set performance evaluation): Expert consensus (mean of five US board-certified radiologists' measurements) was used to assess the algorithm's performance.
- For the training set: The initial annotations were performed by trained non-radiologists and then reviewed by board-certified radiologists. This can be considered a form of expert-verified annotation.
- TimeLens: Not specified, as no clinical ground truth assessment was required.
8. The Sample Size for the Training Set
- OrthoMatic Spine:
- Number of Individual Patients (Training Data): 6,135 unique patients.
- Number of Images (Training Data): A total of 23,464 images were collected within the entire dataset, which was split 60% for training, 20% for validation, and 20% for model selection. Therefore, the training set would comprise approximately 60% of both the patient count and image count. So, roughly 3,681 patients and 14,078 images.
- TimeLens: Not specified.
9. How the Ground Truth for the Training Set Was Established
- OrthoMatic Spine: Most images in the dataset (used for training, validation, and model selection) were annotated using a dedicated annotation tool (Darwin, V7 Labs) by a US-based medical data labeling company (Cogito Tech LLC). Initial annotations were performed by trained non-radiologists and subsequently reviewed by board-certified radiologists. This process was guided by written guidelines and automated workflows to ensure quality and consistency, with annotations including vertebral landmarks and key vertebrae (C7, L1, S1).
- TimeLens: Not specified.
Ask a specific question about this device
Page 1 of 1