Search Results
Found 27 results
510(k) Data Aggregation
(105 days)
Ask a specific question about this device
(22 days)
AI-Rad Companion Prostate MR is indicated for the processing and annotation of DICOM MR prostate images acquired in adult male populations that demonstrate indications of oncological abnormalities in the prostate.
The AI-Rad Companion Prostate MR software aims to support the radiologist and provides the following functionality:
• Viewing, analyzing, evaluating prostate MR images including DCE, ADC, T2 and DWI
• Hosting application for and provides interface to external Prostate MR AI plug-in device
• Accept/reject/edit the results generated by the plug-in software Prostate MR AI
AI-Rad Companion Prostate MR is a diagnostic aid in the interpretation of prostate MRI examinations acquired according to the PI-RADS standard.
AI-Rad Companion Prostate MR provides quantitative and qualitative information based on bi or multiparametric prostate MR DICOM images. It displays information on the segmented gland, prostate volume, and segmented lesions along with their classifications. This information can be used to support the reading and reporting of prostate MR studies, as well as the planning of prostate biopsies in the case of ultrasound guided MR-US fusion biopsies of the prostate gland.
The primary features of AI-Rad Companion Prostate MR include:
• Display of Automatic Segmentation and volume of the prostate gland as well as display of automatic segmentation, quantification and classification of lesions
• Manual Adjustment of gland and lesion segmentation and editing of lesion scores, diameter, and localization of the automated generated lesions
• Marking of new lesions
• Export of results as RTSS format for import into supporting ultrasound or fusion biopsy planning systems
Based on the provided FDA 510(k) clearance letter for AI-Rad Companion Prostate MR (K252608), there is no specific study described that proves the device meets predefined acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy). The document primarily focuses on demonstrating substantial equivalence to a predicate device (AI-Rad Companion Prostate MR K193283) and adherence to non-clinical verification and validation standards for software development and risk management.
The document explicitly states: "No clinical tests were conducted to test the performance and functionality of the modifications introduced within AI-Rad Companion Prostate MR."
Therefore, a table of acceptance criteria and reported device performance, information about sample sizes, expert ground truth establishment, adjudication methods, multi-reader multi-case studies, standalone performance, and training set details are not available in this document as no clinical performance study for the modified device was performed.
The document emphasizes that modifications and improvements were verified and validated through non-clinical tests (software verification and validation, unit, system, and integration tests), which demonstrated conformity to industry standards and the predicate device's existing safety and effectiveness.
Here’s a breakdown of what is stated in the document regarding testing:
1. A table of acceptance criteria and the reported device performance:
- Not provided. The document does not include a table of specific clinical acceptance criteria (e.g., target sensitivity or specificity values) or reported device performance metrics against such criteria. The focus is on demonstrating that software enhancements do not adversely affect safety and effectiveness, assuming the predicate device's performance was already acceptable.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not provided. Since no clinical performance study was conducted for this specific submission, details on test set sample sizes and data provenance are not presented.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. As no clinical study is reported, this information is not available.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not done. The document explicitly states "No clinical tests were conducted."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not explicitly stated for the modified device. While the device description mentions automatic segmentation and classification, the overall context emphasizes a "diagnostic aid" that "aims to support the radiologist" and has functionality to "Accept/reject/edit the results generated by the plug-in software Prostate MR AI." This suggests an interactive workflow where standalone performance is not the primary claim for this particular submission. The separate product, "Prostate MR AI (K241770)," which performs the core AI tasks, is likely where standalone performance would be detailed, but not in this document.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable for this submission, as no new clinical performance study is detailed for the modified device. The original predicate device's performance would have relied on a ground truth, but that information is not part of this document.
8. The sample size for the training set:
- Not provided. Since this submission is for an updated version of an already cleared device and no new clinical performance study is detailed, the training set size for the underlying AI model (likely part of K241770 or the predicate K193283) is not included here.
9. How the ground truth for the training set was established:
- Not provided. This information would typically be detailed in the original submission for the AI algorithm (likely K241770 or K193283), not in this update focused on software enhancements and substantial equivalence.
In summary, the provided document focuses on demonstrating that the enhancements and modifications to the AI-Rad Companion Prostate MR do not adversely affect the safety and effectiveness of the existing predicate device. It relies on non-clinical software verification and validation, and substantial equivalence arguments, rather than presenting a de novo clinical performance study with new acceptance criteria and results.
Ask a specific question about this device
(122 days)
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
The subject device, MAGNETOM Avanto Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Avanto Fit with syngo MR XA50A (K220151).
A high-level summary of the new and modified hardware and software is provided below:
For MAGNETOM Avanto Fit with syngo MR XA70:
Hardware
New Hardware:
myExam 3D Camera
BM Head/Neck 20
Modified Hardware:
Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
myExam Autopilot Brain
myExam Autopilot Knee
3D Whole Heart
HASTE_interactive
GRE_PC
Open Recon
Deep Resolve Gain
Fleet Reference Scan
Physio logging
complex averaging
AutoMate Cardiac
Ghost Reduction
BLADE diffusion
Beat Sensor
Deep Resolve Sharp
Deep Resolve Boost and Deep Resolve Boost (TSE)
Deep Resolve Boost HASTE
Deep Resolve Boost EPI Diffusion
Modified Features and Applications:
SPACE improvement (high band)
SPACE improvement (incr grad)
Brain Assist
Eco power mode
myExam Angio Advanced Assist (Test Bolus)
The subject device, MAGNETOM Skyra Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Skyra Fit with syngo MR XA50A (K220589).
A high-level summary of the new and modified hardware and software is provided below:
For MAGNETOM Skyra Fit with syngo MR XA70:
Hardware
New Hardware:
myExam 3D Camera
Modified Hardware:
Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
Beat Sensor
HASTE_interactive
GRE_PC
3D Whole Heart
Deep Resolve Gain
Open Recon
Ghost Reduction
Fleet Reference Scan
BLADE diffusion
HASTE diffusion
Physio logging
complex averaging
Deep Resolve Swift Brain
Deep Resolve Sharp
Deep Resolve Boost and Deep Resolve Boost (TSE)
Deep Resolve Boost HASTE
Deep Resolve Boost EPI Diffusion
AutoMate Cardiac
SVS_EDIT
Modified Features and Applications:
SPACE improvement (high band)
SPACE improvement (incr grad)
Brain Assist
Eco power mode
myExam Angio Advanced Assist (Test Bolus)
The subject device, MAGNETOM Sola Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA51A (K221733).
A high-level summary of the new and modified hardware and software is provided below:
For MAGNETOM Sola Fit with syngo MR XA70:
Hardware
New Hardware:
myExam 3D Camera
Modified Hardware:
Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
GRE_PC
3D Whole Heart
Ghost Reduction
Fleet Reference Scan
BLADE diffusion
Physio logging
Open Recon
Complex averaging
Deep Resolve Sharp
Deep Resolve Boost and Deep Resolve Boost (TSE)
Deep Resolve Boost HASTE
Deep Resolve Boost EPI Diffusion
AutoMate Cardiac
Implant suite
Modified Features and Applications:
SPACE improvement (high band)
SPACE improvement (incr grad)
Brain Assist
Eco power mode
The subject device, MAGNETOM Viato.Mobile with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato.Mobile with syngo MR XA51A (K240608).
A high-level summary of the new and modified hardware and software is provided below:
For MAGNETOM Viato.Mobile with syngo MR XA70:
Hardware
New Hardware:
n.a.
Modified Hardware:
Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
GRE_PC
3D Whole Heart
Ghost Reduction
Fleet Reference Scan
BLADE diffusion
Physio logging
Open Recon
Complex averaging
Deep Resolve Sharp
Deep Resolve Boost and Deep Resolve Boost (TSE)
Deep Resolve Boost HASTE
Deep Resolve Boost EPI Diffusion
AutoMate Cardiac
Implant suite
Modified Features and Applications:
SPACE improvement (high band)
SPACE improvement (incr grad)
Brain Assist
Eco power mode
Furthermore, the following minor updates and changes were conducted for the subject devices:
Low SAR Protocol minor update (for all subject devices but MAGNETOM Skyra Fit): the goal of the SAR adaptive protocols was to be able to perform knee, spine, heart and brain examinations with 50% of the max allowed SAR values in normal mode for head and whole-body SAR. The SAR reduction was achieved by parameter adaptations like Flip angle, TR, RF Pulse Type, Turbo Factor, concatenations. For cardiac clinically accepted alternative imaging contrasts are used (submitted with K232494).
Implementation of image sorting prepare for PACS (submitted with K231560).
Implementation of improved DICOM color support (submitted with K232494).
Needle intervention AddIn was added all subject device (submitted with K232494).
Inline Image Filter switchable for users: in the subject device, users have the ability to switch the "Inline image filter" (implicite Filter) on or off. This filter is an image-based filter that can be applied to specific pulse sequence types. The function of the filter remains unchanged from the previous device MAGNETOM Sola with syngo MR XA61A (K232535).
SVS_EDIT is newly added for MAGNETOM Skyra Fit, but without any changes (submitted with K203443)
Brain Assist received an improvement and is identical to that of snygo MR XA61A (K232535)
Open Recon is introduced for all systems. The function of Open Recon remains unchanged from the previous submissions (submitted with K221733).
Lock TR and FA in Bold received a minor UI update
Implant Suite is newly introduced for MAGNETOM Sola Fit and MAGNETOM Viato.Mobile, but without any changes (submitted with K232535)
myExam Autopilot Brain and myExam Autopilot Knee are newly introduced for the subject device MAGNETOM AVANTO Fit and are unchanged from previous submissions (submitted with K221733).
myExam Angio Advanced Assist (Test Bolus) received a bug fixing and minimal UI improvements.
The provided text is an FDA 510(k) clearance letter for various MAGNETOM MRI Systems. While it details new and modified software and hardware features, it does not include specific acceptance criteria or a study that "proves the device meets the acceptance criteria" in terms of performance metrics like sensitivity, specificity, or accuracy for a diagnostic task.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices. This is achieved by:
- Stating that the indications for use are the same.
- Listing numerous predicate and reference devices.
- Detailing hardware and software changes.
- Mentioning non-clinical tests like software verification and validation, sample clinical images, and image quality assessment to show that the new features maintain an "equivalent safety and performance profile" to the predicate devices.
- Referencing scientific publications for certain features to support their underlying principles and utility.
- Briefly describing the training and validation data for two AI features: Deep Resolve Boost and Deep Resolve Sharp, but without performance acceptance criteria or detailed results.
Therefore, much of the requested information cannot be extracted from this document because it is not a study report detailing clinical performance against predefined acceptance criteria for a specific diagnostic outcome.
However, I can extract the information related to the AI features as best as possible from the "AI Features/Applications training and validation" section (Page 16).
Acceptance Criteria and Study Details (Limited to AI Features)
1. Table of Acceptance Criteria and Reported Device Performance
| Feature | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Deep Resolve Boost | (Not explicitly stated in the provided document as specific numerical thresholds, but implied through evaluation metrics.) | "The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM). Most importantly, the performance was evaluated by visual comparisons to evaluate e.g., aliasing artifacts, image sharpness and denoising levels." (Exact numerical results not provided). |
| Deep Resolve Sharp | (Not explicitly stated in the provided document as specific numerical thresholds, but implied through evaluation metrics and verification activities.) | "The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. In addition, the feature has been verified and validated by inhouse tests. These tests include visual rating and an evaluation of image sharpness by intensity profile comparisons of reconstructions with and without Deep Resolve Sharp." (Exact numerical results not provided). |
2. Sample size used for the test set and the data provenance
- Deep Resolve Boost:
- Test Set Sample Size: Not explicitly stated as a separate "test set" size. The document mentions "training and validation data" for over 25,000 TSE slices, over 10,000 HASTE slices (for refinement), and over 1,000,000 EPI Diffusion slices. It's unclear what proportion of this was used specifically for final testing, or if the "validation" mentioned includes the final performance evaluation.
- Data Provenance: Retrospective, described as "Input data was retrospectively created from the ground truth by data manipulation and augmentation." Country of origin is not specified.
- Deep Resolve Sharp:
- Test Set Sample Size: Not explicitly stated as a separate "test set" size. The document mentions "training and validation" on more than 10,000 high resolution 2D images. Similar to Deep Resolve Boost, it's unclear what proportion was specifically for final testing.
- Data Provenance: Retrospective, described as "Input data was retrospectively created from the ground truth by data manipulation." Country of origin is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The definition of "ground truth" for the AI features refers to the acquired datasets themselves rather than expert-labeled annotations. Visual comparisons are mentioned as part of the evaluation, but without details on expert involvement or qualifications.
4. Adjudication method for the test set
This information is not provided in the document. While "visual comparisons" and "visual rating" are mentioned, no specific adjudication method (e.g., 2+1, 3+1) is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a MRMC comparative effectiveness study demonstrating human reader improvement with AI assistance is not described in this document. The focus of the AI features (Deep Resolve Boost and Deep Resolve Sharp) is on image quality enhancement (denoising, sharpness) and reconstruction rather than assisting human readers in a diagnostic task that can be quantified by an effect size.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the evaluation of Deep Resolve Boost and Deep Resolve Sharp, based on metrics like PSNR, SSIM, and perceptual loss, and "visual comparisons" or "visual rating" appears to be an assessment of the algorithm's performance in enhancing image quality in a standalone capacity, without direct human-in-the-loop interaction for diagnosis.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Deep Resolve Boost: "The acquired datasets (as described above) represent the ground truth for the training and validation." This implies the original, full-quality, unaltered MRI scan data. Further, "Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition Restricted of noise and mirroring of k-space data."
- Deep Resolve Sharp: "The acquired datasets represent the ground truth for the training and validation." Similar to Boost, this refers to original, high-resolution MRI scan data. For training, "k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation."
8. The sample size for the training set
- Deep Resolve Boost:
- TSE: more than 25,000 slices
- HASTE (for refinement): more than 10,000 HASTE slices
- EPI Diffusion: more than 1,000,000 slices
- Deep Resolve Sharp: more than 10,000 high resolution 2D images.
9. How the ground truth for the training set was established
- Deep Resolve Boost: The ground truth was established by the "acquired datasets" themselves (full-quality MRI scans). The training input data was then derived from this ground truth by simulating degraded images (e.g., under-sampling, adding noise).
- Deep Resolve Sharp: Similarly, the ground truth was the "acquired datasets" (high-resolution MRI scans). The training input data was derived by cropping k-space data to create corresponding low-resolution inputs.
Ask a specific question about this device
(219 days)
syngo Dynamics is a multimodality, vendor agnostic Cardiology image and information system intended for medical image management and processing that provides capabilities relating to the review and digital processing of medical images.
syngo Dynamics supports clinicians by providing post image processing functions for image manipulation, and/or quantification that are intended for use in the interpretation and analysis of medical images for disease detection, diagnosis, and/or patient management within the healthcare institution's network.
syngo Dynamics is not intended to be used for display or diagnosis of digital mammography images in the U.S.
syngo Dynamics is a software only medical device which is used with common IT hardware. Recommended configurations are defined for the hardware required to run the device, and hardware is not considered as part of the medical device.
syngo Dynamics is intended to be used by trained healthcare professionals in a professional healthcare facility to review, edit, and manipulate image data, as well as to generate quantitative data, qualitative data, and diagnostic reports.
syngo Dynamics is a digital image display and reporting system with flexible deployment – it can function as a standalone medical device that includes a DICOM Server or as an integrated module within an Electronic Health Record (EHR) System with a DICOM Archive that receives images from digital image acquisition devices such as ultrasound and x-ray angiography machines. There are three deployments: Standalone, EHR/EHS Integrated, and Multi-Modality Cardiovascular (MMCV). MMCV deployment functions as a standalone medical device with capability of natively support 2D and 3D CT and MR image types.
The use of syngo Dynamics is focused on cardiac ultrasound (echocardiography), angiography (x-ray), cardiac nuclear medicine (NM), CT and MR studies that cover both adult and pediatric medicine. Also supported is vascular ultrasound and ultrasound in Obstetrics/Gynecology and Maternal Fetal Medicine (fetal echocardiography during pregnancy).
syngo Dynamics is based on a client-server architecture. The syngo Dynamics server processes the data from the connected imaging modalities, and stores data and images to a DICOM server and routes them for permanent storage, printing, and review. The client provides the user interface for interactive image viewing, reporting, and processing; and can be installed on network connected workstations.
syngo Dynamics provides various semi-automated anatomical visualization tools.
syngo Dynamics offers multiple access strategies: A Workplace that provides full functionality for reading and reporting; A Remote Workplace that provides additionally compressed images with access to full fidelity images for reading and reporting; and a browser based WebViewer that provides access to additionally compressed images and reports from compatible devices (including mobile devices).
In the United States, monitors (displays) should not be used for diagnosis, unless the monitor (display) has specifically received 510(k) clearance for this purpose.
This FDA 510(k) clearance letter pertains to syngo Dynamics (Version VA41D), a Medical Image Management and Processing System (MIMPS). While the document broadly discusses the device's substantial equivalence to a predicate device (syngo Dynamics VA40F) and its general functionalities, the only specific AI/ML-enabled function for which performance data and acceptance criteria are detailed is the Auto EF algorithm for calculating left ventricular ejection fraction from ultrasound images.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based specifically on the Auto EF algorithm information provided:
1. Table of Acceptance Criteria and Reported Device Performance (Auto EF Algorithm)
The document states that "Additional acceptance criteria were defined with a total of 12 predetermined acceptance criteria," but only explicitly details one primary statistical criterion and provides summarized performance for a few other aspects.
| Acceptance Criterion | Reported Device Performance (syngo Dynamics VA41D) |
|---|---|
| Pearson's correlation coefficient (r) between biplane EF generated by Auto EF and ground truth $\ge 0.800$ | 0.822 (compared to 0.826 for predicate VA40F) |
| Increased percentage of cases with biplane EF results | 93.3% (140 of 150 cases, compared to 92.0% for predicate VA40F) |
| Bias of absolute EF | Minimal, -0.2% (unchanged from predicate VA40F) |
| Percentage of cases where absolute biplane EF delta between Auto EF and GT $\le$ 10% | 87.9% (compared to 83.7% for predicate VA40F) |
| All 12 predetermined acceptance criteria | Exceeded all 12 defined acceptance criteria. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: n = 150 cases.
- Data Provenance: The test data originated from 3 sites in the U.S., representing geographic diversity from 2 different regions. The data was collected retrospectively, as it was independent of the training data. The document states it is "representative of the intended use population for Auto EF" and balanced for gender, covering ages 21-93 years and BMIs 16.5-48.8. It also included data from three ultrasound manufacturers (Philips, GE, and Siemens).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: 2 experienced sonographers.
- Qualifications: "experienced sonographers." Specific details regarding their years of experience or board certifications are not provided in the document.
4. Adjudication Method for the Test Set
- Adjudication Method: The two sonographers worked independently to establish the ground truth. There is no mention of a formal adjudication process (e.g., 2+1, 3+1), arbitration by a third expert, or a consensus meeting after independent readings. They "did not have access to Auto EF when establishing the ground truth."
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No. The document explicitly states, "No clinical studies were carried out for syngo Dynamics (Version VA41D). All performance testing was conducted in a non-clinical fashion as part of the verification and validation activities for the medical device." The evaluation focused on the algorithm's performance against ground truth, not on human reader performance with or without AI assistance.
- Effect Size of Human Reader Improvement: Not applicable, as no MRMC study was performed.
6. Standalone (Algorithm-Only) Performance Study
- Standalone Study: Yes. The performance validation of the Auto EF algorithm was conducted in a standalone manner. The "Auto EF results with the subject device" were compared directly against the established ground truth. The algorithm processed the images and generated biplane EF values without human intervention in the calculation process, although the system allows users to "review, edit or reject the results."
7. Type of Ground Truth Used
- Type of Ground Truth: Expert consensus with a conventional manual method based on the "Method of Disks" (MOD), also known as the Modified Simpson's Rule. The ground truth was established by two independent sonographers calculating left ventricular volumes and ejection fraction.
8. Sample Size for the Training Set
- Training Set Sample Size: Not explicitly stated. The document mentions the algorithm was "re-trained with more training data" compared to the predicate device, but does not provide a specific number.
9. How the Ground Truth for the Training Set Was Established
- Training Set Ground Truth Establishment: Not explicitly detailed. The document only states that the "LV auto contouring algorithm has been updated with pre-training and additional annotated training data." It does not specify the method (e.g., expert consensus, manual contouring) or the number/qualifications of experts involved in annotating the training data. However, given that the test set ground truth was established by sonographers using the Method of Disks, it is highly probable that a similar methodology was used for the training data annotation.
Ask a specific question about this device
(197 days)
AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR pre-defined structures using deep-learning-based algorithms.
Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT.
The outputs of AI-Rad Companion Organs RT are intended to be used by trained medical professionals.
The software is not intended to automatically detect or contour lesions.
AI-Rad Companion Organs RT provides automatic segmentation of pre-defined structures such as Organs-at-risk (OAR) from CT or MR medical series, prior to dosimetry planning in radiation therapy. AI-Rad Companion Organs RT is not intended to be used as a standalone diagnostic device and is not a clinical decision-making software.
CT or MR series of images serve as input for AI-Rad Companion Organs RT and are acquired as part of a typical scanner acquisition. Once processed by the AI algorithms, generated contours in DICOMRTSTRUCT format are reviewed in a confirmation window, allowing clinical user to confirm or reject the contours before sending to the target system. Optionally, the user may select to directly transfer the contours to a configurable DICOM node (e.g., the Treatment Planning System (TPS), which is the standard location for the planning of radiation therapy).
AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept the automatically generated contours. Then the output of AI-Rad Companion Organs RT must be reviewed and, where necessary, edited with appropriate software before accepting generated contours as input to treatment planning steps. The output of AI-Rad Companion Organs RT is intended to be used by qualified medical professionals, who can perform a complementary manual editing of the contours or add any new contours in the TPS (or any other interactive contouring application supporting DICOM-RT objects) as part of the routine clinical workflow.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance Study for AI-Rad Companion Organs RT
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the AI-Rad Companion Organs RT device, particularly for the enhanced CT contouring algorithm, are based on comparing its performance to the predicate device and relevant literature/cleared devices. The primary metrics used are Dice coefficient and Absolute Symmetric Surface Distance (ASSD).
Table 3: Acceptance Criteria of AIRC Organs RT VA50
| Validation Testing Subject | Acceptance Criteria | Reported Device Performance (Summary) |
|---|---|---|
| Organs in Predicate Device | All organs segmented in the predicate device are also segmented in the subject device. | Confirmed. The device continued to segment all organs previously handled by the predicate. |
| The average (AVG) Dice score difference between the subject and predicate device is < 3%. | Confirmed. "For existing organs, the average (AVG) Dice score difference between the subject device and predicate device is smaller than 3%." | |
| New Organs for Subject Device | The subject device in the selected reference metric has a higher value than the defined baseline value. | Confirmed. "The performance results of the subject device for the new CT organs are comparable to the reference literature & cleared devices. Here equivalence for the new organs is defined such that the selected reference metric has a higher value than the defined baseline." |
Table 3: Performance Summary of the Subject Device CT Contouring (Overall Average Dice Coefficients)
| Anatomic Region | Avg Dice (%) | Std Dice (%) | 95% CI |
|---|---|---|---|
| Head & Neck | 76.1 | 14.3 | [75.1, 77.2] |
| Head & Neck lymph nodes | 69.3 | 13.9 | [68.7, 70.0] |
| Thorax | 76.9 | 15.8 | [76.2, 77.6] |
| Abdomen | 87.3 | 10.1 | [86.3, 88.2] |
| Pelvis | 85.7 | 9.6 | [85.0, 86.5] |
| Cardiac | 75.6 | 15.1 | [74.1, 77.1] |
Table 4: Detailed Performance Evaluation of the New Organs in the Subject Device (Selected Examples)
| Organ Name | No. | AVG Dice (%) | STD Dice (%) | MED Dice (%) | 95%CI Dice | AVG ASSD (mm) | STD ASSD (mm) | MED ASSD (mm) | 95%CI ASSD |
|---|---|---|---|---|---|---|---|---|---|
| Left Breast | 30 | 90.4 | 3.8 | 91 | [89, 91.8] | 2.4 | 2.2 | 1.8 | [1.5, 3.2] |
| Right Breast | 30 | 90.2 | 3.7 | 90.8 | [88.8, 91.5] | 1.9 | 0.7 | 1.8 | [1.7, 2.2] |
| Bowel Bag | 33 | 95 | 3.6 | 96.5 | [93.7, 96.3] | 1.9 | 1.5 | 1.4 | [1.4, 2.5] |
| Pituitary | 30 | 75.8 | 7.4 | 77 | [73.1, 78.6] | 0.7 | 0.3 | 0.6 | [0.5, 0.8] |
| Brainstem | 30 | 88.4 | 2.5 | 88.8 | [87.5, 89.3] | 1 | 0.3 | 0.9 | [0.9, 1.1] |
| Esophagus | 30 | 85.6 | 4.2 | 86 | [84, 87.2] | 0.6 | 0.3 | 0.6 | [0.5, 0.7] |
| MEDIASTINAL LN 9L | 31 | 38.3 | 21.1 | 42.9 | [30.6, 46.1] | 5.3 | 4.4 | 3.7 | [3.7, 6.9] |
(Note: The full Table 4 from the document provides detailed performance for all 37 new organs. This table includes a selection for illustrative purposes.)
2. Sample Sizes and Data Provenance
- Test Set Sample Size:
- CT Contouring Algorithm: N = 579 cases
- MR Contouring Algorithm: The MR algorithm is unchanged from the predicate, so its performance is unchanged. The predicate was validated using 66 cases.
- Data Provenance (CT Contouring Algorithm Test Set):
- Geographic Origin (Overall N=579): Data from multiple clinical sites across North American, South American, Asia, Australia, and Europe.
- Example Cohorts (Table 5: Validation Testing Data Information based on Cohort):
- Cohort A.1 (N=73): Germany (14), Brazil (59)
- Cohort A.2 (N=40): Canada (40)
- Cohort A.3 (N=301): South/North America (184), EU (44), Asia (33), Australia (28), Unknown (12)
- Cohort B (N=165): South/North America (100), EU (51), Asia (6), Australia (3), Unknown (5)
- Retrospective/Prospective: "retrospective performance study on CT data previously acquired for RT treatment planning."
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Number of Experts for Ground Truth: "a team of experienced annotators mentored by radiologists or radiation oncologists" for initial manual annotation. "a board-certified radiation oncologist" performed a quality assessment including review and correction of each annotation. The document does not specify an exact number of individuals for these teams, but describes the roles and qualifications.
- Qualifications of Experts:
- "experienced annotators"
- "radiologists or radiation oncologists" (mentors for annotators)
- "board-certified radiation oncologist" (for quality assessment/review)
4. Adjudication Method for the Test Set
The document describes the ground truth establishment process as: "manual annotation" by experienced annotators mentored by radiologists/radiation oncologists, followed by a "quality assessment including review and correction of each annotation was done by a board-certified radiation oncologist." This indicates a hierarchical review/correction process rather than a multi-reader consensus adjudication between equally-weighted readers (e.g., 2+1 or 3+1). The final accepted contour after the board-certified radiation oncologist's review served as the ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was described. The study focused on the standalone performance of the AI algorithm against established ground truth and comparison with a predicate device and literature. The document does not mention an effect size of how much human readers improve with AI vs. without AI assistance. The intended use specifies that the AI-generated contours must be reviewed, edited, and accepted by trained medical professionals, implying a human-in-the-loop workflow, but the validation study presented focuses on the AI's autonomous segmentation accuracy.
6. Standalone (Algorithm Only) Performance Study
Yes, a standalone performance study was done. The performance metrics (Dice coefficient, ASSD) and the comparison to an expert-established ground truth demonstrate the algorithm's autonomous segmentation capability. The study validated the "autocontouring algorithms" and their performance.
7. Type of Ground Truth Used
The ground truth used for the test set was expert consensus / manual annotation based on clinical guidelines. Specifically: "Ground truth annotations were established following RTOG and clinical guidelines using manual annotation." This was further reviewed and corrected by a board-certified radiation oncologist.
8. Sample Size for the Training Set
The document provides the sample sizes for the training set for new organs introduced:
- Table 6: Training Dataset Characteristics (Examples):
- Lacrimal Glands Left/Right: 247
- Pituitary Gland: 247
- Humeral Head Left/Right: 207
- Bowel Bag: 544
- Pelvic Bone Left/Right: 160
- Sacrum: 160
- Mediastinal LN (various): 136
- Femoral Head Left/Right: 160
- Brainstem: 247
- Esophagus: 247
- Breast Left/Right: 172
- Supraglottic Larynx: 247
- Glottis: 247
The total training set size for all organs is not explicitly summed, but these numbers indicate the scale of the training data used for the specific new organs.
9. How the Ground Truth for the Training Set Was Established
"In both the annotation process for the training and validation testing data, the annotation protocols for the OAR were defined following the applicable guidelines. The ground truth annotations were drawn manually by a team of experienced annotators mentored by radiologists or radiation oncologists using an internal annotation tool. Additionally, a quality assessment including review and correction of each annotation was done by a board-certified radiation oncologist using validated medical image annotation tools."
This indicates the same rigorous process of expert manual annotation and review was applied to establish ground truth for the training set as for the test set. The validation testing and training data were explicitly stated to be independent.
Ask a specific question about this device
(258 days)
Prostate MR AI is a plug-in Radiological Computer Assisted Detection and Diagnosis Software device intended to be used · with a separate hosting application · as a concurrent reading aid to assist radiologists in the interpretation of a prostate MRI examination acquired according to the PI-RADS standard · in adult men (40 years and older) with suspected cancer in treatment naïve prostate glands The plug-in software analyzes non-contrast T2 weighted (T2W) and diffusion weighted image (DWI) series to segment the prostate gland and to provide an automatic detection and segmentation of regions suspicious for cancer. For each suspicious region detected, the algorithm moreover provides a lesion Score, by way of PI-RADS interpretation suggestion. Outputs of the device should be interpreted consistently with ACR recommendations using all available MR data (e.g., dynamic contrast enhanced images [if available]). Patient management decisions should not be made solely based on analysis by the Prostate MR AI algorithm.
This premarket notification addresses the Siemens Healthineers Prostate MR AI (VA10A) Radiological Computer Assisted Detection and Diagnosis Software (CADe/CADx). Prostate MR AI is a Computer Assisted Detection and Diagnosis algorithm designed to plug into a hosting workflow that assists radiologists in the detection of suspicious lesions and their classification. It is used as a concurrent reading aid to assist radiologists in the interpretation of a prostate MRI examination acquired according to the PI-RADS standard. The automatic lesion detection requires transversal T2W and DWI series as inputs. The device automatically exports a list of detected prostate regions that are suspicious for cancer (each list entry consists of contours and a classification by Score and Level of Suspicion (LoS)), a computed suspicion map, and a per-case LoS. The results of the Prostate MR AI plug-in (with the case-level LoS, lesion center points, lesion diameters, lesion ADC median, lesion 10th percentile, suspicion map, and non-PZ segmentation considered optional) are to be shown in a hosting application that allows the radiologist to view the original case, as well as confirm, reject, or edit lesion candidates with their contours and Scores as generated by the Prostate MR AI plug-in. Moreover, the radiologist can add lesions with contours and PI-RADS scores and finalize the case. In addition, the outputs include an automatically computed prostate segmentation, as well as sub-segmentations of the peripheral zone and the rest of the prostate (non-PZ). The algorithm will augment the prostate workflow of currently cleared syngo.MR General Engine if activated via a separate license on the General Engine.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Automatic Prostate Segmentation | |
| Median Dice score between AI algorithm results and ground truth masks exceeds 0.9. | The median of the Dice score between the AI algorithm results and the corresponding ground truth masks exceeds the threshold of 0.9. |
| Median normalized volume difference between algorithm results and ground truth masks is within ±5%. | The median of the normalized volume difference between the algorithm results and the corresponding ground truth masks is within a ±5% range. |
| AI algorithm results are statistically non-inferior to individual reader variability (5% margin of error, 5% significance level). | The AI algorithm results as compared to any individual reader are statistically non-inferior based on variabilities that existed among the individual readers within the 5% margin of error and 5% significance level. |
| Prostate Lesion Detection and Classification | |
| Case-level sensitivity of lesion detection ≥ 0.80 for both radiology and pathology ground truth. | The case-level sensitivity of the lesion detection is equal or greater than 0.80 for both radiology and pathology ground truth. |
| False positive rate per case of lesion detection < 1 false positive per case for radiology ground truth. | The false positive rate per case of the lesion detection is smaller than one false positive per case for radiology ground truth. |
| Accuracy of PI-RADS classification of radiology ground truth lesions (detected by algorithm) ≥ 0.8. | The accuracy of the PI-RADS classification of radiology ground truth lesions detected by the algorithm is equal or greater than 0.8. |
| Non-inferior performance in GE vs Siemens and African American vs non-African American cases, and in cases with peripheral zone vs non-peripheral lesions. | The non-inferior performance of the subject device in GE vs Siemens and African American vs non-African American cases, and in cases with peripheral zone vs non-peripheral lesions was demonstrated. (Note: Specific metrics for this non-inferiority are not explicitly stated as distinct numerical criteria but are stated as "met".) |
| Clinical Performance (Reader Study - Case-level discrimination of Gleason Grade Group ≥ 1) | |
| Statistically significant improvement in AUROC for aided reading vs unaided reading. | Fully Inclusive Analysis: AUROC improved from 0.6758 (unaided) to 0.7010 (aided), difference of 0.0252 (95% C.I. [0.0011, 0.0493]; P=0.040). Maximally Restrictive Analysis: AUROC improved from 0.6579 (unaided) to 0.6948 (aided), difference of 0.0368 (95% C.I. [0.0108, 0.0628]; P=0.006). In both analyses, the improvement was statistically significant and the primary endpoint thus met. |
| Clinical Performance (Reader Study - Lesion-level reading performance) | |
| Statistically significant improvement in AUwAFROC for aided reading vs unaided reading. | Fully Inclusive Analysis: AUwAFROC improved in aided reading by 0.0350 (95% C.I.:[0.0020, 0.0681], P=0.037). Maximally Restrictive Analysis: AUwAFROC improved in aided vs. unaided reading by 0.302 (95% C.I.: [0.0080,0.0520], P=0.008). In both analyses, the improvement was statistically significant and the secondary endpoint thus met. |
| Statistically significant improvement in Fleiss' Kappa for interreader agreement in per-case PI-RADS scores for aided reading vs unaided reading. | Fleiss' Kappa improved from 0.283 (unaided) to 0.371 (aided), with a difference of 0.087 (95% C.I. [0.051, 0.125]). The improvement was statistically significant (P<0.0001). |
Study Information
2. Sample size used for the test set and the data provenance:
- Automatic Prostate Segmentation: 222 transversal T2 series.
- Provenance: More than 10 clinical sites.
- Retrospective/Prospective: Not explicitly stated, but the description of comparing against ground truth generated implies retrospective use of existing scans.
- Prostate Lesion Detection and Classification (Standalone Performance):
- 105 cases from 6 sites (against radiology ground truth).
- 115 cases from 6 sites (against pathology ground truth).
- 340 cases from the multi-reader multi-case study (used for evaluation, implied prospective for this part of the evaluation, but the cases themselves were retrospective for the reader study).
- Provenance: 6 sites (for 105 and 115 cases), and two US sites (for 340 cases).
- Retrospective/Prospective: The cases for the lesion detection and classification evaluation were used to compare against established ground truths, suggesting retrospective analysis of existing data. The cases for the reader study were retrospectively selected.
- Multi-Reader Multi-Case (MRMC) Study: 340 cases.
- Provenance: Two US sites. Cases were consecutive and specifically included additional consecutive patient cases from men of African descent to ensure at least 13% Black or African American ethnicity.
- Retrospective/Prospective: Cases were selected retrospectively.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Automatic Prostate Segmentation: 3 expert radiologists. No specific years of experience or subspecialty beyond "radiologists" are mentioned but implied as "expert".
- Prostate Lesion Detection and Classification (Radiology Ground Truth): 3 expert radiologists in prostate MRI reading.
- MRMC Study (Lesion-level reference standard): 3 experienced radiologists acting as Truthers.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Automatic Prostate Segmentation: Pixel-wise consensus among the 3 expert radiologists.
- Prostate Lesion Detection and Classification (Radiology Ground Truth): Consensus reading of the 3 expert radiologists.
- MRMC Study (Case-level reference standard): Biopsy results (Gleason Grade Group GGG ≥ 1), or for cases without biopsy, PSA density and follow-up data.
- MRMC Study (Lesion-level reference standard): Consensus lesions with a consensus PI-RADS of at least 3 from majority voting among the 3 experienced radiologists. (This implies a form of consensus/majority vote).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Yes, an MRMC study was done with a paired split-plot design, combining two fully-crossed MRMC sub-studies.
- Case-level AUROC improvement (discriminating Gleason Score ≥ 1):
- Fully Inclusive Analysis: +0.0252 (from 0.6758 unaided to 0.7010 aided).
- Maximally Restrictive Analysis: +0.0368 (from 0.6579 unaided to 0.6948 aided).
- Lesion-level AUwAFROC improvement:
- Fully Inclusive Analysis: +0.0350.
- Maximally Restrictive Analysis: +0.0302.
- Fleiss' Kappa (interreader agreement in per-case PI-RADS scores) improvement: +0.087 (from 0.283 unaided to 0.371 aided).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, standalone performance was evaluated for:
- Automatic Prostate Segmentation: Compared algorithm results to ground truth generated by radiologists.
- Prostate Lesion Detection and Classification: Compared automatic detection and classification results to radiology ground truth and pathology ground truth.
- MRMC Study (AI Standalone reference): The ROC curves shown graphically include a "grey curve [that] denotes AI standalone performance."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Automatic Prostate Segmentation: Pixel-wise consensus from 3 expert radiologists.
- For Prostate Lesion Detection and Classification:
- Consensus reading of 3 expert radiologists (radiology ground truth).
- Biopsy results for the same patient (pathology ground truth).
- For MRMC Study (Case-level): Biopsy results (Gleason Grade Group GGG ≥ 1), and in cases where biopsy was unavailable, PSA density and follow-up (12 months negative by PSA or MRI).
- For MRMC Study (Lesion-level): Consensus lesions with a consensus PI-RADS of at least 3 from majority voting among 3 experienced radiologists.
8. The sample size for the training set:
The document states: "The cases for the reader study were kept completely separate from those used for the training of the Prostate MR AI algorithm." However, it does not specify the sample size for the training set. It only mentions that the AI algorithm was "trained on a database of prostate MR image series acquired according to the PI-RADS standard (non-contrast T2W and DWI image series), and corresponding radiological and/or biopsy findings."
9. How the ground truth for the training set was established:
The ground truth for the training set was established based on "corresponding radiological and/or biopsy findings." Specific details on the adjudication method (e.g., number of experts, consensus process) for the training set are not provided in this document, only the source of the ground truth.
Ask a specific question about this device
(137 days)
MyAblation Guide is a software application for image processing, 2D/3D visualization, and comparison of medical images imported from multiple imaging modalities.
The software is controlled by the end user interface on a workstation with DICOM connectivity or as an integrated version on a Siemens CT scanner workstation.
The application is used to assist in the preparation and performance of ablative procedures, including of ablation targets, virtual ablation probe placement and contouring of ablated areas, as well as supporting the User in their assessment of the treatment. The application can only be used by trained Users.
The software is not intended for diagnosis and is not intended to predict ablation volumes or predict ablation success.
myAblation Guide is a software medical device that is used in the context of percutaneous ablative procedures with straight instruments. It is used by clinical professionals in a hospital premise; it can be either deployed on compatible CT scanners or a computer workstation.
The application is operated by medical professionals such as Interventional Radiologists and medical technologists with current license and/or certification as required by regional authority. myAblation Guide allows operating functions in an arbitrary sequence. In addition, it includes a structured sequence of steps for ease of utility.
The application supports anatomical datasets from CT, MR, CBCT, as well as PET/CT.
The application includes means and functionalities to support in:
· Multimodality viewing and contouring of anatomical, and multi-parametric images such as CT, CBCT, PET/CT, MRI
· Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
· Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- Manual and semi-automatic registration using rigid and deformable registration
- · Expansion of created contour structures to visualize a safety margin
· Functionality to support the user in creating virtual ablation needle paths and associated virtual ablation zones derived from manufacturer data
- · Export of virtual needle paths in the Dicom SSO format
- · Supports the user in comparing, contouring, and ablation needle planning based on datasets acquired with different imaging modalities
- Supports multimodality image fusion
- · Supports user's procedure flow via a task stepper
Thermal ablation cannot be triggered from myAblation Guide.
The provided text details the 510(k) submission for the myAblation Guide (VB80A) device. It includes information on non-clinical testing performed to demonstrate the device meets established design criteria.
Here's an organized breakdown of the acceptance criteria and study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Implied by Study Target/Reference) | Reported Device Performance (myAblation Guide) |
|---|---|---|
| Lesion Segmentation | Dice Score: 0.82 (from Moltz et al. study) | Dice Score: 0.65 (All Lesion Types) |
| Sensitivity: N/A | Sensitivity: 0.82 (All Lesion Types) | |
| Ablation Zone Segmentation | N/A (no specific numerical target stated) | Dice Score: 0.65 |
| N/A | Sensitivity: 0.95 |
Note on Acceptance Criteria: The document implies the Moltz et al. study's Dice coefficient of 0.82 on liver metastases as a benchmark, stating "the algorithm effectively demonstrated the segmentation of both hyperdense and hypodense lesions... With a Dice coefficient (Dice similarity index) of 0.82". For the internal study, the reported Dice scores and sensitivities appear to be the performance metrics being presented to demonstrate functionality rather than explicitly stated "acceptance criteria" that must be met. However, for the purpose of this exercise, we can infer that these reported values demonstrate the device's acceptable performance.
Study Details
-
Sample Size Used for the Test Set and Data Provenance:
- Lesion Segmentation (Moltz et al. study): 5 different datasets comprising 10 liver metastases. Data provenance is not specified (e.g., country of origin, retrospective/prospective).
- Lesion Segmentation (Internal Study): 50 patients. Data provenance is not specified (e.g., country of origin, retrospective/prospective), but it is referred to as an "internal study," suggesting it was conducted by the manufacturer or an affiliated entity.
- Ablation Zone Segmentation: 33 patients with 41 available ablation zones. Data provenance is not specified.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The document does not specify the number of experts or their qualifications for establishing ground truth for the test sets in either the Moltz et al. study or the internal studies.
-
Adjudication Method for the Test Set:
- The document does not provide details on any adjudication method (e.g., 2+1, 3+1, none) used for the test sets. It only mentions the comparison of algorithm performance against a reference.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical studies were carried out for the subject device, and therefore, no such clinical data is provided within this submission." The study focuses on "algorithm's performance" and "semi-automatic liver ablation zone segmentation."
-
Standalone (Algorithm Only Without Human-in-the-Loop) Performance:
- Yes, the performance data presented (Dice scores, Sensitivity) are indicative of standalone (algorithm only) performance for the semi-automatic segmentation algorithms. The phrasing "To assess the algorithm's performance" and "The internal analysis of the lesion segmentation" supports this. The device is a "software application for image processing," and the described tests evaluate the segmentation algorithms within this software.
-
Type of Ground Truth Used:
- The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). However, for segmentation tasks, ground truth is typically established by expert manual annotation or referencing pathology for pathological confirmation. Given the context of "assessed" cases and "segmentation," it is highly probable that the ground truth was established by expert review/annotation of the medical images.
-
Sample Size for the Training Set:
- The document does not specify the sample size for the training set. The provided information relates only to the test sets used for evaluating the semi-automatic segmentation algorithms.
-
How the Ground Truth for the Training Set Was Established:
- The document does not provide information on how the ground truth for the training set was established, as the size and specifics of the training set are not mentioned.
Ask a specific question about this device
(112 days)
Syngo Carbon Enterprise Access is indicated for display and rendering of medical data within healthcare institutions.
Syngo Carbon Enterprise Access is a software only medical device which is intended to be installed on recommended common IT Hardware. The hardware is not seen as part of the medical device. Syngo Carbon Enterprise Access is intended to be used in clinical image and result distribution for diagnostic purposes by trained medical professionals and provides standardized generic interfaces to connect to medical devices without controlling or altering their functions.
Syngo Carbon Enterprise Access provides an enterprise-wide web application for viewing DICOM, non-DICOM, multimedia data and clinical documents to facilitate image and result distribution.
The provided text is a 510(k) summary for the Syngo Carbon Enterprise Access (VA40A) device. It describes the device, its intended use, and compares it to a predicate device (Syngo Carbon Space VA30A). However, it explicitly states:
"No clinical studies were carried out for the product, all performance testing was conducted in a non-clinical fashion as part of verification and validation activities of the medical device."
Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment for a clinical study.
The document outlines "Non-clinical Performance Testing" and "Software Verification and Validation." These sections describe how the device's functionality was tested and validated to ensure it meets specifications and is substantially equivalent to the predicate device.
Here's what can be extracted about the "study" that proves the device meets "acceptance criteria" from a non-clinical performance testing and software verification/validation perspective:
1. Table of acceptance criteria and reported device performance:
The document broadly states that "The testing results support that all the software specifications have met the acceptance criteria" and "Results of all conducted testing were found acceptable in supporting the claim of substantial equivalence." However, it does not provide a specific, detailed table of acceptance criteria and quantitative performance metrics for each criterion. It rather focuses on demonstrating equivalence through feature comparison and general assertions of testing success.
2. Sample size used for the test set and the data provenance:
The document mentions "non-clinical tests" and "software verification and validation testing." It does not specify the sample size for the test set or the provenance of the data used for this non-clinical testing (e.g., country of origin, retrospective or prospective). This testing would typically involve various types of simulated data, pre-recorded medical images, and functional tests rather than patient studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Since no clinical studies were performed and the testing was non-clinical and technical, there's no mention of "experts" establishing ground truth in the way it would be done for a clinical performance study. The "ground truth" for non-clinical software testing would be derived from the product's functional specifications and expected outputs.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable, as no clinical study with human interpretation/adjudication was conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable, as no clinical studies, especially MRMC studies, were conducted. The device is a medical image management and processing system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The testing was on the device's functionality as a standalone software, but this refers to its technical performance in image management and display, not a diagnostic algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For this type of device and testing, the "ground truth" would be the functional specifications and expected behavior of the software based on its design documents. For example, if a function is to display a DICOM image, the ground truth is that the image should be displayed correctly according to DICOM standards. It's not based on clinical or pathological findings.
8. The sample size for the training set:
Not applicable. This device is a medical image management and processing system, not an AI/ML diagnostic algorithm that requires a "training set."
9. How the ground truth for the training set was established:
Not applicable, as there is no training set for this type of device.
In summary, the provided document focuses on demonstrating substantial equivalence through a comparison of features with a predicate device and general non-clinical verification and validation activities, rather than a clinical performance study with defined acceptance criteria and statistical analysis of human or AI diagnostic performance.
Ask a specific question about this device
(120 days)
AI-Rad Companion (Pulmonary) is image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from specialty care and general practice in the evaluation and assessment of disease of the lungs.
It provides the following functionality:
- Segmentation and measurements of complete lung and lung lobes
- · Identification of areas with lower Hounsfield values in comparison to a predefined threshold for complete lung and lung lobes
- · Providing an interface to external Medical Device syngo.CT Lung CAD
- · Segmentation and measurements of solid and sub-solid lung nodules
- Dedication of found lung nodules to corresponding lung lobe
- Correlation of segmented lung nodules of current scan with known priors and quantitative assessment of changes of the correlated data.
- Identification of areas with elevated Hounsfield values, where areas with elevated versus high opacities are distinquished.
The software has been validated for data from Siemens Healthineers (filtered backprojection and iterative reconstruction), GE Healthcare (filtered backprojection reconstruction), and Philips (filtered backprojection reconstruction).
Only DICOM images of adult patients are considered to be valid input.
The subject device AI-Rad Companion (Pulmonary) is an image processing software that utilizes machine learning and deep learning algorithms to provide quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support qualified clinicians in the evaluation and assessment of disease of the thorax. AI-Rad Companion (Pulmonary) builds on platform functionality provided by the AI-Rad Companion Engine and cloud/edge functionality provided by the Siemens Healthineers teamplay digital platform. AI-Rad Companion (Pulmonary) is an adjunct tool and does not replace the role of a qualified medical professional. AI-Rad Companion (Pulmonary) is also not designed to detect the presence of radiographic findings other than the prespecified list. Qualified medical professionals should review original images for all suspected pathologies.
AI-Rad Companion (Pulmonary) offers:
- Segmentation of lungs, ●
- Segmentation of lung lobes.
- Parenchyma evaluation, ●
- Parenchyma ranges,
- Pulmonary density,
- Visualization of segmentation and parenchyma results,
- Interface to LungCAD,
- Lesion segmentation, ●
- Visualization of lesion segmentation results, ●
- Lesion follow-up
AI-Rad Companion (Pulmonary) requires images of patients of 22 years and older.
AI-Rad Companion (Pulmonary) SW version VA40 is an enhancement to the previously cleared device AI-Rad Companion (Pulmonary) (K213713) that utilizes machine and deep learning algorithms to provide quantitative and qualitative analysis to computed tomography DICOM images to support qualified clinicians in the evaluation and assessment of disease of the thorax.
As an update to the previously cleared device, the following modifications have been made:
- Sub-solid Lung Nodule Segmentation ●
This feature provides the ability to segment and measure all subtypes of lesions including solid and sub-solid lesions.
- . Modified Indications for Use Statement The indications for use statement was updated to include descriptive text for sub-solid lung nodule addition.
- Updated Subject Device Claims List The claims list was updated to reflect the new device functionality
- . Updated Limitations for Use Additional limitations for use has been added to the subject device.
Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Validation Type | Target (Acceptance Criteria) | Reported Device Performance |
|---|---|---|
| Failure Rate | < 1% | Not explicitly stated as a percentage, but implied to have passed since overall performance goals were met. |
| Accuracy for Solid & Calcified Nodules | Average DICE ≥ predicate Bias & RMSE ≤ predicate LoA in line with predicate LoA | Average DICE coefficient was greater than that of the predicate. (Bias, RMSE, LoA not explicitly detailed beyond "met goals") |
| Accuracy for Sub-solid Nodules | Performance was analyzed in relation to the predicate's solid nodule segmentation performance Median nodule size range for sub-solid is 10mm-20mm LoA for 10mm-20mm ≥ 95% for all three diameter metrics | Average DICE coefficient for sub-solid nodules was superior to the average DICE coefficient of the predicate device for solid nodules. (LoA not explicitly detailed beyond "met goals") |
| DICE score | Average DICE score for sub-solid nodules > average DICE for predicate solid nodules | Average DICE coefficient for sub-solid nodules was superior to the average DICE coefficient of the predicate device for solid nodules (repetition of earlier point, but reinforces direct comparison). |
| Consistency of Subgroup results | Average DICE not smaller than DICE of overall cohort minus 1 STD Bias of three metrics not exceed ±1 STD RMSE of three metrics not exceed RMSE of overall cohort +1 STD each | The subject device met its individual subgroup analysis acceptance criterion for all subgroups. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 273 subjects from the United States and 254 subjects from Germany, for a total of 527 subjects.
- Data Provenance: The data originated from the United States (69% of cases) and Germany (31% of cases). The data was retrospective, as it refers to "previously acquired Computed Tomography DICOM images."
- Imaging Vendors: The test data included images from Canon/Toshiba (18%), GE (35%), Philips (15%), and Siemens (32%).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Two board-certified radiologists, with a third radiologist for adjudication.
- Qualifications:
- Radiologist 1: 10 years of experience (board-certified)
- Radiologist 2: 7 years of experience (board-certified)
- Adjudicating Radiologist 3: 9 years of experience
4. Adjudication Method for the Test Set
- Method: 2+1 (Two experts independently established ground truth, and in case of disagreement, a third expert served as an adjudicator.)
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- Was it done?: No, a traditional MRMC comparative effectiveness study involving human readers was not performed in the context of this specific submission. The study focuses on the standalone performance of the AI algorithm in comparison to the predicate device's performance, particularly for the new sub-solid nodule segmentation feature. The device is described as an "adjunct tool," but the presented study validates the algorithm's performance against expert consensus, not against human readers with and without AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Was it done?: Yes, the performance testing described directly evaluates the AI-Rad Companion (Pulmonary) lesion segmentation algorithm's accuracy (measured by DICE score, bias, and RMSE) against established ground truth. This is a standalone performance evaluation of the algorithm.
7. The Type of Ground Truth Used
- Type: Expert Consensus. The ground truth annotations for the test data were established independently by two board-certified radiologists, with a third radiologist serving as an adjudicator in cases of disagreement.
8. The Sample Size for the Training Set
- The sample size for the training set is not explicitly stated in the provided document. However, it is mentioned that "None of the clinical sites providing the test data provided data for training of any of the algorithms. Therefore there is a clear independence on site level between training and test data." This indicates that a distinct training set (or sets) was used.
9. How the Ground Truth for the Training Set was Established
- The document does not explicitly state how the ground truth for the training set was established. It only emphasizes the independence of the training and test data sites.
Ask a specific question about this device
(105 days)
syngo Virtual Cockpit is a software application intended for remote operation, assistance, review, monitoring, and standardization of medical imaging devices. It is a vendor neutral solution allowing read-only or full access control to connected devices. syngo Virtual Cockpit is also intended for training of medical personnel working on the medical imaging devices.
syngo Virtual Cockpit (sVC) is a software solution for geographically distant technologist or radiologists to remotely assist with operating imaging equipment and radiotherapeutic devices. sVC provides a private, secure communication platform for real-time image visualization and crossorganizational collaboration between healthcare professionals across multiple sites. sVC enables remote access to modality consoles and enhances communication capabilities between healthcare professionals across different locations. It is vendor-neutral and applicable to existing multimodalities in a healthcare network, a solution that allows healthcare professionals to share expertise and increase productivity, even when they are not physically present in the same location. sVC is based on a client server architecture, with sVC server as the backbone and 3 different variants of client, based on the user roles, Modality client & Physician client. Modality client as two flavors one windows based client and a web client which can be hosted in a web browser. Steering client establishes remote connection to Modality console / Modality acquisition workplace through KVM (Keyboard, Video and Mouse) switch or Siemens proprietary accessing tool, syngo Expert-i. Steering client can establish connections to more than one (up to 3) Modality console applications. Physician client is the third client for Physician that can be contacted either by Steering technologist or by Modality technologist for assistance regarding scanning in more complex cases, or the Physician can provide expert radiologist knowledge. The connection is possible in full control or read-only mode. The full-control accessibility to CT scanners is limited to the software associated with the modality workplace and is not applicable to the physical switches controlling the equipment operation. The connection to radiotherapeutic equipment is limited to be read-only. In addition to enabling remote access and control of the modality scanners, sVC also supports common communication methods including live videos at the modality site, audio calls and text chats among users.
The provided FDA 510(k) summary for syngo Virtual Cockpit (VB10A) focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study for novel performance claims. As such, it does not contain the typical elements of an acceptance criteria table or a comparative effectiveness study (such as MRMC) that would be expected for a diagnostic AI device requiring a clinical performance evaluation.
The device, "syngo Virtual Cockpit (VB10A)," is categorized as a "Medical Image Management And Processing System" (MIMPS) with product code LLZ and is a software-only solution intended for remote operation, assistance, review, monitoring, and standardization of medical imaging devices, and for training medical personnel. It is explicitly stated that "Images reviewed remotely are not for diagnostic use." This indicates it is not a diagnostic AI device that would produce specific diagnostic outputs requiring performance metrics like sensitivity, specificity, or AUC against a ground truth.
Therefore, the information typically requested in your prompt (e.g., sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC study effect sizes, standalone performance, type of ground truth) is not present in this document because the device's function does not necessitate such a clinical performance evaluation.
Here's an analysis based on the information available in the document:
Acceptance Criteria and Device Performance (as inferred from the document's V&V approach):
Instead of clinical performance metrics, the acceptance criteria are focused on the functional validity, safety, and effectiveness of the remote management system compared to its predicate and the general requirements for medical device software.
| Acceptance Criteria Category | Description (Inferred from document) | Reported Device Performance / Evidence |
|---|---|---|
| Functional Equivalence & Intended Use | Demonstrate that syngo Virtual Cockpit (sVC) enables remote access, assistance, review, monitoring, and standardization of medical imaging devices, similar to or expanded from the predicates, without raising new questions of safety or effectiveness. | "The intended use... is equivalent in that they all enable remote access to medical imaging devices and provide assistance... The subject, predicate and reference devices all allow remote users to help and assist modality technologist to display and review scanning protocols, observe and monitor the image acquisition and therefore help standardize scans in one institution." |
| Technical Performance (Latency) | For remote operation, the system should demonstrate acceptable delay/latency for real-time interaction. | Response time: ≤ 30 ms (for Expert-i method) / 60 ms (for KVM method with one modality). This is stated as "Equivalent to reference device." |
| Connectivity & Compatibility | Ensure stable and secure connections to supported medical imaging devices (including third-party via KVM switch) over a secured clinical network. | "The verification testing demonstrates that the sVC connection to modality scanners via KVM switch can perform as intended, meeting all of the design inputs." Supported modalities include medical imaging devices and radiotherapeutic devices (read-only). Supports multiple vendors. |
| Communication Features | Support essential communication methods for remote collaboration. | Includes "Screen sharing, IP cameras for live video, Audio calls and chat." Stated as "Equivalent. sVC has improved communication features." |
| Software Verification & Validation (V&V) | All software specifications must be met, and risks associated with clinically relevant functions assessed in a simulated clinical environment. Conformity to relevant consensus standards. | "Software validation is performed using externally sourced representative modality scanners... at the Siemens Healthineers training center and at the clinical collaborating site. The system configuration, connectivity, compatibility and operation... are assessed to validate the safety and effectiveness of the system in the simulated clinical environment." "All the software specifications have met the acceptance criteria." |
| Cybersecurity | Address cybersecurity considerations, including prevention of unauthorized access, modification, misuse, denial of use, or unauthorized use of information. | "Siemens Healthineers conforms to cybersecurity requirements by implementing a means to prevent unauthorized access..." |
| Risk Management | Identification and mitigation of identified hazards in compliance with ISO 14971. | "Risk Analysis... was completed and risk control implemented to mitigate identified hazards." |
| Safety and Effectiveness | The device should be safe and effective, performing as well as the predicate device without introducing new safety and effectiveness concerns. Output evaluated by clinicians to identify and intervene in case of malfunction. | "Results of all testing conducted were found acceptable in support to determine similarities to the predicate /previously cleared device." "The device does not come in contact with the patient and is only used by trained professionals. The output of the device is evaluated by clinicians, providing for sufficient review to identify and intervene in the event of a malfunction." |
Details Regarding the Study (as per provided text):
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated as a number of "cases" or "patients" in the traditional sense of a clinical study for diagnostic performance. The testing involved "externally sourced representative modality scanners" and "the entire system."
- Data Provenance: The testing was conducted "in the Siemens Healthineers training center and at the clinical collaborating site." This suggests controlled, simulated environments and potentially real clinical settings, but the origin of any "data" would be the performance of the system itself during validation, not pre-existing patient data. It was effectively a prospective functional validation.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable in the traditional sense. There is no "ground truth" established by experts for diagnostic purposes because the device does not make diagnostic claims or generate diagnostic outputs. The "ground truth" for this device would be its ability to correctly perform its intended functions (remote control, video, audio, chat, latency, etc.) and comply with safety and cybersecurity standards. The "evaluation by clinicians" refers to their use of the system and their ability to detect malfunctions during its operation, not a judgmental "ground truth" on diagnoses.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods are typically used to resolve discrepancies in expert readings or interpretations when establishing ground truth for diagnostic studies. This device's validation focused on functional performance and compliance.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was explicitly NOT done. The document states: "No clinical studies were carried out for syngo Virtual Cockpit (Version VB10A). All performance testing was conducted in a non-clinical fashion as part of the verification and validation activities for the medical device." This device is not designed to assist with human reading of images but rather with operating and managing the imaging process remotely.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not Applicable. The device, by its nature, is a human-in-the-loop system for remote operation. Its functionality is the human-in-the-loop performance (e.g., remote control inputs, video/audio communication). "Stand-alone" performance would refer to diagnostic output generation without human intervention, which is not what this device does.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not Applicable in the traditional sense for diagnostic performance. The "ground truth" relevant to this device's validation is the successful execution of its defined functional requirements and user stories, compliance with design inputs, and adherence to relevant safety and regulatory standards (e.g., IEC 62366-1, ISO 14971, IEC 62304). This is assessed through structured verification and validation testing rather than comparison to a clinical gold standard for disease.
-
The sample size for the training set:
- Not Applicable. This device is a software application for remote operation and management, not a machine learning or AI algorithm that trains on a dataset to learn patterns or make predictions. Therefore, there is no "training set" in the context of an AI model.
-
How the ground truth for the training set was established:
- Not Applicable. As there is no training set for an AI model, there is no ground truth associated with it.
In summary: The K232744 submission for syngo Virtual Cockpit (VB10A) is a 510(k) for a Medical Image Management and Processing System, not a diagnostic AI device. Its validation focuses on functional performance, safety, cybersecurity, and substantial equivalence to a predicate device based on those characteristics, rather than clinical performance metrics like accuracy, sensitivity, or specificity. Therefore, many of the questions regarding clinical study design, ground truth, and AI model training are not relevant to this specific device submission.
Ask a specific question about this device