Search Results
Found 58 results
510(k) Data Aggregation
(245 days)
The Atellica CH Diazo Total Bilirubin (D_TBil) assay is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates using the Atellica CH Analyzer. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Atellica CH Diazo Total Bilirubin is a photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. A specific mixture of detergents enables the determination of the total bilirubin.
N/A
Ask a specific question about this device
(262 days)
The Atellica® IM TSH3‑Ultra II (TSH3ULII) assay is for in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in human serum and plasma (EDTA and lithium heparin) using the Atellica® IM Analyzer. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
This assay is a third-generation assay that employs anti-FITC monoclonal antibody covalently bound to paramagnetic particles, an FITC-labeled anti-TSH capture mouse monoclonal antibody, and a tracer consisting of a proprietary acridinium ester and an anti‑TSH mouse monoclonal antibody conjugated to bovine serum albumin (BSA) for chemiluminescent detection.
N/A
Ask a specific question about this device
(222 days)
Atellica IM total PSA II (tPSAII) assay is for in vitro diagnostic use in the quantitative measurement of total prostate-specific antigen (PSA) in human serum and plasma (EDTA and lithium-heparin) using the Atellica IM Analyzer.
This assay is indicated as an aid in the detection of prostate cancer in conjunction with a digital rectal exam (DRE) in men aged 50 years and older. Prostate biopsy is required for diagnosis of prostate cancer. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer.
The Atellica IM total PSA II (tPSAII) assay consists of:
tPSAII ReadyPack® primary reagent pack
- Lite Reagent (10.0 mL/reagent pack): Unlabeled monoclonal mouse anti-fPSA antibody (~250 ng/mL); monoclonal mouse anti-PSA antibody (~180 ng/mL) labeled with acridinium ester; buffer; bovine serum albumin (BSA); preservative.
- Solid Phase (20.0 mL/reagent pack): Monoclonal mouse anti-PSA antibody (~3.5 μg/mL) labeled with biotin and bound to streptavidin paramagnetic particles; buffer; BSA; bovine gamma globulin (BGG); sodium azide (< 0.1%); preservative.
- Storage: Unopened at 2–8°C (Until expiration date on product), Onboard (42 days).
tPSAII CAL (2.0 mL/vial): Purified PSA from human seminal fluid in buffer; BSA; NaN3 (< 0.1%).
- Storage: Unopened at 2–8°C (Until expiration date on product), Opened at 2–8°C (30 days), On the system at room temperature (8 hours).
The tPSAII assay will have two configurations: 100 tests kit and 500 tests kit (5 x 100T in the carton).
N/A
Ask a specific question about this device
(261 days)
The ADVIA Centaur Anti-Thyroglobulin II (aTgII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroglobulin in human serum and plasma (EDTA, lithium heparin, sodium heparin) using the ADVIA Centaur XP system.
Anti-thyroglobulin (aTg) measurements are used, in conjunction with clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and Graves' disease.
| Component | Volume | Ingredients |
|---|---|---|
| ADVIA Centaur aTgII Primary Reagent ReadyPack (included in assay kit) | ||
| ADVIA Centaur aTgII Lite Reagent | 10.0 mL/pack | Human thyroglobulin labeled with acridinium ester (~1.2 µg/mL); buffered saline; bovine gamma globulin; |
| ADVIA Centaur aTgII Solid Phase Reagent | 20.0 mL/pack | Biotinylated human thyroglobulin bound to streptavidin-coated paramagnetic particles (~0.6 mg/mL); buffered saline; bovine gamma globulin; BSA; sodium azide (< 0.1%); preservative |
| ADVIA Centaur aTgII Ancillary Reagent ReadyPack (included in assay kit) | ||
| ADVIA Centaur aTgII Ancillary Reagent | 17.5 mL/pack | Goat serum; mouse serum; sodium azide (< 0.1%); preservative |
| ADVIA Centaur aTgII Calibrator (included in assay kit) | ||
| ADVIA Centaur aTgII Low and High Calibrators | 1.0 mL/vial Lyophilized | After reconstitution, low or high levels of monoclonal mouse anti-human thyroglobulin; goat serum; mouse serum; sodium azide (< 0.1%); preservative |
N/A
Ask a specific question about this device
(262 days)
The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XP system.
Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and/or Graves' disease.
The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) consists of:
- aTPOII ReadyPack® primary reagent pack (Lite Reagent, Solid Phase)
- aTPOII CAL
Devices sold separately and included in the ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) are: - ADVIA Centaur aTPOII MCM (MCM 1, MCM 2–4)
- ADVIA Centaur aTPOII QC
- ADVIA Centaur aTPOII DIL ReadyPack ancillary reagent pack
N/A
Ask a specific question about this device
(267 days)
The Atellica IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica IM Analyzer.
Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.
The Atellica IM Thyroglobulin (Tg) assay includes:
- Tg ReadyPack primary reagent pack:
- Lite Reagent: mouse monoclonal anti-human Tg antibody labeled with acridinium ester (~1.13 μg/mL); bovine serum albumin (BSA); mouse IgG; buffer; stabilizers; preservatives (7.5 mL/reagent pack).
- Solid Phase: streptavidin-coated paramagnetic microparticles preformed with biotinylated mouse monoclonal antihuman Tg antibody (~267 μg/mL); BSA; mouse IgG; buffer; stabilizers; preservatives (15.0 mL/reagent pack).
- Ancillary Well Reagent: BSA; bovine gamma globulin; buffer; preservatives (6.0 mL/reagent pack).
- Tg CAL: After reconstitution, human thyroglobulin; BSA; buffer; stabilizers; preservatives (2.0 mL/vial).
The following devices are sold separately:
- Atellica IM Tg MCM:
- MCM 1: After reconstitution, bovine serum albumin (BSA); buffer; stabilizers; preservatives (1.0 mL/vial).
- MCM 2–5: After reconstitution, various levels of human thyroglobulin; BSA; buffer; stabilizers; preservatives (1.0 mL/vial).
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Atellica IM Thyroglobulin (Tg) assay:
Device: Atellica IM Thyroglobulin (Tg) Assay
Purpose: Quantitative measurement of thyroglobulin in human serum and plasma as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes various performance characteristics, which serve as acceptance criteria for the device. The reported performance is directly from the summary.
| Acceptance Criteria Category | Specific Acceptance Criteria (implicit from study design) | Reported Device Performance |
|---|---|---|
| Detection Capability | LoB, LoD, LoQ determined per CLSI EP17-A2 | LoB: 0.039 ng/mL (0.059 pmol/L) LoD: 0.044 ng/mL (0.067 pmol/L) LoQ: 0.050 ng/mL (0.076 pmol/L) |
| Precision | Precision determined per CLSI EP05-A3 (within-laboratory and repeatability) | Repeatability (CV%): 1.2% - 6.4% across various concentrations Within-Laboratory Precision (CV%): 2.3% - 9.0% across various concentrations |
| Reproducibility | Reproducibility determined per CLSI EP05-A3 (across sites, runs, days) | Reproducibility (CV%): 1.9% - 5.8% across various concentrations |
| Linearity | Linearity determined per CLSI EP06-ed2 within stated assay range | Linear for 0.050–150 ng/mL (0.076–227 pmol/L) |
| Specimen Equivalence | Performance equivalence across serum, EDTA plasma, lithium heparin plasma | Performance confirmed equivalent across serum, EDTA plasma, lithium heparin plasma, and associated gel barrier tubes. |
| Interferences (HIL) | Bias < 10% for Hemoglobin, Bilirubin, Lipemia at specified concentrations | No bias > 10% observed for tested HIL substances. |
| Interferences (Other Substances) | Bias < 10% for various common substances/medications/biomarkers at specified concentrations | No bias > 10% observed for tested other substances. |
| Cross-Reactivity | Cross-reactivity < 1.0% for specified substances (T3, T4, TSH, Galectin-3, T2) | Cross-reactivity < 1.0% for tested substances. |
| Reagent Stability | Defined on-board and reconstituted calibrator stability | 28 days on-board; Calibrators stable 45 days (2-8°C) / 60 days (≤ -20°C, thaw once). |
| Sample Stability | Defined stability for various sample types and storage conditions | Stable 3-4 days (2-8°C), 4 days (RT), 12-24 months (frozen); ≤ 4 freeze-thaw cycles. |
| High Dose Hook Effect | No hook effect within a specified concentration range | No hook effect up to 80,000 ng/mL (121,200 pmol/L). |
| Expected Values | Reference intervals established per CLSI EP28-A3c | Healthy Adults: 2.44–74.9 ng/mL Post-thyroidectomy adults: < 1.27 ng/mL |
| Clinical Performance | Sensitivity and specificity calculated by comparing assay results to structural disease (SD) at a defined cut-off (0.2 ng/mL). Confidences intervals for these parameters. | Sensitivity: 98.2% (95% CI: 94.6%, 100.0%) Specificity: 53.4% (95% CI: 47.8%, 58.0%) PPV: 10.0% (95% CI: 8.7%, 11.2%) NPV: 99.8% (95% CI: 99.5%, 100.0%) |
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Performance Test Set Sample Size: 291 serum samples collected from 189 subjects.
- Data Provenance:
- The document states "A prospective, multi-center study was conducted." This indicates prospective data collection across multiple sites.
- The country of origin is not explicitly stated in the provided text.
- All samples were from subjects diagnosed with differentiated thyroid cancer, 6 or more weeks following thyroidectomy or radioiodine ablation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- The document does not specify the number of experts or their qualifications for establishing the ground truth (structural disease). It simply states: "SD [Structural Disease] was established and classified as either positive or negative by cross-sectional or functional imaging results."
- This suggests that the ground truth was derived from standard clinical imaging reports rather than a consensus of independent expert readers specifically for this study.
4. Adjudication Method for the Test Set
- The document does not describe an adjudication method for the test set's ground truth (structural disease). It implies that the imaging results themselves provided the classification. This means there was no adjudication process as typically seen with multiple human readers reviewing images.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was not done.
- This study is for an in vitro diagnostic (IVD) assay (a lab test), not an AI-assisted imaging or diagnostic tool where human readers work with or without AI. The performance metrics presented are for the analytical and clinical performance of the assay itself, comparing its results to a ground truth (structural disease status), not to human reader performance or improvement with AI.
6. If a Standalone Performance Study Was Done
- Yes, this is effectively a standalone (algorithm only) performance study.
- The Atellica IM Tg assay is an automated in vitro diagnostic device. Its performance characteristics (sensitivity, specificity, precision, linearity, etc.) are evaluated intrinsically, independent of human interpretation of the assay result values. The output is a quantitative measurement of thyroglobulin.
7. The Type of Ground Truth Used
- Ground truth for clinical performance: Structural disease (SD) status obtained from "cross-sectional or functional imaging results."
- Ground truth for analytical performance (LoB, LoD, LoQ, Precision, etc.): Established through laboratory protocols and reference materials (e.g., CLSI guidelines, certified reference materials like BCR CRM 457, spiked samples, control materials).
8. The Sample Size for the Training Set
- The document does not specify a separate training set or its sample size for the Atellica IM Tg assay.
- For IVD assays like this, the "training" is typically inherent in the assay's development and optimization process (e.g., reagent formulation, calibration curve development), which uses various known samples and standards, rather than a distinct, labeled "training dataset" as would be seen for a machine learning algorithm. The performance characteristics studies presented are akin to a "verification/validation set."
9. How the Ground Truth for the Training Set Was Established
- As a traditional IVD assay, there isn't a "training set" in the sense of a machine learning model.
- Ground truth for assay development and calibration: This would have been established using reference materials (like BCR CRM 457), characterized control samples, and potentially a large panel of clinically characterized patient samples used during the assay's development and optimization phases. These activities are part of the broader product development lifecycle rather than a distinct "training set" with ground truth generated by experts in the context of a clinical study for submission. Standardization is explicitly noted as traceable to BCR CRM 457, which serves as a primary standard for establishing the quantitative accuracy of the assay.
Ask a specific question about this device
(35 days)
Ask a specific question about this device
(192 days)
The ADVIA Centaur® TSH3-Ultra II (TSH3ULII) assay is for in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
This assay is a third-generation assay that employs anti-FITC monoclonal antibody covalently bound to paramagnetic particles, an FITC-labeled anti-TSH capture mouse monoclonal antibody, and a tracer consisting of a proprietary acridinium ester and an anti-TSH mouse monoclonal antibody conjugated to bovine serum albumin (BSA) for chemiluminescent detection.
The provided text describes the ADVIA Centaur® TSH3-Ultra II (TSH3ULII) assay, a device for in vitro diagnostic quantitative determination of thyroid-stimulating hormone (TSH). The document covers the device's indications for use, comparison with a predicate device, and performance characteristics data.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria (Design Goal) | Reported Device Performance |
|---|---|---|
| Detection Capability | N/A (LoB, LoD, LoQ are reported values, not acceptance criteria for determination) | - Limit of Blank (LoB): 0.005 µIU/mL (mIU/L) - Limit of Detection (LoD): 0.008 µIU/mL (mIU/L) - Limit of Quantitation (LoQ): 0.010 µIU/mL (mIU/L) (Lower than predicate device's LoQ of 0.008 µIU/mL, but within acceptable range for the new device as specified in assay range) |
| Precision | - Repeatability (Within-Run): - ≤ 12% CV for 0.020–0.299 µIU/mL (mIU/L) - ≤ 6% CV for ≥ 0.300–90.000 µIU/mL (mIU/L) - ≤ 7% CV for > 90.000 µIU/mL (mIU/L) - Within-Laboratory (Total Precision): - ≤ 16% CV for 0.020–0.299 µIU/mL (mIU/L) - ≤ 8% CV for ≥ 0.300–90.000 µIU/mL (mIU/L) - ≤ 10% CV for > 90.000 µIU/mL (mIU/L) | Reported values (all calculated Repeatability CV and Within-Laboratory CV values are within the specified limits): - Serum A (0.088 µIU/mL): Repeatability CV 2.5%, Within-Lab CV 3.6% - Serum B (0.196 µIU/mL): Repeatability CV 1.8%, Within-Lab CV 3.1% - Serum C (0.507 µIU/mL): Repeatability CV 1.7%, Within-Lab CV 2.6% - Serum D (4.752 µIU/mL): Repeatability CV 2.3%, Within-Lab CV 2.7% - Serum E (46.749 µIU/mL): Repeatability CV 2.4%, Within-Lab CV 4.0% - Serum F (97.929 µIU/mL): Repeatability CV 2.2%, Within-Lab CV 3.5% Similar acceptable results for Plasma and Controls. |
| Reproducibility | - Reproducibility (Total): - ≤ 18.5% CV for 0.020-0.299 µIU/mL (mIU/L) - ≤ 10.5% CV for ≥ 0.300-90.000 µIU/mL (mIU/L) - ≤ 12.5% CV for > 90.000 µIU/mL (mIU/L) | Reported values (all calculated Reproducibility CV values are within the specified limits): - Serum A (0.090 µIU/mL): Reproducibility CV 3.11% - Serum B (0.178 µIU/mL): Reproducibility CV 4.87% - Serum C (0.474 µIU/mL): Reproducibility CV 2.21% - Serum D (4.684 µIU/mL): Reproducibility CV 2.47% - Serum E (56.562 µIU/mL): Reproducibility CV 2.33% - Serum F (99.522 µIU/mL): Reproducibility CV 4.12% Similar acceptable results for Plasma and Controls. |
| Assay Comparison | - Correlation coefficient (r) ≥ 0.95 - Slope of 1.0 ± 0.1 | - Correlation coefficient (r): 0.999 - Regression Equation (Slope): 0.97 (within 1.0 ± 0.1) |
| Specimen Equivalency | - Correlation coefficient (r) ≥ 0.95 - Slope of 0.95-1.05 | - Plasma, EDTA vs. Serum: r = 0.999, Slope = 0.99 (within 0.95-1.05) - Plasma, lithium heparin vs. Serum: r = 0.990, Slope = 1.01 (within 0.95-1.05) |
| Interferences (HIL) | Bias due to hemoglobin, bilirubin (conjugated/unconjugated), and Intralipid does not exceed 10% | Hemoglobin (500mg/dL), Bilirubin (40mg/dL), Intralipid (1000mg/dL) do not exceed 10% bias at TSH concentrations of ~0.900 µIU/mL and ~8.000 µIU/mL. |
| Interferences (Other Substances) | Bias due to various common substances does not exceed 10% | Various substances (e.g., Acetaminophen, Aspirin, Biotin, Heparin, Ibuprofen, Levothyroxine) at specified concentrations do not exceed 10% bias at TSH concentrations of ~0.900 µIU/mL and ~8.000 µIU/mL. |
| Cross-Reactivity | Cross-reactivity of hCG, FSH, and LH does not exceed 5% | hCG (200000 mIU/mL), FSH (1500 mIU/mL), LH (600 mIU/mL) at specified concentrations do not exceed 5% cross-reactivity at TSH concentrations of ~0.400, 5.00, 17.00, and 90.00 µIU/mL. |
| Linearity | Device is linear throughout its measuring interval (0.010–150.000 µIU/mL) | "The assay is linear for the measuring interval of 0.010–150.000 µIU/mL (mIU/L)." |
| High-Dose Hook Effect | Results for TSH concentrations above the measuring interval and up to 3000 µIU/mL should report > 150 µIU/mL (not a paradoxical decrease) | "Patient samples with TSH concentrations above the measuring interval and as high as 3000 µIU/mL will report > 150 µIU/mL (mIU/L)." (This confirms the absence of a significant high-dose hook effect within this specified range, meaning the device displays the result as above the measurable range.) |
The study that proves the device meets the acceptance criteria is detailed across the "Performance Characteristics Data" section (Section 8) of the 510(k) Summary.
2. Sample Size Used for the Test Set and Data Provenance
- Detection Capability (LoQ): Not specified (CLSI Document EP17-A2 was followed).
- Precision: 80 measurements (replicates of 2, 2 runs/day, 20-day protocol) for each of the 6 serum samples, 5 plasma samples, and 5 control samples. Total N for Precision study is 80 x (6+5+5) = 1280 measurements.
- Reproducibility: 225 measurements (replicates of 5, 1 run/day, 5-day protocol) for each of the 6 serum samples, 5 plasma samples, and 5 control samples. Total N for Reproducibility study is 225 x (6+5+5) = 3600 measurements.
- Assay Comparison: 973 samples.
- Specimen Equivalency:
- Plasma, EDTA vs. Serum: 52 samples.
- Plasma, lithium heparin vs. Serum: 57 samples.
- Interferences (HIL and Other Substances): Samples at two TSH concentrations (~0.900 µIU/mL and ~8.000 µIU/mL) were tested for each interfering substance. The exact number of individual samples tested is not given, but multiple samples would be required for the two TSH levels per substance.
- Cross-Reactivity: Samples at four TSH concentrations (~0.400, 5.00, 17.00, and 90.00 µIU/mL) were spiked with hCG, FSH, or LH. The exact number of individual samples is not given.
- Linearity: Not explicitly stated, but performed in accordance with CLSI Document EP06-ed2, which involves testing multiple diluted samples.
- High-Dose Hook Effect: Samples with TSH concentrations up to 3000 µIU/mL were evaluated. The number of samples tested is not explicitly stated.
Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given it's an in vitro diagnostic device, the samples would generally be human biological specimens, likely collected from a clinical laboratory setting. The use of CLSI documents (Clinical and Laboratory Standards Institute) suggests standard laboratory practices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. For an in vitro diagnostic assay like TSH, "ground truth" is typically established by:
- The reference method against which the new assay is compared (for accuracy/assay comparison). In this case, "ADVIA Centaur TSH3-UL assay" is used as the comparative assay (the predicate device).
- Traceability to an international standard (WHO 3rd International Reference Preparation for human TSH (IRP 81/565)), which implies that the TSH values are calibrated against a universally accepted standard, rather than expert consensus on individual patient cases.
4. Adjudication Method for the Test Set
This refers to the process of resolving discrepancies in expert opinions, which is not applicable here as it is an analytical performance study for an IVD, not an interpretative AI device requiring human expert label agreement.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-assisted diagnostic imaging devices where human readers interpret medical images with and without AI assistance. This document pertains to an in vitro diagnostic assay, which involves automated quantitative measurement of a biomarker.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, implicitly. The entire performance characterization (Detection Capability, Precision, Reproducibility, Assay Comparison, Specimen Equivalency, Interferences, Cross-Reactivity, Linearity, High-Dose Hook Effect) is describing the standalone performance of the TSH3-Ultra II assay as an automated laboratory test. There is no mention of "human-in-the-loop" for this device's intended diagnostic function.
7. Type of Ground Truth Used
The ground truth for the performance studies is established by:
- Reference materials/standards: For accuracy, the assay is traceable to the World Health Organization (WHO) 3rd International Reference Preparation for human TSH (IRP 81/565).
- Comparative method: For assay comparison, the predicate device (ADVIA Centaur TSH3-UL assay) results serve as the comparative standard.
- Defined concentrations: For precision, linearity, interferences, and cross-reactivity, samples with known or spiked concentrations of TSH or interfering substances are used.
8. Sample Size for the Training Set
The document does not report a training set sample size. This is because the ADVIA Centaur TSH3-Ultra II is a chemical immunoassay, not a machine learning or AI-based device that would typically involve a "training set" in the computational sense. The "development" process for such an assay involves reagent formulation, assay optimization, and calibration curve development, which are distinct from training an AI model.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" in the context of an AI/ML model, this question is not applicable. The assay's analytical characteristics are established through various studies (precision, accuracy, linearity, etc.) using calibrated materials and established reference methods, as detailed in the performance characteristics.
Ask a specific question about this device
(601 days)
The ADVIA Centaur® NT-proBNPII (PBNPII) assay is for in vitro diagnostic use in the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system.
In the Emergency Department (ED) and Outpatient (OP) populations, measurements of NT-proBNP are used as an aid in the diagnosis of heart failure (HF) in patients with clinical suspicion of new onset or worsening HF.
The ADVIA Centaur® NT-proBNPII (PBNPII) assay kit includes the Primary Reagent ReadyPack and the Calibrator. The Primary Reagent ReadyPack contains Lite Reagent, Solid Phase Reagent, and Ancillary Well Reagent. The Calibrator includes Low and High Calibrators which are lyophilized.
The provided document discusses the ADVIA Centaur® NT-proBNPII (PBNPII) assay, an in vitro diagnostic device for measuring N-terminal pro-brain natriuretic peptide (NT-proBNP) to aid in the diagnosis of heart failure in Emergency Department (ED) and Outpatient (OP) populations. The submission aims to demonstrate substantial equivalence to the predicate device, the Roche Elecsys proBNP II assay (K072437).
Here's an analysis of the acceptance criteria and the studies that support them:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" as a single, consolidated table with pass/fail values for all performance characteristics. Instead, it presents various performance studies with their results. Based on the "Comparison of Technological Characteristics with the Predicate Device" and the "Performance Characteristics" sections, we can infer some criteria and compare the device's performance.
| Performance Characteristic | Acceptance Criteria (Implied/Predicate) | Reported Device Performance (ADVIA Centaur PBNPII) |
|---|---|---|
| Intended Use | Aid in diagnosis of suspected congestive heart failure. | Aid in diagnosis of HF in ED and OP populations with clinical suspicion of new onset or worsening HF. |
| Measurement | Quantitative | Quantitative |
| Technology | Chemiluminescence immunoassay | Chemiluminescence immunoassay (1-step sandwich) |
| Sample Type | Plasma and Serum | Human serum, plasma (EDTA and lithium heparin) |
| Assay Range | 5-35,000 pg/mL (Predicate) | 35-35,000 pg/mL |
| Hook Effect | No hook effect up to 300,000 pg/mL (Predicate) | No hook effect up to 300,000 pg/mL (will report >35,000 pg/mL) |
| Precision (Total CV) | For Serum: e.g., <4.6% @ 44 pg/mL, <1.8% @ 2,410 pg/mL (Predicate) | For Serum: 6.7% @ 116 pg/mL, 4.0% @ 271 pg/mL, 2.3% @ 380 pg/mL, 2.1% @ 806 pg/mL, 2.0% @ 1,597 pg/mL, 2.8% @ 25,073 pg/mL. QCs also within acceptable limits (presumably). |
| Reproducibility | Not explicitly stated, typically within acceptable CV limits across sites. | Total CVs ranging from 3.0% to 5.0% for serum samples (141-10428 pg/mL) and 3.3% to 4.5% for controls. |
| Linearity | N/A (implied to be demonstrated across assay range) | Linear for 35-35,000 pg/mL |
| Limit of Blank (LoB) | LoB < LoD (Predicate) | 13 pg/mL |
| Limit of Detection (LoD) | 5.00 pg/mL (Predicate) | 20 pg/mL |
| Limit of Quantitation (LoQ) | 50.0 pg/mL (Predicate) | 35 pg/mL |
| Interference (HIL & Other) | Bias not to exceed 10% (Implied) | Bias due to hemoglobin, bilirubin, lipemia <10%; various other substances also do not interfere (bias <10%). |
| Cross-Reactivity | Not Detectable / <1.0% for most (Implied, similar to predicate) | Most cross-reactants <1.0% or not detectable. ProBNP (glycosylated) 1.0%-19.0%, ProBNP (non-glycosylated) 13.0%-30.0%. |
| Specimen Equivalency | Regression equation close to y=x+0, r close to 1 (Implied) | Plasma (EDTA, Lithium heparin), SST, RST vs. Serum: Regression equations y=1.00x +/- small constant, r ~0.999-1.000. |
| Clinical Performance (ED Pop.) | Likelihood Ratios (LR+, LR-) and Post-test Risk (Implied to be clinically useful for diagnosis) | Age-dependent rule-in/out cutoffs provided with LR+ and LR- (0.06-0.09 for Negative LR-, suggesting good rule-out). |
| Clinical Performance (OP Pop.) | Sensitivity, Specificity, PPV, NPV (Implied to be clinically useful for diagnosis) | Overall: Sensitivity 86.0%, Specificity 64.3%, PPV 34.4%, NPV 95.5% (at 125 pg/mL cutoff). Detailed breakdown by age/sex also provided. |
Note: The "Acceptable Criteria" values for precision were primarily taken directly from comments in the tables (indicated by "<" followed by a number), which appear to be the applicant's internal criteria rather than directly from the predicate. The predicate's precision data is shown for comparison purposes in the "Comparison of Technological Characteristics."
2. Sample Size Used for the Test Set and Data Provenance
- Precision and Reproducibility:
- Precision: Assayed in duplicate in 2 runs per day for 20 days. Number of samples not explicitly stated but implied to be multiple levels of serum and QCs (6 serum, 3 QC samples listed).
- Reproducibility: N=90 samples assayed in duplicate in 2 runs per day for 5 days at 3 sites.
- Linearity: Not specified.
- Detection Limit (LoB, LoD, LoQ): Not specified.
- Interference (HIL & Other Substances): Not specified.
- Cross-Reactivity: Not specified.
- Specimen Equivalency: N=50 samples for each tube type comparison to serum.
- Expected Values (Reference Study Group): 723 apparently healthy subjects (362 females, 361 males).
- Clinical Performance (ED Population): 3128 subjects with signs and symptoms of acute HF who presented to the ED. 1148 adjudicated as acute HF, 1980 as without HF.
- Clinical Performance (OP Population): 1033 OP subjects with signs and symptoms of new onset HF. 185 adjudicated as new onset HF, 848 as without HF.
Data Provenance: The document does not explicitly state the country of origin for the clinical study data or if it was retrospective or prospective. However, the term "prospectively enrolled" is used for both the ED and OP clinical performance studies, indicating these were prospective studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Clinical Performance (ED Population): An "independent central adjudication panel of expert clinicians (Cardiologists)" determined the diagnosis and severity of HF. The exact number of experts is not stated, nor are their specific qualifications (e.g., number of years of experience).
- Clinical Performance (OP Population): An "independent central adjudication panel of expert clinicians (Cardiologists)" determined the diagnosis of new onset HF. Similar to the ED population, the exact number and specific qualifications are not detailed.
4. Adjudication Method (for the test set)
The adjudication method used by the "independent central adjudication panel of expert clinicians (Cardiologists)" is not explicitly described (e.g., 2+1, 3+1, none). It only states they "determined" the diagnosis.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic assay (laboratory test), not an imaging or interpretive device that typically involves human readers in the same way an MRMC study would apply. Therefore, there is no effect size reported for human readers improving with or without AI assistance.
6. Standalone Performance Study
Yes, a standalone (algorithm only) performance study was performed for the assay. The entire "Performance Characteristics" section (9.1 through 9.12) describes the analytical and clinical performance of the ADVIA Centaur PBNPII assay in isolation, without human interpretation of the assay results in the context of the study design. The clinical utility is assessed by how accurately the assay's quantitative results correlate with the adjudicated clinical diagnosis.
7. Type of Ground Truth Used
- Clinical Performance (ED and OP Populations): Clinical diagnosis adjudicated by an "independent central adjudication panel of expert clinicians (Cardiologists)." This falls under expert consensus clinical diagnosis.
- Analytical Performance (Precision, Linearity, Detection Limits, Interference, Cross-Reactivity, Specimen Equivalency): These are quantitative measurements against established reference methods, standards, or known concentrations, serving as the ground truth for analytical validity.
8. Sample Size for the Training Set
The document describes an in vitro diagnostic assay for measuring biomarkers. These types of assays typically do not have a "training set" in the same sense as machine learning algorithms, which require labeled data for model development. Instead, they involve method development and validation experiments. The document describes studies for:
- Establishing expected values (reference study group, N=723).
- Establishing disease study groups for statistical analysis of correlation (ED population N=1148, OP population N=185).
- Various analytical validation studies such as precision, linearity, and detection limits.
Therefore, the concept of a "training set" for the assay, in the context of algorithm development, is not directly applicable or discussed. The studies mentioned above contribute to the overall validation and characterization of the assay's performance.
9. How the Ground Truth for the Training Set Was Established
As mentioned above, the concept of a "training set" for an assay like this is not directly applicable. If we consider the data used to characterize the assay's performance as analogous to "training data" (i.e., the data used to define the assay's characteristics and clinical cut-offs), then:
- Clinical Cut-offs: The clinical cut-offs for the ED and OP populations were established based on the measured NT-proBNP values in patient populations (ED and OP disease study groups) where the diagnosis of heart failure was determined by an "independent central adjudication panel of expert clinicians (Cardiologists)." This expert consensus clinical diagnosis serves as the ground truth for establishing the clinical utility of the various cut-off values.
- Reference Intervals: The reference interval was established using 723 apparently healthy subjects. The "healthy" status would have been determined through standard medical screening and criteria, serving as the ground truth for defining the normal range.
- Analytical Studies: Ground truth for analytical studies (e.g., linearity, detection limit) is established through reference materials, spiked samples with known concentrations, and comparison to established methods or gold standard instrumentation.
Ask a specific question about this device
(418 days)
The Emit® II Plus Buprenorphine Assay is a homogeneous enzyme immunoassay with a 5 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of buprenorphine in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.
The semiquantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Mass Spectrometry (LC/MS) or permitting laboratories to establish quality control procedures.
The Emit® II Plus Burrenorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method(s) must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/ MS) or LC/MS are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
For Professional Use.
Caution: Federal (USA) law restricts this device to sale by or on the order of a licensed healthcare professional.
For in vitro diagnostic use.
The Emit® II Plus Buprenorphine Assay is a homogeneous enzyme immunoassay technique used for the analysis of specific compounds in human urine. The assay is based on competition between drug in the specimen and drug labeled with the recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not interfere because the coenzyme NAD functions only with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.
The Emit® II Plus Buprenorphine Assay reagents are provided liquid, ready to use and may be used directly from the refrigerator. The product is sold in three (3) kit sizes: 28 mL, 115 mL, and 1000 mL. Reagents 1 and 2 are provided as a matched set. They should not be interchanged with components of kits with different lot numbers.
Antibody/Substrate Reagent 1: Mouse monoclonal antibodies to buprenorphine (0.53 µg/mL)*.NAD (6.9 mM), G6P (10.9 mM), bovine serum albumin, preservatives, and stabilizers. *The antibody titer and enzyme conjugate activity may vary from lot to lot.
Enzyme Reagent 2: Norbuprenorphine labeled with bacterial rG6PDH (0.50 µg/mL), HEPES buffer, bovine serum albumin, preservatives, and stabilizers, where the antibody titer and enzyme conjugate activity may vary from lot to lot.
The marketing submission for the Siemens Healthineers Emit® II Plus Buprenorphine Assay (K221605) demonstrates that the device meets its acceptance criteria through various performance studies. Below is a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as strict numerical thresholds in the provided document, but rather demonstrated through the qualitative agreement and recovery percentages, as well as precision and specificity profiles. Based on the "Method Comparison" results, the implicit acceptance criterion for qualitative agreement against LC/MS/MS is a high percentage, and for recovery, it's generally close to 100%. For specificity, it's low or negligible cross-reactivity with other substances (ideally less than the cutoff concentration when tested at high concentrations). For precision, it's low coefficient of variation (CV).
| Metric / Acceptance Criteria | Reported Device Performance (Emit® II Plus Buprenphine Assay on DxC 500 AU) |
|---|---|
| Method Comparison (Qualitative Agreement with LC/MS/MS) | 92.5% Agreement |
| Recovery vs Nominal Value | Ranges from 95% to 104% (average ~98.5%) |
| Recovery vs LC/MS/MS | Ranges from 86% to 100% (average ~93.8%) |
| Precision (Repeatability %CV) | 1.1% to 6.6% (for relevant concentration ranges) |
| Precision (Within-Lab Precision %CV) | 2.8% to 6.6% (for relevant concentration ranges) |
| Total Reproducibility Precision %CV | 3.0% to 5.3% |
| Specificity (Cross-reactivity with Buprenorphine Metabolites) | Buprenorphine: 98%, Norbuprenorphine: 106%, Glucuronides: 0.10% |
| Specificity (Cross-reactivity with Structurally Related Compounds) | <0.1% for all tested compounds at 100,000 ng/mL |
| Specificity (Cross-reactivity with Structurally Un-Related Compounds) | All tested compounds at high concentrations (µg/mL) showed 'Neg' response for the -40% Control and 'Pos' for the +40% Control, indicating no interference impacting the cutoff. |
| Interference (Endogenous and Exogenous Substances) | All tested substances showed 'Neg' response for the -40% Control and 'Pos' for the +40% Control, indicating no interference impacting the cutoff. |
| Interference (pH and Specific Gravity) | All tested pH and specific gravity levels showed 'Neg' response for the -40% Control and 'Pos' for the +40% Control, indicating no interference impacting the cutoff. |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison: 120 native urine individual patient samples. The samples were collected from external suppliers, received frozen, and thawed before testing. 21% were within -50% of the cutoff and 19% within +50% of the cutoff. The provenance is not explicitly stated in terms of country of origin but implies human urine samples from external suppliers. The study design is retrospective, using previously collected samples.
- Recovery: 9 urine samples prepared by spiking buprenorphine into negative urine pools.
- Precision (Repeatability and Within-Lab): 11 urine samples prepared by spiking buprenorphine into negative urine pools.
- Precision (Reproducibility): 3 urine samples (Negative Control, Cutoff Calibrator, Positive Control from Emit II Plus Specialty Drug Control line).
- Limit of Detection: 4 blank samples and 4 low-level samples (buprenorphine spiked into drug-free urine). A minimum of 60 replicate measurements per sample set.
- Specificity and Cross-reactivity: The number of unique compounds tested is substantial (listed in Tables 9, 10, and 11), with samples assayed in five replicates for each compound.
- Interference: Various endogenous and exogenous substances, as well as pH and specific gravity variations, were tested at specified concentrations. Samples were assayed in five replicates.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- For the Method Comparison study, the ground truth was established by LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry). This is a highly specific and sensitive analytical method considered the gold standard for drug confirmation testing. No human experts are explicitly described as establishing the ground truth; it's based on the instrumental analytical results.
- For Recovery, the ground truth (nominal value) for spiked samples was determined gravimetrically based on stock concentration and confirmed by LC/MS/MS.
- For Precision, Limit of Detection, Specificity, and Interference, the ground truth is based on the known concentrations of spiked substances or the known characteristics of the samples (e.g., drug-free urine, specific pH levels). Instrumental analysis is used to determine the device's measurement against these known values.
4. Adjudication Method for the Test Set
- For the qualitative analysis in the Method Comparison, the results from the Emit® II Plus Buprenorphine Assay on DxC 500 AU were compared directly against the results from LC/MS/MS. A 2x2 box plot was used to assess qualitative agreement. There is no mention of a human adjudication method (e.g., 2+1 or 3+1), as the comparison is between two analytical methods.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic assay, not an AI-assisted interpretation device that would involve human readers.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Yes, the entire performance evaluation presented is a standalone study of the analytical performance of the Emit® II Plus Buprenorphine Assay on the DxC 500 AU Clinical Chemistry Analyzer. There is no human-in-the-loop component for interpreting the assay's results; the device produces qualitative or semi-quantitative analytical results directly.
7. Type of Ground Truth Used
- Analytical Ground Truth:
- LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry) for method comparison and as a reference for recovery studies.
- Known concentrations of spiked urine samples (gravimetrically determined nominal values) for recovery, precision, limit of detection, specificity, and interference studies. This involves preparing samples with precisely known amounts of the analyte or interfering substances.
- Drug-free urine pools for negative controls and spiking experiments.
8. Sample Size for the Training Set
- The document describes performance evaluation studies for a device, not a machine learning algorithm that requires a "training set." Therefore, there is no mention of a training set or its sample size in the context of this device's regulatory submission.
9. How the Ground Truth for the Training Set Was Established
- As this is not a machine learning-based device, there is no training set and thus no ground truth established for a training set. The studies focus on demonstrating the analytical performance of the immunoassay against recognized analytical standards and spiked samples with known concentrations.
Ask a specific question about this device
Page 1 of 6